216 Lab Manual 2003-2004
-
Upload
almoesana-yuph-almusana -
Category
Documents
-
view
216 -
download
0
Transcript of 216 Lab Manual 2003-2004
-
8/6/2019 216 Lab Manual 2003-2004
1/101
The University of Western OntarioFaculty of Engineering Science
DEPARTMENT OF CHEMICAL AND BIOCHEMICAL
ENGINEERING
INDUSTRIAL ORGANIC CHEMISTRY
C.B.E. 216
LABORATORY MANUAL
Instructor: P.A. Charpentier
2nd Edition
Copyright P. Charpentier 2003
-
8/6/2019 216 Lab Manual 2003-2004
2/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
2
TABLE OF CONTENTS
1 INTRODUCTION............................................................................................... 31.1 CONVERSION FACTORS ............................................................................... 4
1.2 COMMON ITEMS OF GLASSWARE AND APPARATUS ........................... 51.3 A NOTE TO THE STUDENT ............................................................................ 91.4 PERSONAL CONDUCT .................................................................................. 111.5 HANDLING CHEMICALS AND EQUIPMENT ............................................ 121.6 GENERAL HOUSEKEEPING ......................................................................... 141.7 GUIDELINES FOR PREPARATION OF LABORATORY REPORTS ......... 14
LABORATORY 1 - THE FRACTIONATION OF HYDROCARBONSAND GAS CHROMATOGRAPHY (GC)......................................................... 20
LABORATORY 2 - ANALYSIS OF PETROLEUM FRACTIONS .............. 29
LABORATORY 3 - REDUCTION OF A KETONE TO A SECONDARYALCOHOL ..48
LABORATORY 4 - CATALYTIC HYDROGENATION OF SQUALENE (APOLYUNSATURATED HYDROCARBON).................................................. 51
LABORATORY 5- EPOXIDATION OF 1-TMP AND IODINETITRATION..60
LABORATORY 6 - SYNTHESIS OF AZO-DYES AND UV-VISSPECTROSCOPY............................................................................................. 73
LABORATORY 7 - SYNTHESIS OF DICUMYL ETHER AND NMR SPECTROSCOPY ....80
LABORATORY 8 - SOXHLET EXTRACTION OF CAFFEINE FROMBEVERAGE PLANTS..87
APPENDIX 190
-
8/6/2019 216 Lab Manual 2003-2004
3/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
3
1 INTRODUCTION
1.1 CONVERSION FACTORS FOR SOMECOMMONLY USED UNITS OF MEASUREMENT
Acceleration m/s 2 ft/s 2 3.281X10 0 3.048X10 -1
g (accel. of gravity) 1.020X10 -1 9.807X10 0 Area m 2 ft2 1.076X10 1 9.290X10 -2
inch 2 1.550X10 3 6.452X10 -4 yard 2 1.197X10 0 8.361X10 -1
Density kg/m 2 g/cm 1.000X10 -3 1.000X10 3 lb/gallon 8.345X10 -3 1.198X10 2
lb/ft3
6.242X10-2
1.602X101
Energy J btu 9.484X10 -1 1.054X10 3 (includes work) calorie 2.387X10 -1 4.184X10 0
(thermochemical)erg 1.000X10 7 1.000X10 -7 ft-lb 7.375X10 -1 1.356X10 0 kW-hr 2.778X10 7 3.600X10 -6
Force N dyne 1.000X10 5 1.000X10 -5 pound force 2.248X10 -1 4.448X10 0
Heat capacity J/kg*K btu/lb* oF 2.390X10 -4 4.184X10 3 (includes entropy) cal/g* oC 2.390X10 -4 4.184X10 3 Length m 1.000X10 10 1.000X10 -10
in 3.937X10 1 2.540X10 -2
ft 3.281X10 0 3.048X10 -1 micron 1.000X10 6 1.000X10 -6 mile 6.213X10 -4 1.609X10 3 yard 1.094X10 0 9.144X10 -1
Mass kg ounce 3.527X10 2 2.835X10 -1 lb 2.205X10 0 4.536X10 -1
Power W btu/hr 3.414X10 0 2.929X10 -1 btu/sec 9.484X10-4 1.054X10 1 cal/sec 2.390X10 -1 4.184X10 0 ft-lb/sec 7.376X10 -1 1.356X10 0 horsepower 1.341X10 -1 7.457X10 2 (550 ft*lb/sec)
Pressure Pa atm 9.869X10 -6 1.013X10 5 (76cm Hg)
bar 1.000X10-5 1.000X10 5 cm of Hg 7.506X10 -3 1.333X10 3
-
8/6/2019 216 Lab Manual 2003-2004
4/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
4
dyne/cm 2 1.000X10 1 1.000X10 -1 in of Hg 2.961X10 -4 3.337X10 3 kg force/cm 2 1.020X10 -5 9.807X10 4
CONVERSION FACTORS FOR SOMECOMMONLY USED UNITS OF MEASUREMENT CONTD
lb/in 2 (psi) 1.450X10 -4 6.895X10 3 torr 7.501X10 -3 1.332X10 2 (mm of Hg)
Volume m 3 ft3 3.531X10 1 2.832X10 2 (includes capacity) in 3 6.102X10 -1 1.639X10 -5
gallon (U.S.) 2.642X10 2 3.785X10 -3 L 1.000X10 3 1.000X10 -3 ounce (U.S.) 3.381X10 4 2.957X10 -5
psig pounds per square inch gauge psia ponds per square inch absolute
-
8/6/2019 216 Lab Manual 2003-2004
5/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
5
1.2 COMMON ITEMS OF GLASSWARE ANDAPPARATUS
Most of the apparatus used will be familiar to you, but the following notes may help youin identifying and using specific pieces.
The ERLENMEYER or CONICAL FLASK is used for handling solutions,
and for titrations. It is designed with a narrow neck to minimize loss of
solution through splashing or evaporation.
The FILTER or BUCHNER FLASK is used in conjunction with the Buchner
funnel for vacuum-assisted filtration. It is heavy-walled to give pressure resistance; for
this reason, a Buchner funnel should never be used to heat a
solution. Attach it to the vacuum line or water aspirator with heavy-
wall rubber tubing. For any operation involving vacuum, always
use heavy walled rubber tubing - never use soft tubes like
Tygon. If a water aspirator is used, an empty flask should come
between the filter and the pump to avoid suck-back if the water
pressure falls. The Buchner assembly is top-heavy and should besupported when in operation.
The BUCHNER FUNNEL is used for filtration. It fits through a rubber bung
or cone into the Buchner flask. To use, assemble the flask and funnel, place
a filter-paper flat across the perforated porcelain plate, and wet the paper
with the solvent being used (usually distilled water). Turn on the vacuum
and make sure the paper is correctly seated in the funnel; the filter paper should be cut
slightly smaller than the funnel, but make sure it covers all the holes. Stir the suspension
to be filtered and quickly pour it onto the center of the paper, using a glass rod to guide it.
Filtration will go more quickly if you keep liquid in the funnel. If all of the liquid is
filtered off, the residual solid will pack down into a solid cake, slowing filtration. To
empty the funnel after the solid cake has been washed and sucked as dry as possible,
-
8/6/2019 216 Lab Manual 2003-2004
6/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
6
loosen the cake around the edge with a spatula, carefully invert the funnel onto a watch-
glass and tap it gently.
The EVAPORATING DISH is used to provide a large surface area to speed up
evaporation. It can be heated on the steam-bath but should never be heated with a
direct flame.
The BURETTE accurately measures volumes to 0.1ml accuracy. When titrating, always
fill the burette to the zero milliliter marking. Your eye should be level with the bottom of
the meniscus in order to take a proper reading of the liquid level.
The TRANSFER PIPETTE accurately delivers one volume (e.g. 5 or10 or25ml).
The PASTEUR or DROPPING PIPETTE is for transferring a few drops.
The SEPARATORY FUNNEL is used for the separation of
liquids with differing densities and for washing. The funnel
should have a properly working stopcock and a stopper of the
correct size. The solution to be separated is poured into the funnel
with the stopcock closed and the funnel stoppered. It is then
shaken vigorously with two hands; one holding the bottom of the
flask between first and second fingers and the other on the stopper
so it does not fall out. This maneuver is performed with the flask
upside down and the stem directed away from people standing by
in case of any splashing. The pressure which may build up inside
the flask is released by holding the funnel upside down and opening the stopcock. Next,
the funnel is placed in a proper size support ring. Enough time is given for the solutions
-
8/6/2019 216 Lab Manual 2003-2004
7/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
7
to separate into two distinct layers after which the bottom layer can be removed and the
procedure repeated until necessary. ( It is assumed that one knows which layer is to be
kept !)
