Bi-Flex IOL RANGE - Domedics AG · Head Office Herceghalmi Road 2072 Zsámbék Hungary...
Transcript of Bi-Flex IOL RANGE - Domedics AG · Head Office Herceghalmi Road 2072 Zsámbék Hungary...
CONTENTS
HYDROPHILIC IOLs
Bi-Flex PIL-MA hydrophilic preloaded . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
677P / PY · 677PT / PTY · 677PMY · 677PMTY
Bi-Flex HL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
677AB / ABY
Bi-Flex T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
677TA / TB · 677TAY / TBY
Bi-Flex M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
677MY
Bi-Flex MT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
677MTY
HYDROPHOBIC IOLs
Bi-Flex POB-MA hydrophobic preloaded . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
877PA / PAY
Bi-Flex HB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
877FAB / FABY
CONSTANTS FOR MEDICONTUR IOLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
INJECTION3rd CLICK READY
2nd CLICK CLOSING1st CLICK LOADING
Head Office
Herceghalmi Road
2072 Zsámbék
Hungary
International Office
Chemin des Aulx 18
1228 Plan-les-Ouates
Switzerland / Geneva
Medicontur
Medical Engineering Ltd
www.medicontur.com
LA-IMCPILMAENG 09 201608
Push the red
stopper forward
to the end
position and
fill visco
Push the plunger
slowly to the
end position'
'
$Insert the injector to the
wet container
Fold up the cartridge
nozzle
Lens types Single-piece aspheric IOLs, clear and yellow Æ Monofocal Æ Toric Æ Diffractive Progressive Æ Toric Diffractive Progressive (Patent pending)
Diameters 6.0 mm (optic) 13.0 mm (overall)
Angulation 0° – asymmetrical design
Material Copolymer of hydrophobic and hydrophilic monomers 25% water content with UV Absorber / + blue light filter (390–470 nm)
Incision size 2.2 mm
PCO protection 360° Special Square Edge
A-Constant 118.9 (SRK/T)
677P/677PYPowers available -10.0 D " -1.0 D · (increment: 1.0 D)
0.0 D " +30.0 D · (increment: 0.5 D) +31.0 D " +35.0 D · (increment: 1.0 D)
677PT/677PTYPowers available -10.0 D " -1.0 D · (increment: 1.0 D) 0.0 D " +30.0 D · (increment: 0.5 D) +31.0 D " +35.0 D · (increment: 1.0 D)Cylinders available 1.00 D, 1.50 D, 2.25 D, 3.00 D, 3.75 D, 4.50 D
677PMYPowers available 0.0 D " +35.0 D · (increment: 0.5 D)
Addition +3.5 D
677PMTYPowers available +10.0 D " +30.0 D · (increment: 0.5 D)Cylinders available 1.00 D, 1.50 D, 2.00 D, 2.50 D, 3.00 D,
3.50 D, 4.00 D, 4.50 DAddition +3.5 D
M
Single use injection system forBi-Flex preloaded monofocal, toric and diffractiveprogressive hydrophilic IOLs
1
Head Office
Herceghalmi Road
2072 Zsámbék
Hungary
International Office
Chemin des Aulx 18
1228 Plan-les-Ouates
Geneva / Switzerland
Medicontur
Medical Engineering Ltd
www.medicontur.com
13m
m
0.4 mm
6 mm
677AB · 677ABYHydrophilic IOL
Technical specification
Type Single-piece Aspheric Hydrophilic Intraocular Lens
for implantation into the capsular bag
Optic Aspheric
PCO protection 360° Special Square Edge (patented)
Estimated incision size from 1.8 to 2.2 mm
Powers available* -10.0 D " +9.0 D · (increment: 1.0 D)
+10.0 D " +30.0 D · (increment: 0.5 D)
+31.0 D " +45.0 D · (increment: 1.0 D)
A-Constant** 118.1 (SRK/T)
Sterilization Steam
Material
• 677AB Copolymer of Hydrophilic and Hydrophobic Acrylic
25% water content with UV Absorber
• 677ABY + blue light filter
Refractive index 1.46
ABBE Nr. 58
Geometry
Optic design Biconvex
Overall lenght 13.0 mm
Optic diameter 6.0 mm
Haptic thickness 0.40 mm
Haptic angulation 0° – asymmetrical design with posterior vaulting
Storage
Temperature +15 – +35°C
Humidity 15% – 50%
Shelf life 5 years (from sterilization)
EA-IMCBHLEN 06 201607
* Other powers upon request
** It is recommended that surgeons personalize the constants they use. Please find more information about IOL constants on ULIB.
