PRESENTED BY LEE FAH HOONG PRINCIPAL TECHNICAL SUPPORT ENGINEER, SEA
840 USER INTERFACE WORKSHOP
![Page 2: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/2.jpg)
WORKSHOP REQUIREMENTS
PB840 Vent
AC
BPS
Inspiratory and Expiratory Filters
Wall Air and Wall Oxygen sources & hoses
![Page 3: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/3.jpg)
Grasp the Oxygen Hose firmly and push the Quick Connect fitting into the Oxygen Outlet:
CONNECTING THE GAS SOURCES:
![Page 4: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/4.jpg)
CONNECT THE PATIENT CIRCUIT, FILTERS & LUNG
NOTE: to eliminate leaks, make sure that the barbs of the patient circuit wye fit snuggly in the patient circuit tubing grooves.
![Page 5: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/5.jpg)
THE RIGHT WAY TO POWER UP THE 840:
The 840 should always be powered on with an open circuit.
By leaving the Expiratory Bacteria Filter Latch open, you’ve created an open circuit condition. You can also leave the test lung disconnected.
![Page 6: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/6.jpg)
NORMAL OPERATION STARTUP SCREENS
On power up the ventilator runs POST.
Default startup screen is displayed there
are no failure conditions
Note what is displayed on the upper
screen:
Vent Startup in progress
Circuit type?
Humidification type?
Time and date?
SST & EST overall test result?
![Page 7: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/7.jpg)
“NORMAL OPERATION” LOWER SCREEN DISPLAY AT START-UP
NOTE: Running EST
wipes out the SAME
PATIENT settings from
the previous setup.
In that case, you may
only see NEW PATIENT
& SST buttons available
in normal operation.
![Page 8: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/8.jpg)
IMPROPER 840 START-UP:
For safety, the ventilator issues a Procedure Error
Alarm and initiates Safety Ventilation Mode if a
patient (or closed circuit condition) is detected
before all of the ventilator settings have been
entered and confirmed (accepted).
Safety ventilation is an emergency mode of
ventilation per the displayed settings.
![Page 9: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/9.jpg)
SAFETY VENTILATION
![Page 10: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/10.jpg)
After POST: 840 generates 10 LPM Flow to check for patient connection.
If the measurement circuitry detects the 10 LPM flow, Procedure Error is initiated.
![Page 11: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/11.jpg)
GUI STATUS INDICATOR PANEL
![Page 12: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/12.jpg)
GUI KEYBOARD
Rotary Encoder Knob
ACCEPT key
CLEAR key
MANUAL INSP key
100% O2 / CAL 2 min key
RESET key
2 min key
VOL key
SCREEN LOCK key
![Page 13: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/13.jpg)
GUI SCREEN SECTIONS Patient Data
Alarm and Status
Miscellaneous Data
SandBox
Primary Settings
Symbols Definitions
Prompt Area
![Page 14: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/14.jpg)
LOWER STARTUP SCREEN
• Press the “Same Patient”
button
![Page 15: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/15.jpg)
LOWER STARTUP SCREEN
• A button is activated when it turns orange.
• Press ACCEPT.
![Page 16: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/16.jpg)
UPPER GUI SCREEN
• Now lower the blue expiratory bacteria filter latch
![Page 17: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/17.jpg)
THE NEXT SCREEN DEFAULTS TO THE APNEA SETUP SCREEN AFTER
SELECTING SAME PATIENT OR COMPLETING A NEW PATIENT SETUP
![Page 18: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/18.jpg)
ALWAYS PRESS THE ALARM SETUP BUTTON TO ADJUST THE ALARM
SETTINGS OR TO TURN THEM OFF IF YOU ARE EVALUATING (ASSESSING) THE
VENT OR RUNNING PVT. HOWEVER, FOR NOW, DO NOT ADJUST THEM.
![Page 19: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/19.jpg)
PATIENT DATA – WHAT DOES IT MEAN?
![Page 20: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/20.jpg)
CIRCUIT & HUMIDIFICATION TYPE AND TIME & DATE
![Page 21: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/21.jpg)
THE MORE INFORMATION SCREEN IF LEAK COMPENSATION IS ENABLED
O2 Sensor Reading – if the reading is out, can calibrate, disable or replace
![Page 22: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/22.jpg)
ALARM LOG ICON
![Page 23: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/23.jpg)
SAMPLE MORE ALARMS SCREEN
![Page 24: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/24.jpg)
ALARMING DOGGIES
![Page 25: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/25.jpg)
PRESS THE OTHER SCREENS BUTTON & TO VIEW THE OTHER
SCREENS. AFTER SELECTING A SCREEN TO VIEW, YOU MUST
PRESS THE OTHER SCREENS BUTTON AGAIN TO BRING BACK
THE SCREEN SELECT MENU BELOW:
![Page 26: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/26.jpg)
PRESS TEST SUMMARY
Next, press the Other Screens button to bring back the screen select menu.
