25369407(liposomas)

download 25369407(liposomas)

of 12

Transcript of 25369407(liposomas)

  • 7/30/2019 25369407(liposomas)

    1/12

    ORIGINAL ARTICLE

    Spatial dose distributions in solid tumors from 186Re

    transported by liposomes using HS radiochromic media

    Luis A. Medina & Beth Goins &

    Mercedes Rodrguez-Villafuerte & Ande Bao &

    Arnulfo Martnez-Davalos & Vibhudutta Awasthi &

    Olga O. Galvn & Cristina Santoyo & William T. Phillips &

    Mara-Ester Brandan

    Received: 27 April 2006 / Accepted: 20 September 2006 / Published online: 8 February 2007# Springer-Verlag 2007

    Abstract

    Purpose A procedure for the measurement of spatial dose

    rate distribution of beta particles emitted by 186Re-lipo-

    somes in tumoral tissue, using HS GafChromic films, is

    presented.

    Methods HNSCC xenografts were intratumorally injected

    with 3.7 or 11.1 MBq of 186Re-liposomes, and planar

    gamma camera images were acquired to determine the

    liposome retention in the tumor. After imaging, rats were

    sacrificed and tumors were excised and processed in slices;

    HS film sections were placed between slices and the tumor

    lobe was reassembled. Tumors and films were kept in the

    dark at 4C for 18 h. After irradiation, films were removed

    and response was read using a transmission scanner. Films

    were analyzed to determine two-dimensional spatial dose

    rate distributions and cumulative dose volume histograms.

    Dose rate distributions were quantified using a 60Co

    calibration curve, the 186Re physical half-life, and a

    perturbation factor that takes into account the effect of the

    film protective layer.

    Results Dose rate distributions are highly heterogeneous

    with maximal dose rates about 0.4 Gy h1 in tumors

    injected with 3.7 MBq and 1.3 Gy h1 in tumors injected

    with 11.1 MBq. Dose volume histograms showed dose

    distributed in more than 95% and 80% of the tumor when

    injected with the lower and the higher activity, respectively.

    Conclusion The described procedures and techniques have

    shown the potential and utility of HS GafChromic film for

    determination of dose rate distributions in solid tumors

    injected intratumorally with 186Re-liposomes. The films

    structure and the liposomes biodistribution must be taken

    into account to obtain quantitative dose measurements.

    Keywords Radionuclide therapy. Liposomes .186

    Re .

    HSGafchromicfilm . Intratumoraldosimetry. MCsimulation

    Introduction

    The interest in the use of liposomes as carriers of

    therapeutic drugs and radionuclides has increased in recent

    years [18]. Several investigators have reported the

    advantages of using intravenously injected liposomes to

    transport and specifically deliver therapeutic drugs intumors of hematopoietic origin [1, 5, 6, 9], and for the

    treatment of solid tumors [3, 6, 10]. However, a major

    limitation for the clinical application of chemotherapeutic

    and radiotherapeutic liposomes in solid tumors has been

    their insufficient penetration and, thus, their inhomoge-

    neous distribution within the tumor tissue [2, 11, 12]. In the

    case of radiotherapeutic liposomes, although the physical

    characteristics of the radionuclide carried by the liposomes

    could limit the level of variability in the spatial deposition

    of the dose inside the tumor, the penetration and distribu-

    Eur J Nucl Med Mol Imaging (2007) 34:10391049

    DOI 10.1007/s00259-006-0297-x

    L. A. Medina (*) : M. Rodrguez-Villafuerte :A. Martnez-Davalos : O. O. Galvn : M.-E. BrandanInstituto de Fisica, UNAM, A.P. 20-364,Mexico 01000 D.F., Mexico

    e-mail: [email protected]

    B. Goins : A. Bao : V. Awasthi : C. Santoyo : W. T. PhillipsDepartment of Radiology,

    University of Texas Health Science Center at San Antonio,

    San Antonio, TX, USA

    A. Bao

    Department of OtolaryngologyHead and Neck Surgery,

    University of Texas Health Science Center at San Antonio,

    San Antonio, TX, USA

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/30/2019 25369407(liposomas)

    2/12

    tion of the liposomes within the tumor still play an

    important role in the spatial distribution of the dose.

    An alternative option for an interstitial pressure-related

    strategy for treatment of locally confined solid tumors, as in

    the case of head and neck tumors, would be to inject the

    liposomes directly into the tumor. This should increase

    the internal pressure at the core of the tumor relative to the

    surrounding tumor tissue, allowing the liposomes to spreadat the interior of the tumor [11]. In this manner a

    radionuclide carried by the liposomes could gradually

    irradiate the tumor depending only on the type and energy

    spectrum of the emitted ionizing particles, producing a

    better spatial dose deposition and a higher intratumoral

    absorbed dose as compared with the intravenous injection

    of the same kind of radiolabeled liposomes [13].

    Our groups are interested in evaluating the therapeutic

    use of a radionuclideliposome system directly injected in

    solid tumors, based on direct dosimetric measurements.

    Recently, a practical method for labeling preformed lipo-

    somes by loading the therapeutic beta-emitting radionuclide186Re (T1/2=89 h; Emax=1.069, 0.932 MeV; Eave=0.359,

    0.306 MeV; mean range 1.8 mm) [14] into the liposome

    interior aqueous space was reported [15].

