Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1 , S. TAJIRI 2 , H. TOMINAGA 2 , K. KAWASAKI 2 , K. KOGO 2 , M. YOSHITANI 3 , N. ARAKI 3 , L. Müller 4 1 Kumamoto University School of Health Sciences, Kumamoto, Japan, 2 Kumamoto Radiosurgery Clinic, 3 Toyo Medic Co., Ltd., Tokyo, Japan, 4 Scanditronix Wellhöfer GmbH, Schwarzenbruck, Germany BACKGROUND A pixel-segmented ionization chamber, MatriXX, (Scanditronix Wellhöfer, Ge rmany), has been designed to ease the two-dimensional (2D) verifications o f fields with complex shapes like intensity modulated radiation therapy, I MRT. The detector features a 24×24 cm 2 active area divided in 1020 indepen dent plane parallel ionization chambers. The sensitive volume of each sign al ionization chamber is 0.07 cm3 (4.5 mm diameter×5 mm height) and chambe r-to-chamber distance is 7.62 mm). PURPOSE This study aims to examine dosimetric characteristics of MatriXX and its u sefulness for dose verification of dynamic IMRT fields with a dedicated No valis shaped beam radiosurgery unit equipped with a micro-multileaf collim ator (mMLC: BrainLAB, Heimstetten, Germany). MARERIALS AND METHOD The absorbed dose for MatriXX was calibrated with a Farmer ionization cham ber in a solid water phantom. The MatriXX response was tested as to dose l inearity and dose rate dependence. Basic parameters of radiation beam (TP Rs, OARs, output factors) were compared with ionization chamber measuremen ts for a 6 MV photon beam. The spatial resolution of MatriXX was tested us ing a bar pattern of mMLC with 3, 4.5, 5.5 and 10 mm wide at isocenter. Th e analysis of measured (MatriXX and ionization chamber) and calculated abs orbed dose distributions were performed for complex dynamic IMRT fields. T he dynamic IMRT dose distribution was calculated with the BrainSCAN 3-D tr eatment planning system (BrainLAB, Heimstetten, Germany) for stereotactic radiotherapy (SRT) fields formed by mMLC. Fig. 1 TPR and output factor data obtained for a 6 MV x-ray beam using an ionization chamber and MatriXX in a solid water phantom. The output factors for small fields less than 2 x 2 cm 2 were measured with a diode detector (SFD). RESULTS AND DISCUSSION The dose response of MatriXX is linear in the range 0.1-10 Gy. The dose rate dependence is within 0.1% in the range 1.6-8 Gy/min. The TPR and OAR data meas ured in a 6 MV photon beam is in good agreement with ionization chamber measur ements (Figs. 1(a) and 2). Output factors measured with MatriXX and an ionizat ion chamber agree within 1% for field sizes in the range from 1.2×1.2 cm 2 to 1 0×10 cm 2 (Fig. 1(b)). The spatial resolution of MatriXX is 7.62 mm, which cannot recognize the bar pattern of 3 mm, as it contains spatial frequencies much higher than Matri XX’s intrinsic resolution (Figs. 3(d) and (e)). The bar pattern of 4.5 and 5.5 mm can be recognized by MatriXX but the dose profile does not contain the full modulation due to limited spatial resolution (Figs. 3(d) and (e)). The MatriXX profile agrees with that of RTPS for the 10 mm bar pattern (Figs. 3(a)~(c)). A direct comparison can be made between dose profiles measured with Mat riXX and the calculated dose profiles of RTPS when they are integrated to the same spatial resolution of 7.62 mm (Fig. 4). With this method, excellent agree ment can be achieved between planned data and measurement with MatriXX. In dos e verification of IMRT fields, MatriXX agrees with those of RTPS in the region where the dose slope is gentle, while it shows some difference from the dose d istributions by RTPS in the steep slope regions (Figs. 5 and 6). This is becau se the dose distribution cannot be faithfully reflected in the linear interpol ation in relation to the resolution due to the arrangement of MatriXX ionizati on chambers (at 7.62 