AMS/ACSM Calibrations Overview
Phil Croteau – [email protected]
August 24, 2019 AMS Users Meeting Lund 1
![Page 2: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/2.jpg)
Outline – Focus on PM1 vs. PM2.5 Lens and Standard vs. Capture Vaporizer Differences
• Inlet flow rate calibration – pressure to flow rate
• Particle lens alignment
• Velocity calibration – particle time-of-flight to size
• Vaporizer temperature calibration
August 24, 2019 AMS Users Meeting Lund 2
![Page 3: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/3.jpg)
mg
m3
mass
volume
From Mass Spectrometer
Ionization Efficiency calibration
From volumetric flow rate
Particle Mass Loading
August 24, 2019 AMS Users Meeting Lund 3
![Page 4: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/4.jpg)
Flow Meter
Low pressure drop
flow meter
Gilibrator™ or
equivalent soap film
meter10 Torr lens
pressure gauge
P
Aerodynamic Lens
Fine metering
valve
AMS Flow Rate Calibration Setup
Poiseuille equation relates pressure drop across Lens tube
to volumetric flow through the lens tube
Flow into
vacuum
August 24, 2019 AMS Users Meeting Lund 4
![Page 5: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/5.jpg)
Flow Calibration
3.0
2.5
2.0
1.5
1.0
0.5
0.0
Flo
w (
cc/s
)
2.52.01.51.00.50.0
Pressure (torr)
U. Tokyo Cal at ARI
NOAA Cal at ARI
NOAA Cal at Boulder
NOAA Scaled to 760 torr
Particle lens pressure is used to measure mass flow rate
August 24, 2019 AMS Users Meeting Lund 5
![Page 6: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/6.jpg)
August 24, 2019 AMS Users Meeting Lund 6
PM1 and PM2.5 flow rate calibration
• PM2.5 lens operates at higher pressure
• Bigger ΔP for smaller Δflow with PM2.5
Standard flow
![Page 7: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/7.jpg)
Research grade
TOF mass spectrometer
Thermal
Vaporization
&
Electron
Impact
Ionization
Aerodynamic
Lens
Beam
Chopper
Particle TOF Region
Pumps (x5)
Fast Data
System
Timers
DACs
Good lens alignment
August 24, 2019 AMS Users Meeting Lund 7
![Page 8: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/8.jpg)
Research grade
TOF mass spectrometer
Thermal
Vaporization
&
Electron
Impact
Ionization
Aerodynamic
Lens
Beam
Chopper
Particle TOF Region
Pumps (x5)
Fast Data
System
Timers
DACs
Bad Lens Alignment
August 24, 2019 AMS Users Meeting Lund 8
![Page 9: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/9.jpg)
AMSDMA
Atomizer
Drier
Setup for particle lens alignment
NH4NO3
August 24, 2019 AMS Users Meeting Lund 9
![Page 10: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/10.jpg)
Vaporizer
Move lens while monitoring NO2+ signal
LENS
August 24, 2019 AMS Users Meeting Lund 10
![Page 11: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/11.jpg)
Vaporizer
Move lens while monitoring NO2+ signal
LENS
August 24, 2019 AMS Users Meeting Lund 11
![Page 12: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/12.jpg)
Vaporizer
Move lens while monitoring NO2+ signal
LENS
Measure here, “min”
Measure here, max
August 24, 2019 AMS Users Meeting Lund 12
![Page 13: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/13.jpg)
Vaporizer
Move lens while monitoring NO2+ signal
LENS
Measure here, min
August 24, 2019 AMS Users Meeting Lund 13
![Page 14: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/14.jpg)
Vaporizer
Move lens while monitoring NO2+ signal
LENS
Measure here, min
Measure here, max
𝐶𝑒𝑛𝑡𝑒𝑟 =min+𝑚𝑎𝑥
2
After calculating, move to Center and repeat for the other axis
August 24, 2019 AMS Users Meeting Lund 14
![Page 15: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/15.jpg)
August 24, 2019 AMS Users Meeting Lund 15Yan Zheng, PKU 2017 AMS users meeting
Lens alignment with the capture vaporizerSame procedure but measure using m/z 30
• Almost all NO3 converts to NO when particles are captured
• NO2+ signal suggests
particles hit edges of vaporizer
![Page 16: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/16.jpg)
Velocity CalibrationHow to go from time (what we measure) to size (what we care about)
August 24, 2019 AMS Users Meeting Lund 16
![Page 17: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/17.jpg)
Velocity CalibrationHow to go from time (what we measure) to size (what we care about)
𝑓 𝐷𝑣𝑎 = 𝑉𝑔𝑎𝑠,𝑙𝑒𝑛𝑠 +𝑉𝑔𝑎𝑠,𝑒𝑥𝑖𝑡 − 𝑉𝑔𝑎𝑠,𝑙𝑒𝑛𝑠
1 + ൗ𝐷𝑣𝑎𝐷∗
𝑏
August 24, 2019 AMS Users Meeting Lund 17
![Page 18: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/18.jpg)
AMSDMA
Atomizer
Drier
Setup for velocity calibration
NH4NO3
This is identical to the lens alignment setup.
