463 PHT Parenteral Quality Control

download 463 PHT Parenteral Quality Control

of 40

Transcript of 463 PHT Parenteral Quality Control

  • 8/3/2019 463 PHT Parenteral Quality Control

    1/40

    463 PHT

    Compendia requirement for parenteral

    dosage forms

    Nahla S Barakat, PhD

    Dept. of Pharmaceutics

    College of Pharmacy

    1430-2009

    11--331 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    2/40

    Parenteral quality Control

    Parenteral articles are preparations intended for injection

    through the skin or other external boundary tissue, rather than

    through the alimentary canal.

    They are prepared by methods designed to ensure that theymeet the pharmacopeial requirements for sterility, Pyrogens,

    Particulate matter and other contaminants

    11--332 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    3/40

    Nomenclature

    [DRUG] injection

    [DRUG] for Injection

    [DRUG] Injectable emulsion

    [DRUG] Injectable suspension

    [DRUG ] for Injectable suspension

    11--333 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    4/40

    Large-volume intravenous solution: is a single

    dose injection that is intended for iv use and ispackaged in container labeled as containing

    more than 100 ml

    Small-volume intravenous injection: is applied

    to an injection that is packaged in containers

    labeled as containing 100 ml or less.

    11--334 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    5/40

    Ingredients Vehicles and added substances

    Aqueous vehicles:

    Types of water: Purified water

    Sterile Purified water

    Sterile water for injection

    Bacteriostatic water for Injection Sterile water for inhalation

    Sterile water for Irrigation

    11--335 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    6/40

    Drinking water: may be used in the early stage of chemical

    synthesis and in the early stages of the cleaning of

    pharmaceutical manufacturing equipments.

    Purified water: (USP monograph), is used in the preparation of

    some bulk pharmaceutical chemicals, do not use purified

    water in preparations intended for parenteral administration.

    Must be protected from microbial proliferation.

    Sterile purified water: Is purified water that is packaged and

    rendered sterile, contains no antimicrobial agent.

    It is not for parenteral administration

    11--336 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    7/40

    Water for injection: is intended for use in the preparation of

    parenteral solutions and in the preparation of some bulkpharmaceutical chemicals. It must be protected from

    microbial contamination.

    It meets the requirements of all of the tests under purified

    water + Bacterial Endotoxin Test.

    Sterile water for injection: It is prepared from water for

    injection that is sterilized and suitably packaged. It contains no

    antimicrobial agent. It is used as diluents for parenteral

    products.

    It is packaged in single dose container not larger than 1-Litresize.

    Bacterial Endotoxin Test + Particulate Matter Test.

    11--337 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    8/40

    Bacteriostatic water for injection: is sterile water for injection

    which contain on or more suitable antimicrobial agents. It is

    intended to be used as diluent for parenteral products.

    It is packaged in single or multiple-dose containers, not larger

    than 30 mL size.

    NOT FOR USE IN NEWBORNS

    + Antimicrobial Preservatives-effectiveness

    Sterile water for irrigation: It contains no antimicrobial agent.

    It is intended to delivered rapidly.

    For irrigation only, Not for injection.

    Packaged in single-dose containers of larger than 1-liter size.

    - Particulate Matter Test.

    11--338 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    9/40

    Sterile water for inhalation: Is intended for use in inhalators

    and in the preparation of inhalation solutions.

    Not use for parenteral administration.

    + Bacterial Endotoxin Test.

    11--339 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    10/40

    Other vehicles:

    Fixed oil used in non- aqueous injection are of vegetable

    origin, have no taste, no odour suggesting rancidity,

    Have a saponification value 185-200,

    Have an Iodine value between 79-141.

    Test for:

    Unsaponifiable matter

    Free Fatty acid

    11--3310 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    11/40

    Added substances:

    Suitable substances may be added to preparations intended

    for injection to increase stability or usefulness.Observe special care in the choice and use of added

    substances in preparations for injection that areadministered in a volume exceeding 5 ml.