GAS CYLINDERS with the safety cap off need to be securely strapped to the wall or a
desk in order to prevent them from falling. There is considerable pressure in these
cylinders and care must be taken to control the flow of gases from them. The main
cylinder or tank valve should be closed when not in operation. This valve measures the
pressure present in the tank (i.e. how much gas is left in the tank). The control valve,
usually the second one, gives the pressure reading present in the line connected to the
tank. This is usually a backwards valve, meaning that to reduce pressure it needs to be
turned in the counterclockwise direction. A BUBBLER is usually inserted in the gas line between the cylinder and the connection to the apparatus to be filled with the gas. This
is advantageous for two reasons: 1) the flow of gas is actually seen as it bubbles through
the oil in the bubbler and 2) this is an outlet for the gas if the pressure becomes too high
so that it exits via the bubbler rather than blowing the glassware or connecting tubes.
When working with gas cylinders it is very important that you know and understand how
everything is connected and what function each piece of equipment has.
There are many different shapes of CONDENSERS available
for use. All of them serve the same purpose in conjunction
with distillation apparatus. Their purpose is to cool the vapors
inside the condenser usually with water as coolant. The
condenser is placed before and is tilted toward the receiving
flasks. The glass is blown so that the cooling liquid is
separated from the vapors which are to be condensed. To have
good cooling cold water should flow through the condenser at
all times. This is achieved by connecting the water inlet to the
bottom end of the condenser and the outlet to the top so the
water flows from the bottom up the condenser and out the top.
-
8/6/2019 216 Lab Manual 2003-2004
8/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
8
ERROR DATA
Volumetric flasks Volumetric pipettes
Volume (ml) Error Volume (ml) Error
10 0.04 1 0.006
25 0.06 2 0.006
50 0.10 5 0.01
100 0.16 10 0.01
-
8/6/2019 216 Lab Manual 2003-2004
9/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
9
1.3 A NOTE TO THE STUDENT
The following introductory paragraphs are intended to be for your benefit and safety.
The objective of this laboratory is to introduce the student to basic organic reactions and
analytical instrumentation, as used in industrial operations and processes. By performing
the prescribed experiments the student will become familiar with a typical organic
chemical laboratory and the operation of typical analytical instruments. She/he will also
obtain a feeling and routine for generation of analytical results and the technical
capabilities of various instruments. The generated knowledge will enable the student to
better understand the basic chemical principles and control of industrial processes, which
is essential for proper operation of individual units in the plant and the management of processes for optimum performance and product quality, and environmental effects.
Whether in management, processing, design or laboratory, an engineer should have good
knowledge and understanding of the chemistry, measurements and instrumentation being
used in the plant. The important decisions and modifications that an engineer must make
in industry will be based on the results obtained from the laboratory. A good
understanding of possible errors in procedures and instruments is also required and
particularly a good understanding of variables that could affect a result. A lack of this
understanding very often results in erroneous judgments that can affect considerably both
production and quality of the final product.
Although this laboratory does not involve extensive manipulation of hazardous
chemicals, materials that will be analyzed are often flammable and volatile as well as
toxic. Each student must therefore follow strictly all safety procedures and not perform
any unauthorized manipulations with these chemicals prior to consultation with the
instructor or the demonstrator.
Remember: ACCIDENTS ARE CAUSED AND CAN BE PREVENTED!
-
8/6/2019 216 Lab Manual 2003-2004
10/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
10
A few rules are listed below. A Handbook of Safety Practices in Engineering is supplied
in the laboratory and each student should be familiar with its contents.
The students will work in groups of two or three and enough time will be provided to
finish all the prescribed experiments. If any piece of equipment fails or does not function
properly, students are required to report the problem immediately to the demonstrator or
the instructor and are not allowed to attempt to fix the instrument on their own.
ALL STUDENTS MUST HAVE A LABORATORY BOOK IN WHICH TO RECORD
ALL LAB DATA DURING THE LABORATORY PERIOD. The lab book used in
industry is a legal document. Loose papers for recording of data or comments are not
allowed and will be removed from the lab.
Each student must read the instructions for the particular experiment PRIOR to coming to
the laboratory. He/she should understand the whole procedure and what must be done in
the experiment. This knowledge will be checked periodically by the instructor or the
demonstrator and it will be evaluated.
Equations for all reactions should be written out in your lab notebook before coming to
the laboratory. Also, any calculations required (e.g. theoretical yield, preparation of
solutions) should be written in full in your laboratory notebook before coming to the
laboratory. Record all observations in the lab book including any color changes,
unexpected events, smells, etc. The notebook will be marked from time to time during the
term. Plan your working time in the laboratory! By doing this your laboratory will be a
useful and pleasant experience rather than a frustrating one.
Develop good working habits. Keep your area clean and tidy. Your working area reflects
your working habits and also the quality of your work and results. You are working with
precise instruments and generating accurate data --keep this in mind!
-
8/6/2019 216 Lab Manual 2003-2004
11/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
11
ALL FLASKS, BEAKERS, CONTAINERS WITH ANY CHEMICALS IN THEM
MUST BE CLEARLY LABELED WITH THE NAME OF THE CHEMICAL AND
DATE!
NO SMOKING, EATING OR DRINKING IS ALLOWED IN THE LABORATORY!
LEAVE YOUR WORKING SPACE, APPARATUS AND GLASSWARE CLEAN AND
IN ORDER BEFORE LEAVING THE LAB!
The demonstrator will sign your lab book ONLY after you have fulfilled this
requirement.
ALL LABORATORY EXPERIMENTS MUST BE DONE DURING THE
ALLOCATED TIME PERIOD. THERE WILL BE NO EXTENSION OF THE LAB
AND NO ADDITIONAL TIME PERIOD AVAILABLE FOR THE EXPERIMENTS,
except under special circumstances such as illness verified by a doctors note.
1.4 PERSONAL CONDUCT
(1) Eye protection (safety glasses with side shields) is mandatory.
(2) Eating and drinking in the lab are strictly forbidden.
(3) Long hair should be tied back.
(4) Laboratory coats are required for cloth and body
protection.
(5) No laboratory work may be performed without the prior permission of Mr.
Milton.
(6) No open toed shoes allowed.
(7) No shorts allowed.
VIOLATION OF ANY OF THESE REGULATIONS WILL MEAN THAT YOU WILL
NOT BE PERMITTED TO WORK IN THE LABORATORY.
-
8/6/2019 216 Lab Manual 2003-2004
12/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
12
1.5 HANDLING CHEMICALS AND EQUIPMENT
1. Do not taste chemicals.
2. Do not pipet any chemicals by mouth. Use rubber bulb (propipette).
3. Do not pour liquids that are flammable or that do not mix with water into sinks. Pour
them into the provided and labeled containers.
4. Do not mix incompatible chemicals.
5. Do not heat flammable liquids with an open flame.
6. Operations involving volatile or toxic materials are to be conducted in the fume hood.
7. Dispose of solid wastes in garbage pails. Do not use the sinks.
8. Clean up spills (solid or liquid) at once.
9. Return chipped or broken glassware to the demonstrator.
10. Be sure the apparatus is placed properly. Do not move instruments without proper
consultation with the demonstrator.
11. When heating a test tube make sure that it is not pointing towards yourself or other
people in the vicinity, so no damage will result if the contents suddenly dump out.
12. Never apply force to any glass apparatus. Many serious cuts are caused by the sudden
fracture of glass under strain from misuse. In particular, never use force in an
attempt to push a thermometer or glass tube through a hole in a cork or rubber.
-
8/6/2019 216 Lab Manual 2003-2004
13/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
13
13. Never heat a tightly stoppered flask even if it is empty-----it will explode.
14. Do not attempt to buttress a laboratory assembly with makeshift supports such as
books, pencils and the like. Use several ring stands if necessary. Round bottom flasks
cannot stand freely on the bench----use a special cork support or place the flask into a
beaker.
15. Do not attempt to break up a solid in the bottom of a flask by punching the solid with
a glass stirring rod. The rod may either fracture in your hand or puncture the bottom
of the flask.
16. Avoid shortcuts! If you have an idea for an improvement talk it over with your
demonstrator; if no objections, try it; if it is successful, tell us about it!
17. There are certain necessary precautions associated with particular chemicals or
experiments. Your demonstrator will point these out when required.
18. If you are not familiar with a piece of apparatus or an experimental procedure ask
for help. Dont just try to muddle through without knowing what you are doing.
19. Chemical waste should be disposed of in the labeled waste containers
provided. Halogenated chemicals must be disposed of separately from non-
halogenated chemicals. Nothing should go down the drain! Dispose of all chemicals
in the bottles marked for the specific lab.
-
8/6/2019 216 Lab Manual 2003-2004
14/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
14
1.6 GENERAL HOUSEKEEPING
1. Keep benches clean and orderly and sinks clean. You must leave your portion of the
bench clean at the end of the lab period.
2. Aisles and floors are to be kept free of obstructions. Keep cupboard doors and
drawers closed when not in use.
3. Hang coats on the rack. No coats are permitted on tables or benches.
4. Laboratory doors MUST be unlocked during lab period.
1.7 GUIDELINES FOR PREPARATION OF LABORATORY REPORTS
Laboratory reports should be written as though they were short technical reports. Thus
Tables and Figures should always be referred to in the prose text of the report, i.e. they
should not appear on their own.
