Ali S. Saad Department of Biomedical Engineering

College of applied medical SciencesKing Saud University

Biopotentials & electrodes

Page 2: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Recording of action potential of an invertebrate nerve axon (a) An electronic stimulator supplies a brief pulse of current to the axon, strong enough to excite the axon. A recording of this activity is made at a downstream site via a penetrating micropipet. (b) The movement artifact is recorded as the tip of the micropipet drives through the membrane to record resting potential. A short time later, an electrical stimulus is delivered to the axon; its field effect is recorded instantaneously at downstream measurement site as the stimulus artifact. The action potential proceeds along the axon at a constant propagation velocity. The time period L is the latent period or transmission time from stimulus to recording site.

Page 3: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

The role of voltage-gated ion channels in the action potential. The circled numbers on the action potential correspond to the four diagrams of voltage-gated sodium and potassium channels in a neuron's plasma membrane.

tMem

bran

e po

tent

ial

(mV

)

+50

0

- 50

- 100

D epolarizing phase

1

R esting phase1

2

R epolarizing phase3

U ndershoot phase4

N a+

O utside ce ll

Ins ide ce ll

P lasm a m em brane+ + + + + + + + +

- - - - - - - - -

K +

+ + + + + + + + +

+ + + + + + + + +

+ + + + + + + + +- - - - - - - - -

- - - - - - - - -

- - - - - - - - -

K +

K +

K +

N a+N a+

N a+

2 3

4

Page 4: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Theoretical action potential and membrane ionic conductance changes for sodium (gNa) and potassium (gK) are obtained by solving the differential equations developed by Hodgkin and Huxley for the giant axon of the squid at a bathing medium temperature of 18.5 ºC. ENa and EK are the Nernst equilibrium potentials for sodium and potassium across the membrane.

Page 5: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Diagram of network equivalent circuit of a small length (z) of an unmyelinated nerve fiber or a skeletal muscle fiber The membrane proper is characterized by specific membrane capacitance Cm (F/cm2) and specific membrane conductances gNa, gK, and gCl in mS/cm2 (millisiemens/cm2). Here an average specific leakage conductance is included that corresponds to ionic current from sources other than Na+ and K+ (for example, Cl-). This term is usually neglected. The cell cytoplasm is considered simply resistive, as is the external bathing medium; these media may thus be characterized by the resistance per unit length ri and ro (/cm), respectively. Here im is the transmembrane current per unit length (A/cm), and i and o are the internal and external potentials at point z, respectively.

Page 6: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

(a) Charge distribution in the vicinity of the active region of an ummyelinated fiber conducting an impulse. (b) Local circuit current flow in the myelinated nerve fiber.

CellNode of Ranvier

Myelinsheath

Activenode

Local closed (solenoidal)lines of current flow

Repolarizedmembrane

Axon

Restingmembrane

External medium

+-

+-

+-

+-

+-

+-

+-

+--+

-+

-+

-+

+-

+-

+-

+-

+-

+-

+-

+-

-+

-+

-+

-+

-+

-+

-+

-+

-+

-+

-+

+-

+-

+-

+--+

-+

-+

-+

-+

-+

-+

-+

+-

+-

+-

Active region

Depolarizedmembrane

(a)

(b)

Direction ofpropagation

Periaxonalspace

Axon + -

Page 7: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Extracellular field potentials (average of 128 responses) were recorded at the surface of an active (1-mm-diameter) frog sciatic nerve in an extensive volume conductor. The potential was recorded with (a) both motor and sensory components excited (Sm + Ss), (b) only motor nerve components excited (Sm), and (c) only sensory nerve components excited (Ss).

Page 8: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Schematic diagram of a muscle-length control system for a peripheral muscle (biceps) (a) Anatomical diagram of limb system, showing interconnections. (b) Block diagram of control system.

Page 9: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Measurement of neural conduction velocity via measurement of latency of evoked electrical response in muscle. The nerve was stimulated at two different sites a known distance D apart.

