Applications of a Novel Sample Preparation Method for the Determination of Sulfonamides in Edible...
-
Upload
muhammad-umar-farooq -
Category
Documents
-
view
225 -
download
7
Transcript of Applications of a Novel Sample Preparation Method for the Determination of Sulfonamides in Edible...
Applications of a Novel Sample PreparationMethod for the Determinationof Sulfonamides in Edible Meat by CZE
Muhammad Umar Farooq, Ping Su, Yi Yang&
College of Science, Beijing University of Chemical Technology, 100029 Beijing, People’s Republic of China; E-Mail: [email protected]
Received: 9 October 2008 / Revised: 15 December 2008 / Accepted: 19 January 2009Online publication: 22 March 2009
Abstract
A simple, rapid and reliable method based on SPE clean-up and CZE separation wasvalidated for the trace determination of sulfonamides (SAs) in meat. Acetonitrile was used forthe extraction of SAs (sulfamethoxazole, sulfamethazine, sulfamerazine, and sulfadime-thoxine) from the samples and 1-propanol was used for the denaturing of the proteinspresent in the sample matrix. SPE procedure was employed for the clean-up and pre-con-centration of SAs prior to CZE analysis. Complete separation was achieved by using45 mmol L-1 phosphate buffer (pH = 6.3) at an applied voltage of 20 kV. Overall obtainedrecoveries were from 83.3 to 94.5% for the SAs. The detection limit of each sulfonamideranges from 4 to 6 lg kg-1. The presented one step SPE clean-up method is highly appli-cable for the determination of the SAs at a residue level below the maximum residue limit.
Keywords
Capillary zone electrophoresisSolid phase extractionSulfonamides residues
Introduction
Sulfonamides (SAs) are analogues of the
para-aminobenzoic acid (PABA) that
inhibit the folic acid synthesis in sus-
ceptible micro-organisms. These com-
pounds are not only employed for
therapeutic and prophylactic purposes,
such as gastrointestinal and respiratory
infections, but also in the livestock
industry to promote growth. Because of
their broad spectrum of activity and low
costs [1], they are often overused or
applied without respecting safety recom-
mendations. As a result, SAs undesirable
residues remain in the animal tissues,
meat, or biofluids such as milk. As it has
been observed that most of the sulfona-
mides are suspected to be carcinogenic
and enhanced the risk of developing
bacterial resistance in human body [2],
the use of these compounds is of great
public health concern and their residues
have to be controlled. Due to their
possible side effects, the usage of SAs in
animals is regulated [3]. The maximum
residue limits (MRLs) for SAs in food-
stuffs have been restricted to a rather low
level in many nations and regions. For
example, MRLs for total SAs have been
established at the level of 100 lg kg-1 [4]
in European Union countries.
The availability of rapid and reliable
cost-effective screening methods for
antimicrobial compounds is often the
most important aspect of implementing
adequate drug residue monitoring
programs. Microbiological [5–7], immu-
noassays [8] and sulfonamide antibody-
based biosensors [9] have been used as
screening methods of SAs in food. These
methods usually save time, provide sim-
plicity and remove the need of laborious
sample pretreatment steps, but they
provide only semiquantitative measure-
ments and sometimes negatively affected
by matrix components [10].
Several analytical techniques for the
quantitative detection of SAs residues
have been reported. Most widely used
technique is liquid chromatography
(HPLC) with UV, fluorescence, electro-
chemical detectors, MS detectors and
chemiluminescence [11–19]. Gas chro-
matography-mass spectrometry (GC-MS)
2009, 69, 1107–1111
DOI: 10.1365/s10337-009-1025-z0009-5893/09/05 � 2009 Vieweg+Teubner | GWV Fachverlage GmbH
Full Short Communication Chromatographia 2009, 69, May (No. 9/10) 1107
has also been used to detect the SAs [20–
23] but complex derivatization processes
have been needed due to the non-volatile
nature of SAs.
