Alhussein Abuhoza, S. Biswas, U. Frankenfeld, J. Hehner, C. Schmidt
GSI, Darmstadt, Germanyand
H.R. Schmidt University of Tübingen, Germany
Construction and commissioning of a setup to study ageing phenomena in
high rate gas detectors
TIPP14 Amsterdam (2-6.6.2014)
![Page 2: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/2.jpg)
Introduction What is the detector ageing? Why ageing studies?
Ageing setupSetup-ISetup-II
Summary
2
Plan
![Page 3: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/3.jpg)
3
Introduction What is the detector ageing? Why ageing studies?
Ageing setupSetup-ISetup-II
Summary
![Page 4: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/4.jpg)
Symptoms
Losses in gas gain
Decrease of resolution
or efficiencyExcessive currents Etc...
Detector ageing
It is the deterioration of the detector performance.
4
![Page 5: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/5.jpg)
5
Introduction What is the detector ageing? Why ageing studies?
Ageing setupSetup-ISetup-II
Summary
![Page 6: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/6.jpg)
~1.5 C/cm2 Estimated accumulated charge in the fourth layer of
Muon Chamber (MuCh) in CBM Ex. in
10 years.
~300 C/cm2 Estimated
accumulated charge in the first
layer of Muon Chamber (MuCh)
in CBM Ex. in 10 years.
~11 C/cm2 Estimated
accumulated charge at TRD1 in CBM Ex. in 10
years.
Experiment like Compressed Baryonic Matter (CBM)
at future FAIR facility at Darmstadt, Germany needs
detectors with long term (~10 years) stable operation
at high rates.
6
![Page 7: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/7.jpg)
7
Introduction What is the detector ageing? Why ageing studies?
Ageing setupSetup-ISetup-II
Summary
![Page 8: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/8.jpg)
Ageing setup
1. Studying ageing phenomena of gaseous detectors.2. Investigating the ageing properties of the materials used
for the construction of gaseous detectors.3. Monitoring the performances of gaseous detectors, in
particular those used in CBM.
8
![Page 9: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/9.jpg)
9
Setup-I
• Two MWPCs in serial were used in setup-I.
• A mixing station fids one gas line.• One sensor for temperature and one
for pressure.• The MWPC configured of two cathodes
(Al-Kapton foils) and on anode (multi-wire plane).
![Page 10: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/10.jpg)
10
Setup-I (Test approach)
t
t
Q Q
MaMa
Mr Mr
Fully Automatic
Ageing Chamber
Reference Chamber
High rate
X-ray55Fe
monitor
X-ray
Ageing Ch.Refernce Ch.
Nor
mal
ized
gai
n
∑Q
No ageing
Ageing
![Page 11: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/11.jpg)
Gain in MWPC
11
Ageing study :1. Gain is the major parameter to monitor detector performance.
2. Gain normalization: Gain ratio of two detectors (ageing and reference).
Extrapolating gas density effect by fitting approach (needs precise measurement of gas temperature and pressure).
3. Variation of normalized gain should be < 1%
p/T
G
![Page 12: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/12.jpg)
12
1. Cathode of Al-Kapton foil has been replaced with a multi-wire plane
2. Premixed gas bottle will be used instead of a mixing station (temperature effect on the MFCs).
3. Moving platforms have to be presented (to avoid the continuous exposure of 55Fe-source and to test two materials or more in parallel).
4. Metal housing of chamber instead of fiberglass .5. Individual gas lines (to avoid any pressure drop in-between the chambers).
Findings from setup-I
[1] A Abuhoza, S Biswas, U Frankenfeld, J Hehner, CJ Schmidt, and HR Schmidt. Physics Procedia, 37:442–447, 2012.[2] A Abuhoza, HR Schmidt, S Biswas, U Frankenfeld, J Hehner, and CJ Schmidt. NIM Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2012.
![Page 13: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/13.jpg)
Introduction What is the detector ageing? Why ageing studies?
Ageing setupSetup-ISetup-II
Summary
13
![Page 14: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/14.jpg)
Setup-II (components)
14
![Page 15: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/15.jpg)
15
Setup-II• Three parallel lines (two for ageing test, one reference).• Premixed gas (no mixing station).• Two motion platforms (collimator and chambers).• Metal housing of the chambers.• Multi wire anode + multi wire cathode.
(components)
![Page 16: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/16.jpg)
16
Setup-II
Dimension of setup box : 110×130×83 cm
Dimension of MWPC: 20×20 cmActive area: 10×10 cm LabView program operates all
the functions automatically
![Page 17: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/17.jpg)
17
Setup-II (Gas system)
Outgassing box
Ageing material
![Page 18: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/18.jpg)
18
Setup-II(Test procedure)
motor
motor
CH-0CH-2CH-1
x-ray
CH-1
CH-0
CH-2
tN
orm
aliz
ed g
ain
∑Q
ageing
No ageing
1. 55Fe2. X-ray (filter)
![Page 19: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/19.jpg)
19
Commissioning and ageing tests of setup-II
Ageing and recovery behavior in the first test.
