ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

28
ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011

Transcript of ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

Page 1: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

ABL80 FLEX CO-OX- OSM version

Product OverviewFebruary 2011

Duff Pickering – February 28, 2011

Page 2: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

2

OSM version of ABL80 FLEX CO-OX – Key Features

CO-OX parameters only Measured parameters: ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHHb Derived parameters: ctO2, Hct, BO2, V(B)Sample volume: 65µL (ABL80 CO-OX: 105µL)Measuring/Cycle time: 60 secs. (ABL80 CO-OX: 140 secs.)New set of Sample Site options for Cardiac Cath Lab applicationsAutomatic QC 2 levels in one QC event (System Cycle) performed once every 24 hrs.

Available System Cycle frequencies are 24, 12, 8, 4, and 2 hours

Extended In-use life of consumables Solution pack: 60 day in-use life

(60 days is maximum in-use life; realized in-use life depends on usage, including number of patient tests and use of Standby Mode)

New Standby mode to maximize in-use life of solution pack 60 days of in-use life can be achieved without Standby, up

to 3 patient tests per day assumes 24 hour system cycle frequency and no manual QC

Sensor cassette: 60 day in-use life 600 patient tests

Extended Shelf life of consumables Solution pack: 180 days Sensor cassette: 180 days

Fast Start-up time of new consumables(approx. 4 minutes for SP and 30 seconds for SC)

Page 3: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

3

Sample Site options – Specific for Cardiac Cath Lab

The OSM version of the ABL80 FLEX CO-OX can be configured with a set of Sample Site options that are specific for clinical applications in the Cardiac Cath Lab

This new set of Sample Sites for the Cardiac Cath Lab include:

Arterial line Mean PA SVC

R. atrium RPA IVC

L. Atrium LPA RINN

R. ventricle PA wedge LINN

L. ventricle Ascending Ao Femoral artery

The use of this new set of sample site options must be set up by a service technician.

The procedures for the service technician to follow are described in the ABL80 Service Manual (revision F), Chapter 8 – Service Procedures, in the “Sample site options” section.

It is also possible to customize the sample site options. This customization can be done only by a service technician and is

also described in the Service Manual (revision F), Chapter 8 – Service Procedures, in the “Sample site options” section.

Page 4: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

4

System Configuration

ABL80 FLEX CO-OX analyzer Now with either OSM software or CO-OX

software Analyzer hardware is the same

ABL80 FLEX CO-OX – OSM version

ABL80 FLEX CO-OX- CO-OX version

OSM SW CO-OX SW

ABL80 FLEX CO-OX analyzer hardware

- plus -

Software version

System Configuration

393-841

Page 5: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

5

System Configuration

Two configurations of analyzer: OSM and CO-OX Two sets of consumables: OSM and CO-OX

An

alyz

erC

on

sum

able

s

?

Not possible to use

OSM consumables on CO-OX analyzer

or

CO-OX consumables on OSM analyzer

OSM CO-OX

CO-OX consumables

OSM consumables

Page 6: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

6

Field Conversions

The ABL80 FLEX CO-OX system configuration can be changed in the field CO-OX to OSM OSM to CO-OX

Items needed for Field Conversion: SW Conversion CD

CO-OX to OSM Conversion CD, v1.50, P/N 933-134 OSM to CO-OX Conversion CD, v1.36, P/N 933-136

Option key – serial number specific CO-OX to OSM Option Key - P/N 933-135 OSM to CO-OX Option Key - P/N 933-137

When a conversion occurs, the consumables on the analyzer before the conversion occurs can not be used on the newly converted analyzer. These pre-conversion consumables will not be compatible with the new, post-conversion version of the analyzer.