mm intervals). CONCLUSIONS The pixel-ionization-chamber 2D-MatriXX is found to be able to measure the giv en dose (absolute dose) and dose distributions simultaneously. It is useful fo r dose verification of dynamic IMRT pre-treatment. As a word of caution, the l imited number of detectors and the dimension of a single detector have to be a lways kept in mind when dealing with very high gradient or very small field si zes formed with mMLC, but within these limitations it can be still used for th e verification of these fields. Reference 1. S. Amerio, A. Boriano, F. Bourhaleb, et al., “Dosimetric characterization of a large area pixel-segmented ionizati on chamber,” Med, phys. 30, 414-20 (2004). Fig. 2 Comparison of beam profiles measured with an ionization chamb er in a water phantom and with MatriXX in a solid water phantom. The i rradiations were performed at d max and 20 cm, and for 4.2 x 4.2 cm 2 an d 10 x 10 cm 2 field sizes in a 6 MV x-ray beam. The results are normal ized to the central beam axis. Fig. 3 Comparison of dose profil es between MatriXX and RTPS teste d using a bar pattern of mMLC wit h 3, 4.5, 5.5 and 10 mm wide at i socenter. Measurements were perf ormed at a depth of 10 cm in a so lid water phantom for a 6 MV x-ra y beam, SSD=95 cm. The dose profi le of RTPS is transferred to the same grid size (7.62 mm) as MariX X in figure of lower left. Fig. 5 Comparison of IMRT dose profiles between MatriXX and RTPS for before and after grid transform. The dose profiles of RTPS are transfer red to the same grid size as MatriXX. Fig. 6 Comparison of IMRT dose profiles between MatriXX and RTPS. The do se profiles are evaluated by D=3% and d=3mm of “Gamma method”. 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 0 5 10 15 20 25 30 D epth in solid w aterphantom (cm) TPR M atriXX Ionization cham ber 6 M V , 10x10 cm 2 , S AD =100 cm 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 0 1 2 3 4 5 6 7 8 9 10 Field size (cm ) O utputfactor M atrixx Ionization cham ber+ SFD 6 M V S SD =100 cm depth=5 cm 5.5 m m 5.5 m m 4.5 m m 4.5 m m 3 m m intervals 10 m m 11 m m 10.5 m m 9 m m intervals 9 m m M atriXX RTPS Y-profile SA D =100cm , D epth=10cm M atriXX RTPS M atriXX RTPS M atriXX RTPS Y-profile SA D =100cm , D epth=10cm 0 20 40 60 80 100 120 -10 -8 -6 -4 -2 0 2 4 6 8 10 X axis (cm ) R elative dose (% ) 6 M V SSD=100 cm d=1.5 cm 0 20 40 60 80 100 120 -10 -8 -6 -4 -2 0 2 4 6 8 10 X axis (cm ) C C01 4.2x4.2 C C01 10x10 M atriXX 4.2x4.2 M atriXX 10x10 6 M V S SD =100 cm d=20 cm M atriXX RTPS Y-profile S A D =100cm , D epth=10cm M atriXX RTPS G rid size ofR TPS Transferto 7.62 m m grid G rid sizes ofR TPS and M atriXX Linearinterpolation to 1 m m grid M atriXX ion cham berarray = 7.62 m m spacing Fluence m ap ofM atriXX Fluence m ap ofR TPS O riginal O riginal:7.62 m m grid Transferto 7.62 m m grid O riginal:7.62 m m grid Fluence m ap ofM atriXX Fluence m ap ofR TPS (b) (d) Linearinterpolation to 1 m m grid Fluence m ap ofM atriXX Fluence m ap ofR TPS (a) Procedure to compare MatriXX with RTPS. M atriXX RTPS Y-profile M atriXX RTPS G am m a M ethod D=3% , d= 3 m m Error:0.28% M atriXX RTPS Y-profile M atriXX RTPS Fig. 4 Procedure to compare fluence map of MatriXX with RTPS’s for I MRT dose verification. Fig. (d) is final dose profiles of MatriXX and RTPS transferred to the same grid size. (c ) (a) Original dose profiles of MatriXX and RTPS. 0 200 400 600 800 1000 -6 -4 -2 0 2 4 6 X axis(cm ) R elative dose R TP S _O riginal M atriX X 0 200 400 600 800 1000 -6 -4 -2 0 2 4 6 X axis(cm ) R elative dose R TP S (after trans) M atriXX (b) After transferring to the same grid size. (a) Original dose profiles. (b) After transferring to the same grid size. (a) TPR (b) Output factor (a) depth=1.5 cm (b) d=20 cm (a) (b) (c) (d) (e)

