2nd Quarter 2019 Audit Report for Shelby County Health ...
Transcript of 2nd Quarter 2019 Audit Report for Shelby County Health ...
2nd Quarter 2019 Audit Report for
Shelby County Health Department
Air Monitoring Network
Prepared for:
P.O.# B032261
Shelby County Health Department
Pollution Control/Air Monitoring
814 Jefferson Ave, Room 438R
Memphis, TN 38105
Prepared by:
1128 NW 39th Drive
Gainesville, FL 32605
Submitted:
June 2019
environmental engineering
& measurement services, inc.
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
EE-MS.com/transfer/clients/Shelby Co i EEMS
Table of Contents
1.0 Introduction ..................................................................................................................1
2.0 Audits of PM Samplers and Gaseous Pollutant Monitors .......................................5
2.1 Audit Methods and Equipment ........................................................................................ 5
2.1.1 Certification of EEMS Standards .......................................................................... 5
2.1.2 EPA Protocol Gas Standards and EEMS Mobile Laboratory ............................... 5
2.1.3 Level-2 Ozone Standard ........................................................................................ 6
2.2 Summary of Field Audit Activities .................................................................................. 6
2.3 Specific PM Sampler and Gaseous Monitor Audit Activities .......................................... 8
2.3.1 PM Sampler Audits ............................................................................................... 8
2.3.2 Gaseous Pollutant Monitor Audits ........................................................................ 9
2.3.3 Ozone Monitor Audits ........................................................................................... 9
2.3.4 Specific NO2 NOy CO and SO2 Audit Activities................................................. 10
2.4 Specific Meteorological Sensor Audit Activities ........................................................... 10
3.0 Audit Results ..............................................................................................................11
3.1 PM Audit Results ........................................................................................................... 11
3.2 NPAP-TTP Ozone Audit Results ................................................................................... 15
3.3 NPAP-TTP Sulfur Dioxide Audit Results ..................................................................... 15
3.4 NPAP-TTP Carbon Monoxide Audit Results ................................................................ 16
3.5 NPAP-TTP Nitrogen Dioxide Audit Results ................................................................. 17
3.6 Meteorological Sensor Audit Results ............................................................................. 18
3.7 Results and Recommendations ...................................................................................... 20
List of Tables
Table 1. Station Locations .............................................................................................................. 2
Table 2. Equipment Audited........................................................................................................... 2
Table 3. OAQPS Approved Audit Levels ...................................................................................... 9
Table 4. Shelby Farm PM10 and TEOM Sampler Verifications ................................................... 11
Table 5. Shelby Farm PM2.5 Samplers Verifications .................................................................... 12
Table 6. Alabama Ave. PM Samplers Verifications .................................................................... 13
Table 7. Near Road PM Sampler Verifications ............................................................................ 14
Table 8. Ozone TTP Audit Results ............................................................................................... 15
Table 9. Sulfur Dioxide TTP Audit Results ................................................................................. 16
Table 10. Carbon Monoxide TTP Audit Results .......................................................................... 16
Table 11. Nitrogen Dioxide TTP Audit Results ........................................................................... 17
Table 12. Meteorological Sensor Audit Results ........................................................................... 18
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
EE-MS.com/transfer/clients/Shelby Co ii EEMS
List of Figures
Figure 1. EEMS Audit Van-3 ......................................................................................................... 5
Figure 2. Edmund Orgill Park Ozone Audit ................................................................................... 7
Figure 3. Shelby Farms CO and SO2 Inlets with Audit Line Connected........................................ 7
Figure 4. Near Road Site with Audit Line Attached ...................................................................... 8
Figure 5. Temperature and Humidity Sensor Plug ....................................................................... 19
Figure 6. Temperature and Humidity Sensor Connector .............................................................. 19
List of Appendices
Appendix A Audit Data Sheets
Appendix B Maps of Locations
Appendix C Audit Standards Certifications
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
EE-MS.com/transfer/clients/Shelby Co iii EEMS
List of Acronyms and Abbreviations
° F Degrees Fahrenheit
% RH Percent Relative Humidity
CFR Code of Federal Regulations
CO carbon monoxide
DAS data acquisition system
EEMS Environmental, Engineering & Measurement Services, Inc.
EPA Environmental Protection Agency
FRM Federal Reference Method
lpm liters per minute
mm Hg millimeters of mercury
MPHD Metro Public Health Department
mps meters per second
NIST National Institute of Standards and Technology
NO nitric oxide
NPAP National Performance Audit Program
OAQPS Office of Air Quality Planning and Standards
O3 Ozone
PE Performance Evaluation
PM particulate matter
PM2.5 particulate matter of 2.5 microns in aerodynamic diameter or less
PM10 particulate matter of 10 microns in aerodynamic diameter or less
ppm parts per million
PSD prevention of significant deterioration
QA quality assurance
rpm revolutions per minute
S/N serial number
SLAMS State or Local Air Monitoring Stations
SO2 sulfur dioxide
SOP standard operating procedure
TTP Through-The-Probe
μg/m3 micrograms per cubic meter
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 1 EEMS
1.0 Introduction
Environmental, Engineering & Measurement Services, Inc. (EEMS) was contracted by the Shelby
County Health Department (SCHD), Pollution Control/Air Monitoring Branch to conduct audits of
the district’s local ambient air quality gaseous and PM pollutant monitoring network. The air
quality monitoring network consists of five stations in the greater Memphis, Tennessee Metro Area
which are operated by the SCHD. The purpose of this network is to fulfill and comply with specific
monitoring requirements for State or Local Air Monitoring Stations (SLAMS) as specified by the
EPA in 40 CFR Part 58. The operation of the monitoring stations must meet the requirements in
40 CFR Part 58 Appendix A, which defines the quality assurance (QA) requirements for gaseous
and PM pollutant ambient air monitoring. The audits performed by EEMS under this contract
fulfilled the requirement for quarterly independent audits of pollutant monitors in the network. The
QA requirements can be found at:
https://www3.epa.gov/ttn/amtic/files/ambient/pm25/qa/APP_D%20validation%20template%20ve
rsion%2003_2017_for%20AMTIC%20Rev_1.pdf
The trained and certified EEMS field scientist followed the National Performance Audit Program
(NPAP) procedures while performing audits of all air quality monitors. The NPAP is a QA program
implemented by the EPA Office of Air Quality Planning and Standards (OAQPS) to conduct audits
of gaseous air pollutant monitors by standard methods throughout each region of the U.S. The
method includes introduction of National Institute of Standards and Traceability (NIST) audit gases
to the station monitors through the ambient sample inlet, including all filters and fittings. This
method evaluates measurement system accuracy including the entire sample train. The audit gas
concentrations are also measured and verified with an audit analyzer on-site which is calibrated at
the time of the audit.
EEMS performed the NPAP Through-The-Probe (TTP) pollutant monitor audits following EPA’s
Quality Assurance Guidance Document – Method Compendium – Field Standard Operating
Procedures (SOP) for the Federal PM2.5 Performance Evaluation Program and NPAP-TTP Audit
SOP. All procedures and guidance documents used to perform these audits can be found at the
EPA OAQPS website: https://www3.epa.gov/ttn/amtic/npepqa.html
This report includes the results of the NPAP-TTP, meteorological sensors, and PM sampler audits
conducted on May 24 through May 30, 2019. The ambient air quality monitors audited were
operating at five stations in the network:
1. Shelby Farms - NCORE
2. SW Tennessee Community College - Near-road
3. Alabama Ave. (PM only)
4. Frayser (Ozone only)
5. Edmund Orgill Park (Ozone only)
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 2 EEMS
The five stations visited are in the Memphis Metro area and in Shelby County, TN. Map images
of the sites are included in Appendix B. The monitoring station locations were obtained during the
audit visits with a GPS and are provided in Table 1.
Table 1. Station Locations
Site Name Site Address Latitude Longitude Elevation (meters)
Alabama Ave. 416 Alabama Ave. 35.151241 -90.041571 75
Edmund Orgill Park 6855 Mudville Rd. 35.377981 -89.834490 118
Frayser 1330 Frayser Blvd. 35.217527 -90.019770 85
Shelby Farms (NCORE) 6388 Haley Rd. 35.151639 -89.850369 90
Southwest Tennessee Community College
(Near-road) 5767 Macon Cove 35.161218 -89.87063 78
The audited monitoring equipment operating at each site is presented in Table 2.
Table 2. Equipment Audited
Site Name Monitor Make and Model S/N AQS ID
Alabama Ave.
PM2.5 (TEOM 1405) 1405A223701302
47-157-0024
FRM 2025i PM2.5 20353
Edmund Orgill Park Ozone 1138 47-157-1004
Frayser Ozone 1304 47-157-0021
Shelby Farms (NCORE)
CO (T-API T300 EU) 246
47-157-0075 SO2 (T-API T100 EU) 135
NOy (T-API T200 EU) 209
O3 (T-API T400 E) 2664
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 3 EEMS
Site Name Monitor Make and Model S/N AQS ID
PM10 (FRM) (R&P 2025) 20917
PM2.5 (FRM-primary) (TEI 2025i) 203531302
PM2.5 (FRM-colo) (R&P 2025) 20921
PM2.5 (continuous) (TEI TEOM 1400A) 140AB231030006
Ambient Temperature (climatronics 100093)
J2850C
Relative Humidity (Met One 083E-0-6)
T25685
Barometric Pressure unknown
Chemical Speciation not tested
Carbon URG not tested
Southwest Tennessee Community College
(Near-road)
CO (T-API T300 U) 174
47-157-0100 NO2 (T-API200 U) 182
FRM 2025i PM2.5 20740
PM2.5 (continuous) (T-API T640) 200
Additional support materials operated at each of the stations include multi-gas dilution systems,
zero air generation systems, and NIST traceable calibration gases. The sites that measure ozone
also have level 3-certified standard photometers to verify Quality Check (QC) concentrations.
Details of the audits are presented in the following sections:
Section 2.0 Audits of PM Samplers and Gaseous Pollutant Monitors
Section 3.0 Audit Results
Appendix A Audit Data Sheets
Appendix B Maps of Locations
Appendix C Audit Standards Certifications
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 4 EEMS
The preparation of this report, and all the activities and tasks described in this report, were
performed by an accredited NPAP-TTP mobile lab Field Scientist. All procedures followed during
the audits were provided by OAQPS and are available at the OAQPS website:
https://www3.epa.gov/ttn/amtic/npepqa.html.