Investigation of setup cleanness.
Improving of gain normalization.
Ageing tests.
![Page 20: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/20.jpg)
Ageing and recovery behavior in the first test.
In this setup all the components are new:+ Detector Lab (D.L.) global gas-line + Mass flow Controller (MFC)+ rigid steel tubes inside the lab + outgassing box (in line-1 and line-2) + flexible steel tube (inlet) + detector + flexible steel tube (outlet)
55Fe1mm
55Fe+x-ray1mm+3cm
Effect of the high rate of x-
ray ??
Where is the contamination source in the new setup??
ΔG=7%
ΔG=4%
20
![Page 21: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/21.jpg)
Investigation of contamination source
21
![Page 22: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/22.jpg)
Investigation of contamination source
(ΔG/G0)CH-0 = 5%(ΔG/G0)CH-1 = 11%
Using two lines
22
![Page 23: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/23.jpg)
Investigation of contamination source
23
![Page 24: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/24.jpg)
Investigation of contamination source
• Two ageing tests using gas scheme-2 revealed that there is still a source of ageing.
ΔG/G0 = 2.5%
Using one line
ΔG/G0 = 20%
24
![Page 25: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/25.jpg)
Investigation of contamination source
25
![Page 26: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/26.jpg)
Investigation of contamination source
Gas system: ball gas flow regulator + flexible steel tube + flexible steel tubes (inlet) + detector + flexible steel tubes (outlet) + outgassing box (some tests)
ΔG/G0 = 5%
26
![Page 27: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/27.jpg)
Investigation of contamination source
27
![Page 28: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/28.jpg)
Investigation of contamination source
Gas system: ball gas flow regulator + copper tube (inlet) + detector + copper tube (outlet)
28
Flexible tubes were cleaned by the supplier with a cleaning and degreasing agent is called
commericialy “A062-K21 hebro®lan TN” contains:
• Phosphoric acid• Alkyl alcohol ethoxylate• Methylethyl ester• Amins• Alhkyldmiethyl• N-oxides
![Page 29: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/29.jpg)
Investigation of contamination source
Cleaning procedures of flexible tubes:1. Distilled water, iso-propanol and dry air are pumped through
the rigid and flexible tubes (without taking them out from the setup).
2. Ultrasonic bath + iso-propanol pumping.
Gain variation = 0.3%
ΔG/G0 = 5%
29
![Page 30: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/30.jpg)
Improving of gain normalization .
30
![Page 31: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/31.jpg)
• RTV-3145
• Al-tape (Gerband 705)
Area=135 cm2
R= 300 %/C/cm
Area=135 cm2
R= 3180 %/C/cm
• Ageing test of RTV-3145 and Al-tape (Gerband 705)
31
![Page 32: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/32.jpg)
Summary
32
Precise measurement of ageing are necessary to implement studies in reasonable time and in real conditions.
Very clean and stable apparatus dedicated to ageing studies is operational in GSI DetLab.
Ageing acceleration by x-ray needs to be investigated. Obtained gain stability is < 0.5% peak-to-peak. Cleaning and degreasing agent “A062-K21 hebro®lan TN” reveals
ageing effect on the MWPC. Ageing test for two materials have been carried out (accelerated by x-
ray generator): RTV-3145 : R=3180 %/C/cm Gerband 705: R=300%/C/cm
![Page 33: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/33.jpg)
33
![Page 34: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/34.jpg)
•Back up
34
![Page 35: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/35.jpg)
35
R= 960 %/C/cm2 , Q=0.117 C/cm2
R= 92 %/C/cm2 , Q=0.037 C/cm2
![Page 36: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/36.jpg)
10 years = 10*2*30*24= 14400 hQ=e·n·r·G·t
36
![Page 37: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/37.jpg)
37
• Polyacrylat
![Page 38: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/38.jpg)
Ageing tests:
1 min 1 min 1 min + 40 min x-ray
Ch-2Ch-1Ch-0
38
![Page 39: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/39.jpg)
![Page 40: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/40.jpg)
Residual instability has been observed due to
change in the gas density and detector geometry
resulting from the change of ambient parameters (T,
P etc.)
40
![Page 41: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/41.jpg)
Modification in the design
Inner windows
Outer windows
41
To avoid the residual fluctuation of the gain resulting from the change of geometry of the
chamber, extra outer windows have been added with two holes in inner windows
• Diethornís formula
![Page 42: Alhussein Abuhoza , S. Biswas , U. Frankenfeld, J. Hehner , C. Schmidt](https://reader031.fdocuments.us/reader031/viewer/2022013012/568164eb550346895dd7573e/html5/thumbnails/42.jpg)
Unstable proportionality of the gain with T/p ratio
T/P (K/bar)
Stable proportionality of the gain with T/p ratio
T/P (K/bar)
With outer windows
Very stable operation after
correction
Unstable after correction
Without outer
windows
42