2.0 Assay 24-Array Format Manual Workflow Introduction

23
Axiom 2.0 Assay 24-Array Format Manual Workflow Catalog Number 902798 Pub. No. MAN0018147 Rev. B.0 Note: For safety and biohazard guidelines, see the “Safety” appendix in the Axiom 2.0 Assay 24-Array Format Manual Workflow User Guide (Pub. No. 703335). Read the Safety Data Sheets (SDSs) and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Introduction Running the Axiom 2.0 Assay 24-Array Format Manual Workflow requires the following sets of steps: 1. Genomic DNA preparation, described in the Axiom 2.0 gDNA Sample Preparation Quick Reference (Pub. No. MAN0017720), the Axiom gDNA Sample Prep for Genome-Wide BOS 1 Array Plate Quick Reference (Pub. No. 702975), Chapter 2 in the Axiom 2.0 Assay 24-Array Format Manual Workflow User Guide (Pub No. 703335), or Chapter 2 in Section 1, in the Axiom Microbiome Assay Protocol in the Axiom Microbiome Solution User Guide (Pub. No. 703408). 2. Manual target preparation of the samples, described in this document. 3. Array processing, described in GeneTitan MC Protocol for Axiom Array Plate Processing Quick Reference (Pub. No. MAN0017718). IMPORTANT! This document contains an abbreviated set of instructions used to perform target preparation. Carefully read all the instructions in the target preparation chapter Axiom 2.0 Assay 24-Array Format Manual Workflow User Guide (Pub. No. 703335) before performing manual target preparation. Note: Array handling and processing protocols still require the use of a GeneTitan MC Instrument, as described in Chapter 5, Array Processing with the GeneTitan Multi-Channel (MC) Instrument in the Axiom 2.0 Assay 24-Array Format Manual Workflow User Guide. Assay notes This manual assay format allows the user to run the Axiom 2.0 Assay for 24 samples 4 times using 1 Axiom 2.0 Reagent Kit 4x24 Reactions (Cat. No. 902798) or 1 Axiom Microbiome Reagent Kit 4x24 Reactions (Cat. No. 902910). All samples are processed in Columns 5, 7, and 9. This manual assay uses disposable divided reservoirs with a “trough-within-a-trough” design, which maximizes the amount of liquid accessible to pipette tips when using small amounts of reagent. During the assay, dispense reagents into either the 8-channel, 4-channel, or both sides, as instructed. We recommend that you prepare your genomic DNA sample plate in a clean room. Remove seals from plates carefully and discard used seals. Do not reuse seals. Use 8-channel pipettes for all sample transfers and additions of reagents and master mixes to the samples and GeneTitan trays. Change pipette tips after each sample transfer or addition to the samples. Unless otherwise specified, all reagent modules are from the Axiom 2.0 Reagent Kit 4x24 Reactions or Axiom Microbiome Reagent Kit 4x24 Reactions. See Chapter 4 of the Axiom 2.0 Assay 24-Array Format Manual Workflow User Guide (Pub. No. 703335) for a complete list of equipment and consumables required for each stage. QUICK REFERENCE For Research Use Only. Not for use in diagnostic procedures.

Transcript of 2.0 Assay 24-Array Format Manual Workflow Introduction

Page 1: 2.0 Assay 24-Array Format Manual Workflow Introduction

Axiom™ 2.0 Assay 24-Array Format Manual WorkflowCatalog Number 902798Pub. No. MAN0018147 Rev. B.0

Note: For safety and biohazard guidelines, see the “Safety” appendix in the Axiom™ 2.0 Assay 24-Array Format Manual Workflow User

Guide (Pub. No. 703335). Read the Safety Data Sheets (SDSs) and follow the handling instructions. Wear appropriate protective eyewear,clothing, and gloves.

Introduction

Running the Axiom™ 2.0 Assay 24-Array Format Manual Workflow requires the following sets of steps:

1. Genomic DNA preparation, described in the Axiom™ 2.0 gDNA Sample Preparation Quick Reference (Pub. No. MAN0017720), theAxiom™ gDNA Sample Prep for Genome-Wide BOS 1 Array Plate Quick Reference (Pub. No. 702975), Chapter 2 in the Axiom™ 2.0Assay 24-Array Format Manual Workflow User Guide (Pub No. 703335), or Chapter 2 in Section 1, in the Axiom™ Microbiome AssayProtocol in the Axiom™ Microbiome Solution User Guide (Pub. No. 703408).

2. Manual target preparation of the samples, described in this document.

3. Array processing, described in GeneTitan™ MC Protocol for Axiom™ Array Plate Processing Quick Reference (Pub. No.MAN0017718).

IMPORTANT! This document contains an abbreviated set of instructions used to perform target preparation. Carefully read all theinstructions in the target preparation chapter Axiom

™ 2.0 Assay 24-Array Format Manual Workflow User Guide (Pub. No. 703335) before

performing manual target preparation.

Note: Array handling and processing protocols still require the use of a GeneTitan™ MC Instrument, as described in Chapter 5, Array

Processing with the GeneTitan™ Multi-Channel (MC) Instrument in the Axiom

™ 2.0 Assay 24-Array Format Manual Workflow User Guide.

Assay notes• This manual assay format allows the user to run the Axiom™ 2.0 Assay for 24 samples 4 times using 1 Axiom™ 2.0 Reagent Kit 4x24

Reactions (Cat. No. 902798) or 1 Axiom™ Microbiome Reagent Kit 4x24 Reactions (Cat. No. 902910). All samples are processed inColumns 5, 7, and 9.

• This manual assay uses disposable divided reservoirs with a “trough-within-a-trough” design, which maximizes the amount of liquidaccessible to pipette tips when using small amounts of reagent. During the assay, dispense reagents into either the 8-channel,4-channel, or both sides, as instructed.

• We recommend that you prepare your genomic DNA sample plate in a clean room.

• Remove seals from plates carefully and discard used seals. Do not reuse seals.

• Use 8-channel pipettes for all sample transfers and additions of reagents and master mixes to the samples and GeneTitan™ trays.

• Change pipette tips after each sample transfer or addition to the samples.

• Unless otherwise specified, all reagent modules are from the Axiom™ 2.0 Reagent Kit 4x24 Reactions or Axiom™ Microbiome ReagentKit 4x24 Reactions. See Chapter 4 of the Axiom™ 2.0 Assay 24-Array Format Manual Workflow User Guide (Pub. No. 703335) for acomplete list of equipment and consumables required for each stage.

QUICK REFERENCE

For Research Use Only. Not for use in diagnostic procedures.

Page 2: 2.0 Assay 24-Array Format Manual Workflow Introduction

Stage 1—DNA amplification

Prepare for Stage 1

Supplies required

Reagents from Axiom™ 2.0 Reagent Kit 4x24 Reactions (Cat. No. 902798) or Axiom™ Microbiome Reagent Kit 4x24 Reactions (Cat. No.902910), Module 1, –20°C, Part No. 901711.

Set up the instruments and prepare the reagents

1. Set an incubator/oven temperature at 37°C.

2. Set the centrifuge to room temperature.

3. Thaw and prepare reagents from Module 1 of the Axiom™ 2.0 Reagent Kit 4x24 Reactions (Cat. No. 902798) or the Axiom™

Microbiome Reagent Kit 4x24 Reactions (Cat. No. 902910).

