BIOE 276 Final Report

14
Microfluidic Platform to Detect Waterborne Pathogens Using DEP and Magnetophoresis Avinash Kumar, Nikhil Makaram, Zina Kurian June 9th, 2015 Abstract This microfluidic device will assist in water pathogen detection. The POC device will be low cost, small, portable, disposable, and require minimal preprocessing to maximize usage in developing countries where water sources are more likely to have contamination. This device should allow multiplexed separation and detection of multiple bacterial pathogens, such as E. Coli or Salmonella. Fabrication will be through standard direct method using a PDMS substrate. The design will utilize a serpentine mixing channel, DEP focusing and separating electrodes, and magnetic complexes to sort cells into channels based on characteristics. Then, using fluorescence microscopy, we will detect expression of different bacterial pathogens from GFP target tags. Through these methods, waterborne pathogens will be detected on site. This prevents future use of contaminated supply by fulfilling two goals of fabricating a novel microfluidic lab- on-a-chip device and detecting a minimum of 70% of pathogens within one small sample. 1

Transcript of BIOE 276 Final Report

Page 1: BIOE 276 Final Report

Microfluidic Platform to Detect Waterborne Pathogens Using DEP and Magnetophoresis

Avinash Kumar, Nikhil Makaram, Zina KurianJune 9th, 2015

Abstract

This microfluidic device will assist in water pathogen detection. The POC device will be low cost, small, portable, disposable, and require minimal preprocessing to maximize usage in developing countries where water sources are more likely to have contamination. This device should allow multiplexed separation and detection of multiple bacterial pathogens, such as E. Coli or Salmonella. Fabrication will be through standard direct method using a PDMS substrate. The design will utilize a serpentine mixing channel, DEP focusing and separating electrodes, and magnetic complexes to sort cells into channels based on characteristics. Then, using fluorescence microscopy, we will detect expression of different bacterial pathogens from GFP target tags. Through these methods, waterborne pathogens will be detected on site. This prevents future use of contaminated supply by fulfilling two goals of fabricating a novel microfluidic lab-on-a-chip device and detecting a minimum of 70% of pathogens within one small sample.

1

Page 2: BIOE 276 Final Report

I. Problem Statement

Water is a necessity for all living things. The World Health Organization (WHO) estimates that about 3.4 million people die from water related diseases. The worst part is the majority of deaths occur in children. (Berman, 2009) This is seen as a high cause of death in developing countries that don’t have the proper infrastructure to handle water and protect it from raw sewage and feces getting into the water supply and contaminating it. This can allow for bacterial growth in the water supply that is then ingested by people who can catch diseases such as cholera, typhoid fever, hepatitis A, gastrointestinal illness, respiratory illness, skin infections, diarrhea, and death. (CDC, 2013 and Berman, 2009) The different types of bacteria that can cause this are E. Coli and Salmonella. These bacteria can not only be transmitted through drinking water but also through eating food prepared by infected individuals, or touching a sick person and then touching your mouth. (Berman, 2009) Building infrastructure such as pipelines can help provide clean water for places that need it. Unfortunately, there is no real way of knowing that even the water that is being transported through these pipelines is in fact bacteria-free.

There are different methods that can be implemented to determine if bacteria is present in a system. In labs, it is easy to undergo immunostaining techniques (or other visual techniques) to see if there is E. Coli or any other type of bacteria in a solution. In determining if a pathogen is there in a system, there must be a proven technique for detection and separation of said pathogen (or multiple pathogens) in a system. Our focus will be in determining a method of fabrication of a microfluidics device that allows for detection of at least 70% of waterborne pathogens as well as implementation of a pathogen separation technique. Detection can potentially be performed through immunostaining light scattering, fluorescence microscopy, or qPCR. The separation technique will be based on fabricated microfluidics system with dielectrophoresis and magnetic labeling. The fabrication method will be soft lithography with multiple microchannels to allow for separation of the pathogen from water. The separation technique will combine fluorescent activated cell sorting with two independent force fields. This device attempts to provide a point-of-care solution to the problem of contaminated water toxins.

