25369407(liposomas)
-
Upload
erick-juarez -
Category
Documents
-
view
212 -
download
0
Transcript of 25369407(liposomas)
-
7/30/2019 25369407(liposomas)
1/12
ORIGINAL ARTICLE
Spatial dose distributions in solid tumors from 186Re
transported by liposomes using HS radiochromic media
Luis A. Medina & Beth Goins &
Mercedes Rodrguez-Villafuerte & Ande Bao &
Arnulfo Martnez-Davalos & Vibhudutta Awasthi &
Olga O. Galvn & Cristina Santoyo & William T. Phillips &
Mara-Ester Brandan
Received: 27 April 2006 / Accepted: 20 September 2006 / Published online: 8 February 2007# Springer-Verlag 2007
Abstract
Purpose A procedure for the measurement of spatial dose
rate distribution of beta particles emitted by 186Re-lipo-
somes in tumoral tissue, using HS GafChromic films, is
presented.
Methods HNSCC xenografts were intratumorally injected
with 3.7 or 11.1 MBq of 186Re-liposomes, and planar
gamma camera images were acquired to determine the
liposome retention in the tumor. After imaging, rats were
sacrificed and tumors were excised and processed in slices;
HS film sections were placed between slices and the tumor
lobe was reassembled. Tumors and films were kept in the
dark at 4C for 18 h. After irradiation, films were removed
and response was read using a transmission scanner. Films
were analyzed to determine two-dimensional spatial dose
rate distributions and cumulative dose volume histograms.
Dose rate distributions were quantified using a 60Co
calibration curve, the 186Re physical half-life, and a
perturbation factor that takes into account the effect of the
film protective layer.
Results Dose rate distributions are highly heterogeneous
with maximal dose rates about 0.4 Gy h1 in tumors
injected with 3.7 MBq and 1.3 Gy h1 in tumors injected
with 11.1 MBq. Dose volume histograms showed dose
distributed in more than 95% and 80% of the tumor when
injected with the lower and the higher activity, respectively.
Conclusion The described procedures and techniques have
shown the potential and utility of HS GafChromic film for
determination of dose rate distributions in solid tumors
injected intratumorally with 186Re-liposomes. The films
structure and the liposomes biodistribution must be taken
into account to obtain quantitative dose measurements.
Keywords Radionuclide therapy. Liposomes .186
Re .
HSGafchromicfilm . Intratumoraldosimetry. MCsimulation
Introduction
The interest in the use of liposomes as carriers of
therapeutic drugs and radionuclides has increased in recent
years [18]. Several investigators have reported the
advantages of using intravenously injected liposomes to
transport and specifically deliver therapeutic drugs intumors of hematopoietic origin [1, 5, 6, 9], and for the
treatment of solid tumors [3, 6, 10]. However, a major
limitation for the clinical application of chemotherapeutic
and radiotherapeutic liposomes in solid tumors has been
their insufficient penetration and, thus, their inhomoge-
neous distribution within the tumor tissue [2, 11, 12]. In the
case of radiotherapeutic liposomes, although the physical
characteristics of the radionuclide carried by the liposomes
could limit the level of variability in the spatial deposition
of the dose inside the tumor, the penetration and distribu-
Eur J Nucl Med Mol Imaging (2007) 34:10391049
DOI 10.1007/s00259-006-0297-x
L. A. Medina (*) : M. Rodrguez-Villafuerte :A. Martnez-Davalos : O. O. Galvn : M.-E. BrandanInstituto de Fisica, UNAM, A.P. 20-364,Mexico 01000 D.F., Mexico
e-mail: [email protected]
B. Goins : A. Bao : V. Awasthi : C. Santoyo : W. T. PhillipsDepartment of Radiology,
University of Texas Health Science Center at San Antonio,
San Antonio, TX, USA
A. Bao
Department of OtolaryngologyHead and Neck Surgery,
University of Texas Health Science Center at San Antonio,
San Antonio, TX, USA
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/30/2019 25369407(liposomas)
2/12
tion of the liposomes within the tumor still play an
important role in the spatial distribution of the dose.
An alternative option for an interstitial pressure-related
strategy for treatment of locally confined solid tumors, as in
the case of head and neck tumors, would be to inject the
liposomes directly into the tumor. This should increase
the internal pressure at the core of the tumor relative to the
surrounding tumor tissue, allowing the liposomes to spreadat the interior of the tumor [11]. In this manner a
radionuclide carried by the liposomes could gradually
irradiate the tumor depending only on the type and energy
spectrum of the emitted ionizing particles, producing a
better spatial dose deposition and a higher intratumoral
absorbed dose as compared with the intravenous injection
of the same kind of radiolabeled liposomes [13].
Our groups are interested in evaluating the therapeutic
use of a radionuclideliposome system directly injected in
solid tumors, based on direct dosimetric measurements.
Recently, a practical method for labeling preformed lipo-
somes by loading the therapeutic beta-emitting radionuclide186Re (T1/2=89 h; Emax=1.069, 0.932 MeV; Eave=0.359,
0.306 MeV; mean range 1.8 mm) [14] into the liposome
interior aqueous space was reported [15].
