Are conversion factors to international units assay related?

14
SoGAT 24-05-2005 24 May 2005 Are conversion factors to international units assay related? Marco Koppelman, Henk Reesink, Nico Lelie, Theo Cuypers

description

Are conversion factors to international units assay related?. Marco Koppelman, Henk Reesink, Nico Lelie, Theo Cuypers. NAT test system for comparison. NucliSens Extraction Cobas Amplicor HCV Procleix Ultrio (Chiron/Gen-Probe) HIV-1, HCV, and HBV Multiprep Cobas AmpliScreen HBV - PowerPoint PPT Presentation

Transcript of Are conversion factors to international units assay related?

SoGAT 24-05-2005

24 May 2005

Are conversion factors to international units assay related?

Marco Koppelman, Henk Reesink, Nico Lelie, Theo Cuypers

SoGAT 24-05-2005

NAT test system for comparison

• NucliSens Extraction Cobas Amplicor HCV

• Procleix Ultrio (Chiron/Gen-Probe) HIV-1, HCV, and HBV

• Multiprep Cobas AmpliScreen HBV

• AmpliPrep TNAI Cobas AmpliScreen HIV vrs 1.5

• AmpliPrep TNAI Cobas AmpliScreen HCV vrs2.0

• AmpliPrep TNAI Cobas AmpliScreen HBV

SoGAT 24-05-2005

NucliSens Extraction Cobas Amplicor HCV

Generic silica extraction‘Boom’ Technology

Cobas AmplicorNuclisens

SoGAT 24-05-2005

Procleix Ultrio HIV-1, HCV, and HBV

ResultsLuminometer

DetectionWater bath

TMAWater baths

PipettingTECAN 2

SampleProcessing

TCS

Specific Capture of Target

SoGAT 24-05-2005

AmpliPrep TNAI Cobas AmpliScreen HIV/HCV/HBV

Generic magnetic silica extraction

AmpliPrep Cobas Amplicor

SoGAT 24-05-2005

Reference materials

Serial dilutions of the WHO International standards for:

HBV genotype A * (NIBSC 97/746)

HIV-1 genotype B* (NIBSC 97/656)

HCV genotype 1* (NIBSC 96/790)

*IU: Calibrated in WHO Collaborative Studies.

PeliCheck reference panels for

HBV genotype A# (S2184)

HIV-1 genotype B# (S2092)

HCV genotype 1# (S2088)#Geq: Calibrated with Bayer bDNA

SoGAT 24-05-2005

Analytical sensitivityWHO International Standard HBV DNA (97/746)

IU/mL Multi-Prep

Ampli-Prep

Ultrio

n=24 n=24 n=63

40 100% 100% 100 %

20 100 % 100 % 100 %

10 100 % 100 % 98 %

5 90 % 96 % 76 %

2.5 86 % 71 % 58 %

1.25 71 % 50 % 27 %

0.63 33 % 46 % 13 %

0.31 36 % 13 % 6 %

0.16 9 % 13 % 8 %

0.08 0 % 0 % 0 %

log (dilution)

50%

% positive

log (dilution)

50%

% positive

log (dilution)

50%

% positive

SoGAT 24-05-2005

Analytical Sensitivity Procleix Ultrio test*

Reference material

50% detection limit 95% detection limit

WHO-IS N1 IU/mL 95%CI IU/mL 95% CI HIV-1 IS2 48 3.4 2.4 - 4.7 26 16 - 58 HCV IS 63 0.9 0.7 - 1.0 4.6 3.7 - 6.5 HBV IS 63 1.9 1.5 - 2.6 11 7.3 - 22 PeliCheck ref. panel

N geq/mL 95%CI geq/mL 95%CI

PeliCh. HIV-1 48 5.7 4.3 - 7.6 65 41 - 119 PeliCh. HCV 48 13 9.0 - 20 86 50 - 204 PeliCh. HBV 48 34 26 - 44 328 213 - 586 1 Number of repeats per dilution 2 Discriminatory HIV-1 test

*Multicenter study; Assal (BC-Tours), Chudy (PEI), Garcia de Villaescusa (BC-Madrid),

Koppelman (Sanquin). Transfusion in press.