This is an issue to anticipate and coordinate when making the conversion

CO-OXOSM

Page 7: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

7

Main Menu: CO-OX vs OSM systems

Appearance of Main Menu

Main Menuwith CO-OX SW

Differences between the two Main Menus:1. OSM Watermark with the OSM SW2. “Enter Standby” button with the OSM SW (No Standby on CO-OX SW)3. Only CO-OX parameters in Parameter Bar

Main Menuwith OSM SW

Page 8: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

8

Identification of SW configuration

ABL80 FLEX CO-OX – OSM version

ABL80 FLEX CO-OX- CO-OX version

Sys

tem

In

form

atio

n s

cree

nP

rin

tou

t

Page 9: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

9

Consumables - Labels

The name “OSM” does not appear on the labels.The differentiation of consumables is achieved by Color on labels Part numbers

SP80 for CO-OX SW

SC80 for CO-OX SW

Existing labels, for ABL80 CO-OX

SP80 for OSM SW

New Labels, for OSM80

SC80 for OSM SW

Page 10: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

10

New steps to order and ABL80 FLEX CO-OX analyzer

Part number 393-841 is still used to order an ABL80 FLEX CO-OX analyzer, either for OSM or CO-OX configuration

Now, a second part number must also be ordered to indicate which software version, OSM or CO-OX, should be pre-installed at SenDx

933-133 Preinstalled OSM SW 933-132 Preinstalled CO-OX SW

An order placed for 393-841 without a corresponding order for the type of software is an incomplete order.

An incomplete order can not be processed at SenDx It will be necessary to order the pre-installed SW version, too.

Page 11: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

11

Process to order a new ABL80 FLEX CO-OX analyzer, after OSM version is released

Item 1All orders for new ABL80 FLEX CO-OX analyzers include Item 1

ABL80 FLEX CO-OX analyzer P/N 393-841 Currently released in Axapta to all sales companies

Item 2All orders for new ABL80 FLEX CO-OX analyzers include either A or B, as an indication of the SW version to be installed and shipped

A. OSM SW – Pre-installed P/N 933-133 Released in Axapta only to countries selling the OSM version

B. CO-OX SW – Pre-installed P/N 933-132 Released in Axapta to all countries

Shipped items

1. ABL80 FLEX CO-OX analyzer with OSM SW preinstalled

2. Plastic bag, with 933-133 label, containing CD with OSM SW

1. ABL80 FLEX CO-OX analyzer with CO-OX SW preinstalled.

2. Plastic bag, with 933-132 label, containing CD with CO-OX SW

Invoiced items1. 393-841 ABL80 FLEX CO-OX analyzer2. 933-133 OSM SW preinstalled

1. 393-841 ABL80 FLEX CO-OX analyzer2. 933-132 CO-OX SW preinstalled

Shipping documents – items listed

1. 393-841 ABL80 FLEX CO-OX analyzer2. 933-133 OSM SW preinstalled

1. 393-841 ABL80 FLEX CO-OX analyzer2. 933-132 CO-OX SW preinstalled

Process to order ABL80 CO-OX analyzer after OSM is released

Page 12: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

12

The analyzer shipping box

Inside the boxOutside the box

CO

-OX

OS

M

Preinstalled CO-OX SW 933-132

Preinstalled OSM SW 933-133

Page 13: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

13

Shelf life and In-use life on Consumables

ABL80 FLEX CO-OX ABL80 FLEX

OSM version CO-OX version BASIC version FLEX version

Shelf life

Solution pack 180 days 90 days150 days

(target 180 days)120 days

Sensor cassette 180 days 90 days120 days

(target 180 days)120 days

In-use life

Solution pack(maximum number of

days)

60 daysDepending on usage

Including the frequency that

Standby Mode is used

30 daysDepending on usage

60 daysDepending on usage Including the frequency

that Standby Mode is used

30 daysDepending on

usage

Sensor cassette 60 days30 days

(some versions have 15 days

60 days(some versions have 30

or 15 days)

30 days(some versions have 15 days)

Part No. Consumables for the OSM configuration of the ABL80 FLEX CO-OX

944-341 Solution pack

945-806 Sensor cassette - 600 tests / 60 days

OSM Consumables

Page 14: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

14

Solution pack construction

All solution packs for the series of ABL80 analyzers contain four pouches

1

2

3

4

Page 15: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

15

OSM80 Solution pack

The solution pack for the OSM80 has only three solutions in the four pouches. Solution 1 is used to flush the system after every sample and to perform

blank calibrations of the oximetry system. Solutions 2 and 3 contain dyes and are used to perform the automatic QC,

involving two QC levels in one System Cycle. The default frequency of a System Cycle is once every 24 hours. The frequency of the System Cycle can be set to be 24, 12, 8, 4, or 2 hours.