Module Reagent and cap color Treatment

Module 1Part No. 901711

–20°C

Axiom™ 2.0 Denat Soln 10X Thaw, vortex, centrifuge, then keep at room temperature.

Axiom™ 2.0 Neutral Soln Thaw (see note below), vortex, then keep at room temperature.

Axiom™ 2.0 Amp Soln Thaw (see note below), vortex, then keep at room temperature.

Axiom™ 2.0 Amp Enzyme[1] Flick tube 3X, centrifuge, then keep in –20°C cooler until ready to use.

Axiom™ Water Thaw (see note below), vortex, then keep at room temperature.

[1] Leave at –20°C until ready to use.

Note: Allow ~1 hour for Axiom™ 2.0 Amp Soln to thaw on the benchtop at room temperature. If the solution is not thawed after 1hour, vortex briefly and return to the benchtop to complete thawing. The bottles can also be thawed in a dish with ultra-pure water.The Axiom™ 2.0 Amp Soln must be thoroughly mixed before use.

4. Thaw samples in the Sample Plate.

a. Bring the samples to room temperature on the benchtop.

b. Centrifuge the plate.

c. Leave at room temperature.

IMPORTANT!· The samples must be brought to room temperature before proceeding with denaturation.

· The sample plate for microbiome studies must have 20 μL of stool gDNA diluted to a concentration of 2.5 ng/μL and/or17.5 μL cDNA must be diluted with 2.5 μL reduced EDTA TE buffer in columns 5, 7, and 9 of the Eppendorf™ 96 DeepwellPlate, 2,000 μL. A no-template control (NTC) must be plated in well G09 and Genomic DNA Standard (Ref 103) must beplated in well H09.

· The sample plate for genotyping studies must have 20 μL of each gDNA diluted to a concentration of 5 ng/μL or 10 ng/μL,as required according to the sample type, in columns 5, 7, and 9 of an ABgene™ 96 square well storage plate, 2.2 mL. Ano template (NTC) control must be plated in well G09 and Genomic DNA Standard (Ref 103) must be plated in well H09.

5. Label the 1.7-mL microcentrifuge tube and the 15 mL conical tube as indicated in the following table.

Label Tube size Temperature Contents

D MM 1.7 mL microcentrifuge tube Leave tube at room temperature Denaturation Master Mix

Amp MM 15 mL conical tube Leave tube at room temperature Amplification Master Mix

6. Label 3 solution basins (Matrix™ Reagent Reservoir with Divider, 25 mL) as indicated in the following table.

Label Temperature Contents Reservoir side

D MM Leave basin at room temperature. Denaturation Master Mix 8-channel

N Soln Leave basin at room temperature. Axiom™ 2.0 Neutral Soln 8-channel

Amp MM Leave basin at room temperature. Amplification Master Mix 8-channel

2 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 3: 2.0 Assay 24-Array Format Manual Workflow Introduction

Prepare the Denaturation Master Mix

Prepare the Denaturation Master Mix at room temperature.

1. To the 1.7-mL microcentrifuge tube labeled “D MM”, use the following table to dilute the appropriate volume of Axiom™ 2.0 DenatSoln 10X using the Axiom™ Water.

Table 1 Denaturation Master Mix.

Reagent and cap color Per sample Master mix 24+

Axiom™ 2.0 Denat Soln 10X 2 µL 116 µL

Axiom™ Water 18 µL 1,044 µL

Total volume 20 µL 1,160 µL

2. Vortex, centrifuge briefly, then leave at room temperature.

Add the Denaturation Master Mix to the samples

Perform these steps at room temperature.

1. Briefly centrifuge the sample plate, if needed.

Samples must be at room temperature for this step.

2. Using a P1000 pipette, gently pipet or pour the Denaturation Master Mix into the 8-channel side of the reagent reservoir labeled “DMM”.

3. Carefully remove the seal from the sample plate, then discard the seal.

4. Using a P20 8-channel pipette, add 20 µL of Denaturation Master Mix to each sample in columns 5, 7, and 9.

• Pipet directly into the liquid of each well. Do not mix by pipetting up and down.

• Change tips between each addition.

• This plate is now known as the Denaturation Plate.

5. Seal, then vortex the Denaturation Plate. Start the timer for a 10-minute incubation.

6. Briefly centrifuge the Denaturation Plate in a room-temperature centrifuge by bringing the centrifuge speed to 1,000 rpm (takes~1 minute).

The centrifuge time is included in the 10-minute incubation.

7. Visually examine the volume in each well (should be 40 μL/well) and:

a. Keep a record of any wells that visually appear to have an unusually low or higher volume, as these samples might need to berepeated.

b. Do not stop to measure volumes. Proceed without delay.

8. Complete the 10-minute incubation on the benchtop at room temperature.

While completing the incubation at room temperature, prepare the Axiom™ 2.0 Neutral Soln as described in step 1 in “Add Axiom™

2.0 Neutral Soln to samples” on page 3.

9. After incubation, immediately add the Axiom™ 2.0 Neutral Soln as described in “Add Axiom™ 2.0 Neutral Soln to samples” onpage 3.

Add Axiom™ 2.0 Neutral Soln to samples

Perform these steps at room temperature.

1. Measure 3.64 mL of Axiom™ 2.0 Neutral Soln, then slowly pipet the reagent into the 8-channel side of the reagent reservoir labeled“N Soln”.

2. Carefully remove the seal from the Denaturation Plate, then discard the seal.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 3

Page 4: 2.0 Assay 24-Array Format Manual Workflow Introduction

3. Use a P200 8-channel pipette to add 130 μL of Axiom™ 2.0 Neutral Soln to each sample in columns 5, 7, and 9.

• Pipet down the wall of each well. Change tips between each addition.

• The plate is now known as the Neutralization Plate.

4. Seal, vortex, then briefly centrifuge the Neutralization Plate.

5. Visually examine the volume in each well. Total volume should be ~170 μL/well.

a. Keep a record of any wells that visually appear to have an unusually low or higher volume as these samples might need to berepeated.

b. Do not stop to measure volumes.

6. Proceed immediately to “Prepare the Amplification Master Mix” on page 4.

Prepare the Amplification Master Mix

Perform these steps at room temperature.

1. Using the following table, pipette the appropriate amount of Axiom™ 2.0 Amp Soln into the 15-mL tube labeled “Amp MM”.

Table 2 Amplification Master Mix.

Reagent and cap color Per sample Master mix 24+

Axiom™ 2.0 Amp Soln 225 µL 6.75 mL

Axiom™ 2.0 Amp Enzyme 5 µL 150 μL

Total volume 230 µL 6.90 mL

Note: The Axiom™ 2.0 Amp Soln is a viscous solution. To ensure that the reagent transfer is accurate:

· Pipet slowly.

· Allow bubbles that are generated from mixing to settle at the top before pipetting.

· Use a 10-mL serological pipette to transfer the Axiom™ 2.0 Amp Soln into the tube labeled “Amp MM”.

2. Remove the Axiom™ 2.0 Amp Enzyme from the freezer, then place in a portable cooler at –20°C.

a. Invert, then flick the Axiom™ 2.0 Amp Enzyme tube 3 times, then briefly centrifuge.

b. Using the table, add the appropriate amount of Axiom™ 2.0 Amp Enzyme to the tube labeled “Amp MM”.

c. Vortex the Amplification Master Mix well, invert the tube 2 times, then vortex again.

