BC-3600&3300 Service Manual V1.0

download BC-3600&3300 Service Manual V1.0

of 242

Transcript of BC-3600&3300 Service Manual V1.0

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    1/242

    BC-3600

    BC-3300

    Auto Hematology Analyzer

    Service Manual

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    2/242

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    3/242

    I

    Copyright

    2008-2011 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.

    For this Service Manual, the issued Date is 2011-02 (Version: 1.0).

    Intellectual Property Statement

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray)

    owns the intellectual property rights to this Mindray product and this manual. This manual may

    refer to information protected by copyrights or patents and does not convey any license under

    the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability

    arising out of any infringements of patents or other rights of third parties.

    Mindray intends to maintain the contents of this manual as confidential information. Disclosure

    of the information in this manual in any manner whatsoever without the written permission of

    Mindray is strictly forbidden.

    Release, amendment, reproduction, distribution, rent, adaptation and translation of this

    manual in any manner whatsoever without the written permission of Mindray is strictly

    forbidden.

    . , are the registered trademarks or trademarks owned by

    Mindray in China and other countries. All other trademarks that appear in this manual are

    used only for editorial purposes without the intention of improperly using them. They are the

    property of their respective owners.

    Responsibility on the Manufacturer Party

    Contents of this manual are subject to changes without prior notice.

    All information contained in this manual is believed to be correct. Mindray shall not be liable for

    errors contained herein nor for incidental or consequential damages in connection with the

    furnishing, performance, or use of this manual.

    Mindray is responsible for safety, reliability and performance of this product only in the

    condition that:

    all installation operations, expansions, changes, modifications and repairs of this

    product are conducted by Mindray authorized personnel;

    the electrical installation of the relevant room complies with the applicable national

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    4/242

    II

    and local requirements;

    the product is used in accordance with the instructions for use.

    This equipment must be operated by skilled/trained medicalprofessionals.

    It is important for the hospital or organization that employs this equipment

    to carry out a reasonable service/maintenanceplan. Neglect of this may

    result in machine breakdown or injury of human health.

    Be sure to operate the analyzer under the situation specified in this manual;

    otherwise, the analyzer will not work normally and the analysis results will

    be unreliable, which would damage the analyzer components and cause

    personal injury.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    5/242

    III

    Warranty

    THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,

    EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR

    FITNESS FOR ANY PARTICULAR PURPOSE.

    Exemptions

    Mindray's obligation or liability under this warranty does not include any transportation or other

    charges or liability for direct, indirect or consequential damages or delay resulting from the

    improper use or application of the product or the use of parts or accessories not approved by

    Mindray or repairs by people other than Mindray authorized personnel.

    This warranty shall not extend to:

    any Mindray product which has been subjected to misuse, negligence or accident;

    any Mindray product from which Mindray's original serial number tag or product

    identification markings have been altered or removed;

    any product of any other manufacturer.

    Safety, Reliability and PerformanceMindray is not responsible for the effects on safety, reliability and performance of BC-2800Vet,

    if:

    Assembly operations, extensions, re-adjusts, modifications or repairs are carried out

    by persons other than those authorized by Mindray.

    Personnel unauthorized by Mindray repairs or modifies the instrument.

    Return Policy

    Return Procedure

    In the event that it becomes necessary to return this product or part of this product to Mindray,

    the following procedure should be followed:

    1. Obtain return authorization: Contact the Mindray Service Department and obtain a

    Customer Service Authorization (Mindray) number. The Mindray number must appear

    on the outside of the shipping container. Returned shipments will not be accepted ifthe Mindray number is not clearly visible. Please provide the model number, serial

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    6/242

    IV

    number, and a brief description of the reason for return;

    2. Freight policy: The customer is responsible for freight charges when this product is

    shipped to Mindray for service (this includes customs charges);

    3. Return address: Please send the part(s) or equipment to the address offered by

    Customer Service department.

    Company Contact

    Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

    Address:Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,

    Nanshan, ShenZhen 518057, P.R.China,

    Phone: +86 755 26582479 26582888

    Fax: +86 755 26582934 26582500

    EC-Representative: Shanghai International Holding Corp. GmbH(Europe)

    Address: Eiffestrae 80, 20537 Hamburg Germany

    Phone: 0049-40-2513175

    Fax: 0049-40-255726

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    7/242

    1

    Table of Contents

    CopyrightIIntellectual Property StatementI

    Responsibility on the Manufacturer Party I

    WarrantyIII

    Exemptions III

    Return Policy III

    1 Using This Manual 1-1

    1.1. Scope 1-2

    1.2. Introduction 1-2

    1.3. Conventions Used in This Manual 1-3

    1.4. Special Terms Used in This Manual 1-3

    1.5. Symbols 1-4

    1.6. Definitions 1-12

    2 Analyzer Structure 2-1

    2.1 Product Name 2-1

    2.2 System Introduction 2-1

    2.3 System Composition 2-7

    2.4 Software Structure 2-82.5 LIS Connection 2-19

    3 Upgrade 3-1

    3.1 Software upgrade 3-1

    4 Mechanical System 4-1

    4.1 Front of the Analyzer 4-1

    4.2 Back of the Analyzer 4-3

    4.3 Left of the Analyzer 4-4

    4.4 Right of the Analyzer 4-6

    5 Fluidic System 51

    5.1 Fluidic Components and Functions 5-1

    5.2 Reagent Consumption 5-2

    5.3 Sample Dilution Flowchart 5-3

    5.4 Introduction of Fluidic Channels 5-5

    5.5 Introduction of Basic Sequence 5-6

    6 Hardware System 6-16.1 Introduction 6-1

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    8/242

    Table of Contents

    2

    6.2 Data Board 6-5

    6.3 Drive Board 6-18

    6.4 Power Board 6-25

    6.5 USB Interface Board 6-30

    6.6 Touchscreen Control Board 6-31

    6.7 Volumetric Board 6-32

    6.8 Sample Compartment Connection Board 6-34

    6.9 Indicator Board 6-35

    7 Servicing 7-1

    7.1 General 7-1

    7.2 Disassembling the Panels 7-2

    7.3 Replacing the Hardware parts and Cable connection 7-6

    7.4 Replacing the Power Supply Assembly 7-16

    7.5 Replacing the Display Assembly 7-17

    7.6 Replacing the Touch Screen 7-18

    7.7 Replacing the Recorder Assembly 7-19

    7.8 Replacing the valve, pumps, and other liquid system parts 7-20

    7.9 Replacing the counting bath assembly 7-29

    7.10 Replacing the syringe assembly 7-35

    7.11 Replacing the sample probe assembly 7-37

    7.12 Replacing the sample compartment assembly 7-47

    7.13 Replacing the start key 7-56

    7.14 Micro-Switch Assembly Replacement 7-57

    7.15 Fluidic Support-bracket Assembly 7-58

    7.16 Front Panel Detection Assembly 7-59

    7.17 Ambient Temperature Sensor Replacement 7-60

    7.18 Cap Assembly 7-61

    7.19 Gain Adjustment 7-62

    7.20 Motor Type, Position and Different Application List 7-64

    7.21 The Check List After Material Replacement 7-64

    8 Maintaining your analyzer 8-1

    8.1 General Guidelines 8-18.2 Operation guidance of Quality Control and Calibration 8-2

    9 Error Information 9-1

    9.1 Error Information and Code 9-1

    9.2 Error analysis and processing 9-2

    9.3 Error diagnosis assisted by abnormal histogram 9-12

    9.4 Error diagnosis assisted by abnormal pulse figure 9-21

    10 Appendices A-1

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    9/242

    1-1

    1 Using This Manual

    This chapter explains how to use this service manual. This manual states in detail the way toservice the BC-3600 and BC-3300 analyzers. Before servicing the BC-3600 and BC-3300,

    read and understand the manual carefully for servicing the equipment properly and for your

    safety.

    This manual is to be used in conjunction with the operator's manuals of BC-3600 and

    BC-3300.It does not contain information and procedures already covered in the operator's

    manuals of BC-3600 and BC-3300.

    Be sure to operate and service the analyzer strictly as instructed in this manual and the

    operator's manuals.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    10/242

    Using This Manual

    1-2

    1.1. Scope

    To use this manual effectively, you need the following capabilities:

    Comprehensive knowledge of circuit and fluidics;

    Comprehensive knowledge of reagents;

    Comprehensive knowledge of controls;

    Comprehensive knowledge of troubleshooting;

    Mastering the way to operate this analyzer;

    Using basic mechanical tools and understand related terminology;

    Using a digital voltmeter (DVM) and an oscilloscope;

    Reading pneumatic/hydraulic schematics and understand related terminology.

    1.2. Introduction

    This manual comprises 9 chapters and 6 appendices. Refer to the table below to find the

    information you need.