    In the present work, we report first results describing a

    dosimetric procedure based on the use of HS GafChromic

    dosimetric media to determine the spatial distribution of

    absorbed doses imparted by 186Re in tumor xenografts,

    using an experimental model of head and neck cancer in

    nude rats by means of intratumoral injection of 186Re-

    liposomes. The features of high spatial resolution, weak

    energy dependence, and good tissue equivalence of

    GafChromic films have made this radiochromic material

    useful for monitoring energy deposition in dosimetric

    procedures with high dose-gradient radiation fields [16],

    as is the case in radionuclide therapy with 186Re. HS films,

    with an active radiochromic layer sandwiched between two

    sheets of transparent 97-m-thick polyester, were designed

    to increase dose response to high-energy photon beams

    (>1 MeV) [17] in a dosimetric range of 0.540 Gy. HS film

    contains the same active component as the previous MD-55

    media, but has a higher sensitivity because of a thicker

    active layer, 40 m in HS versus 15 or 30 m in MD-55-1

    or MD-55-2, respectively [17; D. Lewis, 2006, personal

    communication]. Thus, basic interactions between ionizing

    radiation and the basic active components are the same for

    both types of film, and the known dosimetric properties of

    MD-55 films should apply to GafChromic HS media [18;

    D. Lewis, 2006, personal communication]. Some inves-

    tigators have reported the useful application of HS films in132Ir brachytherapy dosimetric procedures [19, 20]. Also,

    GafChromic films have been used previously by Mayer et

    al. [21] to determine beta-particle dose rate distributions in

    tumor sections previously injected with 90Y. The present

    report describes the dosimetric use of the high-sensitive HS

    film to determine dose rate distributions in tumor sections

    injected with 186Re-labeled liposomes.

    Materials and methods

    SCC-4 cell line

    A human origin head and neck squamous cell carcinoma

    (HNSCC) cell line, SCC-4, was originally obtained from

    American Type Culture Collection (ATCC, Manassas, VA)

    and provided by Martin Thornhill, PhD, of the University

    o f Tex as Health S cience C en ter at S an An to nio

    (UTHSCSA), School of Dentistry. SCC-4 cells were

    cultured in Dulbeccos modified Eagles medium (Invitro-

    gen, San Diego, CA) with 10% fetal bovine serum, 2 mM

    L-glutamine, and antibiotics, penicillin, and streptomycin,

    in 10020 mm2 Corning cell culture dishes (Corning Inc,

    Corning, NY) at 37C in 5% CO2. When near confluence,cells were trypsinized and collected to determine the

    number of cells and viability by trypan blue dye exclusion

    assay. The appropriate volume of cell suspension was

    transferred to a new tube and centrifuged in an Allegra 21R

    Centrifuge (Beckman Coulter, Fullerton, CA) at 800 rpm at

    4C for 5 min. Following aspiration of the supernatant, the

    cell pellets were diluted with saline (Abbott Laboratories,

    Abbott Park, IL) to a concentration of 5106 cells in 0.2 ml

    saline. Aliquots of 0.2 ml cell suspension were drawn into

    tuberculin syringes and used for subcutaneous inoculations

    in nude rats.

    HNSCC xenografts in nude rats

    Animal experiments reported in the present paper were

    performed according to the NIH Animal Use and Care

    Guidelines and were approved by the UTHSCSA Institu-

    tional Animal Care Committee. During all procedures, the

    animals were anesthetized with 13% isoflurane in 100%

    oxygen using an anesthesia inhalation machine (Bickford,

    Wales Center, NY). Male rnu/rnu athymic nude rats

    (Harlan, Indianapolis, IN) at 45 weeks of age (75100 g)

    were inoculated subcutaneously with 5106 SCC-4 tumor

    cells in 0.20 ml of saline on the dorsum at the level of the

    scapulae. The animals were fed and housed following the

    protocol for the nude rat living environment at the animal

    facility at the UTHSCSA, and were checked daily after

    tumor cell inoculation. When the growing tumor in each

    animal was palpable and of sufficient size to be measured,

    the tumor size was obtained via caliper by measuring the

    length (l), width (w), and thickness (t) of each tumor. The

    tumor volumes were subsequently calculated using

    the ellipsoid volume formula, V =6 l w t [22].

    1040 Eur J Nucl Med Mol Imaging (2007) 34:10391049

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/30/2019 25369407(liposomas)

    3/12

    Preparation of liposomes

    Liposomes comprised distearoyl phosphatidylcholine

    (DSPC) (Avanti Polar Lipids, Pelham, AL), cholesterol

    (Calbiochem, San Diego, CA), and -tocopherol (Aldrich,

    Milwaukee, WI) having a molar percentage of 55:44:1. A

    dried film was formed by mixing the lipid ingredients in

    chloroform and then removing the chloroform by rotaryevaporation and vacuum desiccation for 24 h. The dried

    lipid film was rehydrated with 300 mmol/l sucrose (Sigma,

    St. Louis, MO) in sterile water for injection (120 mmol/l total

    lipid) and warmed to 55C and then lyophilized over-

    night. The dried lipidsucrose mixture was rehydrated

    with 200 mmol/l reduced glutathione (GSH) (Sigma) in

    Dulbeccos phosphate-buffered saline (PBS) (pH 6.3).

