Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except...

57
It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD OFFICE: 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) Accredited Testing Laboratory under the terms of ISO/IEC 17025 CE CONFORMANCE REPORT EMC DIRECTIVE 2004/108/EC for PRODUCT: HIGH DEFINITION NETWORK CAMERA “MODEL NAME: LND5110” AUTHORIZED LG Electronics Inc. 222 LG-ro, Jinwi-Myeon, Pyeongtaek -Si, Gyeonggi-Do, 451-713, Korea April 03, 2014 SUMMARY The equipment complies with the standards; • EN 55022: 2010 (Class A) • EN 61000-3-2: 2006 + A1: 2009 + A2: 2009 • EN 61000-3-3: 2008 • EN 50130-4: 2011 (signature) Gea-Won, Lee / Managing Director This is only valid in connection with Test Report: E144R-009

Transcript of Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except...

Page 1: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD OFFICE: 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)

Accredited Testing Laboratory under the terms of ISO/IEC 17025

CE CONFORMANCE REPORT EMC DIRECTIVE 2004/108/EC

for

PRODUCT: HIGH DEFINITION NETWORK CAMERA

“MODEL NAME: LND5110”

AUTHORIZED

LG Electronics Inc. 222 LG-ro, Jinwi-Myeon, Pyeongtaek -Si, Gyeonggi-Do, 451-713, Korea

April 03, 2014

SUMMARY The equipment complies with the standards;

• EN 55022: 2010 (Class A)

• EN 61000-3-2: 2006 + A1: 2009 + A2: 2009

• EN 61000-3-3: 2008

• EN 50130-4: 2011

(signature)

Gea-Won, Lee / Managing Director

This is only valid in connection with Test Report: E144R-009

Page 2: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 1 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

ELECTROMAGNETIC

COMPATIBILITY TEST REPORT

Test Report No. : E144R-009

AGR No. : A143A-070R

Applicant : LG Electronics Inc.

Address : 222 LG-ro, Jinwi-Myeon, Pyeongtaek -Si, Gyeonggi-Do, 451-713, Korea

Manufacturer : LG Electronics Inc.

Address : 222 LG-ro, Jinwi-Myeon, Pyeongtaek -Si, Gyeonggi-Do, 451-713, Korea

Type of Equipment : High Definition Network Camera

Model Name : LND5110

Multiple Model Name : N/A

Serial number : N/A

Total page of Report : 52 pages (including this page)

Date of Incoming : March 10, 2014

Date of Issuing : April 03, 2014

SUMMARY The equipment complies with the standard; EN 55022: 2010 (Class A), EN 61000-3-2: 2006 + A1: 2009 + A2:

2009, EN 61000-3-3: 2008 and EN 50130-4: 2011.

This test report contains only the results of a single test of the sample supplied for the examination.

It is not a general valid assessment of the features of the respective products of the mass-production.

Prepared by: Approved by: Dong-Youl, Lim / Asst. Chief Engineer Gea-Won, Lee / Managing Director ONETECH Corp. ONETECH Corp.

Page 3: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 2 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

CONTENTS

PAGE

1. APPLICANT AND MANUFACTURER INFORMATION ........................................................................................... 7

2. TEST SUMMARY .............................................................................................................................................................. 7

2.1 TEST STANDARDS AND RESULTS ..................................................................................................................................... 7

2.2 ADDITIONS, DEVIATIONS, EXCLUSIONS FROM STANDARDS ........................................................................................... 7

2.3 PURPOSE OF THE TEST ................................................................................................................................................... 7

2.4 TEST FACILITY ............................................................................................................................................................... 7

2.5 ENVIRONMENTAL CONDITIONS ...................................................................................................................................... 8

2.6 INFORMATION OF THE INSTRUCTION FOR CLASS A EQUIPMENT .................................................................................. 8

3. EUT (EQUIPMENT UNDER TEST) ............................................................................................................................... 9

3.1 IDENTIFICATION OF THE EUT ....................................................................................................................................... 9

3.2 ADDITIONAL INFORMATION ABOUT THE EUT .............................................................................................................. 9

3.3 PERIPHERAL EQUIPMENT ............................................................................................................................................... 9

3.4 MODE OF OPERATION DURING THE TEST....................................................................................................................... 9

3.5 CRITERION DESCRIPTION ............................................................................................................................................. 10

3.6 CABLE DESCRIPTION ................................................................................................................................................... 10

3.7 ALTERNATIVE TYPE(S)/MODEL(S) ............................................................................................................................... 10

4. EUT MODIFICATIONS ................................................................................................................................................. 10

5. EMISSION TESTS ........................................................................................................................................................... 11

5.1 MAINS TERMINAL CONTINUOUS DISTURBANCE VOLTAGE ................................................................ 11

5.1.1 Operating environment ......................................................................................................................................... 11

5.1.2 Test set-up ............................................................................................................................................................. 11

5.1.3 Measurement uncertainty ..................................................................................................................................... 11

5.1.4 Test equipment used ............................................................................................................................................. 11

5.2 CONDUCTED COMMON MODE DISTURBANCE AT TELECOMMUNICATION PORTS ................... 14

5.2.1 Operating environment ......................................................................................................................................... 14

5.2.2 Test set-up ............................................................................................................................................................. 14

5.2.3 Measurement uncertainty ..................................................................................................................................... 14

5.2.4 Test equipment used ............................................................................................................................................. 14

5.2.5 Test data ................................................................................................................................................................ 15

5.3 RADIATED ELECTROMAGNETIC FIELD ....................................................................................................... 19

Page 4: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 3 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

5.3.1 Operating environment ......................................................................................................................................... 19

5.3.2 Test set-up ............................................................................................................................................................. 19

5.3.3 Measurement uncertainty ..................................................................................................................................... 19

5.3.4 Test equipment used ............................................................................................................................................. 19

5.3.5 Test data ................................................................................................................................................................ 20

5.4 HARMONICS CURRENT EMISSIONS .................................................................................................................. 23

5.4.1 Operating environment ......................................................................................................................................... 23

5.4.2 Test set-up ............................................................................................................................................................. 23

5.4.3 Measurement uncertainty ..................................................................................................................................... 23

5.4.4 Test equipment used ............................................................................................................................................. 23

5.4.5. Test data ............................................................................................................................................................... 23

