1708230004_Q4-PRESSURE_IMMC_1

download 1708230004_Q4-PRESSURE_IMMC_1

of 1

Transcript of 1708230004_Q4-PRESSURE_IMMC_1

  • 7/22/2019 1708230004_Q4-PRESSURE_IMMC_1

    1/1

    8 October, 2012 Emerson Process Management(India) Private Limited

    Plot A-145/4, TTC Industrial Area

    MIDC Pawane

    Calibration Data Sheet Consistent with ISO 10474 3.1B or EN 10204 3.1

    Navi Mumbai 400 705. India

    Tel : (91) 22 66736000, 66736001

    FAX: (91) 22 66736005

    Customer Information Manufacturer Information

    Name: Sales Order:

    PO: Line:

    Device Information Calibration Information

    Device Type: Pressure Transmitter Factory: INDIATag No: Station Name:

    Serial No: Operator ID:

    Model No: Calibration Date:

    Module Serial No:Output:

    08230004

    PROLIFIC SYSTEMS AND TECHNOLOGIES PVT LT

    PST/JMK/12-13/060A

    2051TG3A2B21AB4M5D4Q4

    11032397

    MRK-2

    1038

    10/8/2012 12:12:04PM

    18383723

    1

    Linear

    Calibration Data

    % of Range Analog Output (mA) Pass/FailApplied Pressure Requested Applied Pressure % Span Error

    Range: 0.0000 TO 20.0000 bar

    2000.000 kPa 20.0000 bar 20.0044 0.0275 PASS100.000

    1600.000 kPa 16.0000 bar 16.8038 0.0238 PASS80.000

    1200.000 kPa 12.0000 bar 13.6026 0.0163 PASS60.000

    800.000 kPa 8.0000 bar 10.4018 0.0113

    PASS40.000

    400.000 kPa 4.0000 bar 7.1972 -0.0175 PASS20.000

    0.000 kPa 0.0000 bar 3.9998 -0.0012 PASS0.000

    This is to certify that the listed product meets the applicable Rosemount Specifications.

    Measuring and test equipment used in the manufacture and inspection of the listed product are

    traceable to the National Institute of Standards and Technology. The calibration system was

    designed to meet the intent of ANSI Z540-1-1994.

    HEMANT SHIRSAT

    ROSEMOUNT REPRESENTATIVE

    1708230004_Q4-PRESSURE_IMMC_1Page 1 of 1