    Before extrusion, the lipid suspension was diluted to

    40 mol/ml with 100 mM GSH in PBS containing

    150 mM sucrose. The diluted lipid suspensions were then

    extruded through a series of Whatman Nucleopore

    polycarbonate filters (Florham Park, NJ) (two passes,2 m; two passes, 400 nm; two passes, 200 nm; five

    passes, 100 nm) at 55C using Lipex Extruder (Northern

    Lipids, Vancouver, Canada). The extruded lipid solution

    was then stored in the refrigerator at 4C for up to 6

    months until needed for radiolabeling studies. Following

    manufacture, the liposome particle diameter was mea-

    sured using a 488-nm laser light scattering instrument

    (Brookhaven Instruments, Holtsville, NY). The measured

    diameter was 122.15.5 nm prior to the radiolabeling

    experiment.

    Preparation of 186Re-BMEDA

    An SNS pattern ligand, N,N-bis(2-mercaptoethyl)-N,N-

    diethylethylenediamine (BMEDA) was synthesized from a

    modification of the method by Corbin et al. [23] and the

    chemical structures were verified using 1H/13C NMR.186Re-aluminum perrhenate [186Re-Al(ReO4)3] was

    obtained from Missouri University Research Reactor

    (Columbia, MO). Stannous chloride (Aldrich) was used as

    the reductant and glucoheptonate (GH) (Sigma) was used as

    an intermediate ligand to make 186Re-BMEDA as previ-

    ously described by Bao et al. [15]. BMEDA (1.0 l;

    1.1 mg) was pipetted into a new vial. Then, 0.50 ml of

    0.17 mol/l GH-0.10 mol/l acetate solution, pH 5.0, was

    added, followed by the addition of 60 l of stannous

    chloride (15 mg/ml). The pH of the solution was adjusted to

    5.0 with 1.0 MNaOH. After flushing the solution with N2gas, 1.3 GBq (35.1 mCi), 0.70 ml of 186Re-Al(ReO4)3(0.4 g Re per mCi 186Re) was added. The vial was sealed

    and heated in an 80C water bath for 1 h. The resultant186Re-BMEDA was used directly for the following 186Re-

    liposome labeling.

    Radiolabeling of preformed liposomes by loading 186Re

    into liposome aqueous space

    Liposomes were prepared with 1.2 ml of GSH neutral

    liposomes (60 mM of total lipids) first undergoing

    separation via Sephadex G-25 column chromatography

    (PD-10 column, Amersham Biosciences, Piscataway, NJ)

    eluted with PBS, pH 7.4. The product was added to the186Re-BMEDA solution and incubated at 37C for 1 h.

    Again, the 186Re-liposomes were separated from any free186Re-BMEDA via Sephadex G-25 column chromatogra-

    phy eluted with PBS buffer, pH 7.4. The labeling efficiency

    (40%) was determined from the186

    Re activity before and

    after separation using a Radix dose calibrator.

    Intratumoral administration of186

    Re-liposomes

    Two groups of nude rats (n=3/group) aged 45 weeks (75

    100 g) and bearing a HNSCC xenograft with an average

    volume of 2.00.4 cm3 were injected with 0.20.05 ml of3.7 MBq (0.1 mCi) (3.7-MBq group), or 0.60.05 ml of

    11.1 MBq (0.3 mCi) (11.1-MBq group) of186Re-liposomes,

    which contained 4.00.7 mg of DSPC and cholesterol. The186Re-liposome doses were equally divided and delivered to

    several separate locations of the tumors, trying to cover the

    central core of the tumor via intratumoral injections.

    Image acquisition and biodistributions

    Static planar images at baseline and 3 h after injection were

    acquired using a micro-SPECT/CT scanner in a 565656

    image matrix, with a pixel size of 2 mm (XSPECT, Gamma

    Medica, Northridge CA). Images of the 186Re-liposomes

    were acquired in a 186Re gamma-ray window (137 keV

    15%) using a high-resolution, small field of view, parallel-

    hole collimator (resolution

  • 7/30/2019 25369407(liposomas)

    4/12

    the baseline image to determine the number of counts

    detected in the body image (N0). Proportionality between

    injected activity in the whole body (A0) and the number of

    counts detected in the body image (N0) is assumed. The

    percentage number of counts at some specific time t

    accumulated in the ROI (%N(t)) was calculated as the ratio

    of the counts in the ROI (NROI(t)) and the total number of

    counts detected in the whole body at baseline, multiplied by100 (%N(t)=100 NROI(t)/N0).

    Tumor sectioning and irradiation

    After imaging, the rats were euthanized by cervical disloca-

    tion under anesthesia. Tumor lobes were excised, washed in

    saline solution, weighed and measured for volume determi-

    nation, and embedded in agarose (3%) (USB, Cleveland,

    OH) solution in TBE buffer concentrate (10) (USB,

    Cleveland OH) to form a rectangular solid matrix with

    33 cm2 transverse section. The agarose matrix helped to

    provide support during the sectioning of the tumors into3-mm-thick slices. Dosimetric HS GafChromic film squares

    (33 cm2) (ISP Technologies Inc, Wayne NJ; Lot

    M0525HS) were placed between slices, and the tumor lobe

    was carefully reassembled to its original shape, maintaining

    the match between tumor slices. When necessary, thin ink

    marks were drawn on the dosimetric film to provide

    registration of the tissue outline. Tumors and films were

    kept in the dark at 4C for 18 h. This time was selected as

    practical for the performance evaluation of the HS Gaf-

    Chromic film dosimetric procedure proposed in this work.

    Dosimetry media handling and readout

    GafChromic films were handled according to recommended

    procedures [16] aimed at obtaining reproducible results of

    the highest sensitivity. In particular: the films were kept in a

    dark environment and, during handling, efforts were made

    to reduce exposure to fluorescent light; films were cut 24 h

    prior to irradiation in order to stabilize their response after

    mechanical stress; calibration was performed using a

    uniform field delivering dose in the range of interest;

    readout took place 11 days after irradiation (sufficient to

    reach color stabilization); and a film scanning procedure

    was followed [24] to assure reproducibility.