5.5 VOLTAGE CHANGES, VOLTAGE FLUCTUATIONS AND FLICKER ........................................................... 25

5.5.1 Operating environment ......................................................................................................................................... 25

5.5.2 Test set-up ............................................................................................................................................................. 25

5.5.3 Measurement uncertainty ..................................................................................................................................... 25

5.5.4 Test equipment used ............................................................................................................................................. 25

5.5.5 Test data ................................................................................................................................................................ 25

6. IMMUNITY TESTS ......................................................................................................................................................... 28

6.1 ELECTROSTATIC DISCHARGE IMMUNITY TEST ......................................................................................... 28

6.1.1 Operating environment ......................................................................................................................................... 28

6.1.2 Test set-up ............................................................................................................................................................. 28

6.1.3 Measurement uncertainty ..................................................................................................................................... 28

6.1.4 Test equipment used ............................................................................................................................................. 28

6.1.5 Test data ................................................................................................................................................................ 29

6.1.6 ESD Test Point & Test Result .............................................................................................................................. 30

6.2 RADIATED RF-ELECTROMAGNETIC FIELD IMMUNITY TEST ................................................................. 31

6.2.1 Operating environment ......................................................................................................................................... 31

6.2.2 Test set-up ............................................................................................................................................................. 31

6.2.3 Measurement uncertainty ..................................................................................................................................... 31

6.2.4 Test equipment used ............................................................................................................................................. 31

6.2.5 Test data ................................................................................................................................................................ 32

6.3 ELECTRICAL FAST TRANSIENT/BURST IMMUNITY TEST ......................................................................... 33

6.3.1 Operating environment ......................................................................................................................................... 33

6.3.2 Test set-up ............................................................................................................................................................. 33

6.3.3 Measurement uncertainty ..................................................................................................................................... 33

6.3.4 Test equipment used ............................................................................................................................................. 33

Page 5: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 4 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.3.5 Test data ................................................................................................................................................................ 34

6.4 SURGE IMMUNITY TEST ....................................................................................................................................... 35

6.4.1 Operating environment ......................................................................................................................................... 35

6.4.2 Test set-up ............................................................................................................................................................. 35

6.4.3 Measurement uncertainty ..................................................................................................................................... 35

6.4.4 Test equipment used ............................................................................................................................................. 35

6.4.5 Test data ................................................................................................................................................................ 36

6.5 CONDUCTED RF FIELDS IMMUNITY TEST ...................................................................................................... 37

6.5.1 Operating environment ......................................................................................................................................... 37

6.5.2 Test set-up ............................................................................................................................................................. 37

6.5.3 Measurement uncertainty ..................................................................................................................................... 37

6.5.4 Test equipment used ............................................................................................................................................. 37

6.5.5 Test data ................................................................................................................................................................ 38

6.6 MAINS SUPPLY VOLTAGE DIPS AND SHORT INTERRUPTIONS IMMUNITY TEST .............................. 39

6.6.1 Operating environment ......................................................................................................................................... 39

6.6.2 Test set-up ............................................................................................................................................................. 39

6.6.3 Measurement uncertainty ..................................................................................................................................... 39

6.6.4 Test equipment used ............................................................................................................................................. 39

6.6.5 Test data ................................................................................................................................................................ 40

6.7 MAINS SUPPLY VOLTAGE VARIATIONS IMMUNITY TEST ........................................................................ 41

6.7.1 Operating environment ......................................................................................................................................... 41

6.7.2 Test set-up ............................................................................................................................................................. 41

6.7.3 Measurement uncertainty ..................................................................................................................................... 41

6.7.4 Test equipment used ............................................................................................................................................. 41

6.7.5 Test data ................................................................................................................................................................ 42

APPENDIX I - TEST SET-UP PHOTOS: (MAINS TERMINAL CONTINUOUS DISTURBANCE VOLTAGE) ... 43

APPENDIX II - TEST SET-UP PHOTOS: (CONDUCTED COMMON MODE DISTURBANCE AT

TELECOMMUNICATION PORTS) ................................................................................................................................. 44

APPENDIX III - TEST SET-UP PHOTOS: (RADIATED ELECTROMAGNETIC FIELD) ..................................... 45

APPENDIX IV - TEST SET-UP PHOTO: (HARMONICS & VOLTAGE FLUCTUATIONS ON AC MAINS) ...... 47

APPENDIX V – TEST SET-UP PHOTO: (ESD) .............................................................................................................. 48

APPENDIX VI – TEST SET-UP PHOTO: (RF E- FIELD) ............................................................................................. 49

Page 6: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 5 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX VII - TEST SET-UP PHOTO: (EFT/BURST) ............................................................................................. 50

APPENDIX VIII - TEST SET-UP PHOTO: (SURGE) .................................................................................................... 51

APPENDIX IX - TEST SET-UP PHOTO: (C.S) ............................................................................................................... 52

APPENDIX XI - TEST SET-UP PHOTO: (MAINS SUPPLY VOLTAGE DIPS, SHORT INTERRUPTIONS) ...... 53

APPENDIX XI - TEST SET-UP PHOTO: (POWER SUPPLY SHORT TERM VARIATION) ................................. 54

APPENDIX XII - PHOTOGRAPHS OF THE PRODUCT ............................................................................................. 55

Page 7: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 6 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

Revision History

Issued Report No. Issued Date Revisions Effect Section

E144R-009 April 03, 2014 Initial Issue All

Page 8: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 7 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

1. APPLICANT AND MANUFACTURER INFORMATION

-. Applicant : LG Electronics Inc.

-. Address : 222 LG-ro, Jinwi-Myeon, Pyeongtaek -Si, Gyeonggi-Do, 451-713, Korea

-. Manufacturer : LG Electronics Inc.

-. Address : 222 LG-ro, Jinwi-Myeon, Pyeongtaek -Si, Gyeonggi-Do, 451-713, Korea

2. TEST SUMMARY

2.1 Test standards and results

STANDARDS RESULTS

EN 55022: 2010

Radiated Electromagnetic Field Met Class A / PASS

Mains Terminal disturbance voltage Met Class A / PASS

Conducted Common Mode Disturbance at Telecommunication Ports Met Class A / PASS

EN 61000-3-2: 2006

+A1: 2009 + A2: 2009 Harmonics Current Emissions Met / PASS

EN 61000-3-3: 2008 Voltage changes, Voltage fluctuations and Flicker Met / PASS

EN 50130-4: 2011

Electrostatic Discharge Immunity PASS

Radiated RF E-Field (80 to 2 700 MHz) PASS

Electrical Fast Transient/Burst Immunity PASS

Surge Immunity PASS

Conducted Disturbance induced by RF fields Immunity PASS

Mains Supply Voltage Dips, Short Interruptions PASS

Mains Supply Voltage Variations PASS Note1. The margin to the limit is within the uncertainty interval of the measured value.

