The World Leader in High Performance Signal Processing Solutions ADIS16228 Evaluation Tutorials Manual FFT, Multiple Sample Rates/Records, Alarms January 23, 2014

Transcript of ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE:...

Page 1: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

The World Leader in High Performance Signal Processing Solutions

ADIS16228 Evaluation Tutorials

Manual FFT, Multiple Sample Rates/Records, Alarms

January 23, 2014

Page 2: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

INTRODUCTION

OBJECTIVE:Demonstrate the Spectral Alarm function and Multi-Sample

Rate/Record functions in the ADIS16228, using the Vibration Evaluation Program.

INITIAL CONDITIONSFactory-default register settings, unless specified otherwise

2

Page 3: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

EVALUATION TOOLS/PHYSICAL SETUP

SOFTWARE:Vibration Evaluation Program, v1.1.9 and v1.2.0

HARDWARE:

3

EVAL-ADIS

ADIS16228/PCBZ

4” ribbon cable1mm, 16-pin

Vibration source

Page 4: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

SUMMARY OF STEPS

Launch softwareSet Device = ‘ADIS16228’Select ‘Manual FFT Mode’ (Main Window)Configure registers (Register Access Window)Copy register settings to the ADIS16228’s flash memory

(Register Access Window)Configure & Enable AlarmsTrigger and display four different FFT results (Main Window)Change vertical scale settings to optimize use of the screen

area (Main Window)

4

Page 5: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Launch software & Set Device = ADIS16228

5

Page 6: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Select mode & Launch Register Access Window

6

NOTES:

• Setting Mode Selection to Manual FFT is the equivalent of setting REC_CTRL1[1:0] = 00

Page 7: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Register Settings

For a tutorial on configuring register values, click on “Back” and then click on the “ADIS16228 Register Configuration Tutorial.”

Using these steps, establish the following register settings:REC_CTRL1 = 0x1F01Hanning Window, Enable SR0/1/2/3, Automatic FFT Mode

REC_CTRL2 = 0x00FFSet range on SR0/1/2/3 to 1g

REC_PRD = 0x00044 seconds between each trigger

AVG_CNT = 0x6420 only the last four bits matters for this exercise

ALM_CTRL = 0x0087The “7” enables alarms for all three axes

7

Page 8: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Click on “Alarms>Alarm Settings”

8

NOTES:

Page 9: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Click on “Read Default Values”

9

NOTES:

• The Vibration Evaluation Program has a set of “default” alarm values, for the purpose of demonstration.

• These values are used to simplify this demonstration.

Page 10: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Click on “Write to DUT”

10

NOTES:

• After seeing the values update for all of the sample rates and bands, write these values to the ADIS16228CMLZ.

• A yellow box will appear for a few moments, indicating that the communication of this data is ongoing. It will disappear when the communication process is complete.

• Exit this Window

Page 11: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Click on “Alarms>Alarm Status Form”

11

NOTES:

• Drag the Alarm Status Window to a convenientplace for viewing, which does not impede clicking on the Start button.

• Click on Start to acquire a spectral record (FFT result).

Page 12: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Click on “Start”

12

NOTES:

• Notice that all of the status flags are green.

• This indicates that all alarms are passing.

Page 13: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Right click in waveforms to adjust scale

13

NOTES:

Page 14: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Right click in waveforms to adjust scale

14

NOTES:

Page 15: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Click on Start

15

NOTES:

• Notice that the frequency range has changed, since the ADIS is now cycling through each SRx setting

• This graphic shows SR1, which supports a sample rate of 5120 SPS and a Nyquist range of 2560 Hz, as the red arrow highlights.

Page 16: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Click on Start

16

NOTES:

• Notice that the frequency range has changed, since the ADIS is now cycling through each SRx setting

• This graphic shows SR2, which supports a sample rate of 1280 SPS and a Nyquist range of 640 Hz, as the red arrow highlights.

Page 17: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Click on Start

17

NOTES:

• Notice that the frequency range has changed, since the ADIS is now cycling through each SRx setting

• This graphic shows SR3, which supports a sample rate of 320 SPS and a Nyquist range of 160 Hz, as the red arrow highlights.

Page 18: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Click on Start

18

NOTES:

• Notice that the frequency range has changed, since the ADIS is now cycling through each SRx setting

• This graphic shows SR0, which supports a sample rate of 20480 SPS and a Nyquist range of 10240 Hz, as the red arrow highlights.

Page 19: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Add Vibration, then cycle through each sample rate setting by clicking Start

19

NOTES:

• SR0 Setting• Sample rate = 20480 SPS• Only one warning alarm

Page 20: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Add Vibration, then cycle through each sample rate setting by clicking Start

20

NOTES:

• SR1 Setting• Sample rate = 5120 SPS• Vibration energy overlaps

with the first alarm band

Page 21: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Add Vibration, then cycle through each sample rate setting by clicking Start

21

NOTES:

• SR2 Setting• Sample rate = 1280 SPS • Notice how the primary

vibration has moved into the first alarm band

Page 22: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Add Vibration, then cycle through each sample rate setting by clicking Start

22

NOTES:

• SR3 Setting• Sample rate = 320 SPS• The vibration energy is

now in the second alarm band

Page 23: ADIS16228 Evaluation Tutorials - Analog Devices · EVALUATION TOOLS/PHYSICAL SETUP SOFTWARE: Vibration Evaluation Program, v1.1.9 and v1.2.0 HARDWARE: 3 EVAL-ADIS ADIS16228/PCBZ 4”

Thank you!