BC-3600&3300 Service Manual V1.0
-
Upload
jonatan-rolong-ibanez -
Category
Documents
-
view
513 -
download
26
Transcript of BC-3600&3300 Service Manual V1.0
-
7/24/2019 BC-3600&3300 Service Manual V1.0
1/242
BC-3600
BC-3300
Auto Hematology Analyzer
Service Manual
-
7/24/2019 BC-3600&3300 Service Manual V1.0
2/242
-
7/24/2019 BC-3600&3300 Service Manual V1.0
3/242
I
Copyright
2008-2011 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
For this Service Manual, the issued Date is 2011-02 (Version: 1.0).
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray)
owns the intellectual property rights to this Mindray product and this manual. This manual may
refer to information protected by copyrights or patents and does not convey any license under
the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability
arising out of any infringements of patents or other rights of third parties.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure
of the information in this manual in any manner whatsoever without the written permission of
Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rent, adaptation and translation of this
manual in any manner whatsoever without the written permission of Mindray is strictly
forbidden.
. , are the registered trademarks or trademarks owned by
Mindray in China and other countries. All other trademarks that appear in this manual are
used only for editorial purposes without the intention of improperly using them. They are the
property of their respective owners.
Responsibility on the Manufacturer Party
Contents of this manual are subject to changes without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be liable for
errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for safety, reliability and performance of this product only in the
condition that:
all installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national
-
7/24/2019 BC-3600&3300 Service Manual V1.0
4/242
II
and local requirements;
the product is used in accordance with the instructions for use.
This equipment must be operated by skilled/trained medicalprofessionals.
It is important for the hospital or organization that employs this equipment
to carry out a reasonable service/maintenanceplan. Neglect of this may
result in machine breakdown or injury of human health.
Be sure to operate the analyzer under the situation specified in this manual;
otherwise, the analyzer will not work normally and the analysis results will
be unreliable, which would damage the analyzer components and cause
personal injury.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
5/242
III
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other
charges or liability for direct, indirect or consequential damages or delay resulting from the
improper use or application of the product or the use of parts or accessories not approved by
Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
any Mindray product which has been subjected to misuse, negligence or accident;
any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
any product of any other manufacturer.
Safety, Reliability and PerformanceMindray is not responsible for the effects on safety, reliability and performance of BC-2800Vet,
if:
Assembly operations, extensions, re-adjusts, modifications or repairs are carried out
by persons other than those authorized by Mindray.
Personnel unauthorized by Mindray repairs or modifies the instrument.
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to Mindray,
the following procedure should be followed:
1. Obtain return authorization: Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must appear
on the outside of the shipping container. Returned shipments will not be accepted ifthe Mindray number is not clearly visible. Please provide the model number, serial
-
7/24/2019 BC-3600&3300 Service Manual V1.0
6/242
IV
number, and a brief description of the reason for return;
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges);
3. Return address: Please send the part(s) or equipment to the address offered by
Customer Service department.
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address:Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan, ShenZhen 518057, P.R.China,
Phone: +86 755 26582479 26582888
Fax: +86 755 26582934 26582500
EC-Representative: Shanghai International Holding Corp. GmbH(Europe)
Address: Eiffestrae 80, 20537 Hamburg Germany
Phone: 0049-40-2513175
Fax: 0049-40-255726
-
7/24/2019 BC-3600&3300 Service Manual V1.0
7/242
1
Table of Contents
CopyrightIIntellectual Property StatementI
Responsibility on the Manufacturer Party I
WarrantyIII
Exemptions III
Return Policy III
1 Using This Manual 1-1
1.1. Scope 1-2
1.2. Introduction 1-2
1.3. Conventions Used in This Manual 1-3
1.4. Special Terms Used in This Manual 1-3
1.5. Symbols 1-4
1.6. Definitions 1-12
2 Analyzer Structure 2-1
2.1 Product Name 2-1
2.2 System Introduction 2-1
2.3 System Composition 2-7
2.4 Software Structure 2-82.5 LIS Connection 2-19
3 Upgrade 3-1
3.1 Software upgrade 3-1
4 Mechanical System 4-1
4.1 Front of the Analyzer 4-1
4.2 Back of the Analyzer 4-3
4.3 Left of the Analyzer 4-4
4.4 Right of the Analyzer 4-6
5 Fluidic System 51
5.1 Fluidic Components and Functions 5-1
5.2 Reagent Consumption 5-2
5.3 Sample Dilution Flowchart 5-3
5.4 Introduction of Fluidic Channels 5-5
5.5 Introduction of Basic Sequence 5-6
6 Hardware System 6-16.1 Introduction 6-1
-
7/24/2019 BC-3600&3300 Service Manual V1.0
8/242
Table of Contents
2
6.2 Data Board 6-5
6.3 Drive Board 6-18
6.4 Power Board 6-25
6.5 USB Interface Board 6-30
6.6 Touchscreen Control Board 6-31
6.7 Volumetric Board 6-32
6.8 Sample Compartment Connection Board 6-34
6.9 Indicator Board 6-35
7 Servicing 7-1
7.1 General 7-1
7.2 Disassembling the Panels 7-2
7.3 Replacing the Hardware parts and Cable connection 7-6
7.4 Replacing the Power Supply Assembly 7-16
7.5 Replacing the Display Assembly 7-17
7.6 Replacing the Touch Screen 7-18
7.7 Replacing the Recorder Assembly 7-19
7.8 Replacing the valve, pumps, and other liquid system parts 7-20
7.9 Replacing the counting bath assembly 7-29
7.10 Replacing the syringe assembly 7-35
7.11 Replacing the sample probe assembly 7-37
7.12 Replacing the sample compartment assembly 7-47
7.13 Replacing the start key 7-56
7.14 Micro-Switch Assembly Replacement 7-57
7.15 Fluidic Support-bracket Assembly 7-58
7.16 Front Panel Detection Assembly 7-59
7.17 Ambient Temperature Sensor Replacement 7-60
7.18 Cap Assembly 7-61
7.19 Gain Adjustment 7-62
7.20 Motor Type, Position and Different Application List 7-64
7.21 The Check List After Material Replacement 7-64
8 Maintaining your analyzer 8-1
8.1 General Guidelines 8-18.2 Operation guidance of Quality Control and Calibration 8-2
9 Error Information 9-1
9.1 Error Information and Code 9-1
9.2 Error analysis and processing 9-2
9.3 Error diagnosis assisted by abnormal histogram 9-12
9.4 Error diagnosis assisted by abnormal pulse figure 9-21
10 Appendices A-1
-
7/24/2019 BC-3600&3300 Service Manual V1.0
9/242
1-1
1 Using This Manual
This chapter explains how to use this service manual. This manual states in detail the way toservice the BC-3600 and BC-3300 analyzers. Before servicing the BC-3600 and BC-3300,
read and understand the manual carefully for servicing the equipment properly and for your
safety.
This manual is to be used in conjunction with the operator's manuals of BC-3600 and
BC-3300.It does not contain information and procedures already covered in the operator's
manuals of BC-3600 and BC-3300.
Be sure to operate and service the analyzer strictly as instructed in this manual and the
operator's manuals.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
10/242
Using This Manual
1-2
1.1. Scope
To use this manual effectively, you need the following capabilities:
Comprehensive knowledge of circuit and fluidics;
Comprehensive knowledge of reagents;
Comprehensive knowledge of controls;
Comprehensive knowledge of troubleshooting;
Mastering the way to operate this analyzer;
Using basic mechanical tools and understand related terminology;
Using a digital voltmeter (DVM) and an oscilloscope;
Reading pneumatic/hydraulic schematics and understand related terminology.
1.2. Introduction
This manual comprises 9 chapters and 6 appendices. Refer to the table below to find the
information you need.
If you want to See
learn about the system and software structures of BC-3600
and BC-3300
Chapter 2 System Structure
learn about the installation requirements and the way to
upgrade the software of BC-3600 and BC-3300
Chapter 3 System Installation
and Software Upgrade
learn about the mechanical system structure of BC-3600 and
BC-3300
Chapter 4 Mechanical System
learn about the fluidic system, reagent dosage, channels and
basic sequence of BC-3600 and BC-3300
Chapter 5 Fluidic System
learn about the hardware system and board composition,
adjustment and test points and general troubleshooting
methods of BC-3600 and BC-3300
Chapter 6 Hardware System
learn about how to service the BC-3600 and BC-3300 Chapter 7 Servicing
learn about how to maintain the BC-3600 and BC-3300 Chapter 8 Maintenance
learn about how to troubleshoot the common errors of
BC-3600 and BC-3300
Chapter 9 Troubleshooting
learn about the main spare parts of BC-3600 and BC-3300 Appendix A List of Spare parts
learn about the main wearing parts of BC-3600 and
BC-3300
Appendix B List of Wearing
Parts
learn about the circuit diagram of BC-3600 and BC-3300 Appendix C Circuit Diagramlearn about the schematic diagram of the fluidic system of Appendix D Fluidic Diagram
-
7/24/2019 BC-3600&3300 Service Manual V1.0
11/242
Using This Manual
1-3
BC-3600 and BC-3300
learn about the function of each valve and pump of BC-3600
and BC-3300
Appendix E Pump and Valve
Function Diagram
learn about the tubing connection of BC-3600 and BC-3300 Appendix F Tubing
1.3. Conventions Used in This Manual
This manual uses certain typographical conventions to clarify meaning in the text:
Format Meaning
[]all capital letters enclosed in [ ] indicate a key name (either on
the pop-up keyboard or the external keyboard)
letters included in " " indicate text you can find on the screen
of BC-3600 and BC-3300
italic letters indicate titles of the chapters that are referred
to
"Menu"the " " button on the screen.