Any questions related to this audit or audit report should be addressed to:
Eric Hebert EEMS Inc.
Telephone: 352-262-0802 P.O. Box 357593
Fax: 352-371-1144 Gainesville, FL 32635
E-mail: [email protected]
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 5 EEMS
2.0 Audits of PM Samplers and Gaseous Pollutant Monitors
2.1 Audit Methods and Equipment
This section describes the steps followed in the performance of these audits. EEMS followed the
document referenced above rigorously. Supplemental guidance and excerpts from the method can
be found at https://www3.epa.gov/ttn/amtic/npepqa.html.
2.1.1 Certification of EEMS Standards
All standards owned and maintained by EEMS undergo annual NIST traceable certification. The
standards include EPA Protocol Gas standards, digital multi-meters, meteorological sensors and
standards, and various flow rate measurement systems including two DeltaCal devices. Copies of
the annual certifications of the EEMS standards used for these audits are included in Appendix C.
2.1.2 EPA Protocol Gas Standards and EEMS Mobile Laboratory
EEMS owns and maintains a Thermo Environmental Instruments Inc. (TEI) 48i TLE CO analyzer
which is used to verify standard audit gas concentrations during NPAP-TTP audits. The CO
analyzer is mounted and operated in a climate controlled mobile laboratory with a multi-gas
dilution system and NIST traceable EPA Protocol Gas standards. The NIST gases include cylinders
of high concentration CO, low concentration CO, and a multi-blend NO, CO, SO2 mixture. The
mobile laboratory is equipped with its own data acquisition system (DAS). The EEMS audit Van-
3 was used for 2nd quarter audits. An image of the audit van is included in Figure 1.
Figure 1. EEMS Audit Van-3
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 6 EEMS
2.1.3 Level-2 Ozone Standard
EEMS owns and maintains a Thermo Environmental Instruments Inc. (TEI) 49i level-2 ozone
standard photometer. The standard photometer is mounted and operated in the climate controlled
mobile laboratory with a multi-gas dilution system and other standards. The digital data output of
the ozone standard is monitored and recorded by the mobile laboratory DAS.
The standard photometer is transported to Research Triangle Park (RTP) North Carolina, or one of
the EPA regional laboratories for verification at least twice per year. The most recent verification
with the Standard Reference Photometer (SRP) from EPA Region 4 is included in Appendix C.
2.2 Summary of Field Audit Activities
On Friday, May 24, 2019, EEMS personnel met SCHD personnel at the Frayser site for the TTP
ozone audit. Following the audit at Frayser, the mobile lab was moved to Edmond Orgill to
complete the TTP audit of that station’s ozone monitor.
On Tuesday, May 28, EEMS personnel met SCHD personnel at the Shelby Farms NCORE site to
complete TTP audits of the CO, SO2, and NOy pollutant monitors.
On Wednesday, May 29, EEMS met SCHD personnel again at the Shelby Farms site to complete
audits of the ozone monitor, meteorological sensors, and PM samplers. Audits of the PM samplers
at both the near road and Alabama Avenue sites were also performed on Wednesday.
On Thursday, May 30, EEMS and SCHD personnel met at the Near-Road site for TTP audits of
the gaseous pollutant monitors and verification of the continuous PM sampler. An exit report and
debriefing interview with the network management team was also conducted. Field activities were
concluded on the afternoon of Thursday, May 30.
The shelter temperature sensor measurements were compared to the EEMS standard. All station
sensors audited, with the exception of the sensor at the near road site, were within 1-degree C of
the standard. The shelter temperature response at the near road site was approximately 1.4 degrees
higher than the temperature measured by the EEMS standard.
Images of some of the sites and samplers are provided in Figures 2 through 4 below.
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 7 EEMS
Figure 2. Edmund Orgill Park Ozone Audit
Figure 3. Shelby Farms CO and SO2 Inlets with Audit Line Connected
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 8 EEMS
Figure 4. Near Road Site with Audit Line Attached
2.3 Specific PM Sampler and Gaseous Monitor Audit Activities
This section describes the procedures used for audits of the gaseous and PM pollutant monitor
parameters at all sites. More detailed NPAP-TTP audit procedures can be found at:
https://www3.epa.gov/ttn/amtic/npepqa.html.
2.3.1 PM Sampler Audits
The sampler dates and times were verified for all samplers and found to be within 1 minute of the
actual time. The PM10 (first cut point) inlets were removed from the samplers and the EEMS leak
check device and DeltaCal or Bios standard were installed at the inlet to the sample train. The
samplers’ operational variables (flow rate, temperature, and pressure) were compared to the
variables as measured by the standard. The variables for all tested samplers were found within
acceptance limits, and all samplers passed leak checks with specific filter cassettes. Audit results
are included in Section 3.
The optical continuous PM2.5 sampler at the near road site was challenged using a filter to verify
the zero and the manufacturer’s standard dust to verify the span. The EEMS standard was used to
verify the inside and outside temperature, pressure, and flow rate. All tests were within acceptable
limits, however the SCHD span dust (lot number 100128-050-004) is out of certification.
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 9 EEMS
2.3.2 Gaseous Pollutant Monitor Audits
The EEMS mobile laboratory audit analyzers and systems were allowed to warm-up overnight prior
to each station audit. Audits of gaseous monitors are performed by providing audit gas to monitor
inlets using Teflon “tee” connectors or Teflon bag for the Near-road sample inlet.
All monitor sample pressures and flow rates were checked prior to, and following the introduction
of audit gas to ensure that changes to the routine sampling variables did not occur as a result of the
addition of audit gas (test atmosphere) to the sampling inlet.
Audit test gas concentrations were selected from the OAQPS list of approved audit levels included
in the 40 CFR part 58 revisions. Table 3 provides the approved list of audit levels. Audit
concentrations from at least three audit levels were selected for each monitor audit. The SCHD
Department Manager provided a list of target audit levels and the EEMS audit personnel attempted
to meet those requirements. The final results of the NPAP-TTP audits are included in Section 3.0.
Table 3. OAQPS Approved Audit Levels
Audit Level Concentration Range, ppm
O3 SO2 NO2 CO
1 0.004 - 0.0059 0.0003 - 0.0029 0.0003 - 0.0029 0.020 - 0.059
2 0.006 - 0.019 0.0030 - 0.0049 0.0030 - 0.0049 0.060 - 0.199
3 0.020 - 0.039 0.0050 - 0.0079 0.0050 - 0.0079 0.200 - 0.899
4 0.040 - 0.069 0.0080 - 0.0199 0.0080 - 0.0199 0.900 - 2.999
5 0.070 - 0.089 0.0200 - 0.0499 0.0200 - 0.0499 3.000 - 7.999
6 0.090 - 0.119 0.0500 - 0.0999 0.0500 - 0.0999 8.000 - 15.999
7 0.120 - 0.139 0.1000 - 0.1499 0.1000 - 0.2999 16.000 - 30.999
8 0.140 - 0.169 0.1500 - 0.2599 0.3000 - 0.4999 31.000 - 39.999
9 0.170 - 0.189 0.2600 - 0.7999 0.5000 - 0.7999 40.000 - 49.999
10 0.190 - 0.259 0.8000 - 1.000 0.8000 - 1.000 50.000 - 60.000
2.3.3 Ozone Monitor Audits
Ozone audit test gas was generated with the ozone generator in the mobile laboratory’s dilution
system. The audit gas was delivered to the station monitor through the station inlet tubing using a
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 10 EEMS
Teflon “tee” to allow the audit gas to vent at the inlet. The audit gas concentration was measured
with the level-2 standard photometer in the mobile lab and recorded by the mobile lab DAS. One
minute averages were reported by the site operator from the station monitor and compared to the
averages for the same time period from the standard.
2.3.4 Specific NO2 NOy CO and SO2 Audit Activities
Prior to beginning the NOx, NOy, CO, and SO2 audits, the EEMS audit CO analyzer was calibrated
using the mobile lab zero air system and CO gases. This was accomplished by supplying the EEMS
mobile laboratory CO analyzer with NIST gas directly from the cylinders and not through the
mobile lab dilution system. Data from this “Pre-audit” calibration was entered into the OAQPS
audit spreadsheet to be used to calculate the concentration of all audit gases generated during the
station audits.
The site audits were accomplished by generating audit test gas using the mobile laboratory dilution
system and the NIST multi-blend gas. The audit gas was delivered to the station monitors at the
station sample inlet, through all filters and fittings using Teflon tubing (and/or) a Teflon bag to
allow the audit gas to vent at the inlet. The generated audit gas CO concentrations were verified
with the mobile laboratory CO analyzer. The other audit gas concentrations were then calculated
based on the ratio of CO to the other gases in the NIST multi-blend audit gas cylinder.
Immediately following the station audits, the mobile laboratory CO analyzer was again challenged
directly with the NIST cylinder gases. The results of this “Post-audit” challenge were entered into
the OAQPS spreadsheet to correct the audit standard gas concentrations, accounting for any drift
of the mobile laboratory CO analyzer. This procedure (Pre-audit CO calibration, multipoint station
audits, and Post-audit CO challenge) is the same OAQPS standard procedure used routinely for all
NPAP-TTP audits.
2.4 Specific Meteorological Sensor Audit Activities
The site temperature and relative humidity sensors were removed from normal operation in order
to be verified with a certified EEMS standard. All station sensor values were obtained from the
station DAS.
The station temperature sensor was placed in three separate water baths with the EEMS standard
and the measurements were compared. The station relative humidity and barometric pressure
sensors were verified using NIST traceable EEMS handheld transfer standards.
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 11 EEMS
3.0 Audit Results
3.1 PM Audit Results
All operational and reporting PM samplers were audited with the EEMS DeltaCal and BIOS
standards. The samplers’ date and time and all operational variables were found to be within
acceptable limits. The Shelby Farm samplers’ results are summarized in Tables 4 and 5. The
Alabama Ave. samplers’ results are included in Table 6. The Near Road samplers’ results are
included in Table 7.
