Agilent HPLC Hardware and solution
January 2015
1
Xavier Maes
Product Specialist HPLC
1290Infinity II LC
LSPD Portfolio Overview
Page 2
1260
Binary
1220
Gradient
1220
Isocr.
SFC1260 SFC, SFC/MS, SFC/FID
1260 Hybrid SFC/UHPLC-MS
Bio-inert LC1260 Bio-inert LC + FC
1260 Multi-Detector Bio-SEC
1290 Infinity II LC
1260
Quat.
1260
Isocr.
Dissolution708/709, 400-DS, BIO-DIS,
850-DS, UV-Dissolution, MQ
App 1, 2, 3, 5, 6, 7
High-end UHPLC
Core UHPLC
HPLC-Chip/MSUHC-NCE,PhosphoPeptide
mAB-Glyco, HiP Protein ID
Lowflow LC1260 Cap/2D-LC
1260 Nano
GPC-SEC1260 GPC, 1260 MDS,
GPC 50, GPC 220
Electrophoresis7100 CE, 2100 Bioanalyzer,
2200TapeStation,3100Offgel
Preparative LC1260 uS, 1260 AS, 1260 PS
218, SD1, A2Prep
HDR-DAD
High-Speed Flexible
Workflow Automation Solutions I2I Method Transfer / ISET, 2D-LC, Method Dev., Multi-method/Walk-up, HT LC/MS, ACR,
Online-SPE, Online Sample Prep, Online Sampling, Impurity Analyzer, A2Prep, Mobile LC…
~ 100 LC Modules
~ 30 Valves
~ 70 Accessories/Kits
~ 30 SW
Agilent LC Division
January 2015
3
Since 1970, the complete
responsibility for all HP and
Agilent LC products is located in
Waldbronn, Germany:
• Product Management
• Research & Development
• Production & Order Fullfilment
• Applications Development
• Marketing Communication
Full module compatibility
• Mix 1220, 1260, 1290 Infinity LC and 1290
Infinity II LC with 1100 and 1200 Series –
and future systems !
Stepwise upgrade – now and in future:
• Exchange Detector only –
for higher sensitivity and detection speed.
• Exchange Pump and Autosampler only –
for ultra-high resolution LC on latest
1.8 µm rapid resolution column technology
• Exchange Column Compartment only –
to upgrade to a fully automated multi-
method or method development solution
supporting up to 8 columns and 24 solvents.
1200 Infinity Series – Agilent Value PromiseInfinitely Higher Investment Protection
January 2015
4
Page 4
0
200
400
600
800
1000
1200
0 1 2 3 4 5
bar
mL/min
Confidential
Agilent 1200 Infinity & Infinity II Series Pumps- Power Range
1300
600
2 5
1220 / 1260 Infinity LC – HPLC Standard
1220 / 1260 Infinity VL
1290 Infinity LC
1290 Infinity II LC
up to 10mL/min @ 200 bar
The Agilent 1220 Infinity LC
January 2015
6
Page 6
Agilent 1220 Infinity LCProduct information
600 bar for UHPLC Capabilities
Isocratic or Gradient Pump:• flow range: 0.2 - 10 mL/min
• low pressure mixing
• integrated degasser
• flow precision 0.07 % RSD
Manual Injector or Autosampler:• 600 bar
• injection range: 0.1-100 µL
• 100 x 2 mL vials or 30 x 6 mL vials
• injection cycle time 50 s
• carryover: < 0.05%
Column Oven:• up to 80°C
• one column, max. 250 mm
VWD or DAD:• 80 Hz data rate
• wavelength range: 190 - 600 nm (VWD)
up to 950 nm (DAD)
• 2.5 µAU (7µAU for DAD) noise at 230 nm
January 2015
7
Integrated, yet flexible…
January 2015
8
Different configurations for different needs available!
Many upgrade paths available!
G4286B G4288B G4290B G4294B
Expandability
January 2015
9
External tray for walk-up:
Together with Agilent‘s Walkup software the 1220 Infinity LC with external tray
is a perfect solution for walk-up environments!
Additional detectors:
Almost all 1260/1290 Infinity detectors and SQ-MS can be added!
