Basic Approach to Dissolution Method Development – Challenges and Regulatory Issues

133
Dr. Harshal Ashok Pawar Assistant Professor & HOD (Quality Assurance), Dr. L. H. Hiranandani College of Pharmacy, Ulhasnagar Email : [email protected] , [email protected]

Transcript of Basic Approach to Dissolution Method Development – Challenges and Regulatory Issues

Dr. Harshal Ashok PawarAssistant Professor & HOD (Quality Assurance),

Dr. L. H. Hiranandani College of Pharmacy, UlhasnagarEmail : [email protected] , [email protected]

CONTENTS

Theoretical aspect

Instrumentation

Pre-requisite

Selection of Dissolution medium

Optimization of dissolution parameters Optimization of dissolution parameters

Specifications /Acceptance criteria

Comparison of Dissolution Profiling

Biovaivers

Theoretical AspectTheoretical Aspect

DISSOLUTION

In the body, a pharmaceuticalactive ingredient must be “insolution” before it can beabsorbed by the blood andabsorbed by the blood andultimately carried to thereceptor site to render atherapeutic effect (in vivo).

Dissolution is the processby which that activeingredient enters intosolvent to yield a solution.

DISSOLUTION SOLUBILITY

Dissolution rate is defined asthe amount of solid substancethat goes into solution perunit time under standard

Absolute solubility is definedas the maximum amount ofsolute dissolved in a givensolvent under standardunit time under standard

conditions of temperature,pH, solvent composition andconstant solid surface area.

solvent under standardconditions of temperature ,pressure and pH.

It is a dynamic process. It is a static process.

MECHAMISM OF DISSOLUTION

Initial mechanical lag

Wetting of dosage form

Penetration of dissolution medium

Disintegration Disintegration

Deaggregation

Dissolution

Occlusion of some particles

Dissolution process of solid dosage Forms :

DISINTEGRATION DISSOLUTION

DISSOLUTION ABSORPTION

IN-VIVO

TABLETS OR CAPSULES

GRANULES OR AGGREGATES

DRUG IN SOLUTION

(IN-VITRO OR IN-

DRUG IN BLOOD,OTHER

FLUIDS,AND IN-VIVO

DISAGGREGATION

DISSOLUTION

AGGREGATES

FINE PARTICLES

(IN-VITRO OR IN-VIVO)

FLUIDS,AND TISSUES

Dissolution TestingDissolution is one the three primary tests used to

release a finished drug product:

Assay – determines the overall potency of the batch and ensuresthe accuracy of the finished drug product.

Dose Uniformity – determines the consistency among the Dose Uniformity – determines the consistency among theindividual dosage units and ensures the precision of themanufacturing process.

Dissolution – ensures that the performance of the finished drugproduct is consistent with the release rates of the API as determinedin bioavailability studies during the clinical trials.

Which product to be tested?USP <1088>:

“No product, including suspensions and chewable tablets,should be developed without dissolution or drug releasecharacterization where a solid phase exists.”

and

“Dissolution testing is required for all solid oral Pharmacopeial“Dissolution testing is required for all solid oral Pharmacopeialdosage forms in which absorption of the drug is necessaryfor the product to exert the desired therapeutic effect.

Exceptions are for tablets meeting a requirement forcompleteness of solution or for rapid (10 to 15 minutes)disintegration for soluble or radiolabeled drugs.”

NEED FOR DISSOLUTION TESTING

Development and Optimization of dosage forms.

Batch to batch drug release uniformity.

Ensures quality, safety, efficacy and stability of the product.

Evaluation of IVIV Correlation / bioavailability .

To support waiver for bioequivalence requirement.

For assessing Pre and Post approval changes in manufacturing

process, or formulation

Dissolution and Drug Development

Drug

Clinical trial (IVIV)

Manufacturing

(Quality control)

Discovery ( Little to no dissolution)

Characterization (Intrinsic dissolution)

Drug development (DMD)

control)

InstrumentationInstrumentation

I.P U.S.P B.P E.P

TYPE 1 Paddleapparatus

Basket apparatus

Basket apparatus

Basket apparatus

TYPE 2 Basket apparatus

Paddleapparatus

Paddleapparatus

Paddleapparatus

TYPE 3 Reciprocating cylinder

Flow through cell

Flow through cell

TYPE 4 Flow through cell

TYPE 5 Paddle over disk

TYPE 6 Rotatingcylinder

TYPE 7 Reciprocating disk

APPARATUS-1(ROTATING BASKET)

DESIGN:

Vessel: -Made of borosilicate glass.

-Semi hemispherical bottom

-Capacity 1000ml

Shaft : -Stainless steel type 316

-Rotates smoothly without

significance wobble(100 rpm)

-Speed regulator

Water bath:-Maintained at 37±0.5ºC

USE: Tablets, capsules, delayed release

suppositories, floating dosage forms.

Apparatus 1 - Basket

Advantages

Full pH change during the test

Can be easily automated which is important for routine investigations.

Disadvantages

Basket screen is clogged with

gummy particles.

Hydrodynamic dead zoneunder the basket

Degassing is particularly important

Mesh gets corroded by HCl solution.

APPARATUS-2 (PADDLE)

DESIGN:

Vessel: -Same as basket apparatus

Shaft: -The blade passes through the shaft so that the bottom of the blade fuses with bottom of the shaft.

Stirring elements: -Made of tefflon

For laboratory purposeFor laboratory purpose

-Stainless steel 316

Water-bath: -Maintains at 37±0.5°C

Sinkers : -Platinum wire used to prevent

tablet/capsule from floating

Advantages

Easy to use

Robust

Can be easily automated which is important for

routine investigations

Disadvantages Disadvantages

pH/media change is often difficult

Sinkers for floating dosage forms

Paddle apparatus:

APPARATUS-3(RECIPROCATING CYLINDER)Tester was designed to test the dissolution rates of extended releaseproducts or any dosage form requiring release profiling at multiple pHlevels.

DESIGN:

Vessel: -Set of cylindrical flat bottom glass vessels

-Set of reciprocating cylinders

-stainless steel fittings(type 316) and

screens made of nonsorbing or screens made of nonsorbing or

non-reactive materials.

Agitation type: -Reciprocating at 5-35 rpm

Volume of dissolution medium:-200-250ml

Water bath:- Maintain at 37±0.5°C

USE: Chewable Tablets, beads, controlled and

Extended release formulations, Soft gel cap,

Non Disintegrating type products

METHOD(Reciprocating cylinder):

Place the stated volume of dissolution medium in each vessel of the apparatus, assemble the apparatus, equilibrate the dissolution medium to 37±0.5 and remove the thermometer

Place one dosage form unit in each of the cylinders taking care to exclude the air bubbles from the surface of each dosage unit and immediately operate the apparatus as specified in the monograph.