(http://www.ocusoft.de/ulib/c1.htm)
2
Head Office
Herceghalmi Road
2072 Zsámbék
Hungary
Export Office
Chemin des Aulx 18
1228 Plan-les-Ouates
Geneva /Switzerland
Medicontur
Medical Engineering Ltd
www.medicontur.com
13m
m
0.4 mm
6 mm
677TA / TB · 677TAY / TBYHydrophilic Toric IOL
Technical specification
Type Single-piece Aspheric Hydrophilic Toric
Intraocular Lens for implantation into the capsular bag
Optic
• 677TA / TAY Toric – Aspheric
• 677TB / TBY Bitoric – Aspheric
PCO protection 360° Special Square Edge (patented)
Estimated incision size from 2.2 to 2.4 mm
Powers available * -10.0 D " -1.0 D · (increment: 1.0 D)
0.0 D " +30.0 D · (increment: 0.5 D)
+31.0 D " +35.0 D · (increment: 1.0 D)
Cylinders available
• 677TA / TAY 1.0; 1.5 D " 9.0 D · (increment: 0.75 D); 10.0 D
• 677TB / TBY +11.0 D " +24.0 D · (increment: 1.0 D)
A-constant** 118.9 (SRK/T)
Sterilization Steam
Material
• 677TA / TB Copolymer of Hydrophilic and Hydrophobic Acrylic,
25 % water content with UV absorber
• 677TAY / TBY + blue light filter
Refractive index 1.46
ABBE number 58
Geometry
Optic design Biconvex
Overall length 13.0 mm
Optic diameter 6.0 mm
Haptic thickness 0.40 mm
Haptic angulation 0° – asymmetrical design with posterior vaulting
Storage
Temperature +15 – +35°C
Humidity 15% – 50%
Shelf life 5 years (from sterilization)
* Other powers upon request
** It is recommended that surgeons personalize the constants they use. Please find more information about IOL constants on ULIB.
(http://www.ocusoft.de/ulib/c1.htm)
EA-IMCBTHLEN 04 201608
3
Head Office
Herceghalmi Road2072 ZsámbékHungary
International Office
Chemin des Aulx 181228 Plan-les-OuatesGeneva/Switzerland
Medicontur
Medical Engineering [email protected]
677MYHydrophilic Progressive Diffractive IOL
Ø 6 mm
Ø 3 mm
0,4 mm
Diffractive zone
Dif
fracti
ve s
ide
Ø13
mm
Technical specification
Type Single-piece Aspheric Hydrophilic Progressive
Diffractive Intraocular Lens for implantation
into the capsular bag
Optic Diffractive – Aspheric
PCO protection 360° Special Square Edge (patented)
Estimated incision size from 1.8 to 2.2 mm
Powers available* 0.0 D " +35.0 D · (increment: 0.5 D)
Addition +3.5 D
A-constant** 118.9 (SRK/T)
Sterilization Steam
Material
677MY Copolymer of Hydrophilic and Hydrophobic Acrylic,
25% water content with UV absorber + blue light filter
Refractive index 1.46
ABBE number 58
Geometry
Optic design Biconvex
Diffractive zone Anterior surface (diameter 3.0 mm)
Overall length 13.0 mm
Optic diameter 6.0 mm
Haptic thickness 0.40 mm
Haptic angulation 0° – asymmetrical design with posterior vaulting
Storage
Temperature +15 – +35°C
Humidity 15% – 50%
Shelf life 5 years (from sterilization)
* Other powers upon request
** It is recommended that surgeons personalize the constants they use. Please find more information about IOL constants on ULIB.