480 cycles of Solution 1 are distributed across two solution pouches. There are 150 cycles each of Sol2 and Sol 3. In contrast, the solution pack for the CO-OX version has four solutions to provide

the multiple levels of automatic QC.

Sol 1 – 230 cycles C8301

Sol 3 – 110 cycles C8303

Sol 2 – 110 cycles C8302

Sol 4 – 110 cycles C8304

FLEX CO-OX Solution pack

48

0 C

yc

les

of

So

l 1

Sol 1 – 330 cycles C8701

Sol 1 – 150 cycles C8703

Sol 2 – 150 cycles C8702

Sol 3 – 150 cycles C8704

OSM80 Solution pack

Page 16: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

16

Cycles remaining in OSM solution pack

Information about the remaining cycles for each solution are found in the Solution Pack tab under System Information. The number of cycles remaining in each of the three Solution 1 pouches are shown on this screen

Sol 1 – 330 cycles C8701

Sol 1 – 150 cycles C8703

Sol 2 – 150 cycles C8702

Sol 3 – 150 cycles C8704

OSM80 Solution pack

Sol 1 – 330 cycles C8701

Sol 1 – 150 cycles C8703

Sol 2 – 150 cycles C8702

Sol 3 – 150 cycles C8704

OSM80 Solution pack

Page 17: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

17

Standby Mode

A Standby Mode feature is included with the OSM version. Standby on the OSM is similar to Standby on the ABL80 BASIC

Key purpose of Standby Mode is to conserve the use of solutions to prolong the in-use life of the solution pack in order to reduce cost of operation as much as possible.

Most important objective of Standby Mode is to conserve Solution 1. Saving available cycles of Solution 1 is achieved by postponing blank

calibrations and System Cycles when possible

The maximum in-use life of 60 days for the OSM solution pack can be achieved even if Standby is not used, if the number of patient tests run per day is 3 or less.

That is different than on the ABL80 BASIC, where the maximum in-use life of 60 days can only be achieved if standby mode is used on the weekend and every night.

Sol 1 – 330 cycles C8701

Sol 1 – 150 cycles C8703

Sol 2 – 150 cycles C8702

Sol 3 – 150 cycles C8704

OSM80 Solution pack

Page 18: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

18

How Standby Mode Works

During the Entry to Standby Mode When entering Standby, air is aspirated through the sensor cassette, but not into the hemolyzer

cuvette. This means that solution remains in the hemolyzer’s cuvette during the Standby period, but not in the sensor cassette segment of the fluidic path.

This is different from the ABL80 BASIC where solution remains in the sensor cassette during Standby.

While In Standby Mode No fluidic activity occurs while the system remains in Standby Mode Specifically, no System Cycles or blank calibrations occur as long as the system remains in Standby

Mode. No cycles of Solutions 1, 2, or 3 are used during the standby period.

This is different from the ABL80 BASIC where one-half flush cycle of Solution 1 occurs every 4 hours when that system is in Standby.

During the Exit from Standby Mode An aspiration / flush routine is run that consumes one half cycle of Solution 1.

After Exiting Standby Mode If the standby period was longer than 48 hours, an extra flush routine is used which consumes a half

cycle of Solution 1. In addition, the hemolyzer is activated for four seconds during the flush sequence.

A Blank Calibration is automatically performed if it has not been performed within the previous 12 hours

A System Cycle is automatically performed if it has not been performed within the previous 24 hours A valid, non-expired sensor cassette and solution pack must be installed on the analyzer for either

the automatic Blank Calibration or System Cycle mentioned above to occur when exiting Standby

The number of tests remaining on the sensor cassette does not affect the standby status.