Table 4. Shelby Farm PM10 and TEOM Sampler Verifications
Delta Cal Version N/A Site Shelby Farm
Time verified Yes PM-10 R&P Partisol 2025 s/n= 20917
Date verified Yes PM-2.5 TEI TEOM s/n= 140AB231030006
DeltaCal S/N 1196 EEMS 01451 DeltaCal Cert Date: 2/15/2019
Date & Site of Verification 5/29/2019 Shelby Farm Partisol 2025 PM-10
Parameter DeltaCal 2025 Site Sampler Difference Acceptance
Criteria Pass/Fail
Flow Rate (Lpm) 16.63 16.67 0.23% ≤ ± 4% Pass
Design Flow Rate (16.67 Lpm)
16.63 -0.23% ≤ ± 4% Pass
Ambient Temperature (oC) 30.7 31 0.3 ≤ ± 2 oC Pass
Barometric Pressure (mm Hg) 749.0 749 0
≤ ± 5 mm Hg
Pass
Filter Temperature (oC) 31.5 31.9 0.4 ≤ ± 2 oC Pass
Leak Check 2 ≤ 25 mm Hg
Pass
Date & Site of Verification 5/29/2019 Shelby Farm TEOM 1400A PM-2.5
Parameter DeltaCal TEOM Difference Acceptance
Criteria Pass/Fail
Flow Rate (Lpm) 16.59 16.66 0.41% ≤ ± 4% Pass
Design Flow Rate (16.67 Lpm) 16.59 -0.47% ≤ ± 4% Pass
Ambient Temperature (oC) 29.7 30.5 0.8 ≤ ± 2 oC Pass
Barometric Pressure (mm Hg) 749.0 746.3 -2.7 ≤ ± 5 mm
Hg Pass
Leak Check (main) 0.09 0.15 lpm Pass
Leak Check (bypass) 0.15 0.6 lpm Pass
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 12 EEMS
Table 5. Shelby Farm PM2.5 Samplers Verifications
Delta Cal Version N/A
Site Shelby Farm
Time verified Yes
PM-2.5 Partisol 2025 Primary s/n= 203531302 POC1
Date verified Yes
PM-2.5 Partisol 2025 colo s/n= 20921 POC2
DeltaCal S/N 1196 EEMS 01451 DeltaCal Cert Date: 2/15/2019
Date & Site of Verification 5/29/2019 Shelby Farm Partisol 2025 PM-2.5 POC1
Parameter DeltaCal 2025 Site Sampler Difference Acceptance
Criteria Pass/Fail
Flow Rate (Lpm) 16.97 16.72 -1.48% ≤ ± 4% Pass
Design Flow Rate (16.67 Lpm) 16.97 1.81% ≤ ± 4% Pass
Ambient Temperature (oC) 29.7 29.4 -0.3 ≤ ± 2 oC Pass
Barometric Pressure (mm Hg) 749.5 749 -0.5 ≤ ± 5 mm
Hg Pass
Filter Temperature (oC) 30.4 31.7 1.3 ≤ ± 2 oC Pass
Leak Check 3 ≤ 25 mm Hg
Pass
Date & Site of Verification 5/29/2019 Shelby Farm Partisol 2025 PM-2.5 POC2
Parameter DeltaCal 2025 Site Sampler Difference Acceptance
Criteria Pass/Fail
Flow Rate (Lpm) 17.00 16.70 -1.77% ≤ ± 4% Pass
Design Flow Rate (16.67 Lpm) 17.00 1.99% ≤ ± 4% Pass
Ambient Temperature (oC) 29.5 28.9 -0.6 ≤ ± 2 oC Pass
Barometric Pressure (mm Hg) 749.5 748 -1.5
≤ ± 5 mm Hg
Pass
Filter Temperature (oC) 32.0 31.7 -0.3 ≤ ± 2 oC Pass
Leak Check 2 ≤ 25 mm
Hg Pass
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 13 EEMS
Table 6. Alabama Ave. PM Samplers Verifications
Delta Cal Version 3.37P Site Alabama Ave
Time verified Yes
PM-2.5 TEI Partisol 2025i s/n= 20737
Date verified Yes
PM-10 TEI TEOM s/n= 1405A223701302
DeltaCal S/N 1196 EEMS # 01451 DeltaCal Cert Date: 2/15/2019
Date & Site of Verification 5/29/2019 Alabama Ave Partisol 2025 PM-2.5
Parameter DeltaCal 2025 Site Sampler Difference Acceptance
Criteria Pass/Fail
Flow Rate (Lpm) 16.69 16.67 -0.13% ≤ ± 4% Pass
Design Flow Rate (16.67 Lpm) 16.69 0.13% ≤ ± 4% Pass
Ambient Temperature (oC) 31.6 31.4 -0.2 ≤ ± 2 oC Pass
Barometric Pressure (mm Hg) 749.5 749 -0.5
≤ ± 5 mm Hg
Pass
Filter Temperature (oC) 34.2 32.7 -1.5 ≤ ± 2 oC Pass
Leak Check 7 ≤ ± 25 mm Hg
Pass
Date & Site of Verification 5/29/2019 Alabama Ave TEOM 1400A PM-10
Parameter DeltaCal TEOM Difference Acceptance
Criteria Pass/Fail
Flow Rate (Lpm) 16.90 16.68 -1.31% ≤ ± 4% Pass
Design Flow (16.67 Lpm) 16.90 1.39% ≤ ± 4% Pass
Bypass Flow (13.67 Lpm) 13.81 13.68 1.01% ≤ ± 4% Pass
Ambient Temperature (oC) 31.5 31.48 -0.02 ≤ ± 2 oC Pass
Barometric Pressure (mm Hg) 749.5 748.6 -0.9 ≤ ± 5 mm
Hg Pass
KO verification 1.53% ≤ ± 4% Pass
Leak Check (main) 0.02 0.15 lpm Pass
Leak Check (bypass) 0.01 0.6 lpm Pass
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 14 EEMS
Table 7. Near Road PM Sampler Verifications
Delta Cal Version N/A Site Near Road
Time & Date Verified Yes PM 2.5 TEI Partisol 2025i s/n= W207401501
Time & Date Verified Yes PM 2.5 T-API T640 s/n= 200
BIOS S/N 287690 & EEMS# 01421 DeltaCal Cert Date: 3/4/2019
DeltaCal S/N 1196 & EEMS# 01451 DeltaCal Cert Date: 2/15/2019
Date & Site of Verification 5/29/2019 Near Road Partisol 2025i PM-2.5
Parameter DeltaCal 2025 Site Sampler
Difference Acceptance
Criteria Pass/Fail
Flow Rate (Lpm) 16.85 16.67 -1.08% ≤ ± 4% Pass
Design Flow Rate (16.67 Lpm) 16.85 1.09% ≤ ± 4% Pass
Ambient Temperature (oC) 29.7 29.3 -0.4 ≤ ± 2 oC Pass
Barometric Pressure (mm Hg) 749.5 750 0.5 ≤ ± 5 mm
Hg Pass
Filter Temperature (oC) 31.0 31.4 0.4 ≤ ± 2 oC Pass
Leak Check 5 mm ≤ 25 mm Hg
Pass
Date & Site of Verification 5/30/2019 Near Road T640 PM-2.5
Parameter BIOS or Std T640 Difference Acceptance
Criteria Pass/Fail
Flow Rate (Lpm) 5.022 5.000 -0.44% ≤ ± 4% Pass
Design Flow Rate (16.67 Lpm) 5.022 0.44% ≤ ± 4% Pass
Ambient Temperature (oC) 26.8 26.4 -0.42 ≤ ± 2 oC Pass
Barometric Pressure (mm Hg) 750 749.3 -0.7 ≤ ± 5 mm
Hg Pass
Dust test 11.3 10.9 -0.4 ≤ ± 0.5 Pass
Zero test 0.0 0.0 Pass
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 15 EEMS
3.2 NPAP-TTP Ozone Audit Results
The audit results of the ozone monitors at the three sites that monitor ozone were within the
acceptance limits of ±10% of any single audit point and also within the warning limit of ±7% (or
1.5ppb for level 2). The results of the O3 audits are summarized in Table 8. The field audit forms
are included in Appendix A.
Table 8. Ozone TTP Audit Results
Site and Audit Level Audit Value (ppm)
Site Value (ppm)
% Difference
Actual Difference
(ppm)
Pass Warning Fail
Shelby Farm – level 6 0.10811 0.10760 -0.5 -0.00051 Pass
Shelby Farm – level 5 0.07999 0.07954 -0.6 -0.00045 Pass
Shelby Farm – level 4 0.06041 0.06053 0.2 0.00012 Pass
Shelby Farm – level 3 0.03729 0.03835 2.8 0.00106 Pass
Shelby Farm – level 2 0.01476 0.01548 4.9 0.00072 Pass
Frayser – level 5 0.07998 0.07926 -0.9 -0.00072 Pass
Frayser – level 4 0.06537 0.06551 0.2 0.00014 Pass
Frayser – level 3 0.02817 0.02854 1.3 0.00037 Pass
Frayser – level 2 0.01637 0.01665 1.7 0.00028 Pass
Edmund Orgill – level 6 0.10700 0.10600 -0.9 -0.00100 Pass
Edmund Orgill – level 5 0.08087 0.08026 -0.8 -0.00061 Pass
Edmund Orgill – level 4 0.05722 0.05695 -0.5 -0.00027 Pass
Edmund Orgill – level 3 0.03656 0.03628 -0.8 -0.00028 Pass
Edmund Orgill – level 2 0.01734 0.01713 -1.2 -0.00021 Pass
3.3 NPAP-TTP Sulfur Dioxide Audit Results
The single SO2 monitor in the network is operated at the Shelby Farm NCORE site. The audit
results were within the acceptance limits of ±15% of any single audit point and within the warning
limit of ±10%. The low-level test points (levels 2 and 1) were within the acceptance limit of
±0.0015ppm. There were no observed problems with the accuracy of the station monitor. The
results of the SO2 audit are summarized in Table 9. The field audit forms are included in Appendix
A.
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 16 EEMS
Table 9. Sulfur Dioxide TTP Audit Results
Site and Audit Level Audit Value (ppm)
Site Value (ppm)
% Difference
Actual Difference
(ppm)
Pass Warning Fail
Shelby Farm – level 6 0.08400 0.0796 -5.2 -0.00437 Pass
Shelby Farm – level 5 0.04198 0.0397 -5.4 -0.00227 Pass
Shelby Farm – level 4 0.01784 0.0165 -7.6 -0.00135 Pass
Shelby Farm – level 2 0.00269 0.0024 N/A -0.00029 Pass
Shelby Farm – level 1 0.00127 0.0013 N/A -0.00001 Pass
3.4 NPAP-TTP Carbon Monoxide Audit Results
There are two CO monitors currently operating in the network. They are located at the Shelby
Farm and Near-road sites. The audit results of all monitors were within the acceptance limits of ±
15% and 0.4ppm of any single audit point. One audit point at the Shelby Farm site was above the
warning criteria of ± 0.030 and ± 10%. There were no observed problems with the accuracy of the
network CO monitors. The results of the CO audits are summarized in Table 10. The field audit
forms are included in Appendix A.