2.2 Additions, deviations, exclusions from standards

No additions, deviations or exclusions have been made from standard.

2.3 Purpose of the test

To determine whether the equipment under test fulfills the EMC requirements of the standards stated in section 2.1.

2.4 Test Facility

The Electromagnetic compatibility measurement facilities are located on at 301-14, Daessangnyeong-ri, Chowol-eup,

Gwangju-si, Gyeonggi-do, 464-862 Korea. Description details of test facilities were submitted to the FCC (KR0013) and

IC (Registration No: IC3736A-2), accredited by KOLAS (Korea Laboratory Accreditation Scheme, No: 85) and approved

by TUV, DNV and KCC (Korea Communications Commission) according to the requirement of ISO 17025.

Page 9: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 8 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

2.5 Environmental conditions

Unless otherwise indicated in the basic standard or test procedure, the tests shall be carried out within the rated supply

voltage for the EUT and the following standard atmospheric conditions for measurements and tests, as specified in IEC

60068-1, Subclause 5.3.1.

-. Temperature: 15 °C to 35 °C.

-. Relative humidity: 25 % R.H. to 75 % R.H.

-. Air pressure: 86.0 kPa to 106.0 kPa

2.6 Information of the instruction for class A equipment

Class A equipment shall be included the following warning in the instructions for use:

WARNING: This is a class A product. In a domestic environment this product may cause radio interference in which

case the user may be required to take adequate measures.

Page 10: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 9 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

3. EUT (Equipment Under Test)

3.1 Identification of the EUT

-. Equipment : High Definition Network Camera

-. Model Name : LND5110

-. Brand Name : LG Electronics

-. Serial number : N/A

-. Manufacturer : LG Electronics Inc.

3.2 Additional information about the EUT The LG Electronics Inc., Model LND5110 (referred to as the EUT in this report) is an High Definition Network Camera.

Product specification information described herein was obtained from product data sheet or user’s manual.

CHASSIS TYPE Plastic & Metal

LIST OF EACH OSC. OR

CRY. FREQ.(FREQ.>=1 MHz) 400 MHz

ELECTRICAL RATING DC 12 V, 490 mA (Adapter : AC 100 - 240 V ~, 50/60 Hz)

NUMBER OF LAYERS -

TEMPERATURE RANGE -40 °C ~ 85 °C

EXTERNAL CONNECTOR Main Input Power, LAN, Audio In/Out, Alarm

3.3 Peripheral equipment Defined as equipment needed for correct operation of the EUT, but not considered as tested:

Model Manufacturer Description Connected to

LND5110 LG Electronics Inc. High Definition Network Camera (EUT) -

XPS 14z DELL Notebook PC EUT

DA90PM11 DELL Adapter Notebook PC

DA-18C-12 LG AC/DC Adapter EUT

3.4 Mode of operation during the test -. The LAN port on the EUT was connected to the notebook PC and then the EUT was observed the state of video output

on the monitor during the testing.

-. The power of the EUT was fed from a power adaptor (230 V~, 50 Hz).

Page 11: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 10 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

3.5 Criterion description There shall be no damage, malfunction or change of status due to the conditioning. Flickering of an indicator during the

conditioning is permissible, providing that there is no residual change in the EUT or any change in outputs, which could be

interpreted by associated equipment as a change,

The EUT shall meet the acceptance criteria for the functional test, after the conditioning.

For Radiated electromagnetic fields

and no such flickering of indicators occurs at a field strength of 3 V/m.

For components of CCTV systems, where the status is monitored by observing the TV picture, then deterioration of the

picture is allowed at 10 V/m, providing:

a) there is no permanent damage or change to the EUT (e.g. no corruption of memory or changes to programmable settings

etc.);

b) at 3 V/m, any deterioration of the picture is so minor that the system could still be used; and

c) there is no observable deterioration of the picture at 1 V/m.

For Conducted disturbances induced by electromagnetic fields

and no such flickering of indicators occurs at U0 = 3 V.

For components of CCTV systems, where the status is monitored by observing the TV picture, then deterioration of the

picture is allowed at U0 = 10 V, providing:

a) there is no permanent damage or change to the EUT (e.g. no corruption of memory or changes to programmable settings

etc.);

b) at U0 = 3 V, any deterioration of the picture is so minor that the system could still be used; and

c) there is no observable deterioration of the picture at U0 = 1 V.

-. Mains supply voltage variations

There shall be no damage, malfunction or change of status due to the different supply voltage conditions.

3.6 Cable Description Ports Name Shielded Ferrite Bead Metal Shell Length (m) Connected to

DC IN Y N N 1.0 AC/DC Adaptor

LAN N N N 3.0 Notebook PC

3.7 Alternative type(s)/model(s)

-. None

4. EUT MODIFICATIONS

-. None

Page 12: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 11 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

5. EMISSION TESTS

5.1 MAINS TERMINAL CONTINUOUS DISTURBANCE VOLTAGE

5.1.1 Operating environment

Temperature : 21 °C

Relative humidity : 35 % R.H.

5.1.2 Test set-up

The EUT was placed on a non-conductive table, 0.8 m height above the floor. The EUT was connected to adapter and the

power of adapter was fed through a 50 Ω/ 50 μH + 5 Ω Artificial Mains Network (AMN). The ground plane was

electrically bonded to reference ground system and all power lines were filtered from ambient.

The test set-up photos are included in appendix I.

5.1.3 Measurement uncertainty

Mains terminal continuous disturbance voltage, quasi-peak detection : ± 2.93 dB

Mains terminal continuous disturbance voltage, average detection : ± 2.93 dB

Measurement uncertainty is calculated in accordance with CISPR 16-4-2. The measurement uncertainty is given with a

confidence of 95 % with the coverage factor, k = 2.