During the upward and downward stroke, the reciprocating cylinder During the upward and downward stroke, the reciprocating cylinder moves through a total distance of 9.9 to 10.1cm.

Within the time interval specified, raise the cylinders and withdraw a portion of the solution under test from a zone midway between the surface of the dissolution medium and bottom of each vessel.

Advantages Easy to change the pH

pH-profiles

Hydrodynamics can be directly influenced by varying the dip rate

Disadvantages Small volume (max. 250 ml)

Little experience

Limited data

APPARATUS-4 (FLOW THROUGH CELL)

DESIGN:

Reservoir : -For dissolution medium

Pump : -Forces dissolution medium through cell

-Holding a sample

-Flow rate 10-100ml/min

-Laminar flow is maintained

-Peristaltic/centrifugal pumps are not recommended

Water bath:- Maintain at 37±0.5°C

USE:

Modified-release dosage forms and immediate-release dosage forms

Soft gelatin capsules, beaded products, suppositories, or injectable-depot dosage forms

Suspension-type extended-release dosage forms for oral or parenteral use, or ocular application.

METHOD(Flow through cell):

The flow through cell is transparent & inert mounted vertically with filters.

Standard cell diameters are 12 & 22.6 mm.

The bottom cone usually filled with glass beads of 1 mm diameter.

Tablet holder used for positioning special dosage form e.g. inlay tablets.

Place the glass beads into the cell as specified in the monograph. Place the glass beads into the cell as specified in the monograph.

Place one dosage unit on top of the beads or on a wire carrier.

Assemble the filter head and fix the parts together by means of a suitable clamping device.

Introduce by the pump of the dissolution medium warmed to 37±0.5 through the bottom of the cell to obtain the flow rate specified and measured with an accuracy of 5%.

Collect the eluate by fractions at each of the times stated.

Advantages

Easy to change media pH

pH-profile possible

Sink conditions maintained.

Different modesa) open systemb) closed system

Disadvantages Deaeration necessary High volumes of media Labor intensive

Cell types:

Tablets 12 mm Tablets 22.6 mm Powders / Granules Implants Suppositories / Soft gelatincapsules

Flow-Through Cell:

APPARATUS-5(PADDLE-OVER-DISK)

DESIGN:

Vessel

Shaft

Stirring elements- rotating speed 25-50 rpm

Sample holder:-disk assembly that hold a product in such a way Sample holder:-disk assembly that hold a product in such a way

that release surface is parallel with paddle

-Paddle is directly attached over disk assembly

-Samples are drawn between surface off the medium

and top of the paddle blade

Volume:900ml

Temperature:32°C (Similar to skin)

USE: Transdermal patches, ointments, floaters , emulsions.

Modification: Disk design and volume

Advantages:

Easy to handle

Sink conditions are maintained.

Membrane effect is minimum.

i.e. drug is placed on a disc at the bottom.

Disadvantages:Disadvantages:

Disk assembly restricts the patch size

17 mesh is standard(others available)

Accommodates patches up to 90mm.

METHOD(Paddle over disk)

This method is used for testing the release of drugs from transdermal products.

The apparatus consists of a sample holder or disc assembly that holds the product.

The entire preparation is placed in a dissolution flask filled with specified medium maintained at 32ºC.specified medium maintained at 32ºC.

The paddle is placed directly over the disc assembly.

The disk assembly holds the system flat and is positioned such that release surface is placed parallel with the bottom of the paddle blade. Vessel is covered to minimize evaporation during test.

Samples are drawn midway between the surface of dissolution medium and the top of the paddle blade at specified times.

APPARATUS-6(ROTATING CYLINDER)

DESIGN:

Vessel:- In place of basket, cylinder is used.

Shaft :-Stainless steel 316

Sample :- Mounted to cuprophan (inner porous cellulosic material)

an entire system adheres to cylinder.an entire system adheres to cylinder.

- Dosage unit is placed in cylinder and release from side out.

Water-bath: maintained at 32±0.5°C

USE:

Transdermal patches cannot be cut into small size.

Solid dosage forms, pH profile , small volumes

METHOD( Rotating cylinder):

Use the assembly from apparatus 1 except to replace the basket and shaft with a stainless steel cylinder stirring element.

The temperature is maintained at 32±0.5°C.

The dosage unit is placed on the cylinder with side out .

The dosage unit is placed to the exterior of the cylinder such that long axis of the system fits around the circumference of the cylinder long axis of the system fits around the circumference of the cylinder and removes trapped air bubbles.

Place the cylinder in the apparatus and immediately rotate at the rate specified in the individual monograph.

Samples are drawn midway between the surface of the dissolution medium and the top of the rotating cylinder for analysis.

Carefully apply theadhesive coated side of thesystem to the exterior ofthe cylinder with the longaxis of the system fittingaxis of the system fittingaround the circumferenceof the cylinder.

Rotating cylinder:

Advantages: -Equipment (apparatus 1)available with the manufacturers can be used with modification as apparatus 6.

Disadvantages:-Large volume of medium is required.

-Drug gets diluted & causes difficulties in analysis

-Difficult to clean the cylinder.

APPARATUS-7(RECIPROCATING-DISK)DESIGN:

Vessel:-Flat bottomed cylindrical vessel

-Volume of dissolution medium

Shaft :

Sample : -Placed on disk shaped holders

Agitation :-Reciprocation

shaft

disk

dissolution medium

constant temp water bath Agitation :-Reciprocation

-Reciprocating frequency 30 cycle/sec

Water-bath:-Maintain at 32±0.5°C

USE:

Non-disintegrating, oral modified-release dosage forms, stents, and implants

Transdermal dosage forms.

METHOD(Reciprocating disk):

The assembly consists of a set of volumetrically calibrated solution containers made of glass or suitable inert material, a motor , a drive assembly used to reciprocate the system vertically.

The samples are placed on the disk shaped holders using cuprophan supports

The test is carried out at 32°C.

The reciprocating frequency is 30cycles/min. The reciprocating frequency is 30cycles/min.

Advantages:-Convenient method for selecting the volume of the

medium. Modifications can been made to accommodate

300 mL vessels

-sink conditions can be maintained.

Disadvantages: -Investment is high because the design is totally different from standard equipment already available in industry.

UNOFFICIAL METHODS

1.ROTATING/STATIC DISK METHOD

Developed by late Eino nelson and described by Levy and Sahli.

