Biomedical Equipment 1Introduction

Prof. Moustafa M. [email protected]

[email protected]

Page 2: Biomedical Equipment 1 Introduction

Course Outlines

Calibration and Maintenance of Medical Equipment

Recording devicesStrip chart recorderMagnetic tape recorderDigital recorder

Ultrasound InstrumentX-Ray InstrumentDialysis Machine CT Scan Magnetic Resonance Imaging

Page 3: Biomedical Equipment 1 Introduction

Calibration and Electrical Safety of Medical Equipment

Why Test & CalibrationAs components age and equipment undergoes

changes in temperature or humidity or sustainsmechanical stress, performance graduallydegrades. This is called drift.

When this happens your test results become unreliable and both design and performance quality suffer.

While drift cannot be eliminated, it can be detected and either corrected or compensated for through the process of calibration.

Page 4: Biomedical Equipment 1 Introduction

Definitions

Calibration: process of comparing an unknown against a reference standard within defined limits, accuracies and Uncertainties

Verification: process of comparing an unknown against a reference standard at usually one data point

Page 5: Biomedical Equipment 1 Introduction

Test & Calibration service

Page 6: Biomedical Equipment 1 Introduction

What to TEST for?

• Performance Testing

• Safety Testing

Page 7: Biomedical Equipment 1 Introduction

When to test equipment

prior to being accepted for use During preventative maintenance.After repairs.

Page 8: Biomedical Equipment 1 Introduction

Need for Medical Equipment Testing

• Medical device incidents resulting in patient injury and death• Ensure that the equipment is performing to the expected standards of accuracy, reliability, free of hysteresis and linear (as designed).• Safe and effective devices need to be available forpatient care– Downtime costs money• Regulations, accreditation requirements and standards.

Page 9: Biomedical Equipment 1 Introduction

Why do we do electrical safety?

Ensure patient safetyProtect against macroshockProtect against microshock

Test for electrical internal breakdown / damage to power cord, AC mains feed, etc.

Meet codes & standards  Association of Medical Instrumentation (AAMI), The International Electrotechnical Commission (IEC), National Federation of Paralegal Associations (NFPA),

etc.Protect against legal liability

In case of a patient incident

Page 10: Biomedical Equipment 1 Introduction

Waveform display devices

Page 11: Biomedical Equipment 1 Introduction

Permanent magnet moving coil instrumentsلمغناطيسمستديم المتحرك الملف أجهزة

Chart recorder: (Recording Oscillographs) used in a medical instrumentation to make permanent recording of the

waveform.

most common type: permanent magnet moving coil (PMMC).

The PMMC is very similar to the Galvanometer movement. A writing pen replaced the meter point.

Current flowing in the moving coil creates a magnetic field that interact with the magnetic field of the permanent magnet. This will cause deflection of the pen.

The tip of the pen is positioned over a strip of chart paper that is pulled under the pen tip at a constant speed. In this mechanism:

• Y-axis is the deflection to the pen and • X-axis is the time base established by moving the chart paper at a constant speed.

This will produce a chart recording of the waveshape of the applied waveform

Page 12: Biomedical Equipment 1 Introduction

Curvilinear recording

The pen assembly sweeps an arched manner and so will write in a curvilinear manner.

Page 13: Biomedical Equipment 1 Introduction

Rectilinear recording

The pivoted pen motor assembly is a solution of curvilinear action.

The pen is connected through a mechanical link that translates the curvilinear motion of the PMMC to the rectilinear motion at the pen tip.

Page 14: Biomedical Equipment 1 Introduction

Pseudo rectilinear writing system

In this type of recorder the pen assembly is very long compared to width of the chart paper.

The pen tip, therefore travels in an arc the length of which is very short compared with the radius.

The trace will appear to be nearly linear

Page 15: Biomedical Equipment 1 Introduction

Writing Methods Used in Strip Chart Recorders

Thermal, and Direct contact

Both of these types use a special writing stylus (rather than a pen and a knife edge, also called writing edge)

The mark on the paper by the contact of the stylus on the paper along the knife edge,

The stylus tip travels in a curvilinear path, but the resulting trace is rectilinear because the knife edge is straight.

