![Page 1: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/1.jpg)
Angular Response Measurements of a Bentham DMc150 Spectroradiometer
Ewan Eadie1, Hannah Oliver1, Julie Smyth2, Harry Moseley1
1Photobiology Unit, Ninewells Hospital and Medical School, Dundee2 Medical Physics, Ninewells Hospital and Medical School, Dundee
![Page 2: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/2.jpg)
Spectroradiometers & Radiometers
![Page 3: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/3.jpg)
Calibration
Correct for errors Human: Poor calibration procedures, improper maintenance, stray light from poor technique, inaccurate lamp orientationEquipment: Non-linearity, aging, directional response, temperature
Evaluate remaining uncertainty
![Page 4: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/4.jpg)
Spectroradiometer Response
Depends on:Direction of incident radiationTemperature of Photomultiplier TubeUniformity of irradiation
Spectroradiometer used for a range of different source geometriesIdeal Cosine response
![Page 5: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/5.jpg)
Angular Responsef2 error: quantifies the quality of the spectroradiometers directional evaluation of the incident radiation
Pye, S.D. and Martin, C.J.: A study of the directional response of ultraviolet radiometers.
![Page 6: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/6.jpg)
f2
*2f
( ) ( )( )
( ) εεε
εεϕε
ϕε
ε
dff
RR
freading
reading
∫=
=
−=
=
309.1
0
22
2
2sin(%)
1cos0
,,
![Page 7: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/7.jpg)
Background
Yearly calibration of our Bentham DMc150 spectroradiometerTwo calibrated lamps
Bentham CL3 30W deuterium lampBentham CL2 100W quartz halogen lamp
Two diffusersPTFE diffuserQuartz diffuser
![Page 8: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/8.jpg)
UV Radiometer Calibration
Deuterium Lamp
Quartz Halogen Lamp
Lamp
Diffuser
PTFE Quartz
200 < λ
< 325
325 < λ
< 600 600 < λ
< 800
-
![Page 9: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/9.jpg)
Angular Response of PTFE and Quartz Diffusers
0
0.2
0.4
0.6
0.8
1
1.2
0 10 20 30 40 50 60 70 80Angle (degrees)
Rela
tive
Inte
nsity
Cosine
PTFE 500nm
PTFE 600nm
PTFE 700nm
PTFE 800nm
Quartz 500nm
Quartz 600nm
Quartz 700nm
Quartz 800nm
![Page 10: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/10.jpg)
Angular Response Measurements
Bentham DMc150 SpectroradiometerPTFE/Quartz diffuserQuartz Halogen LampAngular “Jig”
Distance = 50cmVertical AlignmentHorizontal
±0° - 75°
Vertical ±0° - 70°
![Page 11: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/11.jpg)
Angular Response Measurements
Bentham DMc150 SpectroradiometerPTFE/Quartz diffuserQuartz Halogen LampAngular “Jig”
Distance = 50cmVertical AlignmentHorizontal
±0° - 75°
Vertical ±0° - 70°
![Page 12: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/12.jpg)
Horizontal Measurements
![Page 13: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/13.jpg)
Vertical Measurements
![Page 14: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/14.jpg)
Results
4 readings for each angleHorizontal ±εVertical ±ε
Average R(ε)Relative Response: ( )
( )0=εε
RR
![Page 15: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/15.jpg)
Angular Response of PTFE and Quartz Diffusers
0.000
0.200
0.400
0.600
0.800
1.000
1.200
0 10 20 30 40 50 60 70 80
Angle (degrees)
Rel
ativ
e R
espo
nse
CosinePTFE 500nmPTFE 600nmPTFE 700nmPTFE 750nmPTFE 800nmQuartz 500nmQuartz 600nmQuartz 700nmQuartz 750nmQuartz 800nm
![Page 16: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/16.jpg)
UV Radiometer Calibration
Deuterium Lamp
Quartz Halogen Lamp
Lamp
Diffuser
PTFE Quartz
200 < λ
< 325
325 < λ
< 600 600 < λ
< 800
-
![Page 17: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/17.jpg)
Angular Response of PTFE and Quartz Diffusers
0.000
0.200
0.400
0.600
0.800
1.000
1.200