(http://www.ocusoft.de/ulib/c1.htm)
EA-IMCBMHLEN 04 201608
4
Head Office
Herceghalmi Road2072 ZsámbékHungary
International Office
Chemin des Aulx 181228 Plan-les-OuatesGeneva/Switzerland
Medicontur
Medical Engineering [email protected]
677MTYHydrophilic Progressive Diffractive Toric IOL
Technical specification
Type Single-piece Aspheric Hydrophilic Progressive
Diffractive Toric Intraocular Lens for implantation
into the capsular bag
Optic Progressive Diffractive / Toric – Aspheric
PCO protection 360° Special Square Edge (patented)
Estimated incision size from 2.2 to 2.4 mm
Powers available* +10.0 D " +30.0 D · (increment: 0.5 D)
+31.0 D " +35.0 D · (increment: 1.0 D)
Cylinders available 1.0 D " 4.5 D · (increment: 0.5 D)
5.25 D " 6.0 D · (increment: 0.75 D)
Addition +3.5 D
A-constant** 118.9 (SRK/T)
Sterilization Steam
Material
677MTY Copolymer of Hydrophilic and Hydrophobic Acrylic,
25% water content with UV absorber + blue light filter
Refractive index 1.46
ABBE number 58
Geometry
Optic design Biconvex
Diffractive zone Anterior surface (diameter 3.0 mm)
Overall length 13.0 mm
Optic diameter 6.0 mm
Haptic thickness 0.40 mm
Haptic angulation 0° – asymmetrical design with posterior vaulting
Storage
Temperature +15 – +35°C
Humidity 15% – 50%
Shelf life 5 years (from sterilization)
* Other powers upon request
** It is recommended that surgeons personalize the constants they use. Please find more information about IOL constants on ULIB.
(http://www.ocusoft.de/ulib/c1.htm)
EA-IMCBMTHLEN 04 201606
5
877PA Æ 877PAYHydrophobic intraocular lenspreloaded in a single use injector*
Head Office
Herceghalmi Road
2072 Zsámbék
Hungary
International Office
Chemin des Aulx 18
1228 Plan-les-Ouates
Switzerland / Geneva
Medicontur
Medical Engineering Ltd
www.medicontur.com
LA – IMCPOBMAENG 04 201608
Fill visco in the hole until the optic of thelens is covered (preferably HPMC)
Type Single-piece Aspheric Hydrophobic Acrylic IOL
Material Hydrophobic Acrylic with UV absorber /
+ optional blue light filter
Diameters 6.0 mm (optic) / 13.0 mm (overall)
Angulation 0° – asymmetrical design
Power range -10.0D " +9.0D · (increment: 1.0 D)
+10.0D " +30.0D · (increment: 0.5 D)
+31.0D " +35.0D · (increment: 1.0 D)
A-Constant 118.9 (SRK/T)
Incision size 2.2 mm
1st STEP 2nd STEP
3rd STEP 4th STEP
INJECTION
Rotate the transparent ring to closethe cartridge
Pull out the red stopper Remove the ring-adapter from the injector
Push the plunger slowly to the end position
* Patent pending
6
Head Office
Herceghalmi Road
2072 Zsámbék
Hungary
International Office
Chemin des Aulx 18
1228 Plan-les-Ouates
Geneva / Switzerland
Medicontur
Medical Engineering Ltd
www.medicontur.com
877FAB · 877FABYHydrophobic IOL
13m
m
0,35 mm
6 mm
EA-IMCBHBEN 03 201606
Technical specification
Type Single-piece Aspheric Hydrophobic Intraocular Lens
for implantation into the capsular bag
Optic Refractive – Aspheric
PCO protection 360° Special Square Edge (patented)
Estimated incision size 2.4 mm
Powers available* 0.0 D " +9.0 D · (increment: 1.0 D)
+10.0 D " +30.0 D · (increment: 0.5 D)
+31.0 D " +35.0 D · (increment: 1.0 D)
A-constant** 118.9 (SRK/T)
Sterilization Ethylene-oxide
Material
• 877FAB Hydrophobic Acrylic with UV absorber
• 877FABY + blue light filter
Refractive index 1.47
ABBE number 58
Geometry
Optic design Biconvex
Overall length 13.0 mm
Optic diameter 6.0 mm
Haptic thickness 0.35 mm
Haptic angulation 0° – asymmetrical design with posterior vaulting
Storage
Temperature +15 – +35°C
Humidity 15% – 50%
Shelf life 5 years (from sterilization)
* Other powers upon request
** It is recommended that surgeons personalize the constants they use. Please find more information about IOL constants on ULIB.