In this method ,the drug is compressed in a non-disintegrating disc without excipients.

The disc is mounted in a holder so that only one face of the disc is exposed to the dissolution medium.exposed to the dissolution medium.

The holder and disc are immersed in medium and held in a fixed position as in static disc method and rotated at a given speed in rotating disc method.

Samples are collected at predetermined times.

Surface area of the drug through which dissolution

occurs is kept constant –intrinsic dissolution rate.

2.BEAKER METHOD:

Reported by Levy and Hayes(1960).

Dissolution medium, 250ml of 0.1N HCl at 37°C placed

in a 400ml beaker.

Agitation by three blade polyethylene stirrer, 5cm diameter and rotates at 60 rpm.

Stirrer immersed to a depth of 2.7 cm in medium and in the center.

Tablets are placed in a beaker and test was carried out.Tablets are placed in a beaker and test was carried out.

Samples are removed and assayed for the content.

3.FLASK STIRRER METHOD

Developed by Poole(1969).It includes RBF and a stirring element similar to that of beaker method.

RBF used to avoid the formation of moulds of particles in different positions on the flat bottom of a beaker.

4.PERISTALSIS METHOD:

To stimulate hydrodynamic condition of GIT tract in an in-vitro dissolution device.

It consists of rigid plastic cylindrical tubing fitted with septum and rubber stopper at both ends.

Dissolution chamber consists of a space between septum and lower stopper.

Dissolution medium is pumped with peristaltic action through Dissolution medium is pumped with peristaltic action through the dosage form.

5.ROTATING BOTTLE METHOD:

It consists of rotating rack to hold sample drug products in bottles and they are capped tightly & rotated in 37°C temperature bath.

Sample are decanted through a 40 mesh screen and residue are assayed.

6.DIALYSIS METHOD:

Cell consist of 32mm inflated membrane.

Plugged at the lower end by tight fitting cylindrical perspex box.

Upper end of the tube held by thin perspex ring inserted into the tube and secured by an elastic band.

The cell suspended , from the arm of the tablet disintegration apparatus and containing the dosage form in 150ml of distilled water at 37°C.water at 37°C.

The cell is raised or lowered 30times a min, into 150ml of distilled water at same temperature.

Agitation by slight flexing and stretching of the dialysis membrane as it enters and leaves the bath. Rotated at 60rpm.

.

7.DIFFUSION CELL

Static or flow through diffusion cells are used to characterize in-vitro drug release and drug permeation kinetics from a topical drug product e.g.: Ointment, cream or transdermal drug product.

The Franz diffusion cell is static diffusion system used to characterize drug permeation through skin model.

The skin is mounted on the Franz diffusion cell and the drug product is placed on the skin surface.product is placed on the skin surface.

The drug permeates across the skin into a receptor fluid compartment that may be sampled at various times.

This system is used for selection of appropriate formulation that has optimum drug delivery.

Diffusion cell

Pre-requisitePre-requisite

Instrument should be calibrated /Operational Checks ( document each time of use):

•Basket/shaft examination

•Paddle examination

•Vessel examination

•Vessel Temperature

•Vibration

Analytical balance must be calibrated Analytical balance must be calibrated

pH meter should be calibrated

Pipettes and other glassware's

Volumetric accuracy : ± 1%

Deareation of dissolution medium.

Lighting should be sufficient to perform visual observations.

Calibration of Dissolution Apparatus

Why ?

• To confirm suitability of the equipment and proper operation of the apparatus

How ?How ?

• Mechanical calibration (verification of physical parameters)

• Chemical calibration (Apparatus Suitability Test – USP)

When ?

• Before using new test equipment

• After relocation or major maintenance

• At regular intervals (Every 6 months)

Current Harmonized Physical Parameters and Tolerances USP Mechanical Calibration Parameters include:

Basket/Shaft Wobble (No significant wobble)

Vessel/Shaft Centering (2 mm from centerline)

Height check/Basket or Paddle Depth as measured at basket bottom or Paddle bottom (25 + 2 mm)bottom or Paddle bottom (25 + 2 mm)

No significant vibration

Rotational speed (+ 4%)

Vessel Temperature (37.0 + 0.5 C)

Basket Wobble (bottom rim) (+ 1mm)

USP Performance Verification Test (PVT) Official since 1978, USP Calibration with Prednisone and Salicylic

Acid has been the means of qualifying the dissolution apparatus.

Initially, the primary purpose was to indicate environmentaleffects on the apparatus and vibration since most other parameterscould be controlled by mechanical measurements

The original test was called “Calibration” which was not a trueindication of the test being performed, later changed to“PerformanceVerification” (PVT)

The PVT, has been responsible for detecting problems associatedwith dissolution apparatus that are found to be within mechanicaltolerances

Calibrator Tablets

1970’s : USP Calibrator Tablets Introduced

Disintegrating – 50 mg Prednisone (Upjohn)

Non Disintegrating – 300 mg Salicylic Acid (Hoffman LaRoche)

1997 : 50 mg Prednisone replaced with 10 mg Prednisone manufactured at University of Marylandmanufactured at University of Maryland

2004 : USP begins search for replacement for 10 mg Prednisone tablet

USP: Both Calibrators on a given apparatus (i.e. 4 calibration tests if instrument is used for paddle and basket methods)

JP, BP and EP: No calibrator tablets

Allowable Variations

A basket with gold coating 2.5 μm thick (0.0001 inch) is an allowable variation of the standard 40-mesh basket.

Some changes can be made to the compendial apparatus; for example, a basket mesh size other than the typical 40-mesh basket (e.g., 10-, 20-, or 80-mesh)basket (e.g., 10-, 20-, or 80-mesh)

Larger vessels accommodating up to two and four liters are now allowable variations in the USP. Such vessels are advantageous for poorly soluble drugs.

For example, a small-volume apparatus with mini paddles and baskets may be considered for low-dosage strength products.

De-areation Air bubbles can act as a barrier to the dissolution process

if present on the dosage unit or basket mesh and canadversely affect the reliability of the test results.

Furthermore, bubbles can cause particles to cling to theapparatus and vessel walls.apparatus and vessel walls.

Bubbles on the dosage unit may increase buoyancy, leadingto an increase in the dissolution rate, or may decrease theavailable surface area, leading to a decrease in the dissolutionrate.

Poorly soluble drugs are most sensitive to interferencefrom air bubbles, therefore, deaeration may be needed.

Degassing as per USP

Prepare dissolution media and

properly deaerate.

