Using Automated Software for Improved Results in GC Triple Quadrupole MS Pesticide Analysis
Jason Cole and Paul Silcock
Thermo Fisher Scientific
![Page 2: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/2.jpg)
2
Triple Quadrupole GC-MS/MS
Fast becoming an essential tool in high-throughput, routine laboratories
• Especially true for laboratories performing pesticides analysis in food
• Becoming more so in environmental analysis
• Mainly driven by the selectivity advantages of MS/MS
Environmental
Food Safety
![Page 3: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/3.jpg)
3
http://books.google.com/ngrams 5.2 million books: ~4% of all books ever published
![Page 4: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/4.jpg)
4
![Page 5: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/5.jpg)
5
GC-MS/MS – What’s So Special?
• Low detection limits
• Optimized sample preparation
• Consolidated analytical methods
• Faster, automated data processing
...it’s a high selectivity technique...
![Page 6: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/6.jpg)
6
Selectivity in a Method
McLafferty circa. 1980
![Page 7: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/7.jpg)
7
Method Performance Requirement
• Target compounds
• Matrices
• Sensitivity
Method performance requirement
![Page 8: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/8.jpg)
8
First - Sample Prep..
Ste
p 1
– e
xtr
ac
tio
n
Ste
p 2
- c
lea
n u
p (
1s
t)
Ste
p 3
- c
lea
n u
p (
2n
d)
Method performance requirement
![Page 9: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/9.jpg)
9
...then Instrument Detection...
Ste
p 1
– e
xtr
ac
tio
n
Ste
p 2
- c
lea
n u
p (
1s
t)
Ste
p 3
- c
lea
n u
p (
2n
d)
Ste
p 4
- G
C-M
S d
etec
tio
n (
sin
gle
qu
ad S
IM)
Total method selectivity
Method performance requirement
![Page 10: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/10.jpg)
10
...What about GC-MS/MS?...
Ste
p 1
– e
xtr
ac
tio
n
Ste
p 2
- c
lea
n u
p (
1s
t)
Ste
p 3
- c
lea
n u
p (
2n
d)
Total method selectivity
Ste
p 4
- G
C-M
S/M
S d
etec
tio
n (
Trip
le q
ua
d S
RM
)
Method performance requirement
![Page 11: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/11.jpg)
11
...Use GC-MS/MS to Reduce Clean-Up...
Ste
p 1
– e
xtr
ac
tio
n
Total method selectivity
Ste
p 2
- G
C-M
S/M
S d
etec
tio
n (
Trip
le q
ua
d S
RM
)
Method performance requirement
![Page 12: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/12.jpg)
12
...Use GC-MS/MS to Consolidate Methods...
Ste
p 1
– e
xtr
ac
tio
n
Total method selectivity
Ste
p 2
- G
C-M
S/M
S d
etec
tio
n (
Trip
le q
ua
d S
RM
)
Method 1performance requirement
Method 2performance requirement
Method 3performance requirement
![Page 13: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/13.jpg)
13
...Use GC-MS/MS to Consolidate Methods...
Ste
p 1
– e
xtr
ac
tio
n
Total method selectivity
Ste
p 2
- G
C-M
S/M
S d
etec
tio
n (
Trip
le q
ua
d S
RM
)
Consolidated multi-residue method
![Page 14: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/14.jpg)
14
...so GC-MS/MS is Special as it Delivers...
High Selectivity
• Possibility to reduce selectivity in sample preparation
• Reduced sample prep steps create a more generic sample prep method – more compounds & matrices
• Consolidated GC-MS methods due to high performance – buffer against requirements
• Compressed chromatography possible
• Easy peak evaluation – auto-integrators
![Page 15: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/15.jpg)
15
...and often in Pesticides Analysis Leads to...
High Selectivity
• Possibility to reduce selectivity in sample preparation
• Reduced sample prep steps create a more generic sample prep method – more compounds & matrices
• Consolidated GC-MS methods due to high performance – buffer against requirements
• Compressed chromatography possible
• Easy peak evaluation – auto-integrators
![Page 16: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/16.jpg)
16
Why are we here today?
Discussion of the practical issues that arise in the lab due to the extra capability GC-MS/MS brings and nature of the technique
• Most people working with this instrumentation are looking for ways to easily create, optimize and manage in routine large multi-residue GC-MS/MS methods
• Also, improve analytical performance in pesticides analysis
![Page 17: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/17.jpg)
17
Practical Issues?
The power of the technique is great, but...
• We are consolidating more and more compounds into a single method
• We are having to develop 100s of Selected Reaction Monitoring (SRM) transitions
• We are having to maintain and manage large, high performance methods – in routine!
• All this, in a variety of complex matrices
![Page 18: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/18.jpg)
18
What’s Really Needed...
To really benefit from the productivity advantages of these multi-residue methodologies, we need to:
• Create complex methods - independent of where we begin
• Manage all the complex information associated with large complex methods
• Maintain these methods in routine
• Ensure we are not creating a new bottleneck!