Reference

Velocity = u =

2 ms

V(t)S2

S2

S1

S1

Muscle

+ - + -

D

R

L2

L1- L2

L1

t DV(t)

V(t)

1 m

V

Page 10: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Sensory nerve action potentials evoked from median nerve of a healthy subject at elbow and wrist after stimulation of index finger with ring electrodes. The potential at the wrist is triphasic and of much larger magnitude than the delayed potential recorded at the elbow. Considering the median nerve to be of the same size and shape at the elbow as at the wrist, we find that the difference in magnitude and waveshape of the potentials is due to the size of the volume conductor at each location and the radial distance of the measurement point from the neural source.

Page 11: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

The H reflex The four traces show potentials evoked by stimulation of the medial popliteal nerve with pulses of increasing magnitude (the stimulus artifact increases with stimulus magnitude). The later potential or H wave is a low-threshold response, maximally evoked by a stimulus too weak to evoke the muscular response (M wave). As the M wave increases in magnitude, the H wave diminishes.

Page 12: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Diagram of a single motor unit (SMU), which consists of a single motoneuron and the group of skeletal muscle fibers that it innervates. Length transducers in the muscle activate sensory nerve fibers whose cell bodies are located in the dorsal root ganglion. These bipolar neurons send axonal projections to the spinal cord that divide into a descending and an ascending branch. The descending branch enters into a simple reflex arc with the motor neuron, while the ascending branch conveys information regarding current muscle length to higher centers in the CNS via ascending nerve fiber tracts in the spinal cord and brain stem.

Page 13: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Motor unit action potentials from normal dorsal interosseus muscle during progressively more powerful contractions. In the interference pattern (c ), individual units can no longer be clearly distinguished. (d) Interference pattern during very strong muscular contraction. Time scale is 10 ms per dot.

Page 14: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Distribution of specialized conductive tissues in the atria and ventricles, showing the impulse-forming and conduction system of the heart. The rhythmic cardiac impulse originates in pacemaking cells in the sinoatrial (SA) node, located at the junction of the superior vena cava and the right atrium. Note the three specialized pathways (anterior, middle, and posterior internodal tracts) between the Sa and atrioventricular (AV) nodes. Bachmann's bundle (interatrial tract) comes off the anterior internodal tract leading to the left atrium. The impulse passes from the SA node in an organized manner through specialized conducting tracts in the atria to activate first the right and then the left atrium. Passage of the impulse is delayed at the AV node before it continues into the bundle of His, the right bundle branch, the common left bundle branch, the anterior and posterior divisions of the left bundle branch, and the Purkinje network. The right bundle branch runs along the right side of the interventricular septum to the apex of the right ventricle before it gives off significant branches. The left common bundle crosses to the left side of the septum and splits into the anterior division (which is thin and long and goes under the aortic valve in the outflow tract to the anterolateral papillary muscle) and the posterior division (which is wide and short and goes to the posterior papillary muscle lying in the inflow tract).

Page 15: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.13 Representative electric activity from various regions of the heart. The bottom trace is a scalar ECG, which has a typical QRS amplitude of 1-3 mV. (© Copyright 1969 CIBA Pharmaceutical Company, Division of CIBAGEIGY Corp. Reproduced, with permission, from The Ciba Collection of Medical Illustrations, by Frank H. Netter, M. D. All rights reserved.)

Page 16: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.14 The cellular architecture of myocardial fibers Note the centroid nuclei and transverse intercalated disks between cells.

Page 17: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.15 Isochronous lines of ventricular activation of the human heart Note the nearly closed activation surface at 30 ms into the QRS complex. (Modified from "The Biophysical Basis for Electrocardiography," by R. Plonsey, in CRC Critical Reviews in Bioengineering, 1, 1, p.5, 1971, © The Chemical Rubber Co., 1971. Used by permission of The Chemical Rubber Co. Based on data by D. Durrer et al., "Total excitation of the Isolated Human Heart, "1970, Circulation, 41, 899-912, by permission of the American Heart Association, Inc.)