Table 10. Carbon Monoxide TTP Audit Results
Site and Audit Level Audit Value (ppm)
Site Value (ppm)
% Difference
Actual Difference
(ppm)
Pass Warning Fail
Shelby Farm – level 5 4.7507 4.791 0.8 0.04032 Pass
Shelby Farm – level 4 2.6590 2.639 -0.8 -0.02000 Pass
Shelby Farm – level 4 1.3287 1.311 -1.3 -0.01767 Pass
Shelby Farm – level 3 0.5648 0.540 -4.4 -0.02471 Pass
Shelby Farm – level 2 0.0851 0.052 N/A -0.03309 Warning
Shelby Farm – level 1 0.0401 0.014 N/A -0.02611 Pass
Near-road – level 5 4.8054 4.879 1.5 0.07363 Pass
Near-road – level 4 2.6776 2.681 0.1 0.00344 Pass
Near-road – level 4 1.3311 1.344 1.0 0.01290 Pass
Near-road – level 3 0.6591 0.655 -0.6 -0.00377 Pass
Near-road – level 2 0.1632 0.155 N/A -0.00874 Pass
Near-road – level 1 0.0460 0.038 N/A -0.00797 Pass
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 17 EEMS
3.5 NPAP-TTP Nitrogen Dioxide Audit Results
The Near-road site operates a NOx monitor and the NCORE site operates a NOy monitor. The Near
Road and NCORE monitors were audited utilizing gas phase titration (GPT) with ozone to produce
NO2. An alternative method using NPN gas to challenge the NO2 channel directly was not used.
The posted OAQPS method was followed for the audit. The results were within the acceptance
limits of ± 15% of any single audit point. At the Near Road site, one of the low concentration
converter efficiency values was above 104% (114.5%) and failed. However, the overall Near Road
converter efficiency was 102.4% and passing, using the converter efficiency calculation provided
in OAQPS QA Guidance Doc 2.3.
The low-level points which are used for informational purposes and not for data validation were all
within acceptance limits. OAPQS is evaluating the low-level converter efficiency audit method
due to the difficulty in obtaining stable NOx measurements at very low concentrations during TTP
audits.
The results of the NO2 audits are summarized in Table 11. The field audit forms are included in
Appendix A.
Table 11. Nitrogen Dioxide TTP Audit Results
Site and Audit Level Audit Value (ppm)
Site Value (ppm)
% Difference
Actual Difference
(ppm)
Pass Warning Fail
Shelby Farm – level 5 0.04712 0.04552 -3.4 -0.00160 Pass
Shelby Farm – level 5 0.02021 0.02010 -0.5 -0.00011 Pass
Shelby Farm – level 2 0.00445 0.00380 N/A -0.00065 Pass
Shelby Farm – level 1 0.00261 0.00203 N/A -0.00058 Pass
Shelby Farm Converter Efficiency – level 5 97.2% Pass
Shelby Farm Converter Efficiency – level 5 101.3% Pass
Shelby Farm Converter Efficiency – level 2 97.1% Pass
Shelby Farm Converter Efficiency – level 1 99.6% Pass
Shelby Farm Converter Efficiency – OAQPS QA Document 2.3 (slope/intercept calculation)
97.2% Pass
Near-road – level 6 0.08176 0.08280 1.3 0.00104 Pass
Near-road – level 5 0.04789 0.04870 1.7 0.00081 Pass
Near-road – level 4 0.01917 0.01883 -1.8 -0.00034 Pass
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 18 EEMS
Site and Audit Level Audit Value (ppm)
Site Value (ppm)
% Difference
Actual Difference
(ppm)
Pass Warning Fail
Near-road – level 2 0.00398 0.00365 N/A -0.00033 Pass
Near-road – level 1 0.00155 0.00161 N/A 0.00006 Pass
Near-road Converter Efficiency – level 6 102.9% Pass
Near-road Converter Efficiency – level 5 102.6% Pass
Near-road Converter Efficiency – level 4 102.7% Pass
Near-road Converter Efficiency – level 2 98.3% Pass
Near-road Converter Efficiency – level 1 115.5% Fail
Near-road Converter Efficiency – OAQPS QA Document 2.3 (slope/intercept calculation)
102.4% Pass
3.6 Meteorological Sensor Audit Results
The meteorological sensors operating at the Shelby Farms site were audited for the first time during
the 2nd Quarter 2019 audit visit. The sonic wind sensor was not operational and was not audited.
The temperature, relative humidity, and barometric pressure sensors were challenged with audit
standards. The ambient temperature and humidity sensors are mounted at approximately 3 meters
above the ground. The results are summarized in Table 12.
Table 12. Meteorological Sensor Audit Results
Parameter Audit Value Station Value Difference Pass / Fail
Temperature 31.81 33.2 1.39 (° F) Fail
Temperature 81.15 81.9 0.75 (° F) Fail
Temperature 114.71 115.3 0.59 (° F) Fail
Relative Humidity 58.9 61.7 2.8 (% RH) Pass
Barometric Pressure 749.5 751.6 2.1 (mmHg) Pass
The current procedures require that the sensors be returned to the manufacturer once per year for
calibration and certification. The sensors are not calibrated or verified onsite. Some of the sensor
wiring is beginning to deteriorate and is in poor condition. Some signals were interrupted during
the audit procedures. Images of the temperature and humidity signal cable connections are
presented in Figures 5 and 6.
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 19 EEMS
Figure 5. Temperature and Humidity Sensor Plug
Figure 6. Temperature and Humidity Sensor Connector
Shelby County Health Department Monitoring Network Audit Second Quarter 2019
1. EE-MS.com/transfer/clients/Shelby Co 20 EEMS
It is possible that some of the error observed in the signal could be due to the condition of the
wiring.
3.7 Results and Recommendations
The stations in the SCHD monitoring network are in good condition and very well run. Site
operators are knowledgeable and maintain the site monitors in very good working order.
The meteorological measurement accuracy could be improved by replacing the sensors and
wiring. It is recommended that the non-functional wind sensor be replaced with a mechanical
type that could be calibration and audited onsite. Sensors provided by the R.M. Young Company
are less expensive that those currently in use and can be calibrated on site.
It is suggested that if meteorological measurements are a priority then a program to include
calibrations and audits be developed.
APPENDIX A
Audit Data Sheets
Site Name: Edmund Orgill Audit Date: 5/24/2019
-0.00100 Pass -0.00061 Pass -0.00027 Pass -0.00028 Pass -0.00021 Pass -0.00003 Pass
N/AN/AN/AN/AN/A
N/AN/AN/AN/AN/A
N/AN/AN/AN/AN/A
N/AN/AN/AN/A
N/AN/AN/AN/A
N/AN/AN/AN/AN/A
Post Zero
Post Zero
Post Zero
Pre Zero
Post Zero
Pre Zero
Pre Zero
Pre Zero
Post Zero
Post Zero
NOx Audit Point #5
NO Audit Point #3NO Audit Point #4
NOx Audit Point #1NOx Audit Point #2
NOx Audit Point #4
SO2 Audit level 2
Converter Efficiency NO2 level 1
NO2 Audit level 4NO2 Audit level 2
Converter Efficiency NO2 level 5Converter Efficiency NO2 level 4Converter Efficiency NO2 level 2
Pre Zero
NO2 Audit level 1
NO2 Audit level 5
SO2 Audit level 1
SO2 Audit level 4
Sulfur Dioxide
SO2 Audit level 6SO2 Audit level 5
Oxides of Nitrogen
NO Audit Point #1NO Audit Point #2
NOx Audit Point #3
NO Audit Point #5
CO Audit level 2CO Audit level 1
FINAL SUMMARY AUDIT REPORT CO BASEDEEMS Van-3
Ozone
ParameterNPAP Lab Response
(ppm)Station Response
(ppm)
-0.9
0.017130.00051 0.00048
-0.8
-0.8
Ozone audit level 5Ozone audit level 4
Ozone audit level 2Ozone audit level 3
Pre Zero
Percent Difference
-1.2
Actual Difference
(ppm)Warning
CO Audit level 4CO Audit level 3
Carbon Monoxide
0.080870.057220.03656
0.08026Ozone audit level 6
0.05695
Pass/Fail
0.10700
-0.5
0.017340.03628
0.10600
CO Audit level 5
Site Name: Edmund Orgill Airs ID: 471571004Auditor: Eric Hebert Audit Date: 05/24/19Station Manager: Judy Low / Yong Cai
Instrument: Ozone COManufacturer: Thermo TEIModel: 49i 48i TLESerial Number: 1180030022 1503464200Calibration Date: 09/07/18 CalibratedSlope 1.0029 on Intercept 0.0001098 day of audit