0 10 20 30 40 50 60 70 80
Angle (degrees)
Rel
ativ
e R
espo
nse
CosinePTFE 500nmPTFE 600nmPTFE 700nmPTFE 750nmPTFE 800nmQuartz 500nmQuartz 600nmQuartz 700nmQuartz 750nmQuartz 800nm
![Page 18: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/18.jpg)
PTFE f2f2 error
0
5
10
15
20
25
30
35
(%)
250nm300nm350nm400nm450nm500nm550nm600nm650nm700nm750nm800nm
![Page 19: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/19.jpg)
Quartz f2f2 error
0
5
10
15
20
25
30
35
(%)
250nm300nm350nm400nm450nm500nm550nm600nm650nm700nm750nm800nm
![Page 20: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/20.jpg)
Angular Correction Factor CF(ε)
True irradiance underestimatedApply a correction factorDepends on
Wavelength (PTFE)Average between 250nm and 500nm
Range of anglesSource geometry
![Page 21: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/21.jpg)
Angular Correction Factor CF(ε)
Source Range of Angles
Bank of 6 x 180cm lamps ±
0°
to 70°
Single 180cm lamp ±
0°
to 70°
Ninewells UVA1 Bank 8 x 60cm ±
0°
to 35°
Dr Honle Column ±
0°
to 60°
Single 60cm lamp ±
0°
to 35°
Dr Honle Dermalight Ultra 1 ±
0°
to 20°
Deuterium Lamp n/a
QHT lamp n/a
Dr Honle lamp n/a
![Page 22: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/22.jpg)
Angular Correction Factor CF(ε)
Source CF(θ)
Bank of 6 x 180cm lamps 1.07
Single 180cm lamp 1.04
Ninewells UVA1 Bank 8 x 60cm
1.03
Dr Honle Column 1.03
Single 60cm lamp 1.02
Dr Honle Dermalight Ultra 1 1.01
Deuterium Lamp n/a
QHT lamp n/a
Dr Honle lamp n/a
![Page 23: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/23.jpg)
Conclusions
250nm to 750nm: PTFE diffuser better represents a cosine responseResponse with PTFE is wavelength dependentAngular correction factor, CF(ε), included in UV radiometer calibrationsUncertainty in CF(ε) = 0.4%
![Page 24: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/24.jpg)
Conclusions
250nm to 750nm: PTFE diffuser better represents a cosine responseResponse with PTFE is wavelength dependentAngular correction factor, CF(ε), included in UV radiometer calibrationsUncertainty in CF(ε) = 0.4%
![Page 25: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/25.jpg)
Conclusions
250nm to 750nm: PTFE diffuser better represents a cosine responseResponse with PTFE is wavelength dependentAngular correction factor, CF(ε), included in UV radiometer calibrationsUncertainty in CF(ε) = 0.4%
![Page 26: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/26.jpg)
Conclusions
250nm to 750nm: PTFE diffuser better represents a cosine responseResponse with PTFE is wavelength dependentAngular correction factor, CF(ε), included in UV radiometer calibrationsUncertainty in CF(ε) = 0.4%
![Page 27: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/27.jpg)
References
1. Methods of characterizing the performance of radiometers and photometers. CIE No 53. 1982
2. Methods of characterizing illuminance meters and luminance meters. Performance, characteristics and specifications. CIE No 69. 1987
3. Pye, S. D. and Martin, C. J. A study of the directional response of ultraviolet radiometers: I. Practical evaluation and implications for ultraviolet measurement standards. Phys. Med. Biol. 2000
4. Martin, C. J. and Pye, S. D. A study of the directional response of ultraviolet radiometers: II. Implications for ultraviolet phototherapy derived from computer simulations. Phys. Med. Biol. 2000
![Page 28: Angular Response Measurements of a Bentham DMc150 …resource.npl.co.uk/optical_radiation/ormclub/member/... · 2008. 2. 6. · and photometers. CIE N o 53. 1982 2. Methods of characterizing](https://reader035.fdocuments.us/reader035/viewer/2022071606/6143a8fa6b2ee0265c022f83/html5/thumbnails/28.jpg)
Acknowledgements
Dr Harry MoseleyProfessor James FergusonPhotobiology Unit, Ninewells Hospital and
Medical School, DundeeChristine Wall, Teresa Goodman, Laura
Crane and Paul Miller