Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be...

40
Advances in Pacemakers and ICDs: MRI Compatible Devices Michael S. Panutich, M.D.,F.A.C.C. Newport Heart Hoag Hospital Newport Beach, CA

Transcript of Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be...

Page 1: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Advances in Pacemakers and ICDs: MRI Compatible Devices

Michael S. Panutich, M.D.,F.A.C.C. Newport Heart Hoag Hospital

Newport Beach, CA

Page 2: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Disclosures

• Biosense Webster: Consulting, Speaker’s Honorarium

• Biotronik: Consulting, Speaker’s Honorarium

2

Page 3: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis
Page 4: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

• MRI is the reference standard and imaging modality of choice for imaging in many fields of medicine

• Strong static, gradient and radiofrequency fields used to generate MR images may be detrimental to devices and potentially cause harm to patients.

• Recently, pacemakers specifically designed and tested for the MRI environment have been shown to be safe under certain conditions.

Background

Page 5: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Need for MRI Technology• MRI is the preferred imaging option for brain injuries,

stroke, cancer, orthopedic issues, and many others. • MRI usage is highest in the 55 - 80 age group, which

includes patients more likely to receive a cardiac device.¹

Page 6: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

CT vs MRI

6

MRI Superior for soft tissue Uses magnetic energy and RF Contrast (if used) has low risk Can be confining, noisy, scary High capital costs

CT Better for bony tissue Uses ionizing radiation (x-ray

dose is cumulative) Contrast (if used) Equipment costs less

Page 7: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

MRI Terminology

• Tesla – Measure of magnetic field strength – 70% of all MRI scanners are 1.5T

• Specific Absorption Rate (SAR) – Rate of RF energy absorption by a pt’s body – Measured in watts/kg

7

Page 8: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

MRI Scan Area

• The isocenter is used to position the patient in the MRI scanner (eye level or

hips)

• The Field of View around the isocenter includes the tissue which can be

scanned

• The Field of View varies per MRI manufacturer and model

88% of MRI scans are within field of view

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 9: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Example of MRI scan with isocenter on eye level

MRI Scan Area

isocenter eye level

C2

C3

C4C5C6C7

T1T2

T3

T4

T5

Real patient case with isocenter on eye level and Siemens MRI scanner. Patient length 5 feet 8 inches and scan taken within the ProMRI conditions.

Head, cervicle spine and upper thoracic spine (T4) can be examined

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 10: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Example of MRI scan with isocenter on hip level

MRI Scan Area

S1

L5

L4

L3

L2

Real patient case with isocenter on hip level and Siemens MRI scanner. Patient length 5 feet 8 inches and scan taken according the ProMRI conditions.

Lower lumbar spine can be examined

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 11: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

What could go wrong?

11

Page 12: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Real Device Problems with MRI

12

A 2-second pause was noted…

…diminished battery voltage was noted immediately after MRI…

Significant changes were reported in 9.4% of leads… 1.9% required a change in programmed output.

…increased capture threshold was noted post MRI.

Page 13: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Forces Applied by MRI Could Affect Device Systems

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 14: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Forces Applied by MRI Could Affect Device Systems

14

Medtronic Confidential. Internal Use Only. 29

Static Field –  30,000x Earth’s magnetic field

Static Field Coil

Gradient Coils

RF Coil

Powerful electromagnetic fields

Radio Frequency (RF) Field –  1000’s of watts peak power

Gradient Field –  Creates loud scanner noise

MRI: Three Powerful Fields

Page 15: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

15

Medtronic Confidential. Internal Use Only. 30

Static Gradient RF

Vibration

Case Heating

Force & Torque

Lead Heating

Stimulation

Device Interactions

Summary of Hazards and Risks

Forces Applied by MRI Could Affect Device Systems

Page 16: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

16

Mitigate MRI Risk

Page 17: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Mitigate MRI Risk

• ProMRI® and SureScan® solution – Designed without ferromagnetic components – Lead design – MRI programming mode – MRI scan exclusion zone

Page 18: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Mitigate MRI Risk• MRI testing challenge

– Filter out 64 MHz energy emitted by 1.5 T scanners – Heating at lead tip is the most challenging hazard

• Leads act as an antenna, converting energy into heat • Lead tip is highest risk because it has the smallest area and

direct tissue contact – Many inter-related factors influence the potential risk

of lead tip heating

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 19: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Medtronic Design

19

Medtronic Confidential. Internal Use Only. 33

Pacemaker Design Solutions

• Minimize ferromagnetic content

•  Isolated circuit board

• Hall sensor

• Optimize input circuitry

• Circuit component change

Unintended cardiac stimulation Device interactions Force and Torque

MP – 33

Page 20: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Mitigate MRI Risk

20

Medtronic Confidential. Internal Use Only. 37

5076 5086MRI

Lead Heating - Design Solution •  Lead inner conductor coil design mitigates lead heating

