AACB NSW/ACT QC Subcommittee TnT hs Survey Results Methods Results Introduction Persephone...

1
AACB NSW/ACT QC Subcommittee TnT hs Survey Results Methods Results Introduction Persephone Giannopoulos ¹ and representatives of the AACB NSW/ACT QC Subcommittee. ¹Department of Chemical Pathology, St Vincent’s Hospital, Sydney, Australia. [email protected] With the introduction of the Roche high sensitive Troponin T assay, it allows the measurement of Troponin at very low concentrations of the analyte. One initial challenge was the lack of commercial quality control (QC) and external quality assessment (EQA) material at suitable concentrations for the new assay. Specifically RCPA QAP and value assigned commercial QC materials were not available at low concentrations. The control and assessment of assay performance near the 99 th centile of normal (14ng/L) is important, as vital clinical decisions are made using these results. Method validations had given our laboratories initial confidence to commission the assay however there were some reservations about testing commutability of patient sample precision at the low end of the assay over time. AACB Annual Scientific Meeting, Sydney 2011 Aim Tables and Figures Conclusions Graph 4: Repeat Analysis of human serum pool samples on the 2 separate occasions REFERENCES (1) Giannitsis et al. Clin Chem: 2010;56:254 -261 . The aim of the study was to determine between laboratory assay precision of the TnT hs assay using two pooled serum samples at low levels. Discussion Table 1 & Graph 1: Precision for all samples, human serum pools 1 and 2, Roche QCs , Biorad QCs and QAP material*(QAP data for all TnT hs participants) Graph 2: Box and Whisker plots of sample pool values (1) and (2) for September 2010 data, (---) lines are the mean values. Graph 3: Box and Whisker plots of sample pool values (1) and (2) for February 2011 data, (---) lines are the mean values. Roche Troponin T hs (high sensitive)* assay (05092744 190) - 200 tests reagent (E170, Cobas e 411, cobas e 601, cobas e 602). *Note: Enhancements to 4 th generation by: 1. Detection antibody genetically reengineered. 2. Analytical sensitivity improved by increasing sample volume, increasing ruthenium concentration of detection antibody and lowering background signal via buffer optimization.(1) Two pooled patient samples were distributed to 9 laboratories on two occasions September 2010 and February 2011. Internal QCs from Roche Precicontrol 1 and 2 (#05095107190) and Biorad Cardiac Markers Control LT (Cat # 646, 647) were run according to the laboratory protocols. The General Serum Chemistry RCPA QAP samples for Troponin T hs were in use for routine laboratory monitoring of the assay. QAP data was obtained from the RCPA QAP website and was used with permission. AACB NSW/ACT QC Sub Committee members Excellent precision was obtained for all the human serum pools from all nine laboratories. The September precision for serum pool 1 appeared to be slightly better than the February data. There were no differences with the CVs for serum pool 2. The Roche QC data gave consistently low CVs for the two measurements ( Sept and Feb). The Biorad QC data gave consistent CVs with the two measurements. The Biorad QC1 level measured lower than the Roche low level. The QAP precision was very good in September (n=27 laboratories), the February (n=61 laboratories) CVs poorer in February. The human serum pools for this survey were at very low levels giving us a good indication of the between- laboratory and between instrument performance of the TnT hs assay at these clinically important low concentrations. The precision of the commercial controls tested was excellent. Most of the commercial controls do not measure as low as the human serum pool 1. The repeat analysis of the human sample pools also provided within laboratory precision for each laboratory. Laboratory 3 and 4 did not perform as well as the other laboratories with human serum pool 1 when compared to the other laboratories, however this level is at the LOD of the assay. The QAP data gives us some confidence unfortunately troponin levels of the QAP material with the lowest concentration are still not at the critical and clinical decision making concentrations (i.e.14ng/L). These survey results have given us confidence that the between laboratory precision for low human serum values using the Roche TnT hs assay is appropriate for clinical decision making. We will continue to distribute human serum pools at these low concentrations to ensure ongoing performance of the Roche TnT hs assay through the AACB NSW/ACT QC Sub committee. Acknowledgements: Thank you to the AACB for their financial support, to all participating laboratories and to Jadwiga Francis for the sample preparations and distributions. Troponin T hs -PR EC ISIO N TABLE Sept2010 (average;n=9) Feb 2011 (average;n=9) TnT hs (ng/L) CV% TnT hs (ng/L) CV% Serum Pool 1 6.5 15.3 6.4 19.3 Serum Pool 2 18.7 4.4 21.6 4.6 R oche Q C 1 24.3 4.4 24.5 4 R oche Q C 2 2142 2 2083 2 Biorad C M 1 10.3 9.7 10.6 8.1 Biorad C M 2 52.3 6.4 53.6 5 Q AP 85-13 42* 5 Q AP 86-02 50* 9 Septem ber2010 data 1 2 Troponin T concentration (ng/L) 0 5 10 15 20 25 TnT hs (1) TnT hs (2) February 2011 data Troponin T concentration (ng/L) 0 5 10 15 20 25 TnT hs (1) TnT hs (2)

Transcript of AACB NSW/ACT QC Subcommittee TnT hs Survey Results Methods Results Introduction Persephone...