    If you want to See

    learn about the system and software structures of BC-3600

    and BC-3300

    Chapter 2 System Structure

    learn about the installation requirements and the way to

    upgrade the software of BC-3600 and BC-3300

    Chapter 3 System Installation

    and Software Upgrade

    learn about the mechanical system structure of BC-3600 and

    BC-3300

    Chapter 4 Mechanical System

    learn about the fluidic system, reagent dosage, channels and

    basic sequence of BC-3600 and BC-3300

    Chapter 5 Fluidic System

    learn about the hardware system and board composition,

    adjustment and test points and general troubleshooting

    methods of BC-3600 and BC-3300

    Chapter 6 Hardware System

    learn about how to service the BC-3600 and BC-3300 Chapter 7 Servicing

    learn about how to maintain the BC-3600 and BC-3300 Chapter 8 Maintenance

    learn about how to troubleshoot the common errors of

    BC-3600 and BC-3300

    Chapter 9 Troubleshooting

    learn about the main spare parts of BC-3600 and BC-3300 Appendix A List of Spare parts

    learn about the main wearing parts of BC-3600 and

    BC-3300

    Appendix B List of Wearing

    Parts

    learn about the circuit diagram of BC-3600 and BC-3300 Appendix C Circuit Diagramlearn about the schematic diagram of the fluidic system of Appendix D Fluidic Diagram

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    11/242

    Using This Manual

    1-3

    BC-3600 and BC-3300

    learn about the function of each valve and pump of BC-3600

    and BC-3300

    Appendix E Pump and Valve

    Function Diagram

    learn about the tubing connection of BC-3600 and BC-3300 Appendix F Tubing

    1.3. Conventions Used in This Manual

    This manual uses certain typographical conventions to clarify meaning in the text:

    Format Meaning

    []all capital letters enclosed in [ ] indicate a key name (either on

    the pop-up keyboard or the external keyboard)

    letters included in " " indicate text you can find on the screen

    of BC-3600 and BC-3300

    italic letters indicate titles of the chapters that are referred

    to

    "Menu"the " " button on the screen.

    All illustrations in this manual are provided as examples only. They may not necessarily reflect

    your analyzer setup or data displayed.

    1.4. Special Terms Used in This Manual

    When you read It means

    Clickto press the desired item lightly with your finger; or to

    left-CLICK it with the mouse.

    ENTER

    to CLICKthe desired edit box and use the external keyboard

    or the pop-up keyboard to enter the desired characters or

    digits; or to scan the number by using the bar-code scanner.

    DELETE

    to move the cursor to the character or digit that you want to

    delete by clicking the left button of the mouse or using

    [][][Home][End], and then delete the character after the

    cursor by pressing [Del], or delete the character before the

    cursor by pressing [BackSpace] ([] on the upper right part of

    the soft keyboard).

    SELECT from

    pull-down list

    (for pull-down list)

    to CLICKthe down arrow button of the desired box to display

    the pull-down list, (and DRAG SCROLL BAR) to browse and

    then CLICK the desired item; or to press the keys

    ([][][PageUp][PageDown]) to browse the current list and

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    12/242

    Using This Manual

    1-4

    press [ENTER] to select the desired item.

    1.5. Symbols

    You will find the following symbols in this manual.

    When you see Then

    read the statement below the symbol. The statement is

    alerting you to an operating hazard that can cause

    personnel injury.

    read the statement below the symbol. The statement is

    alerting you to a possibility of analyzer damage or unreliable

    analysis results.

    read the statement below the symbol. The statement is

    alerting you to information that requires your attention.

    read the statement below the symbol. The statement is

    alerting you to a potentially biohazardous condition.

    You may find the following symbols on the analyzer, reagents, controls or calibrators.

    When you see It means

    CAUTION, CONSULT ACCOMPANYING

    DOCUMENTS.

    BIOLOGICAL RISK

    HIGH VOLTAGE

    EXERCISE CAUTION WHEN WORKING

    AROUND TO AVIOD PRICKING

    PROTECTIVE EARTH (GROUND)

    ALTERNATING CURRENT

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    13/242

    Using This Manual

    1-5

    FOR IN VITRO DIAGNOSTIC USE

    BATCH CODE

    USE BY (YYYY-MM-DD)

    SERIAL NUMBER

    MEASUREMENT AUTHORIZATION

    SYMBOL

    DATE OF MANUFACTURE

    MANUFACTURER

    THIS ELECTRO-INFORMATIONAL

    PRODUCT CONTAINS CERTAIN

    POISONOUS OR HARMFUL SUBSTANCE.ITS ENVIRONMENTAL LIFT TIME IS 20

    YEARS. IT CAN BE USED WHTHIN 20

    YEARS, AFTER THAT IT SHALL BE

    RECYCLED.

    Be sure to observe the following precautions for the safety of patients and operators when you

    are servicing the analyzer.

    It is important for the hospital or organization that employs this equipment to

    carry out a reasonable service/maintenance plan. Neglect of this may result

    in machine breakdown or harm to human health.

    Never use combustible gas (e.g. anesthetic) or combustible liquid (e.g.

    ethanol) around the analyzer. Otherwise, the risk of explosion may exist.

    When servicing the analyzer, be sure to turn off the power. Servicing the

    analyzer when it is on may bring risk of electric shock or damage to

    electronic components.

    Connect the analyzer to a socket having sole fuse and protective switch. Do

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    14/242

    Using This Manual

    1-6

    not use the same fuse and protective switch with other equipment (e.g. life

    supporting equipment). Otherwise, the equipment failure, over current or

    impulse current that occurs at the startup moment may lead to tripping.

    To prevent personal injury during maintenance, keep your clothes, hairs and

    hands from the moving parts, such as sample probe, clipper and piercer.

    Possible mechanical movement of the warned position may lead to personal

    injury during the normal operation, removal and maintenance.

    Be sure to dispose of reagents, waste, samples, consumables, etc.

    according to government regulations.

    The reagents are irritating to eyes, skin and mucosa. Wear proper personal

    protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory

    procedures when handling them in the laboratory.

    If the reagents accidentally spill on your skin, wash them off with plenty of

    water and if necessary, go see a doctor; if the reagents accidentally spill into

    your eyes, wash them off with plenty of water and immediately go see a

    doctor.

    Improper maintenance may damage the analyzer. Maintain the analyzer

    strictly as instructed by the service manual and inspect the analyzer

    carefully after the maintenance.

    For problems not mentioned in the service manual, contact Mindray

    customer service department for maintenance advice.

    To prevent personal injury or damage to equipment components, remove

    metal jewelry before maintaining or servicing electronic components of the

    equipment.

    Electrostatic discharge may damage electronic components. If there is a

    possibility of ESD damage with a procedure, then do that procedure at an

    ESD workstation, or wear an antistatic wrist strap.

    When the front cover is open, be sure to close the sample compartment door,

    or the sample compartment assembly or the front cover may be damaged.

    When transporting the analyzer, the screw at its bottom may scratch your

    hands, please exercise caution or wear gloves.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    15/242

    Using This Manual

    1-7

    This equipment must be operated by skilled/trained medical professionals.

    Samples, controls, calibrators and waste are potentially infectious. Wear

    proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow

    safe laboratory procedures when handling them in the laboratory.

    All the analyzer components and surfaces are potentially infectious. Take

    proper protective measures for operation or maintenance.

    The sample probe tip is sharp and may contain biohazardous materials.Exercise caution to avoid contact with the probe when working around it.

    Figure 1-1

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    16/242

    Using This Manual

    1-8

    Warning, potential biohazardous risk.

    Make sure the analyzer is properly grounded.

    To avoid electric shock, disconnect power cord prior to removing or replacing fuse.

    Use fuse only with the type and rating specified.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    17/242

    Using This Manual

    1-9

    Figure 1-2

    Warning, the sample probe is sharp and may contain biohazardous substance. Exercise

    caution to avoid contact with the probe when working around it.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    18/242

    Using This Manual

    1-10

    Figure 1-3

    Warning, make sure the protective cover is properly installed before operation.

    Warning

    Do not put your hands close to the sampling assembly when the analyzer is running.

    Warning, the sample probe is sharp and may contain biohazardous substance.

    Exercise caution to avoid contact with the probe when working around it.

    Warning, the sample probe is sharp and may contain biohazardous substance. Exercise

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    19/242

    Using This Manual

    1-11

    caution to avoid contact with the probe when working around it.

    Figure 1-4

    Warning, risk of electric shock.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    20/242

    Using This Manual

    1-12

    1.6. Definitions

    There are two sampling mode: open vial mode and closed tube mode.

    Open vial mode

    The operator presents the tube or open vial evacuated collection tube containing blood sample

    to the sample probe, and press the "Aspirate" key to collect sample.

    The analyzer of which the sampling mode is open vial is called open vial analyzer.

    Closed tube mode

    The operator puts the capped tube into the sample compartment, the analyzer collects sample

    by piercing through the tube cap.

    The analyzer of which the sampling mode is closed tube is called closed tube analyzer.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    21/242

    2-1

    2 Analyzer Structure

    2.1 Product Name

    Name: Auto Hematology Analyzer.

    Models: BC-3600, BC-3300.

    2.2 System Introduction

    BC-3600 and BC-3300 are in-vitro diagnosis equipments with the functions of blood cell

    counting, WBC 3-differential and measurement of HGB concentration under proper working

    environment. The purpose of the analyzers is to identify the normal patient, with all normal

    system-generated parameters, and to flag or identify patient results that require additional

    studies.

    Scope: the products can be applied for clinical diagnosing reference and scientific

    research.

    Environment of use: the analyzers shall be used in medical laboratories with

    standardized management; they can not be used as portable equipment.

    Operator: the operators must be skilled/trained medical professionals.

    2.2.1 System Configuration

    The whole system includes the analyzer, accessories and reagents; users can choose to

    configure data management software, scanner, printer or PC.

    2.2.2 Electrical Parameters

    Table2-1 Power supply of the analyzer

    Voltage (100V-240V)10%

    Input Power 180 VA

    Frequency 50/601Hz

    Only install fuses of specified specification on the analyzer.