From HPLC to UHPLC on the 1220 Infinity LCVarying Particle Size and Column Length
3/6/2016
1220 Infinity LC Portfolio
10
Column: 3 x 150 mm, 5 µm
Tg = 9 min
Resolution peak 6 = 1.23
Run time = 7.5 min
Flow rate: 1.2 mL/min at 40°C
Column: 3 x 50 mm, 1.8 µm
Tg = 3 min
Resolution peak 6 = 1.09
Run time = 3.2 min
min1 2 3 4 5 6 7
mAU
0
20
60
80
100
120
140
40
Fro
m H
PL
C t
o U
HP
LC
2.4 mL/min at 510bar
3x50 mm, 1.8µm, column at 50°C
Resolution for peak 6 = 2.53
From HPLC to UHPLC on the 1220 Infinity LCVarying Flow Rate, adapting gradient time and column temperature
3/6/2016
1220 Infinity LC Portfolio
11
min0 0.5 1 1.5 2 2.5 3
mAU
0
10
20
30
40
50
60
70
min0 0.5 1 1.5 2 2.5 3
mAU
0
10
20
30
40
50
60
1.2 mL/min at 315bar
3x50 mm, 1.8µm column at 40°C
Resolution for peak 6 = 1.09
Fro
m H
PL
C t
o U
HP
LC
UHPLC with 1220 Infinity LC and Poroshell columns
3/6/2016
1220 Infinity LC Portfolio
12
1220 Infinity LC1260 components in compact housing
Easy Method Transfer
Technical Overview:
„Seamless method transfer from the Agilent 1200 Series to the Agilent 1220 and 1260 Infinity LC systems”
January 2015
13
1220
1260
1100
1200
1200 RRLC
min0 1 2 3 4
mAU
0
200
400
600
800
1000
1200
1400
1200 Series LC
1220 Infinity LC
1260 Infinity LC
RT shift <5%
The Agilent 1260 Infinity LC
January 2015
14
Page 14
1260 Infinity LC - Most flexible LCStandard analytical LC modules
30+ different 1260 Infinity LC modules to chose from and build up YOUR specific solution
March 6, 2016
15
1260 Infinity Quaternary Pump
1260 Infinity Binary Pump
1260 Infinity Autosampler 1260 Infinity Column Compartment 1260 Infinity Diode Array
Detector VL +
1260 Infinity RI Detector
1260 Infinity FLD Spectra
1260 Infinity Variable
Wavelength Detector
1260 Infinity ELSD Detector
1260 Infinity Multisampler
1260 Infinity Manual Injector
Valve options
UHPLC – Advantages & Benefits
Real UHPLC separation
performance (600 bar) benefitting
from the usage of sub-two-micron
(STM) column technology:
• Higher resolution, efficiency and
sensitivity
• Faster analysis - shorter run times –
faster turnaround times
• State-of-the-art, future-proof
instrument being more competitive
• Reduced column diameter which
helps to save solvent
March 6, 2016
16
4.6 x 150, 1.8 µm
Rs = 1.80 (+ 57%)
S/N = 44
7 Impurities
All 7 Baseline
separated!
4.6 x 150, 5 µm
Rs = 1.15
S/N = 42
4 Impurities
2 Not baseline
separated!
4.6 x 150, 3.5 µm
Rs = 1.37
S/N = 50
7 Impurities
6 Not Baseline
separated!
min0 5 10 15 20
mAU
-1
-0.5
0
0.5
1
1.5
2
2.5
DAD1 A, Sig=254,4 Ref =of f (051119A\SIG10003.D)
Isocratic Impurity Method
Column: 4.6 x 150 mm, 5 µm
The value of UHPLC in Impurity ProfilingNumber of theoretical plates – Isocratic analyses
March 6, 2016
17
1100/1200
Series HPLC
Quaternary System
1220/1260 Infinity
LC
Quaternary System
1200 Series
RRLC – Std.