STATION INSTRUMENT INFORMATIONInstrument:Manufacturer/Model #: T-API T400Property Number:Calibration Date:Slope/Intercept: 1.0076 -0.0946Indicated Flow:In-Line Filter Change:Manifold Type:
Pass/Fail Warning Auditor Eric Hebert
Ozone audit level 6 Pass PrintOzone audit level 5 Pass Ozone audit level 4 Pass SignatureOzone audit level 3 Pass Ozone audit level 2 Pass
EPA person notified in case of audit failureAudit Limits
Pass Bias < ±10% OR difference from actual concentration < 24 hour allowable drift (0.003 ppm) Fail Bias > ±10% AND difference from actual concentration > 24 hour allowable drift (0.003 ppm)
Warning Bias > ±7% AND difference from actual concentration > 0.0015 ppm
Comments:0
0.03656 0.03628 -0.8
0.00051 0.00048
0.01734 0.01713 -1.2
0.08087 0.08026 -0.8
0.05722 0.05695 -0.5
FINAL NPAP THROUGH-THE-PROBE AUDIT REPORT
MOBILE PE LAB INSTRUMENTS
05/14/191/4 " Teflon & glass
0.82
113805/22/19
EEMS Van-3
0.10700 0.10600 -0.9
Ozone
OZONE REPORT
NPAP O3 Concentration (ppm)
FINAL OZONE AUDIT RESULTS
Site Response (ppm)
Percent Difference
Site Name: Frayser Audit Date: 5/24/2019
-0.00072 Pass 0.00014 Pass 0.00037 Pass 0.00028 Pass
N/A0.00032 Pass
N/AN/AN/AN/AN/A
N/AN/AN/AN/AN/A
N/AN/AN/AN/AN/A
N/AN/AN/AN/A
N/AN/AN/AN/A
N/AN/AN/AN/AN/A
Warning
CO Audit level 4CO Audit level 3
Carbon Monoxide
0.065370.028170.01637
0.06551Ozone audit level 5
0.02854
Pass/Fail
0.07998
1.30.01665
0.07926
CO Audit level 5
FINAL SUMMARY AUDIT REPORT CO BASEDEEMS Van-3
Ozone
ParameterNPAP Lab Response
(ppm)Station Response
(ppm)
-0.9
0.00051 0.00083
0.2
1.7
Ozone audit level 4Ozone audit level 3
Ozone audit level 1Ozone audit level 2
Pre Zero
Percent Difference
Actual Difference
(ppm)
NO Audit Point #2
NOx Audit Point #3
NO Audit Point #5
CO Audit level 2CO Audit level 1
Oxides of Nitrogen
NO Audit Point #1
SO2 Audit level 1
SO2 Audit level 4
Sulfur Dioxide
SO2 Audit level 6SO2 Audit level 5
SO2 Audit level 2
Converter Efficiency NO2 level 1
NO2 Audit level 4NO2 Audit level 2
Converter Efficiency NO2 level 5Converter Efficiency NO2 level 4Converter Efficiency NO2 level 2
Pre Zero
NO2 Audit level 1
NO2 Audit level 5
NOx Audit Point #5
NO Audit Point #3NO Audit Point #4
NOx Audit Point #1NOx Audit Point #2
NOx Audit Point #4
Post Zero
Post Zero
Post Zero
Pre Zero
Post Zero
Pre Zero
Pre Zero
Pre Zero
Post Zero
Post Zero
Site Name: Frayser Airs ID: 471570021
Auditor: Eric Hebert Audit Date: 05/24/19Station Manager: Betty Brown / Judy Low
Instrument: Ozone COManufacturer: Thermo TEIModel: 49i 48i TLESerial Number: 1180030022 1503464200Calibration Date: 09/07/18 CalibratedSlope 1.0029 on Intercept 0.0001098 day of audit
STATION INSTRUMENT INFORMATIONInstrument:Manufacturer/Model #: T-API T400Property Number:Calibration Date:Slope/Intercept: 1.0156 0.0027Indicated Flow:In-Line Filter Change:Manifold Type:
Pass/Fail Warning Auditor Eric Hebert
Ozone audit level 5 Pass PrintOzone audit level 4 Pass Ozone audit level 3 Pass SignatureOzone audit level 2 Pass Ozone audit level 1 N/A
EPA person notified in case of audit failureAudit Limits
Pass Bias < ±10% OR difference from actual concentration < 24 hour allowable drift (0.003 ppm) Fail Bias > ±10% AND difference from actual concentration > 24 hour allowable drift (0.003 ppm)
Warning Bias > ±7% AND difference from actual concentration > 0.0015 ppm
Comments:
0.07998 0.07926 -0.9
Ozone
OZONE REPORT
NPAP O3 Concentration (ppm)
FINAL OZONE AUDIT RESULTS
Site Response (ppm)
Percent Difference
FINAL NPAP THROUGH-THE-PROBE AUDIT REPORT
MOBILE PE LAB INSTRUMENTS
04/22/191/4 " Teflon & glass
0.76
130404/15/19
EEMS Van-3
0.06537 0.06551 0.2
0.02817 0.02854 1.3
0
0.01637 0.01665 1.7
0.00051 0.00083
Site Name: Shelby Farm - NCORE Audit Date: 5/29/2019
0.00040 Pass-0.00051 Pass -0.00045 Pass 0.00012 Pass 0.00106 Pass 0.00072 Pass 0.00121 Pass
N/AN/AN/AN/AN/A
N/AN/AN/AN/AN/A
N/AN/AN/AN/AN/A
N/AN/AN/AN/A
N/AN/AN/AN/A
N/AN/AN/AN/AN/A
0.00040 0.00080
Post Zero
Post Zero
Post Zero
Pre Zero
Post Zero
Pre Zero
Pre Zero
Pre Zero
Post Zero
Post Zero
NOx Audit Point #5
NO Audit Point #3NO Audit Point #4
NOx Audit Point #1NOx Audit Point #2
NOx Audit Point #4
SO2 Audit level 2
Converter Efficiency NO2 level 1
NO2 Audit level 4NO2 Audit level 2
Converter Efficiency NO2 level 5Converter Efficiency NO2 level 4Converter Efficiency NO2 level 2
Pre Zero
NO2 Audit level 1
NO2 Audit level 5
SO2 Audit level 1
SO2 Audit level 4
Sulfur Dioxide
SO2 Audit level 6SO2 Audit level 5
Oxides of Nitrogen
NO Audit Point #1NO Audit Point #2
NOx Audit Point #3
NO Audit Point #5
CO Audit level 2CO Audit level 1
FINAL SUMMARY AUDIT REPORT CO BASEDEEMS Van-3
Ozone
ParameterNPAP Lab Response
(ppm)Station Response
(ppm)
-0.5
0.015480.00037 0.00158
-0.6
2.8
Ozone audit level 5Ozone audit level 4
Ozone audit level 2Ozone audit level 3
Pre Zero
Percent Difference
4.9
Actual Difference
(ppm)Warning
CO Audit level 4CO Audit level 3
Carbon Monoxide
0.079990.060410.03729
0.07954Ozone audit level 6
0.06053
Pass/Fail
0.10811
0.2
0.014760.03835
0.10760
CO Audit level 5
Site Name: Shelby Farm - NCORE Airs ID: 471570075Auditor: Eric Hebert Audit Date: 05/29/19Station Manager: Joe Maness / Judy Low
Instrument: Ozone COManufacturer: Thermo TEIModel: 49i 48i TLESerial Number: 1180030022 1503464200Calibration Date: 09/07/18 CalibratedSlope 1.0029 on Intercept 0.0001098 day of audit
STATION INSTRUMENT INFORMATIONInstrument:Manufacturer/Model #: T-API T400UProperty Number:Calibration Date:Slope/Intercept: 1.0036 -0.0040Indicated Flow:In-Line Filter Change:Manifold Type:
Pass/Fail Warning Auditor Eric Hebert
Ozone audit level 6 Pass PrintOzone audit level 5 Pass Ozone audit level 4 Pass SignatureOzone audit level 3 Pass Ozone audit level 2 Pass
EPA person notified in case of audit failureAudit Limits
Pass Bias < ±10% OR difference from actual concentration < 24 hour allowable drift (0.003 ppm) Fail Bias > ±10% AND difference from actual concentration > 24 hour allowable drift (0.003 ppm)
Warning Bias > ±7% AND difference from actual concentration > 0.0015 ppm
Comments:0
0.03729 0.03835 2.8
0.00037 0.00158
0.01476 0.01548 4.9
0.07999 0.07954 -0.6
0.06041 0.06053 0.2
FINAL NPAP THROUGH-THE-PROBE AUDIT REPORT
MOBILE PE LAB INSTRUMENTS
03/20/191/4" Teflon
0.78
266404/19/19
EEMS Van-3
0.10811 0.10760 -0.5
Ozone
OZONE REPORT
NPAP O3 Concentration (ppm)
FINAL OZONE AUDIT RESULTS
Site Response (ppm)
Percent Difference
0.0004 0.00080
Site Name: Shelby Farm - NCORE Audit Date: 5/28/2019
N/AN/AN/AN/AN/A
-0.03407 Pass0.04032 Pass
N/AN/AN/AN/A
0.06391 Pass
N/AN/AN/AN/AN/A
N/AN/AN/AN/AN/A
N/AN/AN/AN/A
N/AN/AN/AN/A
-0.00031 PassN/AN/AN/AN/AN/A
0.00076 Pass-0.00036 0.0004
Post Zero -0.00033
Post Zero -0.00034
Post Zero
Pre Zero
Post Zero
Pre Zero
Pre Zero
Pre Zero
Post Zero
Post Zero -0.00001
NOx Audit Point #5
NO Audit Point #3NO Audit Point #4
NOx Audit Point #1NOx Audit Point #2
NOx Audit Point #4
0.00000
0.000260.13851
0.14092
SO2 Audit level 2
Converter Efficiency NO2 level 1
NO2 Audit level 4NO2 Audit level 2
Converter Efficiency NO2 level 5Converter Efficiency NO2 level 4Converter Efficiency NO2 level 2
Pre Zero
NO2 Audit level 1
NO2 Audit level 5
SO2 Audit level 1
0.00000.000290.15008
SO2 Audit level 4
Sulfur Dioxide
SO2 Audit level 6SO2 Audit level 5
Oxides of Nitrogen-0.0113
0.00914.7507
0.053
-0.025
NO Audit Point #1NO Audit Point #2
NOx Audit Point #3
0.00027
NO Audit Point #5
CO Audit level 2CO Audit level 1
FINAL SUMMARY AUDIT REPORT CO BASEDEEMS Van-3
Ozone
ParameterNPAP Lab Response
(ppm)Station Response
(ppm)
Ozone audit level 4Ozone audit level 3