CE has become a powerful separa-
tion technique with its wide applications
due to its high resolution, high efficiency,
rapid analysis and small consumption of
both sample and solvent in comparison
with HPLC. Some methods have been
reported using CE with different detec-
tors for the analysis of SAs from phar-
maceuticals and meat tissues [24–29].
Most of the works are focused on the
separation studies rather than quantita-
tive analysis.
SAs are usually extracted from edible
tissues with organic solvents (acetone,
ethyl acetate, chloroform, acetonitrile
and methanol), alone or combined.
Supercritical fluid extraction (SFE) [30],
matrix solid phase dispersion (MSPD)
[31–33] and accelerated solvent extrac-
tion (ASE) [34] have also been used for
extraction of SAs from pork, calf and
chicken samples. Since there is a need to
remove protein, fat and other potential
interferences from sample matrix, sam-
ple pre-treatment is required. Liquid-
liquid extraction (LLE) combined with
solid-phase extraction (SPE) has been
demonstrated as quite effective proce-
dure. But these methods rather use the
combination of different SPE cartridges
or need tedious clean-up steps [24, 25,
35, 36].
Our basic goal was to achieve sim-
plicity with better selectivity and sensi-
tivity for the isolation of SAs from meat.
LLE was done for the extraction of
sulfamethazine (SMZ), sulfamerazine
(SMR), sulfadimethoxine (SDM), and
sulfamethoxazole (SMX) from meat
samples. SPE was used for the purifica-
tion of desired analytes from sample
matrix. This simplified extraction pro-
tocol is quite sensitive and selective for
SAs. The method was found to be useful
alternative to the commonly applied
techniques for sample preparation. Four
SAs in the sample was analyzed by CZE
within 15 min. The established SPE-CE
method gave a shorter overall analysis
time, provided good resolution for four
SAs and produced excellent linearity
over a wide concentration range.
Experimental
Chemicals
Acetonitrile and methanol were HPLC
grade, purchased from J & K Chemical
(Beijing, China). Ethyl acetate, n-hexane,
sodium sulfate, sodium hydroxide,
potassium dihydrogen phosphate and
1-propanol all were analytical grade
reagents and purchased from Beijing
Chemicals (Beijing, China). Sulfamera-
zine (SMR), sulfamethazine (SMZ),
sulfamethoxazole (SMX) and sulfadi-
methoxine (SDM), with purity 99% were
purchased from Sigma-Aldrich (St
Louis, MO, USA). Cleanert PEP-SPE
cartridges (3 mL, 60 mg) were supplied
by Agela Technologies (Beijing, China).
Food mixer was obtained from Supor
Group (Zhejiang, China). The water
was purified and deionized by a Milli-Q
system (Millipore, Bedford, MA, USA).
The solvents and samples for CE were
filtered through 0.45 lm filters and
degassed in an ultrasonic bath before
injection.
Standard Solution
Individual standard solutions of
100 mg L-1 were prepared in metha-
nol:water (50:50 v/v) and stored in dark
at 4 �C. Working solutions at various
concentrations were freshly prepared by
diluting with water.
Buffer solution was prepared by dis-
solving the appropriate quantity of
KH2PO4 in deionized water and pH was
adjusted by 0.05 mol L-1 H3PO4 or
NaOH solutions.
Apparatus
Experiments were carried out with
Beckman Coulter P/ACE MDQ capil-
lary electrophoresis system equipped
with a photodiode array detector (PDA).
Uncoated fused-silica capillary was used
(60.2 cm 9 75 lm id, 50 cm to detec-
tor). Samples were injected by applying a
pressure of 0.5 psi for 15 s. The applied
voltage is 20 kV and capillary tempera-
ture was maintained at 25 �C. At the
beginning of each run, the capillary was
conditioned by flushing with 1 mol L-1
NaOH for 10 min, water for 5 min and
finally with electrolyte for 20 min. For
pH measurements, a Sartorius PB—10
pH meter was employed.