The report should be written in the past tense, since it is a description and a correlation of
past observations. The present tense may be used in referring to laws of nature, properties
of materials etc. which are independent of time. Thus, for instance, in a particular
experiment The ambient temperature equaled 22 C; on the other hand, The ambient
temperature equals about 20 C .
TITLE PAGE
Title of experiment
Name of person writing the report
Name of experimenters
Date when the experiment was performed
-
8/6/2019 216 Lab Manual 2003-2004
15/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
15
All Formal reports must be TYPED
ABSTRACT
The Abstract should summarize the entire report. It should state clearly and briefly the
objectives, methods, results and conclusions of the lab.
Objective: State the objective or purpose of the lab
Method: In one or two sentences, summarize the methods, including scientific and
common names of organisms and techniques used.
Results: Summarize what was found in the study
Conclusions: State the significance of the results in relation to the objective.
INTRODUCTION
The Introduction should describe the scope and the purpose of the lab and include any
background information necessary to understand the experiment.
State the general problem. Give a brief statement of why the general topic is relevant and
important. Define any specialized terms or concepts (e.g. the concept of distillation)
likely to be encountered later in the lab report. Supply sufficient background (historical
and theoretical) information to allow the reader to evaluate and understand the results of
the study without needing to refer to other publications.
State the specific objective or purpose of the lab and the approach to be used. The
purpose states what you are investigating and why; how you perform the investigation
should be described later in the Methods and Materials.
-
8/6/2019 216 Lab Manual 2003-2004
16/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
16
MATERIALS AND METHODS
The Methods section should describe what was done and how it was done. It should be
written in the past tense with active voice, and in paragraph form.
The Materials and Methods section should provide only enough detail to permit a
competent worker to evaluate the validity of the experiment and to repeat it, if necessary.
It should not be simply a recipe of all the steps involved. State the names (IUPAC if
possible) of the chemicals used, the instruments, equipment and pattern of replication.
Describe any unusual numerical calculations and state the statistical technique used toanalyze the data.
RESULTS
The Results section should present the data collected in a summarized form and describe
only the key features of these data, emphasizing trends or patterns that are relevant to the
hypotheses being tested. Interpretation of the data is reserved for the discussion section.
Do not present the same data in both a table and figure i.e. place table of raw data in an
appendix and place figure in the results section. Titles of tables and figures should
contain enough information to understand the contents without reference to the text. The
number and title are placed at the top of a table, and at the bottom of figure.
Guide the reader through your figure (s) and table (s) in a logical and systematic manner,
pointing out trends and differences that pertain to the objective (s) of the report. Simply
state what you found in your study, without inference or reference to "expected" results.
-
8/6/2019 216 Lab Manual 2003-2004
17/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
17
DISCUSSION
The Discussion section should provide an explanation and interpretation of your results
and indicate the significance of the results to the hypothesis being tested. Results of
previous studies on the same topic should be compared with yours, with an explanation of
why your results are different from previous studies, if necessary.
State how and why your results either support or do not support the objectives and
hypotheses. REMEMBER: results are results, they are never wrong simply by being
different from either your expectations or from other investigations. Draw conclusionsabout the hypotheses (objectives) of your study, based on all data available in the current
studies.
APPENDIX
Includes all raw experimental data, e.g. time vs. temperature data points, sample
calculations, and any other information or data used for the experiment and calculations.
REFERENCES
The Reference section should be a list of all books, journals, and other materials cited in
the body of the paper.
The surname of the author(s) and the year of publication should be inserted in the text at
an appropriate place:
"Smith (1991) compared..." or "... have been recently compared (Smith, 1991)."
If the reference has more than 2 authors, include only the surname of the first author,
followed by "et al."
-
8/6/2019 216 Lab Manual 2003-2004
18/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
18
"Smith et al. (1991) compared..." or "... have been recently compared (Smith et al.,
1991)."
When listing more than one citation at a given point in the text, list them in
chronologically by first author, but for 2 (or more) papers, published in the same year, list
these alphabetically:
"(Jones, 1978; Black et al., 1989; Smith, 1989; Jones and Smith, 1991)"
If an author or group of authors has published more than one article in a given year, you
can distinguish between these articles by placing a letter postscript after the publicationyear:
"(Black and Smith, 1990a; Black and Smith, 1990b)"
List all references in alphabetical order, sorted by the author(s)' last name(s). In cases
where the same author or group of authors has/have published multiple papers that you
have cited, then arrange these references in chronological order.. All authors must be
given in the reference list - the abbreviation "et al." Is used only in the text. The following
are examples of the punctuation, style and abbreviations that may be used for references
(note: the headings given here are not to be included in your reference list).
Journal article:
Jones, R.S., E.J. Gutherz, W.R. Nelson and G.C. Matlock. 1989. Burrow utilization by
yellowedge grouper, Epinephelusflavolimbatus, in the northwestern Gulf of Mexico. Env.
Biol. Fish. 26: 277-284.
Chapter in a Book:
-
8/6/2019 216 Lab Manual 2003-2004
19/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
19
Gross, M.T. 1984. Sunfish, Salmon and evolution of alternative reproductive strategies
and tactics in fishes. Pp. 55-57. In: G.W. Potts and R.J. Wooten (eds.) Fish reproduction:
strategies and tactics. Academic Press, London.
Book:
Siegel, S. 1956. Nonparametric statistics for the behavioral sciences. McGraw-Hill, New
York. pp. 312.
-
8/6/2019 216 Lab Manual 2003-2004
20/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
20
ORGANIC CHEMISTRY LABORATORIES
LABORATORY 1 - THE FRACTIONATION OF HYDROCARBONS AND GAS
CHROMATOGRAPHY (GC)
(2 LAB PERIODS)
I. INTRODUCTION
Mixtures of organic compounds can be separated by distillation. One such mixture is
crude oil. Oil refineries fractionate crude oil by distillation into products such as gasoline.Separation by distillation is based on the difference in boiling points. The composition
and uses of crude oil fractions is summarized in Table 1.1
Table 1.1 Petroleum constituents
FRACTION DISTILLATION
TEMPERATURE (bp),oC
CARBON NUMBER GENERAL USES
Gas below 20 C 1-C4 uses as for natural gas
Petroleum ether 20-60 C 5-C6 solvent, dry-cleaning
Ligroin (light
naphtha)
60-100 C 6-C7 useful as fuel or
chemicals feedstock
Natural gasoline 40-205 C5-C10, cycloalkanes straight run gasoline
Kerosene 175-325 C 12-C18, aromatics jet, tractor, heating fuel
Gas oil above 275 C 12 and higher diesel and heating fuel
Lubricating oil non-volatile liquids long chains attached
to cyclic structures
lubricants or cracked to
give lighter fractions
Asphalt or
petroleum coke
non-volatile solids polycyclic structures paving, coating and
structure uses
-
8/6/2019 216 Lab Manual 2003-2004
21/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
21
SEPARATION BY DISTILLATION
Simple Distillation:
In a simple distillation, the mixture of compounds to be separated is boiled, and the
vapors are condensed. The vapor phase is always richer in the more volatile compound
(lower boiling point). For instance, if we boil the liquid with a composition of X o as
shown in Figure 1.1, the vapor will have a composition of Y o, which is richer in the low
boiler. By condensing the vapors we get a head product rich in the low boiler, while the
bottom product will be nearly pure high boiler. This method gives good separation only
if there
is a reasonably big difference between the boiling points of the high and low boilers. The
method can be further refined by redistilling the condensed head fractions.
Reflux Cooling (Deflegmation):
The head product can be enriched to have more low boilers by using a reflux condenser
shown in Figure 1.2. To achieve this a fraction of the vapors is condensed back to the
Figure 1.1 Simple Distillation
-
8/6/2019 216 Lab Manual 2003-2004
22/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
22
boiler. This can be understood from Figure 1.3. When we boil a liquid with composition
Xo, the vapors would have a composition of Y o. The reflux cooler condenses a fraction of
the vapors, reducing the temperature to P. This point is in the two-phase region, thus the
vapor separates to the refluxing liquid with a composition of X D, going back to the boiler,
and the head product with a composition of Y D. It can be easily seen that the head
product is richer in the low boiler than in a simple distillation (Y o).
Rectification:
To further improve the separation, the refluxing liquid can be contacted with the
evaporating vapor - this is called rectification (Figure 1.4). To understand the principle of
rectification lets assume that the composition of the vapor, A, and the refluxing liquid, B,
Figure 1.3 Reflux coolingFigure 1.2 Reflux cooling setup
-
8/6/2019 216 Lab Manual 2003-2004
23/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
23
is equal, so we can represent it as P on Figure 1.5. The vertical position of P depends on
the ratio of A to B. P separates into two phases, with compositions X L and Y D. This is the
representation of the equilibrium of one theoretical plate. In the industrial distillation
towers there are a number of plates (with lower efficiency than the theoretical plates). For
the separation of components with similar boiling points, as many as 200 plates are used.
In the present lab, a three component system will be separated by a simplified
rectification. The reflux condenser to be used is called a Vigreux column, and the spikes
serve as plates to force the vapors to bubble through the condensing liquid, thereby
improving the separation. A better lab distillation head shown in Figure 1.6 almost
perfectly simulates the industrial towers (Figure 1.7).