    After removal from the tumors, each film was washed

    with water using cotton swabs to remove blood or

    radioactive residuals. After washing, films were inspected

    to assure that there was no sign of damage in their surface

    or presence of radioactive residuals. No signs of surface

    changes due to the direct contact with tissue and agarose

    were observed in the films.

    Film response was read 11 days after the end of

    irradiation employing an Agfa DuoScan T1200 scanner in

    transmission mode using a 36-bit RGB (12 bits per color)

    and saved as tagged image file format. The optical density

    range of the scanner was set to a maximum with all filters

    and image enhancement options turned off. Two scanning

    resolutions were used, 200 dpi for films uniformly

    irradiated (calibration) and 300 dpi for films irradiated with

    a highly non-uniform source (tumor irradiation). Every film

    was scanned using an opaque frame to minimize lightcontributions from areas other than the film. Images were

    analyzed by an in-house-written image manipulation

    routine using MatLab 7.0 (Math Works, Natick, MA) to

    obtain the red, blue, and green components of the image.

    The film response, R, was quantified as:

    R log10Cni

    Ci

    where Cni and Ci are the measured color levels of the

    background (non-irradiated) and irradiated films, respectively

    [24].

    Film calibration

    Film calibration between 0 and 40 Gy was performed using60Co gamma rays under conditions of charged-particle

    equilibrium. Calibration readout took place 11 days after

    exposure to match the conditions in the tumor irradiation

    study.

    It is known [16] that for photon energies in the range

    0.11.33 MeV, and for secondary electrons from 0.1 to

    1.0 MeV, photon mass energy absorption coefficients and

    electron mass stopping powers for the active element in the

    radiochromic films are within 2% of those for water. Monte

    Carlo (MC) estimates (see Fig. 1) for the slowing-down

    electron spectrum on the film active layer for a typical

    arrangement used in this experiment show that15% of the

    electrons have energies below 0.1 MeV. At these low

    energies the electron mass stopping powers for the film are

    within 4% of those for water, and this energy dependence

    should have a minor effect on the water equivalence of the

    film.

    With respect to the use of 60Co gamma rays to calibrate

    the response of the slowing-down beta spectrum, we have

    based our technique on the analysis of Fig. 6 in reference

    [16]. This graph shows the experimental energy depen-

    dence in the response of MD-55-2 films to photons below

    0.1 MeV, and the good data description by the normalized

    mass energy absorption coefficient ratio (film sensitive

    material to water). This agreement implies that the observed

    dependence is totally accounted for by the initial photon

    interactions. Therefore, one can conclude that a possible

    energy dependence of the film response to the secondary

    electrons, within the energy interval from about 15 to

    1042 Eur J Nucl Med Mol Imaging (2007) 34:10391049

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/30/2019 25369407(liposomas)

    5/12

    1,250 keV, is negligible. This has led us to assume that

    calibration with 1.25-MeV gamma rays should be appro-

    priate for the electron spectrum generated in the medium by

    the 186Re beta particles (an MC estimate indicates that 98%

    of the secondary electrons reaching the film active layer

    have energies above 15 keV).

    Effect of the HS film design on the dose measurements

    The dosimetric medium studied in this work has a very

    simple structure, as compared with other radiochromic

    films [16]. However, its use for a continuous electron

    spectrum, as in this case, requires an analysis of the

    possible perturbation caused by the detector features on its

    response as a dosimeter. The 49-m-thick sensitive layer,

    where the deposited energy induces the visible change in

    color, is inserted between two 97-m inert protective

    layers, where no effect is induced. Figure 1 shows MC

    results for the transport of beta particles from 186Re diluted

    in a 3-mm-thick water layer (representing the tumor). The

    geometry simulates the situation of a hypothetical central

    detection plane (dashed line) between two tumor slices.

    Figure 1 displays the slowing-down electron spectra

    incident on A) a central plane in direct contact with both

    water layers, B) the central plane in a 49-m-thick

    radiochromic film without protective layers, and C) the

    central plane in a realistic HS film. As seen in the

    spectra, the main effect of adding material to the ideal

    plane is to reduce the number of electrons without a

    substantial change in the spectral shape. Calculation of

    the absorbed dose at the sensitive layer indicates that,

    because of the protective layers, the HS film detects 68%

    of the dose that would be recorded in the hypothetical

    situation depicted in Fig. 1a. The perturbation factor

    (0.68)1 will be used as a correction to the dose measured

    by the film. This effect has no influence at calibration

    owing to the low absorption of the 60Co gamma rays.

    Calibration curves were used to transform film response

    into integrated dose received by the film during the 18 h of

    dosimetric measurement. Doses were transformed into

    initial dose rates at the tumor slice plane, assuming a fixed

    liposome distribution inside the block, and taking into

    account the 186Re physical half-life and the protective layer

    perturbation factor.

    Energy (MeV)

    0.0 0.2 0.4 0.6 0.8 1.0

    Intensity(a.u.)