Transcript of Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC...

Page 1: Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.

Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD1, S. TAJIRI2, H. TOMINAGA2, K. KAWASAKI2, K. KOGO2, M. YOSHITANI3, N. ARAKI3, L. Müller4

1Kumamoto University School of Health Sciences, Kumamoto, Japan, 2Kumamoto Radiosurgery Clinic, 3Toyo Medic Co., Ltd., Tokyo, Japan, 4Scanditronix Wellhöfer GmbH, Schwarzenbruck, Germany

BACKGROUNDA pixel-segmented ionization chamber, MatriXX, (Scanditronix Wellhöfer, Germany), has been designed to ease the two-dimensional (2D) verifications of fields with complex shapes like intensity modulated radiation therapy, IMRT. The detector features a 24×24 cm2 active area divided in 1020 independent plane parallel ionization chambers. The sensitive volume of each signal ionization chamber is 0.07 cm3 (4.5 mm diameter×5 mm height) and chamber-to-chamber distance is 7.62 mm).

PURPOSEThis study aims to examine dosimetric characteristics of MatriXX and its usefulness for dose verification of dynamic IMRT fields with a dedicated Novalis shaped beam radiosurgery unit equipped with a micro-multileaf collimator (mMLC: BrainLAB, Heimstetten, Germany).

MARERIALS AND METHODThe absorbed dose for MatriXX was calibrated with a Farmer ionization chamber in a solid water phantom. The MatriXX response was tested as to dose linearity and dose rate dependence. Basic parameters of radiation beam (TPRs, OARs, output factors) were compared with ionization chamber measurements for a 6 MV photon beam. The spatial resolution of MatriXX was tested using a bar pattern of mMLC with 3, 4.5, 5.5 and 10 mm wide at isocenter. The analysis of measured (MatriXX and ionization chamber) and calculated absorbed dose distributions were performed for complex dynamic IMRT fields. The dynamic IMRT dose distribution was calculated with the BrainSCAN 3-D treatment planning system (BrainLAB, Heimstetten, Germany) for stereotactic radiotherapy (SRT) fields formed by mMLC.

Fig. 1 TPR and output factor data obtained for a 6 MV x-ray beam using an ionization chamber and MatriXX in a solid water phantom. The output factors for small fields less than 2 x 2 cm2 were measured with a diode detector (SFD).

RESULTS AND DISCUSSIONThe dose response of MatriXX is linear in the range 0.1-10 Gy. The dose rate dependence is within 0.1% in the range 1.6-8 Gy/min. The TPR and OAR data measured in a 6 MV photon beam is in good agreement with ionization chamber measurements (Figs. 1(a) and 2). Output factors measured with MatriXX and an ionization chamber agree within 1% for field sizes in the range from 1.2×1.2 cm2 to 10×10 cm2 (Fig. 1

(b)).    The spatial resolution of MatriXX is 7.62 mm, which cannot recognize the bar pattern of 3 mm, as it contains spatial frequencies much higher than MatriXX’s intrinsic resolution (Figs. 3(d) and (e)). The bar pattern of 4.5 and 5.5 mm can be recognized by MatriXX but the dose profile does not contain the full modulation due to limited spatial resolution (Figs. 3(d) and (e)). The MatriXX profile agrees with that of RTPS for the 10 mm bar pattern (Figs. 3(a)~(c)).   A direct comparison can be made between dose profiles measured with MatriXX and the calculated dose profiles of RTPS when they are integrated to the same spatial resolution of 7.62 mm (Fig. 4). With this method, excellent agreement can be achieved between planned data and measurement with MatriXX. In dose verification of IMRT fields, MatriXX agrees with those of RTPS in the region where the dose slope is gentle, while it shows some difference from the dose distributions by RTPS in the steep slope regions (Figs. 5 and 6). This is because the dose distribution cannot be faithfully reflected in the linear interpolation in relation to the resolution due to the arrangement of MatriXX ionization chambers (at 7.62 mm intervals).

CONCLUSIONSThe pixel-ionization-chamber 2D-MatriXX is found to be able to measure the given dose (absolute dose) and dose distributions simultaneously. It is useful for dose verification of dynamic IMRT pre-treatment. As a word of caution, the limited number of detectors and the dimension of a single detector have to be always kept in mind when dealing with very high gradient or very small field sizes formed with mMLC, but within these limitations it can be still used for the verification of these fields.