1260
Infinity LC
Binary System
Max Flow Rate 10 mL/min 10 mL/min 5 mL/min 5 mL/min
Delay Volume 800-1100 µL 600-900 µL 600-800 µL*600-800 ( up to 900
with VL) µL*
Capillary ID 0.17 mm 0.17 mm 0.17 mm 0.17 mm
Disp. Vol. w/o cell 15 µl 15 µl 15 µl 15 µl
Pressure Range 400 bar 600 bar 600 bar 600 bar
Injection Principle Variable Loop Variable Loop Variable Loop Variable Loop
Inj.Vol. – Std/Ext. 100 / 1500 µL 100 / 1500 µL 100 / 1500 µL 100 / 1500 µL
Area RSD <0.25 % <0.25 % <0.25 % <0.15 %
Max Column Length 300 mm 300 mm 300 mm 300 mm* Smaller delay
volumes possible
Agilent 1260 Infinity LC (full method transferability)I-to-I Method Transferability by Design
Almost
unchanged
critical
specifications
March 2015
Agilent Confidential
18
100% Method TransferabilitySwitching from 1100 Series to 1220/1260 Infinity LC
100% HPLC compatibility: On one system you can run your established and
validated HPLC methods – unchanged – as well as push UHPLC to new levels of
performance.
March 6, 2016
19
The Agilent Value PromiseHigh Investment Protection Offering Stepwise Upgrade
Full module compatibility
• Mix 1220, 1260, 1290 Infinity LC, 1290
Infinity II with 1100 and 1200 Series – and
future systems !
Stepwise upgrade – now and in future:
• Exchange Detector –
for higher sensitivity and detection speed.
• Exchange Column Compartment –
to upgrade to a fully automated multi-method
or method development solution supporting
up to 8 columns and 24 solvents
• Exchange Pump and Autosampler – for
ultra-high resolution LC on latest 1.8 µm
column technology and lowest carry-over
March 6, 2016
20
Stepwise Upgrade - Detector
Agilent 1260 Infinity Series
Diode Array Detector
Benefits
High sensitivity (10x with high sensitivity
6 cm path length cell) but low cell volume.
• Better S/N, lower LOD and LOQ.
Reduced refractive index and thermal
effects.
• Flatter baseline, better and easier
integration results.
No special care instructions for flow cell
required.
Cartridge design for convenient handling.
March 6, 2016
21
History of Sensitivity Gain (DAD)- The last 30 years
Detector Model (Intro Year)
Sensitivity
Gain
(1988)(1986) (1995) (2006) (2010)
1.4 1.8 2.0 1.7
10
0.0
2.0
4.0
6.0
8.0
10.0
12.0
1040 to 1090 1040 to 1050 1040 to 1100 1040 to 1200 1040 to 1290/1260
Agilent 1260 Infinity
Diode Array Detector
March 6, 2016
22
Effects of path length increase
Peak Dispersion
60 mm pathlength
large geom. cell volume (78 µL)
lower light transmission
- Loss of resolution
- Loss of signal height
10 mm pathlength
small geom. cell volume (13 µL)
high light transmission
Conventional flow cells:
Page 23
60 mm pathlength
4 µL σV dispersion volume
Max-Light High Sensitivity cell:
Optofluidic waveguides
(total internal reflection)
March 6, 2016
23
1260/1290 Diode Array Detector Technical principle
Deuterium lamp
Max-Light Cartridge Cell
1290 DAD (G4212A)
Programmable slit
1024 element
diode-array
Grating
Mirror
Check DAD video on the website
1260 DAD (G4212B)
1290 DAD FS (G7117A)
NEW Fixed slit (4 nm)
March 6, 2016
24
11.6 x Higher Sensitivity1290/1260 Infinity DAD compare to 1200 Series VWD A/B
min0 1 2 3 4
mAU
0
5
10
15
20
25
30
Columns: 150 x 4.6mm Zorbax SB C18, 5µm
Sample: Anthracene: 835 pg/µL
Mobile phase: A: Water, B: Acetonitrile
Elution: isocratic 80 % B
Injection volume: 5 µL
Flow: 1.5 mL/min
DAD: 251/4nm, Ref= 450/80nm,
2.5Hz, slit width 4nm
VWD: 251 nm, 2.5 Hz
1290 DAD
60 mm
1200 VWD
10mm
Height [mAU] 28.87579 4.79998
Noise [mAU] 0.009806 0.01894
Signal/ noise 2944 253
Factor +11.6
1290/1260 Infinity DAD
1200 Series VWD A/B
March 6, 2016
25
HDR-DAD
• 30x wider linear UV range
• Quantification of widely different concentration levels in one
single run
1200 Infinity HDR-DAD
With HDR
>30X Linearity
March 6, 2016
26
History of DAD Linearity Gain The last 30 years
Detector Model (Intro Year)