    Specification of the fuse: 250V T3.15AH

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    22/242

    Analyzer Structure

    2-2

    2.2.3 Dimension and Weight

    Figure 2-1Dimensional sketch map

    Table2-2Dimension and weight

    Width (mm) Depth (mm) Height (mm) Weight (Kg)

    395 450 445 (without foot) 28

    2.2.4 ThroughputThe throughput of open vial analyzers under whole blood mode is no less than 60 samples per

    hour; the time needed for the analysis of one sample is no more than 60s (the sample dilution

    time is not included for predilute mode).

    The throughput of closed tube analyzers under whole blood mode is no less than 60 samples

    per hour; the time needed for the analysis of one sample is no more than 60s (the sample

    dilution time is not included for predilute mode).

    2.2.5 ParametersThe analyzer determines 21 parameters (WBC, RBC, PLT, HGB, etc.) and 3 histograms

    (WBC, RBC and PLT) of blood samples.

    Table 2-3Table of parameters

    English name Abbreviation

    White Blood Cell count WBC

    Lymphocyte number Lymph#

    Mid-sized Cell number Mid#

    Granulocyte number Gran#

    Lymphocyte percentage Lymph%

    Width

    Height

    Depth

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    23/242

    Analyzer Structure

    2-3

    Mid-sized Cell percentage Mid%

    Granulocyte percentage Gran%

    Red Blood Cell count RBC

    Hemoglobin concentration HGB

    Mean Corpuscular Volume MCVMean Corpuscular Hemoglobin MCH

    Mean Corpuscular Hemoglobin Concentration MCHC

    Red Blood Cell Distribution Width Coefficient of

    Variation

    RDW-CV

    Red Blood Cell Distribution Width Standard

    Deviation

    RDW-SD

    Hematocrit HCT

    Platelet count PLT

    Mean Platelet Volume MPV

    Platelet Distribution Width PDW

    Plateletcrit PCT

    Platelet Larger Cell Ratio P-LCR

    Platelet Larger Cell Count P-LCC

    Table2-4Table of histograms

    White Blood Cell Histogram WBC Histogram

    Red Blood Cell Histogram RBC Histogram

    Platelet Cell Histogram PLT Histogram

    2.2.6 Performance Specifications

    2.2.6.1 Background or Blank Counting

    Background counting is performed by the analyzer automatically when it is started up. The

    following requirements must be met.

    Blanking counting: test the diluent for 3 consecutive times, and take the highest result. The

    following requirements must be met.

    Table 2-5 Requirements on background or blank counting

    Parameter Requirement

    WBC 0.3 109/ L

    RBC 0.031012

    / L

    HGB 1 g / L

    HCT 0.5 %

    PLT 5109/ L

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    24/242

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    25/242

    Analyzer Structure

    2-5

    RBC 08.001012

    /L 0.051012

    /L or 5 0.051012

    /L or 5

    HGB 0280g/L 2g/L or 2 2g/L or 3

    01000109/L 1010

    9/L or 10 1010

    9/L or 10PLT

    10013000109/L 12 20

    HCT 060%4% (HCT value) or 6%

    (deviation)

    /

    Test suggestion:

    When testing RBC linearity range 7.008.001012

    /L, use sample of which the HCT 60%;

    When testing PLT linearity range 10013000109/L, use sample of which the RBC 6.

    2.2.6.4 Carryover

    The carryover impact of high concentration sample to low concentration sample.

    Test method: mix the high value sample (centrifuged high value control or specialized high

    value linearity control) within the range of Table 2-9, test the sample for 3 consecutive times,

    and then mix and test the low value sample (diluted low value control, dilution rate 1:10) within

    the range of Table 2-9 for 3 consecutive times, then calculate the carryover. The requirements

    in Table 2-10 must be met.

    Table 2-9 Requirements on high and low value controls of carryover calculation

    Parameter Unit High concentration Low concentration

    WBC 109

    /L > 15.0 < 3.0

    RBC 1012

    /L > 6.0 < 2.00

    HGB g/L > 200 < 40

    PLT 109/L > 300 < 100

    Table 2-10 Carryover requirements

    Parameter Carryover

    WBC 0.5

    RBC 0.5

    HGB 0.5

    PLT 1.0

    2.2.7 Functions

    2.2.7.1 Startup and Shutdown

    When starting up the analyzer, initialization will be performed to prepare the analyzer for

    counting. Different startup cleaning sequence will be performed base on the latest shutdown

    status to ensure the success of background counting. The analyzer provides the function to

    skip fluidics initialization to the users of service level and above.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    26/242

    Analyzer Structure

    2-6

    The shutdown procedure will be performed according to different shutdown status.

    2.2.7.2 Password and User Management

    In order to protect the security of product setup and data, you have to possess the password to

    be able to adjust the parameters and perform the extra functions.

    2.2.7.3 Screen Navigation

    The analyzer provides screen navigation function so that users can easily go to their target

    screen.

    2.2.7.4 Status Indication

    The analyzer/system status indication function is provided on the analyzer screen or by

    configuring LED in the hardware system.

    2.2.7.5 Setup

    Users can customize the settings of the analyzer on the software screen.

    1. The configuration files are saved in the analyzer.2. The analyzer must allow users to modify certain configurations. E.g.: Users can set

    up date/time, reference range, parameter unit, communication-related parameters,

    print format, etc.

    3. For the FDA model, users can set whether to display or print RUO parameters

    through the authorization protocol.

    4. For the FDA model, users can set "QC lockout", if this function is activated, sample

    analysis can be started only after the daily QC passes.

    2.2.7.6 Standby

    The analyzer will enter standby mode when no fluidic operation has been performed for a

    certain period of time. The analyzer will exit the standby mode when operation is triggered

    by users.

    2.2.7.7 Self Test

    The analyzer system provides self test and inspection function to ensure proper running.

    2.2.7.8 Sample Review

    After analyzing a sample, the analyzer saves the sample information and analysis result

    automatically. Operators can review all parameters and histograms of the saved samples.

    2.2.8 Sample Compartment

    The tube holder in sample compartment can be fitted with adapters of different specifications.

    The specifications of supported tubes are:

    Evacuated collection tube: diameter: 1013.5mm; height (with cap): 4085mm.

    Capillary collection tube: diameter: 1013.5mm; height (with cap): 4065mm.

    Centrifugal tubes.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    27/242

    Analyzer Structure

    2-7

    2.2.9 Operating, Storing and Running Environment

    Table 2-11 Temperature and Humidity Requirements

    Optimal operating

    environment

    Storage environmentRunning environment

    Temperature 1530 -1040 1535

    Humidity 30%85% 10%93% 30%85%

    Atmospheric

    pressure

    70kPa106kPa 50kPa106kPa70kPa106kPa

    2.3 System Composition

    2.3.1 Functional Composition

    The analyzer system has the following functions: reagent system, sample distribution, sample

    preparation, sample analysis, signal processing, parameter analysis, data management, status

    monitoring, schedule control, information processing, human-machine interface, power source,

    cleaning and maintenance. The relations of the functional modules are shown in Figure 2-2 .

    Under the control of the schedule control and information processing functional module, the

    other modules work together according to the designed procedure and requirement to realize

    the core task of the analyzer - sample analysis.

    Fluid SignalstreamReagentsAirflow

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    28/242

    Analyzer Structure

    2-8

    Figure 2-2 System function block diagram

    2.3.2 Reagents

    In order to ensure the accuracy of analysis results, you shall use the reagents, controls

    and calibrators specified by Shenzhen Mindray Bio-medical Electronics Co., Ltd. For

    details, refer to Table 2-12.

    Table 2-12 Applicable reagents, controls calibrators

    Item Name

    Diluent M-30D DILUENT

    Lyse M-30CFL LYSE

    RinseM-30R RINSE

    Probe cleanser PROBE CLEANSER

    Control B30 or BC-3D

    Calibrator S30 or SC-CAL PLUS

    2.4 Software Structure

    2.4.1 Menu Structure

    The top level structure of the system menu includes 8 functional options: analysis, review, QC,

    calibration, setup, service, logoff and shutdown. Operators can enter the screens to perform

    the functions of the analyzer through the system menu. The " " indicates there is sub-menu

    under the option. Refer to Figure 2-3 -- Figure 2-7 for the full menu structure.

    The menu structure of service access level:

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    29/242

    Analyzer Structure

    2-9

    Figure 2-3 Top level menu and QC sub-menu

    Figure 2-4 Top level menu and calibration sub-menu

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    30/242

    Analyzer Structure

    2-10

    Figure 2-5 Top level menu and setup sub-menu

    Figure 2-6 Top level menu and setup sub-menu(2)

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    31/242

    Analyzer Structure

    2-11

    Figure 2-7 Top level menu and service sub-menu

    2.4.2 Password

    The login service ID is "Service", password is "s 3600!", case sensitive and space required.

    With service access level, you can perform all functions that are open to common users and

    some other special functions:

    1Transfer factor calibration function:

    1Under whole blood mode, click "Menu">"Calibration">"Calibrator", finish the

    calibration successfully, exit and save the calibration factors.

    2Click "Menu">"Calibration">"Transfer factor" to enter the screen in Figure 2-8,

    run the sample used in the above step for 5 consecutive times under whole blood

    mode, check if CV(%) meets the requirements in Table 2-13; if not, cancel a abnormal

    result and calculate again (results of at least 5 analyses required), then check if the

    requirements in Table 2-13 are met.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    32/242

    Analyzer Structure

    2-12

    Figure 2-8Analysis screen of transfer factor under whole blood mode

    Table 2-13 Whole blood CV range

    Param

    eter

    WBC RBC HGB MCV PLT

    CV

    value

    2.0% 1.5% 1.5% 0.5% 4.0%

    3Click the "Predilute" button to switch to the screen of predilute mode, as Figure 2-9

    shows;

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    33/242

    Analyzer Structure

    2-13

    Figure 2-9 Analysis screen of transfer factor under predilute mode

    4Click the Diluent icon, perform diluent dispensing for 7 consecutive times;

    5Add 140l of calibrator in to the diluent, mix thoroughly and then run the mixture for

    more than 5 times under the predilute mode. The CV(%) values must meet the

    requirements of Table 2-14, and the calibration factors are within the range [0.75,

    1.25].