In the present work, we report first results describing a
dosimetric procedure based on the use of HS GafChromic
dosimetric media to determine the spatial distribution of
absorbed doses imparted by 186Re in tumor xenografts,
using an experimental model of head and neck cancer in
nude rats by means of intratumoral injection of 186Re-
liposomes. The features of high spatial resolution, weak
energy dependence, and good tissue equivalence of
GafChromic films have made this radiochromic material
useful for monitoring energy deposition in dosimetric
procedures with high dose-gradient radiation fields [16],
as is the case in radionuclide therapy with 186Re. HS films,
with an active radiochromic layer sandwiched between two
sheets of transparent 97-m-thick polyester, were designed
to increase dose response to high-energy photon beams
(>1 MeV) [17] in a dosimetric range of 0.540 Gy. HS film
contains the same active component as the previous MD-55
media, but has a higher sensitivity because of a thicker
active layer, 40 m in HS versus 15 or 30 m in MD-55-1
or MD-55-2, respectively [17; D. Lewis, 2006, personal
communication]. Thus, basic interactions between ionizing
radiation and the basic active components are the same for
both types of film, and the known dosimetric properties of
MD-55 films should apply to GafChromic HS media [18;
D. Lewis, 2006, personal communication]. Some inves-
tigators have reported the useful application of HS films in132Ir brachytherapy dosimetric procedures [19, 20]. Also,
GafChromic films have been used previously by Mayer et
al. [21] to determine beta-particle dose rate distributions in
tumor sections previously injected with 90Y. The present
report describes the dosimetric use of the high-sensitive HS
film to determine dose rate distributions in tumor sections
injected with 186Re-labeled liposomes.
Materials and methods
SCC-4 cell line
A human origin head and neck squamous cell carcinoma
(HNSCC) cell line, SCC-4, was originally obtained from
American Type Culture Collection (ATCC, Manassas, VA)
and provided by Martin Thornhill, PhD, of the University
o f Tex as Health S cience C en ter at S an An to nio
(UTHSCSA), School of Dentistry. SCC-4 cells were
cultured in Dulbeccos modified Eagles medium (Invitro-
gen, San Diego, CA) with 10% fetal bovine serum, 2 mM
L-glutamine, and antibiotics, penicillin, and streptomycin,
in 10020 mm2 Corning cell culture dishes (Corning Inc,
Corning, NY) at 37C in 5% CO2. When near confluence,cells were trypsinized and collected to determine the
number of cells and viability by trypan blue dye exclusion
assay. The appropriate volume of cell suspension was
transferred to a new tube and centrifuged in an Allegra 21R
Centrifuge (Beckman Coulter, Fullerton, CA) at 800 rpm at
4C for 5 min. Following aspiration of the supernatant, the
cell pellets were diluted with saline (Abbott Laboratories,
Abbott Park, IL) to a concentration of 5106 cells in 0.2 ml
saline. Aliquots of 0.2 ml cell suspension were drawn into
tuberculin syringes and used for subcutaneous inoculations
in nude rats.
HNSCC xenografts in nude rats
Animal experiments reported in the present paper were
performed according to the NIH Animal Use and Care
Guidelines and were approved by the UTHSCSA Institu-
tional Animal Care Committee. During all procedures, the
animals were anesthetized with 13% isoflurane in 100%
oxygen using an anesthesia inhalation machine (Bickford,
Wales Center, NY). Male rnu/rnu athymic nude rats
(Harlan, Indianapolis, IN) at 45 weeks of age (75100 g)
were inoculated subcutaneously with 5106 SCC-4 tumor
cells in 0.20 ml of saline on the dorsum at the level of the
scapulae. The animals were fed and housed following the
protocol for the nude rat living environment at the animal
facility at the UTHSCSA, and were checked daily after
tumor cell inoculation. When the growing tumor in each
animal was palpable and of sufficient size to be measured,
the tumor size was obtained via caliper by measuring the
length (l), width (w), and thickness (t) of each tumor. The
tumor volumes were subsequently calculated using
the ellipsoid volume formula, V =6 l w t [22].
1040 Eur J Nucl Med Mol Imaging (2007) 34:10391049
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/30/2019 25369407(liposomas)
3/12
Preparation of liposomes
Liposomes comprised distearoyl phosphatidylcholine
(DSPC) (Avanti Polar Lipids, Pelham, AL), cholesterol
(Calbiochem, San Diego, CA), and -tocopherol (Aldrich,
Milwaukee, WI) having a molar percentage of 55:44:1. A
dried film was formed by mixing the lipid ingredients in
chloroform and then removing the chloroform by rotaryevaporation and vacuum desiccation for 24 h. The dried
lipid film was rehydrated with 300 mmol/l sucrose (Sigma,
St. Louis, MO) in sterile water for injection (120 mmol/l total
lipid) and warmed to 55C and then lyophilized over-
night. The dried lipidsucrose mixture was rehydrated
with 200 mmol/l reduced glutathione (GSH) (Sigma) in
Dulbeccos phosphate-buffered saline (PBS) (pH 6.3).