All illustrations in this manual are provided as examples only. They may not necessarily reflect
your analyzer setup or data displayed.
1.4. Special Terms Used in This Manual
When you read It means
Clickto press the desired item lightly with your finger; or to
left-CLICK it with the mouse.
ENTER
to CLICKthe desired edit box and use the external keyboard
or the pop-up keyboard to enter the desired characters or
digits; or to scan the number by using the bar-code scanner.
DELETE
to move the cursor to the character or digit that you want to
delete by clicking the left button of the mouse or using
[][][Home][End], and then delete the character after the
cursor by pressing [Del], or delete the character before the
cursor by pressing [BackSpace] ([] on the upper right part of
the soft keyboard).
SELECT from
pull-down list
(for pull-down list)
to CLICKthe down arrow button of the desired box to display
the pull-down list, (and DRAG SCROLL BAR) to browse and
then CLICK the desired item; or to press the keys
([][][PageUp][PageDown]) to browse the current list and
-
7/24/2019 BC-3600&3300 Service Manual V1.0
12/242
Using This Manual
1-4
press [ENTER] to select the desired item.
1.5. Symbols
You will find the following symbols in this manual.
When you see Then
read the statement below the symbol. The statement is
alerting you to an operating hazard that can cause
personnel injury.
read the statement below the symbol. The statement is
alerting you to a possibility of analyzer damage or unreliable
analysis results.
read the statement below the symbol. The statement is
alerting you to information that requires your attention.
read the statement below the symbol. The statement is
alerting you to a potentially biohazardous condition.
You may find the following symbols on the analyzer, reagents, controls or calibrators.
When you see It means
CAUTION, CONSULT ACCOMPANYING
DOCUMENTS.
BIOLOGICAL RISK
HIGH VOLTAGE
EXERCISE CAUTION WHEN WORKING
AROUND TO AVIOD PRICKING
PROTECTIVE EARTH (GROUND)
ALTERNATING CURRENT
-
7/24/2019 BC-3600&3300 Service Manual V1.0
13/242
Using This Manual
1-5
FOR IN VITRO DIAGNOSTIC USE
BATCH CODE
USE BY (YYYY-MM-DD)
SERIAL NUMBER
MEASUREMENT AUTHORIZATION
SYMBOL
DATE OF MANUFACTURE
MANUFACTURER
THIS ELECTRO-INFORMATIONAL
PRODUCT CONTAINS CERTAIN
POISONOUS OR HARMFUL SUBSTANCE.ITS ENVIRONMENTAL LIFT TIME IS 20
YEARS. IT CAN BE USED WHTHIN 20
YEARS, AFTER THAT IT SHALL BE
RECYCLED.
Be sure to observe the following precautions for the safety of patients and operators when you
are servicing the analyzer.
It is important for the hospital or organization that employs this equipment to
carry out a reasonable service/maintenance plan. Neglect of this may result
in machine breakdown or harm to human health.
Never use combustible gas (e.g. anesthetic) or combustible liquid (e.g.
ethanol) around the analyzer. Otherwise, the risk of explosion may exist.
When servicing the analyzer, be sure to turn off the power. Servicing the
analyzer when it is on may bring risk of electric shock or damage to
electronic components.
Connect the analyzer to a socket having sole fuse and protective switch. Do
-
7/24/2019 BC-3600&3300 Service Manual V1.0
14/242
Using This Manual
1-6
not use the same fuse and protective switch with other equipment (e.g. life
supporting equipment). Otherwise, the equipment failure, over current or
impulse current that occurs at the startup moment may lead to tripping.
To prevent personal injury during maintenance, keep your clothes, hairs and
hands from the moving parts, such as sample probe, clipper and piercer.
Possible mechanical movement of the warned position may lead to personal
injury during the normal operation, removal and maintenance.
Be sure to dispose of reagents, waste, samples, consumables, etc.
according to government regulations.
The reagents are irritating to eyes, skin and mucosa. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
Improper maintenance may damage the analyzer. Maintain the analyzer
strictly as instructed by the service manual and inspect the analyzer
carefully after the maintenance.
For problems not mentioned in the service manual, contact Mindray
customer service department for maintenance advice.
To prevent personal injury or damage to equipment components, remove
metal jewelry before maintaining or servicing electronic components of the
equipment.
Electrostatic discharge may damage electronic components. If there is a
possibility of ESD damage with a procedure, then do that procedure at an
ESD workstation, or wear an antistatic wrist strap.
When the front cover is open, be sure to close the sample compartment door,
or the sample compartment assembly or the front cover may be damaged.
When transporting the analyzer, the screw at its bottom may scratch your
hands, please exercise caution or wear gloves.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
15/242
Using This Manual
1-7
This equipment must be operated by skilled/trained medical professionals.
Samples, controls, calibrators and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
All the analyzer components and surfaces are potentially infectious. Take
proper protective measures for operation or maintenance.
The sample probe tip is sharp and may contain biohazardous materials.Exercise caution to avoid contact with the probe when working around it.
Figure 1-1
-
7/24/2019 BC-3600&3300 Service Manual V1.0
16/242
Using This Manual
1-8
Warning, potential biohazardous risk.
Make sure the analyzer is properly grounded.
To avoid electric shock, disconnect power cord prior to removing or replacing fuse.
Use fuse only with the type and rating specified.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
17/242
Using This Manual
1-9
Figure 1-2
Warning, the sample probe is sharp and may contain biohazardous substance. Exercise
caution to avoid contact with the probe when working around it.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
18/242
Using This Manual
1-10
Figure 1-3
Warning, make sure the protective cover is properly installed before operation.
Warning
Do not put your hands close to the sampling assembly when the analyzer is running.
Warning, the sample probe is sharp and may contain biohazardous substance.
Exercise caution to avoid contact with the probe when working around it.
Warning, the sample probe is sharp and may contain biohazardous substance. Exercise
-
7/24/2019 BC-3600&3300 Service Manual V1.0
19/242
Using This Manual
1-11
caution to avoid contact with the probe when working around it.
Figure 1-4
Warning, risk of electric shock.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
20/242
Using This Manual
1-12
1.6. Definitions
There are two sampling mode: open vial mode and closed tube mode.
Open vial mode
The operator presents the tube or open vial evacuated collection tube containing blood sample
to the sample probe, and press the "Aspirate" key to collect sample.
The analyzer of which the sampling mode is open vial is called open vial analyzer.
Closed tube mode
The operator puts the capped tube into the sample compartment, the analyzer collects sample
by piercing through the tube cap.
The analyzer of which the sampling mode is closed tube is called closed tube analyzer.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
21/242
2-1
2 Analyzer Structure
2.1 Product Name
Name: Auto Hematology Analyzer.
Models: BC-3600, BC-3300.
2.2 System Introduction
BC-3600 and BC-3300 are in-vitro diagnosis equipments with the functions of blood cell
counting, WBC 3-differential and measurement of HGB concentration under proper working
environment. The purpose of the analyzers is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that require additional
studies.
Scope: the products can be applied for clinical diagnosing reference and scientific
research.
Environment of use: the analyzers shall be used in medical laboratories with
standardized management; they can not be used as portable equipment.
Operator: the operators must be skilled/trained medical professionals.
2.2.1 System Configuration
The whole system includes the analyzer, accessories and reagents; users can choose to
configure data management software, scanner, printer or PC.
2.2.2 Electrical Parameters
Table2-1 Power supply of the analyzer
Voltage (100V-240V)10%
Input Power 180 VA
Frequency 50/601Hz
Only install fuses of specified specification on the analyzer.
Specification of the fuse: 250V T3.15AH
-
7/24/2019 BC-3600&3300 Service Manual V1.0
22/242
Analyzer Structure
2-2
2.2.3 Dimension and Weight
Figure 2-1Dimensional sketch map
Table2-2Dimension and weight
Width (mm) Depth (mm) Height (mm) Weight (Kg)
395 450 445 (without foot) 28
2.2.4 ThroughputThe throughput of open vial analyzers under whole blood mode is no less than 60 samples per
hour; the time needed for the analysis of one sample is no more than 60s (the sample dilution
time is not included for predilute mode).
The throughput of closed tube analyzers under whole blood mode is no less than 60 samples
per hour; the time needed for the analysis of one sample is no more than 60s (the sample
dilution time is not included for predilute mode).
2.2.5 ParametersThe analyzer determines 21 parameters (WBC, RBC, PLT, HGB, etc.) and 3 histograms
(WBC, RBC and PLT) of blood samples.