Page 18: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.17 Atrioventricular block (a) Complete heart block. Cells in the AV node are dead and activity cannot pass from atria to ventricles. Atria and ventricles beat independently, ventricles being driven by an ectopic (other-than-normal) pacemaker. (B) AV block wherein the node is diseased (examples include rheumatic heart disease and viral infections of the heart). Although each wave from the atria reaches the ventricles, the AV nodal delay is greatly increased. This is first-degree heart block. (Adapted from Brendan Phibbs, The Human Heart, 3rd ed., St. Louis: The C. V. Mosby Company, 1975.)

Page 19: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.18 Normal ECG followed by an ectopic beat An irritable focus, or ectopic pacemaker, within the ventricle or specialized conduction system may discharge, producing an extra beat, or extrasystole, that interrupts the normal rhythm. This extrasystole is also referred to as a premature ventricular contraction (PVC). (Adapted from Brendan Phibbs, The Human Heart, 3rd ed., St. Louis: The C. V. Mosby Company, 1975.)

Page 20: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.19 (a) Paroxysmal tachycardia. An ectopic focus may repetitively discharge at a rapid regular rate for minutes, hours, or even days. (B) Atrial flutter. The atria begin a very rapid, perfectly regular "flapping" movement, beating at rates of 200 to 300 beats/min. (Adapted from Brendan Phibbs, The Human Heart, 3rd ed., St. Louis: The C. V. Mosby Company, 1975.)

Page 21: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.20 (a) Atrial fibrillation. The atria stop their regular beat and begin a feeble, uncoordinated twitching. Concomitantly, low-amplitude, irregular waves appear in the ECG, as shown. This type of recording can be clearly distinguished from the very regular ECG waveform containing atrial flutter. (b) Ventricular fibrillation. Mechanically the ventricles twitch in a feeble, uncoordinated fashion with no blood being pumped from the heart. The ECG is likewise very uncoordinated, as shown (Adapted from Brendan Phibbs, The Human Heart, 3rd ed., St. Louis: The C. V. Mosby Company, 1975.)

Page 22: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.22 The transparent contact lens contains one electrode, shown here on horizontal section of the right eye. Reference electrode is placed on the right temple.

Page 23: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.23 Vertebrate electroretinogram

Page 24: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

The human corneal electroretinogram. A 10 ms flash in quanta/deg2 yields scotopic threshold respone (STR), the postphotoreceptoral b-wave, and the inner layer a-wave.

Page 25: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.27 (a) Different types of normal EEG waves. (b) Replacement of alpha rhythm by an asynchronous discharge when patient opens eyes. (c) Representative abnormal EEG waveforms in different types of epilepsy. (From A. C. Guyton, Structure and Function of the Nervous System, 2nd ed., Philadelphia: W.B. Saunders, 1972; used with permission.)

Page 26: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.28 The 10-20 electrode system This system is recommended by the International Federation of EEG Societies. (From H. H. Jasper, "The Ten-Twenty Electrode System of the International Federation in Electroencephalography and Clinical Neurophysiology," EEG Journal, 1958, 10 (Appendix), 371-375.)

Page 27: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 4.29 The electroencephalographic changes that occur as a human subject goes to sleep The calibration marks on the right represent 50 V. (From H. H. Jasper, "Electrocephalography," in Epilepsy and Cerebral Localization, edited by W. G. Penfield and T. C. Erickson. Springfield, IL: Charles C. Thomas, 1941.)

Page 28: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.1 Electrode-electrolyte interface The current crosses it from left to right. The electrode consists of metallic atoms C. The electrolyte is an aqueous solution containing cations of the electrode metal C+ and anions A-.

Page 29: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.2 A silver/silver chloride electrode, shown in cross section.

Page 30: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.3 Sintered Ag/AgCl electrode.

Page 31: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.4 Equivalent circuit for a biopotential electrode in contact with an electrolyte Ehc is the half-cell potential, Rd and Cd make up the impedance associated with the electrode-electrolyte interface and polarization effects, and Rs is the series resistance associated with interface effects and due to resistance in the electrolyte.