Before extrusion, the lipid suspension was diluted to
40 mol/ml with 100 mM GSH in PBS containing
150 mM sucrose. The diluted lipid suspensions were then
extruded through a series of Whatman Nucleopore
polycarbonate filters (Florham Park, NJ) (two passes,2 m; two passes, 400 nm; two passes, 200 nm; five
passes, 100 nm) at 55C using Lipex Extruder (Northern
Lipids, Vancouver, Canada). The extruded lipid solution
was then stored in the refrigerator at 4C for up to 6
months until needed for radiolabeling studies. Following
manufacture, the liposome particle diameter was mea-
sured using a 488-nm laser light scattering instrument
(Brookhaven Instruments, Holtsville, NY). The measured
diameter was 122.15.5 nm prior to the radiolabeling
experiment.
Preparation of 186Re-BMEDA
An SNS pattern ligand, N,N-bis(2-mercaptoethyl)-N,N-
diethylethylenediamine (BMEDA) was synthesized from a
modification of the method by Corbin et al. [23] and the
chemical structures were verified using 1H/13C NMR.186Re-aluminum perrhenate [186Re-Al(ReO4)3] was
obtained from Missouri University Research Reactor
(Columbia, MO). Stannous chloride (Aldrich) was used as
the reductant and glucoheptonate (GH) (Sigma) was used as
an intermediate ligand to make 186Re-BMEDA as previ-
ously described by Bao et al. [15]. BMEDA (1.0 l;
1.1 mg) was pipetted into a new vial. Then, 0.50 ml of
0.17 mol/l GH-0.10 mol/l acetate solution, pH 5.0, was
added, followed by the addition of 60 l of stannous
chloride (15 mg/ml). The pH of the solution was adjusted to
5.0 with 1.0 MNaOH. After flushing the solution with N2gas, 1.3 GBq (35.1 mCi), 0.70 ml of 186Re-Al(ReO4)3(0.4 g Re per mCi 186Re) was added. The vial was sealed
and heated in an 80C water bath for 1 h. The resultant186Re-BMEDA was used directly for the following 186Re-
liposome labeling.
Radiolabeling of preformed liposomes by loading 186Re
into liposome aqueous space
Liposomes were prepared with 1.2 ml of GSH neutral
liposomes (60 mM of total lipids) first undergoing
separation via Sephadex G-25 column chromatography
(PD-10 column, Amersham Biosciences, Piscataway, NJ)
eluted with PBS, pH 7.4. The product was added to the186Re-BMEDA solution and incubated at 37C for 1 h.
Again, the 186Re-liposomes were separated from any free186Re-BMEDA via Sephadex G-25 column chromatogra-
phy eluted with PBS buffer, pH 7.4. The labeling efficiency
(40%) was determined from the186
Re activity before and
after separation using a Radix dose calibrator.
Intratumoral administration of186
Re-liposomes
Two groups of nude rats (n=3/group) aged 45 weeks (75
100 g) and bearing a HNSCC xenograft with an average
volume of 2.00.4 cm3 were injected with 0.20.05 ml of3.7 MBq (0.1 mCi) (3.7-MBq group), or 0.60.05 ml of
11.1 MBq (0.3 mCi) (11.1-MBq group) of186Re-liposomes,
which contained 4.00.7 mg of DSPC and cholesterol. The186Re-liposome doses were equally divided and delivered to
several separate locations of the tumors, trying to cover the
central core of the tumor via intratumoral injections.
Image acquisition and biodistributions
Static planar images at baseline and 3 h after injection were
acquired using a micro-SPECT/CT scanner in a 565656
image matrix, with a pixel size of 2 mm (XSPECT, Gamma
Medica, Northridge CA). Images of the 186Re-liposomes
were acquired in a 186Re gamma-ray window (137 keV
15%) using a high-resolution, small field of view, parallel-
hole collimator (resolution
-
7/30/2019 25369407(liposomas)
4/12
the baseline image to determine the number of counts
detected in the body image (N0). Proportionality between
injected activity in the whole body (A0) and the number of
counts detected in the body image (N0) is assumed. The
percentage number of counts at some specific time t
accumulated in the ROI (%N(t)) was calculated as the ratio
of the counts in the ROI (NROI(t)) and the total number of
counts detected in the whole body at baseline, multiplied by100 (%N(t)=100 NROI(t)/N0).
Tumor sectioning and irradiation
After imaging, the rats were euthanized by cervical disloca-
tion under anesthesia. Tumor lobes were excised, washed in
saline solution, weighed and measured for volume determi-
nation, and embedded in agarose (3%) (USB, Cleveland,
OH) solution in TBE buffer concentrate (10) (USB,
Cleveland OH) to form a rectangular solid matrix with
33 cm2 transverse section. The agarose matrix helped to
provide support during the sectioning of the tumors into3-mm-thick slices. Dosimetric HS GafChromic film squares
(33 cm2) (ISP Technologies Inc, Wayne NJ; Lot
M0525HS) were placed between slices, and the tumor lobe
was carefully reassembled to its original shape, maintaining
the match between tumor slices. When necessary, thin ink
marks were drawn on the dosimetric film to provide
registration of the tissue outline. Tumors and films were
kept in the dark at 4C for 18 h. This time was selected as
practical for the performance evaluation of the HS Gaf-
Chromic film dosimetric procedure proposed in this work.