Sample Preparation
The meat and liver samples were ground
and homogenized by using a food mixer.
Solvent extraction was used for the
extraction of SAs from the samples.
Acetonitrile was found to be the suitable
solvent for extraction. 5 g (meat or liver)
homogenized sample was placed in a
50 mL centrifuge tube and vortexed with
15 mL acetonitrile, centrifuged for
10 min at 5,000 rpm and filtered through
0.45 lm membrane filter. 5 mL of
1-propanol was added and clear super-
natant was evaporated to dryness under
reduced pressure at 40 �C, reconstitutedwith 10 mL water and passed through
Cleanert PEP-SPE polymer cartridge.
The cartridge was conditioned with
3 mL methanol and water prior to the
loading of sample. After sample loading,
the cartridge was rinsed by water to
wash out the interferences and then dried
for about 10 min. Finally, analytes were
eluted with 4 mL methanol. Eluted
methanol was evaporated to dryness
under a gentle stream of nitrogen and
reconstituted with 1 mL methanol/water
(50/50, v/v) mixture.
Results & Discussion
Optimization of CE
Electropherograms were taken at
254 nm wavelength. pH of the buffer
solution plays a major role in the sepa-
ration of SAs. Phosphate buffer at dif-
ferent pH (5.5–8.5) was investigated
in order to ascertain any influence on
SAs separation and quantification. Best
resolution was obtained at a pH of 6.3.
Ionic strength of the buffer is also a
critical parameter for the separation in
CE. Different concentrations of buffer
(30, 35, 40, 45, 50 and 60 mmol L-1)
were tested. At low concentration, the
1108 Chromatographia 2009, 69, May (No. 9/10) Full Short Communication
migration time was shorter which led to
poor resolutions. Increase in concentra-
tion resulted in increasing of the migra-
tion time. 45 mmol L-1 was selected as
the optimized concentration. The effects
of capillary temperature, applied voltage
and injection volume were also opti-
mized. It was verified that the tempera-
ture (ranging 15–30 �C) and applied
voltage (ranging 15–30 kV) had much
influence on the migration of analytes
and separation efficiencies. Temperature
of 25 �C and 20 kV applied voltage were
chosen as the best compromise between
running time, efficiency of separation
and sensitivity. Moreover, it was also
observed that 15 s injections were most
efficient (5 s injections correspond to
approximately 20 nL injection volume of
sample) and gave a satisfactory separa-
tion in a reasonable analysis time
(15 min). A typical electropherogram of
SMZ, SMR, SDM and SMX at the
optimized conditions is depicted in
Fig. 1.
Isolation of SAs from Meat
Beside the instrument parameters and
buffer composition etc., clean-up step is
the most important factor that influ-
ences the sensitivity. As SAs bind to the
proteins of meat, therefore the effect of
proteins in the extraction and sample
cleaning can not be neglected [37].
Different solvent systems such as, ethyl
acetate, acetonitrile, mixture of aceto-
nitrile and acetone, mixture of ethyl
acetate and acetone were employed for
the extraction of SAs [38, 39]. Acetoni-
trile and ethyl acetate were found better
than mixture solvent because of their
abilities to precipitate proteins present
in the sample matrix. After extraction,
proteins in sample matrix were dena-
tured with two different approaches.
1-propanol was used in one approach
while 1 mol L-1 HCl neutralized with
1 mol L-1 NaOH was used in another
approach. Both of the approaches were
found reasonably well for the denatur-
ing of proteins but 1-propanol was
preferred due to the recovery efficiency.
Due to this fact, the mixture of 1-pro-
panol with ethyl acetate and acetonitrile
was also used for the extraction of SAs
but not much significance was observed.
So 1-propanol was utilized for the
denaturing of proteins in the sample
matrix. Cleanert PEP-SPE polymer
cartridge was employed for the further
cleaning of interferences.