5.1.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal. (Interval)

- ESHS10 Rohde & Schwarz Test Receiver 834467/007 Jul. 02, 2013 (1Y)

- NSLK 8126 Schwarzbeck AMN 8126-404 May 29, 2013 (1Y)

- NSLK 8128 Schwarzbeck AMN 8128-216 Jun. 07, 2013 (1Y)

- 3825/2 EMCO AMN 9109-1867 May 20, 2013 (1Y)

- 3825/2 EMCO AMN 9109-1869 May 20, 2013 (1Y)

All test equipment used is calibrated on a regular basis.

Page 13: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 12 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

5.1.5 Test data

-. Test Date : March 19, 2014

-. Resolution bandwidth : 9 kHz

-. Frequency range : 0.15 MHz ~ 30 MHz

Frequency Line Peak (dBμV) Margin (MHz) Emission level Q.P Limits (dB)

0.15 N 51.39 79.00 27.61

0.18 H 46.85 79.00 32.15

0.56 N 40.23 73.00 32.77

0.57 H 38.13 73.00 34.87

13.39 N 41.73 73.00 31.27

13.58 H 42.03 73.00 30.97

Frequency Line Average (dBμV) Margin (MHz) Emission level Limits (dB)

- - - - -

- - - - -

Remark : “H”: Hot Line, “N”: Neutral Line

See next page for an overview sweep performed with peak detector.

Tested by: Jong-Jin. Lee / Engineer

Page 14: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 13 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

HOT LINE

NEUTRAL LINE

Page 15: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 14 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

5.2 CONDUCTED COMMON MODE DISTURBANCE AT TELECOMMUNICATION PORTS

5.2.1 Operating environment

Temperature : 21 °C

Relative humidity : 35 % R.H.

5.2.2 Test set-up

The EUT and other support equipment were placed on a non-conductive table, 0.8 m height above the floor.

Telecommunication line for the EUT connected to associated equipment through an Impedance Stabilization Network

(ISN) which has a common mode termination impedance of 150 Ω to the telecommunication port under test. The ground

plane was electrically bonded to the reference ground system and all power lines were filtered from ambient.

The test set-up photos are included in appendix II.

5.2.3 Measurement uncertainty

Conducted emission, quasi-peak detection : ± 2.93 dB

Conducted emission, average detection : ± 2.93 dB

Measurement uncertainty is calculated in accordance with CISPR 16-4-2.. The measurement uncertainty is given with a

confidence of 95 % with the coverage factor, k = 2.

5.2.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal.(Interval)

- ESHS10 Rohde & Schwarz EMI Test Receiver 834467/007 Jul. 02, 2013 (1Y)

- NSLK 8126 Schwarzbeck AMN 8126-404 May 29, 2013 (1Y)

- 3825/2 EMCO AMN 9109-1867 May 20, 2013 (1Y)

- ENY41 Rohde & Schwarz ISN 100021 May 30, 2013 (1Y)

All test equipment used is calibrated on a regular basis.

Page 16: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 15 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

5.2.5 Test data

-. Test Date : March 19, 2014

-. Resolution bandwidth : 9 kHz

-. Frequency range : 0.15 MHz ~ 30 MHz

-.Remark : ISN 10 Mbps

Frequency Port Peak (dBμV) Margin Average (dBμV) Margin

(MHz) Emission

Level Detector

Mode Q.P Limits (dB) Emission Level Limits (dB)

0.15 Four 55.32 P 97.00 41.68 - - -

0.59 Four 50.51 P 87.00 36.49 - - -

5.06 Four 58.68 P 87.00 28.32 - - -

8.55 Four 62.05 P 87.00 24.95 - - -

10.0 Four 69.41 P 87.00 17.59 - - -

19.43 Four 51.65 P 87.00 35.35 - - -

Remark : “Four”: Two Unscreened Balance Pair, “P”: Peak detect

See next page for an overview sweep performed with peak detector.

Tested by: Jong-Jin. Lee / Engineer

Page 17: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 16 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

Page 18: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 17 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

-.Remark : ISN 100 Mbps

Frequency Port Peak (dBμV) Margin Average (dBμV) Margin

(MHz) Emission

Level Detector

Mode Q.P Limits (dB) Emission Level Limits (dB)

0.57 Four 50.44 P 87.00 36.56 - - -

1.42 Four 47.86 P 87.00 39.14 - - -

2.19 Four 47.74 P 87.00 39.26 - - -

14.15 Four 53.66 P 87.00 33.34 - - -

18.25 Four 54.77 P 87.00 32.23 - - -

23.13 Four 55.42 P 87.00 31.58 - - -

Remark : “Four”: Two Unscreened Balance Pair, “P”: Peak detect

See next page for an overview sweep performed with peak detector.

Tested by: Jong-Jin. Lee / Engineer

Page 19: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 18 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

Page 20: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 19 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

5.3 RADIATED ELECTROMAGNETIC FIELD

5.3.1 Operating environment

Temperature : 23 °C

Relative humidity : 38 % R.H.

5.3.2 Test set-up

The radiated emissions measurements were on the 10 m, semi anechoic chamber. The EUT and all local support equipment

were placed on a non-conductive turntable approximately 0.8 m above the ground plane.

The frequency spectrum from 30 MHz to 2 000 MHz was scanned and emission levels maximized at each frequency

recorded. The system was rotated 360°, and the antenna was varied in height between 1.0 m and 4.0 m in order to

determine the maximum emission levels. This procedure was performed for both horizontal and vertical polarization of the

receiving antenna.

The test set-up photos are included in appendix III.

5.3.3 Measurement uncertainty

Radiated emission electric field intensity, 30 MHz ~ 300 MHz : ± 4.43 dB

Radiated emission electric field intensity, 300 MHz ~ 1 000 MHz : ± 3.80 Db Radiated emission electric field intensity, 1 GHz ~ 6 GHz : ± 4.20 dB

Measurement uncertainty is calculated in accordance with CISPR 16-4-2. The measurement uncertainty is given with a

confidence of 95 % with the coverage factor, k = 2.

