3D STRESS ECHOCARDIOGRAPHY
Luigi P. Badano**, MD, FESC
Clinica Cardiologica
Università degli Studi di Padova
Direttore: Prof. Sabino Iliceto
**Dr. Badano has received honoraries and research grants from GE
Healthcare, Sorin cardio S.p.A., Actelion, Edwards Lifesciences
*No off-label use of device
![Page 2: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/2.jpg)
European Society of Cardiology copyright -All right reserved
DIAGNOSTIC ACCURACY OF STRESS ECHO
ExEcho DbEcho DypEcho ExEcho DbEcho DypEcho
Highest 97 95 81 86 95 94
Lowest 74 61 61 64 51 90
Mean 88 81 77 79 80 92
SENSITIVITY SPECIFICITY
Marwick TH. Heart 2003
100
80
60
40
20
0
![Page 3: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/3.jpg)
European Society of Cardiology copyright -All right reserved
STRESS-ECHOLimitations of current 2D Imaging for stress echo
• Time required for image acquisition (potential missing of ischemia)• Operator-dependent image acquisition• Difficult to precisely match myocardial segments between baseline and stress (over- or under-estimation of ischemia)• Quad screen display does not encompass the entire left ventricle (risk of missing stress-induced wall motion abnormalities)
![Page 4: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/4.jpg)
European Society of Cardiology copyright -All right reserved
STRESS-ECHOLimitations of current 2D Imaging for stress echo
![Page 5: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/5.jpg)
European Society of Cardiology copyright -All right reserved
REAL-TIME 4D STRESS-ECHO
The Time Saver
4D Echo (multislice)
Multiplane
![Page 6: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/6.jpg)
European Society of Cardiology copyright -All right reserved
4D STRESS-ECHOLimitations of Previous 3D Imaging for Stress-echo
• Limited time resolution• Reduced image quality vs 2D• Analysis time prolonged by manual cropping the full-volume data-sets• Impossibility to side-by-side display of cropped images
Varnero S et al. Cardiovasc Ultrasound 2008
![Page 7: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/7.jpg)
European Society of Cardiology copyright -All right reserved
STRESS-ECHOLimitations of Previous 3D Imaging for Stress-echo
Varnero S et al. Cardiovasc Ultrasound 2008
![Page 8: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/8.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
(4D DSE)
Aminophylline
0 8
Dipyridamole (mg/Kg)
Time (min)
0.84
6
Baseline Peak Recovery
![Page 9: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/9.jpg)
European Society of Cardiology copyright -All right reserved
Variables n= 107 pts
Males 67 (63%)
Age (yrs) 68±13
Hypertension 52 (49%)
Smokers 73 (68%)
Diabetes 25 (23%)
Hypecholesterolemia 46 (43%)
Clinical Indication
Non cardiac surgery pre-op assessment 63 (59%)
Angina pectoris (unable to exercise) 44 (41%)
Previous CAD
Myocardial infarction 25 (23%)
Previous PCI 13 (12%)
Previous CABG 8 (8%)
Medical therapy Withdrawn
Study Population (December 2008 - January 2010)
4D DIPYRIDAMOLE STRESS ECHO
![Page 10: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/10.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
The challenge2D Acquisition 4D Acquisition
Vs.