Page 32: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.5 Impedance as a function of frequency for Ag electrodes coated with an electrolytically deposited AgCl layer. The electrode area is 0.25 cm2. Numbers attached to curves indicate the number of mAs for each deposit. (From L. A. Gedders, L. E. Baker, and A. G. Moore, "Optimum Electrolytic Chloriding of Silver Electrodes," Medical and Biological Engineering, 1969, 7, pp.49-56.)

Page 33: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.6 Experimentally determined magnitude of impedance as a function of frequency for electrodes.

Page 34: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.7 Magnified section of skin, showing the various layers (Copyright © 1977 by The Institute of Electrical and Electronics Engineers. Reprinted with permission, from IEEE Trans. Biomed. Eng., March 1977, vol. BME-24, no. 2, pp. 134-139.)

Page 35: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.8 A body-surface electrode is placed against skin, showing the total electrical equivalent circuit obtained in this situation. Each circuit element on the right is at approximately the same level at which the physical process that it represents would be in the left-hand diagram.

Sweat glandsand ducts

Electrode

Epidermis

Dermis andsubcutaneous layer Ru

Re

Ese

Ehe

Rs

RdCd

EP

RPCPCe

Gel

Page 36: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.9 Body-surface biopotential electrodes (a) Metal-plate electrode used for application to limbs. (b) Metal-disk electrode applied with surgical tape. (c) Disposable foam-pad electrodes, often used with electrocardiograph monitoring apparatus.

Page 37: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.10 A metallic suction electrode is often used as a precordial electrode on clinical electrocardiographs.

Page 38: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.11 Examples of floating metal body-surface electrodes (a) Recessed electrode with top-hat structure. (b) Cross-sectional view of the electrode in (a). (c) Cross-sectional view of a disposable recessed electrode of the same general structure shown in Figure 5.9(c). The recess in this electrode is formed from an open foam disk, saturated with electrolyte gel and placed over the metal electrode.

Double-sidedAdhesive-tapering

Insulatingpackage

Metal disk

Electrolyte gelin recess

(a) (b)

(c)

Snap coated with Ag-AgCl External snap

Plastic cup

Tack

Plastic disk

Foam padCapillary loops

Dead cellular material

Germinating layer

Gel-coated sponge

Page 39: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.12 Flexible body-surface electrodes (a) Carbon-filled silicone rubber electrode. (b) Flexible thin-film neonatal electrode (after Neuman, 1973). (c) Cross-sectional view of the thin-film electrode in (b). [Parts (b) and (c) are from International Federation for Medical and Biological Engineering. Digest of the 10th ICMBE, 1973.]

Page 40: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.13 Needle and wire electrodes for percutaneous measurement of biopotentials (a) Insulated needle electrode. (b) Coaxial needle electrode. (c) Bipolar coaxial electrode. (d) Fine-wire electrode connected to hypodermic needle, before being inserted. (e) Cross-sectional view of skin and muscle, showing coiled fine-wire electrode in place.

Page 41: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.14 Electrodes for detecting fetal electrocardiogram during labor, by means of intracutaneous needles (a) Suction electrode. (b) Cross-sectional view of suction electrode in place, showing penetration of probe through epidermis. (c) Helical electrode, which is attached to fetal skin by corkscrew type action.

Page 42: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.15 Implantable electrodes for detecting biopotentials (a) Wire-loop electrode. (b) Silver-sphere cortical-surface potential electrode. (c) Multielement depth electrode.

Page 43: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.16 Examples of microfabricated electrode arrays. (a) One-dimensional plunge electrode array (after Mastrototaro et al., 1992), (b) Two-dimensional array, and (c) Three-dimensional array (after Campbell et al., 1991).

(c)

Tines

Base

Exposed tip

ContactsInsulated leads

(b)

Base

Ag/AgCl electrodes

Ag/AgCl electrodes

BaseInsulated leads

(a)

Contacts

Page 44: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.17 The structure of a metal microelectrode for intracellular recordings.

Page 45: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.18 Structures of two supported metal microelectrodes (a) Metal-filled glass micropipet. (b) Glass micropipet or probe, coated with metal film.