Ozone audit level 1Ozone audit level 2
Pre Zero
Percent Difference
Actual Difference
(ppm)Warning
CO Audit level 4CO Audit level 3
Carbon Monoxide
Ozone audit level 5
Pass/Fail
4.791CO Audit level 5 0.8
Site Name: Shelby Farm - NCORE Airs ID: 471570075Auditor: Eric Hebert Audit Date: 05/28/19Station Manager: Joe Maness / Judy Low
Instrument: Ozone COManufacturer: 0 TEIModel: 0 48i TLESerial Number: 0 1503464200Calibration Date: 01/00/00 CalibratedSlope 0 on Intercept 0 day of audit
STATION INSTRUMENT INFORMATIONInstrument:Manufacturer/Model #: T-API T300EUProperty Number:Calibration Date:Slope/Intercept: 0.00 0.00Indicated Flow:In-Line Filter Change:Manifold Type:
Pass/Fail Warning Auditor Eric Hebert
CO Audit level 5 Pass PrintCO Audit level 4 N/ACO Audit level 3 N/A SignatureCO Audit level 2 N/ACO Audit level 1 N/A
EPA person notified in case of audit failureAudit Limits
Pass Bias < ±15% OR difference from actual concentration < 24 hour allowable drift (0.4 ppm) Fail Bias > ±15% AND difference from actual concentration > 24 hour allowable drift (0.4 ppm)
Warning Bias > ±10% AND difference from actual concentration > 0.030 ppm
-0.0113 0.053
NPAP CO Concentration (ppm)
Site Response (ppm)
Percent Difference
0.8
-0.0250.0091
4.7507 4.791
FINAL NPAP THROUGH-THE-PROBE AUDIT REPORTEEMS Van-3
MOBIL PE LAB INSTRUMENTS
1/4" Teflon
CO
CARBON MONOXIDE REPORT
24602/27/19
1.574/11/2019
FINAL CARBON MONOXIDE AUDIT RESULTS
Site Name: Shelby Farm - NCORE Audit Date: 5/28/2019
N/AN/AN/AN/AN/A
-0.02023 Pass-0.02000 Pass -0.01767 Pass -0.02471 Pass -0.03309 Pass Warning-0.02611 Pass -0.02114 Pass
N/AN/AN/AN/AN/A
N/AN/AN/AN/AN/A
N/AN/AN/AN/A
N/AN/AN/AN/A
0.00013 Pass-0.00437 Pass -0.00227 Pass -0.00135 Pass-0.00029 Pass-0.00001 Pass0.00031 Pass-0.00012 0.0002
Post Zero -0.00011
Post Zero -0.00011
Post Zero
Pre Zero
Post Zero
Pre Zero
Pre Zero
Pre Zero
Post Zero
Post Zero 0.00000
NOx Audit Point #5
0.00117
0.00119
NO Audit Point #3NO Audit Point #4
0.01647
NOx Audit Point #1NOx Audit Point #2
NOx Audit Point #4
-0.8
0.00000
-0.00014
0.039410.016750.00252
0.077530.03874
0.07887
0.00248
SO2 Audit level 2
Converter Efficiency NO2 level 1
NO2 Audit level 4NO2 Audit level 2
Converter Efficiency NO2 level 5Converter Efficiency NO2 level 4Converter Efficiency NO2 level 2
-10.8
Pre Zero
NO2 Audit level 1
NO2 Audit level 5
SO2 Audit level 10.002690.00127
0.00240.0013
-7.6
0.0000
0.01784
0.07960.03970.0165
-0.000150.084000.04198
-5.2-5.4
SO2 Audit level 4
Sulfur Dioxide
SO2 Audit level 6SO2 Audit level 5
-38.9-65.1
Oxides of Nitrogen-0.0039
-0.00482.65901.3287
-4.40.540
0.04010.0520.014-0.025
-0.025
NO Audit Point #1NO Audit Point #2
NOx Audit Point #3
-0.00014
NO Audit Point #5
CO Audit level 2CO Audit level 1
0.0851
FINAL SUMMARY AUDIT REPORT CO BASEDEEMS Van-3
Ozone
ParameterNPAP Lab Response
(ppm)Station Response
(ppm)
Ozone audit level 4Ozone audit level 3
Ozone audit level 1Ozone audit level 2
Pre Zero
Percent Difference
Actual Difference
(ppm)Warning
CO Audit level 4CO Audit level 3
Carbon Monoxide
Ozone audit level 5
Pass/Fail
0.5648
2.639CO Audit level 4 -0.81.311 -1.3
Site Name: Shelby Farm - NCORE Airs ID: 471570075Auditor: Eric Hebert Audit Date: 05/28/19Station Manager: Joe Maness / Judy Low
Instrument: Ozone COManufacturer: 0 TEIModel: 0 48i TLESerial Number: 0 1503464200Calibration Date: 01/00/00 CalibratedSlope 0 on Intercept 0 day of audit
STATION INSTRUMENT INFORMATIONInstrument:Manufacturer/Model #: T-API T300EUProperty Number:Calibration Date:Slope/Intercept: 0.00 0.00Indicated Flow:In-Line Filter Change:Manifold Type:
Pass/Fail Warning Auditor Eric Hebert
CO Audit level 4 Pass PrintCO Audit level 4 Pass CO Audit level 3 Pass SignatureCO Audit level 2 Pass WarningCO Audit level 1 Pass
EPA person notified in case of audit failureAudit Limits
Pass Bias < ±15% OR difference from actual concentration < 24 hour allowable drift (0.4 ppm) Fail Bias > ±15% AND difference from actual concentration > 24 hour allowable drift (0.4 ppm)
Warning Bias > ±10% AND difference from actual concentration > 0.030 ppm
-1.3
-4.4
-0.0039 -0.025
0.0401 0.014 -65.1
0.0851 -38.9
1.311
0.052
1.3287
0.5648
NPAP CO Concentration (ppm)
Site Response (ppm)
Percent Difference
-0.8
-0.025-0.0048
2.6590 2.639
FINAL NPAP THROUGH-THE-PROBE AUDIT REPORTEEMS Van-3
MOBIL PE LAB INSTRUMENTS
1/4" Teflon
CO
CARBON MONOXIDE REPORT
24602/27/19
1.574/11/2019
FINAL CARBON MONOXIDE AUDIT RESULTS
0.540
Site Name: Shelby Farm - NCORE Airs ID: 471570075Auditor: Eric Hebert Audit Date: 05/28/19Station Manager: Joe Maness / Judy Low
Instrument: Ozone COManufacturer: 0 TEIModel: 0 48i TLESerial Number: 0 1503464200Calibration Date: 01/00/00 CalibratedSlope 0 on Intercept 0 day of audit
STATION INSTRUMENT INFORMATIONInstrument:Manufacturer/Model #: T-API T100EUProperty Number:Calibration Date:Slope/Intercept: 0.000 0.000Indicated Flow:In-Line Filter Change:Manifold Type:
Pass/Fail WarningSO2 Audit level 6 Pass Auditor Eric HebertSO2 Audit level 5 Pass PrintSO2 Audit level 4 PassSO2 Audit level 2 Pass SignatureSO2 Audit level 1 Pass
EPA person notified in case of audit failureAudit Limits
Pass Bias < ±15% OR difference from actual concentration < 24 hour allowable drift (0.003 ppm) Fail Bias > ±15% AND difference from actual concentration > 24 hour allowable drift (0.003 ppm)
Warning Bias > ±10% AND difference from actual concentration > 0.0015 ppm
-5.4
-7.6
NPAP SO2 Concentration (ppm)
0.0013
0.08400
0.0000
-5.2
-0.00012 0.0002
Percent DifferenceSite Response
(ppm)
0.00269
0.00127
0.0024 -10.8
-0.8
SO2
0.63
0.0796
05/22/19
4/10/20191/4" Teflon
FINAL SULFUR DIOXIDE AUDIT RESULTS
-0.00015
0.04198
0.01784
0.0397
0.0165
135
FINAL NPAP THROUGH-THE-PROBE AUDIT REPORTEEMS Van-3
MOBIL PE LAB INSTRUMENTS
SULFUR DIOXIDE REPORT
Site Name: Shelby Farm - NCORE Audit Date: 5/28/2019
N/AN/AN/AN/AN/A
N/AN/AN/AN/AN/A
-0.00008 Pass0.00655 Pass 0.00210 Pass 0.00130 Pass 0.00088 Pass
N/A
-0.00008 Pass0.00280 Pass -0.00006 Pass 0.00024 Pass 0.00012 Pass
N/A
0.00077 Pass-0.00160 Pass -0.00011 Pass -0.00065 Pass -0.00058 Pass
97.2% Pass101.3% Pass97.1% Pass99.6% Pass
Converter Efficiency calculated by OAQPS QA Guidance Doc 2.3 February 2002 = 97.2% Pass
N/AN/AN/AN/AN/A
-0.00035
Post Zero -0.00033
Post Zero -0.00033
Post Zero
Pre Zero
Post Zero
Pre Zero
Pre Zero
Pre Zero
Post Zero
-22.2Post Zero -0.00001
NOx Audit Point #5
NO Audit Point #3NO Audit Point #4
0.02362 5.5
NOx Audit Point #1NOx Audit Point #2
NOx Audit Point #4
2.9
0.01590
-0.11.0
-3.4-0.5
0.8
0.000150.099700.04846
0.00077
5.9
0.00000
0.020210.00445
0.02492
0.00023 0.00015
0.00261 0.00203
4.4
0.048520.024030.01528
0.095250.04770
0.101800.04980
0.09690
0.01502
SO2 Audit level 2
-14.6
0.04712 0.045520.020100.00380
Converter Efficiency NO2 level 1
NO2 Audit level 5NO2 Audit level 2
Converter Efficiency NO2 level 5Converter Efficiency NO2 level 4Converter Efficiency NO2 level 2
Pre Zero
NO2 Audit level 1
NO2 Audit level 5
SO2 Audit level 10.016280.02559
0.000250.103200.05168
SO2 Audit level 4
Sulfur Dioxide
SO2 Audit level 6SO2 Audit level 5
6.9
Oxides of Nitrogen-0.0112
0.00783.26681.6359
NO Audit Point #1NO Audit Point #2
NOx Audit Point #3 0.024270.01540
0.00023
NO Audit Point #5
CO Audit level 2CO Audit level 1
0.5152
FINAL SUMMARY AUDIT REPORT CO BASEDEEMS Van-3
Ozone
ParameterNPAP Lab Response
(ppm)Station Response
(ppm)
Ozone audit level 4Ozone audit level 3
Ozone audit level 1Ozone audit level 2
Pre Zero
Percent Difference
Actual Difference