II. Goals

The goals of this project are twofold. First, we aim to fabricate a device that will be able to separate and detect specific pathogens in a water sample. This method should be rapid and low cost. We want a device to be for point-of-care use. Specifically, cell sorting needs to have high purity, high specificity, high target recovery, high throughput, and with no crossover contamination. We want to be able to separate multiple different pathogens in a water sample. Second, the purpose of this device is to create a biologically relevant model. This device should enable diagnostics for users in remote areas for real-time monitoring. The use of this device should provide pathogen-specific testing and enable a whole community to use this reliable method. If water samples can be tested for deadly diseases, communities will know which water sources are usable. This device should follow the WHO guidelines established for benefits of water testing devices. This includes advantages such as portability, minimal processing using separate lab equipment, independence from power supplies, low cost, low maintenance, rapid testing, sensitivity, purity, and safe operation by everyone in order to meet quality standards.

2

Page 3: BIOE 276 Final Report

III. Background

This high-throughput multiplexing device includes miniaturization of elements required for mixing and synthesis of flow particles. This includes pumps, micro-valves, and mixers in order to separate into microchannels. There are two conditions that will be applied on this device in order to separate bacterial pathogen particles from other non-target water particles: dielectrophoresis and magnetophoresis. Nontarget cells will be separated in spatially different outlets from target cell types. This concept is known as integrated Dielectrophoretic-Magnetic Activated Cell Sorter (iDMACS).

Dielectrophoresis In the presence of electric fields, particles will have dielectrophoretic activity. Polarized

particles move toward a region of the field. This activity is influenced by the surrounding medium. This force comes from the electrode surface. The magnitude of the force depends on the cell’s electrical properties and size1. Low medium conductivity is optimal for these conditions. Different configurations of electrodes affect the control of targets. The height of electrodes in three dimensions, as well as two dimensional placement and angle, can control the force. It has been previously demonstrated that dielectrophoresis activated cell sorters will target cells that are immunochemically labeled with synthetic tags because of the properties of DEP2. Typically, this concept is extended for use with double input and multiple outputs for continuous flow of multitarget operation.

For this type of separation, each target cell type is labeled with a unique DEP tag via receptor-ligand binding in the hybridization chamber. Each tag is specifically chosen for the potential pathogen being assessed. Inside the device, the apparent forces include hydrodynamic force from the sample flow and the dielectrophoretic force from the nonuniform electric field acting perpendicular to the electrodes. Target cells are deflected when they pass through the electric field and the force from DEP is stronger than the hydrodynamic force. By implementing two sets of electrodes at decreasing angles, cells with tag A will be dielectrophoretically separated and guided to channel A. Cells with tag B for a different pathogen will continue along the channel until being deflected by the second set of electrodes and passing into channel B. The liquid motion is controlled by gravity based pressure driven flow, instead of operating by tubing or valves. Having two electrode sets allow two angles to be produced, directing target cells to the

correct outlet. Nontarget cells will also continue to move along the device because they are not tagged and unaffected.

Figure 1: Two electrode sets at different 𝜽1 and 𝜽2 allow separation at two outlets2.

This strategy is used often in microfluidics. The DEP force generated by electrodes is governed by many equations as explained by Krali3. The fluid drag force is described by Stokes

1 Foudeh, Amir M., et al.2 Kim, Unyoung, et al.3 Kralj, J. G., et al.

3

Page 4: BIOE 276 Final Report

law, FHD = 6π𝜇𝑣α. Cells experiencing these forces are affected by the sum total of these forces. This enables accurate size-based separation since the radius is in both of these equations. Further calculations can predict the velocity of particles travelling in the channels, and thus the concentration profile to show how particles with tag A will have a larger complex. This allows target A particles to deflect efficiently into the appropriate channel because of a variation in force magnitude. The electrodes provide adequate electric fields, while the consistent microchannel height ensures cell velocities (and FHD) can be controlled. Also, there is a buffer stream and sample stream at different inlets, so laminar flow is necessary to control the segregation of the target streams. In a single pass through of the device, the particles should form complexes, segregate based on active forces, and collect at magnetic sites. Since we are using bacterial cells, the DEP tag will be similarly sized with the target cell.

Figure 2: Diagram of separating electrodes. (a) shows continuous cell separation. (b) is fluorescence microscopy image of a separation channel coming in while (c) is the separated target cells in the correct channels. 4

Magnetophoresis and Nanoparticle Complexes All materials have their own magnetic field in the

form of each material’s particle’s electron’s spin. When an electrical field is created over an area (by an electrode), the electrons (especially the electrons on the surface of a material) have small forces enacted on them that can cause translation of a material over an area5. DEP commands a linear magnetic response to materials which means all materials will translate away from a DEP electrode equally until they are neither exhibiting a positive or negative DEP6. If there was a need to enact a stronger force on a particle, there would be a need to use a different method as DEP enacts a similar force on all materials.