Figure 1.4 Rectification setup Figure 1.5 Rectification
-
8/6/2019 216 Lab Manual 2003-2004
24/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
24
II. MATERIALS
unknown mixture of hydrocarbons (prepared by the demonstrator) consisting of
toluene, 2-methyl-butene and hexane
III. EQUIPMENT 1000ml round bottom boiling flask
fractionation column (Vigreux)
condenser
distributor (cow)
4, 250ml round bottom receiving flasks
green, plastic glassware clamps
heating mantle with power control unit
thermometer
clamps
Gas Chromatograph
Figure 1.7 Industrial distillationtower
Figure 1.6 Laboratory rectificationhead
-
8/6/2019 216 Lab Manual 2003-2004
25/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
25
-
8/6/2019 216 Lab Manual 2003-2004
26/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
26
IV. PROCEDURE
Before beginning the distillation, make an injection of the hydrocarbon (HC) mixture you
received from the TA into the gas chromatography (GC) unit. Refer to the GC section
under Instrumentation found in this lab manual for background information. The GC will
produce a strip chart with a number of peaks that relate to the volume percent of each of
the HCs in the mixture, as well as some other peaks. Normalize the peak areas of the
three HCs to equal 100 percent of the total peak area. (i.e.: use only the areas of the
peaks that relate to the HC peaks) Then calculate the percentage of each of the fractions
in the mixture as indicated by the GC analysis. These are the values that will determine
the success of your distillation. You can then proceed with the distillation of the HCmixture.
If you do your distillation correctly, the total volume of the fractions collected plus the
remaining bottom fraction (left in the boiling flask) should equal the volume of the
original mixture. Also, the percentage of each HC in the mixture should correspond to
the original GC analysis. Doing this will require several simple but necessary
calculations. The interesting part of the calculation is that your fractions will probably
be a mixture of several HCs, with one HC dominating. You will have to calculated the
total volume of each HC from the volume of that HC in each of the fractions and the
bottom. (The bottom is what is left over in the boiling flask at the end of the distillation)
Include in your report a table listing the volume of each HC collected in each fraction,
and the sum of each HC in the entire mixture. Convert these volumes to percentages, and
compare this to the original GC analysis of the small sample of the mixture, again in table
format. Remember to title each table and figure appropriately. Proper form and style are
very important, even if you are not satisfied with the results of your experiment.
To perform the distillation, set up the equipment as shown in Figure 1. Pour a measured
volume (300 mL) of a prepared HC mixture into the boiling flask. Prepare a table and
-
8/6/2019 216 Lab Manual 2003-2004
27/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
27
take time/temperature measurements throughout the distillation every 60 seconds. This
raw data will be plotted and submitted with the report. (do not submit the raw data)
Slowly heat the mixture until you have achieved a gentle boil and a steady reflux in the
Vigreaux column. As soon as you begin to see condensation occurring in the condenser,
the first temperature plateau has been reached. Maintain this temperature and collect the
liquid (first cut) in a collection flask. Usually a waste cut is taken to eliminate impurities,
but in this case we want to recover as much of the HC mixture as possible. Also, past
distillation experiments have shown that the majority of the low boiler is fractionated
very early in the experiment.
When the dripping stops, you have exhausted the first cut. Increase the temperature and
repeat the procedure. Keep up with the time/temperature data. When the second cut is
exhausted, you will be left with the third cut (bottom cut) in the boiling flask. Never letthe boiling flask run dry! At this point you will have completed the distillation. Turn
off the heat. After the equipment has cooled it can be dismantled and returned to the
storage bins underneath the lab benches. In the meantime, measure and record the
volume of each cut, and prepare a sample of each cut for storage in the lab fridge. Make
sure you have identified your samples in such a manner that anybody would be able to
continue with the experiment.
A portion of the samples you have collected will be injected into the GC. This is a pretty
straightforward operation. You will identify the fractions based on the time it takes for
the material to pass through the system. Different substances pass through the system at
different lengths of time, depending on their interaction with the G.C. column packing
material.. This is the basis of GC. Once you have created a GC trace for each cut, you
can then complete the calculations as outlined, and prepare your report in compliance
with the Guidelines for Preparation of Laboratory Reports included in your lab manual.
V. PRELAB QUESTIONS
Submit to the TA prior to beginning the laboratory session.
-
8/6/2019 216 Lab Manual 2003-2004
28/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
28
1) Draw the structures of toluene, 2-methyl-butene, and hexane. List their physical
properties (molecular weights, density, boiling point, refractive index and melting
point)
VI. SUMMARY OF REQUIREMENTS FOR REPORT
1) In table form show the calculated percentages of each fraction in the mixture prior
to distillation.
2) In table form show the time/temperature measurements recorded throughout the
distillation ( every 60 seconds). Plot the above.
3) Submit the GC trace for each cut and the basis for identification of the fraction
4) Tabulate the volume of each HC cut collected and the volume sum of all the HCfractions in the entire mixture. Convert volumes to percentages, and compare to the
original sample analyzed prior to distillation.
-
8/6/2019 216 Lab Manual 2003-2004
29/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
29
EXPERIMENT 2 : ANALYSIS OF PETROLEUM FRACTIONS
(2 LABORATORY PERIODS)
Purpose: To familiarize the student with the instruments and analysis used in the
petroleum and petrochemical industries.
PART 1. ASTM DISTILLATION OF PETROLEUM PRODUCTS (ASTM D86-61).
This method of testing is intended for use in the distillation of motor gasolines, aviation
gasolines, aviation turbine fuels, naphthas, kerosenes, oils, distillate fuel oils, and similar
petroleum products.
Definitions:
Initial Boiling Point -The thermometer reading which is observed at the instant that the
first drop of condensate falls from the lower end of the condenser tube.
End Point -The maximum thermometer reading attained during the test, which usually
occurs after the evaporation of all liquid from the bottom of the flask. The term
"maximum temperature" is a frequently used synonym.
Dry Point -The thermometer reading which is observed at the instant the last drop of
liquid evaporates from the lower point in the flask. Any drops or film of liquid on the side
of the flask or on the thermometer are disregarded.
Decomposition Point -The thermometer reading which coincides with the first indications
of thermal decomposition of the liquid in the flask.
-
8/6/2019 216 Lab Manual 2003-2004
30/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
30
NOTE: Characteristic indications of thermal decomposition are an evolution of fumes
and erratic thermometer readings which usually show a decided decrease after any
attempt is made to adjust the heat.
Percent Recovered -The volume in milliliters of condensate observed in the receiving
graduate, in connection with a simultaneous thermometer reading.
Percent Recovery -The maximum percent recovered, as observed in accordance with
Procedure.
Percent Total Recovery -The combined percent recovery and residue in the flask, as
observed in accordance with Procedure.
Percent Loss -100 minus the percent total recovery.
Percent Residue -The percent total recovery minus the percent recovery, or the volume of
residue in milliliters if measured directly.
Percent Evaporated -The sum of the percent recovered and the percent loss.
Outline of Method
A lOO-ml sample is distilled under the prescribed conditions which are appropriate to its
nature (Table 1). Systematic observations of thermometer reading and volume of
condensate are made, and from these data the results of the test are calculated and
reported.
Apparatus -A typical assembly of the apparatus is shown in Figure 1.
-
8/6/2019 216 Lab Manual 2003-2004
31/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
31
Preparation of Apparatus
(a) Fill the condenser box to cover the condenser tube with chopped ice. Add sufficient
water to cover the condenser tube. Prepare a similar cold-water bath for the graduate.
(b) Remove any residual liquid in the condenser tube by swabbing with a piece of soft,
lint-free cloth attached to a cord or copper wire.
(c) Measure 100 ml of the sample in the graduated cylinder and transfer it as completely
as practicable to the distillation flask, taking care that none of the liquid flows into the
vapor tube.
(d) Fit the thermometer, provided with a snug-fitting, well-rolled cork, tightly into the
neck of the flask so that the bulk is centered in the neck and the lower end of the capillaryis level with the highest point on the bottom of the inner wall of the vapor tube.
(e) Place the flask containing the 100-ml charge in its support; and by means of a cork
through which the vapor tube has been passed, make a tight connection with the
condenser tube. Adjust the flask so that it is in a vertical position and so that the vapor
tube extends into the condenser tube for a distance of 1 to 2 in.
-
8/6/2019 216 Lab Manual 2003-2004
32/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
32
(f) Place the graduate that was used to measure and charge, without drying, into its bath
under the lower end of the condenser tube so that the end of the condenser tube is
centered in the graduate and extends therein for a distance of at least 1 in, but not below
the 100-ml mark. Cover the graduate closely with a piece of blotting paper so similar
material, suitably weighted, which has been cut to fit the condenser tube snugly. Maintain
the level of the bath around the graduate so that it is at least as high as the 100-ml mark.
(g) Note and record the prevailing barometric pressure, and proceed at once with the
distillation as directed in the following section.
1.4 Procedure
Apply heat to the distillation flask and contents. Immediately after observing the initial boiling point, move the graduate so that the tip of the condenser touches its inner wall.