    The following maximum limit

    For agents containing mercury and cationic SAA, 0.01%

    For those of the types of chlorobutanol, cresol, phenol,0.5%

    For sulfur dioxide, sulfite, bisulfite, metabisulfite, 0.2%

    11--3311 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    12/40

    Substances to prevent the growth of microorganisms must be

    added to preparations intended for injection that are packagedin multiple-dose containers, except the active ingredients are

    themselves antimicrobial.

    Such substances also meet the requirements of Antimicrobial

    preservatives-effectiveness, and Antimicrobial Agents-

    Content.

    The air in the container may be evacuated or be displaced by a

    chemically inert gas, when specified in the monograph,

    information regarding sensitivity of the article to oxygen.

    11--3312 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    13/40

    Volume in container

    An injection container is filled with a volume in slight excess of

    the labeled size

    Determination of filled volume:

    10 mL or more 1 container

    3-10 mL 3 containers

    Less than 3 mL 5 containers11--33463 PHT13

    Viscous liquidMobile liquidLabeled size

    0.12 mL0.1 mL0.5 mL

    0.15 mL0.1 mL1 mL

    0.25 mL0.15 mL2 mL0.5 mL0.3 mL5 mL

    0.7 mL0.5 mL10 mL

    0.9 mL0.6 mL20 mL

    3%2%50 mL or more

  • 8/3/2019 463 PHT Parenteral Quality Control

    14/40

    Take up individually the content of each container into a dry

    hypodermic syringe fitted with 21-guage needle, 2.5 cm

    length.

    Discharge the contents of the syringe into a dry cylinder or in

    dry tared beaker, weigh the solution.

    The content of two or three containers (each 1 or 2-mL) may

    be pooled for the measurement using separate dry syringe.

    The volume is not less than the labeled volume in the case of

    containers examined individually,

    or in the case of pooled samples, is not less than the sum of the

    labeled volumes of the containers taken collectively.

    For injections containing oil, warm the containers and shake

    them thoroughly and remove the content. Cool to 25 C before

    measuring the volume.

    11--33463 PHT14

  • 8/3/2019 463 PHT Parenteral Quality Control

    15/40

    PARTICULATE MATTER IN INJECTIONS

    Particulate matter in injections and parenteral infusions

    consists of mobile undissolved particles, other than gas

    bubbles, unintentionally present in the solutions.

    For the determination of particulate matter, two procedures,

    Method 1 (Light Obscuration Particle Count Test) and Method2 (Microscopic Particle Count Test), are specified hereinafter.

    When examining injections and parenteral infusions for sub-

    visible particles Method 1 is preferably applied. However, it

    may be necessary to test some preparations by the lightobscuration particle count test followed by the microscopic

    particle count test to reach a conclusion on conformance to

    the requirements.

    11--33463 PHT15

  • 8/3/2019 463 PHT Parenteral Quality Control

    16/40

    When Method 1 is not applicable, e.g. in case of preparations

    having reduced clarity or increased viscosity, the test should

    be carried out according to Method 2, Emulsions, colloids, andliposomal preparations are examples.

    Similarly, products that produce air or gas bubbles when

    drawn into the sensor may also require microscopic particlecount testing.

    If the viscosity of the preparation to be tested is sufficiently

    high so as to preclude its examination by either test method, a

    quantitative dilution with an appropriate diluents may be made

    to decrease viscosity, as necessary, to allow the analysis to be

    performed

    11--33463 PHT16

  • 8/3/2019 463 PHT Parenteral Quality Control

    17/40

    METHOD 1. LIGHT OBSCURATION PARTICLE COUNT TEST

    Use a suitable apparatus based on the principle of light

    blockage which allows an automatic determination of the size

    of particles and the number of particles according to size.

