44060ch11.pdf
Transcript of 44060ch11.pdf
-
8/9/2019 44060ch11.pdf
1/5
11-1
11The Use of X-RayFluorescence for CoatWeight Determinations
11.1 Introduction ......................................................................11-1
11.2 Technique...........................................................................11-1
11.3 Method...............................................................................11-2
11.4 Accuracy.............................................................................11-3
11.5 Repeatability and Reproducibility....................................11-3
11.6 Conclusion.........................................................................11-5
11.1 Introduction
The technique of elemental analysis by x-ray fluorescence (XRF) has been applied to the quality control
of coating weights at the plant level. Measurements by nonlaboratory personnel provide precise and rapid
analytical data on the amount and uniformity of the applied coating. XRF has proved to be an effective
means of determining silicone coating weights on paper and film, titanium dioxide loading in paper, and
silver on film.
11.2 Technique
XRF is a rapid, nondestructive, and comparative technique for the quantitative determination of elements
in a variety of matrices. XRF units come in a variety of packages; however, the type of unit most prevalent
in the coating industry is described in this chapter.
The XRF benchtop analyzer makes use of a low level radioisotope placed in close proximity to the
sample. The primary x-rays emitted from the excitation source strike the sample, and fluorescence of
secondary x-rays occurs. These secondary x-rays have specific energies that are characteristic of the
elements in the sample and are independent of chemical or physical state. These x-rays are detected in
a gas-filled counter that outputs a series of pulses, the amplitudes of which are proportional to the energy
of the incident radiation. The number of pulses from silicon x-rays, for example is proportional to the
silicone coat weight of the sample. Because the technique is nondestructive, the sample is reusable for
further analysis at any time.To ensure optimum excitation, alternate radioisotopes may be necessary for different applications. For
silicone coatings and titanium dioxide in paper, an iron-55 (Fe-55) source is used. Fe-55 x-rays are soft
(low energy) and do not penetrate far into a sample. For silver on film, a more energetic americanum-
241 source has been used.
Wayne E. MozerOxford Analytical, Inc.
2007 by Taylor & Francis Group, LLC
-
8/9/2019 44060ch11.pdf
2/5
11-2 Coatings Technology: Fundamentals, Testing, and Processing Techniques
Placing samples just a few millimeters from the excitation source enables high sensitivities to be
obtained. Irradiation of the sample is more efficient the closer the sample is brought to the source. In
the case of low energy x-rays, such as silicon, which are easily absorbed by the atmosphere, a helium
purge should be used. With optimum sample irradiation and helium purging when necessary, measure-
ments within 2 minutes of counting time are typical for most samples.
Calibration curves for different elements and materials can be stored directly in the instrument and
are available for recall. Curves are established by measuring a set of known samples or standards of the
same material. Since XRF is a comparative technique, subsequent analyses are only as good as the quality
of the calibration standards supplied.
The total accumulated intensity is actually a combination of signals from the analyte element and
from the matrix in the form of background. Background intensities may vary depending on the thickness
or the basis weight of the material. These differences may be determined by measuring uncoated or blank
samples that are automatically incorporated into the calibration.
11.3 Method
Three procedural steps must precede analysis. First, spectrum scan helps to set up various calibration
parameters. Next a calibration is established. Almost all the time the instrument is in use, it is dedicated
to analysis of production samples. Typically, spectrum scans are performed on representative standards
to identify the occurrence of elemental peaks of interest in the spectrum, as illustrated in Figure 11.1.
The scan is a qualitative tool and will show the presence of any potential interfacing element. For example,
an Fe-55 source will excite the following range of elements in the periodic scale: aluminum to vanadium
and zirconium to cerium. With the proper combination of sources, it is theoretically possible to measure
from aluminum to uranium. Once any interference has been recognized by a spectrum scan, a calibration
FIGURE 11.1 Spectra scan of silicone coatings on super calendar craft paper.
X-RayCounts/Second
100
75
50
25
100 150 200
Channel
C1Ca
Si
2007 by Taylor & Francis Group, LLC
-
8/9/2019 44060ch11.pdf
3/5
The Use of X-Ray Fluorescence for Coat Weight Determinations 11-3
may be established. A sample is prepared by cutting a disk and placing it onto a nickel-coated paper
holder. The holder ensures that the sample will be held as flat as possible because an uneven surface will
introduce errors in reproducibility, caused by scattering and vacancies at the incident surface. Table 11.1
gives a typical calibration for silicon coat weights.Once a calibration has been established, two samples are selected that bracket the operational range
of the samples measured. These samples will then be used to correct for changes in instrumental
performance with time. The measurement of these two standards and of an uncoated sample for back-
ground correction purposes is a push-button procedure. Termed restandardization is performed once
per shift. Analyses may then be performed: an operator simply cuts a sample to be analyzed, inserts it
into the instrument, and initiates the measurement. Within 23 minutes, the instrument will determine
the concentration of the element of interest, both printing and displaying the results.
11.4 AccuracyTo ensure optimum accuracy, it is important to be aware of three sources of errors: suppliers of coating
material, background changes, and interferences. It has been recognized, although not explained, that
silicones of different suppliers give different sensitivities. When different silicones are analyzed on the
same basis weight paper, lines with different slopes are obtained (Figure 11.2). Thus, to avoid this source
of error, separate calibrations are established for each supplier of silicone. Second, correction for changes
in background will help reduce error.
Figure 11.3illustrates two sets of standards of the same silicone on supercalendar kraft prepared on
different dates. Quite evidently, the samples give the same slopes, they are offset. The analyzer will correct
for this background change by having the operator insert an uncoated sample from the lot of paper to
be coated. Thus, all subsequent analyses will be corrected for the apparent background change.
The final source of error is the presence of another material in high concentrations. Examples include
silicone coatings on polyvinyl chloride (PVC) or titanium dioxide filled films. As the levels in these films
change, the effect on the background for the silicon region of the spectrum will change. Thus, there will
be a raising or a lowering of the apparent silicone coat weight for that film sample. This type of interference
is easily corrected for by software that can be built into the analyzer. The interferences are recognized by
using the spectrum scan and are automatically compensated for during calibration.
11.5 Repeatability and Reproducibility
Repeatability in terms of precision of measurement is a statistical function determining the variability
of repeat measurements on the accumulated intensity and x-ray counts. Since an iron-55 source excites
titanium x-rays more efficiently than silicon x-rays, it follows that sensitivities for titania coatings are
higher, and therefore, measurements are more precise. The precision may always be improved by increas-
ing the coating time, but this does not always result in substantial improvements in accuracy. Typically,
for silicone coatings, reported standard deviations correspond to 0.01 g per square meter of silicone.
TABLE 11.1 Typical Results for a Silicone-on-Paper Calibrationa
Sample
Given Concentration
(g/m2)
X-ray Concentration
(g/m2) Error
1 0.33 0.34 0.01
2 0.41 0.41 0.003 0.54 0.52 0.02
4 0.62 0.62 0.00
5 0.70 0.72 0.02
6 1.50 1.50 0.00
a Standard error = 0.013 g/m2.
2007 by Taylor & Francis Group, LLC
-
8/9/2019 44060ch11.pdf
4/5
-
8/9/2019 44060ch11.pdf
5/5
The Use of X-Ray Fluorescence for Coat Weight Determinations 11-5
11.6 Conclusion
On-site XRF determinations are rapid enough and precise enough for effective quality control programs
for elemental determinations on a variety of substrates. The use of XRF is not limited to measuring
coatings; it is flexible enough to measure related products, such as tin, platinum, and rhodium catalysts,
and other solutions.
2007 by Taylor & Francis Group LLC