![Page 19: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/19.jpg)
19
Demonstrate how we can use automated software
• Integrated into the set-up and operation of the GC-MS/MS system
• Remove the pain (and avoid the bottleneck)
• Method creation
• Method optimization
• Method management
• Method maintenance
• For improved results (and productivity) in pesticide analysis
![Page 20: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/20.jpg)
20
Enabling Technology
“To make the productivity advantages of high performance GC-MS/MS easy to achieve and available routinely; especially for high-throughput laboratories.”
![Page 21: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/21.jpg)
21
![Page 22: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/22.jpg)
22
Anatomy of a Multi-Residue Pesticide Method
Peaks!
• Lots of them, too
• Multiple ions (SRM transitions)
• Multiple co-elutions
• Not much “clear baseline”
• Diverse chemistries
• Also chemistry similarities
• Large difference in response factors
• Different LOD requirements
• Different interference (matrix) pressures
• COMPLEXITY!!
![Page 23: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/23.jpg)
23
Complexity in Developing Multi-Residue MRM Methods
15. Run sequence
16. Examine each data compounds product ion spectrum in each run for each collision energy.
17. (Re-inject for any missed compounds)
18. Recording of best SRM transitions
19. Create MRM method to optimize collision energies
20. Create a pilot method(s) – to test selectivity of transitions in target matrices.
21. Choose final transitions
22. Segment method
23. Calculate appropriate dwell times (depending of number of overlapping transitions)
24. Test final method in matrix.
25. Check for “chopped” or missed peaks
26. Re-adjust method as necessary
1. List compounds (350 pesticides)
2. Arrange standard solutions into vial (s)
3. Set-up GC method
4. Run a full-scan
5. Examine data files to find compounds (extracting ions or using libraries)
6. Record retention times
7. Select and record appropriate pre-cursor ions
8. Create product ion scan methods
9. Segment these methods into windows based on chromatogram
10. Calculate appropriate scan times for good daughter ion experiments
11. Re-segment based on (10)
12. Set-up these methods for all collision energies to see progressive fragmentation
13. Decide on number of injections
14. Set-up sequence list
![Page 24: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/24.jpg)
24
Instrument & Data Processing Method Maintenance
• GC-MS/MS systems in routine pesticide analysis face high volumes of samples with high matrix load
• Cumulative deterioration of the GC column performance
• Backflushing set-up can help to mitigate
• Inevitably compound retention times drift and or GC columns need to be cut or replaced
• Need to locate compounds, update acquisition windows & update RT in data processing method
• Very laborious & time consuming- worse with a large method!
![Page 25: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/25.jpg)
25
Demonstrate how we can use automated software
• Integrated into the set-up and operation of the GC-MS/MS system
• Remove the pain (and avoid the bottleneck)
• Method creation
• Method optimization
• Method management
• Method maintenance
• For improved results (and productivity) in pesticide analysis
![Page 26: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/26.jpg)
26
Software Overview
Automated SRM Development
Timed-SRM Instrument Method
Batch Acquisition, Data Review, and reporting software
![Page 27: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/27.jpg)
27
Complexity in Developing Multi-Residue MRM Methods
15. Run sequence
16. Examine each data compounds product ion spectrum in each run for each collision energy.
17. (Re-inject for any missed compounds)
18. Initial selection of best SRM transitions
19. Create a pilot method(s) – to test selectivity of transitions in target matrices.
20. Choose final transitions
21. Segment method
22. Calculate appropriate dwell times (depending of number of overlapping transitions)
23. Test final method in matrix.
24. Check for “chopped” or missed peaks
25. Re-adjust method as necessary
1. List compounds (350 pesticides)
2. Arrange standard solutions into vial (s)
3. Set-up GC method
4. Run a full-scan
5. Examine data files to find compounds (extracting ions or using libraries)
6. Record retention times
7. Select and record appropriate pre-cursor ions
8. Create product ion scan methods
9. Segment these methods into windows based on chromatogram
10. Calculate appropriate scans times for good daughter ion experiments
11. Re-segment based on (9)
12. Set-up these methods for all collision energies to see progressive fragmentation
13. Decide on number of injections
14. Set-up sequence list
![Page 28: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/28.jpg)
28
Surely, we have these pesticide transitions already!?
![Page 29: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/29.jpg)
29
Surely, we have these pesticide transitions already!?
Of course!