    0

    500

    1000

    1500

    2000

    2500

    a) 3 mm water - 3mmwater

    c) 3 mm water - HS film - 3 mm water

    b) 3 mm water - active layer - 3 mm water

    Capa activa

    Active layer

    Water

    Water

    HS film

    a

    b

    c

    186Re

    Fig. 1 MC simulations (106 histories each) of the electron spectra at the central plane of the three arrangements indicated in the right-hand

    diagrams, as explained in the text

    Eur J Nucl Med Mol Imaging (2007) 34:10391049 1043

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/30/2019 25369407(liposomas)

    6/12

    Results

    Figure 2 shows planar gamma camera images of two tumor-

    bearing rats acquired at baseline and 3 h after intratumoral

    injection with 3.7 or 11.1 MBq (0.1 and 0.3 mCi) of 186Re-

    liposomes. The image analysis indicates that, at baseline,

    506% of the injected material is retained in the tumor; 3 h

    later, the liposome retention in tumor is similar, 456%. Interms of retained activity at 3 h, these values correspond to

    1.7 MBq (0.045 mCi) and 5.0 MBq (0.135 mCi) of the186Re activity for the 3.7-MBq (0.1-mCi) and 11.1-MBq

    (0.3-mCi) groups, respectively. By biodistribution calcula-

    tions we have estimated the percentage injected activity per

    organ as 483 and 4312, for the 3.7- and 11.1-MBq

    groups, respectively, and the percentage injected activity

    per gram as 153 and 2411, respectively, for the same

    groups. The rest of the 186Re-liposomes move into blood

    circulation and other organs (see Table 1). These observa-

    tions are consistent with recent animal studies using

    intratumoral injection of99mTc-liposomes of the same lipidformulation, showing that the amount of liposomes which

    stays in the tumor after the immediate release remains in the

    tumor for prolonged times [13]. Bao et al. [13] reported

    approximately 44% retention of liposomes in the tumor 3 h

    after injection, which was similar to our value. This

    agreement is to be expected since biodistribution results

    should depend only on the liposome formulation and not on

    the carried radionuclide.

    Figure 3 shows the calibration response curve of the HS

    film as a function of absorbed dose in water. The response

    of the film is shown in the three colours of the RGB

    mode. Blue and green components show a linear

    behavior throughout the dose range, suggesting thepossibility of extending the dynamic range of response

    to doses higher than the nominal 40-Gy limit suggested

    by the manufacturer (ISP Technologies Inc.). The red

    component displays the most sensitive response, with

    linear behavior at doses

  • 7/30/2019 25369407(liposomas)

    7/12

    Figure 5 shows the two-dimensional spatial dose rate

    distribution at excision time in six tumor slices from a rat of

    the group injected with 11.1 MBq. These images are

    representative of the distribution encountered in the other

    animals. Dose rate values in each isodose (in this case, iso-

    dose-rate) contour is denoted by a color as indicated by the

    scale next to the graph. In both experimental groups it was

    observed that the two-dimensional spatial distributions are

    highly heterogeneous, with maximal dose rates (in some

    hot spots) of about 0.4 Gy h1 in those tumors injected with

    3.7 MBq and about 1.3 Gy h1 in tumors injected with 11.1

    MBq. The lowest heterogeneity is observed in the central

    slices (3 and 4).

    In Fig. 6, a one-dimensional scan in horizontal and

    vertical directions across the center of the two-dimensional

    dose rate distributions in a central 11.1-MBq tumor slice

    (slice 4 in Fig. 5) is presented. The figure shows relative

    dose rate percentages along a single line through the center

    of the horizontal and vertical directions. From representa-

    tive scans in both groups of tumors it is possible to

    conclude that the dose distributions are highly non-uniform

    and non-symmetrical with respect to the center of the

    tumor, reaching the highest doses in several hot spots

    within the tumor volume, possibly associated with the site

    of injection. Significant dose gradients within distances

    smaller than 2 mm could be identified in the films.

    Comparing dose distributions among the different tumors,

    significant differences (larger than 30%) in the areas

    covered by the iso-dose-rate curves of 0.08 Gy h1 were

    observed.

    In order to display the heterogeneity in the dose

    distribution, Fig. 7 shows a cumulative dose volume

    histogram for a tumor from each group. The tumor dose

    was calculated from initial dose rate values in Fig. 5 (at

    the time of sacrifice), integrated from baseline to 89 h (one186Re physical half-life), and assuming a constant liposome

    retention in the tumor (equal to that at 3 h), as anapproximation to Baos measurements [13]. The histogram

    for the 3.7-MBq group shows doses between 0 and

    17 Gy heterogeneously distributed in more than 95% of

    the tumor, and doses between 17 and 32 Gy in less than

    5% of the volume. For the 11.1-MBq group, the

    histogram shows doses between 0 and 17 Gy heteroge-

    neously distributed in more than 80% of the tumor, and

    doses from 17 to 32 Gy in less than 20% of the tumor

    volume. Similar behavior was observed in the rest of the

    tumors of these groups.

    Dose (Gy)

    0 10 20 30 40

    Re

    sponse

    0.0

    0.2

    0.4

    0.6

    0.8

    1.0

    1.2

    Red

    Green

    Blue

    GafChromic HS

    Fig. 3 Calibration curve response of the HS GafChromic film as a

    function of dose. Calibration was performed with a source of 60Co

    under charged-particle equilibrium conditions

    Table 1 Biodistribution results

    Values represent the

    mean sem (n=3)

    Organ 11.1-MBq group 3.7-MBq group t

    % Inj. activity/organ % Inj. activity/g % Inj. activity/organ % Inj. activity/g t