Continue to regulate the heating so that the rate of condensation into the graduate is 4-5
ml/min.
In the interval between the initial boiling point and the end of the distillation, observe and
record whatever thermometer readings at prescribed percentages recovered, and/or
percentages recovered at prescribed thermometer readings are necessary for the
calculation and reporting of the results of the test as prescribed in the following section.
Record all volumes in the graduate to the nearest 0.5 ml and all thermometer readings to
the nearest 1.0F (0.5C).
NOTE : In cases in which no specific data requirements have been indicated, record the
initial boiling point, the end point or dry point or both, and thermometer readings at 5%
and 95% recovered and at each multiple of 10% recovered from 10% to 90% inclusive.
If either a thermometer reading of 700F (371C) or a decomposition point is observed,
discontinue the heating and resume the procedure as directed in the second paragraph
following. Otherwise, continue the heating until the end point is reached.
-
8/6/2019 216 Lab Manual 2003-2004
33/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
33
Observe and record the end point or dry point or both, as required, and discontinue the
heating. At the end point, observe if all the liquid has evaporated from the bottom of the
flask. If not, include a note of this fact in the report as prescribed in the following section.
While the condenser tube continues to drain into the graduate, observe the volume of
condensate at 2-minute intervals until two successive observations agree. Measure this
volume accurately and record it, to the nearest 0.5 ml, as percent recovery. If the
distillation was previously discontinued under the conditions given in the second
paragraph preceding, deduct the percent recovery from 100, report this difference as
"Percent Residue and Loss", and omit the procedure given in the next two paragraphs.
After the flask has cooled, pour its contents into the condensate in the graduate and allow
to drain until no appreciable increase in the volume of liquid in the graduate is observed.Record this volume, to the nearest 0.5 ml, as percent total recovery.
Deduct the percent total recovery from 100 to obtain the percent loss.
1.5 Calculations and Reporting
Report (and plot) the data specified under section 1.1 and the pressure under which the
measurements were made.
1.6 Question
(a) How many components are present in the sample? What are they?
PART 2. API GRAVITY OF PETROLEUM AND ITS PRODUCTS (ASTM D287-
55)
This method describes a procedure for the determination by means of a glass hydrometer
of the API gravity (in vacuum) of crude petroleum and of petroleum products normally
handled as liquids and having a Reid vapor pressure of 26 lbs. or less. Results are
determined at 60F or converted to values at 60F by means of standard tables.
-
8/6/2019 216 Lab Manual 2003-2004
34/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
34
Definition -API Gravity is defined by the following equation:
API Gravity. deg. = (141.5/ (sp.gr.60/60F )) -131.5
2.1 Apparatus
The following apparatus is required:
Hydrometers , of glass, graduated in degrees API (in vacuum) as listed in Table l and
conforming to the Tentative Specifications for ASTM Hydrometers (ASTM Designation:
EI00). For routine. testing of petroleum products, the long form of hydrometer (IH to
10H) should be used.
Thermometer , having a range of -5F to +215F and conforming to the requirements for
Thermometer 12F as prescribed in ASTM Specifications El.
Hydrometer Cylinders , made of metal, clear glass, or plastic. For convenience in
pouring, the cylinder may have a lip on the rim. The side diameter of the cylinder shall be
at least 25 mm greater than the outside diameter of the hydrometer used in it. The height
of the cylinder shall be such that the length of the column of sample it contains is greater
-
8/6/2019 216 Lab Manual 2003-2004
35/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
35
by at least 25 mm than the portion of the hydrometer which is immersed beneath the
surface of the sample.
Temperature of Test -The gravity determined by the hydrometer method is most
accurate at or near the standard temperature of 60F; use this or any other temperature
between OF and 195F for the test, so far as it is consistent with the type of sample and
necessary limiting conditions shown In Table 2.
-
8/6/2019 216 Lab Manual 2003-2004
36/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
36
2.2 Procedure
Pour the sample into the clean hydrometer jar without splashing so as to avoid the
formation of bubbles and to reduce to a minimum the evaporation of the lower-boiling
constituents of the more volatile samples. For the more volatile samples, transfer to the
hydrometer cylinder by siphoning. Remove any air bubbles formed, after they have
collected on the surface of the sample, by touching them with a piece of clean filter paper
before inserting the hydrometer. Place the cylinder containing the sample in a vertical
position in a location free from air currents. Take precautions to prevent the temperature
of the sample from changing appreciably during the time necessary to complete the test.
During this period, the temperature of the surrounding medium should not change any
more than 5F.
Lower the hydrometer gently into the sample, and when it has settled, depress it about
two scale divisions into the liquid and then release; keep the rest of the stem dry as
unnecessary liquid on the stem changes the effective weight of the instrument and so
affects the reading obtained. With samples of low viscosity, a slight spin imparted to the
instrument on releasing assists in bringing it to rest, floating freely away from the walls
of the hydrometer cylinder. Allow sufficient time for the hydrometer to become
completely stationary and for all air bubbles to come to the surface. This is particularly
necessary in the case of the more viscous samples.
When the hydrometer has come to rest, floating freely, and the temperature of the sample
is constant to O.2F, read the hydrometer to the nearest scale at which the surface of the
liquid cuts the scale. Determine this point by placing the eye slightly below the level of
the liquid and slowly raising it until the surface, first seen as a distorted ellipse, appears to
become a straight line cutting the hydrometer scale.
To make a reading with nontransparent oils, observe the point on the hydrometer scale to
which the sample rises above its main surface, placing the eye slightly above the plane
surface of the liquid. This reading requires a correction. Determine this correction for the
-
8/6/2019 216 Lab Manual 2003-2004
37/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
37
particular hydrometer in use by observing the height above the main surface of the liquid
to which the oil rises on the hydrometer scale when the hydrometer in question is
immersed in a transparent oil having a surface tension similar to that of the sample under
testing.
Observe the temperature of the sample to the nearest O.25F immediately before and
after the observation of the gravity, the liquid in the cylinder being thoroughly but
cautiously stirred with the thermometer, the whole of the mercury thread being immersed.
Should these temperature readings differ by more than 1F, repeat the temperature and
gravity observations when the temperature of the sample has become more stable. Record
the mean of the thermometer reading before and after the final hydrometer reading, to the
nearest degree Fahrenheit, as the temperature of the test.
NOTE: When thermo-hydrometers are used, stir the sample by carefully raising and
lowering the hydrometer. It is satisfactory in this case to read the thermometer scale after
the hydrometer reading has been observed.
2.3 Result
Report the API gravity of the petroleum sample provided.
PART 3. SAYBOLT VISCOSITY (ASTM D88-56)
This method describes procedures for the empirical measurement of Saybolt viscosity of
petroleum products at specified temperatures between 70F and 210F. A special
procedure for waxy and resinous materials is also included.
NOTE: A fundamental and preferred method for measuring viscosity is by the use of
kinematic viscometers as outlined in ASTM Method D445, Test for Kinematic Viscosity.
This method requires smaller samples, less time, and gives greater accuracy.
-
8/6/2019 216 Lab Manual 2003-2004
38/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
38
Saybolt Universal and Saybolt Furol viscosities may be obtained from kinematic
viscosity values by the use of conversion tables given in ASTM Method D446 -
Conversion of Kinematic Viscosity to Saybolt Universal Viscosity; and ASTM Method
D666 -Conversion of Kinematic Viscosity to Saybolt Furol Viscosity, respectively.
Definition
(a) Saybolt Universal Viscosity -The efflux time in seconds of 60 ml of sample flowing
through a calibrated Universal orifice under specified conditions.
(b) Saybolt Furol Viscosity -The efflux time in seconds of 60 ml of sample flowing
through-a calibrated Furol orifice under specified conditions.
The Furol viscosity is approximately one-tenth the Universal viscosity and isrecommended for those petroleum products having viscosities greater than 1000 seconds
(Saybolt Universal), such as fuel oils and other residual materials.
The word "Furol" is a contraction of fuel and road oils.
3.1 Outline of Method
The efflux time in seconds of 60 ml of sample, flowing through a calibrated orifice, is
measured under carefully controlled conditions. This time is corrected by an orifice factor
and reported as the viscosity of the sample at that temperature.
3.2 Apparatus
Viscometer and Bath -The viscometer shall be constructed entirely of corrosion-
resistant metal. The orifice tip, Universal or Furol, may be constructed as a replaceable
unit in the viscometer. Provide a nut at the lower end of the viscometer for fastening it in
the bath. Mount vertically in the bath and test the alignment with a spirit level on the
plane of the gallery rim. Provide a cork or other suitable means to prevent the flow of
sample until the start of the test; a small chain or cord may be attached to the cork to
facilitate rapid removal. The bath serves both as a support to hold the viscometer in a
vertical position as well as the container for the bath medium. Equip the bath with
-
8/6/2019 216 Lab Manual 2003-2004
39/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
39
effective insulation, with a coil for heating and cooling, and with thermostatically
controlled heaters capable of maintaining the bath within the functional parameters given
in Table 3. The heaters and coil should be located at least 3 in. from the viscometer.