USP Method:USP Method:

Heat media to 41°C, vacuumfilter through 0.45μm filter,continue to pull vacuum for 5additional minutes

Common Degassing Methods

Acceptable Methods

USP Vacuum Filtration Method (default unless another approach is validated)

Helium Sparging*

Automated Degassing*

Unacceptable Methods

Nitrogen Sparging

Sonication

Automated Degassing*

Superheating*

Not Degassing At All*

*when validated against USP method

Method DevelopmentMethod Development

Dissolution Method Goals

A successful dissolution method will be:

Discriminatory

Robust and Rugged

Correlated to In Vivo Correlated to In Vivo

Transferrable

Controlled Variability

When to develop method? When to develop method?

Discrimination

Discrimination in Dissolution simply means being able to tell the difference between good difference between good and bad formulations

Development of a Discriminating MethodThe procedure should be capable of distinguishing significantchanges in composition or manufacturing process that mightbe expected to affect in vivo performance.

Factors to consider:

Qualitative and quantitative excipient changes Qualitative and quantitative excipient changes

Manufacturing parameters:

Lubrication

Blend time

Compression force

Drying parameters

Steps for dissolution method development

Literature survey : Drug and drug product knowledge

Selection of Apparatus

Selection of dissolution medium

Optimization of dissolution parameters Optimization of dissolution parameters

Development of suitable analytical method for estimation of content

Validation of analytical method

Knowledge of Drug and Drug Product

Characteristics of the API e.g., particle size, crystal form,bulk density, solubility

Product composition e.g., drug loading /dose, and theidentity, type, and levels of excipients

Manufacturing process e.g., compression forces, Manufacturing process e.g., compression forces,equipment

Effects of stability storage conditions e.g., temperature,humidity

Incompatibility of the drug with certain buffers or salts

Dissolution Tester Choice Paddles and Baskets tend to be the choice for most solid

oral dosage forms.

If pH changes, greater/smaller volumes, or different agitation is needed then Apparatus 3 and 4 are often considered after exhausting Paddle and Basket testing

For Transdermals Apparatus 5-7 are the primary choices.

For semisolid dosage forms, the generally used apparatus include the vertical diffusion cell, apparatus include the vertical diffusion cell, immersion cell, and flow-through cell apparatus with the insert for topical dosage forms

A rotating bottle or dialysis tubes may have utility for microspheres and implants; peak vessels for eliminating coning; and modified flow-through cells for special dosage forms including powders and stents.

Agitation Rate Should be sufficient to allow for

media to interact with dosage form

Decreasing or increasing the apparatus rotation speed may be justified if to achieve an in-vitro–in-vivo correlation (IVIVC)

For better discrimination

Too much agitation can result innon-discriminatory profiles

•Baskets – 50-100 RPM

•Paddles – 25-100 RPM

USP.APPARATUS DESCRIPTION ROT.SPEED DOSAGE FORM

TYPE 1 Basket apparatus 50-100 rpm IDR,DR,ER

TYPE 2 Paddle apparatus 50-75rpm25-50rpm

IDR,DR,ERSuspension

TYPE 3 Reciprocatingcylinder

5-30 dips/min IDR,ER

TYPE 4 Flow through cell 2-50ml/min ER,Poorly soluble API

TYPE 5 Paddle over disk 25-50 rpm TRANSDERMAL

TYPE 6 Rotating cylinder N/A TRANSDERMAL

TYPE 7 Reciprocatingholder

30 rpm ER

The Ideal Dissolution Media

•Meets sink conditions

•Simple preparation

•Drug is Stable in media 24 hrs+

•Uses as little extras as possible possible

–Surfactants

–Alcohol

•Biologically relevant for site of dissolution in vivo

–IR typically in acid

–DR typically in acid, then neutral

–MR typically in neutral solution

Determination of Intrinsic Dissolution

Intrinsic dissolution is defined asthe dissolution of a pure drugsubstance from a specified constantsurface area.

A special punch and die is used tocompress pure drug substances intoa disk or tablet.a disk or tablet.

The disk is placed into a specialholder that allows only one flatsurface to come in contact with adissolution media at any time

during the test.

Application of IDIntrinsic Dissolution

Intrinsic dissolution data is generally used in drugscreening but can provide helpful solubility informationfor method development.

Rates lower than 0.1 mg/min/cm2 generally mean that Rates lower than 0.1 mg/min/cm2 generally mean thatbioavailability will be determined by the dissolutionrate.

Rates higher than 1 mg/min/cm2 mean thatbioavailability will most likely be determined by thedrug permeability.

pH Dependent Solubility Solubility screen in multiple media (pH 1.2-7.5) should be

done to determine optimal solubility.

If needed, use as little surfactant as necessary.

Evaluate multiple surfactants (pay attention to grades and vendors)

Investigations of the stability of the drug substance should Investigations of the stability of the drug substance should be carried out, when needed, in the selected dissolution medium with excipients present, at 37°.

Effect of pH on solubility and stability need to be evaluated.

Sink condition and Solubility Sink condition : Volume of medium at least three times that required in

order to form a saturated solution of drug.

Solubility of the drug substance is usually evaluated by determining the saturation concentration of the drug in different media at 37° using the shake-flask solubility method (equilibrium solubility). Alternative methods for solubility determination may also be used.

In the absence of sink conditions, investigate methods to enhance solubility, e.g. use of a surfactant. If a surfactant is used, its concentration should be properly justified (e.g. typically <2% Sodium Lauryl Sulfate (SLS)).

In certain cases, it may be necessary to evaluate the solubility of the drug at room temperature.

The pH of the clear supernatant should be checked to determine whether the pH changes during the solubility test.

Rules of Thumb for Media Limits Surfactants below 1% tend to be accepted with appropriate checks that lower

limits aren’t acceptable

>1% require greater scrutiny, other surfactants usually

>1.5% tends to be very difficult to handle with automation

Alcohol is generally a last resort – unless doing a dose dumping study specifically.specifically.

Stay within pH 1.1 – pH 7.5 if at all possible

The use of surfactants needs to be justified by data that show lowsolubility in the aqueous media. The chosen concentration of surfactantalso needs to be justified by providing dissolution profiles in mediacontaining the surfactant at concentrations higher and lower than the chosenconcentration.

USP Surfactants

Use of enzymes The use of enzymes in the dissolution medium is permitted,

in accordance with general chapter Dissolution 711 ,whendissolution failures occur as a result of cross-linkingwith gelatin capsules or gelatin-coated products.

A discussion of the phenomenon of cross-linking and methoddevelopment using enzymes can be found in proposedgeneral information chapter Capsules–Dissolution Testing andRelated Quality Attributes 1094 .