Validity of Analysis
The analytical performance parameters
of the method including selectivity, pre-
cision, limits of detection and recovery
were validated. The analysis has been
carried out on the basis of the replicate
analysis of sample according to the
recommendations of EU [40].
Drug free meat samples and spiked
samples were tested. Absence of interfer-
ences from sample matrix in the migra-
tion range of SAs in electropherograms
confirmed the selectivity of method.
Linearity of method was evaluated by
plotting calibration curves in the range
of 0.5–50 lg mL-1 with triplicate anal-
ysis. It was found to be linear over this
range with acceptable correlation coeffi-
cient (r2 = 0.99). Limits of detection
(LOD) were calculated by using S/N
ratio of 3. LOD of each SAs ranged from
3.7 to 6.0 lg kg-1. The LOD values were
comparable to the reported methods
[12, 13]. Precision of method is sufficient
determined by comparing the migration
time and peak area of SAs standards.
Table 1 represents the within-day and
between-day repeatability in terms of
percentage RSD. RSD values of within-
day precision for migration times were
1.3–2.4% and 1.1–3.7% for peak areas.
While between-day RSD values for
migration time were 3.7–5.8% and 4.0–
6.4% for peak areas, showing that the
method is reproducible.
Fig. 1. Electropherogram of four sulfonamides at 25 lg mL-1. Separation conditions:45 mmol L-1 phosphate buffer (pH = 6.3), 20 kV, 25 �C, fused-silica capillary,60.2 cm 9 75 lm id, 50 cm to detector. 1 = SMZ, 2 = SMR, 3 = SDM, 4 = SMX
Table 1. Evaluation of the method
Analyte Within-daya
RSD%
Between-daya
RSD%
Slopeb Intercept r2 LODlg kg-1
LOQlg kg-1
tm Area tm Area
Sulfamethazine 1.3 1.1 4.1 4.0 3537.8 12.1 0.99 5.8 10.7Sulfamerazine 1.7 1.1 5.4 6.4 3251.7 -11.2 0.99 4.5 10.5Sulfadimethoxine 2.1 3.7 3.7 4.8 4111.7 -14.2 0.99 6.0 13Sulfamethoxazole 2.4 2.6 5.8 5.8 4206.6 -4.7 0.99 3.7 24
a n = 5b Calibration data (six points, three replicate of each point)
Full Short Communication Chromatographia 2009, 69, May (No. 9/10) 1109
Recovery
To check the applicability and trueness
of method, extraction recovery was
determined. For recovery studies of
SAs in meat samples, three spiked
samples at concentration of 50, 100
and 500 lg kg-1 were used. The results
are shown in Table 2. Overall recover-
ies were 94.5% (SMZ), 85.0% (SMR),
83.9% (SDM) and 83.3% (SMX)
with RSD less than 10%. The data
confirmed the trueness of proposed
method having adequate sensitivity and
accuracy.
Robustness
Robustness of method was checked by
the eventual variations introduces in
different parameters. Parameters inves-
tigated were buffer composition (pH =
6.3 ± 0.1, ± 0.2), temperature (25 ±
1.0 �C), injection volume (15 ± 1 s),
rinse time, applied voltage (20 ± 1 kV).
No significant variation was observed.
The method was proved to be robust and
could be able to be applied in the routine
analysis.
Analysis of Real Sample
Applicability of the method on real
sample was validated by the analysis of
chicken and beef tissue and liver samples
obtained from local market. Electro-
pherograms of tissue samples spiked at
10 lg kg-1 and 50 lg kg-1 of each of
SAs obtained after the application of
present method are shown in Fig. 2 a
and b. No interferences from the matrix
were observed.
Conclusion
Application of a simple and sensitive
method for the extraction and analysis
of mostly used SAs from chicken and
beef tissue samples were evaluated. CE
separation and determination conditions
of SAs were optimized with phosphate
buffer (pH = 6.3, 45 mmol L-1). Matrix
proteins were denatured by 1-propanol.