Table 2-3Table of parameters
English name Abbreviation
White Blood Cell count WBC
Lymphocyte number Lymph#
Mid-sized Cell number Mid#
Granulocyte number Gran#
Lymphocyte percentage Lymph%
Width
Height
Depth
-
7/24/2019 BC-3600&3300 Service Manual V1.0
23/242
Analyzer Structure
2-3
Mid-sized Cell percentage Mid%
Granulocyte percentage Gran%
Red Blood Cell count RBC
Hemoglobin concentration HGB
Mean Corpuscular Volume MCVMean Corpuscular Hemoglobin MCH
Mean Corpuscular Hemoglobin Concentration MCHC
Red Blood Cell Distribution Width Coefficient of
Variation
RDW-CV
Red Blood Cell Distribution Width Standard
Deviation
RDW-SD
Hematocrit HCT
Platelet count PLT
Mean Platelet Volume MPV
Platelet Distribution Width PDW
Plateletcrit PCT
Platelet Larger Cell Ratio P-LCR
Platelet Larger Cell Count P-LCC
Table2-4Table of histograms
White Blood Cell Histogram WBC Histogram
Red Blood Cell Histogram RBC Histogram
Platelet Cell Histogram PLT Histogram
2.2.6 Performance Specifications
2.2.6.1 Background or Blank Counting
Background counting is performed by the analyzer automatically when it is started up. The
following requirements must be met.
Blanking counting: test the diluent for 3 consecutive times, and take the highest result. The
following requirements must be met.
Table 2-5 Requirements on background or blank counting
Parameter Requirement
WBC 0.3 109/ L
RBC 0.031012
/ L
HGB 1 g / L
HCT 0.5 %
PLT 5109/ L
-
7/24/2019 BC-3600&3300 Service Manual V1.0
24/242
-
7/24/2019 BC-3600&3300 Service Manual V1.0
25/242
Analyzer Structure
2-5
RBC 08.001012
/L 0.051012
/L or 5 0.051012
/L or 5
HGB 0280g/L 2g/L or 2 2g/L or 3
01000109/L 1010
9/L or 10 1010
9/L or 10PLT
10013000109/L 12 20
HCT 060%4% (HCT value) or 6%
(deviation)
/
Test suggestion:
When testing RBC linearity range 7.008.001012
/L, use sample of which the HCT 60%;
When testing PLT linearity range 10013000109/L, use sample of which the RBC 6.
2.2.6.4 Carryover
The carryover impact of high concentration sample to low concentration sample.
Test method: mix the high value sample (centrifuged high value control or specialized high
value linearity control) within the range of Table 2-9, test the sample for 3 consecutive times,
and then mix and test the low value sample (diluted low value control, dilution rate 1:10) within
the range of Table 2-9 for 3 consecutive times, then calculate the carryover. The requirements
in Table 2-10 must be met.
Table 2-9 Requirements on high and low value controls of carryover calculation
Parameter Unit High concentration Low concentration
WBC 109
/L > 15.0 < 3.0
RBC 1012
/L > 6.0 < 2.00
HGB g/L > 200 < 40
PLT 109/L > 300 < 100
Table 2-10 Carryover requirements
Parameter Carryover
WBC 0.5
RBC 0.5
HGB 0.5
PLT 1.0
2.2.7 Functions
2.2.7.1 Startup and Shutdown
When starting up the analyzer, initialization will be performed to prepare the analyzer for
counting. Different startup cleaning sequence will be performed base on the latest shutdown
status to ensure the success of background counting. The analyzer provides the function to
skip fluidics initialization to the users of service level and above.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
26/242
Analyzer Structure
2-6
The shutdown procedure will be performed according to different shutdown status.
2.2.7.2 Password and User Management
In order to protect the security of product setup and data, you have to possess the password to
be able to adjust the parameters and perform the extra functions.
2.2.7.3 Screen Navigation
The analyzer provides screen navigation function so that users can easily go to their target
screen.
2.2.7.4 Status Indication
The analyzer/system status indication function is provided on the analyzer screen or by
configuring LED in the hardware system.
2.2.7.5 Setup
Users can customize the settings of the analyzer on the software screen.
1. The configuration files are saved in the analyzer.2. The analyzer must allow users to modify certain configurations. E.g.: Users can set
up date/time, reference range, parameter unit, communication-related parameters,
print format, etc.
3. For the FDA model, users can set whether to display or print RUO parameters
through the authorization protocol.
4. For the FDA model, users can set "QC lockout", if this function is activated, sample
analysis can be started only after the daily QC passes.
2.2.7.6 Standby
The analyzer will enter standby mode when no fluidic operation has been performed for a
certain period of time. The analyzer will exit the standby mode when operation is triggered
by users.
2.2.7.7 Self Test
The analyzer system provides self test and inspection function to ensure proper running.
2.2.7.8 Sample Review
After analyzing a sample, the analyzer saves the sample information and analysis result
automatically. Operators can review all parameters and histograms of the saved samples.
2.2.8 Sample Compartment
The tube holder in sample compartment can be fitted with adapters of different specifications.
The specifications of supported tubes are:
Evacuated collection tube: diameter: 1013.5mm; height (with cap): 4085mm.
Capillary collection tube: diameter: 1013.5mm; height (with cap): 4065mm.
Centrifugal tubes.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
27/242
Analyzer Structure
2-7
2.2.9 Operating, Storing and Running Environment
Table 2-11 Temperature and Humidity Requirements
Optimal operating
environment
Storage environmentRunning environment
Temperature 1530 -1040 1535
Humidity 30%85% 10%93% 30%85%
Atmospheric
pressure
70kPa106kPa 50kPa106kPa70kPa106kPa
2.3 System Composition
2.3.1 Functional Composition
The analyzer system has the following functions: reagent system, sample distribution, sample
preparation, sample analysis, signal processing, parameter analysis, data management, status
monitoring, schedule control, information processing, human-machine interface, power source,
cleaning and maintenance. The relations of the functional modules are shown in Figure 2-2 .
Under the control of the schedule control and information processing functional module, the
other modules work together according to the designed procedure and requirement to realize
the core task of the analyzer - sample analysis.
Fluid SignalstreamReagentsAirflow
-
7/24/2019 BC-3600&3300 Service Manual V1.0
28/242
Analyzer Structure
2-8
Figure 2-2 System function block diagram
2.3.2 Reagents
In order to ensure the accuracy of analysis results, you shall use the reagents, controls
and calibrators specified by Shenzhen Mindray Bio-medical Electronics Co., Ltd. For
details, refer to Table 2-12.
Table 2-12 Applicable reagents, controls calibrators
Item Name
Diluent M-30D DILUENT
Lyse M-30CFL LYSE
RinseM-30R RINSE
Probe cleanser PROBE CLEANSER
Control B30 or BC-3D
Calibrator S30 or SC-CAL PLUS
2.4 Software Structure
2.4.1 Menu Structure
The top level structure of the system menu includes 8 functional options: analysis, review, QC,
calibration, setup, service, logoff and shutdown. Operators can enter the screens to perform
the functions of the analyzer through the system menu. The " " indicates there is sub-menu
under the option. Refer to Figure 2-3 -- Figure 2-7 for the full menu structure.
The menu structure of service access level:
-
7/24/2019 BC-3600&3300 Service Manual V1.0
29/242
Analyzer Structure
2-9
Figure 2-3 Top level menu and QC sub-menu
Figure 2-4 Top level menu and calibration sub-menu
-
7/24/2019 BC-3600&3300 Service Manual V1.0
30/242
Analyzer Structure
2-10
Figure 2-5 Top level menu and setup sub-menu
Figure 2-6 Top level menu and setup sub-menu(2)
-
7/24/2019 BC-3600&3300 Service Manual V1.0
31/242
Analyzer Structure
2-11
Figure 2-7 Top level menu and service sub-menu
2.4.2 Password
The login service ID is "Service", password is "s 3600!", case sensitive and space required.
With service access level, you can perform all functions that are open to common users and
some other special functions:
1Transfer factor calibration function:
1Under whole blood mode, click "Menu">"Calibration">"Calibrator", finish the
calibration successfully, exit and save the calibration factors.
2Click "Menu">"Calibration">"Transfer factor" to enter the screen in Figure 2-8,
run the sample used in the above step for 5 consecutive times under whole blood
mode, check if CV(%) meets the requirements in Table 2-13; if not, cancel a abnormal
result and calculate again (results of at least 5 analyses required), then check if the
requirements in Table 2-13 are met.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
32/242
Analyzer Structure
2-12
Figure 2-8Analysis screen of transfer factor under whole blood mode
Table 2-13 Whole blood CV range
Param
eter
WBC RBC HGB MCV PLT
CV
value
2.0% 1.5% 1.5% 0.5% 4.0%
3Click the "Predilute" button to switch to the screen of predilute mode, as Figure 2-9
shows;
-
7/24/2019 BC-3600&3300 Service Manual V1.0
33/242
Analyzer Structure
2-13
Figure 2-9 Analysis screen of transfer factor under predilute mode
4Click the Diluent icon, perform diluent dispensing for 7 consecutive times;
5Add 140l of calibrator in to the diluent, mix thoroughly and then run the mixture for
more than 5 times under the predilute mode. The CV(%) values must meet the
requirements of Table 2-14, and the calibration factors are within the range [0.75,
1.25].
Table 2-14 Range of CV value under predilute mode
Param
eter
WBC RBC HGB MCV PLT
CV
value
4.0% 2.0% 2.0% 1.5% 8.0%
2The gain of WBC and RBC can be adjusted, see Figure 2-10;
-
7/24/2019 BC-3600&3300 Service Manual V1.0
34/242
Analyzer Structure
2-14
Figure 2-10 Gain setup screen
3In the service setup screen, you can set up languages, software upgrade, volume of the
volumetric tube, analyzer SN, sample probe positioning adjustment, background
information monitoring (auto adjust of analysis time), rinse photocoupler calibration, etc.
