2100Q and 2100Q IS
Training
D. Slater
N. JanssenSeptember 29, 2009
-
7/30/2019 2100Q Sales Training
2/27
2
2100Q replaces 2100P
2100QIS
replaces 2100P IS
Meets EPA Method 180.1
Meets ISO Method 7027
-
7/30/2019 2100Q Sales Training
3/27
3
Specifications
Measurement Method
Ratio turbidimetric determination using a primary nephelometric light scatter signal (90) to the transmitted lightscatter signal.
Light Source/ Regulatory
2100Q:Tungsten filament lamp. Meets EPA Method 180.1
2100Q IS: Light-emitting diode (LED) @ 860 nm Meets ISO 7027
Range
0 to 1000 NTU (FNU)
Accuracy
2% of reading plus stray light (< 0.02 NTU [FNU])from 0 to 1000 NTU (FNU)
Repeatability
1% of reading, or 0.01 NTU (FNU), whichever isgreater
Resolution
0.01 NTU on lowest range
Signal Averaging
Selectable on/off
Detectors
Silicon Photodiode
Reading Modes (user selectable)
Normal (Push to Read)
Signal Averaging
Rapidly Settling Turbidity
Calibration Options
Single-standard Rapid Cal for low-level measurement from 0-40NTU (FNU)
Full range calibration from 0-1000 NTU (FNU)
Calibration to degrees of turbidity
-
7/30/2019 2100Q Sales Training
4/27
4
Specifications
Calibration Logger
Records the last 25-successful calibrations
Verification Logger
Logs the last 250-successful verifications
Data Logger
500 records
Power RequirementAC 100 240 V, 50/60 Hz (with Power or USB+Power
module)
4 AA alkaline batteries
Rechargeable NiMH (for use with USB+Power module)
Interface
Optional USB
Instrument Enclosure rating
IP67 (closed lid, battery and module compartment
excluded)
Certification
CE certified
Sample Required
10 mL (0.3 oz.)
Sample Cells
Round 60 x 25 mm (2.36 x 1 in.) borosilicate glass withscrew caps
Dimensions
22.9 x 10.7 x 7.7 cm (9.0 x4.2 x 3.0 in.)
Weight530 g (1.17 lb) without batteries
620 g (1.37 lb) with four AA alkaline batteries
-
7/30/2019 2100Q Sales Training
5/27
5
Main Differentiating Features from Alternatives:
Simplified Data Transfer
Data transfer with the 2100Q is simple, flexible, anddoesnt require additional software when used with the
optional USB + Power module. All data can betransferred to the module and easily downloaded to your
computer with a USB connection, providing superiordata integrity and availability. With two different moduleoptions, you can customize connectivity and power tomeet your unique needs.
Accuracy Even for Rapidly Settling SamplesThe innovative Rapid Settling Turbidity mode provides accurate, reproducible
measurements for difficult to measure, rapidly settling sampleshigh measurementconfidence without the need for redundant readings or guessing.
Easiest Calibration and Verification
On-screen assisted calibration and verification save you time andensure accuracy. With an easy-to-follow interface, complicatedmanuals are not needed to perform routine calibrations. Single-
standard RapidCal calibration offers a simplified solution for lowlevel measurements.
-
7/30/2019 2100Q Sales Training
6/27
6
Main Differentiating Features from 2100P:
2100Q retains proven optics, range and resolution Key improvements primarily reside in software
-
7/30/2019 2100Q Sales Training
7/27
7
Easy Calibration and Verification
Menu-guided calibration process
for StabCal or customer
specific standards
Saves calibration
Verification of calibration curve Clear message forpassing/failing
Cal Insert
Standard
Read
Notes:
User may stop at any point in the calibration if testing range is less than 1000 NTU.
Programmable calibration reminder available
-
7/30/2019 2100Q Sales Training
8/27
8
Easy Calibration and Verification
Menu-guided calibration process
forStabCalor customer specificstandards
Saves calibration
Verification of calibration curve
Clear message for passing/failingVerify Read 10 NTU standard
Notes
Regular verification can be carried out any timefrom main menu.
Verify Cal Options
Customizable verification value: 0.5 to 20 NTU
(default is 10 NTU). Customizable acceptance criteria: 1% to 50%(default is 10%). Programmable Reminder
-
7/30/2019 2100Q Sales Training
9/27
9
StablCal Single Standard RapidCal Calibration
Select RapidCal Measure a single 20 NTU
standard
Save Calibration
VerifyRapidCal 20-NTU
Standard
Read
-
7/30/2019 2100Q Sales Training
10/27
10
Evaluation of Rapidly Settling Samples
The 2100Qs RST reading modeuses an exclusive algorithm thatcalculates and continuouslyupdates the turbidity of the sampleto a confidence of 95% based uponthe accumulated trend of the realtime measured values.