(ppm)Warning
CO Audit level 4CO Audit level 3
Carbon Monoxide
Ozone audit level 5
Pass/Fail
0.8100
CO Audit level 5
Site Name: Shelby Farm - NCORE Airs ID: 471570075Auditor: Eric Hebert Audit Date: 05/28/19Station Manager: Joe Maness / Judy Low
Instrument: Ozone CO Instrument:Manufacturer: Thermo TEI Manufacturer/Model #: T-API T200EUModel: 49C-PS 48i TLE Property Number:Serial Number: 0517112167 1503464200 Calibration Date:Calibration Date: 09/12/17 Calibrated Slope/Intercept: 0.000 0.000Slope 1.005036 on Indicated Flow:Intercept 0.00032915 day of audit In-Line Filter Change:
Manifold Type:
Pass/Fail Warning WarningPass Pass Pass Pass N/A
Audit Limits
Pass Bias < ±15% OR difference from actual concentration < 24 houFail Bias > ±15% AND difference from actual concentration > 24 ho
Warning Bias > ±10% AND difference from actual concentration > 0.001
Converter Efficiency Audit LimitsPass Between 96% and 104%Fail <96% or >104%
Warning Between 96%-97% or 103%-104%
Pass/Fail Warning Auditor Eric HebertPass PrintPass Pass SignaturePassPassPassPass EPA person notified in case of audit failurePass Conv eff by slope
Pass 97.2%Converter Efficiency NO2 level 1Converter Efficiency NO2 level 2Converter Efficiency NO2 level 4
Converter Efficiency NO2 level 5
NO2 Audit level 5NO2 Audit level 2
NO2 Audit level 5
NO2 Audit level 1
Converter Efficiency by slope =
0.04712
0.00166
97.2%
0.02010 -0.5
0.00445 0.00380 -14.6 97.1%
0.02021
0.04552 -3.4
-0.00170
101.3%
0.00261 0.00203 -22.2 99.6%
NOx Audit Point #5
NPAP NO2 Concentration (ppm)
Site Response (ppm)
Percent Difference
N/A
Pass/Fail
0.0150
NO Audit Point #3
FINAL NO2 AUDIT RESULTS
NO Audit Point #2
NO Audit Point #4 NOx Audit Point #4 Pass
NO Audit Point #1
NO Audit Point #5
Converter Efficiency
PassPassNOx Audit Point #3
NOx Audit Point #2
NPAP NO Concentration (ppm)
Percent Difference
Site Response (ppm)
NPAP NOX CONCENTRATION (ppm)
0.0153
0.0236
-0.0003 -0.0003
0.0249 1.05.5
0.01540.0159 5.9
0.0243
SITE RESPONSE (ppm)
Percent Difference
0.0002
0.0997
FINAL NPAP THROUGH-THE-PROBE AUDIT REPORTEEMS Van-3
MOBIL PE LAB INSTRUMENTSNO/NOX
FINAL NO AUDIT RESULTS FINAL NOx AUDIT RESULTS
1/4" Teflon
05/02/19
NITROGEN OXIDES REPORT
STATION INTRUMENT INFORMATION
209
0.941/0/1900
NOx Audit Point #1
0.0240
0.8
Pass
0.0002
0.0953
0.0498
0.0002
0.1018
0.0477 4.4
0.0969
0.0485
0.0002
2.9
0.0485 -0.1
6.9
Site Name: SWTCC Near Road Audit Date: 5/30/2019
N/AN/AN/AN/AN/A
0.00974 Pass0.07363 Pass
N/AN/AN/AN/A
-0.00902 Pass
0.00097 Pass0.00319 Pass
N/AN/AN/AN/A
-0.00013 Pass
0.00098 Pass0.00016 Pass
N/AN/AN/AN/A
-0.00019 Pass
0.00007 Pass0.00104 Pass
N/AN/AN/A
0.00002 Pass
102.9% PassN/AN/AN/A
N/AN/AN/AN/AN/A
0.00003
Post Zero 0.00003 -0.00010
Post Zero 0.00003 -0.00016
Post Zero
Pre Zero
Post Zero
Pre Zero
Pre Zero
Pre Zero
Post Zero
Post Zero 0.00000
NOx Audit Point #5
NO Audit Point #3NO Audit Point #4
NOx Audit Point #1NOx Audit Point #2
NOx Audit Point #4
0.1
1.3
0.000510.14270
0.00006
0.00002
-0.00001
-0.00046 0.000510.14011 0.14330
0.14254
SO2 Audit level 2
0.08176 0.08280
Converter Efficiency NO2 level 1
NO2 Audit level 4NO2 Audit level 2
Converter Efficiency NO2 level 6Converter Efficiency NO2 level 4Converter Efficiency NO2 level 2
Pre Zero
NO2 Audit level 1
NO2 Audit level 6
SO2 Audit level 1
-0.000500.15181
SO2 Audit level 4
Sulfur Dioxide
SO2 Audit level 6SO2 Audit level 5
2.3
Oxides of Nitrogen0.0010
-0.01594.8054
-0.008
-0.006
NO Audit Point #1NO Audit Point #2
NOx Audit Point #3
-0.00047
NO Audit Point #5
CO Audit level 3CO Audit level 2
FINAL SUMMARY AUDIT REPORT CO BASEDEEMS Van-3
Ozone
ParameterNPAP Lab Response
(ppm)Station Response
(ppm)
Ozone audit level 5Ozone audit level 4
Ozone audit level 2Ozone audit level 3
Pre Zero
Percent Difference
Actual Difference
(ppm)Warning
CO Audit level 4CO Audit level 4
Carbon Monoxide
Ozone audit level 6
Pass/Fail
4.879CO Audit level 5 1.5
Site Name: SWTCC Near Road Airs ID: 471570100Auditor: Eric Hebert Audit Date: 05/30/19Station Manager: Yong Cai / Judy Low
Instrument: Ozone COManufacturer: 0 TEIModel: 0 48i TLESerial Number: 0 1503464200Calibration Date: 01/00/00 CalibratedSlope 0 on Intercept 0 day of audit
STATION INSTRUMENT INFORMATIONInstrument:Manufacturer/Model #: T-API T300UProperty Number:Calibration Date:Slope/Intercept: 1.00 0.01Indicated Flow:In-Line Filter Change:Manifold Type:
Pass/Fail Warning Auditor Eric Hebert
CO Audit level 5 Pass PrintCO Audit level 4 N/ACO Audit level 4 N/A SignatureCO Audit level 3 N/ACO Audit level 2 N/A
EPA person notified in case of audit failureAudit Limits
Pass Bias < ±15% OR difference from actual concentration < 24 hour allowable drift (0.4 ppm) Fail Bias > ±15% AND difference from actual concentration > 24 hour allowable drift (0.4 ppm)
Warning Bias > ±10% AND difference from actual concentration > 0.030 ppm
FINAL CARBON MONOXIDE AUDIT RESULTS
FINAL NPAP THROUGH-THE-PROBE AUDIT REPORTEEMS Van-3
MOBIL PE LAB INSTRUMENTS
1/4" Teflon
CO
CARBON MONOXIDE REPORT
17404/08/19
1.8 lpm5/15/2019
NPAP CO Concentration (ppm)
Site Response (ppm)
Percent Difference
1.5
-0.006-0.0159
4.8054 4.879
0.0010 -0.008
Site Name: SWTCC Near Road Audit Date: 5/30/2019
N/AN/AN/AN/AN/A
0.00622 Pass0.00344 Pass 0.01290 Pass -0.00377 Pass -0.00874 Pass -0.00797 Pass 0.00236 Pass
-0.00008 Pass-0.00017 Pass -0.00003 Pass -0.00041 Pass -0.00046 Pass
N/A0.00020 Pass
-0.00008 Pass-0.00152 Pass -0.00083 Pass -0.00070 Pass -0.00063 Pass
N/A0.00015 Pass
0.000350.00081 Pass -0.00034 Pass -0.00033 Pass 0.00006 Pass0.00003 Pass
102.6% Pass102.7% Pass98.3% Pass114.5% Fail
Converter Efficiency calculated by OAQPS QA Guidance Doc 2.3 February 2002 = 102.4% Pass
N/AN/AN/AN/AN/A
Warning
CO Audit level 4CO Audit level 3
Carbon Monoxide
Ozone audit level 6
Pass/Fail
0.6591
2.681CO Audit level 4 0.11.344
FINAL SUMMARY AUDIT REPORT CO BASEDEEMS Van-3
Ozone
ParameterNPAP Lab Response
(ppm)Station Response
(ppm)
Ozone audit level 5Ozone audit level 4
Ozone audit level 2Ozone audit level 3
Pre Zero
Percent Difference
Actual Difference
(ppm)
NO Audit Point #2
NOx Audit Point #3 0.018850.00421
0.00014
NO Audit Point #5
CO Audit level 2CO Audit level 1
0.1632 -5.4-17.3
-0.2
Oxides of Nitrogen-0.0104
0.00482.67761.3311
-0.60.655
0.04600.1550.038-0.008
0.011
NO Audit Point #1
1.0
SO2 Audit level 10.005160.00145
0.02082
0.000150.084590.04205
SO2 Audit level 4
Sulfur Dioxide
SO2 Audit level 6SO2 Audit level 5
SO2 Audit level 2
-8.3
0.04789 0.048700.018830.00365
Converter Efficiency NO2 level 1
NO2 Audit level 4NO2 Audit level 2
Converter Efficiency NO2 level 5Converter Efficiency NO2 level 4Converter Efficiency NO2 level 2
Pre Zero
NO2 Audit level 1
NO2 Audit level 50.00000
0.019170.00398
0.01881
0.00014 0.00006
0.00155 0.00161
-0.1
0.039480.019550.00484
0.078070.03881
0.077900.03878
0.07942
0.00476
-1.9
0.00430
-2.1-3.6
1.7-1.8
-13.0
0.000060.077900.03865
0.00035
-9.7
0.00002
NOx Audit Point #5
0.00134
0.00136
NO Audit Point #3NO Audit Point #4
0.01922 -2.1
NOx Audit Point #1NOx Audit Point #2
NOx Audit Point #4
-0.00033
Post Zero -0.00030 -0.00010
Post Zero -0.00031 -0.00016
Post Zero
Pre Zero
Post Zero
Pre Zero
Pre Zero
Pre Zero
Post Zero
3.9Post Zero -0.00001
Site Name: SWTCC Near Road Airs ID: 471570100Auditor: Eric Hebert Audit Date: 05/30/19Station Manager: Yong Cai / Judy Low
Instrument: Ozone COManufacturer: 0 TEIModel: 0 48i TLESerial Number: 0 1503464200Calibration Date: 01/00/00 CalibratedSlope 0 on Intercept 0 day of audit