The following section will introduce the sample probe positioning adjustment and rinse
photocoupler calibration functions.
1) Sample probe positioning adjustment:
Tool: the positioning calibration clamp of sample probe sampling position
Procedure: open the cover as Figure 2-11 shows, open the sample compartment door,
take out the tube holder adapter from the compartment, put the positioning calibration
clamp to the sample compartment, and then close the compartment door.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
35/242
Analyzer Structure
2-15
Figure 2-11 Sample compartment
Click the "Aspirate pos." under the "X direction motor" as shown in Figure 2-12, and click
"Start", then click the "Move Left" or "Move Right" button to move the sample probe to the
top of the positioning hole of the fixture, check if the tip of the probe is right against the
positioning hole, then click "Save" to save the setting.
Figure 2-12 Sample probe positioning adjustment
Note: the tip of the probe must be right against the positioning hole (judge by visual
inspection).
2) Rinse photocoupler calibration
Tool: 101 slot-headed screwdriver
Procedure: as Figure 2-13 shows, click "Menu""Setup""Service
Setup""Photocoupler Cal.", the dialog box shown in Figure 2-14 will pop up.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
36/242
Analyzer Structure
2-16
Figure2-13 Rinse photocoupler calibration menu
Figure2-14 Rinse photocoupler calibration screen
-
7/24/2019 BC-3600&3300 Service Manual V1.0
37/242
Analyzer Structure
2-17
Click "Start" and then click "OK", the dialog box "Priming diluent" will pop up, then an
instruction message will be displayed, use the slot-headed screwdriver to rotate the
potentiometer VR1(RINSE) on the drive board (see Figure 2-15) counter-clockwise
until the D22 (RINSE-LED) turns on. Note: continue to rotate the VR1 for 180 degrees
after the D22 turns on.
Figure 2-15
Connect the connector of the rinse pickup tube to the fitting on the back panel of the
analyzer, click "OK" and then click "Finish", the dialog box "Priming rinse" will pop up,
check the calibration status then: if "Pass" is displayed (see Figure 2-16), that means
the calibration succeeded. The reference voltage and measured voltage are displayed
as "Reference(V)" (1.8-3.0V) and Actual(V) in Figure 2-16.
Figure 2-16
Note:
Rotate the VR1 for 180 degrees more after the D22 turns on, and make sure the LED
is illuminated.
Do not modify the value in the edit box Max(V), Min(V) and Factor(V);
If calibration cannot be done successfully due to the quality of the photocoupler itself
and the PCBA or photocoupler must be replaced, then recalibration is needed. The
dialog box "Remove the Rinse tube from the container and press [OK] to continue."
will pop up, press [OK] to drain the tubing; then do as instructed by the prompt - "1.
Adjust rinse photocoupler potentiometer till the RINSE-LED turns on; 2 Put the Rinse
tube into the container and press [OK]to close the dialog box; 3. Click [Finish] button"
to prime rinse after calibration is done.
"PASS"
displayed
-
7/24/2019 BC-3600&3300 Service Manual V1.0
38/242
Analyzer Structure
2-18
4The reproducibility and carryover functions are available.
Figure 2-17 Menu of reproducibility and carryover
5Check the replacement status of boards and CF card automatically. If the software
system is started and logged in by user of service level, then a prompt will guide the user
to restore the critical data.
Figure 2-18
1) If a new CF card is replaced, follow the guidance to select "Restore" and restore the
critical data to the CF card.
2) If a new data board is replaced, follow the guidance to select "Backup" and backup the
data to the data board.
3) If only the drive board is replaced, no special action shall be taken, the data will not be
-
7/24/2019 BC-3600&3300 Service Manual V1.0
39/242
Analyzer Structure
2-19
lost.
6Users of the service level can export special information, see the figure below.
Figure 2-19
2.5 LIS Connection
2.5.1 LIS Setup and Adjustment
1. Click "Menu""Setup""System Setup""Communication", the following screen will
be displayed.
Figure 2-20 Communication setup screen
-
7/24/2019 BC-3600&3300 Service Manual V1.0
40/242
Analyzer Structure
2-20
2. Select network communication or serial port communication base on the physical
medium used.
3. If serial port communication is selected, communication can be realized when the
parameter settings of the serial port are identical with those of the LIS software. The
handshake function shall also be set up base on the LIS software used.
4. If network communication is selected, it is recommended that direct-coupled network
wire is used, and the major IP addresses such as 192.168.0. or 10.0.0.* are set up.
When connected to LAN, set up the IP address per the IP rules of the LAN. ACK
synchronous transmission shall also be set up base on the LIS software settings. If
you are not sure about that, use the default setup.
2.5.2 Connection Testing
It is recommended that you use Mindray data management software to perform actual
communication testing, or you can use the testing software developed by a compatible
software developer.
2.5.3 Analysis of Communication Failure
1. For abnormal serial port connection, check parameter settings. If the settings are
identical to those of the analyzer, replace serial port wire.
2. For abnormal network connection, check parameter settings. If the settings are correct,
check if the connection is normal by using ping. (If network firewall is installed in the
computer of the LIS end, turn off the firewall). If the ping command fails to restore the
connection, check if the network wire is well connected.
If physical connection is good, and the communication parameters and protocol settings are
correct, but communication failure persists, check if the LIS software complies with the
communication protocol. Refer to the operator's manual for the communication protocols.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
41/242
3-1
3 Upgrade
3.1 Software upgrade
3.1.1 Preparation
1The whole software upgrade procedure will take about 6~11mins.
2What to do if there is an ERRORduring the upgrade
1If there is a Mounting Failureplease restart the analyzer and run software upgrade again.
2If there is a Upgrade Failure, please check the upgrade package in your USB flash disk and
run software upgrade again. Do NOT restart or power off the analyzer immediately.
3Requirements of the USB flash disk used in this upgrade.
1Please use well-known brand USB flash disk.
2Format(FAT32) the USB flash disk before use it.
3Do NOT remove the USB flash disk during the upgrade.
3.1.2 Upgrade procedure
1Copy the update fileto a formatted USB flash disk.
2Insert the USB flash disk into the USB interface of analyzer (See figure 3-1)
Figure 3-1
3Log in the BC-3600 with Service Engineer username and password.
4Run SetupService SetupSetupStart Upgrade (see figure 3-2), and then click Yes
in the dialog box (see figure 3-3).
-
7/24/2019 BC-3600&3300 Service Manual V1.0
42/242
Upgrade
3-2
Figure 3-2
Figure 3-3
NOTES:
1Make sure the UBS storage connected well to the USB interface of the Analyzer.
2After you click Yes to start the upgrade, the system will start to read and analyze the
configuration in the UBS storage and this will last for about 1min. Please wait for the pop-up
upgrade confirmation dialog box.
3Do NOT remove the USB flash disk during the upgrade.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
43/242
Upgrade
3-3
5software upgrade.
1) See the following software upgrade screen:
Figure 3-4
2) Click the Upgrade button to start the upgrade. You will see the upgrade preparation dialog.
(See figure 3-5)
Figure 3-5
3) The upgrade preparation lasts for about 1 min, after that, you will see the upgrade confirm
dialog. (See figure 3-6)
Figure 3-6
4) Click Yes to start the upgrade. (See figure 3-7)
-
7/24/2019 BC-3600&3300 Service Manual V1.0
44/242
Upgrade
3-4
Figure 3-7
5) You may need to restart the instrument for several times during the upgrade. (See figure
3-8)
Figure 3-8
6) Upgrade succeeded. (See figure 3-9)
Figure 3-9
7) Upgrade failed. (See figure 3-10)
-
7/24/2019 BC-3600&3300 Service Manual V1.0
45/242
Upgrade
3-5
Figure 3-10
8) After you finish all the process, please restart the instrument. (See figure 3-11)
Figure 3-11
6After upgrade finished, log in with the service engineer username and password. Run
ServiceVersion to check upgrade items. (See figure 3-12)
Figure 3-12
-
7/24/2019 BC-3600&3300 Service Manual V1.0
46/242
Upgrade
3-6
3.1.3 Notes and Prompts of Errors
1During the upgrade, the software will check whether the hardware and software are
matched. If not, select the proper language, click OK to start upgrade. (See figure 3-13)
Figure 3-13
2If upgrade failed, export the debug data. (See figure 3-14)
Figure 3-14
-
7/24/2019 BC-3600&3300 Service Manual V1.0
47/242
4-1
4 Mechanical System
4.1 Front of the Analyzer
Figure 4-1 Front of the analyzer(closed tube model)
1 --- Touchscreen 2 --- Indicator
3 --- Sample compartment door 4 --- USB interface
5 --- Recorder
-
7/24/2019 BC-3600&3300 Service Manual V1.0
48/242
Mechanical System
4-2
Figure 4-2 Front of the analyzer(open vial model)
1 --- Touchscreen 2 --- Indicator
3 --- Sample probe 4 --- Aspirate key
5 --- USB interface 6 --- Recorder
-
7/24/2019 BC-3600&3300 Service Manual V1.0
49/242
Mechanical System
4-3
4.2 Back of the Analyzer
Figure 4-3 Back of the analyzer
1 --- USB interface 1 2 --- USB interface 2
3 --- Network interface 4 --- Serial port
5 --- Power switch 6 --- Waste outlet
7 --- Rinse outlet 8 --- Diluent outlet
-
7/24/2019 BC-3600&3300 Service Manual V1.0
50/242
Mechanical System
4-4
4.3 Left of the Analyzer
Figure 4-4 Left of the analyzer (open vial model)
1 --- Top cover 2 --- Front cover assembly
3 --- Side door of electrical system 4 --- Back panel assembly
5 --- Bottom panel assembly
-
7/24/2019 BC-3600&3300 Service Manual V1.0
51/242
Mechanical System
4-5
Figure 4-5 Left of the analyzer (closed tube model)
1 --- Top cover 2 --- Front cover assembly
3 --- Side door of electrical system 4 --- Back panel assembly
5 --- Bottom panel assembly
-
7/24/2019 BC-3600&3300 Service Manual V1.0
52/242
Mechanical System
4-6
4.4 Right of the Analyzer
Figure 4-6 Right of the analyzer (open vial model)
1 --- Top cover 2 --- Front cover assembly
3 --- Side door of fluidic system 4 --- Door lock
5 --- Aspirate key 6 --- Back panel assembly
7 --- Bottom panel assembly
-
7/24/2019 BC-3600&3300 Service Manual V1.0
53/242
Mechanical System
4-7
Figure 4-7 Right of the analyzer (closed tube model)
1 --- Top cover 2 --- Front cover assembly
3 --- Side door of fluidic system 4 --- Door lock
5 --- Aspirate key 6 --- Back panel assembly
7 --- Bottom panel assembly
-
7/24/2019 BC-3600&3300 Service Manual V1.0
54/242
-
7/24/2019 BC-3600&3300 Service Manual V1.0
55/242
5-1
5 Fluidic System
5.1 Fluidic Components and Functions
5.1.1 Sample Probe
Aspirating and dispensing sample.