    Table 2-14 Range of CV value under predilute mode

    Param

    eter

    WBC RBC HGB MCV PLT

    CV

    value

    4.0% 2.0% 2.0% 1.5% 8.0%

    2The gain of WBC and RBC can be adjusted, see Figure 2-10;

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    34/242

    Analyzer Structure

    2-14

    Figure 2-10 Gain setup screen

    3In the service setup screen, you can set up languages, software upgrade, volume of the

    volumetric tube, analyzer SN, sample probe positioning adjustment, background

    information monitoring (auto adjust of analysis time), rinse photocoupler calibration, etc.

    The following section will introduce the sample probe positioning adjustment and rinse

    photocoupler calibration functions.

    1) Sample probe positioning adjustment:

    Tool: the positioning calibration clamp of sample probe sampling position

    Procedure: open the cover as Figure 2-11 shows, open the sample compartment door,

    take out the tube holder adapter from the compartment, put the positioning calibration

    clamp to the sample compartment, and then close the compartment door.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    35/242

    Analyzer Structure

    2-15

    Figure 2-11 Sample compartment

    Click the "Aspirate pos." under the "X direction motor" as shown in Figure 2-12, and click

    "Start", then click the "Move Left" or "Move Right" button to move the sample probe to the

    top of the positioning hole of the fixture, check if the tip of the probe is right against the

    positioning hole, then click "Save" to save the setting.

    Figure 2-12 Sample probe positioning adjustment

    Note: the tip of the probe must be right against the positioning hole (judge by visual

    inspection).

    2) Rinse photocoupler calibration

    Tool: 101 slot-headed screwdriver

    Procedure: as Figure 2-13 shows, click "Menu""Setup""Service

    Setup""Photocoupler Cal.", the dialog box shown in Figure 2-14 will pop up.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    36/242

    Analyzer Structure

    2-16

    Figure2-13 Rinse photocoupler calibration menu

    Figure2-14 Rinse photocoupler calibration screen

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    37/242

    Analyzer Structure

    2-17

    Click "Start" and then click "OK", the dialog box "Priming diluent" will pop up, then an

    instruction message will be displayed, use the slot-headed screwdriver to rotate the

    potentiometer VR1(RINSE) on the drive board (see Figure 2-15) counter-clockwise

    until the D22 (RINSE-LED) turns on. Note: continue to rotate the VR1 for 180 degrees

    after the D22 turns on.

    Figure 2-15

    Connect the connector of the rinse pickup tube to the fitting on the back panel of the

    analyzer, click "OK" and then click "Finish", the dialog box "Priming rinse" will pop up,

    check the calibration status then: if "Pass" is displayed (see Figure 2-16), that means

    the calibration succeeded. The reference voltage and measured voltage are displayed

    as "Reference(V)" (1.8-3.0V) and Actual(V) in Figure 2-16.

    Figure 2-16

    Note:

    Rotate the VR1 for 180 degrees more after the D22 turns on, and make sure the LED

    is illuminated.

    Do not modify the value in the edit box Max(V), Min(V) and Factor(V);

    If calibration cannot be done successfully due to the quality of the photocoupler itself

    and the PCBA or photocoupler must be replaced, then recalibration is needed. The

    dialog box "Remove the Rinse tube from the container and press [OK] to continue."

    will pop up, press [OK] to drain the tubing; then do as instructed by the prompt - "1.

    Adjust rinse photocoupler potentiometer till the RINSE-LED turns on; 2 Put the Rinse

    tube into the container and press [OK]to close the dialog box; 3. Click [Finish] button"

    to prime rinse after calibration is done.

    "PASS"

    displayed

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    38/242

    Analyzer Structure

    2-18

    4The reproducibility and carryover functions are available.

    Figure 2-17 Menu of reproducibility and carryover

    5Check the replacement status of boards and CF card automatically. If the software

    system is started and logged in by user of service level, then a prompt will guide the user

    to restore the critical data.

    Figure 2-18

    1) If a new CF card is replaced, follow the guidance to select "Restore" and restore the

    critical data to the CF card.

    2) If a new data board is replaced, follow the guidance to select "Backup" and backup the

    data to the data board.

    3) If only the drive board is replaced, no special action shall be taken, the data will not be

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    39/242

    Analyzer Structure

    2-19

    lost.

    6Users of the service level can export special information, see the figure below.

    Figure 2-19

    2.5 LIS Connection

    2.5.1 LIS Setup and Adjustment

    1. Click "Menu""Setup""System Setup""Communication", the following screen will

    be displayed.

    Figure 2-20 Communication setup screen

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    40/242

    Analyzer Structure

    2-20

    2. Select network communication or serial port communication base on the physical

    medium used.

    3. If serial port communication is selected, communication can be realized when the

    parameter settings of the serial port are identical with those of the LIS software. The

    handshake function shall also be set up base on the LIS software used.

    4. If network communication is selected, it is recommended that direct-coupled network

    wire is used, and the major IP addresses such as 192.168.0. or 10.0.0.* are set up.

    When connected to LAN, set up the IP address per the IP rules of the LAN. ACK

    synchronous transmission shall also be set up base on the LIS software settings. If

    you are not sure about that, use the default setup.

    2.5.2 Connection Testing

    It is recommended that you use Mindray data management software to perform actual

    communication testing, or you can use the testing software developed by a compatible

    software developer.

    2.5.3 Analysis of Communication Failure

    1. For abnormal serial port connection, check parameter settings. If the settings are

    identical to those of the analyzer, replace serial port wire.

    2. For abnormal network connection, check parameter settings. If the settings are correct,

    check if the connection is normal by using ping. (If network firewall is installed in the

    computer of the LIS end, turn off the firewall). If the ping command fails to restore the

    connection, check if the network wire is well connected.

    If physical connection is good, and the communication parameters and protocol settings are

    correct, but communication failure persists, check if the LIS software complies with the

    communication protocol. Refer to the operator's manual for the communication protocols.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    41/242

    3-1

    3 Upgrade

    3.1 Software upgrade

    3.1.1 Preparation

    1The whole software upgrade procedure will take about 6~11mins.

    2What to do if there is an ERRORduring the upgrade

    1If there is a Mounting Failureplease restart the analyzer and run software upgrade again.

    2If there is a Upgrade Failure, please check the upgrade package in your USB flash disk and

    run software upgrade again. Do NOT restart or power off the analyzer immediately.

    3Requirements of the USB flash disk used in this upgrade.

    1Please use well-known brand USB flash disk.

    2Format(FAT32) the USB flash disk before use it.

    3Do NOT remove the USB flash disk during the upgrade.

    3.1.2 Upgrade procedure

    1Copy the update fileto a formatted USB flash disk.

    2Insert the USB flash disk into the USB interface of analyzer (See figure 3-1)

    Figure 3-1

    3Log in the BC-3600 with Service Engineer username and password.

    4Run SetupService SetupSetupStart Upgrade (see figure 3-2), and then click Yes

    in the dialog box (see figure 3-3).

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    42/242

    Upgrade

    3-2

    Figure 3-2

    Figure 3-3

    NOTES:

    1Make sure the UBS storage connected well to the USB interface of the Analyzer.

    2After you click Yes to start the upgrade, the system will start to read and analyze the

    configuration in the UBS storage and this will last for about 1min. Please wait for the pop-up

    upgrade confirmation dialog box.

    3Do NOT remove the USB flash disk during the upgrade.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    43/242

    Upgrade

    3-3

    5software upgrade.

    1) See the following software upgrade screen:

    Figure 3-4

    2) Click the Upgrade button to start the upgrade. You will see the upgrade preparation dialog.

    (See figure 3-5)

    Figure 3-5

    3) The upgrade preparation lasts for about 1 min, after that, you will see the upgrade confirm

    dialog. (See figure 3-6)

    Figure 3-6

    4) Click Yes to start the upgrade. (See figure 3-7)

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    44/242

    Upgrade

    3-4

    Figure 3-7

    5) You may need to restart the instrument for several times during the upgrade. (See figure

    3-8)

    Figure 3-8

    6) Upgrade succeeded. (See figure 3-9)

    Figure 3-9

    7) Upgrade failed. (See figure 3-10)

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    45/242

    Upgrade

    3-5

    Figure 3-10

    8) After you finish all the process, please restart the instrument. (See figure 3-11)

    Figure 3-11

    6After upgrade finished, log in with the service engineer username and password. Run

    ServiceVersion to check upgrade items. (See figure 3-12)

    Figure 3-12

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    46/242

    Upgrade

    3-6

    3.1.3 Notes and Prompts of Errors

    1During the upgrade, the software will check whether the hardware and software are

    matched. If not, select the proper language, click OK to start upgrade. (See figure 3-13)

    Figure 3-13

    2If upgrade failed, export the debug data. (See figure 3-14)

    Figure 3-14

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    47/242

    4-1

    4 Mechanical System

    4.1 Front of the Analyzer

    Figure 4-1 Front of the analyzer(closed tube model)

