A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp...

28
A comparison of A comparison of Spectro::lyser Spectro::lyser , YSI 6 , YSI 6 - - Series, Series, and and Hydrolab Hydrolab Sondes Sondes Is a network’s integrity compromised when using multiple equipment manufacturers; or is diversity acceptable?

Transcript of A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp...

Page 1: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

A comparison of A comparison of Spectro::lyserSpectro::lyser, YSI 6, YSI 6--Series, Series, and and HydrolabHydrolab

SondesSondes

Is a network’s integrity compromised when using multiple equipment manufacturers;

or is diversity acceptable?

Page 2: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

LearyLeary’’s Brook s Brook ––

Atypical Urban StreamAtypical Urban Stream

‘Flashy’ urban stream

Impermeable surface runoff

O’Leary Industrial Park

Avalon Mall Parking lot

Surrounding roads

Page 3: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

00.10.20.30.40.50.60.70.80.9

1

3-DColumn13-DColumn23-DColumn3

Specific Conductivity at Leary's Brook from November 2008 to January 2009

0

500

1000

1500

2000

2500

3000

3500

Sep 30 Oct 15 Oct 30 Nov 14 Nov 29 Dec 14 Dec 29

Spec

ific

Con

duct

ivity

(uS/

cm)

Storm runoff varies according to season

December 1st is the first day of Winter Shift• 18 hour days• More intensive road clearing• More salt

– Greater conductivity flux!

Page 4: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Newfoundland and LabradorNewfoundland and Labrador’’s Networks Network

Water Resources Management Division possesses:

Spectro::lysers

Hydrolabs

YSI 6-Series

Same destination, different routes

Firmware variation

Measurement methodology

Page 5: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Instrument ParametersInstrument Parameters

Hydrolab Datasonde 5x

Temperature

pH

Specific Conductivity

Total Dissolved Solids

Dissolved Oxygen

Turbidity

Phycocyanin

YSI 6-Series

Temperature

pH

Specific Conductivty

Total Dissolved Solids

Dissolved Oxygen

Turbidity

Phycocyanin

Page 6: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Instrument ParametersInstrument Parameters

S::can Spectro::lyser

Turbidity

Nitrate

Total Organic Carbon

Dissolved Organic Carbon

Grab Samples

pH

Specific Conductivity

Turbidity

Nitrate

Total Organic Carbon

Dissolved Organic Carbon

Page 7: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

TemperatureTemperature

Hydrolab_Temp YSI_TempHydrolab_Temp 1YSI_Temp 0.999142354 1

Temperature at Leary's Brook from March to June, 2009

0

5

10

15

20

25

Mar-18 Mar-28 Apr-07 Apr-17 Apr-27 May-07 May-17 May-27 Jun-06 Jun-16

Tem

pera

ture

(oC

)

Hydrolab (oC) YSI (oC)

Page 8: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

YSI_TempHydrolab_Temp

17.5

15.0

12.5

10.0

7.5

5.0

Dat

aBoxplot of Hydrolab_Temp, YSI_Temp

TemperatureTemperature

Page 9: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

pHpH

pH at Leary's Brook from March to June, 2009

0

2

4

6

8

10

12

14

16

Mar-08 Mar-18 Mar-28 Apr-07 Apr-17 Apr-27 May-07 May-17 May-27 Jun-06 Jun-16

pH

Hydrolab YSI Grab Sample

Page 10: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

YSI_pHHydrolab_pH

15

14

13

12

11

10

9

8

7

6

Instrument

pH

Boxplot of pH

pHpH

Page 11: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

pHpH

pH_grab Sample YSI_pH Error Square Mean RMSE6.59 6.57 0.02 0.00046.33 6.36 0.03 0.00096.46 6.51 0.05 0.00256.56 6.68 0.12 0.01446.47 6.63 0.16 0.02566.65 6.68 0.03 0.0009

pH_grab Sample Hydrolab_pH Error Square Mean RMSE6.59 6.78 0.19 0.03616.33 6.67 0.34 0.11566.46 6.75 0.29 0.08416.56 6.6 0.04 0.00166.47 7.78 1.31 1.71616.65 6.74 0.09 0.0081

0.5720.326933

0.00745 0.086

Page 12: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Specific ConductivitySpecific Conductivity