1.4 1.7 2.0 1.42.9
> 30
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
45.0
1040 to 1090 1090 to 1050 1050 to 1100 1110 to 1200 1200 to 1290/60 1200 to HDR
Lin
ea
rity
Ga
in
(1988)(1986) (1995) (2006) (2010) (2011)
1200 Infinity HDR-DAD
With HDR
>30X Linearity
March 6, 2016
27
30x Wider Linear Range with HDR-DAD3.7 mm and 60 mm and Max-Light flow cell
Conc
[µg/mL]
Detector Signal
[AU/cm]
10 x higher sensitivity ~3x higher upper limit
30x lower Limit of Detection (LOD)
30x lower Limit of Detection (LOD) for impurity analysis
- for more reliable automated peak integration
- higher area precision
60 mm
3.7 mm
March 6, 2016
28
1290/1260 High Dynamic Range – HDR Increasing Linear UV Range by >30x
Technology
Cluster of two DADs
DAD 1: 60 mm Cell – for low concentrations
DAD 2: 3.7 mm Cell – for high concentrations
Output: One, combined signal, normalized to 10 mm path
length (HDR range: 0.6 x 10-6 to 6.7 AU/cm) vs for 1200
Series: 7 x 10-6 to 2 AU/cm)
Control / Usability
Like standard 1290/1260 DAD: Control,
data analysis, reporting
Investment Protection
Existing 1290/1260 DAD can be upgraded
(by 2nd DAD)
Existing 1100/1200 Systems can be upgraded
(by two 1290 or 1260 DAD)
3.7mm
60mm
Clustered DAD
DAD 1:
DAD 2:
March 6, 2016
29
Fixed Dose CombinationsFrom Roche Switzerland
API 1
API 2
Fixed dose combinations:
A combination of two or more
APIs in a fixed ratio of dosesAgilent 1200 (black): API 2: 10 ml injected
Agilent 1200 (red): API 1: 100 ml injected
Infinity HDR-DAD (blue): API 1 and 2
simultaneously: 70 ml injected
March 6, 2016
30
Stepwise Upgrade - Column Compartment
Benefits
More flexibility with different valve
solutions.
• Column selection for method
development or multi-method
laboratories.
• Alternating column regeneration
für higher throughput.
• Sample clean-up, pre-con-
centration, sample stripping, etc.
for automated sample preparation.
Investment protection with generic valve
drive/valve head concept.
March 6, 2016
31
Agilent QuickChange Valve HeadsOverview
Available
• in various material
• for variable pressure and flow
ranges
• In addition, tailor made valves
are available on request!
March 6, 2016
32
Higher Sample Throughput
Optimizing analysis times through automated
column regeneration can increase sample
throughput significantly – up to two-fold!
The determination of trace components is
always a challenge in the analysis of drug
metabolites, environmental samples, or foods
and beverages. Enrichment of the analyte prior
to analysis is often required to achieve highest
sensitivity.
Sample cleanup is essential for samples with
complex matrices, such as biological fluids, food
extracts or wastewater. Before injection into an
LC or LC/MS system, the analytes of interest
must be separated from the sample matrix.
March 6, 2016
33
Faster Method ScoutingTo develop an appropriate LC method and determine
the optimum separation conditions – for example, for
chiral screening – samples must be analyzed on
several stationary phases. This process can be
accelerated significantly.
The 12-position/13-port valve can be used as a
solvent selector for analytical flow rates. It provides
access to up to 12 different mobile phases. Combined
with a column selection valve, this solvent selection
valve offers utmost flexibility for developing new
methods.
The design of the 2-position/6-port valve allows you to
switch between two flow paths. Hence it can be used
to direct the column effluent to another detector –
simply select the detector that is best suited for your
application.
March 6, 2016
34
Stepwise Upgrade - Pump and Autosampler
Benefits
• Run legacy methods from 1100 or
1200 Series instrument and get
identical results without method
changes.
• Usage of sub-2-micron particle
columns for better resolution, higher
speed and better sensitivity.