Page 19: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

19

Enter/Exit Standby Mode

1. Manual “on-demand” entry and exit of Standby To enter, press the “Enter Standby” button. Upon entering Standby Mode,

the traffic light will turn to yellow the system will continue to display “Ready” as an indication that the

patient measurements are possible if Standby Mode is exited. “Standby” will appear directly under “Ready”.

The system will stay in Standby until the user manually exits Standby Unless a schedule to exit Standby has been previously set up 30 days is the maximum Standby period

To exit Standby, press the “Exit Standby” button.

Enter Standby, Manually Exit Standby, Manually

Page 20: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

20

Enter/Exit Standby Mode

2. Scheduled automatic entry to and exit from Standby It is possible to set up the automatic entry to and exit from Standby as:

1. Recurring Weekly schedule (up to 7 different schedules)2. One Time event – 30 days is the maximum length of a scheduled

Standby period To access the schedule option, from the Main Menu,

Select Menu Settings Standby

To establish a recurring weekly schedule

To schedule a One Time event

Page 21: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

21

Use of Standby Mode – Impact on In-use life

Examples of usage scenarios to reach the maximum of 60 days in-use life for the solution pack are:

Estimates of In-use life for BASIC solution pack under different usage scenarios

Use of Standby(Weekly basis)

Patient Tests/day

In-use days achieved for Solution Pack

None 3 60

7 Nights only 4 60

2 Full Days only 7 60

2 Full Days + 5 nights 7 60

Assumption: No manual QC/dayFull Day Standby = 24 hour periodNight Standby = 12 hour periods

The Financial Workbook (FWB) version 8.00 includes a calculator feature to determine the in-use life of an OSM solution pack that would be realized under different usage scenarios.

The calculator in the FWB ver. 8.00 can also be used to estimate the in-use life of the solution pack for the ABL80 BASIC, ABL80 FLEX, and ABL80 FLEX CO-OX (CO-OX version) under different usage scenarios.

Page 22: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

22

In-use life of OSM solution pack

 OSM version

Standby Scenario - Weekly ScheduleCO-OX version

 No Standby

Standby7 nights

Standby2 Full days

(Weekend) only

Standby2 Full days (Weekend)

+ 5 nightsNo Standby

Tests per day

Days In-use

No. of packs per year

Days In-use

No. of packs

per year

Days In-use

No. of packs

per yearDays In-use

No. of packs per year

Days In-useNo. of packs

per year

0 60 6.1 60 6.1 60 6.1 60 6.1 30.0 12.2

1 60 6.1 60 6.1 60 6.1 60 6.1 30.0 12.2

2 60 6.1 60 6.1 60 6.1 60 6.1 28.2 12.9

3 60 6.1 60 6.1 60 6.1 60 6.1 25.1 14.5

4 59 6.2 60 6.1 60 6.1 60 6.1 22.6 16.2

5 52 7.0 55 6.6 60 6.1 60 6.1 20.6 17.7

6 47 7.8 49 7.4 60 6.1 60 6.1 18.9 19.3

7 43 8.5 45 8.2 60 6.1 60 6.1 17.5 20.9

8 39 9.3 41 8.9 55 6.6 57 6.4 16.2 22.5

9 36 10.1 38 9.7 51 7.2 53 6.9 15.2 24.0

10 34 10.8 35 10.5 47 7.7 49 7.5 14.2 25.7

11 31 11.6 33 11.2 44 8.3 45 8.0 13.4 27.2

12 30 12.4 30 12.0 41 8.8 42 8.6 12.7 28.7

13 28 13.1 29 12.8 39 9.4 40 9.1 12.0 30.4

14 26 13.9 27 13.5 37 9.9 38 9.7 11.4 32.0

15 25 14.7 26 14.3 35 10.5 36 10.2 10.9 33.5

20 20 18.5 20 18.1 28 13.2 28 13.0 8.8 41.5

Assumptions: 1. No Manual QC performed, and 2. System Cycle Frequency of 24 hours

Page 23: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

23

Cautions about using Standby

Frequent entry and exit of Standby, more than once every 12 hours, can consume more Solution 1 cycles than are saved.