5.1.2 Probe Wipe
Cleaning the exterior and interior walls of the sample probe.
5.1.3 Pump Gas pump (GP): provide pressure to the vacuum chamber for generation of bubbles.
Waste pump (LP): draining probe wipe, WBC bath, RBC bath and vacuum chamber
and building vacuum.
5.1.4 Syringes
Sampling syringe (Asp-Syringe): full measuring range is 100l; aspirating fixed
quantity of sample, dispensing the sample and aspirating the diluted sample again.
Diluent syringe (Dil-Syringe): full measuring range is 10ml; dispensing fixed quantity
of diluent to WBC and RBC bath, providing diluent to probe wipe.
Lyse syringe (Lyse-Syringe): full measuring range is 2.5ml, motivated by the same
motor with the diluent syringe; dispensing lyse.
5.1.5 Valves
Fluid valve and magnetic valve: diaphragm form, controlled by electric magnetic
power. Controlling the directional flow of fluid or air.
5.1.6 Baths
WBC bath: composed of front bath, back bath and aperture. Providing room for the
mixing of WBC sample; measuring HGB and WBC.
RBC bath: composed of front bath, back bath and aperture. Providing room for the
mixing of RBC sample; measuring RBC/PLT.
Vacuum chamber: build and store stable vacuum for the impedance counting of WBC
and RBC, clean the back baths and drain volumetric tubes.
Pressure chamber: build and store stable pressure for generation of bubbles in all
baths.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
56/242
Fluidic System
5-2
WBC isolation chamber: providing air lock to block outside disturbance.
RBC isolation chamber: providing air lock to block outside disturbance.
5.1.7 Volumetric Tubes
WBC volumetric tube: its volume is 300ul, measuring volume of the WBC sample to
be analyzed.
RBC volumetric tube: its volume is 200ul, measuring volume of the RBC sample to
be analyzed.
5.1.8 Filters
Filter of WBC volumetric tube: filtrate the air entering the WBC volumetric tube.
Filter of RBC volumetric tube: filtrate the air entering the RBC volumetric tube.
GP filter: filtrate the air entering the gas pump.
Customized filter: filtrate the impurities and scraps from the probe wipe.
Filter of the isolation chamber: filtrate the impurities and scraps from the baths.
5.2 Reagent Consumption
There are four counting modes: OV-WB, OV-PD, CT-WB and CT-PD. See Table 5- 1 for the
reagent volume consumed in one analysis cycle of a sample.
Table 5- 1Reagent consumed in one analysis cycle of a sample
Reagent consumed in
one analysis cycle of a
sampleItem Name
Whole blood Predilute
Diluent M-30D DILUENT 28mL 28mL
LyseM-30CFL LYSE
0.35mL 0.35mL
Rinse M-30R RINSE 8mL 8mL
Probe
cleanserPROBE CLEANSER
0 0
Control B30BC-3D /
Calibrator S30SC-CAL PLUS /
-
7/24/2019 BC-3600&3300 Service Manual V1.0
57/242
Fluidic System
5-3
5.3 Sample Dilution Flowchart
5.3.1 Whole Blood Mode
The whole blood analysis cycle completes in 60s. The analysis procedure is as follows:
1) Aspirate 17ul of sample, the sample probe ascends and the exterior wall of the probe is
cleaned;
2) Dispense 6ul of sample to the probe wipe to wash it off;
3) Add diluent into the WBC bath and dispense 9ul of sample into the WBC bath, mix them to
get the first dilution sample for WBC and HGB measurement.
4) The sample probe ascends and its exterior wall is cleaned, the residual sample is washed
off by diluent in the probe wipe.
5) The sample probe descends to the WBC bath and aspirate 34ul of sample again, diluent is
added and the second dilution sample is dispensed into the RBC bath, and mix them forRBC/PLT counting.
6) Add lyse into the WBC bath, and mix thoroughly;
7) Analyzing the sample;
8) When the analysis completes, clean and zap the WBC/RBC bath, release the pressure in
the vacuum chamber.
See Figure 5-1 for the dilution procedure of whole blood mode:
Figure 5-1 Dilution procedure of whole blood mode
-
7/24/2019 BC-3600&3300 Service Manual V1.0
58/242
Fluidic System
5-4
5.3.2 Predilute Mode
The predilute analysis cycle includes 3 function sequences (2 for closed tube models): diluent
dispensing sequence, cleaning sequence (open vial models) and analyzing sequence. The
procedure is as follows:1) The sample probe dispense 380ul of diluent to a centrifugal tube and 20ul of sample in the
capillary tube is added in to the centrifugal tube, mix them to form the diluted sample;
2) The sample probe aspirate 146.34ul of sample, and diluent is added into the WBC bath to
form the first dilution sample for WBC and HGB measurement;
3) The sample probe ascends and its exterior and interior walls are cleaned;
4) The sample probe descends to the WBC bath and aspirate 37.39ul of sample again, diluent
is added and the second dilution sample is dispensed into the RBC bath, and mix them for
RBC/PLT counting.
5) Add lyse into the WBC bath, and mix thoroughly;
6) Analyzing the sample;
7) When the analysis completes, clean and zap the WBC/RBC bath, release the pressure in
the vacuum chamber.
You can see from the above procedures that the analysis cycles of predilute mode and whole
blood mode are only different in the operations of sampling and dispensing into WBC bath,
therefore this section will only introduce the different parts.
See Figure 5-2 for the dilution procedure of predilute mode:
-
7/24/2019 BC-3600&3300 Service Manual V1.0
59/242
Fluidic System
5-5
Figure 5-2 Dilution procedure of predilute mode
5.4 Introduction of Fluidic Channels
5.4.1 WBC/HGB Channel
All reagents:1)M-30PCFL lyse (lysing HGB, PLT and separating basophils from
other WBCs ); 2) M-30PD diluent (cleaning, providing reaction and measurement
environment)
Analysis principles: impedance method (analyzing WBC); colorimetric method(analyzing HGB)
Analysis parameters: WBC, HGB
Graphic information: WBC histogram
Dilution ratio:1:302 (WB); 1:369.77 (PD)
Measurement volume: 300l
Function description: Mix 9ul of blood and 2.35ml of diluent in the WBC bath,
provide 34ul of first dilution sample to the RBC bath and add 0.35ml of lyse into it,
mix them; then aspirate the mixed sample into the back bath through the aperture by
the vacuum of the vacuum chamber. The cells are analyzed when they pass the
aperture. Volume of the sample analyzed is measured by the volumetric tube.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
60/242
Fluidic System
5-6
5.4.2 RBC/PLT Channel
All reagents: diluent (diluting, cleaning, providing conductive environment, isometric
processing of cells)
Analysis principles: impedance method
Analysis parameters: RBC, PLT
Graphic information: RBC histogram, PLT histogram
Dilution ratio:1:20000 (WB); 1:22333.07 (PD)
Measurement volume: 200l
Function description: the sample probe aspirates 34ul of sample (dilution ratio:
1:262.11) from the WBC bath, add 0.8ml of mixture of the sample and diluent into the
RBC bath to form the sample of dilution ratio 1:20000 with another 1.8ml of diluent;
Mix them and aspirate the sample by vacuum of the vacuum chamber into the back
bath through the aperture. The cells are analyzed when they pass the aperture.
Volume of the sample analyzed is measured by the volumetric tube.
5.5 Introduction of Basic Sequence
5.1.1 Whole Blood Analysis Sequence
This section introduces the working procedure of sample probe in the measurement sequence
first, then all the other working procedures are introduced in detail.