March 6, 2016
35
1260 Infinity MultisamplerExpanding capacity and flexibility and speed
March 6, 2016
36
Key Features Benefit
Max Pressure: 600 bar
• Highest flow rates > more speed
• Small ID columns, Longer columns > more
resolution
Dual-needle option• Fastest cylce times
• Various injection volumes side by side
Injection Volume at full power range
0.1 -20ul
0.1- 40 ul
0.1-100 ul (single stroke) up to 500µL Multi-
load
• Flexibility according to customers needs
• No need to change hardware for wide volume
range injections
Lowest carryover• 30 ppm (chlorhexidine)
• 9 ppm with Multi-wash option
Flexible drawer setup• Usability
• Scalability of sample capacity and container type
Supports vials, up to 16 wellplates and
other formats as Eppendorf-Cups
• Largest flexibility for sample containers
Cooling inside of module from 4°C to
ambient
• Upgrade with compressor based cooler (slide in)
Carry-over test „Chlorhexidine“
Carry-over: 2.44 ppm
Multi-wash option
one solvent only
Carry-over: 20ppm
Std. Needle flush
one solvent only
Blue: Chlorhexidine injection
Red: Blank run after chlorhexidine injection
37
1260 Infinity AutosamplerExpanding capacity and flexibility and speed
March 6, 2016
38
Key Features Benefit
Max Pressure: 600 bar
• Highest flow rates > more speed
• Small ID columns, Longer columns > more
resolution
Needle-wash station• Outer needle wash for carryover reduction
• Carryover >40 pp,
Injection Volume at full power range
0.1 µL – 100 µL
Extended injection range option 900 µL
analytical head 400 bar
• Flexibility according to customers need
Integrated Column Compartment (ICC)
option
• Optional upgrade, user exchangeable
• Column heating (+5°C above ambient too 80°C)
• Up to two columns (30 cm lenght
• 2 Versions: ICC with 3 µL or 6 µL heater volume
Walk-Up option
• External tray Option
• Up to 5 vials
• Automated loading and disposal
Vial trays (default 2 x 66 vials 2 mL)• 2 individual trays for loading during analysis
• Optional „classic vial tray 100
Cooling inside of module from 4°C to
ambient
• Upgrade with compressor based cooler (slide in)
The New Agilent 1290 Infinity II LC
New 1290 Infinity II VWD
New 1290 Infinity II Multisampler
New 1290 Infinity II High Speed Pump
New 1290 Infinity II Flexible Pump New Accessories New 1290 Infinity II ELSD
New 1290 Infinity II Multicolumn Thermostat New 1290 Infinity II DAD
New 1290 Infinity II DAD-FS
The New Agilent 1290 Infinity II LC
Highest retention time precision for reliablepeak identification
The 1290 Infinity II Flexible Pump is the only low-
pressure mixing pump that combines the ultimate
performance of a high-pressure mixing UHPLC pump with
the flexibility of a quaternary pump. This pump delivers
superior gradient performance over the entire composition
range from 1 to 99 %.
Wide power range up to 1300 bar (2ml/min)/800 bar
(5ml/min) for optimum use of all analytical columns
independent of inner-diameter or particle technology
Flexibility for all applications
Highest retention time precision for reliablepeak identification
Isocratic analysis
The 1290 Infinity II Flexible Pump is the only low-
pressure mixing pump that combines the ultimate
performance of a high-pressure mixing UHPLC pump with
the flexibility of a quaternary pump. This pump delivers
superior gradient performance over the entire composition
range from 1 to 99 %.
Performance and Flexibility
March 6, 2016
44
The four channels of the
1290 Infinity II Flexible Pump give me
highest flexibility at a reasonable
price. Plus the performance of a high-
pressure mixing binary pump.
Technology: The Agilent 1290 Infinity II Flexible Pump
The story of the 1290 Infinity II Flexible Pump is a story about Accuracy!!
January 2015
45
3 Crucial Points to achieve ultimate quaternary pump performance (low
pressure mixing!):
1. For best performance you must know your eluents
2. New InletWeaver using proven Jet Weaver technology
3. Pump intelligence - Active Damping - with high pump
resolution (~300 pL minimum step)
The 1290 Infinity Quaternary Pumpe offers the highest accuracy and
precision over the full application range!
More automation, less manual work
March 6, 2016
46
Manual mixing of different buffer
and/or modifier concentrations is
tedious and error-prone. BlendAssist
is doing that automatically and
reliably.
Technology: Blend Assist for the 1290 Infinity II Flexible
Pump
More automation, less manual work
March 6, 2016
47
Influence of TFA concentration on separation of impurity
2 and on peak shape of tramadol.
Blend Assist mixes different buffer/modifier
concentrations automatically from a stock solution!