The result is a measurement ofturbidity that is more accurate andrepeatable than can be obtainedusing traditional smoothingtechniques-such as averaging.
What is the righttime point formeasuring?
2100Q starts taking measurements
Continuous reading shows settling of sample
First RST reading displayed
-
7/30/2019 2100Q Sales Training
11/27
11
Optional Modules: Data Transfer and Power
Instrument ships with blank module
Two Module Accessory Options:
USB + Power Module (LZV813)
Line power: 110 - 230Vac, 50/60 Hz Charges NiMH batteries Data transfer to computer without additional
software (stored in module in .xls format) Enables firmware updates
Power Only Module (LZV804)
Line power: 110 - 230Vac, 50/60 Hz
-
7/30/2019 2100Q Sales Training
12/27
12
Battery Options
1. 2100Q provided with 4 AAalkaline batteries.
2. USB + Power module allows usewith rechargeable NiMH batteries. Charges NiMH in instrument when
connected to AC power.
3. Power module allows AC powerwhen plugged into wall. Will not charge batteries
-
7/30/2019 2100Q Sales Training
13/27
13
2100Q Target Customers/Markets
Customer Market Application
Drinking Water Grab sample to compare to process turbs. Sourcewater monitoring
Environmental Monitoring (USGS,US Fish & Wildlife, etc.)
Monitoring cleanliness and pollution
Contractors/Construction Control of runoff
General Industrial Product control and consistency. Monitor corrosionin boilers
Aquariums, zoos Clarity of water. Cleanliness
Beverage Consistency of product
Waste Water Discharge control
Power Demineralization/RO. Source water monitoring.
-
7/30/2019 2100Q Sales Training
14/27
14
Competitive Comparison
All prices are in US dollars
-
7/30/2019 2100Q Sales Training
15/27
15
Shipping unitAvailable for quotes/orders November 2009
-
7/30/2019 2100Q Sales Training
16/27
16
Part Numbers - Instrument and AccessoriesAvailable for quotes/orders October 2009
Hach portable turbidimeters are supplied with four AA alkaline batteries, a carrying case with insert,StablCal primary calibration standards in 1 sealed vials (20, 100, 800 NTU), 10 NTU primary
verification standard, 6 sample cells with screw-tops, instrument manual (printed and on CD-ROM),quick start guide, silicone oil and oiling cloth.
Part # Description A Price
2100Q-01 2100Q Portable Turbidimeter (meets EPA method 180.1) $930.002100Qis-01 2100Q is Portable Turbidimeter (meets ISO 7027) $930.00
AccessoriesPart # Description Quantity
LZV813USB+Power module (includes: universal power supply, USBcable, instruction sheet)
1
LZV804Power module (includes: universal power supply, instructionsheet)
1
2971304 Battery, NiMH AA pk/4
4397500 Degassing Kit 1
4397510 Sample Filtration and Degassing Kit 1
2971210 StablCal 100mL calibration kit, 2100Q pk/ 4 bottles
2971200 StablCal 500mL calibration kit, 2100Q pk/ 4 bottles
2464105 Gelex Secondary Standard Set pk/ 4 vials
-
7/30/2019 2100Q Sales Training
17/27
17
Part Numbers - Replacement PartsAvailable for quotes/orders October 2009
Replacement Parts
Part # Description Quantity
2971205StablCal ampule calibration kit, 2100Q (10, 20, 100, 800NTU)
pk/ 4 vials
2961701 10 NTU Verification Standard 1
126936 Silicone Oil, 15 mL 12971507 Insert, molded bottom, 2100Q 1
4707600 Sample Cell Oiling Cloth 1
2434706 1" glass sample cells (10ml ) w/cap (Turb) pkg/6
2971500 Carrying case for 2100Q (includes inserts) 1
-
7/30/2019 2100Q Sales Training
18/27
18
Key Takeaways
2100Q is best forportable use
Convenient operation and great display that functions as dashboard forperformance verification and visualization of key attributes
Quick and easy assisted calibration
Data storage and easy transfer to PC
Unique reading mode forrapidly settling samples
Quick readings (6 seconds vs 15 seconds in Normal mode)
-
7/30/2019 2100Q Sales Training
19/27
19
Appendix - Turbidimeter
Nephelometry techniqueMeasurement of the amount of light, which is scattered by a sample
Scattered light at 90 angle
The more light is scattered, the higher it results in turbidity!