Cleanert PEP-SPE polymer cartridge
was used for the cleaning of potential
interferences. The proposed simplified
clean-up procedure enables the quantifi-
cation of SAs residues at concentration
lower than the MRL established by EU.
Obtained recoveries were over 94%. The
presented method is feasible to study the
occurrence of selected compounds in
edible meat.
Acknowledgments
This work was supported by Natural
Science Foundation of China (No.
20375004).
References
1. Brugere H, Brunere-Picux J, Villenin P(1985) Rec Med Vet 161:1241–1246
2. Littlefield NA, Sheldon WG, Allen R,Gaylor DW (1990) Food Chem Toxicol28:157–167. doi:10.1016/0278-6915(90)90004-7
Table 2. Recovery at different spiking levels
Analyte Concentrationadded (lg kg-1)
Mean recovery(%) (RSD %)
Sulfamethazine 500 97 (5.74)100 92 (3.92)50 94.6 (6.00)
Sulfamerazine 500 84.6 (2.97)100 83.6 (7.30)50 86.7 (9.43)
Sulfadimethoxine 500 99 (6.06)100 75.3 (5.98)50 77.3 (5.38)
Sulfamethoxazol 500 85.3 (5.78)100 86.6 (2.40)50 78 (5.12)
Fig. 2. Electropherograms of chicken meat samples spiked at (a) 10 lg kg-1 and (b) 50 lg kg-1
of each of the sulfonamides. Separation conditions were same as Fig. 1. 1 = SMZ, 2 = SMR,3 = SDM, 4 = SMX
1110 Chromatographia 2009, 69, May (No. 9/10) Full Short Communication
3. Van KV, Woulters MFA, Van FXRL(1993) WHO food additives series. WorldHealth Organization, Geneva
4. Establishment of Maximum ResidueLevels of Veterinary Medical Products inFoodstuffs and Animal Origin, CouncilRegulation No. 237790 of EEC, Euro-pean Economic Community, 26 June1990
5. Pena A, Serrano C, Reu C, Baeta L,Calderon V, Silveira I, Sousa JC, Peixe L(2004) Food Addit Contam 21:749–755.doi:10.1080/02652030410001712493
6. Dey BP, Reamer RP, Thaker NH, ThalerAM (2005) J AOAC Int 88:440–446
7. Dey BP, Reamer RP, Thaker NH, ThalerAM (2005) J AOAC Int 88:447–453
8. He J, Shen J, Suo X, Jiang H, Hou X(2005) J Food Sci 70:C113–C117
9. McGrath T, Baxter A, Ferguson J,Haughey S, Bjurling P (2005) Anal ChimActa 529:123–127. doi:10.1016/j.aca.2004.10.054
10. Font G, Garcia AJ, Pico Y (2007) JChromatogr A 1159:233–241. doi:10.1016/j.chroma.2007.03.062
11. Ikai Y, Oka H, Kawamura N, HayakawaJ, Yamada M, Harada KI, Suzuki M(1991) J Chromatogr A 541:393–400. doi:10.1016/S0021-9673(01)96011-X
12. Balizs G, Benesch-Grike L, Borner S,Hewitt SA (1994) J Chromatogr B661:75–84. doi:10.1016/0378-4347(94)00324-6
13. Jen JF, Lee HL, Lee BN (1998) J Chro-matogr A 793:378–382. doi:10.1016/S0021-9673(97)00894-7
14. Porter S (1994) Analyst (Lond) 119:2753–2756. doi:10.1039/an9941902753
15. Hartig C, Storm T, Jekel M (1999) JChromatogr A 854:163–173. doi:10.1016/S0021-9673(99)00378-7
16. Heller DN, Ngoh MA, Donoghue D,Podhorinak L, Righter H, Thomas MH(2002) J Chromatogr B 774:39–52. doi:10.1016/S1570-0232(02)00187-3