•  4 filars to 2 filars increases inductance and reduces heating

0 10 20 30 40 50 60 70 80 90

100

0 5 10 15 20 25 30 35 40 45 50

Perc

ent o

f Max

imum

Lead Path

5076 vs. 5086MRI Comparison

5076 - 52 5086 - 52

Page 21: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Mitigate MRI Risk

21

Medtronic Confidential. Internal Use Only. 38

5086MRI Lead Performance & Reliability •  5076 Lead Survivability, 99%

@ 8 years*

•  Materials identical to 5076**

•  5086MRI lead flex testing –  Connector / body –  Lead body –  Tip / body - 10 year

equivalent

•  5086MRI clinical implant experience starting Feb 2007 (928 leads implanted)

Lead Tip Connector Lead Body *Medtronic Product Performance Report - 2009 Issue 2 **Exception of MRI Marker band and electrode coating

Page 22: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Mitigate MRI RiskBIOTRONIK leads designed for MRI:

– Maximized inductance of conductor coil • Diameter and coil pitch are optimized • Capacitive coupling of cable

– Correct proximity to conductor

– Thousands of measurements are performed to make sure multiple lead device combinations are covered

• Various lead lengths and positions • Lead wrapping around device • Different implant sites

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 23: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Data

23

Page 24: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Medtronic Advisa

24

Page 25: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Medtronic Data

25MEDTRONIC CONFIDENTIAL – FOR INTERNAL USE ONLY

Advisa MRI Clinical Trial confirms safety in MRI environment1 Objectives: •  Prospective, randomized controlled, multi-center/263 patients •  Designed to confirm safety and effectiveness of the Advisa MRI SureScan

Pacing System

Results: •  Similar pacing capture thresholds changes between MRI and control groups •  No MRI related complications

1 Medtronic, Inc. Advisa MRI SureScan Pacing System clinical report, in support of FDA premarket approval.

Page 26: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Medtronic Data

26MEDTRONIC CONFIDENTIAL – FOR INTERNAL USE ONLY

SureScan is supported by extensive evidence and experience

• Industry proprietary computer model evaluated more than 400,000 scenarios1

• Over 2,600 patients studied in three prospective clinical studies2-4

• Available internationally for more than four years with over 100,000 SureScan devices sold5

• 10% SureScan patients scanned at 18 months post-implant in the United States6

1 Park SM, et al. A path to safe MRI scanning of cardiac pacemaker patients: A role for computer modeling. WCHD. 2012;AB241.

2 Wilkoff BL, et al. Heart Rhythm. 2011;8:65-73. 3 Medtronic, Inc. Advisa MRI SureScan Pacing System clinical report, in support of FDA premarket approval.

4 Clinicaltrials.gov NCT01299675. 5 Medtronic data on file. November 27, 2012. 6 Data on file. November 5, 2012.

Page 27: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Biotronik Entovis

27

Page 28: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Pro-MRI Phase A

28

Atrial Threshold

Ventricular Threshold

Atrial Sensing

Ventricular Sensing

Page 29: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Pro-MRI Phase B

16 Not available for sale in the USA. CAUTION – Investigational Device. Limited by United States Law to Investigational Use.

Phase B Results

Five primary endpoints were evaluated to establish safety and efficacy of the ProMRI system: Primary Endpoint 1: SADE Free Rate

!  A rejection of the null hypothesis would demonstrate that there is evidence that the SADE-free rate at 1 month post-MRI is greater than 90%

!  One SADEs was adjudicated as related or possibly related to both the implanted pacing system and the MRI procedure resulting in an SADE-free rate of 99.5% (202/203), p < 0.001, 95% CI: (97.3%, 100.0%)

!  The null hypothesis is rejected and Primary Endpoint 1 is met

Page 30: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Pro-MRI Phase B

30

17 Not available for sale in the USA. CAUTION – Investigational Device. Limited by United States Law to Investigational Use.

Phase B Results

Primary Endpoint 2: Atrial Pacing Threshold

!  A rejection of the null hypothesis would demonstrate evidence that pPT (atrial) is greater than 95%

!  The null hypothesis is rejected and Primary Endpoint 2 is met

Results N=178

P Value and CI

Difference in atrial pacing threshold (V)

Mean ± SD Range

0.00 ± 0.12 -0.7, 0.4

Proportion of subjects with atrial pacing threshold success

178 (100.0%)

P<0.001

(97.9%, 100.0%)

18 Not available for sale in the USA. CAUTION – Investigational Device. Limited by United States Law to Investigational Use.

Phase B Results

Primary Endpoint 3: Ventricular Pacing Threshold

!  A rejection of the null hypothesis would demonstrate evidence that pPT (ventricular) is greater than 95%

!  The null hypothesis is rejected and Primary Endpoint 3 is met

Results N=189

P Value and CI

Difference in ventricular pacing threshold (V)

Mean ± SD Range

0.01 ± 0.10 -0.2, 0.3

Proportion of subjects with ventricular pacing threshold success

189 (100%)

P < 0.001, (98.1%, 100.0%)

19 Not available for sale in the USA. CAUTION – Investigational Device. Limited by United States Law to Investigational Use.