5.5.1.1 Sampling and dispensing procedure
The working procedure of sample probe is indicated in Figure 5- 3. See section 5.3 for the
sample dilution procedure. The working procedure of sample probe:
a) Sample aspirating
piercing(closed tube models)
aspirating 17ul of sample.
b) First dilution
When aspiration finishes, the sample probe ascends;
The sample probe moves to the top of the WBC bath; The sample probe descends to the WBC bath and dispenses 9ul of sample.
c) Aspirating sample from the WBC bath
The sample probe ascends from the WBC bath, and removes the residual
sample in the probe wipe;
The sample probe descends to the WBC bath and aspirates 34ul of first dilution
sample.
d) Second dilution in the RBC bath
When aspiration finishes, the sample probe ascends from the WBC bath;
The sample probe moves to the top of the RBC bath;
The sample probe descends to the RBC bath, and dispenses 0.8ml of mixture of
the sample and diluent.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
61/242
Fluidic System
5-7
Initialize position of the sample probe by command INIT.
RBC WBC
?
?
?
RBC WBC
? ?
RBC WBC
?
a) b)
c) d)
?
?
Secon
d
dilutionsam le
First
dilutio
n
sampl
DiluentBlood
sample
Figure 5- 3 Sampling and dispensing procedure
5.5.1.2 The measurement procedure
This section highlights the whole blood measurement sequence of the closed tube models.
The measurement sequence consists of sample aspiration, dispensing, dilution, mixing,
counting, and cleaning.
Sampling (0~12.6s)
The actions are:
Pre-piercing: the pre-piercing process of sample probe (for closed tube models)
to release the pressure in the tube before sample aspiration.
Cleaning the exterior and interior walls of sample probe: when piercing finishes, the sample
probe ascends to home position, meanwhile its exterior and interior walls are washed. See the
following figure.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
62/242
Fluidic System
5-8
Diluent
Waste
V01
V04V02
100ul
10ml
V19
Diluent
Waste
V01
V04V02
V19
b) Wash interior wall of
sample probe
a) Wash exterior wall of
sample probe
100ul
10ml
Figure 5- 4 Fluidic passages of sample probe cleaning
The sample probe descends to the sampling position, and then moves to the
piercing position, the syringe withdraws for 2ul to create separating bubbles.
Then the sample probe descends to the lowest aspirating position to prepare for
sample aspirating. Besides, the fluidic passages of draining and dispensing of
the WBC bath are indicated in Figure 5- 5 and Figure 5- 6; Under the effect of
vacuum, rinse flows through and washes the WBC and RBC back baths and
their volumetric tubes, the fluidic passage of washing the WBC back bath and
volumetric tube is indicated in Figure 5- 7; When the washing progress ends,
Valve 5 and 6 opens to drain the volumetric tubes.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
63/242
Fluidic System
5-9
Diluent
Waste
V01
Isolationchamber
WBC
bath
V16
V03
Lyse
V04
V02
10mL
SR3
2.5mL
(SR1)
Figure 5- 5 Fluidic passage of draining of the WBC bath
DiluentV01
WBC
bath
V03
lyse
V04
V02
10mL
SR32.5mL
(SR1)
Figure 5- 6 Fluidic passage of dispensing of the WBC bath
-
7/24/2019 BC-3600&3300 Service Manual V1.0
64/242
Fluidic System
5-10
Rinse
Waste
V09Isolation
chamber
WBC
bath
V10
V16
V14
Volumetricboard
V05 V07
V18
Vacuum
chamber
Pressure
Atmosp
heric
pressure
Figure 5- 7Fluidic passages of washing WBC back bath and volumetric tube
Aspirating sample: the sampling syringe aspirates 17ul of sample at 6.5~8.5s;
and 800ul of diluent is added to the RBC bath after it is drained;
Ascending to home position: the sample probe ascends to home position at
9.3~11.6s, and its exterior wall is washed; by now, the sampling procedure isfinished.
Other actions finished during the sampling procedure:
After sample dispensing of the WBC bath finishes (7s) and the fluid level
stabilizes, the HGB voltage and WBC aperture voltage (the constant-current
source is opened at 10.5s in advance) is read at 10.5s.
Sample Dispensing, Dilution and Mixing (11.5~30.8s)
The actions are:
Dispensing sample for the first time: the sample probe removes 6ul of sample in
the probe wipe at 11.5~12.3s.
Dispensing sample for the second time: the sample probe moves to the top of
WBC bath and descends into the bath; the diluent syringe dispenses 1150 ul of
diluent into the WBC bath, then the sample probe dispensed 9 ul of sample and
swings; when sample dispensing finishes, the diluent syringe dispenses 1200ul
of diluent into the WBC bath again; Valve 10 opens, and blows one small bubble
and 3 big bubbles into the WBC bath in turn. By now, the first dilution of sample
finishes.
The sample probe ascends with its exterior and interior walls washed: when the
second sample dispensing finishes, the sample probe ascends to home position
with its exterior and interior walls washed; besides, 2700ul of diluent isdispensed into the RBC bath.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
65/242
Fluidic System
5-11
Second aspiration in the WBC bath: the sample probe descends to the WBC
bath and aspirates 34 ul of fluid at 19.9~21s.Valve 17 opens, and the RBC bath
is drained.
The sample probe ascends: when the second aspiration finishes, the sample
probe ascends to home position with its exterior wall washed; besides, 1800 ul
of diluent is added to the RBC bath at 24.5~25.5s.
Adding sample to RBC bath, and lyse to WBC bath: the sample probe moves to
the RBC bath and descends into the bath; the lyse syringe dispenses 350ul of
lyse into the WBC bath, the fluidic passage is indicated in Figure 5- 8; the diluent
syringe discharges 800 ul of fluid (mixture of first dilution sample and diluent)
through the sample probe. Valve 10 opens at 26.65~27.25s, the pressure
chamber blows bubbles into the WBC bath; then a bubble that can last for 0.05s
is blowed into the WBC bath, and 3 bubbles are blowed into the RBC bath in
turn; Finally, the sample probe restores to the original position with its exterior
wall washed.
WBC
bath
V03
Lyse
2.5mL
(SR1)
Figure 5- 8 Fluidic passage of lyse dispensing of the WBC bath
Other actions finished during the sample dispensing, dilution and mixing
procedure:
Build the vacuum needed for analysis, total time consumed is 7.9s;
Build the pressure needed for bubble generation;
RBC aperture voltage inspection: same as WBC aperture voltage inspection;
The sample compartment door opens at 12s to prepare for the next analysis
cycle.
Analysis
The actions are:
Measuring volume of volumetric tube: valve 7 opens after WBC bubbling
finishes for a while and valve 8 opens after RBC bubbling finishes for a while,
the fluidic passage is indicated in Figure 5- 9; the closing of valve 7 and valve 8
is finished by the ending cleaning cycle started when the WBC and RBC downphotocouplers are triggered.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
66/242
Fluidic System
5-12
RBC
bath
Volumetric
board
Rinse
Waste
V06 V08
V15
WBC
bath
V14
Volumetric
board
V05 V07
V18
Vacuum
chamber
Figure 5- 9 Fluidic passages of WBC and RBC analysis
The photocoupler of volumetric board is opened at 25.7s.
Cleaning when analysis ends
The actions are:
The ending cleaning sequence is an independent sequence, it is started by the
software automatically base on the information obtained during the analysis;
Fluid discharging of RBC and WBC baths: open valve 17, the RBC bath starts to
discharge fluid; open valve 16, the WBC bath starts to discharge fluid;
Fluid dispensing of RBC and WBC baths: 1400 ul of fluid is dispensed into the
RBC and WBC bath respectively;
Washing RBC and WBC back baths: when valve 8 and valve 15 open at the
same time, the RBC back bath will be washed by rinse under the effect ofvacuum; when valve 7 and valve 14 open at the same time, the WBC back bath
will be washed by rinse under the effect of vacuum.
Draining the vacuum chamber: open valve 9 to drain the vacuum chamber.
Resetting of the diluent syringe and sample probe assembly: the diluent syringe
aspirates 2100 ul of fluid, then valve 1 opens, the syringe returns to home
position and dispenses fluid to the RBC bath at the same time; the sample probe
assembly returns to home position.
The sample probe aspirates air bar and returns to home position: the sample
probe moves to the piercing position and withdraws for 2ul and then returns to
home position.
Zapping: zapping starts when fluid dispensing of WBC bath completes for a
while.
Calculating the volume of rinse consumed: the software reads from the
command to calculate the volume of rinse consumed.
Detecting rinse status: sensor in the analyzer detects if the rinse container is
empty; when alarming for no rinse, the residual volume of rinse must be able to
support the on-going analysis.
The major difference between the whole blood analysis sequences of open vial models and
closed tube models lies in the sampling procedure, the open vial sequence involves nopiercing action, but its sampling time point and the cleaning action after sampling are the same
-
7/24/2019 BC-3600&3300 Service Manual V1.0
67/242
Fluidic System
5-13
with those of the closed tube sequence, therefore the whole blood analysis sequences of open
vial models will not introduced in detail.
5.1.2 Predilute Measurement Sequence
The predilute analysis cycle includes 3 function sequences (2 for closed tube models): diluent
dispensing sequence, cleaning sequence (open vial models) and analyzing sequence.