Add the Amplification Master Mix to the samples

Perform these steps at room temperature.

1. Slowly pour the Amplification Master Mix to the 8-channel side of the reagent reservoir labeled “Amp MM”.

2. Carefully remove the seal from the Neutralization Plate, then discard the seal.

3. Using a P1200 8-channel pipette, slowly add 230 μL of Amplification Master Mix to columns 5, 7, and 9 of the Neutralization Plate.

• Pipet down the wall of the well (total volume: 400 µL/well). Do not mix by pipetting up and down.

• Change tips between each addition.

Note: After adding the Amplification Master Mix, the plate is now known as the Amplification Plate.

4. Blot the top of the plate with a laboratory tissue. Seal tightly, vortex twice, then centrifuge the Amplification Plate for 1 minute at1,000 rpm.

5. Place the sealed Amplification Plate in an oven set at 37°C, then leave undisturbed for 23 ±1 hours.

Note: If using a GeneChip™ Hybridization Oven, place the plate on the bottom of the oven. Plates do not rotate.

6. Gather all the reagents from Module 1, tighten all caps, then mark the reagent pouches, tubes, and bottles to track use. Store at–20°C.

4 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 5: 2.0 Assay 24-Array Format Manual Workflow Introduction

Freeze the plate or proceed to Stage 2After the incubation finishes, do one of the following:

• Proceed to “Stage 2—Fragmentation and precipitation” on page 6.

• Store the Amplification Plate at –20°C.

Note: If freezing, do not perform the stop amplification reaction step described in Stage 2 before you store the sample plate at –20°C. Thestop amplification reaction step is performed after thawing the frozen plate.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 5

Page 6: 2.0 Assay 24-Array Format Manual Workflow Introduction

Stage 2—Fragmentation and precipitation

Prepare for Stage 2

Supplies required

• Selected reagents from Axiom™ 2.0 Reagent Kit 4x24 Reactions or Axiom™ Microbiome Reagent Kit 4x24 Reactions.– Module 2‑1, –20°C, Part No. 901528

– Module 2‑2, 2-8°C, Part No. 901529

• Isopropanol (supplied by user)

Instrument setup

• Prepare the following instruments for this stage before you begin the assay:– One oven at 65°C

– One oven at 37°C

– One centrifuge at room temperature

Note: If the plate of amplified DNA samples was frozen at the end of Stage 1, thaw the plate before beginning Stage 2. See instructionsin Chapter 4 of the Axiom™ 2.0 Assay 24-Array Format Manual Workflow User Guide (Put No. 703335) for notes on thawing and spinningdown prior to changing the seal to avoid cross contamination.

Note: Keep a balance plate ready to avoid delays during the fragmentation steps.

Thaw and prepare the reagents

Thaw and prepare the fragmentation and precipitation reagents from Module 2‑1 and Module 2‑2 of the Axiom™ 2.0 Reagent Kit 4x24Reactions (Cat. No. 902798) or the Axiom™ Microbiome Reagent Kit 4x24 Reactions (Cat. No. 902910)..

Module Reagent and cap color Treatment

Module 2‑1Pouch 1 of 2

Part No. 901528

–25°C to –15°C

Axiom™ Frag Enzyme Leave at –20°C until ready to use. Gently flick the tube 3 times to mix. Centrifugebefore use.

Axiom™ 10X Frag Buffer Thaw on the benchtop to room temperature. When thawed, place on ice. Vortexbefore use.

Axiom™ Precip Soln 2 Thaw on the benchtop to room temperature. Vortex and centrifuge before use.

Module 2‑2Pouch 2 of 2

Part No. 901529

2°C to 8°C

Axiom™ Frag Diluent Keep in refrigerator or place on ice.

Axiom™ Frag Rxn Stop Warm on the benchtop to room temperature. Vortex before use.

Axiom™ Precip Soln 1 Warm on the benchtop to room temperature. Vortex before use.

N/A Isopropanol Keep in room temperature.

Label tubes and reagent reservoirs

1. Label the 15-mL conical tube as indicated in the following table.

Label Temperature Contents

Frag MM Place tube on ice Fragmentation Master Mix

2. Label the 4 reagent reservoirs as indicated in the following table.

Label Size Temperature Contents

Frag MM 25 mL Room temperature Fragmentation Master Mix

Stop 25 mL Room temperature Frag Rxn Stop

Precip MM 25 mL Room temperature Precipitation Master Mix

ISO 100 mL Room temperature Isopropanol

6 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 7: 2.0 Assay 24-Array Format Manual Workflow Introduction

Incubate samples in preheated ovens

Note: (Optional) Remove samples for quantifying amplification yield by the PicoGreen™ assay at a later time. See Chapter 4, 24-array

manual target preparation, in the Axiom™ 2.0 Assay 24-Array Format Manual Workflow User Guide.

1. Stop the DNA amplification reaction.

a. Place the Amplification Plate in the 65°C oven.

• If proceeding directly from the end of Stage 1, transfer the Amplification Plate from the 37°C oven to the 65°C oven.Ensure that the seal is still securely attached to the plate to minimize evaporation.

• If working with a frozen Amplification Plate, follow the guidelines in “Thaw and prepare the reagents” on page 6 beforeplacing it in the 65°C oven.

b. Incubate for 20 minutes.

c. Prepare the fragmentation reagents as detailed in step 1 in “Thaw and prepare the reagents” on page 6 after starting the 65°Cincubation of the Amplification Plate.

2. Prepare for fragmentation.

a. Remove the Amplification Plate from the 65°C oven, then transfer the plate to the 37°C oven.

b. Incubate for 45 minutes.

Prepare the Fragmentation Master Mix

1. Start making the Fragmentation Master Mix when there are 5 minutes to the finish of the 37°C incubation.

2. Transfer the Axiom™ Frag Enzyme to a –20°C portable cooler until ready to use.

3. Use the appropriate single-channel pipettes to add the reagents to the 15-mL tube labeled “Frg MM” tube, in the order that is shownin the following table.

Table 3 Fragmentation Master Mix.

Reagent and cap color Per sample Master mix 24+

Axiom™ 10X Frag Buffer 45.7 µL 1.69 mL

Axiom™ Frag Diluent 10.3 µL 381 µL

Axiom™ Frag Enzyme 1.0 µL 37 µL

Total volume 57 μL 2.11 mL

a. Just before the end of the 45-minute 37°C incubation, flick the Axiom™ Frag Enzyme tube 2 to 3 times, then centrifuge.

b. At the end of the 45-minute 37°C incubation, add the Axiom™ Frag Enzyme to the Fragmentation Master Mix.

Note: Leave the Axiom™ Frag Enzyme at –20°C until ready to use.

4. Vortex twice, then place on ice.

5. Using a P1000 pipette, slowly transfer the Fragmentation Master Mix in the 8-channel side of the reagent reservoir labeled “FRGMM” placed at room temperature.

Add the Fragmentation Master Mix to the samples

IMPORTANT! Work quickly to perform this set of steps to minimize the time that the Fragmentation Plate is out of the 37°C oven.