17. Cavaliere C, Curini R, Di Corcia A,Nazzari M, Samperi R (2003) J AgricFood Chem 51:558–566. doi:10.1021/jf020834w
18. Soto-Chinchilla JJ, Gamiz-Gracia L,Garcıa-Campana AM, Imai K, Garcıa-Ayuso LE (2005) J Chromatogr A1095:60–67. doi:10.1016/j.chroma.2005.07.123
19. Sabatino MD, Pietra AMD, Benfenati L(2007) J AOAC Int 90:598–603
20. Takatsuki K, Kikuchi T (1990) J AssocOff Anal Chem 73:886–892
21. Carignan G, Carrier K (1991) J Assoc OffAnal Chem 74:479–482
22. Tarbin JA, Clarke P, Shearer G (1999) JChromatogr B 729:127–138. doi:10.1016/S0378-4347(99)00142-5
23. Soto-Chinchilla JJ, Garcıa-CampanaAM, Gamiz-Gracia L (2007) Electro-phoresis 28:4164–4172. doi:10.1002/elps.200600848
24. Fuh MS, Chu SY (2003) Anal Chim Acta499:215–221. doi:10.1016/S0003-2670(03)00721-9
25. Soto-Chinchilla JJ, Garcia-CampanaAM, Gamiz-Garcia L, Cruces-Blanco C(2006) Electrophoresis 27:4060–4068. doi:10.1002/elps.200600166
26. Lamba S, Sanghi SK, Asthana A, ShelkeM (2005) Anal Chim Acta 552:110–115.doi:10.1016/j.aca.2005.05.084
27. Santos B, Lista A, Simonet BM, Rıos A(2005) Electrophoresis 26:1567–1575. doi:10.1002/elps.200410267
28. Wang A, Gong F, Li H, Fang Y (1999)Anal Chim Acta 386:265–269. doi:10.1016/S0003-2670(98)00848-4
29. Ma M, Zhang HS, Xiao LY, Xiao L,Wang P, Cui HR, Wang H (2007) Elec-trophoresis 28:4091–4100. doi:10.1002/elps.200700317
30. Arancibia V, Valderrama M, RodrıguezP, Hurtado F, Segura R (2003) J Sep Sci26:1710–1716. doi:10.1002/jssc.200301501
31. Kishida K, Furusawa N (2001) J Chro-matogr A 937:49–55. doi:10.1016/S0021-9673(01)01307-3
32. Bogialli S, Curini R, Di Corcia A, Nazz-ari M, Samperi R (2003) Anal Chem75:1798–1804. doi:10.1021/ac0262816
33. Kishida K, Furusawa N (2003) J LiquidChromatogr Relat Technol 26:2931–2939.doi:10.1081/JLC-120025054
34. Gentili A, Perret D, Marchese S, Sergi M,Olmi C, Curini R (2004) J AgricFood Chem 52:4614–4624. doi:10.1021/jf0495690
35. Ben WW, Qiang ZM, Adams C, ZhangHQ, Chen LP (2008) J Chromatogr A1202:173–180. doi:10.1016/j.chroma.2008.07.014
36. Ito Y, Oka H, Ikai Y, Matsumoto H,Miyazaki Y, Nagase H (2000) J Chro-matogr A 898:95–102. doi:10.1016/S0021-9673(00)00828-1
37. Stoev G, Mlchailova AI (2000) J Chro-matogr A 871:37–42. doi:10.1016/S0021-9673(99)00904-8
38. Diaz GJ, Skovdos KW, Yost GS, SquiresEJ (1999) Drug Metab Dispos 27:1150–1156
39. Spapan CV, Lundblad RL, Price NC(1999) Biotechnol Appl Biochem 29:99–108
40. European Commission Decision 2002/657/EC of 12 August 2002 implementingCouncil Directive 96/23EC concerningthe performance of analytical methodsand interpretation of results, OJ L, 2002,p 221
Full Short Communication Chromatographia 2009, 69, May (No. 9/10) 1111