The stylus can write anywhere along its length, so by keeping the knife edge straight under the paper, we obtain the rectilinear recording that shows waveshape as well as amplitude

Page 16: Biomedical Equipment 1 Introduction

PMMC writing system

Several writing systems are commonly used on PMMC recorders:1- direct contact,2- thermal pen3- ink pen4- ink jet5- optical

Recorder that use any type of pen (ink) or stylus (Direct contact and thermal) have a relatively low-frequency response due to the inertia of the pen or stylus assembly (100-200 Hz) Ink jet and optical type

Lighter weight fixtures and writing stylus system have a frequency response of 1000-3000 Hz

Page 17: Biomedical Equipment 1 Introduction

Direct contact

The direct contact uses a special types of chart paper that is chemically treated to have a carbonized back.

When a pressure applied to that front of the paper a black mark will appear.

Most of these instrument have a frequency response of less than 25 Hz and so are not commonly used in medical instrumentation.

Page 18: Biomedical Equipment 1 Introduction

The thermal recorder

The thermal recorder also uses special paper, but in this case it is waxed or treated with paraffin so that it will turn black when treated

The thermal recorder is the most commonly employed in medical instrumentation, especially in cardiovascular instruments such as the ECG and pressure monitors

Page 19: Biomedical Equipment 1 Introduction

The Stylus

The stylus in a thermal system is little more than a heated resistance wire connected to a low voltage ac or dc power supply.

Early models formed a U-shaped electrical resistance element

Modern models use a wire inside a cylindrical metal stylus

In both cases a low voltage electrical power supply energizes the element, causing the tip to become heated.

The black mark is made at the points where heated stylus touches the paper

Page 20: Biomedical Equipment 1 Introduction
Page 21: Biomedical Equipment 1 Introduction

Ink pen writers

use a hollow pen and an ink supply to write on the chart paper.

In some machines the ink pressured by an atomizer-like hand pump EEG system use this type of writing system.

More automatic machines use a thick, viscous ink in a special cartridge that is placed under pressure by a spring-driven piston

In multichannel instruments, the ink may be distributed to all pens from the same cartridge by connecting the line from the cartridge to the input side of a special ink manifold and additional pressure is applied to the ink by the solenoid-operated manifold bladder

Page 22: Biomedical Equipment 1 Introduction

Ink jet recorder

Higher-frequency response than are the types mentioned previously.

This type is popular on European instrument

Low-viscosity ink is directed to a nozzle mounted on a PMMC galvanometer in place of the pen assembly.

The ink jet produced by the nozzle is directed at the paper.

When the system is properly adjusted, it will produce a recording that is very linearly rectilinear.

Only a small amount of trace fuzziness due to ink splattering is apparent

للتردادات تستجيب التى االنواعالسابقة االنواع من اكثر المرتفعة

اوروبا فى مشهورة فوهة الى يوجه اللزوجة منخفض حبر

القلم مكان الجلفانومتر فى الى يوجه الفاتحة من المتدفق الحبر

الورقة خط ينتج سوف تماما النظام ضبط عند

االستواء تام الحبر بعثرة نتيجة التشوش بعض ينتج

Page 23: Biomedical Equipment 1 Introduction

Optical recorder

There are two types of optical recorder:- PMMC type uses small inertia mirror in place

of the pen assembly or stylus. - The other uses a photographic paper that is

pulled across a cathode ray tube and is called a CRT camera recorder.

Most of these recorders use wide paper.On multichannel optical recorder it is possible

to examine the time relationships between different traces more easily because the traces can be allowed to overlap each other.

Page 24: Biomedical Equipment 1 Introduction

The paper in the optical recorder is often develop by exposure to an ultraviolet lamp as the paper comes out from the recorder.

Unless the paper is either wet developed following the recording session or stored in a light-light box.