SoGAT 24-05-2005

0 2 4 6 8log (dilution)

Conversion Factor between two data-sets

95%

50%

% positivePotency calculated with SPSS

SoGAT 24-05-2005

Conversion factors IU to Geq

AmpliScreen

 WHO

studies Ultrio1 AmpliPrep2 MultiPrep2 AmpliCor3

HIV-1 1 1.7 (1.2-4.3) 1.6 (0.9-2.7) - -

HCV 3.8 17.0 (5.8-84) 9.4 (3.3-43) - 5.1 (3.1-9.1)

HBV 7.4 20.7 (9.1-60) 12.5 (5.4-40) 11.0 (4.9-33) -

1Koppelman et al. Transfusion 2005 in press.

2Koppelman et al. 2005 Submitted.

3Cuypers et al.; Biologicals. 1999 27 303-14.

SoGAT 24-05-2005

Conversion factors between IU and copies/ml are assay specific

• Conversion factors of the WHO collaborative studies; the mean of different nucleic acid extraction techniques and amplification/detection systems (considerable interlaboratory variation)

• Molecular composition of the reference material

• Physical composition reference materials. Liquid versus lyophilized reference materials

Commutability of reference materials; knowledge of molecular characteristics of preparations is necessary

Preliminary conclusion: conversion factors from IU to VQC-geq assay specific

SoGAT 24-05-2005

Conversion factor Dutch guilder to Euro

What was the conversion factor between Euro and Guilder in 2001?

It makes no difference !

SoGAT 24-05-2005

Analytical Sensitivity AmpliScreen HIV-1/HCV/HBVAmpliPrep MultiPrep

Standard Nucleic acid extraction

method 50% limit of detection 95% limit of detection

HBV-IS AmpliPrep 1.0 IU/mL (0.7-1.3) 6.7 IU/mL (4.3-13) MultiPrep 0.7 IU/mL (0.5-1.0) 5.2 IU/mL (3.3-10) MultiPrepa 1.3 IU/mL (0.9-1.7) 5.5 IU/mL (4.0-10) HCV-IS AmpliPrep 2.9 IU/mL (1.7-5.5) 23 IU/mL (11-106) MultiPrepa 7.3 IU/mL (2.2-10) 29 IU/mL (21-86) HIV-1-IS AmpliPrep 18 IU/mL (13-26) 187 IU/mL (108-422) MultiPrepa 22 IU/mL (13-29) 78 IU/mL (60-137) HBV-PeliCheck AmpliPrep 12 geq/mL (8.2-18) 123 geq/mL (68-301) MultiPrep 8.2 geq/mL (5.4-13) 101 geq/mL (52-295) MultiPrep 8.0 geq/mL (6.0-13) 100 geq/mL (55-291) HCV-PeliCheck AmpliPrep 36 geq/mL (17-65) 289 geq/mL (142-1061) MultiPrepb 9.9 geq/mL (6.3-15) 77 geq/mL (43-203) HIV-1-PeliCheck AmpliPrep 28 geq/mL (20-41) 183 geq/mL (108-434) MultiPrepb 6.6 geq/mL (4.4-10) 92 geq/mL (49-238) a data from kit insert. b data from Lelie et al. Transfusion et al. 2002; 42: 527-536.

SoGAT 24-05-2005

Conversion factors IU to Geq

AmpliScreen

 WHO

studies Ultrio1 AmpliPrep2 MultiPrep2 AmpliCor3

HIV-1 1 1.7 1.6 - -

HCV 3.8 17.0 9.4 - 5.1

HBV 7.4 20.7 12.5 11.0 -

1Koppelman et al. Transfusion 2005 in press.

2Koppelman et al. 2005 Submitted.

3Cuypers et al.; Biologicals. 1999 27 303-14.