Specific Conductivity at Leary's Brook from March to June, 2009

0

1000

2000

3000

4000

5000

6000

7000

Mar-08 Mar-18 Mar-28 Apr-07 Apr-17 Apr-27 May-07 May-17 May-27 Jun-06 Jun-16

Spec

ific

Con

duct

ivity

(uS/

cm)

Hydrolab (uS/cm) YSI (uS/cm) Grab (uS/cm)

Page 13: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Specific ConductivitySpecific Conductivity

YSI_SpCon (uS/cm)Hydrolab_SpCon (uS/cm)

6000

5000

4000

3000

2000

1000

0

Instruments

Sp

Co

n (

uS

/cm

)Boxplot of SpCon (uS/cm)

Page 14: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

SpCon (uS/cm)_Grab Hydrolab_SpConError Square Mean RMSE700 546 154 23716.00310 269 41 1681.00470 398 72 5184.00500 482 18 324.00550 547 3 9.00580 593 13 169.00

SpCon (uS/cm)_Grab YSI_SpCon Error Square Mean RMSE700 639 61 3721.00310 294 16 256.00470 436 34 1156.00500 484 16 256.00550 548 2 4.00580 586 6 36.00

71.98

30.08904.83

5180.50

Specific ConductivitySpecific Conductivity

Page 15: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Total Dissolved SolidsTotal Dissolved Solids

Total Dissolved Solids at Leary's Brook from March to June, 2009

0

0.5

1

1.5

2

2.5

3

3.5

4

Mar-08 Mar-18 Mar-28 Apr-07 Apr-17 Apr-27 May-07 May-17 May-27 Jun-06 Jun-16

Tota

l Dis

solv

ed S

olid

s (m

g/l)

Hydrolab (mg/l) YSI (mg/l) Grab (mg/l)

Page 16: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Total Dissolved SolidsTotal Dissolved Solids

TDS (g/l)_YSITDS (g/l)_Hydrolab

4

3

2

1

0

Instruments

TDS

(g

/l)

Boxplot of TDS (g/l)

Page 17: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Total Dissolved SolidsTotal Dissolved Solids

TDS (g/l)_Grab Hydrolab_TDS Error Square Mean RMSE0.348 0.362 0.014 0.0001960.161 0.171 0.01 0.00010.231 0.255 0.024 0.0005760.2240.259 0.35 0.091 0.0082810.3 0.377 0.077 0.005929

TDS (g/l)_Grab YSI_TDS Error Square Mean RMSE0.348 0.416 0.068 0.0046240.161 0.191 0.03 0.00090.231 0.284 0.053 0.0028090.2240.259 0.356 0.097 0.0094090.3 0.381 0.081 0.006561

0.069718

0.054921760.003016

0.004861

Page 18: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

YSI DO mg/l Hydrolab DO mg/lYSI DO mg/l 1Hydrolab DO mg/l 0.996421334 1

Dissolved OxygenDissolved Oxygen

Dissolved Oxygen at Leary's Brook from March to June, 2009

0

2

4

6

8

10

12

14

16

Mar-18 Mar-28 Apr-07 Apr-17 Apr-27 May-07 May-17 May-27 Jun-06 Jun-16

Dis

solv

ed O

xyge

n (m

g/l)

Hydrolab (mg/l) YSI (mg/l)

Page 19: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Dissolved OxygenDissolved Oxygen

YSI_ODO (mg/l)Hydrolab_LDO (mg/l)

15

14

13

12

11

10

9

Instrument

DO

(m

g/l

)

Boxplot of DO (mg/l)

Page 20: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

TurbidityTurbidity

Turbidity at Leary's Brook from March to June, 2009

0

500

1000

1500

2000

2500

3000

Mar-08 Mar-18 Mar-28 Apr-07 Apr-17 Apr-27 May-07 May-17 May-27 Jun-06 Jun-16

Turb

idity

S::can (ftu) Hydrolab (ntu) YSI (ntu) Grab Sample (ntu)

Page 21: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

YSI_Turb(ntu)Spectro_Turb(ftu)Hydro_Turb(ntu)

3000

2500

2000

1500

1000

500

0

Instrument

Turb

idit

y

Boxplot of Turbidity

TurbidityTurbidity

Page 22: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

TurbidityTurbidityTurbidity_Grab Sample (ntu) Turbidity_Hydrolab (ntu) Error Square Mean RMSE