August 24, 2019 AMS Users Meeting Lund 18
![Page 19: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/19.jpg)
Velocity Calibration
We measure particle time-of-flight as a function of particle diameter.
August 24, 2019 AMS Users Meeting Lund 19
![Page 20: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/20.jpg)
Velocity Calibration
We measure particle time-of-flight as a function of particle diameter.
August 24, 2019 AMS Users Meeting Lund 20
![Page 21: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/21.jpg)
Velocity Calibration
𝑉𝑒𝑙𝑜𝑐𝑖𝑡𝑦 =𝑑𝑖𝑠𝑡𝑎𝑛𝑐𝑒
𝑝𝑡𝑜𝑓
215-series:Distance = 0.295 m
255-series:Distance = 0.395 m
August 24, 2019 AMS Users Meeting Lund 21
![Page 22: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/22.jpg)
Velocity Calibration
𝑓 𝐷𝑣𝑎 = 𝑉𝑔𝑎𝑠,𝑙𝑒𝑛𝑠 +𝑉𝑔𝑎𝑠,𝑒𝑥𝑖𝑡 − 𝑉𝑔𝑎𝑠,𝑙𝑒𝑛𝑠
1 + ൗ𝐷𝑣𝑎𝐷∗
𝑏
August 24, 2019 AMS Users Meeting Lund 22
![Page 23: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/23.jpg)
PM1 and PM2.5 lens velocity calibration comparison
August 24, 2019 AMS Users Meeting Lund 23
• NaNO3 has higher density (2.26 g cm-3), which allows us to measure higher Dvawith same Dmob range
• Exercise caution interpreting measured ptof with PM2.5 for small particles; singly charged particles may not dominate or even be observed
![Page 24: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/24.jpg)
Summary_VapT.pxp
August 24, 2019 AMS Users Meeting Lund 24
Standard Vaporizer – optimum temperature ~600C
![Page 25: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/25.jpg)
Capture Vaporizer – optimum temperature ~550C
8/24/2019 AMS Users Meeting 206 25
![Page 26: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/26.jpg)
Summary_VapT.pxp
August 24, 2019 AMS Users Meeting Lund 26
Red points – good thermocouple connectionBlue, Green, Pink – poor thermocouple connectionA disconnected thermocouple measurement can be off by 100s of degrees
Vaporizer power temperature relationship
![Page 27: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/27.jpg)
AMSDMA
Atomizer
Drier
Setup for vaporizer temperature calibration
NaNO3
This is identical to the lens alignment setup and the velocity calibration setup, but with NaNO3.
August 24, 2019 AMS Users Meeting Lund 27
![Page 28: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/28.jpg)
PToF traces for NaNO3 (m/z 30) give good temperature indication.
VapT_SN_100805.pxp
444°C
540°C
625°C
705°C
August 24, 2019 AMS Users Meeting Lund 28
![Page 29: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/29.jpg)
Fit falling edge of PToF with single exponential. Plot tau vs amps. PToF are narrow by
1.1 A, or 700 to 750 C. Vaporizer is glowing dull orange from back (front is cooler than
back).
Summary_VapT.pxp
August 24, 2019 AMS Users Meeting Lund 29
![Page 30: AMS/ACSM Calibrations Overviewcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · AMS/ACSM Calibrations Overview Phil Croteau –croteau@aerodyne.com August 24, 2019 AMS Users](https://reader036.fdocuments.us/reader036/viewer/2022070706/5e9e1e15bbdecc13af15613e/html5/thumbnails/30.jpg)
August 24, 2019 AMS Users Meeting Lund 30
Comparison of Standard Vaporizer and Capture Vaporizer heater power and temperature calibration
Weiwei Hu, CU, AMS Users Meeting 2016