    1 --- Touchscreen 2 --- Indicator

    3 --- Sample compartment door 4 --- USB interface

    5 --- Recorder

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    48/242

    Mechanical System

    4-2

    Figure 4-2 Front of the analyzer(open vial model)

    1 --- Touchscreen 2 --- Indicator

    3 --- Sample probe 4 --- Aspirate key

    5 --- USB interface 6 --- Recorder

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    49/242

    Mechanical System

    4-3

    4.2 Back of the Analyzer

    Figure 4-3 Back of the analyzer

    1 --- USB interface 1 2 --- USB interface 2

    3 --- Network interface 4 --- Serial port

    5 --- Power switch 6 --- Waste outlet

    7 --- Rinse outlet 8 --- Diluent outlet

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    50/242

    Mechanical System

    4-4

    4.3 Left of the Analyzer

    Figure 4-4 Left of the analyzer (open vial model)

    1 --- Top cover 2 --- Front cover assembly

    3 --- Side door of electrical system 4 --- Back panel assembly

    5 --- Bottom panel assembly

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    51/242

    Mechanical System

    4-5

    Figure 4-5 Left of the analyzer (closed tube model)

    1 --- Top cover 2 --- Front cover assembly

    3 --- Side door of electrical system 4 --- Back panel assembly

    5 --- Bottom panel assembly

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    52/242

    Mechanical System

    4-6

    4.4 Right of the Analyzer

    Figure 4-6 Right of the analyzer (open vial model)

    1 --- Top cover 2 --- Front cover assembly

    3 --- Side door of fluidic system 4 --- Door lock

    5 --- Aspirate key 6 --- Back panel assembly

    7 --- Bottom panel assembly

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    53/242

    Mechanical System

    4-7

    Figure 4-7 Right of the analyzer (closed tube model)

    1 --- Top cover 2 --- Front cover assembly

    3 --- Side door of fluidic system 4 --- Door lock

    5 --- Aspirate key 6 --- Back panel assembly

    7 --- Bottom panel assembly

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    54/242

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    55/242

    5-1

    5 Fluidic System

    5.1 Fluidic Components and Functions

    5.1.1 Sample Probe

    Aspirating and dispensing sample.

    5.1.2 Probe Wipe

    Cleaning the exterior and interior walls of the sample probe.

    5.1.3 Pump Gas pump (GP): provide pressure to the vacuum chamber for generation of bubbles.

    Waste pump (LP): draining probe wipe, WBC bath, RBC bath and vacuum chamber

    and building vacuum.

    5.1.4 Syringes

    Sampling syringe (Asp-Syringe): full measuring range is 100l; aspirating fixed

    quantity of sample, dispensing the sample and aspirating the diluted sample again.

    Diluent syringe (Dil-Syringe): full measuring range is 10ml; dispensing fixed quantity

    of diluent to WBC and RBC bath, providing diluent to probe wipe.

    Lyse syringe (Lyse-Syringe): full measuring range is 2.5ml, motivated by the same

    motor with the diluent syringe; dispensing lyse.

    5.1.5 Valves

    Fluid valve and magnetic valve: diaphragm form, controlled by electric magnetic

    power. Controlling the directional flow of fluid or air.

    5.1.6 Baths

    WBC bath: composed of front bath, back bath and aperture. Providing room for the

    mixing of WBC sample; measuring HGB and WBC.

    RBC bath: composed of front bath, back bath and aperture. Providing room for the

    mixing of RBC sample; measuring RBC/PLT.

    Vacuum chamber: build and store stable vacuum for the impedance counting of WBC

    and RBC, clean the back baths and drain volumetric tubes.

    Pressure chamber: build and store stable pressure for generation of bubbles in all

    baths.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    56/242

    Fluidic System

    5-2

    WBC isolation chamber: providing air lock to block outside disturbance.

    RBC isolation chamber: providing air lock to block outside disturbance.

    5.1.7 Volumetric Tubes

    WBC volumetric tube: its volume is 300ul, measuring volume of the WBC sample to

    be analyzed.

    RBC volumetric tube: its volume is 200ul, measuring volume of the RBC sample to

    be analyzed.

    5.1.8 Filters

    Filter of WBC volumetric tube: filtrate the air entering the WBC volumetric tube.

    Filter of RBC volumetric tube: filtrate the air entering the RBC volumetric tube.

    GP filter: filtrate the air entering the gas pump.

    Customized filter: filtrate the impurities and scraps from the probe wipe.

    Filter of the isolation chamber: filtrate the impurities and scraps from the baths.

    5.2 Reagent Consumption

    There are four counting modes: OV-WB, OV-PD, CT-WB and CT-PD. See Table 5- 1 for the

    reagent volume consumed in one analysis cycle of a sample.

    Table 5- 1Reagent consumed in one analysis cycle of a sample

    Reagent consumed in

    one analysis cycle of a

    sampleItem Name

    Whole blood Predilute

    Diluent M-30D DILUENT 28mL 28mL

    LyseM-30CFL LYSE

    0.35mL 0.35mL

    Rinse M-30R RINSE 8mL 8mL

    Probe

    cleanserPROBE CLEANSER

    0 0

    Control B30BC-3D /

    Calibrator S30SC-CAL PLUS /

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    57/242

    Fluidic System

    5-3

    5.3 Sample Dilution Flowchart

    5.3.1 Whole Blood Mode

    The whole blood analysis cycle completes in 60s. The analysis procedure is as follows:

    1) Aspirate 17ul of sample, the sample probe ascends and the exterior wall of the probe is

    cleaned;

    2) Dispense 6ul of sample to the probe wipe to wash it off;

    3) Add diluent into the WBC bath and dispense 9ul of sample into the WBC bath, mix them to

    get the first dilution sample for WBC and HGB measurement.

    4) The sample probe ascends and its exterior wall is cleaned, the residual sample is washed

    off by diluent in the probe wipe.

    5) The sample probe descends to the WBC bath and aspirate 34ul of sample again, diluent is

    added and the second dilution sample is dispensed into the RBC bath, and mix them forRBC/PLT counting.

    6) Add lyse into the WBC bath, and mix thoroughly;

    7) Analyzing the sample;

    8) When the analysis completes, clean and zap the WBC/RBC bath, release the pressure in

    the vacuum chamber.

    See Figure 5-1 for the dilution procedure of whole blood mode:

    Figure 5-1 Dilution procedure of whole blood mode

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    58/242

    Fluidic System

    5-4

    5.3.2 Predilute Mode

    The predilute analysis cycle includes 3 function sequences (2 for closed tube models): diluent

    dispensing sequence, cleaning sequence (open vial models) and analyzing sequence. The

    procedure is as follows:1) The sample probe dispense 380ul of diluent to a centrifugal tube and 20ul of sample in the

    capillary tube is added in to the centrifugal tube, mix them to form the diluted sample;

    2) The sample probe aspirate 146.34ul of sample, and diluent is added into the WBC bath to

    form the first dilution sample for WBC and HGB measurement;

    3) The sample probe ascends and its exterior and interior walls are cleaned;

    4) The sample probe descends to the WBC bath and aspirate 37.39ul of sample again, diluent

    is added and the second dilution sample is dispensed into the RBC bath, and mix them for

    RBC/PLT counting.

    5) Add lyse into the WBC bath, and mix thoroughly;

    6) Analyzing the sample;

    7) When the analysis completes, clean and zap the WBC/RBC bath, release the pressure in

    the vacuum chamber.

    You can see from the above procedures that the analysis cycles of predilute mode and whole

    blood mode are only different in the operations of sampling and dispensing into WBC bath,

    therefore this section will only introduce the different parts.

    See Figure 5-2 for the dilution procedure of predilute mode:

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    59/242

    Fluidic System

    5-5

    Figure 5-2 Dilution procedure of predilute mode

    5.4 Introduction of Fluidic Channels

    5.4.1 WBC/HGB Channel

    All reagents:1)M-30PCFL lyse (lysing HGB, PLT and separating basophils from

    other WBCs ); 2) M-30PD diluent (cleaning, providing reaction and measurement

    environment)

    Analysis principles: impedance method (analyzing WBC); colorimetric method(analyzing HGB)

    Analysis parameters: WBC, HGB

    Graphic information: WBC histogram

    Dilution ratio:1:302 (WB); 1:369.77 (PD)

    Measurement volume: 300l

    Function description: Mix 9ul of blood and 2.35ml of diluent in the WBC bath,

    provide 34ul of first dilution sample to the RBC bath and add 0.35ml of lyse into it,

    mix them; then aspirate the mixed sample into the back bath through the aperture by

    the vacuum of the vacuum chamber. The cells are analyzed when they pass the

    aperture. Volume of the sample analyzed is measured by the volumetric tube.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    60/242

    Fluidic System

    5-6

    5.4.2 RBC/PLT Channel

    All reagents: diluent (diluting, cleaning, providing conductive environment, isometric

    processing of cells)

    Analysis principles: impedance method

    Analysis parameters: RBC, PLT

    Graphic information: RBC histogram, PLT histogram

    Dilution ratio:1:20000 (WB); 1:22333.07 (PD)

    Measurement volume: 200l

    Function description: the sample probe aspirates 34ul of sample (dilution ratio:

    1:262.11) from the WBC bath, add 0.8ml of mixture of the sample and diluent into the

    RBC bath to form the sample of dilution ratio 1:20000 with another 1.8ml of diluent;

    Mix them and aspirate the sample by vacuum of the vacuum chamber into the back

    bath through the aperture. The cells are analyzed when they pass the aperture.

    Volume of the sample analyzed is measured by the volumetric tube.