Dosimetry media handling and readout
GafChromic films were handled according to recommended
procedures [16] aimed at obtaining reproducible results of
the highest sensitivity. In particular: the films were kept in a
dark environment and, during handling, efforts were made
to reduce exposure to fluorescent light; films were cut 24 h
prior to irradiation in order to stabilize their response after
mechanical stress; calibration was performed using a
uniform field delivering dose in the range of interest;
readout took place 11 days after irradiation (sufficient to
reach color stabilization); and a film scanning procedure
was followed [24] to assure reproducibility.
After removal from the tumors, each film was washed
with water using cotton swabs to remove blood or
radioactive residuals. After washing, films were inspected
to assure that there was no sign of damage in their surface
or presence of radioactive residuals. No signs of surface
changes due to the direct contact with tissue and agarose
were observed in the films.
Film response was read 11 days after the end of
irradiation employing an Agfa DuoScan T1200 scanner in
transmission mode using a 36-bit RGB (12 bits per color)
and saved as tagged image file format. The optical density
range of the scanner was set to a maximum with all filters
and image enhancement options turned off. Two scanning
resolutions were used, 200 dpi for films uniformly
irradiated (calibration) and 300 dpi for films irradiated with
a highly non-uniform source (tumor irradiation). Every film
was scanned using an opaque frame to minimize lightcontributions from areas other than the film. Images were
analyzed by an in-house-written image manipulation
routine using MatLab 7.0 (Math Works, Natick, MA) to
obtain the red, blue, and green components of the image.
The film response, R, was quantified as:
R log10Cni
Ci
where Cni and Ci are the measured color levels of the
background (non-irradiated) and irradiated films, respectively
[24].
Film calibration
Film calibration between 0 and 40 Gy was performed using60Co gamma rays under conditions of charged-particle
equilibrium. Calibration readout took place 11 days after
exposure to match the conditions in the tumor irradiation
study.
It is known [16] that for photon energies in the range
0.11.33 MeV, and for secondary electrons from 0.1 to
1.0 MeV, photon mass energy absorption coefficients and
electron mass stopping powers for the active element in the
radiochromic films are within 2% of those for water. Monte
Carlo (MC) estimates (see Fig. 1) for the slowing-down
electron spectrum on the film active layer for a typical
arrangement used in this experiment show that15% of the
electrons have energies below 0.1 MeV. At these low
energies the electron mass stopping powers for the film are
within 4% of those for water, and this energy dependence
should have a minor effect on the water equivalence of the
film.
With respect to the use of 60Co gamma rays to calibrate
the response of the slowing-down beta spectrum, we have
based our technique on the analysis of Fig. 6 in reference
[16]. This graph shows the experimental energy depen-
dence in the response of MD-55-2 films to photons below
0.1 MeV, and the good data description by the normalized
mass energy absorption coefficient ratio (film sensitive
material to water). This agreement implies that the observed
dependence is totally accounted for by the initial photon
interactions. Therefore, one can conclude that a possible
energy dependence of the film response to the secondary
electrons, within the energy interval from about 15 to
1042 Eur J Nucl Med Mol Imaging (2007) 34:10391049
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/30/2019 25369407(liposomas)
5/12
1,250 keV, is negligible. This has led us to assume that
calibration with 1.25-MeV gamma rays should be appro-
priate for the electron spectrum generated in the medium by
the 186Re beta particles (an MC estimate indicates that 98%
of the secondary electrons reaching the film active layer
have energies above 15 keV).
Effect of the HS film design on the dose measurements
The dosimetric medium studied in this work has a very
simple structure, as compared with other radiochromic
films [16]. However, its use for a continuous electron
spectrum, as in this case, requires an analysis of the
possible perturbation caused by the detector features on its
response as a dosimeter. The 49-m-thick sensitive layer,
where the deposited energy induces the visible change in
color, is inserted between two 97-m inert protective
layers, where no effect is induced. Figure 1 shows MC
results for the transport of beta particles from 186Re diluted
in a 3-mm-thick water layer (representing the tumor). The
geometry simulates the situation of a hypothetical central
detection plane (dashed line) between two tumor slices.
Figure 1 displays the slowing-down electron spectra
incident on A) a central plane in direct contact with both
water layers, B) the central plane in a 49-m-thick
radiochromic film without protective layers, and C) the
central plane in a realistic HS film. As seen in the
spectra, the main effect of adding material to the ideal
plane is to reduce the number of electrons without a
substantial change in the spectral shape. Calculation of
the absorbed dose at the sensitive layer indicates that,
because of the protective layers, the HS film detects 68%
of the dose that would be recorded in the hypothetical
situation depicted in Fig. 1a. The perturbation factor
(0.68)1 will be used as a correction to the dose measured
by the film. This effect has no influence at calibration
owing to the low absorption of the 60Co gamma rays.
Calibration curves were used to transform film response
into integrated dose received by the film during the 18 h of
dosimetric measurement. Doses were transformed into
initial dose rates at the tumor slice plane, assuming a fixed
liposome distribution inside the block, and taking into
account the 186Re physical half-life and the protective layer
perturbation factor.
Energy (MeV)
0.0 0.2 0.4 0.6 0.8 1.0
Intensity(a.u.)