1. Carefully remove the Sample Plate from the 37°C oven and place on the bench top at room temperature.

Do not place the Sample Plate on ice.

2. Add 57 μL of Fragmentation Master Mix to each sample, pipetting directly into the liquid. Do not mix by pipetting up and down.

3. Seal the plate, then vortex twice.

4. Start the timer for 30 minutes.

5. Briefly centrifuge the Sample Plate in the room temperature plate centrifuge.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 7

Page 8: 2.0 Assay 24-Array Format Manual Workflow Introduction

6. Quickly transfer the plate to a 37°C oven, then incubate for 30 minutes.

CAUTION! Be watchful for the end of the 30-minute incubation step. Fragmentation is an exact 30-minute incubation step.Longer and shorter incubation times may lead to poor performance of the assay.

Add the stop solution to the samples

Perform the following steps at room temperature.

1. A few minutes before the end of the 30-minute incubation, carefully transfer 988 μL of the Axiom™ Frag Rxn Stop solution into the8-channel side of the reagent reservoir labeled “Stop”.

Leave the stop solution reservoir at room temperature.

2. Remove the Fragmentation Plate from the oven and place on the bench top at room temperature.

3. At the end of the 30-minute fragmentation incubation period, carefully remove the seal from the Fragmentation Plate, then discardthe seal.

4. Using a P20 8-channel pipette, end the fragmentation reaction by adding 19 μL of Axiom™ Frag Rxn Stop solution to each sample.

• Do not mix by pipetting up and down.

• Pipette directly into the liquid of each well,.

• Change tips after each addition.

• Proceed immediately to the next step.

5. Seal, vortex, then centrifuge at 1,000 rpm.

6. Leave the Fragmentation Plate on the bench top at room temperature while you prepare the Precipitation Master Mix.

Prepare the Precipitation Master Mix

Perform the following steps at room temperature.

1. Prepare the Precipitation Master Mix in a 15-mL conical tube labeled “Precip MM”. Add the reagents in the order and volumesshown in the following table.

Table 4 Precipitation Master Mix.

Reagent Per sample Master mix 24+

Axiom™ Precip Soln 1 238 µL 6.19 mL

Axiom™ Precip Soln 2 2 µL 52 µL

Total volume 240 µL 6.24 mL

Note: Use a 5-mL serological pipette to pipet Axiom™ Precip Soln 1.

2. Vortex the tube labeled “Precip MM”, then place on the benchtop at room temperature.

3. Pour the Precipitation Master Mix into the 8-channel side of the reagent reservoir labeled “Precip MM”.

4. Carefully remove the seal from the Fragmentation Plate, then discard the seal.

5. Using a P1200 8-channel pipette, add 240 μL Precipitation Master Mix to each sample. Rest each pipette tip against the wall of eachwell while delivering.

• You do not need to mix up and down.

• Change tips after each addition.

• After adding the Precipitation Master Mix, the plate is now known as the Precipitation Plate.

6. Seal the Precipitation Plate, vortex, then centrifuge briefly.

8 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 9: 2.0 Assay 24-Array Format Manual Workflow Introduction

Prepare and add isopropanol to the Precipitation Plate

1. Remove the Precipitation Plate from the centrifuge and place on the benchtop at room temperature.

2. Pour 20 mL of isopropanol into both sides of the reservoir.

3. Carefully remove the seal from the Precipitation Plate, then discard the seal.

4. Use a P1200 8-channel pipette to add 600 µL isopropanol to each sample, then mix well by pipetting up and down 6 or 7 timeswithin the solution to ensure mixing.

Observe the solution while it is within the tips—it should look homogeneous after pipetting 5 to 7 times. If not, repeat mixing a fewmore times until the solution looks homogeneous.

• Do not vortex the plate after isopropanol addition to avoid cross-contamination of the samples.

• Change the tips after each addition.

5. Blot the top of the plate with laboratory tissue, then seal tightly.

6. Carefully transfer the sample plate into the –20°C freezer, then incubate overnight (16–24 hours).

7. Gather all the reagents from Module 2‑1 and Module 2‑2 and tighten all caps. Mark reagent pouches, tubes, and bottles to track use.Store Module 2‑1 at –20°C and Module 2‑2 at 4°C.

8. After incubation, proceed to “Stage 3—Centrifuge and drying, resuspension and hybridization preparation, and sample QC” onpage 10.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 9

Page 10: 2.0 Assay 24-Array Format Manual Workflow Introduction

Stage 3—Centrifuge and drying, resuspension and hybridization preparation, and sample QC

Prepare for Stage 3

CAUTION! Some of the steps in this stage should be performed under a fume hood.

Supplies required

• Selected reagents from the Axiom™ 2.0 Reagent Kit 4x24 Reactions or Axiom™ Microbiome Reagent Kit 4x24 Reactions (see Table3.1):Module 2‑1, –20°C, Part No, 901528Module 2‑2, 2–8°C, Part No. 901529

• Other reagents required for QC steps (optional):– Invitrogen™ TrackIt™ Cyan/Orange Loading Buffer (Cat. No. 10482028)

– Applied Biosystems™ 25 bp DNA Ladder (Cat. No. 931343)

– Water, Nuclease-free, Molecular Biology Grade, Ultrapure (Cat. No. 71786)

– Invitrogen™ E‑Gel™ 48 Agarose Gels, 4% (Cat. No.G800804)

Instrument setup

• Prepare the following instruments for this stage:– Oven preheated to 37°C

– Plate centrifuge set at 4°C

– Jitterbug™ or Microplate shaker

Prepare the reagents for Stage 3

1. Prepare the Gel Diluent for Sample QC (100-fold dilution of TrackIt™ Cyan/Orange Loading Buffer): Mix 49.95 mL of nuclease-freewater with 500 µL of TrackIt™ Cyan/Orange Loading Buffer.

2. Thaw and prepare the reagents from Module 2‑1 and Module 2‑2 of the Axiom™ 2.0 Reagent Kit 4x24 Reactions (Cat. No. 902798) orthe Axiom™ Microbiome Reagent Kit 4x24 Reactions (Cat. No. 902910)..

Module Reagent and cap color Treatment

Module 2‑1Pouch 1 of 2

Part No. 901528

–25°C to –15°C

Axiom™ Hyb Buffer Vortex, then keep at room temperature.

Axiom™ Hyb Soln 1 Thaw, vortex, spin, then keep at room temperature.

Module 2‑2Pouch 2 of 2

Part No. 901529

2°C to 8°C

Axiom™ Hyb Soln 2 Vortex, spin, then keep at room temperature.

Axiom™ Resusp Buffer Warm to room temperature (1 hour).

CAUTION! Some of the steps in this stage should be performed under a fume hood.

Stage 3A—Centrifuge the precipitation plate and dry the DNA pellet

Note: Keep the centrifuge ready at 4°C.

1. Transfer the Precipitation Plate from the –20°C freezer to a prechilled centrifuge.

2. Centrifuge the plate at 4°C at 3,200 x g for 40 minutes.

10 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 11: 2.0 Assay 24-Array Format Manual Workflow Introduction

3. Immediately after the 40-minute centrifugation time, empty the liquid from the plate.

CAUTION! During this step, handle the plate gently to avoid disturbing the pellets. Do not bump or bang the plate.

a. Carefully remove the seal from the Precipitation Plate, then discard the seal.

b. Invert the plate over a clean waste container and allow the liquid to drain. Collect liquid, then discard the liquid according tolocal, state, and federal regulations.

c. While still inverted, gently press the plates on a pile of laboratory tissues on a bench, then leave them to drain for 5 minutes.Transfer the plates to a new pile of tissues twice during the 5-minute time frame.