Antifoaming Agents and antioxidants The hydrodynamics are influenced by the cylinder's

reciprocating motion and the resulting movement of thesample in the medium. The reciprocating motion of thecylinder and screen may cause foaming if the mediumcontains surfactants. Addition of an anti-foaming agent suchcontains surfactants. Addition of an anti-foaming agent suchas simethicone or n-octanol may be useful for avoidingfoaming from surfactants.

In some cases, antioxidants such as ascorbic acid may be used in the dissolution medium to stabilize the drug.

Media Cautions Be careful with water

–No buffering capacity

–Quality can differ b/w sites

–Quality can differ b/w DI systems, filters, etc.

Check pH before and after run to ensure buffering capacity is Check pH before and after run to ensure buffering capacity is acceptable

Beware of methods needing tight pH limits

Do not use SLS with Potassium Phosphate Buffers –Sodium Phosphate only

Bio-relevant Dissolution Medium Bio-relevant media is the media that represent the

conditions same as that of the in-vivo condition.

The fed and fasted state may have significant effects on the absorption or solubility of a compound.

Composition of media that simulate the fed and fasted Composition of media that simulate the fed and fasted condition is necessary to establish in-vivo in-vitro correlations.

This media reflect changes in the pH, bile concentration and osmolarity after meal intake and therefore have a different composition than that of typical compendial media.

Why the Need for Better Biorelevant Information Better IVIVC

Fewer Clinical Trials

Better Predictive tools

Shorter Development CyclesCycles

Choice of bio-relevant media is usually based upon: A mechanistic approach that considers the absorption site, if

known.

Whether the rate-limiting step to absorption is the dissolutionor permeability of the compound.

Fed and fasted states may have significant effects on theFed and fasted states may have significant effects on theabsorption or solubility of a compound.

These media are primarily used to establish in vitro-in vivocorrelation during formulation development and to assess potentialfood effects; they are not always intended for Quality Controlpurposes.

Selection of Dissolution medium Primary requirement for selection of dissolution media is

that, it should be able to reflect in vivo situations when it isused to establish an IVIVC.

For Class I and III drugs, use of simple aqueous media such asSGF without enzymes or SIF without enzymes isSGF without enzymes or SIF without enzymes isrecommended.

For Class II and III drugs, use of biorelevant media fordissolution testing is recommended. They are: 1) SGF plussurfactant 2) Milk with 3.5 % fat to stimulate fed statecondition 3)FaSSIF is used for poorly soluble drugs.

Temperature of Dissolution Medium

The standard temperature- 37±0.5 °C for oral dosage form .

Slightly increased temperatures such as 38±0.5 °C have been recommended for dosages forms such as suppositories.

Lower temperatures such as 32±0.5 °C are utilized for Lower temperatures such as 32±0.5 °C are utilized for topical dosage forms such as trans-dermal patches and topical ointments.

Media temperature readings must be taken at least twice during the dissolution test, at the start and end of a test.

Appropriate Test Duration Time points should be selected to adequately characterize the

ascending and plateau phases of the dissolution curve.

Typically:

Immediate release, 85% in <15 minutes, one time point

Immediate release, 15, 20, 30, 45, 60 minutes

For some products, including suspensions, useful information mayFor some products, including suspensions, useful information maybe obtained from earlier points, e.g., 5–10 min.

Delayed release,Acid 1-2 hours, Buffer +30 minutes

Extended release, minimum of three points

– Initial (1-2 hours) to show potential dose dumping

– Intermediate point to define similar in vivo profile

– Final point to show that essentially complete release (>80%) of thedrug is achieved.

When to add dosage form?

Tablet introduction may beperformed manually orautomatically

Dosage forms may beintroduced simultaneously orintroduced simultaneously orsequentially but they must beintroduced into non-rotating media.

Evaporation cover shown withDosage Delivery Module(DDM)

How to handle sample?

Always handle dosage unitswith gloves (not cotton),forceps or tweezers whichwill not scratch or damagethe surface of the dosagethe surface of the dosageunit.

Examine the six dosageunits.

Do not use chipped,cracked or capped tablets

Weighing of the sample?

Option: Record the dosageunit weights?

Weight is forinformation andinvestigation purposesonly.

Dosage units are to bechosen at random and maynot be selected ordiscarded based on weight.

Sinkers (USP Apparatus 2) In addition to sinking, floating dosage forms, sinkers may

assist in keeping a dosage form from sticking to the vesselinappropriately as in the case with some film coated tablets.

Sinkers Specifications must be adequately described inthe method to eliminate hydrodynamic variation associatedwith different sinker devices.

Sinkers may be fabricated by wrapping around a corkbore.

Should not be too tight – will restrict any disintegratingrelease mechanism

Too loosely, the dosage form may escape soon after the testbegins.

USP Recommendations about sinkers

Examples of typical observations Uneven distribution of particles throughout the

vessel

1. Particles cling to the sides of the vessel,

2. coning or mounding directly under the apparatus, e.g., below the basket or paddle,below the basket or paddle,

3. Particles float at the surface of the medium,

4. Film-coated tablets stick to the vessel

Examples of typical observations Air bubbles on the inside of the vessel or on the

apparatus or dosage unit

1. Air bubbles on the inside of the vessel or on the apparatus or dosage unit.

2. Sheen on the apparatus is also a sign of air bubbles.2. Sheen on the apparatus is also a sign of air bubbles.

This observation would typically be made when

assessing the need to deaerate the medium

Examples of typical observations Dancing or spinning of the dosage unit, or the dosage

unit being hit by the paddle.

Adhesion of particles to the paddle or the inside of thebasket, which may be observed upon removal of the

stirring device at the end of the run.stirring device at the end of the run.

Pellicles or analogous formations, such as transparentsacs or rubbery, swollen masses surrounding the capsulecontents.

Examples of typical observations Observation of the disintegration rate (e.g.,

percentage reduction in size of the dosage unit within a certain time frame).

Complex disintegration of the coating of modified or enteric-coated products. modified or enteric-coated products.

Whether the dosage form lands in the vessel center or off-center, and if off-center, whether it sticks there.

Time required for the complete dissolution of the capsule shell or for tablet disintegration.

Usual Remedies Visual observations

Changing any of the followingfactors:

1) Apparatus type,

2) Speed of agitation,

3) Level of deaeration,3) Level of deaeration,

4) Sinker type, or

5) composition of the medium.

For dosage forms that exhibitconing (mounding) under thepaddle at 50 rpm, the coning can bereduced by increasing the paddlespeed to 75 rpm

How To Sample Properly?