0
500
1000
1500
2000
2500
a) 3 mm water - 3mmwater
c) 3 mm water - HS film - 3 mm water
b) 3 mm water - active layer - 3 mm water
Capa activa
Active layer
Water
Water
HS film
a
b
c
186Re
Fig. 1 MC simulations (106 histories each) of the electron spectra at the central plane of the three arrangements indicated in the right-hand
diagrams, as explained in the text
Eur J Nucl Med Mol Imaging (2007) 34:10391049 1043
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/30/2019 25369407(liposomas)
6/12
Results
Figure 2 shows planar gamma camera images of two tumor-
bearing rats acquired at baseline and 3 h after intratumoral
injection with 3.7 or 11.1 MBq (0.1 and 0.3 mCi) of 186Re-
liposomes. The image analysis indicates that, at baseline,
506% of the injected material is retained in the tumor; 3 h
later, the liposome retention in tumor is similar, 456%. Interms of retained activity at 3 h, these values correspond to
1.7 MBq (0.045 mCi) and 5.0 MBq (0.135 mCi) of the186Re activity for the 3.7-MBq (0.1-mCi) and 11.1-MBq
(0.3-mCi) groups, respectively. By biodistribution calcula-
tions we have estimated the percentage injected activity per
organ as 483 and 4312, for the 3.7- and 11.1-MBq
groups, respectively, and the percentage injected activity
per gram as 153 and 2411, respectively, for the same
groups. The rest of the 186Re-liposomes move into blood
circulation and other organs (see Table 1). These observa-
tions are consistent with recent animal studies using
intratumoral injection of99mTc-liposomes of the same lipidformulation, showing that the amount of liposomes which
stays in the tumor after the immediate release remains in the
tumor for prolonged times [13]. Bao et al. [13] reported
approximately 44% retention of liposomes in the tumor 3 h
after injection, which was similar to our value. This
agreement is to be expected since biodistribution results
should depend only on the liposome formulation and not on
the carried radionuclide.
Figure 3 shows the calibration response curve of the HS
film as a function of absorbed dose in water. The response
of the film is shown in the three colours of the RGB
mode. Blue and green components show a linear
behavior throughout the dose range, suggesting thepossibility of extending the dynamic range of response
to doses higher than the nominal 40-Gy limit suggested
by the manufacturer (ISP Technologies Inc.). The red
component displays the most sensitive response, with
linear behavior at doses
-
7/30/2019 25369407(liposomas)
7/12
Figure 5 shows the two-dimensional spatial dose rate
distribution at excision time in six tumor slices from a rat of
the group injected with 11.1 MBq. These images are
representative of the distribution encountered in the other
animals. Dose rate values in each isodose (in this case, iso-
dose-rate) contour is denoted by a color as indicated by the
scale next to the graph. In both experimental groups it was
observed that the two-dimensional spatial distributions are
highly heterogeneous, with maximal dose rates (in some
hot spots) of about 0.4 Gy h1 in those tumors injected with
3.7 MBq and about 1.3 Gy h1 in tumors injected with 11.1
MBq. The lowest heterogeneity is observed in the central
slices (3 and 4).
In Fig. 6, a one-dimensional scan in horizontal and
vertical directions across the center of the two-dimensional
dose rate distributions in a central 11.1-MBq tumor slice
(slice 4 in Fig. 5) is presented. The figure shows relative
dose rate percentages along a single line through the center
of the horizontal and vertical directions. From representa-
tive scans in both groups of tumors it is possible to
conclude that the dose distributions are highly non-uniform
and non-symmetrical with respect to the center of the
tumor, reaching the highest doses in several hot spots
within the tumor volume, possibly associated with the site
of injection. Significant dose gradients within distances
smaller than 2 mm could be identified in the films.
Comparing dose distributions among the different tumors,
significant differences (larger than 30%) in the areas
covered by the iso-dose-rate curves of 0.08 Gy h1 were
observed.
In order to display the heterogeneity in the dose
distribution, Fig. 7 shows a cumulative dose volume
histogram for a tumor from each group. The tumor dose
was calculated from initial dose rate values in Fig. 5 (at
the time of sacrifice), integrated from baseline to 89 h (one186Re physical half-life), and assuming a constant liposome
retention in the tumor (equal to that at 3 h), as anapproximation to Baos measurements [13]. The histogram
for the 3.7-MBq group shows doses between 0 and
17 Gy heterogeneously distributed in more than 95% of
the tumor, and doses between 17 and 32 Gy in less than
5% of the volume. For the 11.1-MBq group, the
histogram shows doses between 0 and 17 Gy heteroge-
neously distributed in more than 80% of the tumor, and
doses from 17 to 32 Gy in less than 20% of the tumor
volume. Similar behavior was observed in the rest of the
tumors of these groups.