5.3.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal. (Interval)

- ESCI Rohde & Schwarz EMI Test Receiver 101013 Oct. 14, 2013 (1Y)

- ESU Rohde & Schwarz EMI Test Receiver 100261 May 27, 2013(1Y)

- 310N Sonoma Instrument Amplifier 312544 May 21, 2013 (1Y)

- 310N Sonoma Instrument Amplifier 312545 May 21, 2013 (1Y)

- VULB9163 Schwarzbeck TRILOG Broadband Antenna 9163-420 Mar. 27, 2012(2Y)

- VULB9163 Schwarzbeck TRILOG Broadband Antenna 9163-255 Apr. 24, 2012 (2Y)

- BBHA9120D Schwarzbeck Horn Antenna BBHA9120D295 Sep. 05, 2013 (2Y)

- SCU 18 Rohde & Schwarz Signal Conditioning Unit 10041 Nov. 07, 2013 (1Y)

- CO2000 Innco System Controller 619/27030611/L N/A

- DT3000 Innco System Turn Table 930611 N/A

- MA4000-EP Innco System Antenna Master 3320611 N/A

- MA4000-EP Innco System Antenna Master 3350611 N/A

All test equipment used is calibrated on a regular basis.

Page 21: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 20 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

5.3.5 Test data

-. Test Date : March 19, 2014

-. Resolution bandwidth : 120 kHz

-. Frequency range : 30 MHz ~ 1 000 MHz

-. Measurement distance : 10 m

Graphical representation of Radiated Disturbance

Tabulated Results for Radiated Disturbance

Remark: Margin (dB) = Limit – Result and Result = Reading Peak + Antenna Factor + Loss – Gain Loss and Gain in above table means Cable Loss and Pre-amplifier gain.

Page 22: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 21 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

-. Resolution bandwidth : 1 MHz

-. Frequency range : 1 000 MHz ~ 2 000 MHz

-. Measurement distance : 3 m

Detector Mode: Peak

Graphical representation of Radiated Disturbance

Tabulated Results for Radiated Disturbance

Remark: Margin (dB) = Limit – Result and Result = Reading Peak + Antenna Factor + Loss – Gain Loss and Gain in above table means Cable Loss and Pre-amplifier gain.

Page 23: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 22 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

Detector Mode: Average

Graphical representation of Radiated Disturbance

Tabulated Results for Radiated Disturbance

Remark: Margin (dB) = Limit – Result and Result = Reading Average + Antenna Factor + Loss – Gain Loss and Gain in above table means Cable Loss and Pre-amplifier gain.

Tested by: Jong-Jin. Lee / Engineer

Page 24: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 23 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

5.4 HARMONICS CURRENT EMISSIONS

5.4.1 Operating environment

Temperature : 24 °C

Relative humidity : 34 % R.H.

5.4.2 Test set-up

Harmonics of the fundamental current were measured up to 2 kHz using a universal power analyzer. The measurements

were carried out under steady conditions and using average.

Before making measurements, it is necessary for the EUT to decide which class the EUT fulfills into; A, B, C or D.

The test set-up photo is included in appendix IV.

5.4.3 Measurement uncertainty

The uncertainty of our equipments for harmonic measurement is ± 0.2 %.

Measurement uncertainty is calculated in accordance with EN 61000-4-7. The measurement uncertainty is given with a

confidence of 95 % with the coverage factor, k = 2.

5.4.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal.(Interval)

- PM6000 Voltech Universal Power Analyzer 1000067 00196 May 28, 2013 (1Y)

- 134350 Pacific Power Magnetics Module 1512 Jun. 24, 2013 (1Y)

- UPC32-413 Pacific Power Waveform Generator 0002 Apr. 15, 2013 (1Y)

- 360AMXT/UPC-32 Pacific Power AC Power Source 0222 Jun. 24, 2013 (1Y)

- 360AMXT/UPC-32 Pacific Power AC Power Source 1631 Jun. 24, 2013 (1Y)

All test equipments used are calibrated on a regular basis.

5.4.5. Test data

-. Test Date : March 22, 2014

-. Class : A

-. Remark : The input power of the EUT is lower than 75 W.

According to the cause 7 in EN 61000-3-2, no limit applies for the EUT.

See next page for Harmonics on AC Mains and DC test data.

Tested by: Jong-Jin. Lee / Engineer

Page 25: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 24 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

Page 26: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 25 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

5.5 VOLTAGE CHANGES, VOLTAGE FLUCTUATIONS AND FLICKER

5.5.1 Operating environment

Temperature : 24 °C

Relative humidity : 34 % R.H.

5.5.2 Test set-up The voltage changes at the supply terminals were measured across the complex reference impedance Z = (0.4 + j0.25) Ω.

The short-term flicker values are measured during a time interval of 10 min. Dc = relative voltage change between two

steady states and Dmax = maximum single voltage change, are measured over the reference impedance.

The test set-up photo is included in appendix IV.

5.5.3 Measurement uncertainty The uncertainty of our equipment for flicker measurement is ± 5 %.

Measurement uncertainty is calculated in accordance with EN 61000-3-3. The measurement uncertainty is given with a

confidence of 95 % with the coverage factor, k = 2.

5.5.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal.(Interval)

- PM6000 Voltech Universal Power Analyzer 1000067 00196 May 28, 2013 (1Y)

- IEC 61000-3 Voltech Reference Impedance Network IG 203/3401 May 28, 2013 (1Y)

- IEC 61000-3 Voltech Reference Impedance Network IG 119/3228 May 28, 2013 (1Y)

- IEC 555 Voltech Reference Impedance Network IB13/9886 May 28, 2013 (1Y)

- 134310 Pacific Power Magnetics Module 0283 Jun. 24, 2013 (1Y)

- 134350 Pacific Power Magnetics Module 1512 Jun. 24, 2013 (1Y)

- UPC32-413 Pacific Power Waveform Generator 0002 Apr. 15, 2013 (1Y)

- 360AMXT/UPC-32 Pacific Power AC Power Source 0222 Jun. 24, 2013 (1Y)

- 360AMXT/UPC-32 Pacific Power AC Power Source 1631 Jun. 24, 2013 (1Y)

All test equipments used are calibrated on a regular basis.

5.5.5 Test data -. Test Date : March 22, 2014

-. Test Result : Pass

-. Remark : See next page for Voltage Fluctuations on AC mains test data.

Tested by: Jong-Jin, Lee / Engineer

Page 27: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 26 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

Page 28: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 27 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

Page 29: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 28 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6. IMMUNITY TESTS

6.1 ELECTROSTATIC DISCHARGE IMMUNITY TEST

The measurement of the Immunity Against Electrostatic Discharge was performed in a shield room. Date: March 21, 2014

6.1.1 Operating environment Ambient temperature 22 [°C] Relative humidity 34 [% R.H.] Atmospheric pressure 100.8 [kPa]

6.1.2 Test set-up

The EUT and all local support equipment were placed on a non-conductive support 0.1 m above a reference ground plane

(RGP) and were put into operation according to the specified operating mode.