4. Turn the plate top-side up and place in an oven for 20 minutes at 37°C to dry.

5. After 20 minutes, remove the plate from the oven, even if some droplets of liquid remain, then do one of the following:

• Proceed directly to “Stage 3B—Resuspension and hybridization preparation” on page 11, even if some droplets of liquidremain. Leave the sample plate at room temperature. It is helpful to start preparing reagents for stage 3A and 3B whilecentrifuging and drying pellets.

• Tightly seal the plates and store the plates for resuspension later in the same day.– If resuspension is carried out within 4 hours, keep the plates at room temperature.

– If resuspension is carried out in more than 4 hours, store the plates in a refrigerator (2–8°C).

• Store the plates for resuspension on another day. Tightly seal the plate and store at –20°C.

Stage 3B—Resuspension and hybridization preparation

Prepare for resuspension and hybridization

• If a plate was stored at –20°C after drying the pellets, allow the plate to sit at room temperature for 1.5 hour before carrying outresuspension.

• Make sure the Axiom™ Resusp Buffer has equilibrated to room temperature before adding to the dry pellets in Step 1.

• Perform these steps at room temperature.

1. Pipet 1.4 mL of Axiom™ Resusp Buffer into the 8-channel side of a reagent reservoir. Transfer 35 µL Axiom™ Resusp Buffer to eachwell of the sample plate with a dry pellet. Avoid touching pellets with the pipette tips.

After adding resuspension buffer, the plate is known as the Resuspension Plate.

2. Seal the Resuspension Plate, then place the plate on one of the following shakers:

• Thermo Scientific™ Compact Digital Microplate Shaker: at speed 900 rpm for 10 minutes

• Jitterbug™: at speed 7 for 10 minutes

CAUTION! Perform the rest of the steps in this stage under a fume hood.

3. While the Resuspension Plate is shaking, prepare the Hybridization Master Mix in a 15 mL tube as shown in the following table.Vortex well to mix and pour contents in the 8-channel side of a reagent reservoir.

Table 5 Hybridization Master Mix.

Reagent and cap color Per sample Master mix 24+

Axiom™ Hyb Buffer 70.5 µL 2.26 mL

Axiom™ Hyb Soln 1 0.5 µL 16 µL

Axiom™ Hyb Soln 2 9 µL 288 µL

Total Volume 80 µL 2.56 mL

4. Inspect the Resuspension Plate from the bottom. If the pellets are not dissolved, repeat Step 2 on page 11. Briefly centrifuge.

5. Select a PCR plate appropriate to the type of approved thermal cycler that you will use in Stage 4, then label the plate “Hyb ReadyPlate [plate ID]”.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 11

Page 12: 2.0 Assay 24-Array Format Manual Workflow Introduction

6. Transfer the entire contents of each well in columns 5, 7, and 9 of the Resuspension Plate to the corresponding wells of the labeledHyb-Ready Plate.

7. Add 80 µL of the Hybridization Master Mix to each well in columns 5, 7, and 9 of the Hyb-Ready Plate.

8. Seal tightly, vortex, then briefly centrifuge.

9. Gather all the reagents from Module 2‑1 and Module 2‑2 and tighten all caps. Mark reagent pouches, tubes, and bottles to track use.Store Module 2‑1 at –20°C and Module 2‑2 at 4°C.

Stage 3C—Recommended: Perform quantitation and fragmentation quality control checks

Before proceeding to Stage 4, we recommend that you perform quantitation and fragmentation QC checks.

1. Prepare the Dilution QC Plate.

a. Add 33 µL nuclease-free water to columns 5, 7, and 9 of a PCR plate labeled “Dil QC”.

b. Transfer 3 µL of the hybridization-ready sample from each well of the Hyb-Ready Plate to the corresponding well of the Dil QCplate.

c. Seal, vortex, and briefly centrifuge.

2. Prepare and read the OD Plate.

a. Add 90 µL nuclease-free water to columns 5, 7, and 9 of the OD Plate (96-well UV-Star™ plate).

b. Transfer 10 µL of each Dilution QC Plate sample to the OD Plate and mix by pipetting up and down.

c. Read absorbance on a plate reader.

See Appendix B, Sample Quantitation after Resuspension of the Axiom™ 2.0 Assay 24-Array Format Manual Workflow UserGuide.

3. Prepare and run Gel QC samples.

a. Add 120 µL Gel Diluent (100-fold dilution of TrackIt™ Cyan/Orange Loading Buffer) to columns 5, 7, and 9 of the Gel QC Plate.

b. Transfer 3 µL of each Dilution QC Plate sample to the Gel QC Plate.

c. Seal, vortex, and briefly centrifuge.

d. Run the Gel:

See Appendix A, Fragmentation Quality Control Gel Protocol of the Axiom™ 2.0 Assay 24-Array Format Manual Workflow UserGuide.

Freeze or proceed to Stage 4At this point you can:

• Proceed to “Stage 4—Denaturation and hybridization” on page 13, or

• Store the hybridization-ready samples at –20°C.

12 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 13: 2.0 Assay 24-Array Format Manual Workflow Introduction

Stage 4—Denaturation and hybridization

Prepare for Stage 4

Supplies required

• Reagents from the Axiom™ 2.0 Reagent Kit 4x24 Reactions or Axiom™ Microbiome Reagent Kit 4x24 Reactions, Module 3, WashBuffer A (Part No. 901446), Wash Buffer B (Part No. 901447), Axiom™ Water (Part No. 901578).

• Axiom™ 24-array plate in a protective base.

• Hybridization tray from the Axiom™ GeneTitan™ Consumables Kit (Cat. No. 901606)

Instruments and setup

• GeneTitan™ MC Instrument

• Approved thermal cycler.– Must be programmed with the Axiom™ 2.0 Denature protocol of 95°C for 10 minutes; 48°C for 3 minutes; 48°C for hold.

– Use the heated lid option when setting up or running protocols.

• Hybridization-ready samples in plate appropriate to the thermal cycler model used.

• 96-well metal chamber pre-heated in a 48°C oven

CAUTION! Some of the steps of this stage should be performed under a fume hood.

Prepare the hybridization-ready samples stored at –20°C

1. Warm up the Hyb-Ready Plate at room temperature for 5 minutes.

It is not necessary to equilibrate the plate for a longer duration.

2. Ensure sure that the Hyb-Ready Plate is sealed well. If the plate is not sealed well:

a. Centrifuge the plate, then carefully remove the old seal.

b. If condensation is on the top of the plate, blot dry gently with a laboratory tissue.

c. Use a fresh seal to tightly reseal the plate.

3. Vortex the Hyb-Ready Plate briefly, then centrifuge at 1,000 rpm for 30 seconds.

4. Place the Hyb-Ready Plate at room temperature.

Prepare the GeneTitan™ MC Instrument and denature the hybridization-ready sample plate

1. Warm up the array plate on the bench top for a minimum of 25 minutes before setting up hybridization on the GeneTitan™ MCInstrument.