Dose (Gy)
0 10 20 30 40
Re
sponse
0.0
0.2
0.4
0.6
0.8
1.0
1.2
Red
Green
Blue
GafChromic HS
Fig. 3 Calibration curve response of the HS GafChromic film as a
function of dose. Calibration was performed with a source of 60Co
under charged-particle equilibrium conditions
Table 1 Biodistribution results
Values represent the
mean sem (n=3)
Organ 11.1-MBq group 3.7-MBq group t
% Inj. activity/organ % Inj. activity/g % Inj. activity/organ % Inj. activity/g t
Blood 20.64.14 1.50.34 21.03.25 1.50.21 tSkin 1.60.18 0.050.01 1.40.08 0.040.01 tHeart 0.10.01 0.10.02 0.10.005 0.10.01 tLiver 8.81.02 0.90.12 8.30.68 0.80.05 tSpleen 4.71.08 7.32.86 3.80.29 4.31.24 tKidney 0.90.09 0.90.10 0.80.03 0.80.03 tBowel 2.40.34 0.20.03 1.90.17 0.20.01 tBladder 1.20.20 1.80.42 0.60.16 1.20.17 tTestes 0.030.001 0.020.001 0.030.005 0.020.01 tSkin of neck 7.21.88 3.00.85 2.00.46 0.90.14 tBone 3.50.52 0.10.02 2.20.35 0.10.01 tMuscle 1.30.18 0.010.001 1.30.13 0.010.01 tStomach 0.60.12 0.10.01 0.60.01 0.10.02 tCecum 1.70.24 0.30.04 1.50.05 0.20.01 tLung 0.40.11 0.30.08 0.40.04 0.30.02 t
Eur J Nucl Med Mol Imaging (2007) 34:10391049 1045
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/30/2019 25369407(liposomas)
8/12
Discussion
Success of radionuclide therapy for solid tumors with the
use of radiolabeled carrier systems, such as 186Re-lipo-
somes, is determined by an efficient delivery of cytotoxic
levels of absorbed dose throughout all neoplastic cells
within the tumor, with minimal irradiation of the surround-
ing normal tissues [4, 25]. However, factors such as the
limited tumoral penetration of the radiolabeled carrier
system, non-uniform activity distribution inside the tumor,
and the physical characteristics of the transported radionu-
clide lead to subcytotoxic absorbed doses of radiation to
neoplastic cells in the central region of the tumor, which
could result in sublethal damage repair [26, 27].
y(cm)
-1.0
-0.5
0.0
0.5
1.0
y(cm)
-1.0
-0.5
0.0
0.5
1.0
x (cm)
-1.0 -0.5 0.0 0.5 1.0
y(cm)
-1.0
-0.5
0.0
0.5
1.0
y(cm)
-1.0
-0.5
0.0
0.5
1.0
y(cm)
-1.0
-0.5
0.0
0.5
1.0
1
2
3
4
5
0.44
0.881.32
0.18
0.26
0.09
Dose rate
(Gy/h)
x (cm)
-1.0 -0.5 0.0 0.5 1.0
y(cm)
-1.0
-0.5
0.0
0.5
1.0
6
Fig. 5 HS films showing six
spatial dose distributions of
tumor slices from a tumor lobe
injected with 11.1 MBq of186Re-liposomes, and the
two-dimensional dose rate dis-
tributions calculated from the
digitalization of the HS films, as
described. Each iso-dose-rate
contour is denoted by a color as
indicated by the color-dose rate
scale. Films were irradiated
between tumor slices for 18 h
Fig. 4 Photographs of a) an
excised tumor, b) the agarose
matrix where the tumor was
embedded with the HS films, c)
the same agarose matrix viewed
from the bottom, and d) tumor
slices and films after removal
from the agarose matrix
1046 Eur J Nucl Med Mol Imaging (2007) 34:10391049
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/30/2019 25369407(liposomas)
9/12
In order to achieve a better intratumoral penetration that
could result in a more homogeneous dose deposition inside
the tumor, intratumoral injection of the radionuclide system
could offer an alternative treatment option. However, a
better understanding of the dose deposition after intra-
tumoral injection of a radiolabeled carrier is necessary. A
goal of this study has been to describe and evaluate spatial
dose distributions in solid tumors from the direct measure-ment of dose rate in explanted tumoral sections of HNSCC
xenografts injected with 186Re-liposomes, using the recent-
ly available HS GafChromic film and digital image analysis
to evaluate the spatial dose distributions.
Image analysis during the first 3 h after186
Re-liposome
injection (Fig. 2 and ref. [13]) suggested that about 55% of
the injected activity was removed very quickly from the
tumor. It is known that pressure in the interstitial tissue of
solid tumors is uniformly elevated except in the outermost
rim, where it drops precipitously [11]. An intratumoral
injection of186Re-liposomes in the core of the tumor would
increase the pressure at that location, relative to thesurrounding tumor tissue, spreading the 186Re-liposomes
along the induced pressure gradient formed by convection
from the surrounding region to the periphery. In the
experiments reported here, HNSCC xenografts with vol-
umes of 2.030.37 cm3 and mass of 2.90.6 g were in-
jected with either 3.7 MBq/0.2 ml or 11.1 MBq/0.6 ml (i.e.,
18.5 MBq/ml) of 186Re-liposomes in several locations of
the tumor, aiming at producing a uniform distribution of
the injected activity. More than likely the increment in the
interstitial pressure of the tumor, due to the volume of the
injected 186Re-liposomes, promoted a high pressure gradi-
ent that induced a rapid removal of the liposomes from the
tumor into the blood circulation. However, the liposomes
that were not initially removed from the tumor (45%)
stayed within the volume and produced heterogeneous and
non-symmetrical dose distributions. The analysis of the
films also showed that the degree of heterogeneity was
lower in the central slices. Similar behavior was observed in
all tumors studied.