Acquisition Time = 65±30 s
Analysis Time= 176±63 s
Temporal resolution= 75±5 fps
Acquisition Time = 16±3 s (p<0.0001)
Analysis Time= 91±5 s (p<0.0001)
Temporal resolution= 41±5 vps (p<0.0001)
Badano LP, Muraru D et al. J Am Soc Echocardiogr 2010
![Page 11: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/11.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHOImage Acquisition
![Page 12: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/12.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO2D Image Analysis
Baseline Stress-echo
![Page 13: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/13.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
Baseline 4D Image Analysis
9 slices 12 slices
![Page 14: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/14.jpg)
European Society of Cardiology copyright -All right reserved
2D 4D
4D DIPYRIDAMOLE STRESS-ECHO
Stress-Echo 4D Image Analysis (4CH-view)
![Page 15: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/15.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
Stress-Echo 4D Image Analysis (2CH view)
2D 4D
![Page 16: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/16.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
Stress-Echo 4D Image Analysis (ApLAX view)
2D 4D
![Page 17: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/17.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
Stress-Echo 4D Image Analysis (SAX Basal)
![Page 18: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/18.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
Stress-Echo 4D Image Analysis (Sax Mid)
![Page 19: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/19.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
Stress-Echo 4D Image Analysis (SAX Apex)
![Page 20: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/20.jpg)
European Society of Cardiology copyright -All right reserved
Variables n= 107 pts
Adequate 4D imaging 84 (79%)
Reasons for test termination
End of the protocol 46 (55%)
New or worsening wall motion 38 (45%)
LV segments available for analysis 1428 x 2
Uninterpretable segments at 2DE 311
Baseline 197 (14%)
Peak stress 114 (8%)
Unininterpretable segments at 4DE 213 (p<0.001 vs 2DE)
Baseline 127(9%) (p<0.03 vs 2DE)
Peak stress 81 (6%)
Badano LP, Muraru D et al. J Am Soc Echocardiogr 2010
4D DIPYRIDAMOLE STRESS ECHO
Results
![Page 21: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/21.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
0
10
20
30
40
50
60
70
80
90
100
Apical Anteroseptal Inferior Ant.-lateral
3D Bas
2D Bas
3D Pk
2D Pk
Su
cces
s R
ate
of
Ad
equ
ate
Im
ag
e A
cqu
isis
tio
n (
%)
p=0.02 p=0.03 p=1.00 p=0.51 p=1.00 p=1.00 p=1.00 p=0.29
Badano LP, Muraru D et al. J Am Soc Echocardiogr 2010
![Page 22: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/22.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO
Results
WMSI 2DE 3DE p Value
Baseline 1.041±0.015 1.049±0.01 NS
Peak stress 1.21±0.025 1.29±0.0253 0.011
Apical 1.34±0.057 1.55±0.078 <0.0001
k= 0.71 k= 0.73
Badano LP, Muraru D et al. J Am Soc Echocardiogr 2010
![Page 23: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/23.jpg)
European Society of Cardiology copyright -All right reserved
Sensitivity (%) Specificity (%)
RT3DE 2DE p RT3DE 2DE p
All territories 80 78 NS 87 91 NS
LAD territory 87 78 0.02 90 93 NS
RCA territory 82 77 NS 85 88 NS
LCX territory 65 63 NS 94 92 NS
Coronary angiography 35 pts.
1 vessel disease 9
2 vessel disease 17
3 vessel disease 9
Badano LP, Muraru D et al. J Am Soc Echocardiogr 2010
4D DIPYRIDAMOLE STRESS ECHO
Results
![Page 24: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/24.jpg)
European Society of Cardiology copyright -All right reserved
Variables n= 107 pts
Adequate 4D Imaging 84 (79%)
Reasons for test termination
End of the protocol 46 (55%)
New or worsening wall motion 38 (45%)
LV segments available for analysis 1428 x 2
Uninterpetable segments at 2DE 311
Baseline 197 (14%)
Peak stress 114 (8%)
Unininterpretable segments at 4DE 213 (p<0.001 vs 2DE)
Baseline 127 (9%) (p<0.03 vs 2DE)
Peak stress 81 (6%)
4D DIPYRIDAMOLE STRESS-ECHO
Results
Badano LP, Muraru D et al. J Am Soc Echocardiogr 2010
![Page 25: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/25.jpg)
European Society of Cardiology copyright -All right reserved
4D DIPYRIDAMOLE STRESS-ECHO Conclusions
PROS:
- Simultaneous acquisition of all views;
- Quick (around 30 s);
- Comprehensive assessment of the whole LV
(long. and circumf. extension of ischemia) ;
- Precise anatomically correct tomographic LV
sections;
- No foreshortening of the apex;
- Reduced operator skill needed for acquisition
- Image quality competing with 2D
- Cost/effectiveness
CONS:
-(Temporal resolution (41 vps))
- Atrial fibrillation
- Nearly real time
- (Visual assessment only)
-No contrastFeasibility 78%
![Page 26: 3D STRESS ECHOCARDIOGRAPHY - European Society …assets.escardio.org/.../23.Badano-3D-stress-echo.pdf · •Difficult to precisely match myocardial segments between baseline and stress](https://reader031.fdocuments.us/reader031/viewer/2022011723/5a7c492c7f8b9a49588c9286/html5/thumbnails/26.jpg)