0.10.80.6 4.79 4.19 17.560.6 6.92 6.32 39.940.20.4 0 0.4 0.160.7 0 0.7 0.490.1 0 0.1 0.01

Turbidity_Grab Sample (ntu) Turbidity_Spectrolyser (ftu) Error Square Mean RMSE0.10.80.6 8.403 7.803 60.890.6 1.667 1.067 1.140.20.4 1.68 1.28 1.640.7 1.281 0.581 0.340.1 0.66 0.56 0.31

Turbidity_Grab Sample (ntu) Turbidity_YSI (ntu) Error Square Mean RMSE0.10.80.6 5.1 4.5 20.250.6 1.7 1.1 1.210.20.4 2 1.6 2.560.7 2 1.3 1.690.1 1.2 1.1 1.21

5.38

11.63

12.86

2.32

3.59

3.41

Page 23: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Grab_NO3 (mg/l) Spectro::lyser_NO3 Error Square Mean RMSE0.36 0.166 0.194 0.0376360.38 0.186 0.194 0.0376360.33 0.076 0.254 0.0645160.16 0 0.16 0.02560.26 0.096 0.164 0.0268960.21 0.057 0.153 0.0234090.28 0.196 0.084 0.0070560.32 0.21 0.11 0.0121

0.029356 0.171

NitrateNitrateNitrate at Leary's Brook from March to June, 2009

0

0.2

0.4

0.6

0.8

1

1.2

1.4

Mar-08 Mar-18 Mar-28 Apr-07 Apr-17 Apr-27 May-07 May-17 May-27 Jun-06 Jun-16

Nitr

ate

(mg/

l)

Spectro::lyser (mg/l) Grab Samples (mg/l)

Page 24: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Total Organic CarbonTotal Organic Carbon

Grab_TOC (mg/l) Spectro_TOC Error Square Mean RMSE1.8 1.76 0.04 0.0016

1.5 4.686 3.186 10.15063 1.185 1.815 3.294225

1.9 2.81 0.91 0.82811.6 3.121 1.521 2.3134412.1 5.449 3.349 11.21581.9 2.59 0.69 0.4761

4.03998 2.01

Total Organic Carbon at Leary's Brook from march to June, 2009

0

5

10

15

20

25

30

Mar-08 Mar-18 Mar-28 Apr-07 Apr-17 Apr-27 May-07 May-17 May-27 Jun-06 Jun-16

Tota

l Org

anic

Car

bon

(mg/

l)

Spectro::lyser (mg/l) Grab (mg/l)

Page 25: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

Dissolved Organic CarbonDissolved Organic Carbon

Grab_DOC (mg/l) Spectro::lyser_DOC Error Square Mean RMSE1.8 1.76 0.04 0.00162.1 1.382 0.718 0.5155241.5 1.774 0.274 0.0750760.5 0 0.5 0.252.8 2.305 0.495 0.2450251.8 2.228 0.428 0.1831842.2 3.753 1.553 2.4118091.9 2.021 0.121 0.014641

0.679780.462107375

Dissolved Organic Carbon at Leary's Brook from March to June, 2009

0

2

4

6

8

10

12

14

16

Mar-08 Mar-18 Mar-28 Apr-07 Apr-17 Apr-27 May-07 May-17 May-27 Jun-06 Jun-16

Dis

solv

ed O

rgan

ic C

arbo

n (m

g/l)

Spectro::lyser (mg/l) Grab (mg/l)

Page 26: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

ConclusionsConclusions

A major assumption:

Grab samples report the true value of each parameter

RMSE is an indicator of accuracy:

Known values versus modelled values

Caution must be used:

RMSE values are derived from <10 samples

Page 27: A comparison of Spectro::lyser, YSI 6-Series, and Hydrolab ... · Hydrolab_Temp YSI_Temp Hydrolab_Temp 1 YSI_Temp 0.999142354 1 Temperature at Leary's Brook from March to June, 2009

ConclusionsConclusions

RMSE Differences can distort perceptions

Spectro::lyser, Hydrolab, and YSI are in close agreement

A quick glance at a time-series plot indicates different sondes are comparable

Accuracy is important, but perfection is not sought

Most important: change of variable over time