In general, the results showed that average absorbed
doses were higher at the central core of the tumor and lower
at the surface of the tumor. This result is significant when
compared with intratumoral absorbed dose distribution
when intravenous injection of the radionuclide carrier is
performed. A previous theoretical work that considered
intravenous injection of high-energy beta-emitters [28] has
reported that higher radiation doses are achieved at the
surface than in the core of the tumor, as a function of the
tumor diameter and the energy of the beta particle. In
the case of the intratumoral injection of 186Re-liposomes,
the size of the tumor, the volume and activity of the186Re-liposomes injected, the number and locations of the
sites of injection, and the retention time of the radionuclide
Dose (Gy)
0 10 20 30
V
olume(%)
0
20
40
60
80
100
3.7 MBq
11.1 MBq
Fig. 7 Cumulative (integral) dose volume histogram for a tumor from
each group. Doses were calculated using the two-dimensional iso-
dose-rate distributions during one 186Re half-life, as explained in the
text
Slice 4, Group 11.1 MBq
Horzontal axis (x/dx)
-1.0 -0.5 0.0 0.5 1.0
Relativedoserate(%)
Relativedoserate
(%)
0
10
20
30
40
50
Vertical axis (y/dy)
-1.0 -0.5 0.0 0.5 1.0
0
10
20
30
40
50
Fig. 6 One-dimensional scan, in horizontal and vertical directions
(dx=dy=2 cm), across two-dimensional dose rate distributions for the11.1-MBq group (slice 4 in Fig. 5). Arrows mark the physical edges of
the tumor
Eur J Nucl Med Mol Imaging (2007) 34:10391049 1047
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/30/2019 25369407(liposomas)
10/12
inside the tumor will determine variations in the levels of
dose depositions between the core of the tumor and its
surface as well as the surrounding normal tissue. In terms of
dosimetric evaluations, a more detailed study that may
consider these parameters should be pursued.
Our method permits the estimation of the dose rate
spatial distribution at the moment of sacrifice. Complemen-
tary information, such as that provided by biodistributionstudies, e.g., the radionuclide effective half-life, might
permit the evaluation of the integrated total dose (and its
inhomogeneity) received by a tumor after the radionuclide
injection. Finally, the excellent spatial resolution associated
with radiochromic films may permit the use of the general
concept of this technique in smaller lesions, too.
We have presented the use of a commercial radiochromic
film, originally designed for penetrating gamma rays, in the
dosimetry of a distributed source of a beta emitter. We have
evaluated the effect of the particular design parameters in
its performance. In spite of the perturbation effect caused
by the convenient protective layer, we have shown itspotential use as a high-resolution image receptor. These
results could lead to improvements in the development of
new radiochromic media in the future.
Conclusion
This work has shown that the use of HS Gafchomic film
provid es experimental information that could help to
improve models and computational calculations of intra-
tumoral dose distribution in radionuclide therapy with high
energy beta-emitters. The use of the HS Gafchromic film
with the procedures and techniques described in the present
work is of value in the determination of dose rate
distributions and their degree of heterogeneity and symme-
try and in the evaluation of significant dose gradients in
small areas. Accurate use of these films under electron
irradiation requires specific consideration of the interaction
between the film structural design and the physi cal
properties of the radiation field.
Acknowledgements Luis A. Medina acknowledges the National
Institute of Cancerology (Mexico) for hospitality during the develop-
ment of this study, and the Instituto de Ciencias Nucleares, UNAM,
for the calibration of the films. The authors would like to thank Dr.Mohan Natarajan and Xiangpeng Zheng for help with SCC-4 cell
culture, Anuradha Soundararajan, Cristina Zavaleta and Maxwell
Amurao for help during the image acquisition, and Jos-Manuel
Lrraga for assistance with the Monte Carlo simulations. The authors
would also like to thank Dr. Cathy Cutler at the University of Missouri
Research Reactor for sponsoring our research under the Reactor
Sharing Grant (US Department of Energy grant DE-FG07-
02ID14380). This work was also supported in part by DGAPA-
UNAM grants IN-108906 and IN-110204.
References
1. Carlsson J, Aronsson-Forssell E, Hietala S-O, Stigbrand T,
Tennvall J. Tumor therapy with radionuclides: assessment of
progress and problems. Radiother Oncol 2003;66:10717.
2. Drummond DC, Meyer O, Hong K, Kirpotin DB, Papahadjopoulos
D. Optimizing liposomes for delivery of chemotherapeutic agents
to solid tumors. Pharmacol Rev 1999;51:691743.
3. Emfietzoglou D, Kostarelos K, Papakostas A, Yang W-H,Ballangrud A, Song H, et al. Liposome-mediated radiotherapeu-
tics within avascular tumor spheroids: comparative dosimetry
study for various radionuclides, liposome-systems, and a targeting
antibody. J Nucl Med 2005;46:8997.
4. Emfietzoglou D, Kostarelos K, Sgouros G. An analytical
dosimetry study for the use of radionuclide-liposomes conjugates
in internal radiotherapy. J Nucl Med 2001;42:499504.
5. Harrington KJ, Mohammadtaghi S, Uster P, Glass D, Peters AM,
Vile R, et al. Effective targeting of solid tumors in patients with
locally advanced cancers by radiolabeled pegylated liposomes.
Clin Canc Res 2001;7:24354.