Page 1: AACB NSW/ACT QC Subcommittee TnT hs Survey Results Methods Results Introduction Persephone Giannopoulos¹ and representatives of the AACB NSW/ACT QC Subcommittee.

AACB NSW/ACT QC Subcommittee TnT hs Survey Results

Methods

Results

Introduction

Persephone Giannopoulos¹ and representatives of the AACB NSW/ACT QC Subcommittee. ¹Department of Chemical Pathology, St Vincent’s Hospital, Sydney, [email protected]

With the introduction of the Roche high sensitive Troponin T assay, it allows the measurement of Troponin at very low concentrations of the analyte.

One initial challenge was the lack of commercial quality control (QC) and external quality assessment (EQA) material at suitable concentrations for the new assay.

Specifically RCPA QAP and value assigned commercial QC materials were not available at low concentrations.

The control and assessment of assay performance near the 99th centile of normal (14ng/L) is important, as vital clinical decisions are made using these results.

Method validations had given our laboratories initial confidence to commission the assay however there were some reservations about testing commutability of patient sample precision at the low end of the assay over time.

AACB Annual Scientific Meeting, Sydney 2011

Aim

Tables and Figures

Conclusions

Graph 4: Repeat Analysis of human serum pool samples on the 2 separate occasions

REFERENCES(1) Giannitsis et al. Clin Chem: 2010;56:254 -261.

The aim of the study was to determine between laboratory assay precision of the TnT hs assay using two pooled serum samples at low levels.

Discussion

Table 1 & Graph 1: Precision for all samples, human serum pools 1 and 2, Roche QCs , Biorad QCs and QAP material*(QAP data for all TnT hs participants)

Graph 2: Box and Whisker plots of sample pool values (1) and (2) for September 2010 data, (---) lines are the mean values.

Graph 3: Box and Whisker plots of sample pool values (1) and (2) for February 2011 data, (---) lines are the mean values.

Roche Troponin T hs (high sensitive)* assay (05092744 190) - 200 tests reagent (E170, Cobas e 411, cobas e 601, cobas e 602).

*Note: Enhancements to 4th generation by:

1. Detection antibody genetically reengineered.

2. Analytical sensitivity improved by increasing sample volume, increasing ruthenium concentration of detection antibody and lowering background signal via buffer optimization.(1)

Two pooled patient samples were distributed to 9 laboratories on two occasions September 2010 and February 2011.

Internal QCs from Roche Precicontrol 1 and 2 (#05095107190) and Biorad Cardiac Markers Control LT (Cat # 646, 647) were run according to the laboratory protocols.

The General Serum Chemistry RCPA QAP samples for Troponin T hs were in use for routine laboratory monitoring of the assay.

QAP data was obtained from the RCPA QAP website and was used with permission.

AACB NSW/ACT QC Sub Committee members

Excellent precision was obtained for all the human serum pools from all nine laboratories. The September precision for serum pool 1 appeared to be slightly better than the February data. There were no differences with the CVs for serum pool 2.

The Roche QC data gave consistently low CVs for the two measurements ( Sept and Feb).

The Biorad QC data gave consistent CVs with the two measurements. The Biorad QC1 level measured lower than the Roche low level.

The QAP precision was very good in September (n=27 laboratories), the February (n=61 laboratories) CVs poorer in February.

The human serum pools for this survey were at very low levels giving us a good indication of the between-laboratory and between instrument performance of the TnT hs assay at these clinically important low concentrations.

The precision of the commercial controls tested was excellent. Most of the commercial controls do not measure as low as the human serum pool 1.

The repeat analysis of the human sample pools also provided within laboratory precision for each laboratory. Laboratory 3 and 4 did not perform as well as the other laboratories with human serum pool 1 when compared to the other laboratories, however this level is at the LOD of the assay.

The QAP data gives us some confidence unfortunately troponin levels of the QAP material with the lowest concentration are still not at the critical and clinical decision making concentrations (i.e.14ng/L).

These survey results have given us confidence that the between laboratory precision for low human serum values using the Roche TnT hs assay is appropriate for clinical decision making.

We will continue to distribute human serum pools at these low concentrations to ensure ongoing performance of the Roche TnT hs assay through the AACB NSW/ACT QC Sub committee.

Acknowledgements:

Thank you to the AACB for their financial support, to all participating laboratories and to Jadwiga Francis for the sample preparations and distributions.

Troponin T hs - PRECISION TABLESept 2010 (average;n=9) Feb 2011 (average;n=9)

TnT hs (ng/L) CV% TnT hs (ng/L) CV%Serum Pool 1 6.5 15.3 6.4 19.3Serum Pool 2 18.7 4.4 21.6 4.6Roche QC 1 24.3 4.4 24.5 4Roche QC 2 2142 2 2083 2Biorad CM 1 10.3 9.7 10.6 8.1Biorad CM 2 52.3 6.4 53.6 5QAP 85-13 42* 5QAP 86-02 50* 9

September 2010 data

1 2

Tro

pon

in T

con

cen

tra

tion

(ng/

L)

0

5

10

15

20

25

TnT hs (1) TnT hs (2)

February 2011 data

Tro

poni

n T

con

cent

ratio

n (

ng/L

)

0

5

10

15

20

25

TnT hs (1) TnT hs (2)