In CRT camera, the CRT sweeps only the vertical axis. The time base is provided by pulling the photosensitive

paper in the front of the CRT screen. The frequency response of the CRT camera recorder is

better than that of any of the other types, being limited mostly by the writing speed of the photosensitive paper.

Page 25: Biomedical Equipment 1 Introduction
Page 26: Biomedical Equipment 1 Introduction
Page 27: Biomedical Equipment 1 Introduction
Page 28: Biomedical Equipment 1 Introduction

Servo recorders and recording potentiometers In potentiometeric measurements a three-terminal variable resistor

(potentiometer) is connected to produce an output voltage that is a function of both a reference potential and position of the variable resistor’s wiper arm.

A galvanometer will read zero when the unknown voltage and potentiometer voltage are equal.

Page 29: Biomedical Equipment 1 Introduction
Page 30: Biomedical Equipment 1 Introduction
Page 31: Biomedical Equipment 1 Introduction

Servo recorder

A servo recorder is a self-nulling potentiometer that records the waveshape of the applied signal on graph paper.

The pen attached to a string that is wound around a pair of idler pulleys and drive pulley that is on the shaft of a dc servomotor.

The pen assembly is also linked to a potentiometer (R1) in such a way that the position of the wiper arm on the resistance element is proportional to the pen position.

Page 32: Biomedical Equipment 1 Introduction

The potentiometer element is connected across a reference potential Eref, so potential E is the electric analog of pen position.

The pen position is controlled by the dc servomotor, which in turn driven by the output of the servo amplifier.

The amplifier has a differential inputs: Ein (unknown)

is connected to one input, and the position signal E is connected to the other input

The difference signal (Ein – E) represents the error between the actual pen position and the position the pen should be in for the applied voltage.

Page 33: Biomedical Equipment 1 Introduction

If the error is zero , meaning that the amplifier output is also zero and the pen is correctly positioned, then the motor remain turned off.

If Ein# E then the amplifier creates an output signal that turns on the motor

The motor drive the pen and potentiometer in such a direction as to cancel the error signal

When the input signal and position signal are equal, then the motor turns on the motor.

Page 34: Biomedical Equipment 1 Introduction

A paper drive motor

A paper drive motor forms time base because it pulls the paper underneath the pen at constant rate.

Most high-quality servorecorders use a stepper motor to drive the paper supply.

Such motor will rotate only a few degree every time a pulse applied to its windings

A few models use a continuously running motor that drive the sprocket through a speed-reducing gear box.

Page 35: Biomedical Equipment 1 Introduction

The stepper motor system is capable very good accuracy because a crystal oscillator or the ac power mains are used to drive the pulses used to advance the motor.

Digital integrated circuit frequency dividers (counter circuit) can be used to reduce the clock frequency to the frequency required to drive the motor at desired speed.

The reference potentiometer used to measure the pen position may be any of the following devices:

- Slide wire - Rectilinear, or - Rotary.

The slide-wire system is often used because it can be built with less friction and no mechanical linkage.

Page 36: Biomedical Equipment 1 Introduction

A slide-wire potentiometer

A resistance wire and a shorting wire are stretched taut parallel to each other and the direction of the pen travel.

A shorting bar on the pen assembly serves as a wiper on the resistance element and also connects the shorting wire.

A,B, and C in Figure refer to the potentiometer terminals

the shorting wire serves as terminal B of the potentiometer (the wiper)

Page 37: Biomedical Equipment 1 Introduction
Page 38: Biomedical Equipment 1 Introduction

X-Y recorders

The X-Y recorder uses two servo mechanisms connected to the same writing assembly, but at right angles to each other.

The X-axis servomechanism moves the pen-bar assembly back and forth across the paper in the horizontal plane, while the Y-axis servomechanism moves the pen vertically up and down along the bar.

The paper itself does not move. It is held in place either by clamp or, in high-quility instruments by a vacuum pump that is used to evacuate a hollow chamber below the paper platform.

Holes in the platform create the negative pressure needed to keep the paper in place.

