A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS
PRODUCED BY THE HUMAN BODY &
DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE
![Page 2: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/2.jpg)
B.HEMA KUMAR
Under the guidance of
Mr.G.BALASUBRAMANIANSenior Lecturer,
Biomedical Signal Processing
&
Instrumentation
SASTRA Deemed university
![Page 3: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/3.jpg)
FES for heart sounds, lung sounds, intestinal sound analysis
Design of an electronic circuit to detect and amplify sounds from human body.
Comparison of electronic stethoscope with conventional stethoscope.
![Page 4: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/4.jpg)
SOFTWARE
Twenty volunteers and twenty patients in each category (for cardiac sounds, lung sounds and intestinal sounds) randomly selected with different age, sex and clinical symptom, who were referred to the outpatient department of a major hospital.
Auscultatory findings were given to the engineer for designing a FES for each category of cases.
Fuzzy expert system design was done using fuzzy logic toolbox of MATLAB 6.1
MATERIALS AND METHODS
![Page 5: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/5.jpg)
MATERIALS AND METHODS
HARDWARE
The electronic stethoscope was developed under the guidance of an electronic expert.
On a scale of three (1. Good 2. Better 3. Best) the doctors were asked to evaluate the overall performance of both the conventional and electronic stethoscope as per the questionnaire given to them.
![Page 6: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/6.jpg)
![Page 7: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/7.jpg)
STRUCTURE OF FUZZY EXPERT SYSTEM
SOUND ANALYSIS
(MAMDANI)
CONDITION
INPUT 1
INPUT 2
INPUT 3
INPUT 4
![Page 8: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/8.jpg)
HEART SOUNDS ANALYSIS
INPUT VARIABLES
Heart sound absence or presence (HSA, HSP)
Intensity (NOR, DEC, INC)
Number of heart sounds heard (L2S, M2S)
Presence or absence of added sounds (ASP,ASA)
Character of the heard sound (NOR, ABN)
OUTPUT VARIABLE
Patient condition (CDN)
![Page 9: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/9.jpg)
DISEASES
PERICARDIAL EFFUSION
LEFT VENTRICULAR FAILURE
AORTIC STENOSIS
MITRAL STENOSIS
SAMPLE RULE
If (HSP/A is HSP) and (INT is INC) and (CHAR is ABN) then (CDN is MS)
![Page 10: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/10.jpg)
CDN = AORTIC STENOSIS
![Page 11: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/11.jpg)
LUNGS SOUNDS ANALYSIS
INPUT VARIABLES
Breath sound absence or presence (BSA, BSP)
Normal vesicular or Bronchial (NORV, NORB)
Presence or absence of added sounds (ASP,ASA)
If added sounds present (rals or rhonchi) (ASRA, ASRO)
Nature of the heard sound ( tubular or cavernous ) (T, C)
OUTPUT VARIABLE
Patient condition (CDN)
![Page 12: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/12.jpg)
DISEASES
Lung Consolidation
Lung cavity
Lung Parenchymal Disease
Lung Bronchial Disease
SAMPLE RULE
If (BSP/A is BSP) and (ASA/P is ASP) and (ASRA/RO is ASRO) then (CDN is LBD)
![Page 13: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/13.jpg)
CDN =Lung bronchial disease
![Page 14: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/14.jpg)
INTESTINAL SOUNDS ANALYSIS
INPUT VARIABLES
Intestine sound absence or presence (ISA, ISP)
Intensity (NOR, DEC, INC)
Duration (SD, LD)
Site of the heard sound (UA, LA)
OUTPUT VARIABLE
Patient condition (CDN)
![Page 15: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/15.jpg)
DISEASES
PERITONITIS PERFORATION LARGE INTESTINAL OBSTRUCTION SMALL INTESTINAL OBSTRUCTION
SAMPLE RULE
If (ISA/P is ISP) and (INT is INC) and (DUR is LD) and (SITE is UA) then (CDN is SIO)
![Page 16: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/16.jpg)
CDN = Small intestinal obstruction
![Page 17: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/17.jpg)
![Page 18: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/18.jpg)
![Page 19: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/19.jpg)
![Page 20: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/20.jpg)
PREAMPLIFIER Amplify low level signals
BUFFER AMPLIFIER Second stage of amplification
FILTER CIRCUIT To suppress whistling sound due to high feedback
MAIN AMPLIFIER TBA 810 versatile audio amplifier
![Page 21: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/21.jpg)
![Page 22: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/22.jpg)
1 Intensity 1 2
2 Number 1 2
3 Opening snap 1 2
4 Murmur 1 2
5 Click 1 2
![Page 23: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/23.jpg)
1 Vesicular 1 2
2 Bronchial breath sound 1 2
3 Added sounds 1 2
1 Intensity 1 2
2 Character 1 2
![Page 24: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/24.jpg)
The paper describes a design of a fuzzy expert system for determination of the possibility of the diagnosis of the abnormal body sounds, which can be used by the doctors with hearing impairment and by the students for learning the sounds.
Developed FES for various body sounds were compared with the clinical data given by the doctor under different clinical settings
![Page 25: A FUZZY EXPERT SYSTEM DESIGN FOR ANALYSIS OF SOUNDS PRODUCED BY THE HUMAN BODY & DESIGN OF AN UNIQUE ELECTRONIC STETHOSCOPE.](https://reader036.fdocuments.us/reader036/viewer/2022081504/5697bfa21a28abf838c960a6/html5/thumbnails/25.jpg)
The electronic stethoscopes scored over conventional stethoscopes by its better audibility, quality etc.
A good CS cost about Rs 125/- whereas a good ES will cost any where between 7000 to 8000 (About seventy times) whereas ours -- 1500/-
One distinct advantage of ES is its extent of audibility of sounds for in a CS only the examining doctor can hear the sound.