![Page 27: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/27.jpg)
PRESS VENTILATOR CONFIGURATION Ventilator S/W
POST S/W
GUI & Ventilator
serial numbers (S/N)
Compressor S/N
SAAS S/W
Options
Next, press the Other Screens button to bring back the screen select menu.
SAAS = Safety Audible Alarm System
![Page 28: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/28.jpg)
PRESS DIAGNOSTIC CODE LOG
Next, press the Other Screens button to bring back the screen select menu.
![Page 29: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/29.jpg)
PRESS OPERATIONAL TIME LOG
Next, press the Other Screens button to bring back the screen select menu.
![Page 30: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/30.jpg)
PRESS SST RESULTS LOG
Next, press the Other Screens button to
bring back the screen select menu.
![Page 31: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/31.jpg)
PRIMARY SETTINGS AREA: CONTAINS VENTILATOR
SETTINGS DEFINING HOW A BREATH IS DELIVERED TO THE
PATIENT.
10
A/C VC V-TRIG 50
365 44 21
RAMP
1. Which button represents how often the vent will deliver a
controlled breath?
2. Which button represents how much oxygen is delivered?
3. Which button represents how much volume is delivered?
4. Which button represents how much positive pressure is
maintained during exhalation?
5. Which button represents the rate of volume delivery?
![Page 32: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/32.jpg)
SYMBOLS DEFINITIONS AREA – PRESS THE PEAK FLOW BUTTON TO VIEW ITS NAME:
When a button is pressed, you know it is activated
because it changes to an
orange color.
10 44 21
AC
RAMP
![Page 33: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/33.jpg)
CHANGING THE IBW FROM VENT SETUP TOUCH THE IBW BUTTON:
50 110 50
110
CURRENT
SETUP
Current Vent Settings
Sandbox
![Page 34: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/34.jpg)
CHANGING THE IBW FROM VENT SETUP: TURN THE KNOB TO LOWER THE IBW BELOW RECOMMENDED S/W LIMITS:
12
A/C VC V-TRIG 50
400 50 21
RAMP 0
25 55 Press OK
& continue to
turn the knob.
![Page 35: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/35.jpg)
TOUCH CONTINUE & PRESS THE ACCEPT KEY
12
A/C VC V-TRIG 50
400 50 21
RAMP 0
9 20
![Page 36: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/36.jpg)
UPDATED IBW SETTING
12
A/C VC V-TRIG 50
400 50 21
RAMP 0
9 20
9
Now change the IBW
Back to 50 kg!
![Page 37: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/37.jpg)
Peak Pressure = PEEP + Pi
20
Now change the PC breath type back to VC and PEEP
from 10 to 3cmH2O.
Pressure Control Ventilation
![Page 38: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/38.jpg)
ALARM SETUP ICON
Ppeak (High circuit pressure limit) – sets the maximum circuit pressure allowed during inspiration
![Page 39: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/39.jpg)
ALARM SETUP ICON
fTot (Breath rate) – sets the maximum breath rate limit (may be an indicator of breath stacking, autocycling or patient distress)
VeTot (Minute Volume) – sets the min & max exhaled minute volume for mandatory & spontaneous breath types
![Page 40: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/40.jpg)
ALARM SETUP ICON
Vte mand (Mandatory exhaled tidal volume limit) – sets the min and max mandatory exhaled tidal volume limit
![Page 41: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/41.jpg)
ALARM SETUP SCREEN
Medium urgency alarm High urgency alarm
![Page 42: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/42.jpg)
LAST 200 BREATHS TO HELP CLINICIANS SET APPROPRIATE ALARM LIMITS
![Page 43: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/43.jpg)
TURNING OFF THE 840 ALARMS
Touch the alarm setup icon in the lower screen to activate that screen if it is not already selected.
Turn off all alarms by pressing the gray alarm buttons and turning the knob to raise the upper limit alarms until they read OFF or 100 cmH2O for Ppeak (it’s max setting).
Turn off all the lower limit alarms until they read OFF or 0.050 ml for Vetot (it’s min setting).
![Page 44: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/44.jpg)
OTHER SCREENS ICON
Click on each of
the buttons
to view the
associated
screen
![Page 45: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/45.jpg)
WORK?
![Page 46: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/46.jpg)
Performed during normal operation when ventilating – should be performed DAILY when in use
When the 100% O2 Cal 2 MIN key is pressed, vent performs a single-point oxygen sensor calibration during the 100% suctioning procedure
Allows you to calibrate the sensor without having to disconnect the patient
Oxygen Sensor Calibration
![Page 47: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/47.jpg)
CALIBRATE THE OXYGEN SENSOR Press the 100% O2/CAL 2min key on the keyboard to
calibrate the Oxygen sensor.