    Blood 20.64.14 1.50.34 21.03.25 1.50.21 tSkin 1.60.18 0.050.01 1.40.08 0.040.01 tHeart 0.10.01 0.10.02 0.10.005 0.10.01 tLiver 8.81.02 0.90.12 8.30.68 0.80.05 tSpleen 4.71.08 7.32.86 3.80.29 4.31.24 tKidney 0.90.09 0.90.10 0.80.03 0.80.03 tBowel 2.40.34 0.20.03 1.90.17 0.20.01 tBladder 1.20.20 1.80.42 0.60.16 1.20.17 tTestes 0.030.001 0.020.001 0.030.005 0.020.01 tSkin of neck 7.21.88 3.00.85 2.00.46 0.90.14 tBone 3.50.52 0.10.02 2.20.35 0.10.01 tMuscle 1.30.18 0.010.001 1.30.13 0.010.01 tStomach 0.60.12 0.10.01 0.60.01 0.10.02 tCecum 1.70.24 0.30.04 1.50.05 0.20.01 tLung 0.40.11 0.30.08 0.40.04 0.30.02 t

    Eur J Nucl Med Mol Imaging (2007) 34:10391049 1045

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/30/2019 25369407(liposomas)

    8/12

    Discussion

    Success of radionuclide therapy for solid tumors with the

    use of radiolabeled carrier systems, such as 186Re-lipo-

    somes, is determined by an efficient delivery of cytotoxic

    levels of absorbed dose throughout all neoplastic cells

    within the tumor, with minimal irradiation of the surround-

    ing normal tissues [4, 25]. However, factors such as the

    limited tumoral penetration of the radiolabeled carrier

    system, non-uniform activity distribution inside the tumor,

    and the physical characteristics of the transported radionu-

    clide lead to subcytotoxic absorbed doses of radiation to

    neoplastic cells in the central region of the tumor, which

    could result in sublethal damage repair [26, 27].

    y(cm)

    -1.0

    -0.5

    0.0

    0.5

    1.0

    y(cm)

    -1.0

    -0.5

    0.0

    0.5

    1.0

    x (cm)

    -1.0 -0.5 0.0 0.5 1.0

    y(cm)

    -1.0

    -0.5

    0.0

    0.5

    1.0

    y(cm)

    -1.0

    -0.5

    0.0

    0.5

    1.0

    y(cm)

    -1.0

    -0.5

    0.0

    0.5

    1.0

    1

    2

    3

    4

    5

    0.44

    0.881.32

    0.18

    0.26

    0.09

    Dose rate

    (Gy/h)

    x (cm)

    -1.0 -0.5 0.0 0.5 1.0

    y(cm)

    -1.0

    -0.5

    0.0

    0.5

    1.0

    6

    Fig. 5 HS films showing six

    spatial dose distributions of

    tumor slices from a tumor lobe

    injected with 11.1 MBq of186Re-liposomes, and the

    two-dimensional dose rate dis-

    tributions calculated from the

    digitalization of the HS films, as

    described. Each iso-dose-rate

    contour is denoted by a color as

    indicated by the color-dose rate

    scale. Films were irradiated

    between tumor slices for 18 h

    Fig. 4 Photographs of a) an

    excised tumor, b) the agarose

    matrix where the tumor was

    embedded with the HS films, c)

    the same agarose matrix viewed

    from the bottom, and d) tumor

    slices and films after removal

    from the agarose matrix

    1046 Eur J Nucl Med Mol Imaging (2007) 34:10391049

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/30/2019 25369407(liposomas)

    9/12

    In order to achieve a better intratumoral penetration that

    could result in a more homogeneous dose deposition inside

    the tumor, intratumoral injection of the radionuclide system

    could offer an alternative treatment option. However, a

    better understanding of the dose deposition after intra-

    tumoral injection of a radiolabeled carrier is necessary. A

    goal of this study has been to describe and evaluate spatial

    dose distributions in solid tumors from the direct measure-ment of dose rate in explanted tumoral sections of HNSCC

    xenografts injected with 186Re-liposomes, using the recent-

    ly available HS GafChromic film and digital image analysis

    to evaluate the spatial dose distributions.

    Image analysis during the first 3 h after186

    Re-liposome

    injection (Fig. 2 and ref. [13]) suggested that about 55% of

    the injected activity was removed very quickly from the

    tumor. It is known that pressure in the interstitial tissue of

    solid tumors is uniformly elevated except in the outermost

    rim, where it drops precipitously [11]. An intratumoral

    injection of186Re-liposomes in the core of the tumor would

    increase the pressure at that location, relative to thesurrounding tumor tissue, spreading the 186Re-liposomes

    along the induced pressure gradient formed by convection

    from the surrounding region to the periphery. In the

    experiments reported here, HNSCC xenografts with vol-

    umes of 2.030.37 cm3 and mass of 2.90.6 g were in-

    jected with either 3.7 MBq/0.2 ml or 11.1 MBq/0.6 ml (i.e.,

    18.5 MBq/ml) of 186Re-liposomes in several locations of

    the tumor, aiming at producing a uniform distribution of

    the injected activity. More than likely the increment in the

    interstitial pressure of the tumor, due to the volume of the

    injected 186Re-liposomes, promoted a high pressure gradi-

    ent that induced a rapid removal of the liposomes from the

    tumor into the blood circulation. However, the liposomes

    that were not initially removed from the tumor (45%)

    stayed within the volume and produced heterogeneous and

    non-symmetrical dose distributions. The analysis of the

    films also showed that the degree of heterogeneity was

    lower in the central slices. Similar behavior was observed in

    all tumors studied.