One advantage of the X-Y recorder is that almost any type of paper may be used.

Page 39: Biomedical Equipment 1 Introduction
Page 40: Biomedical Equipment 1 Introduction

Problem of recorder design

1- A dead band signal: largest signal to which the recorder will not respond due to pen assembles mass (inertia). Sufficient preamplification of the signal is required

2- Overshooting and undershooting of the recorded trace.

3- undercritically damped reader will overshoot the correct point and then hunt back and forth across the correct point for a few cycle until it hones in and settles properly

4- over critically damped recorder is sluggish the pen approaches the correct position very slowly .

Page 41: Biomedical Equipment 1 Introduction

Deadband

Page 42: Biomedical Equipment 1 Introduction
Page 43: Biomedical Equipment 1 Introduction

Pen assembles are often damaged if they strike the limits-of-travel steps at a high speed.

A pair of Zener diodes connected back to back across the PMMC coil are sometimes used to accomplish the same job.

The Zenner potential of the diodes is selected so that the bodies break over and conduct current only when a voltage greater than the normal full-scale potential is applied to the input of the amplifier

Page 44: Biomedical Equipment 1 Introduction

Maintenance of the PMMC writing stylus and pens Figure shows how to remove an ink blockage from an ink pen recorder that has

been allowed to stand too long without being used.

As a general rule, such recorders should be run for about five minutes or so once a week when not in regular service.

Fill a 3- t0 10-cc syringe with water (or acetone for certain type of ink) Insert end into the ink inlet

The pen has to removed from the machine for this operation.

Quickly, and with a single sharp motion, drive the plunger “home” so that a high pressure jet of water or acetone is forced into the pen. The ink clot should be forced out the other end.

Page 45: Biomedical Equipment 1 Introduction
Page 46: Biomedical Equipment 1 Introduction

Precautions

Always wear protective-goggles and protective clothing.

Make sure that the pen is aimed downward into a sink

As always when dealing with needles, be careful not to stick yourself

The thick high-viscosity ink strains every thing it touches and a nearly impossible to remove.

Ink pen tips are designed to operate parallel to the paper surface

Page 47: Biomedical Equipment 1 Introduction

If the pen is worn or when a new pen is installed, it is necessary to lap the tip in order to reestablish the parallelism.

The sign that lapping is needed will be either (or both) of the following: A- a blob f ink when the machine first stats a waveform, or B- a too-thick trace

To lap the pen, place a piece of fine Emery cloth (sandpaper) under the tip.

Work the pen tip back and forth 5 to 10 times to sand the tip parallel to the paper

Page 48: Biomedical Equipment 1 Introduction
Page 49: Biomedical Equipment 1 Introduction

The pressure of the stylus or pen is also important.

If the pressure is not correct, then the waveform may be distorted . In medical equipment it is possible to make a normally healthy lead-1

ECG as though the patient has.

Suitable stylus ECG gages can be purchased from ECG machine manufactures.

The stylus pressure adjustment is made using a screw that is usually located on the rear of the stylus or the assembly that holds it in place.

Page 50: Biomedical Equipment 1 Introduction
Page 51: Biomedical Equipment 1 Introduction
Page 52: Biomedical Equipment 1 Introduction

Dot matrix analog recorders

The original dot matrix printers used a 5 X 7 matrix of dots to form alphanumeric data.

The dot matrix machine used a print head to cause the correct dot element

Two different methods:A- Thermal based machine: Some of the earliest machines were thermally based. The dots were thermally connected to heating coils and could be heated

when needed. Special sensitive paper was used to receive the text. This method is no longer widely used.

Page 53: Biomedical Equipment 1 Introduction

B- an array of seven print hammers (pins). The pins would either extend or retract depending whether or not that

particular data was active for the character being printed. An advantage of the pin method is that ordinary paper can be used The pen impacted at inked ribbon to leave the impression Higher resolution models are now valuable with 9, 18, 24 pins

Page 54: Biomedical Equipment 1 Introduction