    11--33463 PHT17

  • 8/3/2019 463 PHT Parenteral Quality Control

    18/40

    11--33463 PHT18

    General precautions

    The test is carried out under conditions limiting particulate

    matter, preferably in a laminar-flow cabinet.

    Very carefully wash the glassware and filtration equipment

    used, with a warm detergent solution and rinse with abundant

    amounts of water to remove all traces of detergent.

    Immediately before use, rinse the equipment from top tobottom, outside and then inside, with particle-free water.

    Take care not to introduce air bubbles into the preparation to

    be examined, especially when fractions of the preparation are

    being transferred to the container in which the determination is

    to be carried out.

  • 8/3/2019 463 PHT Parenteral Quality Control

    19/40

    Method

    Mix the contents of the sample by slowly inverting the

    container 20 times successively. If necessary, cautiouslyremove the sealing closure. Clean the outer surfaces of the

    container opening using a jet of particle-free water and

    remove the closure, avoiding any contamination of the

    contents. Eliminate gas bubbles by appropriate measures such

    as allowing to stand for 2 min or sonicating.

    For large-volume parenterals, single units are tested.

    For small-volume parenterals less than 25 ml in volume, the

    contents of 10 or more units is combined in a cleaned

    container to obtain a volume of not less than 25 ml;

    11--33463 PHT19

  • 8/3/2019 463 PHT Parenteral Quality Control

    20/40

    the test solution may be prepared by mixing the contents of a

    suitable number of vials and diluting to 25 ml with particle-

    free water or with an appropriate particle-free solvent when

    particle-free water is not suitable.

    Small volume parenterals having a volume of 25 ml or more

    may be tested individually. Remove four portions, each of not less than 5 ml, and count

    the number of particles equal to or greater than 10 m and 25

    m.

    Disregard the result obtained for the first portion, and calculatethe mean number of particles for the preparation to be

    examined.

    11--33463 PHT20

  • 8/3/2019 463 PHT Parenteral Quality Control

    21/40

    Evaluation

    For preparations supplied in containers with a nominalvolume of more than 100 ml, apply the criteria of test 1.A.

    The preparation complies with the test if the average number

    of particles present in the units tested does not exceed 25 per

    mL equal to or greater than 10 m and does not exceed 3 per

    mL equal to or greater than 25 m.

    11--33463 PHT21

  • 8/3/2019 463 PHT Parenteral Quality Control

    22/40

    For preparations supplied in containers with a nominal

    volume of less than 100 ml, apply the criteria of test 1.B. The preparation complies with the test if the average number

    of particles present in the units tested does not exceed 6000

    per container equal to or greater than 10 m and does not

    exceed 600 per container equal to or greater than 25 m.

    If the average number of particles exceeds the limits, test the

    preparation by the Microscopic Particle Count Test.

    11--33463 PHT22

  • 8/3/2019 463 PHT Parenteral Quality Control

    23/40

    METHOD 2. MICROSCOPIC PARTICLE COUNT TEST

    The microscope is equipped with an ocular micrometer

    calibrated with an objective micrometer, a mechanical stage

    capable of holding and traversing the entire filtration area of

    the membrane filter, two suitable illuminators to provide

    episcopic illumination in addition to oblique illumination, andis adjusted to 100 10 magnifications.

    11--33463 PHT23

  • 8/3/2019 463 PHT Parenteral Quality Control

    24/40

    Test 2.A Solutions for parenteral infusion or solutions for

    injection supplied in containers with a nominal content of

    more than 100 mL. The preparation complies with the test if the average number

    of particles present in the units tested does not exceed 12 per

    mL equal to or greater than 10 m and does not exceed 2 per

    mL equal to or greater than 25 m. Test 2.B Solutions for parenteral infusion or solutions for

    injection supplied in containers with a nominal content of less

    than 100 ml.

    The preparation complies with the test if the average number

    of particles present in the units tested does not exceed 3000 per

    container equal to or greater than 10 m and does not exceed

    300 per container equal to or greater than 25 m.