    5.5 Introduction of Basic Sequence

    5.1.1 Whole Blood Analysis Sequence

    This section introduces the working procedure of sample probe in the measurement sequence

    first, then all the other working procedures are introduced in detail.

    5.5.1.1 Sampling and dispensing procedure

    The working procedure of sample probe is indicated in Figure 5- 3. See section 5.3 for the

    sample dilution procedure. The working procedure of sample probe:

    a) Sample aspirating

    piercing(closed tube models)

    aspirating 17ul of sample.

    b) First dilution

    When aspiration finishes, the sample probe ascends;

    The sample probe moves to the top of the WBC bath; The sample probe descends to the WBC bath and dispenses 9ul of sample.

    c) Aspirating sample from the WBC bath

    The sample probe ascends from the WBC bath, and removes the residual

    sample in the probe wipe;

    The sample probe descends to the WBC bath and aspirates 34ul of first dilution

    sample.

    d) Second dilution in the RBC bath

    When aspiration finishes, the sample probe ascends from the WBC bath;

    The sample probe moves to the top of the RBC bath;

    The sample probe descends to the RBC bath, and dispenses 0.8ml of mixture of

    the sample and diluent.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    61/242

    Fluidic System

    5-7

    Initialize position of the sample probe by command INIT.

    RBC WBC

    ?

    ?

    ?

    RBC WBC

    ? ?

    RBC WBC

    ?

    a) b)

    c) d)

    ?

    ?

    Secon

    d

    dilutionsam le

    First

    dilutio

    n

    sampl

    DiluentBlood

    sample

    Figure 5- 3 Sampling and dispensing procedure

    5.5.1.2 The measurement procedure

    This section highlights the whole blood measurement sequence of the closed tube models.

    The measurement sequence consists of sample aspiration, dispensing, dilution, mixing,

    counting, and cleaning.

    Sampling (0~12.6s)

    The actions are:

    Pre-piercing: the pre-piercing process of sample probe (for closed tube models)

    to release the pressure in the tube before sample aspiration.

    Cleaning the exterior and interior walls of sample probe: when piercing finishes, the sample

    probe ascends to home position, meanwhile its exterior and interior walls are washed. See the

    following figure.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    62/242

    Fluidic System

    5-8

    Diluent

    Waste

    V01

    V04V02

    100ul

    10ml

    V19

    Diluent

    Waste

    V01

    V04V02

    V19

    b) Wash interior wall of

    sample probe

    a) Wash exterior wall of

    sample probe

    100ul

    10ml

    Figure 5- 4 Fluidic passages of sample probe cleaning

    The sample probe descends to the sampling position, and then moves to the

    piercing position, the syringe withdraws for 2ul to create separating bubbles.

    Then the sample probe descends to the lowest aspirating position to prepare for

    sample aspirating. Besides, the fluidic passages of draining and dispensing of

    the WBC bath are indicated in Figure 5- 5 and Figure 5- 6; Under the effect of

    vacuum, rinse flows through and washes the WBC and RBC back baths and

    their volumetric tubes, the fluidic passage of washing the WBC back bath and

    volumetric tube is indicated in Figure 5- 7; When the washing progress ends,

    Valve 5 and 6 opens to drain the volumetric tubes.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    63/242

    Fluidic System

    5-9

    Diluent

    Waste

    V01

    Isolationchamber

    WBC

    bath

    V16

    V03

    Lyse

    V04

    V02

    10mL

    SR3

    2.5mL

    (SR1)

    Figure 5- 5 Fluidic passage of draining of the WBC bath

    DiluentV01

    WBC

    bath

    V03

    lyse

    V04

    V02

    10mL

    SR32.5mL

    (SR1)

    Figure 5- 6 Fluidic passage of dispensing of the WBC bath

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    64/242

    Fluidic System

    5-10

    Rinse

    Waste

    V09Isolation

    chamber

    WBC

    bath

    V10

    V16

    V14

    Volumetricboard

    V05 V07

    V18

    Vacuum

    chamber

    Pressure

    Atmosp

    heric

    pressure

    Figure 5- 7Fluidic passages of washing WBC back bath and volumetric tube

    Aspirating sample: the sampling syringe aspirates 17ul of sample at 6.5~8.5s;

    and 800ul of diluent is added to the RBC bath after it is drained;

    Ascending to home position: the sample probe ascends to home position at

    9.3~11.6s, and its exterior wall is washed; by now, the sampling procedure isfinished.

    Other actions finished during the sampling procedure:

    After sample dispensing of the WBC bath finishes (7s) and the fluid level

    stabilizes, the HGB voltage and WBC aperture voltage (the constant-current

    source is opened at 10.5s in advance) is read at 10.5s.

    Sample Dispensing, Dilution and Mixing (11.5~30.8s)

    The actions are:

    Dispensing sample for the first time: the sample probe removes 6ul of sample in

    the probe wipe at 11.5~12.3s.

    Dispensing sample for the second time: the sample probe moves to the top of

    WBC bath and descends into the bath; the diluent syringe dispenses 1150 ul of

    diluent into the WBC bath, then the sample probe dispensed 9 ul of sample and

    swings; when sample dispensing finishes, the diluent syringe dispenses 1200ul

    of diluent into the WBC bath again; Valve 10 opens, and blows one small bubble

    and 3 big bubbles into the WBC bath in turn. By now, the first dilution of sample

    finishes.

    The sample probe ascends with its exterior and interior walls washed: when the

    second sample dispensing finishes, the sample probe ascends to home position

    with its exterior and interior walls washed; besides, 2700ul of diluent isdispensed into the RBC bath.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    65/242

    Fluidic System

    5-11

    Second aspiration in the WBC bath: the sample probe descends to the WBC

    bath and aspirates 34 ul of fluid at 19.9~21s.Valve 17 opens, and the RBC bath

    is drained.

    The sample probe ascends: when the second aspiration finishes, the sample

    probe ascends to home position with its exterior wall washed; besides, 1800 ul

    of diluent is added to the RBC bath at 24.5~25.5s.

    Adding sample to RBC bath, and lyse to WBC bath: the sample probe moves to

    the RBC bath and descends into the bath; the lyse syringe dispenses 350ul of

    lyse into the WBC bath, the fluidic passage is indicated in Figure 5- 8; the diluent

    syringe discharges 800 ul of fluid (mixture of first dilution sample and diluent)

    through the sample probe. Valve 10 opens at 26.65~27.25s, the pressure

    chamber blows bubbles into the WBC bath; then a bubble that can last for 0.05s

    is blowed into the WBC bath, and 3 bubbles are blowed into the RBC bath in

    turn; Finally, the sample probe restores to the original position with its exterior

    wall washed.

    WBC

    bath

    V03

    Lyse

    2.5mL

    (SR1)

    Figure 5- 8 Fluidic passage of lyse dispensing of the WBC bath

    Other actions finished during the sample dispensing, dilution and mixing

    procedure:

    Build the vacuum needed for analysis, total time consumed is 7.9s;

    Build the pressure needed for bubble generation;

    RBC aperture voltage inspection: same as WBC aperture voltage inspection;

    The sample compartment door opens at 12s to prepare for the next analysis

    cycle.

    Analysis

    The actions are:

    Measuring volume of volumetric tube: valve 7 opens after WBC bubbling

    finishes for a while and valve 8 opens after RBC bubbling finishes for a while,

    the fluidic passage is indicated in Figure 5- 9; the closing of valve 7 and valve 8

    is finished by the ending cleaning cycle started when the WBC and RBC downphotocouplers are triggered.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    66/242

    Fluidic System

    5-12

    RBC

    bath

    Volumetric

    board

    Rinse

    Waste

    V06 V08

    V15

    WBC

    bath

    V14

    Volumetric

    board

    V05 V07

    V18

    Vacuum

    chamber

    Figure 5- 9 Fluidic passages of WBC and RBC analysis

    The photocoupler of volumetric board is opened at 25.7s.

    Cleaning when analysis ends

    The actions are:

    The ending cleaning sequence is an independent sequence, it is started by the

    software automatically base on the information obtained during the analysis;

    Fluid discharging of RBC and WBC baths: open valve 17, the RBC bath starts to

    discharge fluid; open valve 16, the WBC bath starts to discharge fluid;

    Fluid dispensing of RBC and WBC baths: 1400 ul of fluid is dispensed into the

    RBC and WBC bath respectively;

    Washing RBC and WBC back baths: when valve 8 and valve 15 open at the

    same time, the RBC back bath will be washed by rinse under the effect ofvacuum; when valve 7 and valve 14 open at the same time, the WBC back bath

    will be washed by rinse under the effect of vacuum.

    Draining the vacuum chamber: open valve 9 to drain the vacuum chamber.

    Resetting of the diluent syringe and sample probe assembly: the diluent syringe

    aspirates 2100 ul of fluid, then valve 1 opens, the syringe returns to home

    position and dispenses fluid to the RBC bath at the same time; the sample probe

    assembly returns to home position.

    The sample probe aspirates air bar and returns to home position: the sample

    probe moves to the piercing position and withdraws for 2ul and then returns to

    home position.

    Zapping: zapping starts when fluid dispensing of WBC bath completes for a

    while.

    Calculating the volume of rinse consumed: the software reads from the

    command to calculate the volume of rinse consumed.

    Detecting rinse status: sensor in the analyzer detects if the rinse container is

    empty; when alarming for no rinse, the residual volume of rinse must be able to

    support the on-going analysis.

    The major difference between the whole blood analysis sequences of open vial models and

    closed tube models lies in the sampling procedure, the open vial sequence involves nopiercing action, but its sampling time point and the cleaning action after sampling are the same

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    67/242

    Fluidic System

    5-13

    with those of the closed tube sequence, therefore the whole blood analysis sequences of open

    vial models will not introduced in detail.