And gives the same result as for a manually mixed
solution!
Agilent innovative ISET technology allows you to use legacy
methods of virtually any LC and achieve the same results on
latest 1290 Infinity II LCs. You can retire your old systems
without retiring your precious methods.
All 1290 Infinity II pumps have ISET included.
Non-disruptive transition by ISET
ISET not only takes care for the various delay volumes in
different LC systems but also for different mixing behavior
– this is only available from Agilent and make instrument-
to-instrument method transfer easier than ever before.
Non-disruptive transition by ISET
Method Transfer to UHPLC instrumentsImpact of delay volume and mixing behavior...
Page 50
Programmed
gradient step
Gradient slop/
Mixing behavior
Delay volume
1290 Infinity
Binary Pump
1260 Infinity
Binary Pump
time
response
What happens with a
programmed gradient?
Total delay volume of the system (sum
of capillaries, mixer, cells, valves..)
1200 Series
1290 Infinity
10 %
90 %
Programmed gradient
1290 Infinity LC gradient
1200 Series gradient
0 0.5 1 1.5 2 2.5 3 3.5
0
50
100
150
200
250
300
350
400
min
mAU
Injection
Same gradient conditions by moving
the gradient and by emulating the
gradient mixing behavior.
Page 51
Method Transfer Possibilities Today- C) Intelligent System Emulation Technology (ISET)
Select pump
Select sampler
Page 52
Intelligent System Emulation Technology ISETImplementation
Infinitely More Adaptive- Intelligent System Emulation Technology (ISET)
January 2015
53
Without ISETDifferent retention time
and resolution
With ISETEmulate other (U)HPLC system
resulting in similar retention time
and resolution.
54
Method transfer: Waters Alliance to 1290 Binary LC4.6 x 100 mm Zorbax Eclipse Plus, 5 µm, flow: 2 mL/min, 25-95% ACN-Water
Waters
Alliance
2695
345 bar
1290 Infinity
Binary LC
1200 bar
1290 Infinity
Binary LC
1200 bar
with ISET
55
Method Transfer: Shimadzu LC-20AB to 1290 Binary4.6 x 100 mm Zorbax EP C-18, 5 μm, 1.0 ml/min, 5-95% ACN-Water, 20 min, 40°C, compressibility A0.45-B1.2
Shimadzu
LC-20AB
1290 Infinity
Binary LC
1290 Infinity
Binary LC
with ISET
With the Instrument Control Frame- work (ICF) and
RapidControl drivers you can operate the 1290 Infinity II LC
under any Agilent CDS and under most 3rd party CDS.
Dealing with various CDS by ICF & RC-drivers
High capacity for different sample containers within small footprint
The 1290 Infinity II Multisampler takes a maximum load of
16 shallow well MTPs. Giving a maximum capacity of 6144
samples within the footprint of an Agilent LC stack.
Besides the maximum capacity also a high flexibility to
accept many different sample containers without re-
configuration of the system is possible.
Deep-well MTPs, large vials
Medium-well MTPs (96/384),
std. 2ml vials
Shallow-well MTPs (96/384)
High capacity, different sample containers
March 6, 2016
58
Technology: Sample hotel of the 1290 Infinity II
Multisampler
Lowest carryover for highest data quality
Wash inner needle
Flow through principle washes inner needle
thoroughly and constant with eluent.
Wash outer needle
Choice of 3 external solvents to program
washing steps for the outer needle.
Needle-seat back flush
Choice of 3 external solvents to back
flush the seat.
Using the new Multiwash feature you can flush the inside
and outside surfaces of the injection needle with up to three
different solvents, as well as backflush the needle seat.
Lowest carryover for highest data quality
The 1290 Infinity II Multisampler with multiwash reduces
carryover to less than 9 ppm!
Flexibility for all applications
The dual-needle setup of the 1290 Infinity II Multisampler
enhances flexibility by providing two injectors in a single
instrument. You can optimize one path for large volume
injections and the other for low delay volumes.
20µl
500µl
Fast injection cycles with dual-needle injection
By fitting the 1290 Infinity II Multisampler with a second
injection needle, you can reduce cycle times to mere seconds
– virtually eliminating wait times for conventional injections and
even for large-volume loading or flushing procedures.