    In general, the results showed that average absorbed

    doses were higher at the central core of the tumor and lower

    at the surface of the tumor. This result is significant when

    compared with intratumoral absorbed dose distribution

    when intravenous injection of the radionuclide carrier is

    performed. A previous theoretical work that considered

    intravenous injection of high-energy beta-emitters [28] has

    reported that higher radiation doses are achieved at the

    surface than in the core of the tumor, as a function of the

    tumor diameter and the energy of the beta particle. In

    the case of the intratumoral injection of 186Re-liposomes,

    the size of the tumor, the volume and activity of the186Re-liposomes injected, the number and locations of the

    sites of injection, and the retention time of the radionuclide

    Dose (Gy)

    0 10 20 30

    V

    olume(%)

    0

    20

    40

    60

    80

    100

    3.7 MBq

    11.1 MBq

    Fig. 7 Cumulative (integral) dose volume histogram for a tumor from

    each group. Doses were calculated using the two-dimensional iso-

    dose-rate distributions during one 186Re half-life, as explained in the

    text

    Slice 4, Group 11.1 MBq

    Horzontal axis (x/dx)

    -1.0 -0.5 0.0 0.5 1.0

    Relativedoserate(%)

    Relativedoserate

    (%)

    0

    10

    20

    30

    40

    50

    Vertical axis (y/dy)

    -1.0 -0.5 0.0 0.5 1.0

    0

    10

    20

    30

    40

    50

    Fig. 6 One-dimensional scan, in horizontal and vertical directions

    (dx=dy=2 cm), across two-dimensional dose rate distributions for the11.1-MBq group (slice 4 in Fig. 5). Arrows mark the physical edges of

    the tumor

    Eur J Nucl Med Mol Imaging (2007) 34:10391049 1047

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/30/2019 25369407(liposomas)

    10/12

    inside the tumor will determine variations in the levels of

    dose depositions between the core of the tumor and its

    surface as well as the surrounding normal tissue. In terms of

    dosimetric evaluations, a more detailed study that may

    consider these parameters should be pursued.

    Our method permits the estimation of the dose rate

    spatial distribution at the moment of sacrifice. Complemen-

    tary information, such as that provided by biodistributionstudies, e.g., the radionuclide effective half-life, might

    permit the evaluation of the integrated total dose (and its

    inhomogeneity) received by a tumor after the radionuclide

    injection. Finally, the excellent spatial resolution associated

    with radiochromic films may permit the use of the general

    concept of this technique in smaller lesions, too.

    We have presented the use of a commercial radiochromic

    film, originally designed for penetrating gamma rays, in the

    dosimetry of a distributed source of a beta emitter. We have

    evaluated the effect of the particular design parameters in

    its performance. In spite of the perturbation effect caused

    by the convenient protective layer, we have shown itspotential use as a high-resolution image receptor. These

    results could lead to improvements in the development of

    new radiochromic media in the future.

    Conclusion

    This work has shown that the use of HS Gafchomic film

    provid es experimental information that could help to

    improve models and computational calculations of intra-

    tumoral dose distribution in radionuclide therapy with high

    energy beta-emitters. The use of the HS Gafchromic film

    with the procedures and techniques described in the present

    work is of value in the determination of dose rate

    distributions and their degree of heterogeneity and symme-

    try and in the evaluation of significant dose gradients in

    small areas. Accurate use of these films under electron

    irradiation requires specific consideration of the interaction

    between the film structural design and the physi cal

    properties of the radiation field.

    Acknowledgements Luis A. Medina acknowledges the National

    Institute of Cancerology (Mexico) for hospitality during the develop-

    ment of this study, and the Instituto de Ciencias Nucleares, UNAM,

    for the calibration of the films. The authors would like to thank Dr.Mohan Natarajan and Xiangpeng Zheng for help with SCC-4 cell

    culture, Anuradha Soundararajan, Cristina Zavaleta and Maxwell

    Amurao for help during the image acquisition, and Jos-Manuel

    Lrraga for assistance with the Monte Carlo simulations. The authors

    would also like to thank Dr. Cathy Cutler at the University of Missouri

    Research Reactor for sponsoring our research under the Reactor

    Sharing Grant (US Department of Energy grant DE-FG07-

    02ID14380). This work was also supported in part by DGAPA-

    UNAM grants IN-108906 and IN-110204.

    References

    1. Carlsson J, Aronsson-Forssell E, Hietala S-O, Stigbrand T,

    Tennvall J. Tumor therapy with radionuclides: assessment of

    progress and problems. Radiother Oncol 2003;66:10717.

    2. Drummond DC, Meyer O, Hong K, Kirpotin DB, Papahadjopoulos

    D. Optimizing liposomes for delivery of chemotherapeutic agents

    to solid tumors. Pharmacol Rev 1999;51:691743.

    3. Emfietzoglou D, Kostarelos K, Papakostas A, Yang W-H,Ballangrud A, Song H, et al. Liposome-mediated radiotherapeu-

    tics within avascular tumor spheroids: comparative dosimetry

    study for various radionuclides, liposome-systems, and a targeting

    antibody. J Nucl Med 2005;46:8997.

    4. Emfietzoglou D, Kostarelos K, Sgouros G. An analytical

    dosimetry study for the use of radionuclide-liposomes conjugates

    in internal radiotherapy. J Nucl Med 2001;42:499504.

    5. Harrington KJ, Mohammadtaghi S, Uster P, Glass D, Peters AM,

    Vile R, et al. Effective targeting of solid tumors in patients with

    locally advanced cancers by radiolabeled pegylated liposomes.

    Clin Canc Res 2001;7:24354.