The test set-up photo is included in appendix V.

6.1.3 Measurement uncertainty

It has been demonstrated that the ESD generator meets the specified requirements in the standard with at least a 95 %

confidence.

6.1.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal.(Interval)

- ESS-2002EX NOISEKEN ESD Simulator ESS0898812 May 28, 2013(1Y)

All test equipment used is calibrated on a regular basis.

Page 30: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 29 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.1.5 Test data

Test levels : Contact discharge 6 kV, Air discharge 2 / 4 / 8 kV

Number of discharges : 10 at all test points.

Polarity : Positive / Negative

EUT-position : Floor Standing (Wall and Ceiling mounted Equipment)

Required performance criterion : See section 3.5 of this report

Test result : Pass

The EUT met the acceptance criteria for the functional test, after the conditioning.

Monitoring of the EUT : Observed the state of video output of the EUT on screen of notebook pc via LAN.

The test points of the EUT are each location on the surface touchable by hand (see test point in next page) and VCP -0.1

m from the four sides of the EUT.

The results of selected test points of the EUT are listed in below table.

Point Test level [± kV] Pass / Fail Description

Enclosure 2 / 4 / 8 (Air) Pass There was no deviation from normal operating condition

during and after test. VCP 6 (Indirect) Pass

Tested by: Jong-Jin. Lee / Engineer

Page 31: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 30 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.1.6 ESD Test Point & Test Result

ESD Point Discharge voltage[± kV] Results

(1) Enclosure 2 / 4 / 8 (Indirect) Pass

VCP 6 (Indirect) Pass

(1) (1)

Page 32: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 31 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.2 RADIATED RF-ELECTROMAGNETIC FIELD IMMUNITY TEST

The measurement of the Immunity Against Radiated RF-Electromagnetic Field was performed in an anechoic

chamber. Date: April 01, 2014

6.2.1 Operating environment Ambient temperature 21 [°C] Relative humidity 40 [% R.H.] Atmospheric pressure 100.0 [kPa]

6.2.2 Test set-up

The EUT and all local support equipment were placed on a non-metallic support 0.8 m above a reference ground plane

(RGP) and were put into operation according to the specified operating mode.

The test set-up photo is included in appendix VI.

6.2.3 Measurement uncertainty

The measurement uncertainty is ± 0.23 V/m for 1 V/m, ± 0.70 V/m for 3 V/m and ± 2.30 V/m for 10 V/m.

Measurement uncertainty is calculated in accordance with UKAS Lab34. The measurement uncertainty is given with a

confidence of 95 %. The measurement uncertainty is calculated as the uncertainty of the electric field intensity detected by

the probe(s). The uncertainty calculations exclude influence of phenomena like inhomogeneity of the electric field intensity.

6.2.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal. (Interval)

- SML 03 Rohde & Schwarz Signal Generator 102602 May 21, 2013 (1Y)

- NRVD Rohde & Schwarz Power Meter 101448 May 21, 2013 (1Y)

- HL046 Rohde & Schwarz Log Periodic Antenna 100053 Jan. 28, 2014 (1Y)

- AT4002A Amplifier Research Microwave Horn Antenna N/A Jan. 28, 2014 (1Y)

- 120S1G4M1 Amplifier Research Power Amplifier 333101 Feb. 22, 2013 (1Y)

- 500W1000A Amplifier Research Power Amplifier 332911 Jan. 28, 2014 (1Y)

- DC6180A Amplifier Research Directional Coupler 332598 Jan. 28, 2014 (1Y)

- DC7144M1 Amplifier Research Directional Coupler 313662 Jan. 28, 2014 (1Y)

- TA-EMS Rohde & Schwarz System Communication 11017 Jan. 28, 2014 (1Y)

All test equipment used is calibrated on a regular basis.

Page 33: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 32 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.2.5 Test data

Test level : 1 V/m, 3 V/m, 10 V/m

(AM 80 %, 1 kHz, sine wave / PM 50%, 1 Hz (0.5 s ON: 0.5 s OFF))

Frequency range : 80 MHz ~ 2 700 MHz

Frequency step : 1 %

Dwell time at each frequency : 3 s

Exposed side : Front / Back / Left / Right

Distance of antenna – EUT : 3 m

EUT-position : Table top

Required performance criterion : See section 3.5 of this report

Test result : Pass

The EUT met the acceptance criteria for the functional test, after the conditioning.

Monitoring of the EUT : Observed the state of video output of the EUT on screen of notebook pc via LAN.

The results of test are listed in below table.

Freq. Range [MHz]

Exposed side Pass/Fail Description

80 ~ 1 000 Front / Back / Left / Right

Pass There was no deviation from normal operating condition

during and after test. 1 000 ~ 2 700 Pass

Tested by: Jong-Jin. Lee / Engineer

Page 34: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 33 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.3 ELECTRICAL FAST TRANSIENT/BURST IMMUNITY TEST

The measurement of the Immunity Fast Transient/Burst was performed in a shield room. Date: March 24, 2014

6.3.1 Operating environment Ambient temperature 20 [°C] Relative humidity 33 [% R.H.] Atmospheric pressure 100.5 [kPa]

6.3.2 Test set-up

The EUT was placed on non-metallic support 0.1 m above a reference ground plane (RGP) and was put into operation

according to the specified operating mode. If the manufacturer provides a non-detachable supply cable more than 0.5 m

long with the equipment, the excess length of this cable shall be folded to avoid a flat coil and situated at a distance of 0.1

m above the ground reference plane.

The test set-up photo is included in appendix VII.

6.3.3 Measurement uncertainty

It has been demonstrated that the burst generator meets the specified requirements in the standard with at least a 95 %

confidence.

6.3.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal. (Interval)

- UCS 500 M4 S1 EM Test Ultra Compact Generator 0304-20 May 30, 2013(1Y)

- CNI 503-32A EM Test Combined Coupling Decoupling Net 0304-02 Jul. 16, 2013(1Y)

- H.F.K EM Test Capacitive Coupling Clamp N/A May 28, 2013(1Y)

- AC2716 EM Test Power Source 93. 30. 1037 May 30, 2013(1Y)

All test equipment used is calibrated on a regular basis.