Filtration must occur at USP location and at appropriate time

±2 % from time point or 15 minutes (whichever less)

Halfway between top of paddle Halfway between top of paddle or basket and media

No closer than 1cm to vessel wall

Recommend not sampling close to shaft due to poor hydrodynamics

Acceptable Method Requirements

Low variability (<20% at initial time point,<10% at later points)

Complete Release (85%+ or Asymptote)

Proper understanding of dissolution release

Discrimination between batches -Challengedwith other formulations

Robust/Rugged/Reproducible results

SpecificationsSpecifications

Specification Parameters -Dissolution Dissolution is considered product-specific.

The method and limits should be appropriate for the proposed product.

It is useful to have the parameters (medium, apparatus, speed) in specs.specs.

Dissolution specs at release and shelf-life should be identical.

Surfactant use should be exceptional and appropriate.—not exceed 2% normally

Dissolution Acceptance Criteria

Q –Value :

Define as a percentage of drug content dissolved in a given time period.

It is commonly used in the USP for immediate release and It is commonly used in the USP for immediate release and delayed release dosage forms.

The quantity of Q is the amount of dissolved activeingredient specified in the individual monographexpressed as a percentage of the labeled content.

Q - Value For highly soluble and rapidly dissolving drug products (BCS

classes 1 and 3), a single-point dissolution test specification of NLT 85% (Q=80%) in 60 minutes or less is sufficient as a routine quality control test for batch-to-batch uniformity.

For slowly dissolving or poorly water soluble drugs (BCS class 2), a two-point dissolution specification, one at 15 minutes to include a dissolution range (a dissolution window) and the other at a later point (30, 45, dissolution range (a dissolution window) and the other at a later point (30, 45, or 60 minutes) to ensure 85% dissolution, is recommended to characterize the quality of the product.

For products containing water insoluble APIs, it is recommended to have a two tire dissolution limit. For example Artemether dissolution: NLT 40% in 1 hour and NLT 60% at the 3rd hour.

Immediate Release Forms – AcceptanceTable USP <711>

STAGE No. of Dosage units tested

Acceptance criteria

S1 6 Each unit is ≥ Q + 5%.

S2 6 Average of 12 units (S1 + S2) is ≥Q and no unit is < Q – 15%.unit is < Q – 15%.

S3 12(6+6+12=24) Average of 24 units (S1 + S2 + S3) is ≥ Q,not more than 2 units are < Q – 15% and no single unit is less than Q – 25%.

99

If a sample fails either Stage S1 or S2, proceed to the next stage and testthe number of units indicated.

Delayed Release Forms – Method AUSP <711> USP Delayed Release Dosage Forms – Method A:

Media is 750 mL of 0.1N HCl

Samples are removed for analysis after 2 hours ± 2%

Within 5 minutes of withdrawing the acid stage sample aliquots, add 250 mL of 0.20-M tribasic sodium phosphate,

aliquots, add 250 mL of 0.20-M tribasic sodium phosphate, adjusting to pH + 0.05 at 37 °C while stirring at the specified rate.

Dissolution continues for 0.75 h (or per monograph) or less.

Sample aliquots are then analyzed with Q being the total % dissolved for both acid and buffer stages.

Delayed Release Forms – Acid Stage Acceptance Table USP <711>

STAGE No. of Dosage units tested

Acceptance criteria

A1 6 No individual value exceeds 10%dissolved.

A2 6 Average of 12 units (A1 + A2) is no morethan 10% dissolved and no single unit isthan 10% dissolved and no single unit isgreater than 25% dissolved.

A3 12(6+6+12=24) Average of 24 units (A1 + A2 + A3) is notmore than 10% dissolved and no singleunit is more than 25% dissolved.

101

If a sample fails either Stage A1 or A2, proceed to the next stage and testthe number of units indicated.

Delayed Release Forms - Buffer Stage Acceptance Table USP <711>

STAGE No. of Dosage units tested

Acceptance criteria

B1 6 Each unit is ≥ Q + 5%.

B2 6 Average of 12 units (B1 + B2) is ≥ Q and no unit is < Q– 15%.– 15%.

B3 12(6+6+12=24) Average of 24 units (B1 + B2 + B3) is ≥ Q, not morethan 2 units are < Q – 15% and no single unit is less thanQ – 25%.

102

If a sample fails either Stage B1 or B2, proceed to the next stage and testthe number of units indicated.

Delayed Release Forms – Method B USP <711> USP Delayed Release Dosage Forms – Method B:

Media is 1000 mL of 0.1N HCl

Samples are removed for analysis after 2 hours ―

USP Delayed Release Dosage Forms – Method B:

Drain the acid from the original vessel and replace with

Drain the acid from the original vessel and replace with

1000mL pH6.8 phosphate buffer pre equilibrated at 37°C

or switch the dosage form to a second vessel containing the phosphate buffer

Disssolution specification – Case study -1 Ethambutol hydrochloride 400mg tablets. The applicant

claimed USP standard for the product.

Bioequivalent of the product is accepted as per BCS class 3 based biowavier

The applicant set the dissolution specification limits as below: The applicant set the dissolution specification limits as below:

NLT 80% (Q) in 45min at release

NLT 75% (Q) in 45min at shelf life, which is in line with the requirement of USP monograph.

Is it acceptable? What limits should be applied?

Disssolution specification – Case study -1Answer: not acceptable

The dissolution limits at release and shelf life should be the same.

A limit of NLT 80% (Q) in 15min should be set for A limit of NLT 80% (Q) in 15min should be set for both release and shelf life as for the BCS class 3 biowaiver.

Comparison of Dissolution Profiling

Why Profiling / Comparision?

It reflects its release pattern under the selected condition sets. i.e. either sustained release or immediate release of the formulated formulas.

For optimizing the dosage formula by comparing the dissolution profiles of various formulas of the same drug.of various formulas of the same drug.

Dissolution profile comparison between pre change and post change products for SUPAC (scale up post approval change ) related changes or with different strengths, helps to assure the similarity in the product performance and green signals to bioequivalence.

Modify formulation as per Brand Leader.

Requirement for Comparative dissolution testing

Two or more products or batches containing the same APIare compared

The strength of products / batches may or may not be the same (depending on purpose of test)

The dissolution conditions are similar, e.g.

Apparatus, medium, volume, rotation speed & temp.• Apparatus, medium, volume, rotation speed & temp.

• Minimize possible experimental differences in conditions

Samples are taken at the same time points and the data (dissolution profiles) compared

How to Compare? Aim: To show similar in vitro dissolution under physiologically

relevant experimental pH conditions.

Advisable to investigate more than one batch of test and referenceproducts; must include bioequivalence batches.