The predilute measurement procedure: (1)The sample probe dispense 380ul of diluent to a
centrifugal tube and 20ul of sample in the capillary tube is added in to the centrifugal tube, mix
them to form the diluted sample. (2)The sample probe aspirate 146.34ul of sample, and diluent
is added into the WBC bath to form the first dilution sample for WBC and HGB measurement.
(3)The sample probe ascends and its exterior and interior walls are cleaned. (4)The sample
probe descends to the WBC bath and aspirate 37.39ul of sample again, diluent is added and
the second dilution sample is dispensed into the RBC bath, and mix them for RBC/PLT
counting. (5)Add lyse into the WBC bath, and mix thoroughly; (6)Analysis. (7)When the
analysis completes, clean and zap the WBC/RBC bath, release the pressure in the vacuum
chamber.
You can see from the above procedures that the analysis cycles of predilute mode and whole
blood mode are only different in the operations of sampling and dispensing into WBC bath,
therefore this section will only introduce the different parts.
See section 5.3.2 for the dilution procedure of predilute mode.
5.5.2.1 Dilution dispensing sequence
The procedure is indicated in Figure 5- 10:
a) closed tube models b) open vial models
Figure 5- 10 Diluent dispensing flowchart
5.5.2.2 Cleaning sequence after diluent dispensing
Introduction:
-
7/24/2019 BC-3600&3300 Service Manual V1.0
68/242
Fluidic System
5-14
1. This sequence applies to open vial models only.
2. The sequence cleans the exterior wall of the sample probe, initializes diluent syringe
by command INIT; the sampling syringe withdraws for 2ul to create a separating air
bar, the system restores to "Ready" status.
5.5.2.3 Predilute Measurement Sequence
Introduction:
Sampling:
1. The diluent syringe withdraws for 100ul through the probe wipe tube;
2. WBC/RBC bath is zapped for 3s;
3. The diluent syringe aspirates 146.34ul of sample;
4. When aspirating finishes, valve 01 closes 0.1s after valve 04 closes.
Sample dispensing of WBC bath:
1. The diluent syringe pushes sample in the sampling tube into the WBC bath to form
the first dilution sample;
2. Sample volume of the second aspiration is 37.39ul;
3. The predilute measurement takes 1.5s more than the whole blood measurement due
to the zapping action added before analysis;
4. The other parts of sequence are the same as those of the whole blood measurement
sequence, only the fluid volumes are different in some steps.
5. The major difference between the closed tube and open vial models is the position of
the sample probe under the "Ready" status.
5.1.3 Startup/Shutdown Sequence
5.5.3.1 Startup sequence
There are three types of startup sequence: 1 startup after normal shutdown; 2 startup after
abnormal shutdown; 3 startup after performing "Prepare To Ship" procedure.
Startup after normal shutdown
This sequence includes the troubleshooting sequence 1, fluidics initializing sequence and
startup sequence after normal shutdown.
1. Troubleshooting sequence 1, refer to the corresponding section.
2. Fluidics initializing sequence, refer to the corresponding section.
3. Startup after normal shutdown
-
7/24/2019 BC-3600&3300 Service Manual V1.0
69/242
Fluidic System
5-15
Figure 5- 11Startup procedure after normal shutdown
Introduction:
1 There are zapping and flushing operations at the beginning to avoid clogging after
long term of probe cleanser soak.
2 0.5ml of lyse is discharged to make sure the tube mouth is filled with reagent.
3 Push back 200ul of diluent to the cleanser tube to wash off the cleanser at the
closed end of valve 12.
4 Empty RBC charging line to avoid crystallizing of the cleanser spills on the wall of
the bath which may influence the PLT background.
5 Empty the sample probe and dispense fluid: after emptying the sampling tube,
-
7/24/2019 BC-3600&3300 Service Manual V1.0
70/242
Fluidic System
5-16
dispense fluid at a higher speed so that the bubbles in the sampling tube can be
removed effectively.
6 The sample probe dispenses fluid in the WBC bath: removing all the bubbles in
the sampling tube.
7 Cleaning the two baths: cleanser volumes of WBC bath 4ml, 4.6ml, 5.5ml;cleanser volumes of RBC bath 4ml, 4ml, 5ml, generating bubbles during the
cleaning process to strengthen the cleaning effect.
8 Emptying WBC, RBC baths and the back baths, switching among valve 16 ,17, 7
and 8 to strengthen the cleaning effect.
Startup after abnormal shutdown
This sequence includes the troubleshooting sequence 1, fluidics initializing sequence and
startup sequence after normal shutdown.
1. Troubleshooting sequence 1, refer to the corresponding section.
2. Fluidics initializing sequence, refer to the corresponding section.
3. Startup after normal shutdown
Figure 5- 12 Startup procedure after abnormal shutdown
Introduction:
1 When starting up the analyzer after abnormal shutdown, there may be some
-
7/24/2019 BC-3600&3300 Service Manual V1.0
71/242
Fluidic System
5-17
residual sample in the fluidic components, so the exterior and interior walls of
the sample probe must be cleaned repeatedly by fluid discharging of the
sample probe and probe wipe.
2 0.5ml of lyse is discharged to make sure the tube mouth is filled with reagent.
3 Push back 200ul of diluent to the cleanser tube to wash off the cleanser at theclosed end of valve 12.
4 Remove the bubbles in the sampling tube through fluid discharging for multiple
times.
5 Clean the two baths repeatedly to remove the possible residual sample.
6 Zapping for 4s to avoid clogging.
7 Wash tubes of the back baths twice to remove the residual sample in the tubes
or volumetric tubes.
Startup after performing "Prepare To Ship" procedure
This sequence includes the troubleshooting sequence 1, fluidics initializing sequence and
overall reagent priming sequence.
1. Troubleshooting sequence 1, refer to the corresponding section.
2. Fluidics initializing sequence, refer to the corresponding section.
3. See section 5.5.3.5 for the overall reagent priming sequence.
5.5.3.2 Shutdown sequence
The probe cleanser soak sequence (general and intensified, refer to 5.5.4.7 and 5.5.4.8 for
details), intensified shutdown unclogging sequence (refer to 5.5.4.2 for details) and shutting
down sequence will be performed in turn during the shutdown process.
As per the different probe cleanser soak sequences, the shutdown sequence can be classified
into two kinds: 1 shutdown after general probe cleanser soak, 2 shutdown after intensified
probe cleanser soak.
The operations of the shutting down sequence are: washing the exterior wall of the sample
probe and soaking the probe in the WBC bath.
5.1.4 Cleaning/Maintenance Sequence
5.5.4.1 Unclogging sequence
Introduction:
1. Drain the two baths.
2. Valve 9 opens to connect the pressure chamber and vacuum chamber and build
pressure.
3. Flush the aperture for 4s and drain the fluid.
4. Add 3.5ml of fluid in the two baths respectively.5. Zap the two baths for 4s respectively.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
72/242
Fluidic System
5-18
6. Wash the tubes of the back baths. Empty the vacuum chamber and release vacuum.
5.5.4.2 Zapping sequence
Introduction:
1. Build vacuum in the vacuum chamber, zap the two baths for 4s.
2. Drain the two baths and add 3.5ml of fluid in the two baths respectively.
3. Wash the tubes of the back baths. Empty the vacuum chamber and release vacuum.
5.5.4.3 Flushing sequence
Introduction:
1 The flushing sequence is almost the same as the unclogging sequence, only it involves no
zapping operation.
5.5.4.4 Draining sequence of the baths
Introduction:
1 Drain the two baths for 3s.
5.5.4.5 Priming sequence of the baths
Introduction:
1. Drain the two baths and add 3.5ml of fluid in the two baths respectively.
2. The function is activated after the baths are drained. The baths may need to be
serviced during the draining process, the tubes of the back baths need to be washed.
3. Drain the vacuum chamber and release vacuum.
5.5.4.6 Cleaning sequence of the baths
Introduction:
1. 4.5ml of fluid is added to the two baths respectively, and bubbles are generated to
strengthen the cleaning effect.
2. Drain the two baths and add 3.5ml of fluid into the two baths respectively, open valve
18 and the fluidic pump for a while to release pressure in the pressure chamber.
5.5.4.7 General probe cleanser soak sequence
This sequence includes soaking sequence and the cleaning sequence after soaking.
Soaking procedure:
-
7/24/2019 BC-3600&3300 Service Manual V1.0
73/242
Fluidic System
5-19
Figure 5-13 Soaking procedure
Introduction: the soaking process consists of three steps. Step 1: soak the WBC aperture
and interior wall of the sample probe with 100% probe cleanser. Step 2: soak the WBC and
RBC front/back baths with 30.77% and 38.46% probe cleanser respectively. Step 3: soak the
WBC and RBC front/back baths with 26.32% and 21.05% probe cleanser respectively. The
operations are:
1. The sampling syringe withdraws for 2ul to form a separating air bar.
2. The sample probe aspirates 1.8ml of probe cleanser, the two baths are drained,
the sample probe ascends with its exterior wall washed.
3. The sample probe dispenses 1ml of probe cleanser into the WBC bath to form
100% probe cleanser.
4. The sample probe dispenses 2.6ml of mixture of probe cleanser and diluent into
the RBC bath to form the high concentration compound (30.77%) for the first
soak.