    6. Harrington KJ, Lewasnki CR, Stewart JSW. Liposomes as vehicles

    for targeted therapy of cancer. Part 1: preclinical development. Clin

    Oncol 2000;12:215.

    7. Torchilin VP. Recent advances with liposomes as pharmaceuticalcarriers. Nat Rev Drug Discov 2005;4:14560.

    8. Belhaj-Tayeb H, Briane D, Vergote J, Kothan S, Leger G,

    Bendada SE, et al. In vitro and in vivo study of 99mTc-MIBI

    encapsulated in PEG-liposomes: a promising radiotracer for tumor

    imaging. Eur J Nucl Med Mol Imaging 2003;30:5029.

    9. Medina OP, Zhu Y, Kairemo K. Targeted liposomal drug delivery

    in cancer. Current Pharm Design 2004;10:29819.

    10. Harrington KJ, Rowlinson-Busza G, Syrigos KN, Vile RG, Uster

    PS, Peters AM, et al. Pegylated Liposome-encapsulated doxoru-

    bicin and cisplatin enhance the effect of radiotherapy in tumor

    xenograft model. Clin Canc Res 2000;6:493949.

    11. Jain RK. Barriers to drug delivery in solid tumors. Sci Am

    1994;271:5865.

    12. Murray JC, Carmichael J. Targeting solid tumors: challenges,

    disappointments, and opportunities. Adv Drug Deliv Rev1995;17:11727.

    13. Bao A, Phillips WT, Goins B, Zheng X, Sabour S, Natarajan M, et

    al. Potential use of drug carried-liposomes for cancer therapy via

    direct intratumoral injection. Int J Pharm 2006;316:1629.

    14. NNDC (National Nuclear Data Center, Brookhaven National

    Laboratory: http://www.nndc.bnl.gov/chart/), 2006.

    15. Bao A, Goins B, Klipper R, Negrete G, Phillips WT. 186Re-

    liposome labeling using 186Re-SNS/S complexes: in vitro stability,

    imaging, and biodistribution in rats. J Nucl Med 2003;44:19929.

    16. Niroomand-Rad A, Blackwell CR, Coursey BM, Gall KP, Galvin

    JM, McLaughlin WL, et al. Radiochromic film dosimetry:

    Recommendations of AAPM Radiation Therapy Committee Task

    Group 55. Med Phys 1998;25:2093115.

    17. ISP (International Speciality Products, Inc.: http://www.ispcorp.

    com), 2006.18. Hirata EY, Cunningham C, Micka JA, Keller H, Kissick MW,

    DeWerd LA. Low dose fraction behavior of high sensitivity

    radiochromic film. Med Phys 2005;32:105460.

    19. Piermattei A, Fidanzio A, Azario L, Russo A, Perrone F, Capote

    R, et al. A standard dosimetry procedure for192Ir sources used for

    endovascular brachytherapy. Phys Med Biol 2002;47:420521.

    20. Sharma SD, Bianchi C, Conte L, Novario R, Bhatt BC. Radio-

    chromic film measurement for anisotropy function for high-dose-

    rate Ir-192 brachytherapy source. Phys Med Biol 2004;49:406572.

    1048 Eur J Nucl Med Mol Imaging (2007) 34:10391049

    http://www.nndc.bnl.gov/chart/http://www.ispcorp.com/http://www.ispcorp.com/http://www.ispcorp.com/http://www.ispcorp.com/http://www.nndc.bnl.gov/chart/
  • 7/30/2019 25369407(liposomas)

    11/12

    21. Mayer R, Dillenhay LE, Shao Y, Song S, Zhang Y-G, Bartholomew

    RM, et al. A new method for determining dose rate distribution

    from radioimmuno-therapy using radiochromic media. Int J Radiat

    Oncol Biol Phys 1993;28:50513.

    22. Tomayko MM, Reynolds CP. Determination of subcutaneous

    tumor size in athymic (nude) mice. Cancer Chemother Pharmacol

    1989;24:14854.

    23. Corbin JL, Miller KF, Pariyadath N, Wherland S, Bruce AE.

    Preparation and properties of tripodal and linear tetradentate N, S-

    donor ligands and their complexes containing the MoO22+ core.

    Inorg Chim Acta 1984;90:4151.

    24. Alva H, Mercado-Uribe H, Rodriguez-Villafuerte M, Brandan

    ME. The use of a reflective scanner to study radiochromic film

    response. Phys Med Biol 2002;47:292533.

    25. Flynn AA, Pedley RB, Green AJ, Boxer GM, Boden R, Bhatia J,

    et al. Antibody and radionuclide characteristics and the enhance-

    ment of effectiveness of radioimmunotherapy by selective dose

    delivery to radiosensitive areas of tumour. Int J Radiat Biol

    2002;78:40715.

    26. Hall EJ. Radiobiology for the radiologist, 4th ed. New York: J.B.

    Lippincott; 1994.

    27. ODonoghue M, Bardies M, Wheldon TE. Relationship between

    tumor size and curability for uniformly targeted therapy with beta-

    emitting radionuclides. J Nucl Med 1995;36:19029.

    28. Bao A, Zhao X, Phillips WT, Woolley FR, Otto RA, Goins B, et

    al. Theoretical study of the influence of a heterogeneous activity

    distribution on intratumoral absorbed dose distribution. Med Phys

    2005;32:2008.

    Eur J Nucl Med Mol Imaging (2007) 34:10391049 1049

  • 7/30/2019 25369407(liposomas)

    12/12