1 59059-88 INSTRUCTION SHEET Odyssey DR/2500 Illuminator Module Replacement Replacing the Illuminator Module 1. Turn off the power switch and remove the cell holder. Disconnect the power supply, the pH probe, and the computer or printer cables. 2. Place the instrument upside down on a padded surface. Use the T10 L-Key supplied in the Illuminator Replacement Module Kit (Cat. No. 59059-00) to remove the three Torx screws that secure the illuminator module (Figure 1). Warning: Some internal components may be damaged by static electricity. Use static discharge protection. 3. Remove the illuminator module by pushing on the cell holder cavity from the top of the instrument and pushing the module out the bottom of the instrument (Figure 1). Figure 1 Removing the Illuminator Module Torx Screws (3 places) Push Down T10 L-Key

description

DR 2500 Lamp Replacement

Transcript of 5905988

Page 1: 5905988

59059-88

INSTRUCTION SHEETOdyssey DR/2500 Illuminator Module Replacement

Replacing the Illuminator Module1. Turn off the power switch and remove the cell holder. Disconnect the power

supply, the pH probe, and the computer or printer cables.

2. Place the instrument upside down on a padded surface. Use the T10 L-Key supplied in the Illuminator Replacement Module Kit (Cat. No. 59059-00) to remove the three Torx screws that secure the illuminator module (Figure 1).

Warning: Some internal components may be damaged by static electricity. Use static discharge protection.

3. Remove the illuminator module by pushing on the cell holder cavity from the top of the instrument and pushing the module out the bottom of the instrument (Figure 1).

Figure 1 Removing the Illuminator Module

Torx Screws (3 places)Push Down

T10 L-Key

1

Page 2: 5905988

Odyssey DR/2500 Illuminator Module Replacement, continued

Note: Do not touch the three optical surfaces of the illuminator module! (See Figure 2.)

4. Insert the new illuminator module from the bottom of the instrument. Make sure the connector pins on the module line up with the connector inside the Odyssey. Do not force the module into place. Secure the module with the three Torx screws (Figure 1).

Figure 2 Inserting New Illuminator Module

5. Return the instrument to the upright position and restore power and other cable connections. No calibration is necessary for the illuminator module.

ComponentsDescription Cat. No.Illuminator Module Replacement................................................................................................................ 59059-00Instruction Sheet ............................................................................................................................................. 59059-88T10 L-Key......................................................................................................................................................... 59117-00Torx Screws...................................................................................................................................................... 49440-00

Optical Surfaces (3 places)3

1

2

Connector Pins

HACH COMPANYWORLD HEADQUARTERSTelephone: (970) 669-3050FAX: (970) 669-2932

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:Call 800-227-4224

Contact the HACH office or distributor serving you.www.hach.com [email protected]

In the U.S.A. –Outside the U.S.A. –On the Worldwide Web – ; E-mail –

toll-free

© Hach Company, 2001. All rights reserved. Printed in the U.S.A. te/dk 06/01 1 ed