![Page 1: Bioengineering 208 Magnetic Resonance Imaging Winter 2008 ...fmri.ucsd.edu/ecwong/BE208W08Lec05.pdf · 1 E. Wong, BE208, UCSD Winter 2008 Bioengineering 208 Magnetic Resonance Imaging](https://reader034.fdocuments.us/reader034/viewer/2022052000/6011f25fb6fc40155767f590/html5/thumbnails/1.jpg)
1
E. Wong, BE208, UCSD Winter 2008
Bioengineering 208Magnetic Resonance Imaging
Winter 2008Lecture 5
• MRI Artifacts– Noise spikes– Clipping– Gibbs Ringing– Quadrature ghost– Wraparound– Motion– Chemical Shift
• SNR in MRI– RF Coil– Magnetization– Sampling time
E. Wong, BE208, UCSD Winter 2008
Normal Image
K-space Image space
![Page 2: Bioengineering 208 Magnetic Resonance Imaging Winter 2008 ...fmri.ucsd.edu/ecwong/BE208W08Lec05.pdf · 1 E. Wong, BE208, UCSD Winter 2008 Bioengineering 208 Magnetic Resonance Imaging](https://reader034.fdocuments.us/reader034/viewer/2022052000/6011f25fb6fc40155767f590/html5/thumbnails/2.jpg)
2
E. Wong, BE208, UCSD Winter 2008
Noise Spike
K-space Image space
•Localized in K-space•Extends outside object in image space•Come from arcing, loose connections, ground spikes
E. Wong, BE208, UCSD Winter 2008
More Noise Spikes
K-space Image space
•Multiple spikes create multiple sinusoids and generate‘herringbone’ patterns
![Page 3: Bioengineering 208 Magnetic Resonance Imaging Winter 2008 ...fmri.ucsd.edu/ecwong/BE208W08Lec05.pdf · 1 E. Wong, BE208, UCSD Winter 2008 Bioengineering 208 Magnetic Resonance Imaging](https://reader034.fdocuments.us/reader034/viewer/2022052000/6011f25fb6fc40155767f590/html5/thumbnails/3.jpg)
3
E. Wong, BE208, UCSD Winter 2008
Data clipping
K-space Image space
•Center of K-space over-ranges ADC and clips•Image is (correct image) - (low frequency image)
E. Wong, BE208, UCSD Winter 2008
Data clipping
K-space Image space
•Here is the data that was clipped
![Page 4: Bioengineering 208 Magnetic Resonance Imaging Winter 2008 ...fmri.ucsd.edu/ecwong/BE208W08Lec05.pdf · 1 E. Wong, BE208, UCSD Winter 2008 Bioengineering 208 Magnetic Resonance Imaging](https://reader034.fdocuments.us/reader034/viewer/2022052000/6011f25fb6fc40155767f590/html5/thumbnails/4.jpg)
4
E. Wong, BE208, UCSD Winter 2008
Gibbs Ringing
K-space Image space
•Data is truncated before it decays into the noise•Result is an image convolved with FT of the window ink-space
E. Wong, BE208, UCSD Winter 2008
Quadrature Ghost
K-space Image space
•K-space data is a superposition of good_data andgood_data*•=>Image space is a superposition of good_image(x,y)and good_image(-x,-y)
![Page 5: Bioengineering 208 Magnetic Resonance Imaging Winter 2008 ...fmri.ucsd.edu/ecwong/BE208W08Lec05.pdf · 1 E. Wong, BE208, UCSD Winter 2008 Bioengineering 208 Magnetic Resonance Imaging](https://reader034.fdocuments.us/reader034/viewer/2022052000/6011f25fb6fc40155767f590/html5/thumbnails/5.jpg)
5
E. Wong, BE208, UCSD Winter 2008
Wraparound
http://www.fmrib.ox.ac.uk/~peterj/lectures/kspace/img034.GIF
Phase Encode•K-space sampled discretely•Susceptible to aliasing
Frequency Encode•K-space traveled continuously,but data is digitized•Wraparound can be preventedby either analog or digitalbandpass filter
‘No Phase Wrap’
E. Wong, BE208, UCSD Winter 2008
Motion Artifact
http://wwwrad.pulmonary.ubc.ca
•Motion between TR periods generatesinconsistency between lines of K-space•Ghosts propagate in the phase encodedirection•Period motion generates structuredghosts (analogous to EPI NyquistGhosts)
Magnitude Phase
FT (columns)FT (rows)
Normal Peristalsis Paralyzed Bowel
![Page 6: Bioengineering 208 Magnetic Resonance Imaging Winter 2008 ...fmri.ucsd.edu/ecwong/BE208W08Lec05.pdf · 1 E. Wong, BE208, UCSD Winter 2008 Bioengineering 208 Magnetic Resonance Imaging](https://reader034.fdocuments.us/reader034/viewer/2022052000/6011f25fb6fc40155767f590/html5/thumbnails/6.jpg)
6
E. Wong, BE208, UCSD Winter 2008
Chemical Shift•Magnetic field of electron clouds shields nucleusfrom external magnetic field•=>Actual magnetic field experienced by nucleus issmaller than applied field•Differences in local field are called chemical shift,and are measured in PPM•Water and fat differ in chemical shift by 3.5 PPM= 440Hz at 3T•Chemical shift causes phase twist across readout•Fourier shift theorem tells you how far things areshifted
http://chickscope.beckman.uiuc.edu/roosts/carl/artifact/cs_a001.gif
http://www.mr-tip.com/serv1.php?type=art&sub=Chemical%20Shift%20Artifact
E. Wong, BE208, UCSD Winter 2008
CONVENTIONAL IMAGING
0ms
8ms1 / 0ms = ∞ Hz / pixel
1 / 8ms = 125 Hz / pixel
![Page 7: Bioengineering 208 Magnetic Resonance Imaging Winter 2008 ...fmri.ucsd.edu/ecwong/BE208W08Lec05.pdf · 1 E. Wong, BE208, UCSD Winter 2008 Bioengineering 208 Magnetic Resonance Imaging](https://reader034.fdocuments.us/reader034/viewer/2022052000/6011f25fb6fc40155767f590/html5/thumbnails/7.jpg)
7
E. Wong, BE208, UCSD Winter 2008
Off Resonance Behavior : EPI
0.5ms
40ms
1 / 40ms = 25 Hz / pixel
1 / 0.5ms = 2000 Hz / pixel
E. Wong, BE208, UCSD Winter 2008
Signal to Noise Ratio in MRI
!
SNR" (coil_ factor) # (magnetization _ factor) # (sampling_ factor)
•Proportional to local B1•Sample noise increaseswith coil size•Depends on coil geometry,coil quality (internal coillosses)
•Proportional to Mxy at timeof readout•Depends on B0, pulsesequence, TR, TE, PD, T1,T2, T2*, voxel volume
•Proportional tosqrt(total readout time)
!
SNR" (voxel_volume) total_ readout _ time