Description ACTIFLOW®, is a new sampling device with a

magnetically activated piston to take a sample of a liquid content from a reactor or storage tank.

This device is specially designed for sampling under difficult conditions like extreme vacuum 10-7 [mmHg]. Also this system guarantees the safety of the operating staff.

Technical caracteristics

• Nominal pressure - 1 à 16 [bar] • Nominal temperature - 20 à 200 [°C] • Standard outlet diameter 10 [mm] • Raised volume / lift 100 [ml] • Volume delivery 400 [ml/min] • Viscosity 2000 [mPa/sec]

Advantages

• The liquid is never sucked but pushed up. • No aspiration problems. • Avoids the use of a vacuum pump. • Guaranteed sampling even under extreme

vacuum conditions 10-7 [mmHg]. • Magnetic drive in a two chamber system which

avoids the physical contact as well as dynamic seals.

• Small piston lift. • Compact design. • Simple operation. • Self-cleansing. • Closed sampling system. • Excellent draining.

What's new ?

Its principle of function is to push the liquid by a piston and chamber to the outside of the reactor. As a result of gravity, the liquid flows via a loop back into the reactor. In accordance with the requirements a sampler and / or an analytical apparatus for the measurement of certain values for example, pH, viscosity, Redox, Conductivity, etc. can be installed in this circuit.

Materials

• Standard stainless steel 1.4435 or 316 L. • All parts in contact with the product are

supplied in accordance with the customers choice.

• PTFE; (Tantalum); Titanium; Hastelloy®; etc.

Page 3: BIAR SA - mitec-eng.it · I:\Prospectus\Anglais2.doc Description ACTIFLOW®, is a new sampling device with a magnetically activated piston to take a sample of a liquid content from

I:\Prospectus\Anglais2.doc

Principle of function

a) Filling the cylinder (1)

b) Setting in motion the pneu-magnetic cylinder (2), connected to the piston (4) by a rigid wire (3)

c) Piston (4) moving in the cylinder (1)

d) Liquid rising in the pipe (5)

e) Liquid flowing into the sampling valve (6)

f) Taking the sample into the bottle(7)

g) Surplus liquid flowing back into the vessel (8)

h) Resumption of the cycle

Page 4: BIAR SA - mitec-eng.it · I:\Prospectus\Anglais2.doc Description ACTIFLOW®, is a new sampling device with a magnetically activated piston to take a sample of a liquid content from

I:\Prospectus\Anglais2.doc

Rinsing circuit

BIAR S.A. 5, rue des Epineys BIAR INC. 2506 S. Philippe Ave. CH-1948 Lourtier - Switzerland Gonzales LA 70737 USA Tél. +41 27 779 11 11 Phone (504) 647-4300 Fax +41 27 779 11 12 Fax (504) 647-4309

E-mail: [email protected] Web: http://www.biar.com

Page 5: BIAR SA - mitec-eng.it · I:\Prospectus\Anglais2.doc Description ACTIFLOW®, is a new sampling device with a magnetically activated piston to take a sample of a liquid content from

Contact details

*Company: Department: *Contact:

Street: Zip Code / City: State:

*Phone: Extension: Date:

Fax: *Email: BIAR:

ACTIFLOW®

Maximum available height h1: mm

Diving depth h2: mm

Connection flange B1 (Min. DN100/ANSI 4"): Unit DN/Size PN/Class

DN

AINSI

Material Stainless steel 1.4435/1.4571 with PTFE seals

Hastelloy® 2.4602/HC22 with PTFE seals

Working pressure Min. Available Max.

2 bar

6 bar

Volume of desired sample: ml

Sampling frequency: / day

Reactor / tank

Connection flange B2: Unit DN/Size PN/Class

DN

AINSI

Internal diameter d1: mm

Manufacturer:

Type:

Standard:

Volume: m3

Material:

Allowable temperature:

Min Max °C

°F

Allowable pressure:

Min Max bar

PSI

Agitator system:

Height: mm Diameter: mm

Page 6: BIAR SA - mitec-eng.it · I:\Prospectus\Anglais2.doc Description ACTIFLOW®, is a new sampling device with a magnetically activated piston to take a sample of a liquid content from

Characteristics of the product, sampling, analysis

1. Chemical composition: Formula: Name:

2. Temperature of the liquid: Min: °C °F Max: °C °F

3. Solidification temperature of the liquid: °C °F

4. Working pressure: bar lbs

5. Viscosity and vapor pressure of the product in the working area:

Temperature °C °F 20°C / 68°F

Viscosity cP SSU

Vapor pressure mbar psi

6. Toxicity of the product: low high

7. Can the liquid come into contact with the atmosphere? yes no

8. Does the liquid sampled solidify in the receptacle? yes no

Necessary to pre-heat receptacle before filling? yes no

Necessary to re-heat receptacle to empty? yes no

9. How is removed sample used? emptied into another receptacle injected using a needle

10. What happens to the remainder of the sample?

Additional note

BIAR Switzerland

BIAR SA 12, Rue des Epineys CH-1948 Lourtier

Tel. +4127 779 11 11 Fax +4127 779 11 12 [email protected]

BIAR USA

BIAR, Inc. 2506 S. Philippe Ave. Gonzales, LA 70737

Tel: 225-647-4300 Fax: 225-647-4309 [email protected]

Copyright © BIAR Sampling Systems. All rights reserved.