    5.1.2 Predilute Measurement Sequence

    The predilute analysis cycle includes 3 function sequences (2 for closed tube models): diluent

    dispensing sequence, cleaning sequence (open vial models) and analyzing sequence.

    The predilute measurement procedure: (1)The sample probe dispense 380ul of diluent to a

    centrifugal tube and 20ul of sample in the capillary tube is added in to the centrifugal tube, mix

    them to form the diluted sample. (2)The sample probe aspirate 146.34ul of sample, and diluent

    is added into the WBC bath to form the first dilution sample for WBC and HGB measurement.

    (3)The sample probe ascends and its exterior and interior walls are cleaned. (4)The sample

    probe descends to the WBC bath and aspirate 37.39ul of sample again, diluent is added and

    the second dilution sample is dispensed into the RBC bath, and mix them for RBC/PLT

    counting. (5)Add lyse into the WBC bath, and mix thoroughly; (6)Analysis. (7)When the

    analysis completes, clean and zap the WBC/RBC bath, release the pressure in the vacuum

    chamber.

    You can see from the above procedures that the analysis cycles of predilute mode and whole

    blood mode are only different in the operations of sampling and dispensing into WBC bath,

    therefore this section will only introduce the different parts.

    See section 5.3.2 for the dilution procedure of predilute mode.

    5.5.2.1 Dilution dispensing sequence

    The procedure is indicated in Figure 5- 10:

    a) closed tube models b) open vial models

    Figure 5- 10 Diluent dispensing flowchart

    5.5.2.2 Cleaning sequence after diluent dispensing

    Introduction:

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    68/242

    Fluidic System

    5-14

    1. This sequence applies to open vial models only.

    2. The sequence cleans the exterior wall of the sample probe, initializes diluent syringe

    by command INIT; the sampling syringe withdraws for 2ul to create a separating air

    bar, the system restores to "Ready" status.

    5.5.2.3 Predilute Measurement Sequence

    Introduction:

    Sampling:

    1. The diluent syringe withdraws for 100ul through the probe wipe tube;

    2. WBC/RBC bath is zapped for 3s;

    3. The diluent syringe aspirates 146.34ul of sample;

    4. When aspirating finishes, valve 01 closes 0.1s after valve 04 closes.

    Sample dispensing of WBC bath:

    1. The diluent syringe pushes sample in the sampling tube into the WBC bath to form

    the first dilution sample;

    2. Sample volume of the second aspiration is 37.39ul;

    3. The predilute measurement takes 1.5s more than the whole blood measurement due

    to the zapping action added before analysis;

    4. The other parts of sequence are the same as those of the whole blood measurement

    sequence, only the fluid volumes are different in some steps.

    5. The major difference between the closed tube and open vial models is the position of

    the sample probe under the "Ready" status.

    5.1.3 Startup/Shutdown Sequence

    5.5.3.1 Startup sequence

    There are three types of startup sequence: 1 startup after normal shutdown; 2 startup after

    abnormal shutdown; 3 startup after performing "Prepare To Ship" procedure.

    Startup after normal shutdown

    This sequence includes the troubleshooting sequence 1, fluidics initializing sequence and

    startup sequence after normal shutdown.

    1. Troubleshooting sequence 1, refer to the corresponding section.

    2. Fluidics initializing sequence, refer to the corresponding section.

    3. Startup after normal shutdown

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    69/242

    Fluidic System

    5-15

    Figure 5- 11Startup procedure after normal shutdown

    Introduction:

    1 There are zapping and flushing operations at the beginning to avoid clogging after

    long term of probe cleanser soak.

    2 0.5ml of lyse is discharged to make sure the tube mouth is filled with reagent.

    3 Push back 200ul of diluent to the cleanser tube to wash off the cleanser at the

    closed end of valve 12.

    4 Empty RBC charging line to avoid crystallizing of the cleanser spills on the wall of

    the bath which may influence the PLT background.

    5 Empty the sample probe and dispense fluid: after emptying the sampling tube,

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    70/242

    Fluidic System

    5-16

    dispense fluid at a higher speed so that the bubbles in the sampling tube can be

    removed effectively.

    6 The sample probe dispenses fluid in the WBC bath: removing all the bubbles in

    the sampling tube.

    7 Cleaning the two baths: cleanser volumes of WBC bath 4ml, 4.6ml, 5.5ml;cleanser volumes of RBC bath 4ml, 4ml, 5ml, generating bubbles during the

    cleaning process to strengthen the cleaning effect.

    8 Emptying WBC, RBC baths and the back baths, switching among valve 16 ,17, 7

    and 8 to strengthen the cleaning effect.

    Startup after abnormal shutdown

    This sequence includes the troubleshooting sequence 1, fluidics initializing sequence and

    startup sequence after normal shutdown.

    1. Troubleshooting sequence 1, refer to the corresponding section.

    2. Fluidics initializing sequence, refer to the corresponding section.

    3. Startup after normal shutdown

    Figure 5- 12 Startup procedure after abnormal shutdown

    Introduction:

    1 When starting up the analyzer after abnormal shutdown, there may be some

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    71/242

    Fluidic System

    5-17

    residual sample in the fluidic components, so the exterior and interior walls of

    the sample probe must be cleaned repeatedly by fluid discharging of the

    sample probe and probe wipe.

    2 0.5ml of lyse is discharged to make sure the tube mouth is filled with reagent.

    3 Push back 200ul of diluent to the cleanser tube to wash off the cleanser at theclosed end of valve 12.

    4 Remove the bubbles in the sampling tube through fluid discharging for multiple

    times.

    5 Clean the two baths repeatedly to remove the possible residual sample.

    6 Zapping for 4s to avoid clogging.

    7 Wash tubes of the back baths twice to remove the residual sample in the tubes

    or volumetric tubes.

    Startup after performing "Prepare To Ship" procedure

    This sequence includes the troubleshooting sequence 1, fluidics initializing sequence and

    overall reagent priming sequence.

    1. Troubleshooting sequence 1, refer to the corresponding section.

    2. Fluidics initializing sequence, refer to the corresponding section.

    3. See section 5.5.3.5 for the overall reagent priming sequence.

    5.5.3.2 Shutdown sequence

    The probe cleanser soak sequence (general and intensified, refer to 5.5.4.7 and 5.5.4.8 for

    details), intensified shutdown unclogging sequence (refer to 5.5.4.2 for details) and shutting

    down sequence will be performed in turn during the shutdown process.

    As per the different probe cleanser soak sequences, the shutdown sequence can be classified

    into two kinds: 1 shutdown after general probe cleanser soak, 2 shutdown after intensified

    probe cleanser soak.

    The operations of the shutting down sequence are: washing the exterior wall of the sample

    probe and soaking the probe in the WBC bath.

    5.1.4 Cleaning/Maintenance Sequence

    5.5.4.1 Unclogging sequence

    Introduction:

    1. Drain the two baths.

    2. Valve 9 opens to connect the pressure chamber and vacuum chamber and build

    pressure.

    3. Flush the aperture for 4s and drain the fluid.

    4. Add 3.5ml of fluid in the two baths respectively.5. Zap the two baths for 4s respectively.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    72/242

    Fluidic System

    5-18

    6. Wash the tubes of the back baths. Empty the vacuum chamber and release vacuum.

    5.5.4.2 Zapping sequence

    Introduction:

    1. Build vacuum in the vacuum chamber, zap the two baths for 4s.

    2. Drain the two baths and add 3.5ml of fluid in the two baths respectively.

    3. Wash the tubes of the back baths. Empty the vacuum chamber and release vacuum.

    5.5.4.3 Flushing sequence

    Introduction:

    1 The flushing sequence is almost the same as the unclogging sequence, only it involves no

    zapping operation.

    5.5.4.4 Draining sequence of the baths

    Introduction:

    1 Drain the two baths for 3s.

    5.5.4.5 Priming sequence of the baths

    Introduction:

    1. Drain the two baths and add 3.5ml of fluid in the two baths respectively.

    2. The function is activated after the baths are drained. The baths may need to be

    serviced during the draining process, the tubes of the back baths need to be washed.

    3. Drain the vacuum chamber and release vacuum.

    5.5.4.6 Cleaning sequence of the baths

    Introduction:

    1. 4.5ml of fluid is added to the two baths respectively, and bubbles are generated to

    strengthen the cleaning effect.

    2. Drain the two baths and add 3.5ml of fluid into the two baths respectively, open valve

    18 and the fluidic pump for a while to release pressure in the pressure chamber.

    5.5.4.7 General probe cleanser soak sequence

    This sequence includes soaking sequence and the cleaning sequence after soaking.

    Soaking procedure:

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    73/242

    Fluidic System

    5-19

    Figure 5-13 Soaking procedure

    Introduction: the soaking process consists of three steps. Step 1: soak the WBC aperture

    and interior wall of the sample probe with 100% probe cleanser. Step 2: soak the WBC and

    RBC front/back baths with 30.77% and 38.46% probe cleanser respectively. Step 3: soak the

    WBC and RBC front/back baths with 26.32% and 21.05% probe cleanser respectively. The

    operations are:

    1. The sampling syringe withdraws for 2ul to form a separating air bar.

    2. The sample probe aspirates 1.8ml of probe cleanser, the two baths are drained,

    the sample probe ascends with its exterior wall washed.

    3. The sample probe dispenses 1ml of probe cleanser into the WBC bath to form

    100% probe cleanser.