    11--33463 PHT24

  • 8/3/2019 463 PHT Parenteral Quality Control

    25/40

    Pyrogen testing

    11--33463 PHT25

    Is designed to limit to a acceptable level the risks offebrile reaction in the patient to the injectable

    administration of the product concerned.

    It involves measuring the rise in temperature of

    rabbits following i.v injection (into ear vein) of a test

    solution in a dose 10 ml/kg within 10 min.

  • 8/3/2019 463 PHT Parenteral Quality Control

    26/40

    Use temperature-sensing probe to measure the rabbits

    body temp. 30 min before the injection of the test dose.

    (use three rabbits).

    Do not use a rabbit for pyrogen test more frequently than

    once every 48 hours, do not use any rabbit having a temp.

    exceeding 39.8C.

    Record the temp. at 30-min intervals between 1 and 3

    hours subsequent to the injection.

    11--33463 PHT26

  • 8/3/2019 463 PHT Parenteral Quality Control

    27/40

    Consider any temp. decrease as zero rise. If no rabbit show a

    rise in temp. of 0.5 or more its respective control temp. i.e

    the product is pyrogen-free.

    If any rabbit show rise in temp. of 0.5 or more, continue the

    test using 5 other rabbits.

    If not more than 3 rabbits of the 8 rabbits show individual

    rise in temp. of 0.5 or more and if the sum of the 8 individual

    maximum temp. rises doesnt exceed 3.3C ; the material is

    pyrogen-free

    11--33463 PHT27

  • 8/3/2019 463 PHT Parenteral Quality Control

    28/40

    Packaging

    containers for injections The container is made of material that permits

    inspection of the contents.

    Containers are closed by fusion or byapplication of suitable closures.

    11--3328 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    29/40

    Elastomeric closures for Injections are made of materials

    obtained by vulcanization (cross-linking) polymerization,

    polyaddition, or polycondensation of macromolecular organic

    substances (elastomers).

    Closure formulations contain natural or synthetic elastomers

    and inorganic and organic additives to aid or control

    vulcanization, impart physical and chemical properties orcolor, or stabilize the closure formulation.

    Such closure are typically used as part of a vial, bottle, or pre-

    fill syringe package system.

    11--3329 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    30/40

    Criteria for the selection of an elastomeric closure should alsoinclude a careful review of all the ingredients to assure that

    no known or suspected carcinogens, or other toxic substances

    are added.

    DefinitionAn elastomeric closure is a packaging component

    that is, or may be, in direct contact with the drug.

    11--33463 PHT30

  • 8/3/2019 463 PHT Parenteral Quality Control

    31/40

    Functionality tests

    Includes: penetrability, fragmentation, self-sealing capacity Functionality tests are performed on closures intended to be

    pierced by a hypodermic needle.

    The self-sealing capacity is required only for closures intendedfor multiple-dose containers.

    The needle specified for each test is a lubricated long bevelhypodermic needle*.

    * Sterile hypodermic needle with an external diameter of

    0.8mm (21 Gauge).

    11--3331 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    32/40

    Penetrability

    ProcedureFill 10 suitable vials to the nominal volume with

    water, fit the closures to be examined, and secure with a cap.

    Using a new hypodermic needle for each closure, pierce the

    closure with the needle perpendicular to the surface.

    RequirementThe force for piercing is no greater than 10 N (1

    kgf) for each closure, determinedwith an accuracy of 0.25 N

    (25 gf).

    11--3332 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    33/40

    Fragmentation

    Closures for LiquidPreparationsFill 12 clean vials with water

    to 4 mL less than the nominal capacity. Fit the closures to beexamined, secure with a cap, and allow to stand for 16 hours.

    Closures for Dry PreparationsFit closures to be examined

    into 12 clean vials, and secure each with a cap.

    ProcedureUsing a hypodermic needle fitted to a cleansyringe, inject into each vial 1 mL of water while removing 1

    mL of air. Repeat this procedure 4 times for each closure,

    piercing each time at a different site.