2. At the end of the array warm up time, open the pouch and scan the array plate barcode into the Batch Registration file.

3. Before you denature the hybridization ready samples:

a. Prepare the reagents from Module 3 by inverting the bottles 2 to 3 times to mix.

b. Upload the Batch Registration File.

See Appendix B, Register samples in GeneChip™ Command Console™, of the Axiom™ 2.0 Assay 24-Array Format ManualWorkflow User Guide.

c. Set up the GeneTitan™ MC Instrument.

See GeneTitan™ MC Protocol for Axiom™ Array Plate Processing Quick Reference (Pub. No. MAN0017718).

See Chapter 5, Array Processing with the GeneTitan™ MC Instrument of the Axiom™ 2.0 Assay 24-Array Format ManualWorkflow User Guide (P/N 703335).

4. Place the Hyb-Ready Plate in the thermal cycler block, secure the lid, and start the Axiom 2.0 Denature protocol.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 13

Page 14: 2.0 Assay 24-Array Format Manual Workflow Introduction

Prepare the hybridization tray and load into the GeneTitan™ MC Instrument

The following instructions assume familiarity with techniques and procedures that are associated with loading and operating theGeneTitan

™ MC Instrument. If needed, review procedures before starting.

CAUTION! Perform these steps under a fume hood.

1. After the Axiom 2.0 Denature protocol has completed, remove the Hyb-Ready Plate from the thermal cycler and place it into a96-well metal chamber that has been prewarmed in an oven at 48°C.

2. Move the metal chamber containing the denatured Hyb-Ready Plate to a fume hood.

3. Remove the seal from the Hyb-Ready Plate, then discard.

4. Remove the hybridization tray from packaging.

5. Label the hybridization tray.

IMPORTANT! It is critical that you write only on the proper location of the hybridization tray (on the edge in front of wells A1 andB1). Do not write on any other side, because the writing can interfere with sensors inside the GeneTitan™ MC Instrument and result inexperiment failure.

1

3

2

Figure 1 Label the GeneTitan™ hybridization tray.• Do not label trays on the long side of the tray.

• Notched corner of the hybridization tray should face the front.

• Label the hybridization tray in this area.

6. Place the hybridization tray under the fume hood.

7. Using a P200 8-channel pipette set at 105 µL, slowly transfer the denatured samples in columns 5, 7, and 9 from the Hyb-ReadyPlate into the corresponding wells of the hybridization tray. Dispense to the first stop to avoid creating bubbles.

• Change pipette tips after each transfer. Discard the tip even if it shows some remaining volume.

• Ensure that there are no air bubbles present in the hybridization tray. Puncture any air bubbles that you see using a clean pipettetip.

• There is no need to spread the sample on the bottom of the hybridization tray wells. Sample distribution across wells occurswhen the array plate is stacked together with the hybridization tray by the GeneTitan™ MC Instrument.

14 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 15: 2.0 Assay 24-Array Format Manual Workflow Introduction

8. Load the array plate and hybridization tray into the GeneTitan™ MC Instrument.

The hybridization tray must be loaded on the right side. The array plate must be loaded on the left side on its protective blue base.The clear plastic cover on top of the array plate should not be loaded into the GeneTitan™ MC Instrument.

1 2

Figure 2 Array plate on protective blue base and the hybridization tray properly loaded into drawer 6.

1 Array plate on protective base. 2 Hybridization tray.

IMPORTANT! After the GeneTitan™ MC Instrument has stacked the array plate and hybridization tray, the instrument extends thedrawer. Manually check the stacking by gently pressing the 6 latching points to ensure that the 2 parts are clamped properly, andcheck underneath the arrays to ensure that there are no bubbles. If bubbles are found, gently tap the plate on top to eliminate thebubbles. Do not tip/tilt the array plate/hybridization tray sandwich as you are inspecting the bottom for bubbles.

9. Follow the prompts in the software..

Hybridization continues on the GeneTitan™ MC Instrument for 23.5–24 hours before you can load the ligation, staining, andstabilization reagent trays into the GeneTitan™ MC Instrument.

10. Near the end of the 23.5–24 hour hybridization in the GeneTitan™ MC Instrument and approximately 1.5 hours from completion(22 hours after the start of hybridization), proceed to “Stage 5—GeneTitan™ reagent preparation” on page 16.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 15

Page 16: 2.0 Assay 24-Array Format Manual Workflow Introduction

Stage 5—GeneTitan™ reagent preparation

Prepare for Stage 5

Equipment required

Quantity Item

Equipment

1 GeneTitan™ MC Instrument

1 GeneTitan™ ZeroStat AntiStatic Gun

1 Microcentrifuge

1 Pipet-Aid™ Pipette Controller

1 each Pipettes• Single-channel P200

• Single-channel P1000

• Multichannel P200

As required Disposable 10-mL serological pipets (VWR™ Cat. No. 89130-898)

1 Vortexer

Consumables required

Quantity Item

As required Aluminum foil (optional)

• 1

• 5

• 6

Axiom™ GeneTitan™ Consumables Kit (Cat. No. 901606)• Scan Tray (Part No. 501006)

• Stain Tray (Part No. 501025)

• Covers for trays (Part No. 202757)

As required Pipette tips

5 Matrix™ 25-mL Sterile Reagent Reservoir with Divider (Cat. No. 8095)

4 15-mL conical tube

16 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 17: 2.0 Assay 24-Array Format Manual Workflow Introduction

Thaw and prepare the reagents

Prepare the reagents that are required for Stage 5 according to the following table.

Module Reagent and cap color Thaw, thenplace on ice

Place onice

Place at room temperature Reagent handling

Module 4-1Pouch 1 of 2

Part No. 901278–20°C

Axiom™ Ligate Buffer[1]

Place on bench top at room temp for 30minutes. Vortex twice for 30 seconds.

Examine for precipitate. If any, warm bottlewith your hands and vortex again for 30seconds.

Axiom™ Ligate Enzyme[2] Do not thaw. Keep at −20°C until readyto use.

Immediately before use: Gently flick tube3 times, then centrifuge briefly. Place in a –20°C portable cooler until use.

Axiom™ Ligate Soln 1 Vortex, then centrifuge briefly.

Axiom™ Probe Mix 1 Vortex, then centrifuge briefly.

Axiom™ Stain Buffer Vortex, then centrifuge briefly.

Axiom™ Stabilize Soln Vortex, then centrifuge briefly.

Module 4-2Pouch 2 of 2

Part No. 9012762°C to 8°C

Axiom™ Ligate Soln 2 Vortex, then centrifuge briefly. When thawed,do not place on ice.

Axiom™ Probe Mix 2[2] Gently flick tube 3 times, then centrifugebriefly.

Axiom™ Wash A[1]

for 30minutes

Vortex twice. Place on Bench for 30 minutes.

Look for precipitate. Vortex again ifnecessary.

Axiom™ Stain 1‑A[2] Gently flick tube 3 times, then centrifugebriefly.

Axiom™ Stain 1‑B[2] Gently flick tube 3 times, then centrifugebriefly.

Axiom™ Stain 2‑A[2] Gently flick tube 3 times, then centrifugebriefly.