Magnetophoresis is another method similar to DEP that causes an magnetic gradient across an area. While most materials do have a magnetic permeability, only

some materials can be actuated by the magnetic field when it is applied on an area with different materials and particles. This process allows for actuation of particles that are strongly affected by a magnetic field to be moved with a non-linear force response to magnetic permeability of a particle7. The magnetic permeability (or the dipole moment) is expressed in the equation, m = (4/3)πα3xHϕ where B = μoH8. When a magnetic field is applied on a group of particles, the magnetic permeability of the medium, particle, and magnetic field combine to express the magnetophoretic force on a particle; This force dictates how much pull a magnet might have on a particle that has a magnetic permeability. This is expressed as Fmag=(μmag,mmeff×grad(f))H9. The magnetic force depends on all these different forces but due to their non-linear relationship, a

4 Park, S., et al.5 Kirby, Brian J., et al.6 Pethig, Ronald, et al.7 Kirby, Brian J., et al.8 Ramadan, Qasem, et al.9 Kirby, Brian J., et al

4

Page 5: BIOE 276 Final Report

small change in any one of the variables can have a huge shift in the magnetic force. However this force is usually very small when being impacted on particles in a medium. Due to the fact that most materials that exhibit a magnetic permeability have values that are very small, using particles that have higher values is how this process is exemplified in microfluidic systems.

Figure 3: Magnetic bead attaches to RNA and allows for a stronger magnetic pull to magnet.10

Using magnetic particles with a high magnetic permeability, this allows for a strong magnetic pull and therefore can cause an object to be actuated and possibly stopped in a high flow solution. Using this principle, magnetic beads can be conjugated with antibodies for different pathogens so when they come near each other or into contact, they will bind with each other

and will become a complex conjugate. When the complex conjugate comes near a magnetic field, the magnetic bead creates a strong dipole moment which causes the attraction between the magnet and the magnetic bead. That draws the complex conjugate to the magnet holding it in place. This method would allow for anything attached to the magnetic bead conjugate to be held in place in a high flow system.

Fluorescence ImagingA method for detection of cells in a system is by fluorescent imaging. This method

involves illuminating an area that contains cells that are either tagged or leak fluorophores through a microscope. The light source from the microscope is a higher intensity which causes the fluorescent species to emit a longer wavelength light of lower energy which is viewed in the microscope. Depending on the wavelength of the background and the inspected fluorophores, that would dictate what color is viewed in the microscope. When viewed, the image is taken through 2 filters which prevent the combination of excited and fluorescing rays. The first filter only lets through light that fits the wavelength of the tagged species that is being viewed. The second filter then lets in the light that the entire sample produces and it separates that light from the emitted light from the tagged species11. That is what is seen in the final image. This type of imaging technique is used today due to a number of favorable properties including: high contrast (little to no noise seen after signal processing), high sensitivity, cheap, and easy to use12.

IV. Methods and Materials

10 Zaytseva, Natalya V., et al11 Rice, George., et al12 Alander, Jarmo T., et al

5

Page 6: BIOE 276 Final Report

Figure 4: Schematic of Proposed Microfluidic System Design. Dual-inlet channel takes in both the solution of target pathogens & associated tag + micro-bead complexes, and the subsequently run surfactant solution. Particles are then separated through DEP into corresponding outlet channels where they are imaged after magnetic containment.

The design of the microfluidic device is made up of multiple inlet and outlet ports and sequential regions of first DEP separating action and subsequent magnetic separation action (Figure 4). The first inlet port is designed to bring in a pre-prepared solution composed of the target RNA of the pathogens of interest along with a mixture of reporter and capture probe-complexes corresponding to each pathogen. These complexes contain magnetic beads attached to the capture probes and different liposomes attached to each of the reporter probes for each pathogen type. Within the liposomes of these reporter probes are fluorophore molecules; each liposome type contains one of three differently colored fluorophores to correspond with each pathogen type in this tri-pathogen detecting microfluidic system. The mechanism of action of these reporter and capture probes can be seen in the example of Figure 3 mentioned earlier. As the solution runs through the serpentine portion of the first inlet channel, the probe-complexes hybridize with their appropriate target pathogens. Then these hybridized complexes pass through the DEP region and are separated into each of their corresponding outlet channels based on the respective electrical properties of each complex. Afterward, these complexes are attracted to magnets at the end of each of their outlet channels and are collected there. At this point, a surfactant solution is sent through the second inlet to wash over all the liposomes in each of the complexes of the outlet channels, causing them to erupt to release the differently colored fluorophores within so that each pathogen can be fluorescently imaged for detection and quantification.