    Use a new needle for each closure, checking that it is not

    blunted during the test.

    11--3333 463 PHT

  • 8/3/2019 463 PHT Parenteral Quality Control

    34/40

    Filter the total volume of liquid in all the vials through a single

    filter with a nominal pore size no greater than 0.5 m. Count the rubber fragments on the surface of the filter visible

    to the naked eye.

    Requirement:

    there are no more than 5 fragments visible.

    This limit is based on the assumption that fragments with adiameter 0.5 m are visible to the naked eye.

    In case of doubt, the particles are examined microscopically to

    verify the nature and size.

    11--33463 PHT34

  • 8/3/2019 463 PHT Parenteral Quality Control

    35/40

    Self-Sealing Capacity

    ProcedureFill 10 suitable vials with water to the nominal

    volume. Fit the closures that are to be examined, and cap.

    Using a new hypodermic needle as described above for each

    closure, pierce each closure 10 times, piercing each time at a

    different site.

    Immerse the 10 vials in a solution of 0.1% (1 g per L)methylene blue, and reduce the external pressure by 27 kPa

    for 10 minutes.

    Restore to atmospheric pressure, and leave the vials

    immersed for 30 minutes. Rinse the outside of the vials. RequirementNone of the vials contain any trace of blue

    solution.

    11--33463 PHT35

  • 8/3/2019 463 PHT Parenteral Quality Control

    36/40

    Type of glass containers:

    Type I, II: Soda-lime glass is suitably dealkalized are usually

    used for packaging neutral and acidic parenteral

    preparations.

    Type III is soda-lime glass containers are not used for

    parenteral preparations.

    Type NP glass: for packaging nonparenteral articles, for oral

    and topical use.

    11--33463 PHT36

  • 8/3/2019 463 PHT Parenteral Quality Control

    37/40

    Powdered glass Test:

    Use crushed glass containers in 250-ml conical flask, add 50ml high purity water, cap the flask with borosilicate glass

    beaker

    Place the containers in the autoclave and close it securely hold

    temperature at 1212rC for 30 min., counting from the time

    this temperature is reached.

    cool the flask, decant the water from the flask into a clean

    vessel, and wash the residual powdered glass with high purity

    water, add 5 drops methyl red solution, titrate immediately

    with 0.02 N sulfuric acid . Record the volume of 0.02N Sulfuric acid used to neutralize

    the extract from 10 g of the prepared specimen of glass.

    11--33463 PHT37

  • 8/3/2019 463 PHT Parenteral Quality Control

    38/40

    Water attack at 121C

    Rinse 3 or more containers with high purity water

    Fill each container to 90% of its capacity with high purity

    water

    Cap all the flasks with borosilicate glass beaker, place in the

    autoclave at 121 C for 60 min

    11--33463 PHT38

    mL of 0.02 N acidGeneral

    description

    Type

    1.0High resistant

    borosilicate glass

    I

    0.7

    0.2

    Treated soda lime

    glass

    II

    8.5

    Soda lime glassIII

    15.0

    All purpose glassNP

  • 8/3/2019 463 PHT Parenteral Quality Control

    39/40

    Labels and labeling

    The label states the name of the preparation (in case of a liquid

    preparation)

    The percentage content of drug in a specified volume (in case

    of dry preparation) The route of administration

    Storage condition

    Expiration date

    Name of the manufacturer

    The lot number

    11--33463 PHT39

  • 8/3/2019 463 PHT Parenteral Quality Control

    40/40

    Containers for injection that are intended for use as dialysis, orirrigation solution are labeled to indicate that the contents are

    not intended for use by iv infusion

    Injection intended forveterinary use are labeled to that effect

    The containers are so labeled that a sufficient area of thecontainer remains uncovered for its full length to permit

    inspection of the contents

    11--33463 PHT40