Provide a means for maintaining the bath medium at least 1/4 in. above the overflow rim.
The bath media are given in Table 3.
-
8/6/2019 216 Lab Manual 2003-2004
40/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
40
-
8/6/2019 216 Lab Manual 2003-2004
41/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
41
3.3 Preparation of Apparatus
Use a Universal orifice for lubricants and distillate materials with efflux times greater
than 32 seconds to give the desired accuracy. Liquids with efflux times over 1000
seconds are not conveniently tested with this orifice.
Use a Furol orifice for residual materials with efflux times greater than 25 seconds. The
Furol eff1ux is approximately one-tenth the Universal efflux time.
-
8/6/2019 216 Lab Manual 2003-2004
42/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
42
-
8/6/2019 216 Lab Manual 2003-2004
43/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
43
NOTE : The Universal orifice is used at 70, 100, 130 and 210F. The Furol orifice is
used at 77, 110, 122 and 210F.
Set up the viscometer and bath where they will be free from drafts and rapid changes in
air temperature. Locate them so that the sample will not be contaminated by dust or
vapors during the test.
Viscosity determinations shall not be made at temperatures below the dew point of the
room's atmosphere. Room temperatures up to 100F will not introduce errors in excess of 1.0 percent. For standardization and referee tests, the room temperature shall be kept
between 68 and 86F and the actual temperature recorded.
Fill the bath at least 1/4 in. above the overflow rim of the viscometer. Table 4 lists
recommended bath media for each test temperature.
Provide adequate stirring and thermal control for the bath so that the sample will not
fluctuate more than 0.05F after reaching the test temperature.
Clean the viscometers with an effective nontoxic solvent and remove all solvent from the
gallery and viscometer.
NOTE : The plunger commonly supplied with the viscometer should never be used for
cleaning as the overflow rim and walls of the viscometer may be damaged by its use.
-
8/6/2019 216 Lab Manual 2003-2004
44/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
44
Calibration of Viscometer
Calibrate the Saybolt Universal viscometer at periodic intervals by measuring the efflux
time at 100F of an appropriate viscosity standard, following the procedure for standards
given below.
Viscosity standards are available from two sources. These standards may be used with
equal confidence provided they are used immediately after opening and not stored for re-
use as permanent viscosity standards.
Standards Conforming to ASTM Saybolt Viscosity Standards
Viscosity Oil Standards conforming to the requirements of ASTM Viscosity Oil
Standards for viscometer calibration having certified Saybolt viscosity values established
by cooperative determinations of kinematic viscosity values. The kinematic values are
converted to Saybolt Universal and Saybolt Furo1 viscosity values by means of
conversion tables given in ASTM Methods 0446 and 0666, respectively. The Viscosity
Oil Standards are oils with approximate Saybolt viscosities as shown in Table 5.
3.4.2 NBS Viscosity Standards
National Bureau of Standards liquid viscosity standards having accurate values supplied
with each sample. Standard SB has a Saybolt Universal viscosity of approximately 300
sec. at 100F. Standard SF has a Saybolt Furol viscosity of approximately 100 seconds at
122F.
3.4.3. Routine Calibrations
The viscosity standards may also be used for routine calibrations at other temperatures as
shown in Table 5.
-
8/6/2019 216 Lab Manual 2003-2004
45/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
45
3.4.4 Other Viscosity Standards
Other reference liquids, suitable for routine calibrations, may be established by selecting
stable oils covering the desired range and determining their viscosities in a viscometer
calibrated with a standard conforming to ASTM requirements or an NSB standard as
described herein.
The efflux time should equal the certified Saybolt viscosity value. If the efflux timediffers from the certified value by more than 0.2%, calculate a correction factor, F, for the
viscometer as follows:
F = V/T
where F = certified Saybolt viscosity of the standard, and
-
8/6/2019 216 Lab Manual 2003-2004
46/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
46
t = efflux time in seconds at 100F.
NOTE : The correction factor applies to all viscosity levels and for all temperatures,
provided the calibration is based on viscosity standard having an efflux time between 200
and 600 sec.
Calibrate the Saybolt Furol viscometer at 122F in the same manner as above, using a
viscosity standard having a minimum efflux time of 90 seconds.
Viscometers or orifices which have corrections in excess of 1.0 percent shall not be used
for routine testing.
3.5 Procedure
If the test temperature is above room temperature, the test may be expedited by
preheating the sample to not more than 3F of its flash point (see ASTM Method D93).
Test for Flash Point by Means of the Pensky-Martens Closed Tester), as volatility losses
may alter its composition.
Insert a cork stopper, having a cord attached for its easy removal, into the air chamber at
the bottom of the viscometer. The cork shall fit tightly enough to prevent the escape of
air, as evidenced by the absence of oil on the cork when it is withdrawn.
Filter the prepared sample through a 100-mesh screen directly into the viscometer until
the level is above the overflow rate.
Stir the sample until its temperature remains constant within 0.05F of the test
temperature during 1 minute of continuous stirring. Stir with a viscosity thermometer
equipped with a thermometer support (Figure 3). Use a circular motion at 30 to 50 r.p.m.
in a horizontal plane.
-
8/6/2019 216 Lab Manual 2003-2004
47/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
47
NOTE : Never adjust the temperature by immersing hot or cold bodies into the sample.
Such thermal treatment may affect the sample and the precision of the test.
Remove the thermometer from the sample. Quickly remove the oil from the gallery until
its level is below the overflow rim. This is done by placing the tip of the withdrawal tube
at one point in the gallery and applying suction. Do not touch the overflow rim with the
withdrawal tube, or the effective head of the sample will be reduced.
Place the receiving flask where the stream of oil from the bottom of the viscometer will
just strike the neck of the flask. The graduation mark on the flask shall be between 10 and
13 cm from the bottom of the viscometer tube.
Snap the cork from the viscometer using the attached cord. At the same instant start the
timer. Stop the timer the instant the bottom of the meniscus reaches the graduation mark.
Record the efflux time in seconds.
Report the corrected time in seconds as the Saybolt Universal viscosity or Saybolt Furol
viscosity of the oil at the temperature at which the test was made.
Report the values below 200 seconds to the nearest 0.1 second. Report all values of 200
seconds or higher to the nearest whole second.
-
8/6/2019 216 Lab Manual 2003-2004
48/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
48
EXPERIMENT 3 : REDUCTION OF A KETONE TO A SECONDARY ALCOHOL
(1 LABORATORY PERIOD)
INTRODUCTION
This experiment is intended to serve as an example of how to carry out an organic
reaction. You will find a reaction which may appear as one line in a textbook actually
involves several steps, including the essential ones of isolating and purifying the product.
You should review the technique for taking a melting point and recrystallizing a
compound BEFORE COMING TO THE LABORATORY. The procedure is explained
in the Appendix I.
The carbonyl group of aldehydes and ketones may be converted to a hydroxyl group by a process known as reduction. Aldehydes may be reduced to primary alcohols, while
ketones yield secondary alcohols.
You will notice that the process of reduction involves the addition of a molecule of
hydrogen across the double bond of the carbonyl group.
The most general method of reducing compounds with carbon-oxygen double bonds is
reaction with lithium aluminum hydride. In this experiment, however, you will use
sodium borohydride to reduce a ketone (benzophenone) to an alcohol (benzhydrol).
Lithium aluminum hydride and sodium borohydride will both reduce aldehydes and
ketones to alcohols; the main difference between the two reagents is that lithium
aluminium hydride is a much more reactive reducing agent than sodium borohydride,
which makes it more difficult to handle safely. For example, lithium aluminium hydride
-
8/6/2019 216 Lab Manual 2003-2004
49/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
49
reacts explosively with water and readily decomposes in moist air, whereas with sodium
borohydride the hydrolysis in water is sufficiently slow at room temperature to allow its
use as a reducing agent in an aqueous medium.
Very many organic compounds are insoluable in water, so it is necessary to use ethanol
as a co-solvent in sodium borohydride reductions. However, sodium borohydride reacts
rapidly with ethanol and it is therefore necessary to use a large excess of sodium
borohydride to overcome the effect of this solvolysis and to ensure that enough is present
to reduce the carbonyl group.
Theoretically, one mole of sodium borohydride will reduce four moles of benzophenone
to yield a borate ester.
4C 6H5COC 6H5 + NaBH 4 Na[(C 6H5)2CHO] 4B (3)
This borate ester may be hydrolyzed with aqueous sodium hydroxide to yield benzhydrol.
The sodium hydroxide acts as a catalyst for the hydrolysis.
Na[(C 6H5)2CHO] 4B + 4H 2O 4C 6H5-CH.C 6H5 + NaB(OH) 4 (4)
|
OH
The product benzhydrol is soluble in the ethanol-water mixture which was used for the
reaction but is insoluble in water. Thus, if the reaction mixture is diluted with cold water
the product will precipitate. Sodium borate remains in solution since it is a salt and very
soluble in water.
