A Temperature-Pressure Calibration Grid for Multiauvil ...
Transcript of A Temperature-Pressure Calibration Grid for Multiauvil ...
Rev. High Pressure Sci. Technol., Vol. 7 (1998) 9•`11
A Temperature-Pressure Calibration Grid for Multiauvil Experriments Based on
Phase Relations in the System CaO-MgO-SiO2
T. GasparikCenter for High Pressure Research and Department of Geosciences, State University of New York at Stony Brook,
Stony Brook, New York 11794, USA
The subsolidus phase boundaries and melting curves in the system CaO-MgO-SiO2 produce a grid of sufficient
density and coverage at temperatures up to 2600•Ž and pressures up to 25 GPa to be suitable as a calibration
grid for multianvil experiments. The relative positions of these phase boundaries have been determined with high
precision in high-pressures experiments using the same multianvil apparatus, sample assembly, experimental
procedures and calibrations to maximize the internal consistency of the results. The grid can be easily adjusted
in the future to reflect a consensus based on new advances in our understanding of the absolute temperatures and
pressures for selected calibration points.
[calibration, phase relations, high-pressure experiments, multianvil press, system CaO-MgO-SiO2]
1. Introduction
In experiments conducted with solid-media high-pres-
sure devices, the relationships between the reading of the
thermocouple emf and the sample temperature and between
the reading of the pressure gauge and the sample pressure are
unique for every apparatus and sample assembly, and have
to be calibrated. A good calibration should assure that
the experimental temperatures and pressures are independent
of the experimental conditions, including the apparatus,
sample assembly, laboratory or the scientist. This can be
achieved if the same calibration grid is used in all high-
pressure laboratories, and the density of the calibration points
and the coverage of the entire temperature-pressure space
explorable in the experiments is sufficient for carrying out
a detailed calibration. A grid based on phase relations in the
system CaO-MgO-SiO2 appears to be particularly suitable for
calibration. The grid was determined in a large number of
experiments over an extended period of time using the
same multianvil apparatus, sample assembly, experimental
procedures and calibrations to maximize the internal consis
tency of the data. This emphasis allowed to locate, with high
precision, the relative positions of many melting curves and
subsolidus phase boundaries, thus producing a grid with suffi
cient density and coverage at temperatures up to 2600•Ž and
pressures up to 25 GPa.
2. Multianvil Apparatus and Sample Assembly
The experiments were carried out using the split-sphere an-
vil apparatus at the Stony Brook High Pressure Laboratory
(USSA-2000). The apparatus consists of a 2000-ton uniaxial
press with a two-stage anvil system capable of achieving pres
sures up to 25 GPa at temperatures in excess of 2600•Ž . The
first stage is a steel sphere split in six parts, which enclose a
cubic cavity holding the second stage. The second stage con
sists of 8 tungsten carbide cubes separated by pyrophyllite gas
kets, teflon back-up gaskets and balsa wood spacers. The cubes
are truncated at the corners and enclose an octahedral cavity
which holds the sample assembly. The sample assembly was
made of an MgO octahedron with the edge length of 10 mm .
A lanthanum chromite sleeve was used as the resistance heater.
Temperature was measured by a W3%Re vs. W25% Re ther
mocouple introduced axially. Additional details were re-
ported in [1]. This assembly was used in two configurations:
the 10/5 assembly for the experiments at 7-16.5 GPa [1], and
the 10/4 assembly for 16-23 GPa [2]. The 10/5 assembly
uses Kennametal Grade K313 tungsten carbide cubes with 5
mm truncation edge length and pyrophyllite gaskets 2.4 •~ 3.5
mm in cross section, while the 10/4 assembly uses Toshiba
Tungaloy Grade F tungsten carbide cubes with 4 mm trunca
tion edge length and gaskets 3.4 •~ 3.4 mm.
3. Temperature calibration
The relationship between the thermocouple electromotive
force (emf) and the measured temperature is based on the 1
bar calibration provided by the manufacturer (Engelhard).
This calibration is typically performed to 1800•Ž and in
cludes extrapolation to 2300•Ž. Further extrapolation can be
carried out by the user for experiments at higher temperatures.
The pressure effect on the thermocouple emf is not known
and, therefore, not included. The temperature (emf) is con-
trolled during an experiment by a Eurotherm controller, which
has a limit set by the manufacturer at 39.99 mV, approxi-
mately corresponding to 2380•Ž. Experiments at higher tem-
peratures are rarely needed or performed, because the
temperatures in such experiments are not controlled and are
based on extrapolations far beyond the calibrated range. An
electronic ice point is used at the cold junction. The contact
at the hot junction is provided by a rhenium sample capsule.