5. The sample probe dispenses 1.6ml of diluent into the WBC bath to form the highconcentration compound (38.46%) for the first soak.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
74/242
Fluidic System
5-20
6. After soaking for about 2 minutes, dispenses 1.2ml of diluent to the two baths to
dilute the probe cleanser, which will be used for the long term soaking of the front
and back baths (WBC:26.32%; RBC:21.05%).
7. Build vacuum and aspirate fluid: aspirate for 6s from the WBC bath, the volume
aspirated is about 232ul; aspirate for 8s from the RBC bath, the volume aspiratedis about 236ul; When the aspiration finishes, release the vacuum in the vacuum
chamber.
8. Meanwhile, the sample probe aspirates 50ul of fluid from the RBC bath for
soaking of the interior wall of sample probe and the tube.
9. Soak the tubes for 2 minutes and then start the cleaning procedure.
Cleaning procedure:
-
7/24/2019 BC-3600&3300 Service Manual V1.0
75/242
Fluidic System
5-21
Figure 5-14Cleaning procedure
Introduction: the front and back baths, sampling tube, waste valve shall be cleaned to remove
residual probe cleanser.
1. Add probe cleanser to the WBC bath; when draining the WBC bath, switch on/off
valve 16 for multiple times to soak the waste valve with probe cleanser (for about
5s) to remove residual probe cleanser.
2. Add probe cleanser to the RBC bath; when draining the RBC bath, switch on/off
valve 17 for multiple times to soak the waste valve with probe cleanser (for about
5s) to remove residual probe cleanser.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
76/242
Fluidic System
5-22
3. The sample probe dispenses 7.8ml of probe cleanser to the RBC bath to clean
the sampling tube and remove bubbles in the tubing.
4. Discharge fluid to the probe wipe twice, meanwhile the sample probe ascends to
avoid residual probe cleanser on the exterior wall of the sample probe or in the
probe wipe.5. Build vacuum, wash tube of the back bath, switch on/off valve 7 and 8 for
multiple times.
6. Empty the vacuum chamber, switch on/off valve 18 for multiple times.
7. Drain the baths and build pressure to 40kpa.
8. Flushing the apertures.
9. Drain the two baths and add 3.5ml of fluid into the baths respectively, zap the
baths for 4s and build vacuum to -24kp.
10. Wash the back baths, switch on/off valve 7 and 8 for multiple times to strengthen
the cleaning effect of the valves and remove residual probe cleanser.
11. Empty the vacuum, switch on/off valve 18 for multiple times to strengthen the
cleaning effect of the valves and remove residual probe cleanser.
12. Release the pressure, the analyzer restores to "Ready" status.
5.5.4.8 Intensified probe cleanser soak sequence
Soaking procedure:
-
7/24/2019 BC-3600&3300 Service Manual V1.0
77/242
Fluidic System
5-23
Figure 5-15 Soaking procedure
Introduction: the intensified probe cleanser soak procedure is almost the same as the generalprobe cleanser soak procedure, the only difference lies in the dilution ratio of probe cleanser.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
78/242
Fluidic System
5-24
Cleaning procedure:
The cleaning operations of the intensified and general probe cleanser soak procedure
are almost the same, the only difference lies in the occurrence time of the operation.
5.5.4.9 Draining sequence of the tubing
Procedure:
Figure 5- 16Draining procedure of the tubing
Introduction:
1. Build vacuum in the vacuum chamber to drain the back bath and rinse
tubes.
2. Drain the RBC bath, the diluent syringe withdraws for 1.3ml to drain the
charging line of RBC bath.
3. The lyse syringe aspirates and discharges lyse to the WBC bath for
multiple times to drain the lyse tube.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
79/242
Fluidic System
5-25
4. The diluent syringe aspirates and discharges diluent to the WBC bath for
multiple times to drain the diluent tube.
5. The sampling syringe aspirates 80ul of fluid.
6. The sample probe goes inside the WBC bath to drain the sampling tube.
7. Drain the charging line and the waste tube of the probe wipe.
8. Push the 80ul of fluid to the syringe inlet.
9. After build vacuum in the vacuum chamber, open the rinse tubes and empty
tubes of the back baths.
10. Empty the vacuum chamber and build vacuum at the same time, open V5
and V6 of the volumetric tube and dry the volumetric tube.
11. Release vacuum and pressure, the syringes return to home position, the
sample probe returns to the sampling position.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
80/242
-
7/24/2019 BC-3600&3300 Service Manual V1.0
81/242
6-1
6 Hardware System
6.1 Introduction
The hardware system consists of 8 boards, which are power board, data board, drive board,
indicator board, touch screen control board, volumetric boars, sample compartment connector
board, and USB interface board; the system also includes the drive and detect components
that require power supply (such as motors, valves, pumps, sensors, display, and filter of input
power) and the wires connecting boards and components.
6.1.1 Function DiagramSee Figure 6-1 for the function diagram of the hardware system.
Figure 6-1Function diagram of the hardware system
The hardware system consists of f ive modules, which are system power, data stream channel,
central system, drive components and peripherals. The function of each module is as follows:
1. System power: providing power of various specifications to the boards, components
and devices.
2. Data stream channel: extracting, adjusting, amplifying and collecting and
preprocessing of signals.
3. Central control system: data collection and processing, result display, sample
storage, etc. Besides, the central control system controls and responds to all
peripherals.
4. Drive/detect components: controlling valves, pumps and motors, monitoringphotocouplers and other major parameters, collecting detection information and
-
7/24/2019 BC-3600&3300 Service Manual V1.0
82/242
Hardware System
6-2
sending alarms.
5. Peripherals and interfaces: display/touchscreen, recorder, USB interfaces (for printer,
keyboard and barcode scanner), Ethernet interface and LIS serial port. Besides,
peripherals includes status indicator, key-presses and aspirate key input.
6.1.2 Electric Connection Diagram
See Figure 6-2 for the electric connection diagram of the hardware system.
-
7/24/2019 BC-3600&3300 Service Manual V1.0
83/242
Hardware System
6-3
Figure 6-2 Electric connection diagram of the hardware system
-
7/24/2019 BC-3600&3300 Service Manual V1.0
84/242
Hardware System
6-4
6.1.3 Troubleshooting the System
The errors of the hardware system include board errors, wire errors and component errors.
Generally ,the troubleshooting methods of the three types of errors can be found in the following
sections; however, if the power supply of the hardware system cannot be ensured (the analyzer
cannot be electrified, or self-shields right after power on), the errors shall be fixed on the system
level.
1 Troubleshooting procedure when the analyzer cannot be electrified (the indicators of the drive
board and the data board are both off)
(1) Check if the AC input wire outside the analyzer is firmly connected, if not, re-plug the wire.
(2) Power off the analyzer, check if the wire of the power input filter is firmly connected, if not,
re-plug the wire.
(3) If step (1) and (2) cannot remove the error, replace the power input filter (see Figure 6-3);
if the error persists, do step (4).
Figure 6-3 Power input filter
(4) Replace the power board or assembly.
2 Troubleshooting procedure when the analyzer self-shields right after power on
(1) First, unplug the wires of the drive board and data board (see Figure 6-4), see if the power
can be started; if not, check if the wire is damaged; if so, replace the wire; otherwise replace
the power board or assembly. If the power can be started normally, the problem of power
assembly can be excluded.
Figure 6-4Sockets of drive board and power board
-
7/24/2019 BC-3600&3300 Service Manual V1.0
85/242
Hardware System
6-5
(2) After completing step (1), plug the two wires on of the drive board in turn to see if the power
can be started; if yes, the problem of the drive board can be excluded, the data board needs to be
replace; if no, unplug the wire of the volumetric board; if the power can be started, that means the
volumetric board or its wire is short-circuited, the volumetric board or its wire shall be replaced; if
not, that means the drive board is short-circuited, the drive board shall be replaced.
If the power can be started normally, troubleshoot the error per the following sections.
6.2 Data Board
6.2.1 Introduction
The data board consists of the analog and digital parts, the analog part filters, amplifies, outputs
and processes signals to obtain the signals that are suitable for A/D data conversion. The digitalpart processes, outputs, controls and transmits data. It is the core part of the data board or even
the entire hardware system.
Section 6.2 is the servicing and troubleshooting guideline of the data board.
6.2.2 Structure
Introduction of Functions
See Figure 6-5 for the structural diagram of the data board.
Figure 6-5Structural diagram of the data board
You can see from Figure 6-5, the data board consists of the analog circuit and digital circuit. The
analog circuit adjusts and amplifies the WBC, RBC, PLT and HGB signals to make sure the
signals are turn and suitable for A/D conversion; the A/D module is the interface of the analog
circuit and digital circuit, it collects the sensing signals and other monitoring signals, and converts
analog signals into digital signals; the digital circuit processes data, stores and outputs results.
Besides, the digital circuit takes on the control and communication tasks, it is the core part of the
data board or even the entire hardware system.
1. Analog module
-
7/24/2019 BC-3600&3300 Service Manual V1.0
86/242
Hardware System
6-6
Figure 6-6 Structural diagram of the analog module
The analog module of the data board consists of RBC/PLT channel, WBC channel, HGB
channel and A/D conversion part, its functions are:
Adjusting analog power
Adjusting volume signal
Data collection
2. Digital module
Figure 6-7 Structural diagram of the digital module
The structure of the digital module is "CPUFPGA", see Figure 6-7.The functions of the