PROCEDURE:
In a 100 ml Erlenmeyer flask, dissolve 6.0 g of benzophenone in 50 ml of ethanol. In a 50
ml beaker dissolve 0.6 g of sodium borohydride in 25 drops of distilled water, and using a
medicine dropper add this solution drop-wise, with swirling, to the solution of
benzophenone. Swirl the mixture from time to time for 15 minutes. Add 4 ml of a 6N
aqueous sodium hydroxide solution and boil the reaction mixture, using a boiling chip, on
-
8/6/2019 216 Lab Manual 2003-2004
50/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
50
a steam bath for 10 minutes. Pour the reaction mixture into 100 ml of cold water and ice
(100 ml total) and
collect the resulting precipitate of benzhydrol on a Buchner funnel. Dry the crude product
for 15 min. at 60C under suction, weigh it, and then recrystallize from a minimum
volume of 95% ethanol (b.p. 78.2C).
Weigh your recrystallized product and determine the percentage yield of crude product
and of recrystallized product and the melting point of your pure benzhydrol product.
Submit a sample for grading.
Take an infrared spectrum of your starting material (benzophenone) and your purified
final product (benzhydrol). Discuss differences and point out characteristic absorption peaks.
-
8/6/2019 216 Lab Manual 2003-2004
51/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
51
LABORATORY 4 - CATALYTIC HYDROGENATION OF SQUALENE (A
POLYUNSATURATED HYDROCARBON)
(1 LAB PERIOD)
I. INTRODUCTION
Hydrogenation is the general method of converting a carbon-carbon double bond or triple
bond ( -bonds) into a carbon-carbon single bond ( -bond). The energy of activation for
the reaction is very high. Therefore, a catalyst is needed to reduce this activation energy
and enable the reaction to proceed at room temperature.
Hydrogenation is frequently used in the food industry to convert unsaturated fats such as
soybean oil and cottonseed oil into saturated solids. This is done to increase the shelf-
life of oils by removing the oxidation-sensitive double bonds. Saturated oils (fats) are
solids having a consistency similar to that of lard. However, from the nutritional point of
view, saturated fats are less healthy because they are harder to break down and cause
deposition on the arterial wall; unsaturated oils, especially polyunsaturated oils, areconsidered to be healthier.
Table 2.1 lists the fatty acid content of several commonly consumed oil products.
Table 2.1 Approximate fatty acid content of some fats (percentages)
ACIDmelting point.
(oC)
BUTTER FAT
COTTONSEEDOIL
OLIVEOIL
PALMOIL
LINSEED
OILButyric, C 3H7COOH 3-4Caproic, C 5H11COOH 1-2Caprylic, C 7H15COOH 1-2Capric, C 9H19COOH 2-4Lauric, C 11H23COOH 3-5Myristic, C 13H27COOH 9-11 1-4 0.2Palmitic, C 18H31COOH62.9
22-30 20-23 7-15 35-45 5
-
8/6/2019 216 Lab Manual 2003-2004
52/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
52
Stearic, C 17H35COOH69.9
7-15 1-3 1-4 1-8 4
Oleic, CH 3(CH 2)7CH=CH(CH 2)7COOH16
29-40 24-30 70-85 40-50 6-12
Linoleic, C 17H33COOH
-5
3-5 42-54 4-12 2-11 26-46
Linolenic,CH 3(CH 2)4CH=CHCH 2CH=CH(CH 2)7COOH
36-50
For the purpose of the following lab a polyunsaturated hydrocarbon, squalene, will be
hydrogenated in the presence of 10% palladium on carbon support as a catalyst.
Squalene is a naturally occurring polyunsaturated hydrocarbon. Whale oil is rich in
squalene. It is built from polyisoprene blocks. Note that isoprene (2-methyl-butadiene)
is the basic building block of several natural compounds, including cholesterol and
steroids. The multiple double bonds of squalene (6) will disappear due to the addition of
hydrogen, and IR analysis results will show an obvious distinction between the
hydrogenated (saturated), and unhydrogenated (unsaturated) squalene.
Squalene
CH 3
CH 3
CH 3 CH 3 CH 3
CH 3 CH 3
H 3 C
-
8/6/2019 216 Lab Manual 2003-2004
53/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
53
Squalane
Figure 2.1. Squalene & Squalane
By knowing the stoichiometry of one hydrogen molecule reacting with each double bond,
hydrogenation can also be used as a quantitative tool to determine the number of double
bonds in an organic compound.
II. MATERIALS
squalene
hydrogen (already set up)
chloroform
catalyst (10% Pd on carbon support)
III. EQUIPMENT
Hydrogenation Apparatus as seen in Figure 2.2
H3
CCH 3
CH 3 CH 3 CH 3
CH 3 CH 3 CH 3
-
8/6/2019 216 Lab Manual 2003-2004
54/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
54
IV. PROCEDURES
BE CAREFUL WITH HYDROGEN GAS. IT IS HIGHLY FLAMMABLE! Mixed
with oxygen in a 2:1 ratio it forms an explosive gas.
Before you start, study the apparatus and compare with the one in the notes. Stopcock A
is 3-way so gas can be allowed to fill different parts of the system. Also read the
introduction on GAS CYLINDERS found at the beginning of this manual.
The hydrogenation reaction takes place in a closed system made up of the gas burette
connected to the reaction flask. However, the system is full of air and this needs to be
exchanged with hydrogen before the reaction is started.
The hydrogen cylinder is equipped with a two-stage regulator. The high pressure gauge
(closer to the cylinder) shows the pressure in the cylinder when you open the main valve
on the cylinder.
The low pressure gauge (second stage) will show the pressure of the gas going to your
apparatus. Open the first stage to tank pressure and set the second stage to about 2.5
pounds per square inch pressure. DO NOT EXCEED 6 POUNDS/INCH 2 ON THE
SECOND STAGE. Open the needle valve which connects the cylinder to your apparatus.
The pressure on the low gauge may drop. Readjust it.
To clear air out of the system: first open stopcock B and turn stopcock A (3-way) so that
it connects the gas burette to the exhaust line. See Figure 4 of 2.2. Expel all air present in
the gas burette by holding the reservoir up until the oil pushes all the air out;
approximately 1cm below stopcock A. Then turn stopcock A (3-way) so that H 2 gas can
flow through the system and exit via the exhaust tube without entering the gas burette.
See Figure 2 of 2.2. Flush the system with hydrogen and shut off flow at the regulator
needle valve.
To prepare the gas burette with hydrogen ready for the reaction:
-
8/6/2019 216 Lab Manual 2003-2004
55/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
55
Turn the three way stopcock A to the position which allows H 2 to enter the gas burette
See Figure 3 of 2.2. Carefully turn on the hydrogen flow and fill the burette with
hydrogen gas 5-6cm from the bottom of the burette.
-
8/6/2019 216 Lab Manual 2003-2004
56/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
56
Figure 2.2 Hydrogenation apparatus
-
8/6/2019 216 Lab Manual 2003-2004
57/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
57
Once the burette is full of hydrogen, turn stopcock A back to the bypass position (Figure
2 of 2.2).
To set up the reaction flask: pipette 1ml of squalene into 19 ml. Chloroform. Place 5 ml.
Of this mixture into the reaction flask along with a magnetic stirrer. Add about 80mg of
10% palladium on charcoal to the reaction flask. This is the catalyst.
Flush the flask with hydrogen for 2-3 minutes using the gas inlet. When flushing, be sure
to hold the glass stopper loosely to allow the hydrogen to enter the flask and the air to
leave out the sides.
Before starting the reaction record the volume of H 2 in the gas burette. This is done bymatching the levels of oil in the burette and the reservoir. Then move the reservoir to its
maximum height to exert a small pressure on the hydrogen in the closed system.
To start the reaction close the main valve on the H 2 cylinder. Place the gas inlet tube
snugly into the flask and turn the three-way stopcock A to connect the gas burette with
the reaction flask (see Figure 4 of 2.2). Stopcock C should also be open to allow
hydrogen into the reaction flask. Turn on the magnetic stirrer and let the reaction proceed
until there is no longer a change in H 2 volume (about one-half hour). Check the extent of
the reaction by matching the oil in the burette and reservoir. Record the amount of H 2
every 5 minutes.
During the reaction observe whether the flask is cooling or warming up.
Stop the stirrer when there is no longer a decrease in H 2 over a 5-10 minute period.
Match the oil level in the gas burette with that in the oil reservoir and record the final
volume of hydrogen and the room temperature. Close the stopcocks and remove the
reaction flask. By matching the oil levels before and after the reaction, the final volume
reading is obtained at a pressure of 1 atm.
-
8/6/2019 216 Lab Manual 2003-2004
58/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
58
Release the pressure from the apparatus through the exhaust line by opening stopcocks A
and B and loosening the membrane valve (second stage) all the way. Both gauges should
read zero psig.
Filter off the catalyst using vacuum filter apparatus and buchner funnel. Place the
catalyst in the vessel for Pd residues.
V. INFRARED ANALYSIS
For the theory of infrared analysis refer to the Appendix.
Procedure:1. Rinse IR cell with CHCl 3 and run CHCl 3 sample, scanning from 300 to 700.