The, fluctuations around the set point of the temperature con-
troller are usually trivial with respect to other uncertainties.
The sample enclosed in a metal capsule is placed in the
assembly slightly off center, so that the hot spot is in the hot
end of the sample. Calibration of the temperature distribution
in the sample was performed at the nominal temperatures of
1400, 1600 and 1700•Ž using the compositions of two
coexisting pyroxenes on the join Mg2Si2O6,-CaMgSi2O6 [1]. It
was found that the temperature controlled by the thermocouple
10
Fig. 1. Schematic cross-section of an experimental charge in
a rhenium capsule. Solid isotherms show the tempera-ture distribution in a standard-size sample, dashed rectangle
outlines the distribution in a reduced-size sample. After [9].
(nominal T) was approximately the same as the temperature
in the center of the sample. The temperature in the hot spot
was about 50•Ž higher than the nominal temperature, and it
was 150•Ž lower at the cold end. The resulting temperature
distribution shown in Fig. 1 can be used to estimate the tem-
perature at various levels in the sample. Due to the low
temperature gradient, the temperature estimates around the
hot spot are more reliable than the estimates from the cold
end. Hence, for example, the best constraints on melting
temperatures are obtained from experiments showing minimal
melting. The reproducibility of the experimental temperatures
around the hot spot was estimated at •}30•Ž [1]. In some ex-
periments, particularly with Na or H2O, chemical gradients can
develop in response to the temperature gradient [3,4]. If this
is deemed unsuitable with respect to the intended goals, the
simplest solution is to decrease the size of the sample, as i
ndicated in Fig. 1, thus reducing the total temperature
range in the sample from 200•Ž to 50•Ž or less .
4. Pressure Calibration
Phase transitions in metals (Bi, Ba, Pb) or semiconductors
(ZnTe, ZnS, GaAs, GaP) are commonly used for the pres-
sure calibration at room temperature . However, such calibra-
tion is of a limited use in the high-temperature experiments,
because the assemblies are different and the temperature ef-
fect on the sample pressures at high temperatures can be a
complex function of temperature with little apparent relation
to the room temperature calibration. The pressure calibration
of the 10/5 assembly at high temperatures (Fig. 2) is based
on 3 univariant transitions [1]. Reversals of the coesite to
stishovite transformation were carried out at 1200•Ž (9 .3
GPa) and 1400•Ž (9.5 GPa). Although, the original calibration
by [5] turned out to be incorrect and was replaced later by
[6], the change in the investigated temperature range was
Fig. 2. Pressure calibration of the 10 mm assembly at high
temperatures in the 10/5 and 10/4 configurations. After [1, 2].
negligible. The transition from forsterite to beta phase
(Mg2SiO4) was used at 1400•Ž (14.3 GPa) and 1600•Ž (15.2
GPa), based on [7]. The transition from enstatite (MgSiO) to
beta phase and stishovite was observed at 1400•Ž, and asigned
to 16.5 GPa according to [8]. The pressure calibration of the
10/4 assembly (Fig. 2) was anchored at 16.6 GPa by the
observation of the transition from enstatite to beta phase
and stishovite at 1500•Ž [2]. The formation of MgSiO,
perovskite from majorite was observed at 2000•Ž and later
asigned to 21.5 GPa [9]. The transition from beta phase to
perovskite and periclase was observed at 2000•Ž and the cor-
responding reading of the pressure gauge was asigned to 22.4
GPa [2]. Later experiments [10] indicated that the formation
of perovskite from ilmenite at 1800•Ž occurred at about the
same gauge pressure as previously found at 2000°C [2], de-
spite the well documented negative slope of the reaction [11].
Hence, the efficiency of the pressure generation decreased with
temperature between 1800 and 2000•Ž. It was assumed that
this temperature dependence of the pressure calibration was
limited only to the maximum pressures and the pressure cali-
bration was independent of temperature at lower pressures.
The reproducibility of the experimental pressures was esti-
mated at •}0.3 GPa [1].