Because one cycle of Solution 1 is used for every exit from Standby.

To save solutions, the Standby period should be at least 12 hours in order to postpone at least one Blank Calibration

This extends the availability of Solution 1.

The conservation of Solutions 2 and 3 is the important benefit of the Standby Mode, especially for low volume users.

Optimal benefit from Standby Mode is achieved by extended standby periods, such as a weekend or longer.

Information about Standby to End-users

It may be necessary to actively remind and inform end-users about the need to use Standby to maximize the in-use life of the solution pack

Active use of the Scheduling options to setup automatic entry to and exit from Standby may be the best approach to ensure regular use of the Standby feature.

Page 24: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

24

How to Enter/Exit Standby Mode

To enter Standby, the system requires that a sensor cassette be connected (though not necessarily installed) a solution pack be installed on the analyzer A minimum of 2 cycles of Solution 1 must be available for the system to

enter Standby

The entry to and exit from Standby Mode can be achieved in three ways:1. Manual “on-demand” steps2. Scheduled automatic steps3. Remotely enter (and exit) standby from RADIANCE

Page 25: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

25

While in Standby

The user can access all areas of the software including logs and settings.

The system will display information on the Main Menu about the status of the consumables and the length of time that the system will support Standby.

The system can remain in Standby for a period up to 30 days. For periods less than 30 days, the system can remain in standby as long as cycles

remain in the solution pack, even if the solution pack is past its expiration date. When in a “one-time” scheduled standby event, the main menu will display the

date and time the analyzer is scheduled to exit standby. The system will not allow

1. patient analysis 2. manual QC, 3. manual System Cycle 4. sensor cassette installation.

Entering and exiting standby is recorded in the Event log.

Page 26: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

26

QUALICHECK 5+ is used with the OSM version

There are unique insert ranges for QUALICHECK 5+ for use with the OSM. Due to unique usage conditions of the consumables on the OSM, in

particular the extended in-use life and smaller sample volume.

QC5+ insert ranges for the CO-OX version should not be used with the OSM version, or vice versa.

The use of CO-OX ranges on a OSM system, and OSM ranges on a CO-OX system, will likely result in out-of-range results.

These incorrect out-of-range results would falsely indicate that there were problems with the OSM system.

Supplementary inserts for QC5+ with OSM ranges are available on the RMED extranet for the following lots:

QUALICHECK 5+ Control Ranges Lots

Level 1 S7730 0083-0099

Level 2 S7740 0088-0104

Level 3 S7750 0085-0101

Level 4 S7760 0082-0097

Page 27: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

27

Range+ material and the OSM version

Range+ QUALICHECK

Control Ranges

Lots

Level 1 S7930 0007-0009

Level 2 S7940 0009-0011

Level 3 S7950 0006-0008

There are unique insert ranges for Range+ for use with the OSM. Due to unique usage conditions of the consumables on the OSM, in

particular the extended in-use life and smaller sample volume.

Range+ insert ranges for the CO-OX version should not be used with the OSM version, or vice versa.

The use of CO-OX ranges on a OSM system, and OSM ranges on a CO-OX system, will likely result in out-of-range results. These incorrect out-of-range results would falsely indicate that there

were problems with the OSM system.

Supplementary inserts for Range+ with OSM ranges are available on the RMED extranet for the following lots:

Page 28: ABL80 FLEX CO-OX - OSM version Product Overview February 2011 Duff Pickering – February 28, 2011.

28

tHb calibration solution (S7770) and the OSM version

The use of the tHb calibration solution (S7770) is the same on all ABL80 FLEX CO-OX analyzers, with either current CO-OX software or the new OSM software.

The barcode on the tHb calibration solution insert identified for the ABL80 FLEX CO-OX is used on analyzers with either software version, CO-OX and OSM.

There isn't a new or unique barcode for an ABL80 FLEX CO-OX analyzer with OSM software.

The recommended frequency of tHb calibration frequency is once every three months for a system with OSM software, as it is for a system with CO-OX software.