2. Run sample of original unsaturated squalene and CHCl 3.
3. Run sample of hydrogenated squalene and CHCl 3.
Compare differences in peaks around area where double bonds are detected. Note, this
will only provide you with a qualitative analysis, not a quantitative analysis.
Make photocopies of the three IR charts for each member of the lab group and submit
them along with the formal lab report. Remember to properly label the figures and
clearly indicate the observed changes in the peaks.
VI. PRE-LAB QUESTIONS
To be completed before starting the laboratory please check with T.A. for proper
calculations
1) Write down the structure of isoprene and squalene. List the physical properties of
squalene ( mol. wt., density, boiling point, melting point) and write the stochiometric
equation for the hydrogenation of squalene.
2) Calculate the theoretical volume of H 2 required to completely hydrogenate your
-
8/6/2019 216 Lab Manual 2003-2004
59/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
59
squalene sample (hint: use ideal gas law).
VII. SUMMARY OF REQUIREMENTS FOR LABORATORY REPORT
1) From the volume of hydrogen used, calculate the number of grams and moles of hydrogen consumed in the reaction.
2) Given that squalene has a density of 0.858 g/cm 3 at 25 0C, how many grams of hydrogen were taken up per 100g of squalene? How many moles of H 2 are taken up
by 1 mole of squalene? (Use the molar weight, MW = 410.73gmol -1 of squalene for
your calculations.)
3) How many moles of hydrogen reacted per mole of palladium present?Is hydrogenation exothermic or endothermic and why?
-
8/6/2019 216 Lab Manual 2003-2004
60/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
60
LABORATORY 5 EPOXIDATION OF 1-TMP AND IODINE TITRATION
(2 LAB PERIODS)
I. INTRODUCTION AND THEORY
Compounds containing the three-membered ring CHCH
O
, are called
epoxides. Epoxides are ethers with unusual properties due to the strained ring structure
present. One method to prepare them is to use peroxybenzoic acid to oxidize alkenes or
double bonds in chloroform, dichloromethane or ether as solvent. Once prepared, the
three-membered ring easily undergoes acid- and base-catalyzed cleavage. An example of
the mechanism and orientation during the cleavage of isobutylene oxide, a substituted
epoxide is shown:
H3C C CH 2
CH 3
O
+ H3C C CH 2OHCH 3
OH
H3C C CH 2
CH 3
O
+ CH 3 O HCH 3ONa
H+
H3C C CH 2
CH 3
OH
OCH 3
Acid-catalyzed cleavage:
Base-catalyzed cleavage:
HOH
The following experiment makes use of the fact that each double bond reacts with one
mole of peroxyacid to form the epoxide ring. The reaction for 2,4,4-trimethylpentene-1
(TMP-1) is shown:
-
8/6/2019 216 Lab Manual 2003-2004
61/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
61
The mol% unsaturation can be calculated by back-titration of any unreacted peracid. This
is done by reacting the leftover (unreacted) acid with potassium iodide, which will release
I2 exactly equivalent to 2 mols of peroxiacid:
The liberated I 2 is then titrated with Na 2S2O3. The reaction for the titration is:
I2 + 2 Na 2S2O3 Na 2S4O6 + 2 NaI (3)
From reaction (3) it is clear that 2 mol Na 2S2O3 is equivalent with one mol I 2 (1N
Na 2S2O3 = 0.5M Na 2S2O3)
Iodine titration is used very often in industry for double bond titration; it is also used for water analysis at the ppm level (Carl-Fischer titration).
H 3 CC
CH 3H 3 C
CH 2
C
CH 2
CH 3
+ ClC O
OHO
H 3 CC
CH 3H 3C
CH 2
C
H 2 C
CH 3
O
+ ClC
OHO
1-TMP3-Chloroperoxybenzoic acid
1-TM POxide 3-Chlorobenzoic acid
ClC O
OHO
+2KICl
C OKO
+ I2 + +H2O 1/2O 22 2
-
8/6/2019 216 Lab Manual 2003-2004
62/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
62
II. MATERIALS
3-Chloroperoxybenzoic acid, (CPBA)
sodium phosphate
potassium dihydrogen phosphate
dichloromethane
glacial acetic acid
10w% potassium iodide
0.1N sodiumthiosulfate
starch indicator
water bottle with distilled water
III. EQUIPMENT
3, 1L beakers
stirring plate and magnetic stirrers
vacuum oven or desiccators
100ml volumetric flask
3, 250ml Erlenmeyer flasks
1ml, 10ml pipettes
timer
4, 5ml and 1, 10ml graduated cylinder
watch-glasses
burette
nitrile gloves
-
8/6/2019 216 Lab Manual 2003-2004
63/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
63
IV. PROCEDURES
PART A (WEEK 1)
A) PURIFICATION OF PERACID:
During storage 3-Chloroperoxybenzoic acid decomposes to give a mixture of the peroxy
(a) and the carboxylic (b)
acids. Therefore, purification
is needed to remove the
carboxylic acid by a buffer
which selectively dissolves
the acid but leaves the desired peroxyacid or peracid insoluble.
Step 1: Prepare the buffer solution
Dissolve 67.1g of sodium phosphate (Na 2HPO 4.7H 2O) in 500ml of distilled water. Do the
same with 34g of potassium dihydrogen phosphate (KH 2PO 4) in a separate beaker. Once
the solids are completely dissolved mix the two solutions.
Step 2: Washing of the peracid
Take 50g of raw ( 50% purity) 3-Chloroperoxybenzoic acid (CPBA) and place it in a
large beaker. Add about 200ml of buffer and stir it well using a stirring rod. Let the solid
settle down and decant the liquid. Repeat 4 more times. (The settling of the solid takes
time so set up the Buchner funnel). Next, wash the peracid 3 times the same way with
distilled water. The last 2 washes should be done on a Buchner funnel to filter the solid
(see discussion on Buchner funnel at the beginning of the lab notes).
Cl
C O
O
OH
Cl
C OH
O
3-Chloroperoxybenzoic acid and 3-Chlorobenzoic acid(a) (b)
-
8/6/2019 216 Lab Manual 2003-2004
64/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
64
Step 3: Drying of the peracid
Weigh an empty beaker, sufficient in size to hold all of your peroxyacid. Transfer the
peroxyacid from the Buchner funnel into the beaker. This is the pure peroxyacid, which
should be dried. Label the beaker with the name of compound, your name, group number
and date. Place in a desiccator. The following week the peracid should be dry and ready
to use. Measure the amount recovered and do a %yield calculation.
PART B (WEEK 2)
B) EPOXIDATION OF 1-TMP AND IODINE TITRATION
FOR YOUR PROTECTION PLEASE WEAR NITRILE GLOVES (BLUE) FOR
STEP 4/a OF THIS EXPERIMENT. The dichloromethane used in this experiment
dissolves the latex gloves and is harmful to your skin.
Step 4a) Preparation of 0.1M CPBA
In a weighing boat measure out the amount needed to make 50ml of 0.1M CPBA from
the one that you made last week. Record the exact weight used to 4 decimal point
accuracy. UNDER A FUMEHOOD transfer this amount into a clean, dry, 50ml
volumetric flask using a funnel. Rinse the boat and the funnel with dichloromethane
(CH 2Cl2), and add some more CH 2Cl2 to dissolve the solid. Wait until complete
dissolution and fill up the volumetric flask to the mark. Calculate the exact molarity of
your peracid solution and record it in your notebook. Stopper the flask, label it, wrap it in
aluminum foil and label it on the outside again. Peroxyacid is heat and light sensitive and
every precaution should be taken to prevent it from decomposing.
Step 4b) Epoxidation and iodine titration
-
8/6/2019 216 Lab Manual 2003-2004
65/101
E.S. 216 Industrial Organic Processes Laboratory ManualProf. P. Charpentier 2003-2004
65
The TMP-1 is epoxidized by reacting it with excess CPBA in a dichloromethane solution.
After 10 minutes reaction time the amount of leftover peracid is determined by reacting it
with excess KI and backtitrating the liberated I 2 with sodium thiosulfate (Na 2S2O3). By
knowing the stoichiometry of each reaction, the extent of the epoxidation can be
determined.
2 mols of peracid releases 1 mol of I 2. 1 mol of I 2 reacts with 2 mols of Na 2S2O3. Thus, 1
mol of peracid will be equivalent to 1 mol of Na 2S2O3. The Na 2S2O3 is normalized
against I 2 so 1 mol I 2 is equivalent to 2 mols Na 2S2O3. Therefore, the normality of 1 M
Na 2S2O3 = 2 N. Also, 10ml of 1 M m-CPBA will be equivalent to 10ml of 1 M Na 2S2O3
or 20ml of 1 N Na 2S2O3. (See reaction equations in the INTRODUCTION section).
Step 4b.1) Blank titration
Since one can never make an exact 1 M solution, the exact amount of the peracid in the
0.1 M solution made in Step 4 a will be determined by a blank titration. Thus, 10ml of
peracid solution is reacted by excess KI, and the liberated I 2 is backtitrated by Na 2S2O3. If
the peracid solution were perfect, 10ml of peracid should be equivalent to 20ml of 0.1 N
sodium