Page 46: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.19 A glass micropipet electrode filled with an electrolytic solution (a) Section of fine-bore glass capillary. (b) Capillary narrowed through heating and stretching. (c) Final structure of glass-pipet microelectrode.

Page 47: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.20 Different types of microelectrodes fabricated using microelectronic technology (a) Beam-lead multiple electrode. (Based on Figure 7 in K. D. Wise, J.B. Angell, and A. Starr, "An Integrated Circuit Approach to Extracellular Microelectrodes." Reprinted with permission from IEEE Trans. Biomed. Eng., 1970, BME-17, pp. 238-246. Copyright (C) 1970 by the institute of Electrical and Electronics Engineers.) (b) Multielectrode silicon probe after Drake et al. (c) Multiple-chamber electrode after Prohaska et al. (d) Peripheral-nerve electrode based on the design of Edell.

Bonding pads

Si substrateExposed tips

Lead viaChannels

Electrode

Silicon probe

Silicon chip

Miniatureinsulatingchamber

Contactmetal film

Hole

SiO2 insulatedAu probes

Silicon probe

Exposedelectrodes

Insulatedlead vias

(b)

(d)

(a)

(c)

Page 48: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.21 Equivalent circuit of metal microelectrode (a) Electrode with tip placed within a cell, showing origin of distributed capacitance. (b) Equivalent circuit for the situation in (a). (c) Simplified equivalent circuit. (From L. A. Geddes, Electrodes and the Measurement of Bioelectric Events, Wiley-Interscience, 1972. Used with permission of John Wiley and Sons, New York.)

Metal rod

Tissue fluidMembranepotential

N

C

B

B

EmpMembraneandactionpotential

Cma

Rma

Cd + Cw

Ema - Emb

E

0

A

(a)

N = NucleusC = Cytoplasm

(b)

(c)

B

A

A

Referenceelectrode

RmbRma

EmbEma

EmpRi Re

CmbCma

Cdi

Cd2

Rs Cw

Insulation

CdCellmembrane

++ +

++

+++

++

++++++++++++ - -

- -

- - ---

--- ------

---

-

Page 49: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.22 Equivalent circuit of glass micropipet microelectrode (a) Electrode with its tip placed within a cell, showing the origin of distributed capacitance. (b) Equivalent circuit for the situation in (a). (c) Simplified equivalent circuit. (From L. A. Geddes, Electrodes and the Measurement of Bioelectric Events, Wiley-Interscience, 1972. Used with permission of John Wiley and Sons, New York.)

Ema

Rma

Rt

Ri Re(b) Emp

Emb

RmbCmb

Ej

Et

Cma

Cd

A B

Rt

Em

A

B

Membraneandactionpotential

(c)

Emp

Em = Ej + Et + Ema- Emb

Cd = Ct0

Cellmembrane

Tip+-+

-+

++

--

+ -

-+ -

+-

+ + + + +- - - - -+

++

---

+--++++

---

Taper

Internal electrode

Glass

A BTo amplifier

Electrolyte inmicropipet

Stem

(a)

Referenceelectrode

Cell membrane

CytoplasmN = Nucleus

N

Environmentalfluid

Cd

Page 50: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Figure 5.23 Current and voltage waveforms seen with electrodes used for electric stimulation (a) Constant-current stimulation. (b) Constant-voltage stimulation.

(a)

Polarizationpotential

Polarization

Polarization

i

i

t

t

t

t

Ohmicpotential

(b)

Polarizationpotential

Page 51: Ali S. Saad Department of Biomedical Engineering College of applied medical Sciences King Saud University Biopotentials & electrodes.

Questions for Biopotentials & electrodes. You should be able to:

Explain the origin of biopotentials Explain nerve propagationExplain cardiac excitationExplain cardiac arrhythmiasDescribe brain wavesExplain electrode offset voltagesExplain electrode polarizationModel electrode impedance versus frequencyMinimize skin-electrode motion artifactsDescribe good electrode constructionDescribe microelectrodesSketch electrode voltages during current flowJ. G. Webster (ed.), Medical instrumentation: application and design, 3rd ed., John Wiley & Sons, 1998.