Capacity per benchspace
January 2015
63
System Width 52 cm 86 cm 54 cm 96 cm 60 cm 40 cm
Vial Capacity 216 486 316 648 528 432
DW Capacity 4 12 6 12 7 8
SW Capacity 4 23 6 12 21 16
CompetitorThermo
Vanquish
Thermo
Vanquish plus
rack charger
Shimadzu
Nexera 30
ACMP
Shimadzu
Nexera Sil 30
AC plus rack
changer
Waters Acquity
plus Sample
Organizer
Agilent 1290
Infinity II with
1290
Multisampler
Flexibility for all applications
New 1290 Infinity II Multicolumn Thermostat with space
for up to eight short columns, four 30 cm columns or
any combinations plus column selection valves allows
simplest method-switching.
Lowest dispersion for highest resolution
1290 Infinity II Diode Array Detector
• Ultra-low dispersion flow cell
• Data rates up to 240 Hz for small peak width
even in ultra-fast separations
1290 Infinity II Multicolumn Thermostat
• Quick Change heat exchangers down to 1.0 µl
• A-Line Quick-Connect fittings
• Capillary side-outlets for shorter connections to
LC/MS
1290 Infinity II Multisampler
• Optimized needle seat
min0 0.1 0.2 0.3 0.4
mAU
0
10
20
30
40
50
60
The optimized new ULD flow cell and data rates up to 240
Hz deliver uncompromised chromatographic resolution –
even for your fastest separations.
Lowest dispersion for highest resolution
min0 0.1 0.2 0.3 0.4
mAU
0
20
40
60
80
100
Data rate = 240 Hz
Data rate = 20 Hz
More automation capabilities
Quick-Change valves for column regeneration, or sample
clean-up and enrichment. And many more applications!
More automation capabilities
Automated column and mobile phase screening with up to
8 columns within one column compartment and up to 26
solvents.
More automation capabilities
Method Scouting Wizard software for most easiest method
and sequence generation for any combinations of column,
solvent, gradient and temperature screening.
Agilent A-Line Quick Connect UHPLC column fittings for
truly dead-volume-free fluidic connections
• Tool-free connection up to 1300 bar
• Spring loaded design for easy and truly zero-dead
volume connection
• Removable and reusable for all column types
Lowest dispersion for highest resolution
Agilent A-Line Quick Connect UHPLC column fittings for
truly dead-volume-free fluidic connections
• Tool-free connection up to 1300 bar
• Spring loaded design for easy and truly zero-dead
volume connection
• Removable and reusable for all column types
Lowest dispersion for highest resolution
Agilent A-Line Quick Connect UHPLC column fittings for
truly dead-volume-free fluidic connections
• Tool-free connection up to 1300 bar
• Spring loaded design for easy and truly zero-dead
volume connection
• Removable and reusable for all column types
Lowest dispersion for highest resolution
Unique detection capabilities reveal more
The 60 mm path length of the Agilent Max-Light flow cell
enables the 1290 Infinity II DAD to achieve typical noise
levels of less than ± 0.6 μAU/cm for highest detection
sensitivity and lowest detection limits.
1290 Infinity II DAD with
High Sensitivity cell
Conventional detector
Unique detection capabilities reveal more
For your samples with widely different concentration levels,
the 1290 Infinity II High Dynamic Range DAD solution
offers a 30x wider linear range and up to 30x higher sensitivity,
allowing you to detect and quantify all components within a
single run.
Two wavelengths, less injections
March 6, 2016
75
The 1290 Infinity II VWD I can
measure at two wavelengths. I can
now use a VWD for samples that had
required a DAD so far.
Technology: 1290 Infinity II Variable Wavelength Detector
Wider dynamic range, less injections
March 6, 2016
76
With the 1290 Infinity II HDR-ELSD I
can reliably quantify minor and major
compounds in my sample in a single
run, even if they do not have a
chromophore.
Technology: 1290 Infinity II High Dynamic Range ELSD
Select your separation performance – match system power with sample
complexity at the flip of a switch between single-dimension UHPLC and
the ultimate chromatographic power of 2D-LC.
Flexibility for all applications
General thoughts in separation science
78
„Resolving power is what it is all
about in analytical separation
science.“
March 6, 2016
Bringing in a 2nd Dimension - how?