    4. The sample probe dispenses 2.6ml of mixture of probe cleanser and diluent into

    the RBC bath to form the high concentration compound (30.77%) for the first

    soak.

    5. The sample probe dispenses 1.6ml of diluent into the WBC bath to form the highconcentration compound (38.46%) for the first soak.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    74/242

    Fluidic System

    5-20

    6. After soaking for about 2 minutes, dispenses 1.2ml of diluent to the two baths to

    dilute the probe cleanser, which will be used for the long term soaking of the front

    and back baths (WBC:26.32%; RBC:21.05%).

    7. Build vacuum and aspirate fluid: aspirate for 6s from the WBC bath, the volume

    aspirated is about 232ul; aspirate for 8s from the RBC bath, the volume aspiratedis about 236ul; When the aspiration finishes, release the vacuum in the vacuum

    chamber.

    8. Meanwhile, the sample probe aspirates 50ul of fluid from the RBC bath for

    soaking of the interior wall of sample probe and the tube.

    9. Soak the tubes for 2 minutes and then start the cleaning procedure.

    Cleaning procedure:

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    75/242

    Fluidic System

    5-21

    Figure 5-14Cleaning procedure

    Introduction: the front and back baths, sampling tube, waste valve shall be cleaned to remove

    residual probe cleanser.

    1. Add probe cleanser to the WBC bath; when draining the WBC bath, switch on/off

    valve 16 for multiple times to soak the waste valve with probe cleanser (for about

    5s) to remove residual probe cleanser.

    2. Add probe cleanser to the RBC bath; when draining the RBC bath, switch on/off

    valve 17 for multiple times to soak the waste valve with probe cleanser (for about

    5s) to remove residual probe cleanser.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    76/242

    Fluidic System

    5-22

    3. The sample probe dispenses 7.8ml of probe cleanser to the RBC bath to clean

    the sampling tube and remove bubbles in the tubing.

    4. Discharge fluid to the probe wipe twice, meanwhile the sample probe ascends to

    avoid residual probe cleanser on the exterior wall of the sample probe or in the

    probe wipe.5. Build vacuum, wash tube of the back bath, switch on/off valve 7 and 8 for

    multiple times.

    6. Empty the vacuum chamber, switch on/off valve 18 for multiple times.

    7. Drain the baths and build pressure to 40kpa.

    8. Flushing the apertures.

    9. Drain the two baths and add 3.5ml of fluid into the baths respectively, zap the

    baths for 4s and build vacuum to -24kp.

    10. Wash the back baths, switch on/off valve 7 and 8 for multiple times to strengthen

    the cleaning effect of the valves and remove residual probe cleanser.

    11. Empty the vacuum, switch on/off valve 18 for multiple times to strengthen the

    cleaning effect of the valves and remove residual probe cleanser.

    12. Release the pressure, the analyzer restores to "Ready" status.

    5.5.4.8 Intensified probe cleanser soak sequence

    Soaking procedure:

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    77/242

    Fluidic System

    5-23

    Figure 5-15 Soaking procedure

    Introduction: the intensified probe cleanser soak procedure is almost the same as the generalprobe cleanser soak procedure, the only difference lies in the dilution ratio of probe cleanser.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    78/242

    Fluidic System

    5-24

    Cleaning procedure:

    The cleaning operations of the intensified and general probe cleanser soak procedure

    are almost the same, the only difference lies in the occurrence time of the operation.

    5.5.4.9 Draining sequence of the tubing

    Procedure:

    Figure 5- 16Draining procedure of the tubing

    Introduction:

    1. Build vacuum in the vacuum chamber to drain the back bath and rinse

    tubes.

    2. Drain the RBC bath, the diluent syringe withdraws for 1.3ml to drain the

    charging line of RBC bath.

    3. The lyse syringe aspirates and discharges lyse to the WBC bath for

    multiple times to drain the lyse tube.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    79/242

    Fluidic System

    5-25

    4. The diluent syringe aspirates and discharges diluent to the WBC bath for

    multiple times to drain the diluent tube.

    5. The sampling syringe aspirates 80ul of fluid.

    6. The sample probe goes inside the WBC bath to drain the sampling tube.

    7. Drain the charging line and the waste tube of the probe wipe.

    8. Push the 80ul of fluid to the syringe inlet.

    9. After build vacuum in the vacuum chamber, open the rinse tubes and empty

    tubes of the back baths.

    10. Empty the vacuum chamber and build vacuum at the same time, open V5

    and V6 of the volumetric tube and dry the volumetric tube.

    11. Release vacuum and pressure, the syringes return to home position, the

    sample probe returns to the sampling position.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    80/242

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    81/242

    6-1

    6 Hardware System

    6.1 Introduction

    The hardware system consists of 8 boards, which are power board, data board, drive board,

    indicator board, touch screen control board, volumetric boars, sample compartment connector

    board, and USB interface board; the system also includes the drive and detect components

    that require power supply (such as motors, valves, pumps, sensors, display, and filter of input

    power) and the wires connecting boards and components.

    6.1.1 Function DiagramSee Figure 6-1 for the function diagram of the hardware system.

    Figure 6-1Function diagram of the hardware system

    The hardware system consists of f ive modules, which are system power, data stream channel,

    central system, drive components and peripherals. The function of each module is as follows:

    1. System power: providing power of various specifications to the boards, components

    and devices.

    2. Data stream channel: extracting, adjusting, amplifying and collecting and

    preprocessing of signals.

    3. Central control system: data collection and processing, result display, sample

    storage, etc. Besides, the central control system controls and responds to all

    peripherals.

    4. Drive/detect components: controlling valves, pumps and motors, monitoringphotocouplers and other major parameters, collecting detection information and

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    82/242

    Hardware System

    6-2

    sending alarms.

    5. Peripherals and interfaces: display/touchscreen, recorder, USB interfaces (for printer,

    keyboard and barcode scanner), Ethernet interface and LIS serial port. Besides,

    peripherals includes status indicator, key-presses and aspirate key input.

    6.1.2 Electric Connection Diagram

    See Figure 6-2 for the electric connection diagram of the hardware system.

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    83/242

    Hardware System

    6-3

    Figure 6-2 Electric connection diagram of the hardware system

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    84/242

    Hardware System

    6-4

    6.1.3 Troubleshooting the System

    The errors of the hardware system include board errors, wire errors and component errors.

    Generally ,the troubleshooting methods of the three types of errors can be found in the following

    sections; however, if the power supply of the hardware system cannot be ensured (the analyzer

    cannot be electrified, or self-shields right after power on), the errors shall be fixed on the system

    level.

    1 Troubleshooting procedure when the analyzer cannot be electrified (the indicators of the drive

    board and the data board are both off)

    (1) Check if the AC input wire outside the analyzer is firmly connected, if not, re-plug the wire.

    (2) Power off the analyzer, check if the wire of the power input filter is firmly connected, if not,

    re-plug the wire.

    (3) If step (1) and (2) cannot remove the error, replace the power input filter (see Figure 6-3);

    if the error persists, do step (4).

    Figure 6-3 Power input filter

    (4) Replace the power board or assembly.

    2 Troubleshooting procedure when the analyzer self-shields right after power on

    (1) First, unplug the wires of the drive board and data board (see Figure 6-4), see if the power

    can be started; if not, check if the wire is damaged; if so, replace the wire; otherwise replace

    the power board or assembly. If the power can be started normally, the problem of power

    assembly can be excluded.

    Figure 6-4Sockets of drive board and power board

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    85/242

    Hardware System

    6-5

    (2) After completing step (1), plug the two wires on of the drive board in turn to see if the power

    can be started; if yes, the problem of the drive board can be excluded, the data board needs to be

    replace; if no, unplug the wire of the volumetric board; if the power can be started, that means the

    volumetric board or its wire is short-circuited, the volumetric board or its wire shall be replaced; if

    not, that means the drive board is short-circuited, the drive board shall be replaced.

    If the power can be started normally, troubleshoot the error per the following sections.

    6.2 Data Board

    6.2.1 Introduction

    The data board consists of the analog and digital parts, the analog part filters, amplifies, outputs

    and processes signals to obtain the signals that are suitable for A/D data conversion. The digitalpart processes, outputs, controls and transmits data. It is the core part of the data board or even

    the entire hardware system.

    Section 6.2 is the servicing and troubleshooting guideline of the data board.

    6.2.2 Structure

    Introduction of Functions

    See Figure 6-5 for the structural diagram of the data board.

    Figure 6-5Structural diagram of the data board

    You can see from Figure 6-5, the data board consists of the analog circuit and digital circuit. The

    analog circuit adjusts and amplifies the WBC, RBC, PLT and HGB signals to make sure the

    signals are turn and suitable for A/D conversion; the A/D module is the interface of the analog

    circuit and digital circuit, it collects the sensing signals and other monitoring signals, and converts

    analog signals into digital signals; the digital circuit processes data, stores and outputs results.

    Besides, the digital circuit takes on the control and communication tasks, it is the core part of the

    data board or even the entire hardware system.

    1. Analog module

  • 7/24/2019 BC-3600&3300 Service Manual V1.0

    86/242

    Hardware System

    6-6

    Figure 6-6 Structural diagram of the analog module

    The analog module of the data board consists of RBC/PLT channel, WBC channel, HGB

    channel and A/D conversion part, its functions are:

    Adjusting analog power

    Adjusting volume signal

    Data collection

    2. Digital module

    Figure 6-7 Structural diagram of the digital module

    The structure of the digital module is "CPUFPGA", see Figure 6-7.The functions of the