Investigate within pH 1-6.8 (normally pH 1.2, 4.5 and 6.8) and QC media

(if different).

Water may be used as an additional medium, especially when the APIWater may be used as an additional medium, especially when the APIis unstable in buffered media to the extent that data is unusable.

Additional investigations may be required at pH values in which thedrug has minimum solubility.

Use12 units to enable statistical evaluation

For each condition, present comparative dissolution profiles(mean valuesvs.time) together with statistics (max, min, mean, RSD; f2 similarity factor ifcalculated; individual values)

METHODS TO COMPARE DISSOLUTION PROFILEMETHODS TO COMPARE DISSOLUTION PROFILE

Graphical methodGraphical method Statistical Analysis

Statistical Analysis

Model Dependent method

Model Dependent method

Model Independent Method

Model Independent Method

t-Testt-Test ANOVAANOVA

Zero Zero First First Hixson-Hixson- Higuchi Higuchi Weibull Weibull Korsemeyar Korsemeyar Baker-Baker-Zero order Zero order

First orderFirst

orderHixson-crowell

law

Hixson-crowell

law

Higuchi model

Higuchi model

Weibull model

Weibull model

Korsemeyar and peppas model

Korsemeyar and peppas model

Baker-Lonsdale

model

Baker-Lonsdale

model

Ratio Test ProcedureRatio Test Procedure

Pair Wise ProcedurePair Wise Procedure

Multivariate Confidence Region Procedure

Multivariate Confidence Region Procedure

Index of RescignoIndex of Rescigno

Paired Wise Procedure DIFFERENCE FACTOR (f1) & SIMILARITY FACTOR (f2)

The difference factor (f1) as defined by FDA calculates the % difference between 2 curves at each time point and is a measurement of the relative error between 2 curves.

n

TtRtf1 = × 100

where, n = number of time points

Rt = % dissolved at time t of reference product (pre change)

Tt = % dissolved at time t of test product (post change)

n

t

t

Rt

TtRt

1

1

The similarity factor (f2) as defined by FDA is logarithmic reciprocal square root transformation of sum of squared error and is a measurement of the similarity in the percentage (%) dissolution between the two curves

f2 = 50 ×

100

1

log )(1

1

5.0n

r

TtRtwtn 1rn

Guidance for Industry

A specific procedure to determine difference and similarity factors is as follows:

1. Determine the dissolution profile of two products (12 units each) of the test (postchange) and reference (prechange) products.

2. Using the mean dissolution values from both curves at each time interval, calculate the difference factor (f1 ) and similarity factor (f2) using the above equations.equations.

3. For curves to be considered similar, f1 values should be close to 0, and f2 values should be close to 100. Generally, f1 values up to 15 (0-15) and f2 values greater than 50 (50-100) ensure equivalence of the two curves and thus, of the performance of the test (postchange) and reference (prechange) products.

4. This model independent method is most suitable for dissolution profile comparison when three to four or more dissolution time points are available.

The following recommendations should also be considered:

The dissolution measurements of the test and reference batches should bemade under exactly the same conditions.

The dissolution time points for both the profiles should be the same (e.g., 15, 30, 45, 60 minutes).

The reference batch used should be the most recently manufactured prechange product.

A minimum of three points required for comparison of profile. Only one measurement should be considered after 85% dissolution of both the Only one measurement should be considered after 85% dissolution of both the

products. To allow use of mean data, the percent coefficient of variation at the earlier

time points (e.g., 10 minutes) should not be more than 20%, and at other time points should not be more than 10%.

The mean dissolution values for R can be derived either from (1) last prechange (reference) batch or (2) last two or more consecutively manufactured prechange

batches.

Example Determination of similarity of profiles

Example 1-B

% API dissolved

Time(min)

Tablet D(Ref)

Tablet E(Test)

10 55 57

15 72 78

Example 1-A

% API dissolved

Time(min)

Tablet A(Ref)

Tablet B(Test)

10 87 94

15 96 99 15 72 78

20 85 91

30 97 100

45 102 100

60 103 101

f2 required? Yes

f2 (n = 3 ?) 64 (similar)

15 96 99

20 99 99

30 100 99

45 101 99

60 101 99

f2 required? No, ≥ 85% in 15 min

f2 (n = N/A ?) profiles similar

ExampleDetermination of similarity of profiles (cont.)

Example 1-D

% API dissolved

Time(min)

Tablet A(Ref)

Tablet Y (Test)

10 87 55

15 96 72

Example 1-C

% API dissolved

Time(min)

Tablet X(Ref)

Tablet Y(Test)

10 29 34

15 38 41 15 96 72

20 99 85

30 100 97

45 101 102

60 101 103

f2 required? Yes

f2 (n = 3 ?) 31 (not similar)

15 38 41

20 47 50

30 63 64

45 80 79

60 95 91

f2 required? Yes

f2 (n = 6 ?) 74 (similar)

Comparison to reference medicinal product Immediate Release Tablets/Capsules

> 85% dissolved within15 minutes : test and reference similar without any further calculationcalculation

≤ 85 % dissolved within15minutes: calculate f2 similarity factor

Modified Release Preparations

Calculate f2 similarity factor

Comparison to reference medicinal product

Prolonged Release Preparations

Minimum of 3 time points, but may be prudent to do more

Particularly important if desired release profile not uniform Particularly important if desired release profile not uniform

(e . g. immediate release outer coat and prolonged release core)

If only3 time points: expected to mirror final specification timepoints (20-30%(dose dumping),50%(defines profile),>80%)

More time points =↑confidence of bioequivalence

e.g. for once daily preparations(1,2,4,8,12,16,20 & 24 hours)

Alternative methods to dissolution testing

In ICH Q6A permits use of disintegration testing as a surrogatefor conventional Compendial dissolution tests, provided

highly soluble drug substances

intrinsic rate of solubilization is rapid intrinsic rate of solubilization is rapid

overall drug release rate is dominated by cohesive propertiesof the formulation

DT in place of dissolution?ICH Q6A decision trees #7 can be used to assess the proposed dissolution criteria, however:

Highly soluble throughout physiological pH range.

Solubility at 37°C ± 0.5°C, dose + solubility < 250 ml, pH 1.2 - 6.8.

For considering /accepting DT in place of dissolution: all the considerations For considering /accepting DT in place of dissolution: all the considerations should be carefully assessed: highly soluble and very rapidly dissolving, plus significant supporting development data – including

when DT is more discriminating or

has a demonstrated relationship to dissolution, robustness of the formulation/manufacturing process have been demonstrated wrt DT, etc.

Dissolution may not be necessary or proposed as a skip test.

