Advanced Systems, Inc.
Chromatography Integration and Trouble-shooting
Presented by: Marlene Moore [email protected]
October 2007
Advanced Systems, Inc.
Problems Baseline Disturbances Irregular Peak Shapes, Sizes Retention Time Shifts Loss of Separation or Resolution Quantitation Difficulties Rapid Column Deterioration Ghost Peaks Broad Solvent Front
Advanced Systems, Inc.
Spiking Particulate Matter Passing Through
Detector Loose Connections
Cables Circuit boards
Advanced Systems, Inc.
Noise Contaminated Injector/column Physical Defect Detector Defective Detector Board Carrier Gas Leak at Septum or Column FID, NPD, FPD
Column in Flame Incorrect Combustion Gases or Flow Rate/leak
ECD, TCD Air Leak
Advanced Systems, Inc.
Wander Contaminated Carrier Gas
Isothermal Conditions Contaminated Gas Chromatograph Poor Control of Carrier Gas or
Detector Gas Flows Poor Thermal Control of Detector
Advanced Systems, Inc.
Drift Septum Bleed Upward
GC or column contamination Damaged stationary phase
Downward Incomplete conditioning of column Unequilibrated detector
Advanced Systems, Inc.
Offset Contamination:
Injector, column, detector Carrier gas - change/install impurity
traps Column Inserted Into Flame (FID,
NPD, FPD) Malfunctioning or Improperly Set
Recording Device
Advanced Systems, Inc.
QuickTime™ and aTIFF (Uncompressed) decompressor
are needed to see this picture.
Peak Shape
Peak Shape tells the story of your chromatography
Advanced Systems, Inc.
No Peaks Plugged Syringe Broken Column Injecting Sample Into Wrong Injector Column Installed in Wrong Detector Integrator Reading Wrong Signal or Not Connected Detector Not on Temperature
Oven not heated Column temperature too low Injection port temperature too low
Detector Gases Improperly Set or Not on Very Low or No Carrier Gas Flow
Advanced Systems, Inc.
Reduced Peaks Plugged Syringe Change in Injection Technique Large Leak - Injector Split Ratio Too High Too Short of Purge Activation Time for
Splitless Injections Very High Septum Purge Flow Too Low Injector Temperature Column Temperature Too Low Sample Components Absorbed by
Column or Inlet Liner
Advanced Systems, Inc.
Reduced Peaks Initial Temperature Column Too High for
Splitless or On-column Injections High Background Signal -
Contamination Excessive column bleed Check auto zero function and setting
Improperly Operated Detectors Impurities Carrier Gas or Detector Gas Detector Compound Mismatch Excessive Attenuated Integrator Signal Sample Concentration or Integrity Problems
Advanced Systems, Inc.
Tailing Peaks Active Injector Liner or Column Contaminated
Injector liner Column
Dead Volume Caused by Poorly Installed Liner, Column or Union
Poorly Cut Column End Polarity Mismatch - Stationary Phase,
Solute or Solvent Cold Spot in the Flow Path
Advanced Systems, Inc.
Tailing Peaks Solid Debris in Liner or Column Poor Injection Technique - Too Slow Too Low of a Split Ratio Overloading on a PLOT Column
Decrease sample amount Amines and carboxylic acids tail on most
columns Need ph modified stationary phase
Some Peaks Always Exhibit Tailing
Advanced Systems, Inc.
Rounded or Flat Topped Peaks Overloaded Detector Exceeding Range of Integrator
Advanced Systems, Inc.
Split Peaks Detector Overloaded Poor Injection Technique Poorly Installed Column Into Injector Column Temperature Fluctuations Coelution of Two or More Compounds Mixed Sample Solvent for Splitless or
On-column Injections.
Advanced Systems, Inc.
Negative Peaks All Peaks Negative - Polarity Check
Connectors TCD - Select Peaks
Thermal conductivity of compound greater than carrier gas
ECD - After Positive Peak Dirty or old ECD cell
Advanced Systems, Inc.
Retention Time Shifts Different Column Temperature Different Carrier Gas Flow or Linear
Velocity Leak in Injector - Septum Contaminated Column Change in Sample Solvent
Advanced Systems, Inc.
Loss of Separation/Resolution Contaminated Column Damaged Stationary Phase Different Column Temperature,
Carrier Flow Rate, Column Large Changes in Sample
Concentrations Improper Injector Operation
Advanced Systems, Inc.
Quantitation Difficulties Injection Techniques Split Discrimination Different Purge Activation Time for
Splitless Injection Baseline Disturbances Improper Integrator Settings Inconsistent Detector Gas Flows or
Temperatures Column or Liner Activity
Advanced Systems, Inc.
Rapid Column Deterioration Exposure of Column to Air at
Elevated Temperatures Exceeding Upper Temperature
Limit of Column Chemical Damage Contamination of Column With
High Molecular Weight Materials Column Breakage
Advanced Systems, Inc.
Ghost Peaks Contamination
Injector Column
Septum Bleed Dirty Syringe Sample Too Large Previous Run Terminated Too Soon
BOO!
Advanced Systems, Inc.
Broad Solvent Front Poorly Installed Column Leak in the Injector Solvent Interacting with Detector Too Low
Split Ratio (Flow) Injector Temperature
Too Long Purge Activation for Splitless Injections Large Injection Volume Low Column Temperature and High Boiling Solvent High Column Temperature and Low Boiling Solvent
Advanced Systems, Inc.
Tools Flowmeter Syringe (NEW) Methane or Nonretained Compound Septa, Ferrules, Injection Liners (NEW) Leak Detector Column Test Mixture (Reference) Checkout Column Instrument Manuals
Advanced Systems, Inc.
NO!
WRONG ANSWER!!!
SolutionsQuickest Solution ----
Improve Chromatography ---
Do Manual Integration ---
Get Results out Fast !!
Advanced Systems, Inc.
Solutions Preventive Maintenance
Clean detector Clean injector Clean and repair parts Clean and tighten connections Follow instrument manual suggestions
Allow Time for Optimizing System Integration parameters Programs