Press the More Information icon in the Upper Screen to view the O2 Sensor reading:
![Page 48: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/48.jpg)
UNDERSTANDING THE VT VS VTE RELATIONSHIP Locate Vt in lower screen and compare to Vte in
upper screen
Vte Must be +/-10mls +/- 10% of set
VtVte = Spirometry = Patient’s Exhaled Tidal Volume
Set Tidal Volume (Vt) 365 mls
+ 10 mls
375 mls
+ 36.5 mls (+ 10%)
High Limit 411.5 mls
Set Tidal Volume (Vt) 365 mls
- 10 mls
355 mls
- 36.5 mls (- 10%)
Low Limit 318.5 mls
![Page 49: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/49.jpg)
5 CONDITIONS THAT MAY CAUSE THE SPIROMETRY READING TO BE INACCURATE.
1. A leak 2. An obstruction 3. Wrong humidifier type 4. Wrong circuit type 5. A bad Q3 exhalation flow
sensor.
![Page 50: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/50.jpg)
Time &
Date Stamps
Scroll Bar
Alarm Log
Can store up to 80 of the most recent events
Tracks PATIENT RELATED ALARMS, Device Alerts, and System Faults such as Low/Loss AC, Low/Loss of O2 and/or Air, Circuit Disconnects, and Occlusions
Completing a new patient setup clears log memory
![Page 51: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/51.jpg)
ALARM LOG The Alarm Log shows
alarm events in order of occurrence, with the most recent event at the top of the list:
Time stamped alarms
Alarm silences
Alarm resets
The question mark in a triangle appears if the log contains an event that has/had not been viewed yet.
![Page 52: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/52.jpg)
ALARMS
High-urgency alarm indicator
Medium-urgency alarm indicator
Low-urgency alarm indicator
Alarms on the 840 Ventilator
System are classified as
High, Medium, or Low
urgency
! !
!
![Page 53: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/53.jpg)
High Priority Alarm: 5 tones
SMART ALARMS
![Page 54: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/54.jpg)
High urgency- requires immediate attention.
Red high urgency !!! indicator flashes rapidly
High urgency audible alarm sounds (5 tones repeated twice)
Upper screen flashes alarm message
ALARMS
![Page 55: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/55.jpg)
CIRCUIT DISCONNECT ALARM
Press the Alarm Log icon in the upper GUI screen and the Alarm setup icon in the lower GUI screen.
Create a tubing disconnect by removing the test lung from the patient wye.
What urgency alarm is generated? What is displayed in the upper GUI screen? Does pressing the Alarm Silence button silence the alarm? What new entry is in the Alarm Log?
Reconnect the test lung. What happened? Does the Alarm Log reflect the series of events?
![Page 56: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/56.jpg)
Medium Priority Alarm: 3 tones
SMART ALARMS
![Page 57: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/57.jpg)
Alarms
Medium urgency requires prompt attention
Yellow medium urgency !! indicator flashes slowly
Medium urgency audible alarm (a repeating 3 tone sound) is enunciated
The upper screen flashes an alarm message
![Page 58: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/58.jpg)
Low Priority Alarm: 2 tones
Smart Alarms
![Page 59: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/59.jpg)
Low urgency alarms tell you there has been a change in patient-ventilator system
Yellow low urgency !! indicator lights
Low urgency audible alarm (two tone non-repeating sound) is enunciated
Upper screen displays an alarm message
ALARMS
![Page 60: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/60.jpg)
Alarm handling:
Medium urgency alarms may escalate to high or de-escalate to low urgency
Low urgency alarms may escalate to medium or high urgency
All patient alarms will auto-reset if the criteria is reached
ALARMS
![Page 61: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/61.jpg)
2 of last 4 mand breaths < set limit.
Check for leaks, changes in patient’s R & C.
V TE MAND Base Message:
Analysis:
Remedy:
ALARMS
![Page 62: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/62.jpg)
GOLDEN RULE #1
If possible, always review and, if applicable, make
note of the Alarm Log, Ventilator Settings, Alarm Settings, and
Apnea Settings, before doing anything to the ventilator:
Turn to the Sample Alarm Log on pages 4 & 5 of the Reference Manual Addendum
End!
![Page 63: 840 USER INTERFACE WORKSHOP - Mr Bagaimana · Alarms on the 840 Ventilator System are classified as High, Medium, or Low urgency ! ! ! High Priority Alarm: 5 tones SMART ALARMS High](https://reader034.fdocuments.us/reader034/viewer/2022051918/600a938707471c24ce4eebf8/html5/thumbnails/63.jpg)
QUESTIONS? Q&A