Example

APIs with good solubility at gastric pH levels may begranted BCS Class I and III classification i.e. may becharacterized by disintegration testing alone.

In liquid filled capsules, drug dissolved in solubilisation aidsoffering a true mechanism for drug release is likely to be therupture of the capsule use disintegration as a surrogate forthe QC dissolution test

Bio vaiver (In vitro equivalence testing)

The term biowaiver is applied to a regulatory drug approvalprocess where the efficacy and safety part of the dossier(application) is approved based on evidence of equivalence otherthan through in vivo equivalence testing.

A biowaiver can be applied only for products which meetrequirements on pharmaceutical similarity, as well as similarity inrequirements on pharmaceutical similarity, as well as similarity incomparative dissolution tests.

A BCS-based biowaiver has become an important andcost-saving tool in approval of generic drugs.

The bio-relevance of the BCS properties and the in vitro releaseare best expressed through a correlation between in vitro and invivo data.

BCS Classification

Class 1 Class 2 Class 3 Class 4

Highly Poorly Highly Poorly Highly Soluble

Poorly Soluble

Highly Soluble

Poorly Soluble

Highly Permeable

Highly Permeable

Poorly Permeable

Poorly Permeable

‘High solubility

The definition of ‘high solubility’ refers to the highest dosestrength of an immediate release product, which has to besoluble in 250 ml or less of aqueous media over the pHrange of 1 – 7.5,. Solubility measurements should berange of 1 – 7.5,. Solubility measurements should beperformed at 37 °C using a stability indicating, validatedmethod.

High permeability

The classification regarding high permeability refers to the extent of absorption in humans, i.e. high permeability is concluded if the extent of absorption in humans reaches at least 90 % of an orally administered dose. reaches at least 90 % of an orally administered dose. This conclusion may be based on either pharmacokinetic studies in humans (e.g. mass balance, or absolute bioavailability studies) or intestinal permeability methods like e.g. in vivo intestinal perfusion studies in humans or validated in vitro permeation studies across a monolayer of cultured epithelial cells.

Requirement

Comparative in vitro dissolution investigations should ensure that no less than 85 % of the labeled amount is dissolved within 30 min in each of the required media: 0.1 N HCl, pH 4.5 and 6.8 buffers. Regarding experimental requirements, reference is made to the US Pharmacopoeia and the US-FDA guidance for industry on Dissolution Testing of Immediate Release Solid Oral Dosage Forms (August 1997) 3. Oral Dosage Forms (August 1997) 3.

Resulting profiles should be compared using the similarity factor (f2), unless 85% or more of the labeled amount dissolves within 15 min from both products.

The latter case would allow the conclusion that the investigated products are similar without requiring any further statistical calculations.

Restrictions

BCS based waivers are not applicable for the initial in vivo bioavailability characterization for NDAs.

Other restrictions of application include:

1) narrow therapeutic index (NTI) drug; and 1) narrow therapeutic index (NTI) drug; and

2) drug products intended to be absorbed in the oral cavity.

3) Similarly, prodrugs and excipients require special consideration. In case of prodrugs, whether to measure the prodrug or the drug for permeability determination will depend on where the conversion occurs.

4

Comparison of FDA, EU and WHO guidance on BCS based biowaiver.

Parameters US-FDA EU WHO

Allowed classes 1 1 and 3 1, 2 (weak acids), and 3

High solubilityHighest strength completely dissolved in 250mL of aqueous media at 37 ˚C ± 1 ˚C.Highest strength completely dissolved in 250mL of aqueous media at 37 ˚C ± 1 ˚C.

pH range pH 1-7.5 pH 1-6.8 pH 1.2-6.8

High permeability >90% absolute BA ormass balance study

>85% absolute BA or mass balance study

Comparison of FDA, EU and WHO guidance on BCS based biowaiver.Parameters FDA EU WHO

Rapid dissolution

Media (studies should beconducted at 37 ±

900 mL or less aqueousmedia (0.1N HCl

900 mL or less aqueousmedia (pH 1.0-1.2

900 mL or less aqueousmedia (pH 1.2conducted at 37 ±

1 C)media (0.1N HClorSGF; pH 4.5 buffer; andpH 6.8 buffer or SIF)

media (pH 1.0-1.2buffer, usually 0.1NHCl or SGF; pH 4.5buffer; and pH 6.8buffer or SIF)

media (pH 1.2HCl solution; pH 4.5acetate buffer; and pH6.8 phosphate buffer)

Apparatus (APP) USP APP I - 100 rpmUSP APP II - 50 rpm

Paddle APP - 50 rpmBasket APP - 100 rpm

Paddle APP - 75 rpmBasket APP - 100 rpm

Comparison of FDA, EU and WHO guidance on BCS based biowaiver.Parameters FDA EU WHO

Rapid dissolution

Criteria >85% in 30 min in 3 media

Class 1: >85% in30 min in 3 media

Class 1: >85% in 30 minin 3 media (Rapid)in 3 media 30 min in 3 media

(Rapid)Class 3: >85% in 15min in 3 media (VeryRapid); or, >85% in30 min and similardissolution profile toRLD (Similarly Rapid)

in 3 media (Rapid)Class 2: >85% in 30 minin pH 6.8 medium andsimilar dissolutionprofile in 3 mediaClass 3: >85% in 15 minin 3 media (Very Rapid)

Restrictions Narrow therapeutic drugsOral products intended to be absorbed in the oral cavityModified release drug products

CONCLUSION: In vitro dissolution is the best available tool today which can at

least quantitatively assure about the biological availability of drug.

Great Scope for New Method Development / Pharmacopoeial monographs

Systematic and scientific Approach is needed. Systematic and scientific Approach is needed.

Academic standards / Research ?????? Basic infrastructure / facilities / sophisticated instrument /

positive efforts / attitude and honesty is the need of time to improve the quality of research.

Guidelines Referred BCS Guidance (Waiver of In Vivo Bioavailability and

Bioequivalence Studies for Immediate-Release Solid Oral Dosage Forms Based on a Biopharmaceutics Classification System”); August 2000

IR Dissolution Guidance (Dissolution Testing of Immediate Release Solid Oral Dosage Forms); August 1997

IVIVC Guidance (Extended Release Oral Dosage Forms: IVIVC Guidance (Extended Release Oral Dosage Forms: Development, Evaluation, and Application of In Vitro/In Vivo Correlations); September 1997

General BA/BE Guidance (Bioavailability and Bioequivalence Studies for Orally Administered Drug Products - General Considerations); 2003

WHO, US FDA and EMEA guidelines

ICH guidelines

Thank you