STATION INSTRUMENT INFORMATIONInstrument:Manufacturer/Model #: T-API T300UProperty Number:Calibration Date:Slope/Intercept: 1.00 0.01Indicated Flow:In-Line Filter Change:Manifold Type:
Pass/Fail Warning Auditor Eric Hebert
CO Audit level 4 Pass PrintCO Audit level 4 Pass CO Audit level 3 Pass SignatureCO Audit level 2 Pass CO Audit level 1 Pass
EPA person notified in case of audit failureAudit Limits
Pass Bias < ±15% OR difference from actual concentration < 24 hour allowable drift (0.4 ppm) Fail Bias > ±15% AND difference from actual concentration > 24 hour allowable drift (0.4 ppm)
Warning Bias > ±10% AND difference from actual concentration > 0.030 ppm
FINAL CARBON MONOXIDE AUDIT RESULTS
0.655
EEMS Van-3
MOBIL PE LAB INSTRUMENTS
1/4" Teflon
CO
CARBON MONOXIDE REPORT
17404/08/19
1.8 lpm5/15/2019
NPAP CO Concentration (ppm)
Site Response (ppm)
Percent Difference
0.1
0.0110.0048
2.6776 2.681
1.0
-0.6
-0.0104 -0.008
0.0460 0.038 -17.3
0.1632 -5.4
1.344
0.155
1.3311
0.6591
Site Name: SWTCC Near Road Airs ID: 471570100Auditor: Eric Hebert Audit Date: 05/30/19Station Manager: Yong Cai / Judy Low
Instrument: Ozone CO Instrument:Manufacturer: 0 TEI Manufacturer/Model #: T-API T200UModel: 0 48i TLE Property Number:Serial Number: 0 1503464200 Calibration Date:Calibration Date: 01/00/00 Calibrated Slope/Intercept: 0.999 -0.531Slope 0 on Indicated Flow:Intercept 0 day of audit In-Line Filter Change:
Manifold Type:
Pass/Fail Warning WarningPass Pass Pass Pass N/A
Audit Limits
Pass Bias < ±15% OR difference from actual concentration < 24 houFail Bias > ±15% AND difference from actual concentration > 24 ho
Warning Bias > ±10% AND difference from actual concentration > 0.001
Converter Efficiency Audit LimitsPass Between 96% and 104%Fail <96% or >104%
Warning Between 96%-97% or 103%-104%
Pass/Fail Warning Auditor Eric HebertPass PrintPass Pass SignaturePass PassPassPass EPA person notified in case of audit failureFail Conv eff by slope
Converter Efficiency by slope = Pass 102.4%
NOx Audit Point #1
0.00136
0.01955
-13.0
Pass
NOx Audit Point #5
NPAP NO2 Concentration (ppm)
Site Response (ppm)
Percent Difference
N/A
Pass/Fail
0.00476
NO Audit Point #3
FINAL NO2 AUDIT RESULTS
NO Audit Point #2
NO Audit Point #4 NOx Audit Point #4 Pass
NO Audit Point #1
0.00014
0.07807
0.0388
0.0001
0.0779
0.03881 -0.1
0.07942
0.03948
0.00014
FINAL NPAP THROUGH-THE-PROBE AUDIT REPORTEEMS Van-3
MOBIL PE LAB INSTRUMENTSNO/NOX
FINAL NO AUDIT RESULTS FINAL NOx AUDIT RESULTS
1/4" Teflon
04/08/19
NITROGEN OXIDES REPORT
STATION INTRUMENT INFORMATION
182
1.1 lpm3/21/2019
SITE RESPONSE (ppm)
Percent Difference
0.0001
0.0779
-0.0002
-1.9
0.0387 -2.1
-0.2
PassPassNOx Audit Point #3
NOx Audit Point #2
0.04789
NPAP NO Concentration (ppm)
Percent Difference
Site Response (ppm)
NPAP NOX CONCENTRATION (ppm)
0.00484
0.01922
-0.0001-0.00030 -0.00031
0.0188 -3.6-2.1
0.00420.0043 -9.7
0.0189
0.0
0.0487 1.7
0.0001
102.7%
NO Audit Point #5
Converter Efficiency
0.00134
0.00136
102.6%
0.0188 -1.8
0.00398 0.0037 -8.3 98.3%
0.01917
0.00155 0.0016 3.9 114.5%
Converter Efficiency NO2 level 4
Converter Efficiency NO2 level 5
0.0000
Converter Efficiency NO2 level 1
NO2 Audit level 4NO2 Audit level 2
NO2 Audit level 5
NO2 Audit level 1
Converter Efficiency NO2 level 2
0.00000
Shelby County Meteorological QA Test Results
Unadjusted Temperature
Site: Shelby Farms Date: 29-May-2019Field Scientist : E. Hebert
Site Sensor(s) Mfg.: Climatronics Ref. Std. Mfg.: Extech Ref. Std. Slope: 1.0091Model: 100093 Ref. Std. Model: 407907 Ref. Std. Intercept: -0.1104
Measurement Units: ° F Ref. Std. ID No: EEMS 01228 Certification Date: 2/12/2019
3 m probe Ref. Temp. Raw Ref. Temp. Corr. Site Temp. DAS DifferenceSensor ID. No: J2850C 31.99 31.81 33.2 1.39
81.78 81.15 81.9 0.75115.64 114.71 115.3 0.59
Average 0.91
RemarksFailFailFail
Page 1 of 1
Shelby County Meteorological QA Test Results
Unadjusted Relative Humidity/Barometric Pressure
Site: Shelby Farms Date: 29-May-2019Field Scientist : E. Hebert
Relative Humidity
Site Sensor Make: Met One Ref. Std. Mfg.: AZModel: 083E-0-6 Ref. Std. Model: AZ 8723 Ref. Std. Slope: 1.0154
Serial No: T25685 Ref. Std. ID No: EEMS 01223 Ref. Std. Intercept: -1.3456
Salt Solution Ref. Raw Ref. Corr. Site DAS Difference Remarks% RH % RH % RH % RH % RH
58.5 58.9 61.7 2.8 PASS
Average 2.76
Barometric Pressure
Site Sensor Make: RM Young Transfer Make BGIModel: Unknown Model DeltaCal Ref. Std. Slope: 1.0000
Serial No: Unknown Serial No: EEMS 01451 Ref. Std. Intercept: 0.0000
Ref. BP Raw Ref. BP Cor. Site BP DAS Error RemarksmmHg mmHg749.5 749.5 751.6 2.1 PASS
Page 1 of 1
APPENDIX B
Maps of Locations
Topo North America™ 10
Data use subject to license.
© DeLorme. Topo North America™ 10.
www.delorme.com
TN
MN (1.4°W)0 1 2 3 4 5
0 1 2 3 4 5
mikm
Scale 1 : 200,000
1 cm = 2.00 km Data Zoom 10-0
Topo North America™ 10
Data use subject to license.
© DeLorme. Topo North America™ 10.
www.delorme.com
TN
MN (1.5°W)0 400 800 1200 1600 2000
0 160 320 480 640 800
ftm
Scale 1 : 17,600
1 cm = 176.0 m Data Zoom 13-5
Topo North America™ 10
Data use subject to license.
© DeLorme. Topo North America™ 10.
www.delorme.com
TN
MN (1.3°W)0 600 1200 1800 2400 3000
0 200 400 600 800 1000
ftm
Scale 1 : 22,400
1 cm = 224.0 m Data Zoom 13-2
APPENDIX C
Audit Standards Certifications
Date2/12/2019 - - Calibration and verification of three RTD meters with most recent certification of EEMS RTD
RTD RTD RTDAt Date 01230 / 01231 01227 / 1 01228 / 2
TMI EEMS 2/12/2019 EEMS EEMS EEMSSTD AER van3 van1
Cert # A2380069diff corrected raw corrected raw corrected raw corrected raw corrected
-25.00 -25.02 0.020 -25.016 0.02 0.03 0.04 0.01 0.15 0.00 -0.09 0.020.00 0.01 -0.010 0.016 10.32 10.33 10.35 10.34 10.56 10.09 10.31 10.57
100.00 99.99 0.010 100.007 21.10 21.11 21.12 21.12 21.43 20.89 21.19 21.35150.00 149.97 0.030 149.993 30.30 30.31 30.32 30.33 30.67 30.30 30.47 30.30
40.00 40.01 39.98 40.00 40.46 40.02 40.28 40.0347.91 47.92 47.89 47.92 48.40 47.90 48.23 47.9025.00 25.01 25.00 25.00 25.34 25.01 25.14 25.02
2019 correction: slope= 0.99989313intercept= -0.0064885
corr= 1.0000000
0.998872 1.007333 1.0090922/12/2019 0.026147 0.144973 -0.11036
1.0000 1.0000 1.0000
RTD 01229
slope =intercept =
correlation =
TMI Cert data -- 1/23/2019
EEMSRTD RTD
01229 01229
Field Scientist : E. HebertSite ID: Shelby FarmsAudit/Calibration Date: 5/29/2019
Audit Equipment List Assigned to: Van 1
Reference Standard Mfg. Model SN ID Number Slope Intercept Certification DateMultimeter 1 0Voltage Source 1 0RTD (Temperature Reference) Extech 407907 EEMS 01228 1.009092 -0.11036 2/12/2019Hygrometer (RH Reference) AZ AZ 8723 EEMS 01223 1.0154 -1.3456 2/6/2019Pressure Reference BGI DeltaCal EEMS 01451 1 0 2/15/2019Wind Speed Motor - Low RM Young 18802 EEMS 01254 1 0 7/11/2018Wind Speed Motor - High RM Young 18811 EEMS 01255 1 0 7/11/2018Solar Radiation Translator 1 0Solar Radiation Sensor 1 0Compass Ushikata 190037 EEMS 01272 1 0 1/23/2019Hi-vol orifice (QA) Tisch TE-5028A 2927 EEMS 01462 0.99477 -0.00057 4/8/2019Hi-vol orifice (QS) Tisch TE-5028A 2927 EEMS 01462 1.58863 -0.00092 4/8/2019Manometer EEMS 01461 1.00000 0.00000GPS Delorme PN-60 EEMS 01006 1 0
Field Scientist Certification
Eric Hebert
Has satisfactorily completedThe US Environmental Protection Agency’s
“National Performance Audit Program (NPAP)Field Scientist Re-certification Course”
Office of Air Quality Planning and Standards
Research Triangle Park, NC
Course Dates: April 13-14, 2017
Gregory W. NoahNPAP National Coordinator
USEPA, OAQPS, AAMG