A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering- energy transfer analytical platform for detection of iodide and H 2 O 2 Yaohui Wang, Xinghui Zhang, Qingye Liu, Guiqing Wen, Aihui Liang * , Zhiliang Jiang * (Key Laboratory of Ecology of Rare and Endangered Species and Environmental Protection of Ministry Education, Guangxi Key Laboratory of Environmental Pollution Control Theory and Technology, Guangxi Normal University, Guilin 541004, China.) 0 0.4 0.8 1.2 1.6 200 300 400 500 600 700 Wavelength(nm) Absorbance(a.u.) Fig. 1S Absorption spectra of the GO/GN-KI-KIO 3 system (a) pH 2.8 Na 2 HPO 4 -citric acid-4mmol/LKI-4.78μg/mL GO/GN; (b) a+5μmol/L KIO 3 ; (c) a+10μmol/L KIO 3 ; (d) a+20μmol/L KIO 3 ; (e) a+25 μmol/L KIO 3 . e d c b a Electronic Supplementary Material (ESI) for RSC Advances. This journal is © The Royal Society of Chemistry 2014

Transcript of A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold...

Page 1: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide and H2O2

Yaohui Wang, Xinghui Zhang, Qingye Liu, Guiqing Wen, Aihui Liang*, Zhiliang Jiang*

(Key Laboratory of Ecology of Rare and Endangered Species and Environmental Protection of Ministry Education, Guangxi Key Laboratory of Environmental Pollution Control Theory and

Technology, Guangxi Normal University, Guilin 541004, China.)

0

0.4

0.8

1.2

1.6

200 300 400 500 600 700

Wavelength(nm)

Absorbance(a.u.)

Fig. 1S Absorption spectra of the GO/GN-KI-KIO3 system(a) pH 2.8 Na2HPO4-citric acid-4mmol/LKI-4.78μg/mL GO/GN; (b) a+5μmol/L KIO3; (c)

a+10μmol/L KIO3; (d) a+20μmol/L KIO3; (e) a+25 μmol/L KIO3.

e

d

c

b

a

Electronic Supplementary Material (ESI) for RSC Advances.This journal is © The Royal Society of Chemistry 2014

Page 2: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

0

0.2

0.4

0.6

0.8

1

1.2

1.4

200 300 400 500 600 700

Wavelength(nm)

Absorbance(a.u.)

Fig. 2S Absorption spectra of the GN-KI-KIO3 system(a)pH 2.8 Na2HPO4- citric acid-4mmol/L KI-4.782µg/ml GN; (b) a+15μmol/L KIO3;

(c) a+20μmol/L KIO3; (d) a+25μmol/L KIO3.

0

0.2

0.4

0.6

0.8

1

200 300 400 500 600 700

Wavelength(nm)

Absorbance(a.u.)

Fig. 3S Absorption spectra of the GO-KI-KIO3 system

d

c

b

a

ed

c

b

a

Page 3: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

(a)pH 2.8 Na2HPO4-citric acid-4mmol/L KI-15μg/L GO; (b) a+5.5μmol/L KIO3; (c)a+7μmol/L KIO3; (d)a+8.5μmol/L KIO3; (e)a+10μmol/L KIO3.

0

500

1000

1500

2000

2500

3000

3500

200 300 400 500 600 700

Wavelength(nm)

Intensity(

a.u.)

Fig. 4S RRS spectra of the GN-KI-KIO3 system(a)pH 2.8 Na2HPO4- citric acid-4mmol/L KI-4.782µg/ml GN; (b)a+0.25μmol/L

KIO3;(c) a+1μmol/L KIO3; (d) a+2μmol/L KIO3; (e) a+2.5μmol/L KIO3;(f) a+4μmol/L KIO3; (g) a+4.5μmol/L KIO3;(h)a+5μmol/L KIO3.

0

1000

2000

3000

4000

5000

200 300 400 500 600 700Wavelength(nm)

Inte

nsity

(a.

u.)

a

h

a

i

Page 4: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

Fig. 5S RRS of the GN-KI-H2O2 system(a) pH 3.4 Na2HPO4-citric acid-3mmol/L KI-4.782µg/ml GN; (b)a+10μmol/L H2O2;(c)

a+20μmol/L H2O2; (d) a+30μmol/L H2O2; (e) a+40μmol/L H2O2; (f) a+50μmol/L H2O2; (g) a+60μmol/L H2O2; (h)a+70μmol/L H2O2; (i)a+80μmol/L H2O2

0

500

1000

1500

2000

2500

3000

3500

200 300 400 500 600 700

Fig. 6S RRS of the GN-KI-H2O2-Fe3+ catalytic system(a)pH 2.8 Na2HPO4- citric acid-4mmol/L KI-25mmol/L H2O2-Fe3+-4.782µg/ml GN; (b)

a+0.025mmol/L Fe3+; (c) a+0.1mmol/L Fe3+; (d) a+0.25mmol/L Fe3+; (e) a+0.5mmol/L Fe3+; (f) a+1mmol/L Fe3+.

a

f

Page 5: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

0

500

1000

1500

2000

200 300 400 500 600 700

Wavelength(nm)

Inte

nsity

(a.

u.)

