3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU...

44
Clear Results Accuracy You Can Count On Eric Link LabtronX

Transcript of 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU...

Page 1: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

ClearResultsAccuracy You Can Count On

Eric LinkLabtronX

Page 2: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

I. Accuracy vsprecision

Page 3: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

A digital alarm clock assures precision…but not accuracy

Page 4: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Accuracy Is…

Getting close to the bullseye

Page 5: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Precision is…

Gettingthe same results

Page 6: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

When is it close enough?

Page 7: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

II. Standards vs Readability

Page 8: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

What is an Acre?

8 Furlongs Or 80 chains Or 320 Rods

1 Furlong long by1 Chain wide (4 Rods)

LENGTH OF A MILE?

Or 660 x 66 feet

Or 5280 feet

Page 9: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

An Inch, A Foot, and A Yard

Page 10: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

The Metrics of water

1 cubic cm of water =

1mL =

1g @ 25C

@ 1hpa

(Or 1mbar/22.92inhg)

@sea level

Page 11: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

How do you report? Standard must be

more accurate than normal reported reading

Can not report more information than you have

Page 12: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Do not report Precision

Sensor +/-2.2C

Meter Resolution +/-0.1C

Meter Accuracy +/-0.1% + 0.7C

Each degree of Ambient Temperature above 28C or below 18c add 0.01% + 0.03C

Page 13: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Benchtop Turbidimeter

TL2300Tungsten

TU5200650nm

Range 0-4000NTU 0-40NTU

Resolution 0.001NTU 0.0001NTU

Repeatability 1% or 0.01NTU

1% or 0.002NTU

Accuracy 2% + 0.01NTU

2% + 0.01NTU

Readability

Precision

Page 14: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

III. Being Proactive

Page 15: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

In 1492 Columbus Sailed the Ocean Blue

He landed in the

Bahamas!

He proposed to sail west from the Canary Islands in hopes to find Japan and the riches of the East Indies.

Page 16: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Steven Callahan in 1982

In January , Steven Callahan set sail from the Canary Islands to solo the Columbus voyage

Page 17: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Steven Callahan in 1982• 6 days later, in the middle of the Atlantic Ocean, the boat sank

• He was left with a 5 foot life raft, 3 pounds of food, 8 pints of water, a solar still, and a makeshift spear

• 6 days later, in the middle of the Atlantic Ocean, the boat sank

• He was left with a 5 foot life raft, 3 pounds of food, 8 pints of water, a solar still, and a makeshift spear

Page 18: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Adrift : 76 Days Lost at Sea• He had to get food and fresh water and contend with the sun and

sharks

• 43 days later his raft sprung a leak that he could not patch

• He kept the boat afloat manually 33 more days, until his rescue

• Adrift on the currents, he traveled over miles

• He had to get food and fresh water and contend with the sun and sharks

• 43 days later his raft sprung a leak that he could not patch

• He had to get food and fresh water and contend with the sun and sharks

• 43 days later his raft sprung a leak that he could not patch

Page 19: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

If you do nothing you WILL get somewhere (it may be miles)

If you do something you WILL get closer to where you want to go

Page 20: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU
Page 21: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Being reactive…dealing withthe day to day problems,is just surviving

Page 22: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Instead,decide where you want to goand chart a course

Page 23: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Where are you now?

Page 24: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU
Page 25: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

IV. As FoundvsAs Left

Page 26: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

The Basics of Navigation

He may have landed in Trinidad instead

of the Bahamas

If Columbus would have steered 270° 0

90

180

270 225

270

240

Page 27: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

We know that accuracy and reliability

of lab equipmentdrifts overtime

Page 28: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Temperature Change

MoistureCorrosive Gasesand Chemicals

DustDirtInsects

StainedScratchedLenses & Glassware

Page 29: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Mechanical Parts

Consumable Parts

Vibration

Electrical Problems

Page 30: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

A calibration and maintenance programMakes course correctionsBefore you drift too farFrom where you want to be

Clean, Calibrate, and Perform Periodic Maintenance

Page 31: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

AS FOUND readingsWhere you were may not beWhere you wanted to be

Page 32: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

AS FOUND

AS LEFT

DRIFT

The Difference betweenAS FOUND and AS LEFTEquals the DRIFT

DRIFT can help determineThe frequency of calibrationFor reliable and accurate readings

Page 33: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

LabtronX Calibration RecommendationsTwice per year: All critical and reporting equipment

Frequent AS FOUND failures: may require additional calibration or investigation into the cause of the drift.

For systems using continuous turbidity monitoring, Chapter 109.301(1)(i)(B) requires at a minimum,calibration with an EPA-approved primary standard shall be conducted at least quarterly.

Verification should be done regularly to ensure accuracy

Page 34: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

V. How do you know?

Page 35: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

How do you know?I see the reported number but…

You did the proper test but…

You calibrated the meter but…

You used standards but…

Page 36: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Paperwork

Diploma of ‘How you know’

Page 37: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Unique ID

Equipment and Standards

Make/Model

Type

Serial Number

Page 38: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Nominal Value

Stated Value

Units

(Calculated)

Page 39: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

AS FOUND

Meter Reading

Units

Must not damage standard

Page 40: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

AS LEFT

Meter Reading

Units

Even without calibration

Page 41: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Pass/Fail

Agreed Standard deviation

Uncertainty caused by test method or standard

Page 42: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Correction Factor

Add or Subtract

Units

Not Deviation

Page 43: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Standard Documentation

Certificate with ID

Origin Date

Replacement Date

Last Verified

Test Method and/or equipment

conditions and uncertainty

Page 44: 3 hour tour - Home - One Water Ohio...2018/08/28  · TL2300 Tungsten TU5200 650nm Range 0-4000NTU 0-40NTU Resolution 0.001NTU 0.0001NTU Repeatability 1% or 0.01NTU 1% or 0.002NTU

Questions?

LabtronX800-830-2554