Page 35: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 34 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.3.5 Test data

Test level : 2 kV (AC-mains supply), 1 kV (Other supply / Signal lines > 3 m)

Burst frequency : 100 kHz

Rise time/Hold time : 5/50 ns

Polarity : Positive / Negative

Coupling time : 72 s

Coupling methods : AC mains - Coupling/decoupling network (CDN)

Signal lines - Capacitive Coupling Clamp (CCC)

Lines for test : AC Mains of the Adaptor

Signal lines of the EUT

Type of line & length : Unshielded, 0.5 m (AC Mains)

Unshielded, 3.0 m (RJ-45 - Signal line)

EUT-position : Table top

Required performance criterion : See section 3.5 of this report

Test result : Pass

The EUT met the acceptance criteria for the functional test, after the conditioning.

Monitoring of the EUT : Observed the state of video output of the EUT on screen of notebook pc via LAN.

The results of test are listed in below table.

Line for test Coupling Method

Test level [± kV]

Pass/Fail

Description

AC-mains (L/N) CDN 2 PassThere was no deviation from normal operating

condition during and after test. AC-mains (L+N) CDN 2 Pass

Signal line CCC 1 Pass

Tested by: Jong-Jin. Lee / Engineer

Page 36: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 35 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.4 SURGE IMMUNITY TEST

The measurement of the Immunity Surge Transients was performed in a shield room. Date: March 24, 2014

6.4.1 Operating environment Ambient temperature 20 [°C] Relative humidity 33 [% R.H.] Atmospheric pressure 100.5 [kPa]

6.4.2 Test set-up

The EUT and all local support equipment were placed on a non-metallic support 0.8 m above a reference ground plane

(RGP) and were put into operation according to the specified operating mode.

The test set-up photo is included in appendix VIII.

6.4.3 Measurement uncertainty

It has been demonstrated that the surge generator meets the specified requirements in the standard with at least a 95 %

confidence.

6.4.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal. (Interval)

- UCS 500 M4 S1 EM Test Ultra Compact Generator 0304-20 May 30, 2013(1Y)

- CNI 503-32A EM Test Combined Coupling Decoupling Net 0304-02 Jul. 16, 2013(1Y)

- CNV508 EM Test Coupling Network N/A N/A

- AC2716 EM Test Power Source 96. 30. 1037 May 30, 2013(1Y)

- HV Gun SQS HV Gun N/A N/A

All test equipments used is calibrated on a regular basis.

Page 37: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 36 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.4.5 Test data

Test level : 0.5 kV, 1 kV (Common mode of AC mains)

0.5 kV, 1 kV (Signal lines > 30 m)

Number of surge : 5 surges/ polarity

Polarity : Positive / Negative

Angle : 0° / 90° / 180° / 270°

Repetition rate : 60 s

Coupling methods : AC mains - Coupling/decoupling network (CDN)

Signal lines – Coupling/decoupling network (CDN)

Lines for test : AC Mains of the Adapter

Signal line of the EUT

Type of line & length : Unshielded, 0.5 m (AC Mains)

Shielded, 30.0 m (Signal cable - RJ-45)

EUT-position : Table top

Required performance criterion: See Item 3.5 of this report

Test result : Pass

The EUT met the acceptance criteria for the functional test, after the conditioning.

Monitoring of the EUT : Observed the state of video output of the EUT on screen of notebook pc via LAN.

The results of test are listed in below table.

Line for test Coupling Method

Test level [± kV]

Pass/ Fail

Description

AC-mains (L+N) CDN 0.5 / 1 Pass There was no deviation from normal operating

condition during and after test Signal lines CDN 0.5 / 1 Pass

Tested by: Jong-Jin. Lee / Engineer

Page 38: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 37 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.5 CONDUCTED RF FIELDS IMMUNITY TEST

The measurement of the Immunity Against Conducted RF Field was performed in the anechoic chamber. Date: March 24, 2014

6.5.1 Operating environment Ambient temperature 20 [°C] Relative humidity 33 [% R.H.] Atmospheric pressure 100.5 [kPa]

6.5.2 Test set-up

The EUT and all local support equipment were placed on a non-metallic support 0.1 m above a reference ground plane

(RGP) and were put into operation according to the specified operating mode.

The test set-up photo is included in appendix IX.

6.5.3 Measurement uncertainty

The measurement uncertainty is ± 0.17 V for 1 V, ± 0.50 V for 3 V and ± 1.70 V for 10 V.

Measurement uncertainty is calculated in accordance with EN 61000-4-6 Annex G. The measurement uncertainty is given

with a confidence of 95 %.

6.5.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal. (Interval)

- CWS 500 N1 EM Test Continuous Wave Generator P1247105431 Aug. 27, 2013 (1Y)

- CDN-M2/M3 EM Test Coupling Decoupling Net P1312115835 Aug. 27, 2013 (1Y)

- EM 101 Fisher Custom Injection Clamp 35364 May 06, 2013 (1Y)

- 100-A-FFN-06 Bird Technologies Group Attenuator 751 May 21, 2013 (1Y)

All test equipments used is calibrated on a regular basis.

Page 39: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 38 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.5.5 Test data

Test level : 1 V, 3 V, 10 V

(AM 80 %, 1 kHz sine wave / PM 50 %, 1 Hz (0.5 s ON: 0.5 s OFF))

Frequency range : 0.15 MHz ~ 100 MHz

Frequency step : 1 %

Dwell time at each frequency : 3 s

Coupling methods : AC mains - Coupling/decoupling network (CDN)

Various input/output ports >3 m - EM clamp

Lines for test : Power line of the Adaptor

Signal line of the EUT

Type of line & length : Unshielded, 0.3 m (AC Mains)

Shielded, 3.0 m (Signal cable - RJ-45)

EUT-position : Table top

Required performance criterion: See Item 3.5 of this report

Test result : Pass

The EUT met the acceptance criteria for the functional test, after the conditioning.

Monitoring of the EUT : Observed the state of video output of the EUT on screen of notebook pc via LAN.

The results of test are listed in below table.