Fig. 7S RRS spectra of the GO-KI-KIO3 system(a) pH 2.8 Na2HPO4-citric acid-4mmol/L KI-15μg/L GO; (b)a+0. 5μmol/L KIO3; (c) a+1μmol/L

KIO3; (d) a+2.5μmol/L KIO3; (e) a+7μmol/L KIO3; (f) a+10μmol/L KIO3.

0

50

100

150

200

250

100 300 500 700 900 1100 1300 1500 1700

Raman shift (cm-1)

Raman Intensity(

cps)

Fig. 8SA SERS spectra of KIO3-KI-GN systema: pH 2.8Na2HPO4-citric acid-4mmol/L KI-1.0μmol/L VBB-12μg/mL GN; b:

a

e

ab

c

d

e

f

0

400

800

1200

1600

0 5 10c(µmol/L)

ΔI

Page 6: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

a+1μmol/L KIO3; c: a+2.0μmol/L KIO3; d: a+3.0μmol/L KIO3 ; e: a+4.0μmol/L KIO3.

0

50

100

150

100 300 500 700 900 1100 1300 1500 1700

Raman shift (cm-1)

Raman Intensity(

cps)

Fig. 8SB SERS spectra of KIO3-KI-GO/GN systema: pH 2.8Na2HPO4-citric acid-4mmol/L KI-1.0μmol/L VBB-12μg/mL GO/GN; b:

a+0.25μmol/L KIO3; c: a+0.5μmol/L KIO3 ; d: a+1μmol/L KIO3; e: a+2.5μmol/L KIO3; f: a+4.0μmol/L KIO3; g: a+5.5μmol/L KIO3.

0

5

10

15

0.1 1 10 100 1000 10000

Inten

sity (

Perce

nt)

Size (d.nm)

Size Distribution by Intensity

Record 1139: 3 1

0

5

10

15

20

0.1 1 10 100 1000 10000

Inte

nsity

(Per

cent

)

Size (d.nm)

Size Distribution by Intensity

Record 125: 0-1 1

a

b

g

a

Page 7: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

0

5

10

15

20

25

0.1 1 10 100 1000 10000

Inte

nsity

(Per

cent

)

Size (d.nm)

Size Distribution by Intensity

Record 1141: 4-1 1

Figure 9S Laser scattering graph of the NG/GO systemsa: GO; b: GO/GN; c: GO/GN-2.8 Na2HPO4-citric acid-4 mmol/L KI-2.5μmol/L KIO3.

0

400

800

1200

1600

2 2.5 3 3.5 4 4.5 5

pH

ΔI

Fig. 10S Effect of pH Na2HPO4-citric acid buffer 3 mmol/L KI-2.5 μmol/L KIO3-7.17μg/mL GN

c

Page 8: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

0

400

800

1200

1600

2000

0 1 2 3 4 5 6 7

c(mmol/L)

ΔI

Fig. 11S Effect of KI concentrationpH 2.8Na2HPO4-citric acid buffer-2.5μmol/L KIO3-7.173μg/mL GN

0

500

1000

1500

2000

2500

3000

0 2 4 6 8 10 12 14

c(μg/mL)

ΔI

Fig. 12S Effect of GN concentrationpH 2.8Na2HPO4-citric acid buffer-4mmol/L KI-2.5μmol/L KIO3

Page 9: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

0

200

400

600

800

1000

1200

0 5 10 15 20 25 30 35

c(μg/L)

ΔI

Fig. 13S Effect of GO concentrationpH 2.8Na2HPO4-citric acid buffer-4mmol/L KI-2.5μmol/L KIO3.

0

500

1000

1500

2000

2500

3000

0 2 4 6 8 10 12 14c(μg/mL)

ΔI

Fig. 14S Effect of GO/GN concentrationpH 2.8Na2HPO4-citric acid buffer-4mmol/L KI-2.5μmol/L KIO3

Page 10: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

0

500

1000

1500

2000

2500

2 2.5 3 3.5 4 4.5 5

pH

ΔI

Fig. 15S Effect of pH Na2HPO4-citric acid buffer 3mmol/L KI-50μmol/L H2O2-7.173 μg/mL GN-50mmol/L NaCl

0

500

1000

1500

2000

2500

3000

0 1 2 3 4 5 6

c(mmol/L)

ΔI

Fig. 16S Effect of KI concentrationpH3.4Na2HPO4-citric acid buffer-KI-50μmol/L H2O2-7.173μg/mL GN-50mmol/L

NaCl

Page 11: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

0

500

1000

1500

2000

2500

3000

3500

0 5 10 15

c(μg/mL)

ΔI

Fig. 17S Effect of GN concentrationpH3.4Na2HPO4-citric acid buffer-3 mmol/L KI-50μmol/L H2O2- GN-50mmol/L NaCl

0

500

1000

1500

2000

2500

20 40 60 80 100

Fig. 18S Effect of temperaturepH 3.4Na2HPO4-citric acid-3 mmol/L KI-50 μmol/L H2O2-4.782μg/mL GN.