Axiom™ Stain 2‑B[2] Gently flick tube 3 times, then centrifugebriefly.

Axiom™ Stabilize Diluent Vortex, then centrifuge briefly.

Look for precipitate. If any, warm tube toroom temperature and vortex again.

Axiom™ Water

Axiom™ Hold Buffer[2], [3] Vortex for 30 seconds. Pour in reservoir.

Estimated reagent thawing time is ~1 hour.

[1] Check for precipitate. If precipitate is present, repeat the vortex and centrifuge step.[2] These solutions are light sensitive. Keep tubes out of direct light for a prolonged period of time.[3] Axiom™ Hold Buffer for preparing the scan tray for the 2nd, 3rd, and 4th plate are provided in Module 5, Pouch 2 of 2.

Note: The presence of some precipitate in Axiom™ Ligate Buffer will not adversely impact assay performance. Follow the instructions

above to resuspend any precipitate before use

Note: Occasionally, crystals are observed in Axiom™ Wash A and Axiom

™ Stabilize Diluent upon removal from 2-8°C storage. Before using

these solutions, the crystals should be dissolved by warming the solutions to room temperature and then vortexing.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 17

Page 18: 2.0 Assay 24-Array Format Manual Workflow Introduction

Label master mix tubes and reagent reservoirs

1. Label side of each conical tube as indicated in the following table.

Tube size Label Contents

15 mL S1 Stain 1 Master Mix

15 mL S2 Stain 2 Master Mix

15 mL Stbl Stabilization Master Mix

15 mL Lig Ligation Master Mix

2. Place the 4 tubes on ice.

3. Label the five Matrix™ Reagent Reservoirs with Divider, 25-mL, as indicated in the following table.

Label Contents

S1 Stain 1 Master Mix

S2 Stain 2 Master Mix

Stbl Stabilization Master Mix

Lig Ligation Master Mix

Stop Axiom™ Hold Buffer

Prepare the Stain, Ligation, and Stabilization Master Mixes

Prepare the Stain 1 Master Mix

1. Use appropriate serological and single-channel pipettes to add reagents to the 15-mL tube labeled “S1” in the order shown in thefollowing table.

This recipe provides enough for both S1 reagent trays.

Table 6 Stain 1 Master Mix.

Reagent and cap color Per array Master mix 24+

Axiom™ Wash A 201.6 µL 5.242 mL

Axiom™ Stain Buffer 4.2 µL 109.2 µL

Axiom™ Stain 1‑A 2.1 µL 54.6 µL

Axiom™ Stain 1‑B 2.1 µL 54.6 µL

Total 210 µL (105 µL x 2) 5.46 mL

2. Gently invert the tube 10 times to mix. Do not vortex.

3. Place on ice, then protect from direct light (for example, cover with aluminum foil or ice bucket lid).

18 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 19: 2.0 Assay 24-Array Format Manual Workflow Introduction

Prepare Stain 2 Master Mix

1. Use appropriate serological and single-channel pipettes to add reagents to the 15-mL tube labeled “S2” in the order shown in thefollowing table.

Table 7 Stain 2 Master Mix.

Reagent and cap color Per array Master mix 24+

Axiom™ Wash A 100.8 µL 2.62 mL

Axiom™ Stain Buffer 2.1 µL 54.6 μL

Axiom™ Stain 2‑A 1.05 µL 27.3 µL

Axiom™ Stain 2‑B 1.05 µL 27.3 µL

Total 105 µL 2.73 mL

2. Gently invert the S2 MM tube 10 times to mix. Do not vortex.

3. Place on ice, then protect from direct light (for example, cover with aluminum foil or ice bucket lid).

Prepare Stabilization Master Mix

1. Use appropriate serological and single-channel pipettes to add reagents to the 15-mL tube labeled “Stbl” in the order shown in thefollowing table.

Table 8 Stabilization Master Mix.

Reagent and cap color Per array Master mix 24+

Axiom™ Water 93.19 µL 2.38 mL

Axiom™ Stabilize Diluent 10.50 µL 268 μL

Axiom™ Stabilize Soln 1.31 µL 33.4 µL

Total 105 µL 2.68 mL

2. Vortex the master mix at high speed for 3 seconds.

3. Place on ice.

Prepare Ligation Master Mix—part 1

The Ligation Master Mix is prepared in 2 parts.

1. Place the 15-mL conical tube marked “Lig” on ice.

2. Use appropriate serological and single-channel pipettes to add reagents to the 15-mL tube labeled “Lig” in the order shown in thefollowing table.

Table 9 Ligation Master Mix—part 1.

Reagent and cap color Per array Master mix 24+

Axiom™ Ligate Buffer 66.15 µL 1.75 mL

Axiom™ Ligate Soln 1 13.12 µL 348 μL

Axiom™ Ligate Soln 2 3.15 µL 83 µL

Total 82.42 µL 2.18 mL

3. Vortex the master mix tube at high speed for 3 seconds.

4. Place on ice.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 19

Page 20: 2.0 Assay 24-Array Format Manual Workflow Introduction

Prepare Ligation Master Mix—part 2

The Ligation Master Mix is prepared in 2 parts.

IMPORTANT! Module 5 of the Axiom™ 2.0 Reagent Kit 4x24 Reactions contains an extra tube of Axiom

™ Ligate Enzyme, which has

been provided for back-up purposes. If there is insufficient volume of the Axiom™ Ligate Enzyme for the preparation of the fourth (4th)

24-sample Ligation Master Mix, discard the tube and its contents and proceed to use the second tube of Axiom™ Ligate Enzyme. The

remaining contents of the second tube of Axiom™ Ligate Enzyme may be discarded or stored in the freezer for future use.

1. Remove the Axiom™ Ligate Enzyme from the –20°C freezer, then place in a cooler chilled to –20°C.

2. Use appropriate serological and single-channel pipettes to add reagents to the 15-mL tube labeled “Lig” in the order shown in thefollowing table.

Note: Gently flick the Axiom™ Ligate Enzyme tube 2—3 times, then centrifuge briefly immediately before adding the enzyme to themaster mix.

Table 10 Ligation Master Mix—part 2.

Reagent Per array Master mix 24+

Ligation Master Mix from Stage 1 82.42 µL 2.18 mL

Axiom™ Probe Mix 1 10.5 µL 278 μL

Axiom™ Probe Mix 2 10.5 µL 2.78 μL

Axiom™ Ligate Enzyme 1.58 µL 42 µL

Total 105 µL 2.78 mL

3. Gently invert the master mix tube 10 times to mix. Do not vortex.

4. Place on ice, then protect from direct light (for example, cover with aluminum foil or ice bucket lid).

Aliquot master mixes and Axiom™ Hold Buffer into trays

Note: It is not necessary to change pipette tips between additions of the same reagents to stain trays and scan trays.

Prepare trays and covers

1. Label 2 stain trays “S1” (for Stain 1 Master Mix)

2. Label the remaining stain trays:

• “S2” (for Stain 2 Master Mix)

• “Stbl” (for Stabilization Master Mix)

• “Lig” (for Ligation Master Mix)

3. Destatic the inside of each tray and cover.

See Deionization of GeneTitan™ trays and covers in Appendix A of the Axiom™ 2.0 Assay 24-Array Format Manual Workflow UserGuide for the recommended technique.