Lines for test

Freq. Range [MHz]

Test level [V]

Pass/Fail

Coupling Method Description

AC mains 0.15 ~ 100 1 / 3 / 10 Pass CDN There was no deviation from normal operating

condition during and after test. Signal lines 0.15 ~ 100 1 / 3 / 10 Pass EM clamp

Tested by: Jong-Jin. Lee / Engineer

Page 40: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 39 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.6 MAINS SUPPLY VOLTAGE DIPS AND SHORT INTERRUPTIONS IMMUNITY TEST

The measurements of the Immunity Mains Supply Voltage Dips and Interruptions were performed in a shield room. Date: March 24, 2014

6.6.1 Operating environment Ambient temperature 20 [°C] Relative humidity 33 [% R.H.] Atmospheric pressure 100.5 [kPa]

6.6.2 Test set-up

The EUT and all local support equipment were placed on a non-metallic support 0.8 m above a reference ground plane

(RGP) and were put into operation according to the specified operating mode.

The test set-up photo is included in appendix X.

6.6.3 Measurement uncertainty

It has been demonstrated that the voltage dips and interruptions generator meets the specified requirements in the standard

with at least a 95 % confidence.

6.6.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal. (Interval)

- UCS 500 M4 S1 EM Test Ultra Compact Generator 0304-20 May 30, 2013(1Y)

- AC2716 EM Test Power Source 93. 30. 1037 May 30, 2013(1Y)

All test equipments used is calibrated on a regular basis.

Page 41: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 40 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.6.5 Test data

Nominal Mains Voltage (VNOM) : 230 V~, 50 Hz

Level of Reduction (dip) : 200 ms at 60 % of VNOM, 500 ms at 30 % of VNOM, 5 000 ms at 20 % of VNOM

No. of dips/interruption : 3

Interval : 10 s

Type of line & length : AC mains of the adapter – Unshielded, 1.0 m

EUT-position : Table top

Required performance criterion : See section 3.5 of this report

Test result : Pass

The EUT met the acceptance criteria for the functional test, after the conditioning.

Monitoring of the EUT : Observed the state of video output of the EUT on screen of notebook pc via LAN.

The results of test are listed in below table.

Test Reduction (% of VNOM)

Duration in time (ms)

Pass / Fail Description

Voltage dips

20 5 000 Pass There was no deviation from normal operating condition

during and after test. 30 500 Pass

60 200 Pass

Voltage Interruptions 100 5 000

N/A

(See Note)

The power of EUT is off during the test. After the test,

EUT is getting back to normal operation.

Note.: “It fully recorded using ancillary Power source equipment to content with Manufacturer`s set up manual.”

Tested by: Jong-Jin. Lee / Engineer

Page 42: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 41 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.7 MAINS SUPPLY VOLTAGE VARIATIONS IMMUNITY TEST

The measurement of the Immunity Mains Supply Voltage Variations was performed in a shield room. Date: March 24, 2014

6.7.1 Operating environment Ambient temperature 20 [°C] Relative humidity 33 [% R.H.] Atmospheric pressure 100.5 [kPa]

6.7.2 Test set-up

The EUT and all local support equipment were placed on a non-metallic support 0.8 m above a reference ground plane

(RGP) and were put into operation according to the specified operating mode.

The test set-up photo is included in appendix XI.

6.7.3 Measurement uncertainty

It has been demonstrated that the Mains Supply Voltage Variations generator meets the specified requirements in the

standard with at least a 95 % confidence.

6.7.4 Test equipment used

Model Number Manufacturer Description Serial Number Last Cal. (Interval)

- UCS 500 M4 S1 EM Test Ultra Compact Generator 0304-20 May 30, 2013(1Y)

- AC2716 EM Test Power Source 93. 30. 1037 May 30, 2013(1Y)

All test equipments used is calibrated on a regular basis.

Page 43: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 42 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

6.7.5 Test data

Nominal Mains Voltage (VNOM) : 230 V~

Voltage variations : VNOM + 10 %, VNOM – 15 %

Interval : 10 s

Type of line & length : AC mains of the Adaptor – Unshielded, 1.0 m

EUT-position : Table top

Required performance criterion : See section 3.5 of this report

Test result : Pass

The EUT met the acceptance criteria for the functional test, after the conditioning.

Monitoring of the EUT : Observed the state of video output of the EUT on screen of notebook pc via LAN.

The results of test are listed in below table.

Test Voltage (% of VNOM) Pass / Fail Description

Voltage Variations +10 Pass There was no deviation from normal operating condition

during and after test. -15 Pass

Tested by: Jong-Jin. Lee / Engineer

Page 44: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 43 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX I - TEST SET-UP PHOTOS: (Mains Terminal Continuous Disturbance Voltage)

Page 45: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 44 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX II - TEST SET-UP PHOTOS: (Conducted Common Mode Disturbance at Telecommunication Ports)

Page 46: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 45 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX III - TEST SET-UP PHOTOS: (Radiated Electromagnetic Field)

30 MHz ~ 1 000 MHz

Page 47: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 46 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

1 000 MHz ~ 2 000 MHz

Page 48: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 47 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX IV - TEST SET-UP PHOTO: (Harmonics & Voltage Fluctuations on AC Mains)

Page 49: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 48 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX V – TEST SET-UP PHOTO: (ESD)

Page 50: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 49 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX VI – TEST SET-UP PHOTO: (RF E- FIELD)

Page 51: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 50 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX VII - TEST SET-UP PHOTO: (EFT/BURST)

Page 52: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 51 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX VIII - TEST SET-UP PHOTO: (SURGE)

Page 53: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 52 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX IX - TEST SET-UP PHOTO: (C.S)

Page 54: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 53 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX XI - TEST SET-UP PHOTO: (MAINS SUPPLY VOLTAGE DIPS, SHORT INTERRUPTIONS)

Page 55: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 54 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX XI - TEST SET-UP PHOTO: (Power Supply Short Term Variation)

Page 56: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 55 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)

APPENDIX XII - PHOTOGRAPHS OF THE PRODUCT

Page 57: Accredited Testing Laboratory under the terms of ISO/IEC ... · It should not be reproduced except in full, without the written approval of ONETECH Corp. SAFETY-CE:DoC(Rev.2) HEAD

Page 56 of 52 Report No.: E144R-009

It should not be reproduced except in full, without the written approval of ONETECH Corp. EMC-009(Rev.2) HEAD OFFICE : 301-14 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599) EMC Testing Div. : 307-51 Daessangnyeong-ri, Chowol-eup, Gwangju-si, Gyeonggi-do 464-862 Korea (TEL: 82-31-765-8289, FAX: 82-31-766-2904)