Advanced Modes of CMV
RC 270
![Page 2: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/2.jpg)
Pressure Support = mode that supports spontaneous
breathing
A preset pressure is applied to the airway with each spontaneous
inspiration
![Page 3: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/3.jpg)
Pressure Support
Pure assist mode Patient determines rate, Vt, and
inspiratory time Inspiration is flow cycled
Most ventilators flow cycle a pressure support breath when inspiratory flow drops to 25% of the peak flow for that inspiration
PB 7200 flow cycles when pressure support flow drops to 5-10 lpm
![Page 4: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/4.jpg)
Indications/Advantages: Pressure Support Initially used to overcome the increased
W.O.B. when breathing spontaneously through an E-T tube
Also may be used during spontaneous breaths during IMV
Weaning Assisted ventilation (instead of A/C)
PSVmax
![Page 5: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/5.jpg)
Initial Settings and Adjustments: Pressure Support
To overcome resistance of E-T tube, start at 5-10 cmH2O
For PSVMax, set pressure to level that gives an exhaled Vt of 10-12 ml/kg
![Page 6: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/6.jpg)
Advantages: Pressure Support
Supports spontaneous breathing with decreased W.O.B.(with or without an E-T tube)
Can be done with a face mask Usually less barotrauma and
hemodynamic compromise Patients like it!
![Page 7: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/7.jpg)
Disadvantage: Pressure Support
A leak in the system prevents flow cycling Will cause a CPAP effect
![Page 8: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/8.jpg)
Pressure Controlled Ventilation (PCV)
A set pressure is applied to the airway during inspiration and the
breath time cycles
![Page 9: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/9.jpg)
Pressure Controlled Ventilation Can be used in A/C or control Flow tapers – if it drops to zero before
time cycling occurs, the pressure plateaus
Besides pressure, RCP also sets rate and either inspiratory time or I:E ratio
Vt may vary from breath to breath
![Page 10: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/10.jpg)
Pressure Controlled Ventilation Indications are same as for any type of CMV:
Apnea Acute ventilatory failure Impending ventilatory failure Acute respiratory failure (Oxygenation failure)
Often used when volume cycling (volume control) is causing high airway pressures
Has been used to ventilate neonates since the 60s
![Page 11: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/11.jpg)
PCV: Initial Settings and Adjustments
Initially choose a pressure (PIP) that gives desired exhaled Vt
If switching from volume cycling (volume control), use a PIP that is less than PIP during volume cycling
Adjustment in rate, PIP, and I:E (or inspiratory time based on ABGs, oximetry, and capnography
A change in PIP or I:E/insp time will change Vt
![Page 12: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/12.jpg)
PC-IRV: Pressure Control with Inverse I:E Ratio
Control mode only
Patient is paralyzed
Settings like PCV except for inverse I:E (gives long insp time)
![Page 13: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/13.jpg)
PC-IRV used in diseases with high elastic resistance, eg
ARDS
Prolonged insp time helps O2
To increase PaO2: increase rate, PIP or insp time
To decrease PaCO2: decrease rate or PIP
![Page 14: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/14.jpg)
Airway Pressure Release Ventilation (APRV)
Alternating levels of CPAP in a spontaneously breathing patient
![Page 15: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/15.jpg)
APRV
Like PC-IRV but patient is breathing spontaneously and is not paralyzed
Also used for high elastic resistance High CPAP level is applied longer than
low CPAP level Is NOT synchronized with inspiration and
expiration
![Page 16: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/16.jpg)
APRV: Settings and Adjustments
Low CPAP usually between 2-10 cmH2O
High CPAP usually between 10-30 cmH2O
RCP also sets the time for each CPAP level Low CPAP is usually only for 1-2 seconds
![Page 17: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/17.jpg)
Bilevel Positive Airway Pressure (BIPAP)
IPAP + EPAP
Differs from APRV – IPAP only during inspiration, EPAP only
during expiration
Rate and I:E ratio can also be set
![Page 18: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/18.jpg)
Indications : BIPAP
Sleep apnea Ventilatory Assist without intubation
Can be done via face mask Often used to keep COPDers from being
tubed and put on A/C
Popular mode for NIPPV (Non-invasive Positive Pressure Ventilation)
![Page 19: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/19.jpg)
High Frequency Ventilation (HFV)
A form of ventilation utilizing high rates and small Vt that seems to enhance diffusion of gases into
and out of the lung
![Page 20: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/20.jpg)
History of CMV
HFV should not work based on classical respiratory physiology!
![Page 21: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/21.jpg)
HFV: High Frequency Jet Ventilation (HFJV) Vt usually 20-150 ml Frequency (rate) 60-400 breaths per
minute Usually a catheter is inserted via ET
tube or transnasally to apply jet bursts to airway
Adjust rate, driving pressure, and insp time, and FIO2
![Page 22: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/22.jpg)
HFV: High Frequency Oscillation (HFO) Vt between 5-50ml Frequency between 400-3000
Frequency expressed in Hertz (Hz) 10 Hz equals 600 breaths per minute
![Page 23: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/23.jpg)
HFO Techniques
![Page 24: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/24.jpg)
HFV (both HFJV and HFO)
Strict FIO2 and humidification can be variable
Both appear to cause diffusion to occur from proximal airway to alveoli How does
spontaneous breathing work?
Coaxial flow Inspiration and
expiration may be occurring simultaneously
HFV seems to stimulate mucociliary clearance
![Page 25: Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.](https://reader036.fdocuments.us/reader036/viewer/2022062721/56649f225503460f94c3ac07/html5/thumbnails/25.jpg)
Enough already!