The creation of the microfluidics system will be an inexpensive, easy to build system. The main portions of the system are the different buffer solutions, the surfactant, the liposome tags, the magnetic beads, and fabricating the microfluidic system. Each provides a function in separation and detection of the pathogens in the system. The fabrication method involved for the device will be soft lithography. This is a popular method due to the inexpensive nature of it as well as the ability to work in combination with hard material fabrication. The combination of photolithography and dry etch lithography will be used to obtain a master pattern which will then be to create a mold made up of PDMS (polydimethylsiloxane) and a silicon wafer. The structure (once constructed) will be cured and the PDMS will be peeled off to produce a replica of the system. Any extra holes needed to be made (for injecting buffer solutions) can be fabricated too13. The device being made will have two inlets; one for a combination of magnetic beads, water and the waterborne pathogens mixed, and the liposome tags while the other inlet will have a surfactant that cause the liposome tags to burst releasing fluorophores. A special part of the device is the serpentine region where hybridization of the magnetic beads, liposome tags, and pathogen occurs. The region is serpentine to allow for time for the binding to occur.

13 Zaytseva, Natalya V., et al

6

Page 7: BIOE 276 Final Report

There will be 3 outlets with magnets placed in each outlet channel. The magnets will have a strong enough pull that it can stop the magnetic bead complexes that will be travelling through the channel. There will be 3 types of magnetic beads of different sizes (this way the DEP separation will separate the complexes based on size into the right outlets). The magnetic beads will be streptavidin coated superparamagnetic beads14 created as determined by a third party. The magnetic beads will be bound with the DNA capture probes during this process. Liposome tags will be made using a modified version of a reverse phase evaporation technique. They will also be given a DNA capture probe that will allow them to bind to the target pathogen when going through the hybridization process in the serpentine region. The surfactant solution will be a running buffer OG solution. An imaging software (by Roper Scientific Inc) and Pro Express software combined with Photometerics CCD camera is how detection of the different types of pathogens attached to liposome tags15.

V. Potential Problems

There are limitations with this solution. While it is ideal to create a low cost device, there may be elements that are impossible to miniaturize on the chip. There may be post or pre processing for each sample required. We aim to minimize this condition of necessary lab operations. There may be concerns of unnecessary capture with external magnetic pads. We chose to use magnets to hold the target cells in place so that the complexes can be easily collected or imaged. These external magnets do not need to be manually moved. However, this can also be done by letting the target pathogens accumulate at the outlet. This may allow easier withdrawal of cells using a flow-through solution after processing.

Additionally, this design should give the exact concentration of cells based on GFP expression. Cell viability is usually measured via flow cytometry. An alternative would be to just give a YES/NO signal if a pathogen is present in the sample. This would most likely be faster, and fulfill the criteria for detection. However, our solution chose to implement various concentrations as a user output because small concentrations of pathogens such as nitrate are sometimes allowed and not toxic at that level. Length of the device is another critical factor. A longer channel length ensures purity of separation. However, this increases size and cost of the device. Other parameters include electrode spacing, dielectric constant, and viscosity of solution. Electrode material is also up to debate. Our solution uses platinum plated electrode arrays 3. It is also practical to use a low operating voltage at the electrodes because the viability of the cells is sensitive and it is limited by electrolysis. This is a minor pitfall because throughput would be increased with higher voltage, but cells may die. Quantum Dots are another strategy for detection, but fluorescence imaging has been shown to be more reliable for all cell types16. The most obvious problem that we hope to avoid is the reduction of purity in separated outlets. This likelihood is reduced by capturing pathogens at magnetic collection points, and overall by using two force fields to double the separation idea. These potential problems are minimized with our current strategy. This device has the best characteristics for optimal separation and detection for the requirements desired. It will serve its purpose efficiently.References

14 Zaytseva, Natalya V., et al15 Zaytseva, Natalya V., et al16 Deerinck, Thomas J.

7

Page 8: BIOE 276 Final Report

Centers for Disease Control and Prevention. Centers for Disease Control and Prevention, 22 Oct. 2013. Web. 08 June 2015.