5. Calibration grid based on the system CaO-MgO-SiO2
Fig. 3 shows the proposed calibration grid based on the phase
relations in the system CaO-MgO-SiO2. They were determined
during a systematic investigation of the phase relations relevant
to the Earth's mantle, which is the primary focus of research
for many multianvil experiments. The most important among
the simple chemical systems for the Earth's mantle is the sys-
11
Fig. 3. Temperature-pressure phase diagram showing the pro-posed calibration grid. Symbols: Bt, Mg2SiO4 beta phase; CaPv, C
aSiO3 perovskite; CEn, clinoenstatite; Cpx, clinopyroxene; Cs, coesite; Di, diopside; Fo, forsterite; Ga, garnet; Ii, MgSiO3 ilmenite; L, liquid; MgPv, MgSiO3 perovskite; Mj3, majorite; OEn, orthoenstatite; Opx, orthopyroxene; Pc, periclase; Pv, perovskite; Sp, Mg2SiO4 spinel; St, stishovite.
tem MgO-SiO2, which constitutes over 80% of the mantle composition. Building on the early pioneering works [7, 8, 11], the phase diagram for the system MgO-SiO2 was finalized in a series of experimental studies [1, 2, 9,12-16]'. The recom-mended temperature-pressure conditions for selected univariant
points are summarized in Table 1. While the phase relations in the system MgO-SiO2 have been used for calibration in most multianvil studies, the addition of CaO increases the com-
plexity of the calibration grid to allow calibration to a much greater detail. The phase diagram for the CaSiO3 system was reported by [17], and the experimental study of the join Mg2Si2O6 CaMgSi2O6 [9] completed the phase diagram for the system CaO-MgO-SiO2. The resulting calibration grid (Fig. 3) can be constructed using the parameters and information summarized in [9]. For the calibration at lower temperatures, the grid can be improved by including the melting curves of Mg(OH)2 [10], Na2Mg2Si2O, and Na2MgSiO4 [18]. Additional phase relations fully consistent with the calibration grid in Fig. 3 were reported by [19-21]. The proposed calibration
grid can be easily adjusted in the future to reflect a consen-sus based on new advances in our understanding and knowl-edge of the absolute temperatures and pressures for selected
calibration points. Such absolute calibration points could result from the correct identification of the phase transitions responsible for the discontinuities in the seismic velocity pro-files of the Earth's mantle.
Acknowledgements The high-pressure experiments reported in this paper were
performed in the Stony Brook High Pressure Laboratory which is jointly supported by the National Science Foundation Sci-ence and Technology Center for High Pressure Research (EAR 89-20239) and the State University of New York at Stony Brook.
References[1] T. Gasparik, Contrib. Mineral. Petrol., 102, 389 (1989).[2] T. Gasparik, J. Geophys. Res., 95, 15751 (1990).[3] C. E. Lesher, D. Walker, J Geophys. Res., 93, 10295 (1988).[4] T. Gasparik, M.1. Drake, Earth Planet. Sci. Lett.,134, 307 (1995).[5] T. Yagi, S. Akimoto, Tectonophysics, 35, 259 (1976).[6] J. Zhang, B. Li, W. Utsumi, R. C. Liebermann, Phys. Chem. Minerals, 23,1(1996).[7] T. Katsura, E. Ito, J. Geophys. Res., 94, 15663 (1989).[8] H. Sawamoto, In: M. H. Manghnani, Y. Syono (eds.), High-Pressure Research in Mineral Physics, Terrapub., Tokyo, 209 (1987).[9] T. Gasparik, Contrib. Mineral. Petrol., 124, 139 (1996).[10] T. Gasparik, J. Geophys. Res., 98,4287 (1993).[11] E. Ito, E. Takahashi, J. Geophys. Res., 94, 10637 (1989).[12] R. E. G. Pacalo, T. Gasparik, J. Geophys. Res., 95, 15858 (1990).[13] D. C. Presnall, T. Gasparik, J. Geophys. Res., 95, 15771 (1990).[14] T. Gasparik, J. Geophys. Res., 97, 15181 (1992).[15] D. C. Presnall, M. J. Walter, J. Geophys. Res., 98, 19777 (1993).[16] J. Zhang, R. C. Liebermann, T Gasparik, C. T. Herzberg, J. Geophys. Res., 98, 19785 (1993).[17] T. Gasparik, K. Wolf, C. M. Smith, Am. Mineral., 79, 1219 (1994).[18] T. Gasparik, Yu. A. Litvin, Eur. J. Mineral., 9, 311(1997).[19] T. Gasparik, Contrib. Mineral. Petrol., 111, 283 (1992).[20] T. Gasparik, High Temp. High Pres., 25, 245 (1993).[21] T. Gasparik, Phys. Chem. Minerals., 23, 476 (1996).