79
Pump Autosampler Column Detector
Pump
Detector
Co
lum
n
Valve
(„Sampler“)
1D-LC System
2D-LC System
+
1D-LC System
„Just like 1D-LC with a chemically selective detector“
March 6, 2016
2D-LC – Heart-cutting vs Comprehensive 2D-LC
Heart-cutting 2D-LC (LC-LC):
LC1
LC2
Parts of the 1D effluent are injected onto
2D system
Long 2D gradients possible good
data quality
Limited 2D information
80
March 6, 2016
2D-LC – Heart-cutting vs Comprehensive 2D-LC
Comprehensive 2D-LC (LCxLC):
LC1
LC2
1st peak from
1st dimension
2nd peak from
1st dimension
3rd peak from
1st dimension
The whole 1D effluent is injected onto
2D system
(Ultra)Short 2D gradients necessary
Good data quality with fast pumps &
detector
Full („comprehensive“) 2D information!
81
March 6, 2016
2D-LC – Heart-cutting vs Comprehensive 2D-LC
Comprehensive 2D-LC (LCxLC):
LC1
LC2
1st peak from
1st dimension
2nd peak from
1st dimension
3rd peak from
1st dimension
LC1
LC2
min
sec
82
March 6, 2016
Bringing in a 2nd Dimension - how?The unique 2D-LC valve
83
March 6, 2016
2D-LC MULTIPLE HEART CUTTING
A new tool closing closing the gap between comprehensive and
heart-cutting 2D-LC
March 6, 2016
Confidentiality Label
84
1290 Infinity 2D-LC SolutionMultiple heart-cutting
Heart-cutting Data Viewer
85
March 6, 2016
20min
4.5min
Peaks lost for extra 2nd dimension analysis
1D
2D 4.5min
1290 Infinity II 2D-LC Solution Multiple heart-cutting
86
March 6, 2016
20min
1D
2D
4.5min4.5min 4.5min 4.5min 4.5min
Unmatched multiple heart-cutting 2D-LC Usability
Pre-aligned loop-valve kits, just add
to the existing 2D-LC system
Online status monitoring
Smart Valve-Loop Setup with 12 loops
2D-LC valve + two 6/14 valves
87
March 6, 2016
1290 Infinity II 2D-LC Solution – ScalabilityMultiple heart-cutting
88
March 6, 2016
20min
4.5min
4.5min
4.5min
4.5min
4.5min
#2: 7.41 min#2: 7.41 min
#3: 9.80 min#3: 9.80 min#3: 9.80 min
#6: 12.58 min
#3: 14.39 min
#4: 12.58 min
#5: 14.52 min#5: 14.52 min
#4: 12.58 min
#5: 14.52 min#5: 14.58 min
Most intuitive software to set
up and edit methods within
seconds:
• 1st dimension gradient
• 2nd dimension gradient
• Gradient shift
• Time-segments
• Method parameter
• Method set-up calculator
• Reference chromatogram
overlay for heart-cutting
• Time-based or peak-based
mode
2D-LC Acquisition SoftwareOne easy-to-use software for all operation modes
89
March 6, 2016
New multi heart-cut data viewer for smartest heart-cut data evaluation and
automatic reporting
Data Analysis Software
90
March 6, 2016
1290 Infinity 2D-LC Solution – ScalabilityThe right solution from medium to highly complex samples
91
Heart-cutting 2D-LC
Sa
mp
le c
om
ple
xity
Multiple heart-cutting 2D-LC
Comprehensive 2D-LC
Comprehensive 2D-LC/(IMS)-MS/MS
API impurity
profiling
Complex
formulations
Bio-
pharmaceuticals
Natural products,
biological
samples....
March 6, 2016
Achiral-chiral 2D-LC of Pharmaceutical SubstancesIbuprofen (racemic)
92
1st Dim:
C18
2nd Dim:
Chiral column
Method Development for Impurity AnalysisPotential Co-elutions in Onedimensional Analysis
93
Main compound
ImpuritiesImpurities
• Impurities hidden underneath the main compound peak?
• Coelutions of impurities?
Method Development for Impurity AnalysisTime-based Heart-cutting of Main Compound and Impurities
94
Method Development for Impurity AnalysisHeart-cutting 2D-LC Reveals Co-elutions of Impurities
95
Method Development for Impurity AnalysisMultiple Cuts Across the Main Compound Peak
96
• Two impurities
completely hidden
under main
compound peak
• Partial coelution of
a third impurity
with the main
compound
January 2015
97