6. Harrington KJ, Lewasnki CR, Stewart JSW. Liposomes as vehicles
for targeted therapy of cancer. Part 1: preclinical development. Clin
Oncol 2000;12:215.
7. Torchilin VP. Recent advances with liposomes as pharmaceuticalcarriers. Nat Rev Drug Discov 2005;4:14560.
8. Belhaj-Tayeb H, Briane D, Vergote J, Kothan S, Leger G,
Bendada SE, et al. In vitro and in vivo study of 99mTc-MIBI
encapsulated in PEG-liposomes: a promising radiotracer for tumor
imaging. Eur J Nucl Med Mol Imaging 2003;30:5029.
9. Medina OP, Zhu Y, Kairemo K. Targeted liposomal drug delivery
in cancer. Current Pharm Design 2004;10:29819.
10. Harrington KJ, Rowlinson-Busza G, Syrigos KN, Vile RG, Uster
PS, Peters AM, et al. Pegylated Liposome-encapsulated doxoru-
bicin and cisplatin enhance the effect of radiotherapy in tumor
xenograft model. Clin Canc Res 2000;6:493949.
11. Jain RK. Barriers to drug delivery in solid tumors. Sci Am
1994;271:5865.
12. Murray JC, Carmichael J. Targeting solid tumors: challenges,
disappointments, and opportunities. Adv Drug Deliv Rev1995;17:11727.
13. Bao A, Phillips WT, Goins B, Zheng X, Sabour S, Natarajan M, et
al. Potential use of drug carried-liposomes for cancer therapy via
direct intratumoral injection. Int J Pharm 2006;316:1629.
14. NNDC (National Nuclear Data Center, Brookhaven National
Laboratory: http://www.nndc.bnl.gov/chart/), 2006.
15. Bao A, Goins B, Klipper R, Negrete G, Phillips WT. 186Re-
liposome labeling using 186Re-SNS/S complexes: in vitro stability,
imaging, and biodistribution in rats. J Nucl Med 2003;44:19929.
16. Niroomand-Rad A, Blackwell CR, Coursey BM, Gall KP, Galvin
JM, McLaughlin WL, et al. Radiochromic film dosimetry:
Recommendations of AAPM Radiation Therapy Committee Task
Group 55. Med Phys 1998;25:2093115.
17. ISP (International Speciality Products, Inc.: http://www.ispcorp.
com), 2006.18. Hirata EY, Cunningham C, Micka JA, Keller H, Kissick MW,
DeWerd LA. Low dose fraction behavior of high sensitivity
radiochromic film. Med Phys 2005;32:105460.
19. Piermattei A, Fidanzio A, Azario L, Russo A, Perrone F, Capote
R, et al. A standard dosimetry procedure for192Ir sources used for
endovascular brachytherapy. Phys Med Biol 2002;47:420521.
20. Sharma SD, Bianchi C, Conte L, Novario R, Bhatt BC. Radio-
chromic film measurement for anisotropy function for high-dose-
rate Ir-192 brachytherapy source. Phys Med Biol 2004;49:406572.
1048 Eur J Nucl Med Mol Imaging (2007) 34:10391049
http://www.nndc.bnl.gov/chart/http://www.ispcorp.com/http://www.ispcorp.com/http://www.ispcorp.com/http://www.ispcorp.com/http://www.nndc.bnl.gov/chart/ -
7/30/2019 25369407(liposomas)
11/12
21. Mayer R, Dillenhay LE, Shao Y, Song S, Zhang Y-G, Bartholomew
RM, et al. A new method for determining dose rate distribution
from radioimmuno-therapy using radiochromic media. Int J Radiat
Oncol Biol Phys 1993;28:50513.
22. Tomayko MM, Reynolds CP. Determination of subcutaneous
tumor size in athymic (nude) mice. Cancer Chemother Pharmacol
1989;24:14854.
23. Corbin JL, Miller KF, Pariyadath N, Wherland S, Bruce AE.
Preparation and properties of tripodal and linear tetradentate N, S-
donor ligands and their complexes containing the MoO22+ core.
Inorg Chim Acta 1984;90:4151.
24. Alva H, Mercado-Uribe H, Rodriguez-Villafuerte M, Brandan
ME. The use of a reflective scanner to study radiochromic film
response. Phys Med Biol 2002;47:292533.
25. Flynn AA, Pedley RB, Green AJ, Boxer GM, Boden R, Bhatia J,
et al. Antibody and radionuclide characteristics and the enhance-
ment of effectiveness of radioimmunotherapy by selective dose
delivery to radiosensitive areas of tumour. Int J Radiat Biol
2002;78:40715.
26. Hall EJ. Radiobiology for the radiologist, 4th ed. New York: J.B.
Lippincott; 1994.
27. ODonoghue M, Bardies M, Wheldon TE. Relationship between
tumor size and curability for uniformly targeted therapy with beta-
emitting radionuclides. J Nucl Med 1995;36:19029.
28. Bao A, Zhao X, Phillips WT, Woolley FR, Otto RA, Goins B, et
al. Theoretical study of the influence of a heterogeneous activity
distribution on intratumoral absorbed dose distribution. Med Phys
2005;32:2008.
Eur J Nucl Med Mol Imaging (2007) 34:10391049 1049
-
7/30/2019 25369407(liposomas)
12/12