Phase B Results

Primary Endpoint 4: P-wave Sensing Attenuation

!  A rejection of the null hypothesis would demonstrate evidence that the P-wave attenuation free rate is greater than 90%

!  The null hypothesis is rejected and Primary Endpoint 4 is met

Results N=157

P Value and CI

Difference in P-wave sensing amplitude between 1-Month post-MRI and pre-MRI (mV)

Mean ± SD Range

-0.21 ± 0.89 -2.7, 2.8

Proportion of subjects with attenuation-free P-wave sensing

154 (98.1%)

P < 0.001, (94.5%, 99.6%)

20 Not available for sale in the USA. CAUTION – Investigational Device. Limited by United States Law to Investigational Use.

Phase B Results

Primary Endpoint 5: R-wave Sensing Attenuation

!  A rejection of the null hypothesis would demonstrate evidence that the R-wave attenuation free rate is greater than 90%

!  The null hypothesis is rejected and Primary Endpoint 5 is met

Results N=173

P Value and CI

Difference in R-wave sensing amplitude between 1-Month post-MRI and pre-MRI (mV)

Mean ± SD Range

-0.26 ± 1.37 -6.1, 4.5

Proportion of subjects with attenuation-free R-wave sensing

173 (100.0%)

P < 0.001, (97.9%, 100.0%)

Atrial Threshold Atrial Sensing

VentricularThreshold

Ventricular Sensing

Page 31: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Conditions for MRI

31

Page 32: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

MRI Conditions for Advisa

32

Medtronic Confidential. Internal Use Only. 61

System Requirements for MRI •  A complete SureScan pacing system including a Revo

MRI SureScan pacemaker and two SureScan leads is required for use in the MRI environment.

–  Any other combination may result in a hazard to the patient during an MRI scan.

•  The SureScan feature must be programmed to On prior to

scanning a patient according to the specified conditions of use.

•  The Revo MRI SureScan pacing system has been designed to be MR-Conditional.

Medtronic Confidential. Internal Use Only. 63

Conditions for Use: Radiology Requirements •  Horizontal cylindrical bore magnet, clinical MRI systems with a static magnetic field of 1.5 Tesla (T) must be used. •  Gradient systems with maximum gradient slew rate performance per axis of ≤ 200 Tesla per meter per second (T/m/s) must be used. •  The scanner must be operated in Normal Operating Mode:

- The whole body averaged SAR must be ≤ 2.0 watts per kilogram (W/kg). - The head SAR must be < 3.2 W/kg.

•  The patient must be positioned within the bore such that the isocenter (center of the MRI bore) is superior to the C1 vertebra or inferior to the T12 vertebra. •  The proper patient monitoring must be provided during the MRI scan.

-  The methods include visual and verbal contact with the patient, electrocardiography, and pulse oximetry (plethysmography).

Page 33: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

MR Conditions for Entovis ProMRI®

• Entovis device and Setrox S 53 cm or 60 cm lead for 6 weeks

• There are no other implants in the patient’s body • The lead-tissue interface is stable • The system is operating normally • The specific device programming for the MRI

scan has been performed • The MRI is performed in a standard 1.5 Tesla,

closed bore scanner in the normal operating mode • Mean slew rate is less than or equal to 200 T/m/s

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 34: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

MRI and ICD’s

34

Page 35: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Pro-MRI TrialPhase A

Pacemaker(Scan Exclusion Zone)

Phase B Pacemaker

(No Scan Exclusion Zone)

Phase CICD

(No Scan Exclusion Zone)

First Enrollment March 2013 December 2013 May 2014

Enrollment/Follow-up Status

Closed to Enrollment and Follow-up

Closed to enrollement an follow-up. Submitted to FDA

Closed to enrollment. Follow-up ongoing

Estimated Completion FDA Approval May 4, 2014 FDA Approval March 15, 2015 Q1 2015

Page 36: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Investigational Sites

25 centers12 centers

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 37: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

Phase C Study Design

3. Post-MRI procedure

2. MRI scanning procedure*

1. Pre-MRI procedure

Performed on the same day

Baseline Evaluation(At least)

Observation via Home Monitoring

1-Month follow-up (1 month after the MRI procedure (FU window:

-2 weeks; + 4 weeks)

3-Month Follow-up ( 3 months after the MRI procedure (FU

window: - 2 weeks; + 4 weeks)

*Not earlier than 7 days after enrollment and not later than 2 months after enrollment/baseline evaluation

Device implanted for at least 6 weeks

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 38: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

MRI Procedure--- CAUTION ---

Risks associated with failing to program a subject to MRI mode:

• Device may sense EMI by MRI and trigger an inappropriate shock

• Noise from MRI may cause battery drain, resulting in the EOS voltage of 1.75 V

• Wires inside the device may heat and cause damage to internal components, causing the device to become uninterrogatable

Used by permission ©2014 BIOTRONIK, Inc. All rights reserved. PPT184

Page 39: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis
Page 40: Advances in Pacemakers and ICDs: MRI Compatible Devices · magnetic field of 1.5 Tesla (T) must be used. • Gradient systems with maximum gradient slew rate performance per axis

What does the future hold?

3T MRI with exclusion zones for PM and ICD

1.5T MRI without exclusion zones for PM and ICD

1.5T MRI with exclusion zone for BiV ICD and PM