(http://www.ocusoft.de/ulib/c1.htm)
7
CONSTANTS FOR BI-FLEX IOLsFOR OPTICAL OR IMMERSION ULTRASOUND MEASUREMENTS
BI-FLEX IOLs
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 0.325a1 = 0.255a2 = 0.141
pACD = 5.01 sf = 1.25 A = 118.1
Bi-Flex – 677AB/677ABY
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 1.243a1 = 0.400a2 = 0.100
pACD = 5.46 sf = 1.67 A = 118.9
Bi-Flex PIL-MA – 677P/677PY
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 1.243a1 = 0.400a2 = 0.100
pACD = 5.46 sf = 1.67 A = 118.9
Bi-Flex M – 677MY
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 1.243a1 = 0.400a2 = 0.100
pACD = 5.46 sf = 1.67 A = 118.9
Bi-Flex PIL-MA – 677PMY
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 1.243a1 = 0.400a2 = 0.100
pACD = 5.46 sf = 1.67 A = 118.9
Bi-Flex T – 677TA/TB - 677TAY/TBY
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 1.243a1 = 0.400a2 = 1.000
pACD = 5.46 sf = 1.67 A = 118.9
Bi-Flex PIL-MA – 677PT/677PTY
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 1.243a1 = 0.400a2 = 0.100
pACD = 5.46 sf = 1.67 A = 118.9
Bi-Flex MT – 677MTY
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 1.320a1 = 0.400a2 = 0.100
pACD = 5.46 sf = 1.70 A = 118.9
Bi-Flex POB-MA – 877PA/877PAY
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 1.320a1 = 0.400a2 = 0.100
pACD = 5.46 sf = 1.70 A = 118.9
Bi-Flex HB – 877FAB/877FABY
Haigis HofferQ Holladay 1SRK/T
Constante A
a0 = 1.243a1 = 0.400a2 = 0.100
pACD = 5.46 sf = 1.67 A = 118.9
Bi-Flex PIL-MA – 677PMTY
It is recommended that surgeons personalize the constants they use based on their techniques, equipment and post-operative results. IOL constants were calculated using previous experienceswith SRK/T formula and relations between constants. (Source:(http://www.ocusoft.de/ulib/c1.htm). For further assistance please contact our Scientific Group at [email protected].
EA-IMCAKONSTEN 01 201606
Head Office
Herceghalmi Road2072 ZsámbékHungary
International Office
Chemin des Aulx 181228 Plan-les-OuatesGeneva/Switzerland
Medicontur
Medical Engineering [email protected]
8
Vision of expertise
An independent European company, ever growing and evolving
since 27 years.
Consistent high quality with more than 4 million intraocular
implants manufactured and implanted.
The most diverse portfolio of Intraocular Lenses, both hydrophilic
and hydrophobic, all premium categories, state of the art
preloaded systems.
A team of more than 250 dedicated professionals constantly
pushing the limits of quality.
Facilities located near Geneva (Switzerland), Limas (France),
Brno (Czech Republic) and near Budapest (Hungary).
Distribution in more than 60 countries with a growing
share worldwide.
Head Office
Herceghalmi Road2072 ZsámbékHungary
International Office
Chemin des Aulx 181228 Plan-les-OuatesSwitzerland
Medicontur
Medical Engineering [email protected]