"PACE REPORT." Microfluidic Mutli-layer Fabrication by Soft Lithography. PACE Report, n.d. Web. 08 June 2015.

Agrawal, Shailaja, Amit Morarka, Dhananjay Bodas, and K. M. Paknikar. "Multiplexed Detection of Waterborne Pathogens in Circular Microfluidics." Applied Biochemistry and Biotechnology 167.6 (2012): 1668-677. Appl Biochem Biotechnol. Web.Berman, E. S. F., M. Gupta, C. Gabrielli, T. Garland, and J. J. McDonnell (2009), High-frequency field-deployable isotope analyzer for hydrological applications, Water Resour. Res., 45, W10201, doi:10.1029/2009WR008265.Berman, Jessica. "WHO: Waterborne Disease Is World's Leading Killer." VOA. Voice of America, 29 Oct. 2009. Web. 09 June 2015. Caputo, Domenico, Giampiero De Cesare, Luisa Stella Dolci, Mara Mirasoli, Augusto Nascetti, Aldo Roda, and Riccardo Scipinotti. "Microfluidic Chip With Integrated A-Si:H Photodiodes for Chemiluminescence-Based Bioassays." IEEE Sensors Journal 13.7 (2013): 2595-602. Web.

Deerinck, Thomas J. "The application of fluorescent quantum dots to confocal, multiphoton, and electron microscopic imaging." Toxicologic pathology 36.1 (2008): 112-116.

Foudeh, Amir M., et al. "Microfluidic designs and techniques using lab-on-a-chip devices for pathogen detection for point-of-care diagnostics." Lab on a Chip 12.18 (2012): 3249-3266.

Jarmo T. Alander, Ilkka Kaartinen, Aki Laakso, et al., “A Review of Indocyanine Green Fluorescent Imaging in Surgery,” International Journal of Biomedical Imaging, vol. 2012, Article ID 940585, 26 pages, 2012. doi:10.1155/2012/940585

Kralj, Jason G., et al. "Continuous dielectrophoretic size-based particle sorting." Analytical chemistry 78.14 (2006): 5019-5025.

Kim, Unyoung, et al. "Multitarget dielectrophoresis activated cell sorter." Analytical chemistry 80.22 (2008): 8656-8661.Kirby, Brian J. "Kirby Research Group at Cornell: Microfluidics and Nanofluidics :." Kirby Research Group at Cornell: Microfluidics and Nanofluidics :. Kirby Research Group at Cornell, n.d. Web. 08 June 2015.

Mairhofer, Jürgen, Kriemhilt Roppert, and Peter Ertl. "Microfluidic systems for pathogen sensing: a review." Sensors 9.6 (2009): 4804-4823.

Park, Seungkyung, et al. "Advances in microfluidic PCR for point-of-care infectious disease diagnostics." Biotechnology advances 29.6 (2011): 830-839.

Park, Seungkyung, et al. "Continuous dielectrophoretic bacterial separation and concentration from physiological media of high conductivity." Lab on a Chip 11.17 (2011): 2893-2900.

8

Page 9: BIOE 276 Final Report

Pethig, Ronald et al. “Dielectrophoresis: A Review of Applications for Stem Cell Research.” Journal of Biomedicine and Biotechnology 2010 (2010): 182581. PMC. Web. 8 June 2015.

Qasem Ramadan, Lay Christophe, William Teo, Li ShuJun, Feng Han Hua, Flow-through immunomagnetic separation system for waterborne pathogen isolation and detection: Application to Giardia and Cryptosporidium cell isolation, Analytica Chimica Acta, Volume 673, Issue 1, 12 July 2010, Pages 101-108, ISSN 0003-2670, http://dx.doi.org/10.1016/j.aca.2010.05.025.

Rice, George. "Fluorescent Microscopy." Microscopy. Science Education Research Center @ Carleton College, n.d. Web. 09 June 2015.

Vahey, M. D., and J. Voldman. 'High-Throughput Cell And Particle Characterization Using Isodielectric Separation'. Anal. Chem. 81.7 (2009): 2446-2455. Web.

Zaytseva, Natalya V., Vasiliy N. Goral, Richard A. Montagna, and Antje J. Baeumner. "Development of a Microfluidic Biosensor Module for Pathogen Detection." Lab on a Chip 5.8 (2005): 805. Web.

9