Reference1. S. Amerio, A. Boriano, F. Bourhaleb, et al., “Dosimetric characterization of a large area pixel-segmented ionization chamber,” Med, phys. 30, 414-20 (2004).2. M. Stasi, S. Giordan, R. Cirio, “D-IMRT verification with a 2D pixel ionization chamber: dosimetric and clinical results in hed and neck caner.” Phys. Med. Biol. 50, 4681-94 (2005).

Fig. 2 Comparison of beam profiles measured with an ionization chamber in a water phantom and with MatriXX in a solid water phantom. The irradiations were performed at dmax and 20 cm, and for 4.2 x 4.2 cm2 and 10 x 10 cm2 field sizes in a 6 MV x-ray beam. The results are normalized to the central beam axis.

Fig. 3 Comparison of dose profiles between MatriXX and RTPS tested using a bar pattern of mMLC with 3, 4.5, 5.5 and 10 mm wide at isocenter.   Measurements were performed at a depth of 10 cm in a solid water phantom for a 6 MV x-ray beam, SSD=95 cm. The dose profile of RTPS is transferred to the same grid size (7.62 mm) as MariXX in figure of lower left.

Fig. 5 Comparison of IMRT dose profiles between MatriXX and RTPS for before and after grid transform. The dose profiles of RTPS are transferred to the same grid size as MatriXX.

Fig. 6 Comparison of IMRT dose profiles between MatriXX and RTPS. The dose profiles are evaluated by D=3% and d=3mm of “Gamma method”.

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0 5 10 15 20 25 30

Depth in solid water phantom (cm)

TP

R

MatriXXIonization chamber

6 MV, 10x10 cm2, SAD=100 cm

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0 1 2 3 4 5 6 7 8 9 10

Field size (cm)

Ou

tpu

t fa

cto

r

Matrixx

Ionization chamber + SFD

6 MVSSD=100 cmdepth=5 cm

5.5 mm

5.5 mm

4.5 mm

4.5 mm

3 mm intervals

10 mm

11 mm

10.5 mm

9 mm intervals

9 mm

MatriXX

RTPS

Y-profile SAD=100cm,Depth=10cm

ー MatriXXー RTPS

ー MatriXXー RTPS

MatriXX

RTPS

Y-profile SAD=100cm,Depth=10cm

0

20

40

60

80

100

120

-10 -8 -6 -4 -2 0 2 4 6 8 10

X axis (cm)

Rela

tive d

ose (%

)

6 MVSSD=100 cmd=1.5 cm

0

20

40

60

80

100

120

-10 -8 -6 -4 -2 0 2 4 6 8 10

X axis (cm)

CC01 4.2x4.2CC01 10x10MatriXX 4.2x4.2MatriXX 10x10

6 MVSSD=100 cmd=20 cm

MatriXX

RTPS

Y-profile SAD=100cm,Depth=10cm

ー MatriXXー RTPS

Grid size of RTPS Transfer to 7.62 mm grid

Grid sizes of RTPS and MatriXX

Linear interpolation to 1 mm grid

MatriXX ion chamber array = 7.62 mm spacingFluence map of MatriXXFluence map of RTPS

Original Original: 7.62 mm grid

Transfer to 7.62 mm grid Original: 7.62 mm grid

Fluence map of MatriXXFluence map of RTPS

(b)

(d)Linear interpolation to 1 mm grid

Fluence map of MatriXXFluence map of RTPS

(a) Procedure to compare MatriXX with RTPS.

MatriXX

RTPS

Y-profileー MatriXXー RTPS

Gamma Method

D=3%, d= 3 mm

Error: 0.28%

MatriXX

RTPS

Y-profileー MatriXXー RTPS

Fig. 4 Procedure to compare fluence map of MatriXX with RTPS’s for IMRT dose verification. Fig. (d) is final dose profiles of MatriXX and RTPS transferred to the same grid size.

(c)

(a) Original dose profiles of MatriXX and RTPS.

0

200

400

600

800

1000

-6 -4 -2 0 2 4 6

X axis(cm)

Re

lati

ve

do

se

RTPS_Original

MatriXX

0

200

400

600

800

1000

-6 -4 -2 0 2 4 6X axis(cm)

Re

lati

ve

do

se

RTPS (after trans)MatriXX

(b) After transferring to the same grid size.

(a) Original dose profiles. (b) After transferring to the same grid size.

(a) TPR (b) Output factor

(a) depth=1.5 cm (b) d=20 cm

(a)

(b)

(c)

(d)

(e)