![Page 30: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/30.jpg)
30
Thermo Scientific TraceFinder Compound Data Store
![Page 31: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/31.jpg)
31
Compound-Based Method Creation
Selecting your compounds from CDS…
and processing methods
populates synced acquisition…
![Page 32: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/32.jpg)
32
Thoughts on Fishing and Triple Quadrupoles
"Give a man a fish; feed him for a day. Teach a man to fish; feed him for a lifetime“
Lao Tzu circa. 5th Century BC
![Page 33: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/33.jpg)
33
Thoughts on Fishing and Triple Quadrupoles
"Give a man a fish; feed him for a day. Teach a man to fish; feed him for a lifetime“
Lao Tzu circa. 5th Century BC
![Page 34: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/34.jpg)
34
AutoSRM Overview
1) Precursor ion selection
2) Product ion selection
3) Collision energy optimization
SR
M C
reation
Wo
rkflow
![Page 35: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/35.jpg)
35
Step 1 – Pick Your Precursor Ions
![Page 36: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/36.jpg)
36
Step 1 – Pick Your Precursor Ions
![Page 37: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/37.jpg)
37
Step 1 – Pick Your Precursor Ions
![Page 38: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/38.jpg)
38
Step 2 – Pick Your Product Ions
![Page 39: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/39.jpg)
39
Step 2 – Pick Your Product Ions
![Page 40: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/40.jpg)
40
Step 3 – Optimize Your Transitions
![Page 41: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/41.jpg)
41
Step 3 – Optimize Your Transitions
![Page 42: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/42.jpg)
42
AutoSRM Use Case
• Created and optimized > 250 transitions for > 80 compounds• Minimal user interaction over 24 hours
![Page 43: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/43.jpg)
43
Export from AutoSRM to Instrument Method
![Page 44: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/44.jpg)
44
Timed-SRM Method Overview
![Page 45: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/45.jpg)
45
Timed-SRM Method Overview
Acquisition windows centered around retention time
![Page 46: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/46.jpg)
46
Timed-SRM Method Overview
Acquisition windows allowed to overlap
![Page 47: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/47.jpg)
47
Timed-SRM Advantages
Segmented SRM
Timed SRM
![Page 48: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/48.jpg)
48
Timed-SRM Advantages
Acquisition Windows
Segmented SRM
Timed SRM
![Page 49: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/49.jpg)
49
Timed-SRM Advantages
• Removes wasted dwell time
• Allow higher overall dwell times
• Leads to higher sensitivity
Wasted Dwell Time
![Page 50: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/50.jpg)
50
Timed-SRM Advantages
• Peaks centered in acquisition window
• No peak elutes near acquisition break
• Allows for retention time shift (e.g. due to heavy matrix)
![Page 51: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/51.jpg)
51
Thermo Scientific TSQ 8000 GC-MS/MS Timed-SRM Case Study
• Previous method: Segmented SIM acquisition on single quad
• Required five injections for full list of pesticides
• Needed to analyze more pesticides (350 total), but current methodology took too long
• Wanted to consolidate to a single injection
![Page 52: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/52.jpg)
52
• Segmented SRM
• Closest compound to segment break:
5 seconds
• Average number of simultaneous transitions:
55
• Timed SRM
• Closest compound to segment break:
15 seconds
• Average number of simultaneous transitions:
15 (4X higher dwell times)
TSQ™ 8000 GC-MS/MS Timed-SRM Case Study
![Page 53: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/53.jpg)
53
TSQ 8000 GC-MS/MS Timed-SRM Case Study
Tea Analysis: 4 pg on-column
Terbacil
Alachlor
Tolylfluanid
Pyridaben
![Page 54: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/54.jpg)
54
Export from Instrument Method to TraceFinder™ Software
![Page 55: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/55.jpg)
55
TraceFinder Software Overview
Batch Creation
Data Review
Report Generation
Routine Workflow
![Page 56: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/56.jpg)
56
Instrument & Data Processing Method Maintenance
• GC-MS/MS systems in routine pesticide analysis face high volumes of samples with high matrix load
• Cumulative deterioration of the GC column performance
• Backflushing set-up can help to mitigate
• Inevitably compound retention times drift and or GC columns need to be cut or replaced
• Need to locate compounds, update acquisition windows & update RT in data processing method
• Very laborious & time consuming- worse with a large method!
![Page 57: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/57.jpg)
57
TraceFinder Software Method Sync
• Links TraceFinder Software Method with instrument method
• Enables• Compound based acquisition setup
• Automated update of acquisition windows
![Page 58: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/58.jpg)
58
Method Sync – Automated RT Update
Updating retention times in data review…
![Page 59: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/59.jpg)
59
Method Sync – Automated RT Update
…updates both TraceFinder Software Method and Timed-SRM Method
![Page 60: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/60.jpg)
60
Summary
We can use integrated and automated software
• Easily adopt large multi-residue methods
• Remove the pain (and avoid bottlenecks)
• Method creation
• Method optimization
• Method management
• Method maintenance
• To create improved results (and productivity) in routine pesticide analysis
![Page 61: Automated Software for Improved Results in Triple Quadrupole Gas Chromatography-Mass Spectrometry Pesticide Analysis](https://reader036.fdocuments.us/reader036/viewer/2022062300/554e79a3b4c9054a698b4fde/html5/thumbnails/61.jpg)
61
Thank You for Your Attention!
Questions?
Stay connected with us
Twitter @ChromSolutions
Chromatography Solutions Blog http://chromblog.thermoscientific.com/blog
YouTubehttp://www.youtube.com/ChromSolutions
Facebookhttp://www.facebook.com/ChromatographySolutions
Pinteresthttp://pinterest.com/chromsolutions/