-
7/30/2019 2100Q Sales Training
20/27
20
Appendix - Turbidimeter
Definition
Turbidity is the interaction between light, suspended particles and the medium.
The turbidity may be determined by measuring the attenuation of a radiant flux
as it passes through the liquid orby measuring the intensity ofdiffused radiation.
Measurement is made at an angle of 90 degree ( NEPHELOMETRY)
Two methods based on different regulations :
USEPA 180.1
ISO7027
transmission
radiation
-
7/30/2019 2100Q Sales Training
21/27
21
Appendix - Turbidimeter
Definition USEPA Method 180.1
1. Most common regulatory method used in the world
2. Standard Methods for the Examination of Water and Wastewater Method 2130B is very similar
3. Following design criteria are required:
a) Primary detector for nephelometric (90) measurement30
b) Light source tungsten filament lamp with a colour temperature between 2200 and3000 K.
c) Spectral response peak for the detectorbetween 400 and 600 nm, the primarywavelengths of light
d) Measurement range 0 to 40 NTU.
e) Any sample above this range must be diluted until it is within this range.
-
7/30/2019 2100Q Sales Training
22/27
22
Appendix - Turbidimeter
Definition USEPA Method 180.1
Many factors affect light scattering:
Particle Size
Particle Shape
Refractive Index ( water, organic solvents)
Colour of Particles and/or Fluids
Particle Concentration.
Nephelometers attempt to minimize the effects of these factors
(size, shape, color of particles) while maximizing measurement accuracy!
-
7/30/2019 2100Q Sales Training
23/27
23
Appendix - Turbidimeter
Influences on turbidity
Measurement of light scattering at 90 in order to minimize effects causedby particle size!
small
particles
larger
particleseven larger
particles
-
7/30/2019 2100Q Sales Training
24/27
24
Appendix - Turbidimeter
Formazin is the most closely mimic for real world samples (0.01m...10m)
Formazin particle distribution is similar to natural water samples, size distributions isapproximate the 1/d3 relationship:
particle size decreases by a factor 10, number of particles increases by factor 1000
Benefits:
highly reflective standards efficiently scatter incident light Light absorption by the formazin polymer is minimal
Can be prepared reproducible from traceable raw materials
formazin polymer has a density similar to that of water
Formazin is the accepted primary turbidity standard (US EPA, ISO etc.)
Stable turbidity suspension (in high concentrations like 4000 NTU) that produces stablemeasurements over time
Drawbacks:
Diluted Formazin Suspensions are unstable and must be prepared freshly
Raw chemicals are harmful (carcinogenic) and allergenic
Formazin and StablCal
Formazin
-
7/30/2019 2100Q Sales Training
25/27
25
Appendix - Turbidimeter
Formazin and StablCal
Formazin is an aqueous suspension of an insoluble polymer formed by the condensation reactionbetween hydrazine sulfate and hexamethylenetetramine
-
7/30/2019 2100Q Sales Training
26/27
26
Appendix - Turbidimeter
Proper standard dilution: Stabilizing the formazin polymer based on changes that are made to the polymerand the matrix relative to a dilution.
When a stock standard is diluted, the dilution is traditionally performed using ultrapure water.
During this dilution, both the turbidity-causing polymer and any other components in the matrix are
diluted! Theoretically, only the turbidity-causing formazin polymer should be diluted.
The concentrations of all other components in the matrix should remain unchanged.
Thus, when a concentrated formazin stock solution is traditionally diluted to make a low-turbidity standardusing ultrapure dilution water, all components in the matrix are also diluted. The equilibrium then shiftsand the products are less stable.
Thus, the matrix of a formazin standard is significantly different in a low-turbidity standard versus thehighly concentrated 4000 NTU stock standard.
Formazin and StablCal
Stability research
-
7/30/2019 2100Q Sales Training
27/27
27
Appendix - Turbidimeter
Formazin and StablCal
Hachs Calibration Philosophy
Linear relationship exists between turbidity and nephelometric response up to 40 NTU.
The zero point of the calibration curve is taken as the dark current in the instrumentbefore the lamp is turned on.
This enables accurate measurement even for very low level turbidity samples.
Preparation of low level standards for calibration results in high error at the low
NTU values.
The calibration curve is linear from 0-40 NTU therefore best calibration accuracy ismaintained at lower concentrations.