0

400

800

1200

1600

2000

0 5 10 15 20 25 30

ΔIΔI

Page 12: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

Fig. 19S Effect of reaction timepH 3.4 Na2HPO4-citric acid-3 mmol/L KI-50μmol/L H2O2-4.782μg/mL GN.

Table 1S The reproducibility for some nanoparticle systems

System Single value I Average RSD (%)

4.78µg/ml GN197,210,235,201,1

94207 8.0

pH 2.8 Na2HPO4-citric acid-4mmol/L KI-4.78µg/ml GN

4365,3918,3537,3343,3803

3793 10.3

15μg/L GO810,887,875,800,8

23839 4.7

pH 2.8 Na2HPO4-citric acid-4mmol/L KI-15μg/L GO

906,900,872,917,869

893 2.4

4.78μg/mL GO/GN1394,1425,1282,13

51,14921389 5.7

pH 2.8 Na2HPO4-citric acid-4mmol/LKI-4.78μg/mL GO/GN

1578,1727,1742,1607,1558

1642 5.2

Table 2S Comparison of the reported methods for I

Method Principle LR(μmol/L)

DL(μmol/L) Comments Ref.

SM *

Iodate reacts with iodide to form I3

- that combined with starch to form blue complex with an absorption peak at 580 nm.

1.6-7.0 0.8 Simple, but low sensitivity 37

ISE

Using standard addition method and iodine ISE, iodine ion in iodine fortified beverages were determined.

0.5-1.0×105 0.63

Simple and automatic, but low sensitivity

38

IC-ICP/MS

Alkaline extraction and IC-ICP /MS were applied as the sample pretreat mentmethod and the detection technique respectively, for iodate and iodide determ ination.

0.002-7.9 0.0008

Highly sensitive, but high-cost and

complex.

39

Sensor

The colorimetric iodide sensor was based on the anti-aggregation of GNs that relied upon the distance-dependent optical properties of GNs, the combination of mercapto-functionalized thymine on GNs, and the stronger affinity between I− and Hg2+.

0.02-0.6 0.01

Simple and sensitive, but Hg2+ used and

complex.

33

RRS

Iodate reacts with iodide to form I3

- that combined with rhodamine 6G to form ion- associated particles with a RRS peak at 400 nm.

0.1-2.0 0.03 Simple, but low sensitivity 35

Page 13: A simple and sensitive graphene oxide/nanogold SPR ...A simple and sensitive graphene oxide/nanogold SPR Rayleigh scattering-energy transfer analytical platform for detection of iodide

RRS-ET

The SPRRS energy of GO/GNs transferred to I3

- complex that the decreased RRS intensity is linear to IO3

- concentration.

0.025-5 0.008Simple, fast,

highly sensitive.

This paper

* SM-spectrophotometric method, ISE-ion selective electrode, IC-ICP/MS-ion chromatography-inductively

coupled plasma mass spectrometry.

Table 3S Effect of foreign substances on the determination of iodide

Coexistent substance

Tolerance Concentratio

n(µmol/L)

Relative error(%)

Coexistent substance

ToleranceConcentratio

n(µmol/L)

Relative error(%)

K+ 250 8.7 Ni2+ 250 3.0

Ca2+ 250 3.0 MoO42- 200 -0.9

Zn2+ 250 2.4 SO42- 250 -4.9

Glucose 200 -6.1 BrO3 - 250 -2.0

Co2+ 250 -5.4 Cu2+ 200 0.8

Al3+ 250 1.8 Mn2+ 250 7.5

Ba2+ 250 1.6 Cr3+ 12.5 -7.1

Mg2+ 250 1.1 SeO32- 12.5 -5.4

Table 4S Effect of foreign substances on the determination of H2O2

Coexistent substance

Tolerance Concentratio

n(µmol/L)

Relative error(%)

Coexistent substance

ToleranceConcentratio

n(µmol/L)

Relative error(%)

K+ 1000 2.0 Mg2+ 1000 0.7

Ca2+ 500 -8.9 SeO32- 2000 0.4

Zn2+ 1000 1.4 SO42- 1000 2.8

Glucose 500 2.4 BrO3 - 2500 3.6

BPO 250 6.1 Cu2+ 1000 3.1

Al3+ 250 -7.8 Mn2+ 75 8.6

Ba2+ 500 -8.7 Cr3+ 500 -7.6