20 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 21: 2.0 Assay 24-Array Format Manual Workflow Introduction

FOR RESEARCH

USE ONLY

1

32

Figure 3 Stain tray with cover.

1 Stain tray.

2 Notched corner.3 Label the stain tray here.

IMPORTANT! It is critical that you write only on the proper location of the proper edge of the stain tray, as shown in the figure. Do NOTwrite on any other side, because this can interfere with sensors inside of the GeneTitan

™ MC Instrument and result in experiment failure.

About aliquoting reagents to GeneTitan™ trays

IMPORTANT! Always aliquot reagents to the bottom of the reagent tray. Avoid touching the sides or the top of the wells with the pipettetips. Droplets close to or on the top of the well dividers can cause the cover to stick to the tray during GeneTitan™ MC Instrumentprocessing.

For all trays, pipet into trays on the benchtop. If the trays are not being used immediately, protect them from light by covering with foil orplacing in a cabinet.

IMPORTANT! Remember to deionize the stain trays and the covers before aliquoting master-mixes.

When aliquoting ligation, staining, and stabilization reagents to the trays, it is not necessary to spread the reagent to each corner ofthe well. The reagent spreads evenly when the array plate is inserted into the reagent tray during processing with the GeneTitan™ MCInstrument.

Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference 21

Page 22: 2.0 Assay 24-Array Format Manual Workflow Introduction

Aliquot reagents to stain trays

You will need to add the appropriate master mix into columns 5, 7, and 9 of the “S1”, “S2”, “Stbl”, and “Lig” trays labeled in the previousstep.

1. Pipet or pour the master mix into the 4- or 8-channel side of the reagent reservoir:

• Stain 1 Master Mix into the 8-channel side of the “S1” reagent reservoir.

• Stain 2 Master Mix into the 4-channel side of the “S2” reagent reservoir.

• Stabilization Master Mix into the 4-channel side of the “Stbl” reagent reservoir.

• Ligation Master Mix into the 4-channel side of the “Lig” reagent reservoir.

2. Add 105 µL per well of each master mix into the appropriate stain tray. Dispense to the first stop only to avoid creating bubbles.

3. If:

• Bubbles are present, puncture them with a pipette tip.

• Droplets of liquid splashed onto the well dividers, place a laboratory wipe on top of the tray to blot and remove.

4. Place covers on the trays. Orient cover correctly on the tray with the notched corners together.

5. Protect the trays from light if not immediately loading onto the GeneTitan™ MC Instrument.

Aliquot the Axiom™ Hold Buffer to the scan tray

The scan tray is shipped with 2 covers, a bottom protective base and a top cover.

32

1

Figure 4 Scan tray with top cover and black protective base.

1 Black protective base (remove before loading).

2 Scan tray.3 Top cover.

The top cover is removed to fill the tray during the target prep process, while the scan tray is left on the protective base during this part ofthe process.

Note: Module 5 (Part No. 902797) has 3 Axiom™ Hold Buffer bottles that should be used to prepare the scan tray for the second, third,

and fourth plate.

1. Pour all the contents of the Axiom™ Hold Buffer into both sides of the 25 mL divided reagent reservoir, placed on the bench top atroom temperature.

2. Remove the scan tray from its pouch.

3. Remove the scan tray cover, but leave the scan tray on its protective black base.

CAUTION! Do not remove the scan tray from its protective black base until loading onto the GeneTitan™ MC Instrument. Toavoid scratching, do not touch the bottom of the tray with pipette tips.

4. Deionize the barcoded scan tray cover (Part No. 202757) that came with the scan tray.

22 Axiom™ 2.0 Assay 24-Array Format Manual Workflow Quick Reference

Page 23: 2.0 Assay 24-Array Format Manual Workflow Introduction

5. Use a 12-channel P200 pipette with new tips to aliquot 150 µL of Axiom™ Hold Buffer to each well of the 96-plate scan tray.Dispense to the first stop and avoid touching the bottom of the tray.

You do not need to change pipette tips between additions of the Axiom™ Hold Buffer.

IMPORTANT! The scan tray has an open-bottom design, so it is very important that all 96 wells of the scan tray receive 150 µL ofAxiom™ Hold Buffer.

6. If droplets of liquid splashed onto the well dividers, place a laboratory tissue on top of the tray to blot and remove.

7. Cover the tray by orienting the notched corner of the scan tray cover over the notched edge of the tray and the flat side of the coveragainst the scan tray, then leave on the bench top.

8. Continue to the array processing ligate, wash, stain, and scan section for instructions on loading the reagent trays and scan tray.

Store remaining reagentsGather all the reagents from Module 4‑1 and Module 4‑2, then tighten all caps. Mark reagent pouches, tubes, and bottles to track use.Store Module 4‑1 at –20°C and Module 4‑2 at 4°C.

Limited product warrantyLife Technologies Corporation and/or its affiliate(s) warrant their products as set forth in the Life Technologies' General Terms andConditions of Sale at www.thermofisher.com/us/en/home/global/terms-and-conditions.html. If you have any questions, pleasecontact Life Technologies at www.thermofisher.com/support.

Manufacturer: Thermo Fisher Scientific Baltics UAB |V.A. Graiciuno 8, LT-02241 |Vilnius, Lithuania

Products:Axiom™ 2.0 Reagent Kit 4x24 Reactions

Manufacturer: Affymetrix Pte Ltd |7 Gul Circle #2M-01 |Keppel Logistics Building |Singapore 629563

Products:Axiom™ 24F Array PlatesAxiom™ Microbiome Array PlatesAxiom™ myDesign™ Array Plates

For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.

The information in this guide is subject to change without notice.

DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT,PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.

Revision history: Pub. No. MAN0018147

Revision Date Description

B.0 8 March 2021 Added the following product options:• Bio-Rad™ HSP9631, HSP9601, HSS9601, and HSS9641 96-well PCR plates.

• Applied Biosystems™ 25 bp DNA Ladder.

• Applied Biosystems™ ProFlex™ Thermal Cycler.

• Sorvall™ Legend™ XTR Centrifuge.

• Thermo Scientific™ Digital Microplate Shaker.

A.0 30 November 2018 Initial release in Thermo Fisher Scientific document control system.Supersedes legacy Affymetrix publication number 703337.Updated to the current document template, with associated updates to trademarks, logos, licensing, and warranty.

Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these products, you accept the terms and conditions of allapplicable Limited Use Label Licenses.

©2021 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. ABgene is atrademark of Thomas Scientific. Allegra is a trademark of Beckman Coulter, Inc. BINDER is a trademark of GINDER GmbH. Bio-Rad, Microseal, DNA Engine Tetrad, DNA Engine, andHard-Shell are registered trademarks of Bio-Rad Laboratories, Inc. Diversified Biotech is a trademark of Diversified Biotech. Eppendorf is a registered trademark of Eppendorf AG.Jitterbug is a trademark of Boekel Scientific. Microsoft and Excel are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries.Molecular Devices and SpectraMax are trademarks of Molecular Devices Corporation. Sigma-Aldrich is a registered trademark of Sigma-Aldrick, Inc. VWR and Pipet-Aid are trademarksof VWR International, LLC.

thermofisher.com/support | thermofisher.com/askaquestion

thermofisher.com

8 March 2021