12706A Test Fixture OPERATION MANUAL

32
© 2009 ADC CORPORATION First printing March 31, 2009 All rights reserved. Printed in Japan 12706A Test Fixture OPERATION MANUAL Manual Number FOE-00000048A00

Transcript of 12706A Test Fixture OPERATION MANUAL

Page 1: 12706A Test Fixture OPERATION MANUAL

© 2009 ADC CORPORATION First printing March 31, 2009 All rights reserved. Printed in Japan

12706A

Test Fixture

OPERATION MANUAL

Manual Number FOE-00000048A00

Page 2: 12706A Test Fixture OPERATION MANUAL
Page 3: 12706A Test Fixture OPERATION MANUAL

C-1*

12706A Test Fixture Operation Manual

Table of Contents

1. Introduction .................................................................................................. 1-1

1.1 Product Overview ......................................................................................................................... 1-1

1.2 Before Using the Product ............................................................................................................. 1-2

1.2.1 Appearance Check .............................................................................................................. 1-2

1.2.2 Operating Environment ...................................................................................................... 1-2

1.2.3 Parts with Limited Service Life .......................................................................................... 1-3

2. Part Descriptions .......................................................................................... 2-1

3. Methods of Use ............................................................................................ 3-1

3.1 Attaching a Sample ....................................................................................................................... 3-1

3.2 Connecting to the Measuring Instruments .................................................................................... 3-2

3.2.1 Preparation before Connection ........................................................................................... 3-2

3.2.2 Connecting with the 8340A ................................................................................................ 3-3

3.2.3 Connecting with the 8240 ................................................................................................... 3-5

3.2.4 Connecting with the 8252 ................................................................................................... 3-7

3.3 Measurement ................................................................................................................................ 3-8 3.4 Sample Program ........................................................................................................................... 3-9 3.5 Replacing the Test Socket ........................................................................................................... 3-12

4. Specifications ............................................................................................... 4-1

Appendix ......................................................................................................... A1-1

A.1 Internal Connection ................................................................................................................... A1-1

A.2 12706A External View .............................................................................................................. A1-2

Page 4: 12706A Test Fixture OPERATION MANUAL
Page 5: 12706A Test Fixture OPERATION MANUAL

F-1*

12706A Test Fixture Operation Manual

List of Illustrations

No. Title Page

2-1 12706A Part Descriptions ............................................................................ 2-2 3-1 Attaching a Sample ..................................................................................... 3-1 3-2 Connecting with the 8340A ......................................................................... 3-3 3-3 Connecting with the 8240 and the 6161 ...................................................... 3-4 3-4 Replacing the Test Socket ............................................................................ 3-11

Page 6: 12706A Test Fixture OPERATION MANUAL
Page 7: 12706A Test Fixture OPERATION MANUAL

T-1*

12706A Test Fixture Operation Manual

List of Tables

No. Title Page

1-1 Parts with Limited Service Life ...................................................................... 1-3 3-1 Connectable Measuring Instruments and Necessary Cables/Accessories........ 3-2 3-2 Test Socket Standard ....................................................................................... 3-10

Page 8: 12706A Test Fixture OPERATION MANUAL
Page 9: 12706A Test Fixture OPERATION MANUAL

1-1

12706A Test Fixture Operation Manual

1.1 Product Overview

1. Introduction

1.1 Product Overview The 12706A test fixture is a sample box that connects to a measuring instrument such as 8340A, 8240 or 8252 to perform insulation resistance measurement and VSIM (voltage source current measurement) of capacitors and other electrical components.

Features:

(1) A shielded and insulated structure is adopted, allowing stable measurement of low cur-rent and high resistance.

(2) Voltage is applied by closing the cover and is shut off by opening the cover. (3) A contact signal according to opening and closing of the cover controls the measuring

instrument.

Page 10: 12706A Test Fixture OPERATION MANUAL

1-2

12706A Text Fixture Operation Manual

1.2 Before Using the Product

1.2 Before Using the Product 1.2.1 Appearance Check

When receiving the 12706A, inspect it for any damage during transportation. If there is any damage in appearance, please contact ADC Corporation or your sales represent-ative. Our address and phone number are listed at the end of this manual.

1.2.2 Operating Environment

(1) Use the 12706A in an area free from dust, corrosive gas and vibrations and away from di-rect sunlight.

(2) The ambient temperature should be -20°C to + 80°C, the relative humidity should be 80% or less.

(3) Storage

The storage temperature range is from -25°C to +80°C. When not in use for a long time, cover the 12706A with a plastic sheet or put in a card-board box, and store it in a low humidity area away from direct sunlight.

(4) Transportation

When transporting the 12706A, pack with the packaging materials used at delivery. If the packaging materials have been lost, pack the 12706A by using the following procedure:

a) Wrap the 12706A with a plastic sheet.

b) Put it in a cardboard box (with thickness of 5 mm or more) and protect it with cu-shioning materials.

c) Fix the outside with packing tape.

Page 11: 12706A Test Fixture OPERATION MANUAL

1-3

12706A Test Fixture Operation Manual

1.2 Before Using the Product

1.2.3 Parts with Limited Service Life

The 12706A use the following parts with limited service life. Check these parts by reference to the durability in the table. If any abnormality such as a con-tact failure is found, the parts need replacing. Please contact ADC Corporation or your sales representative. For how to replace the test socket, refer to Chapter 3.5 "Test Socket Replacement."

Table 1-1 Parts with Limited Service Life

Part name Durability (times) ADC stock No. Manufacturer Standard name Test socket Mechanical:

100,000 or more Electrical: 50,000 or more

JCB-AB004EX01-1 Textool (3M) 201-0080-00-0605

Micro switch Mechanical: 100,000 or more Electrical: 50,000 or more

KSP-000698 OMRON SS-5GL2

VS contact switch Mechanical: 250,000 or more Electrical: 50,000 or more

MBN-68725A ADC -

Page 12: 12706A Test Fixture OPERATION MANUAL
Page 13: 12706A Test Fixture OPERATION MANUAL

2. P

1

2

Part Desc

1. INPUT: t

ConnectsTo be con

2. V SOURC

HI: ConnLO: Conn

criptions

terminal for m

to the MEAnnected to th

CE HI/LO: t

nects to the Vnects to the c

s

Figure 2-1

measuremen

ASURE of thehe input termi

terminal for v

V SOURCE ochassis.

12706A Pa

t

e test socket inal of a mea

voltage appli

of the test soc

12706A Te

art Descriptio

[4]. asuring instru

ication

cket [4].

est Fixture

2

ons

ument.

Operation M

2. Part Des

2-1

Manual

scriptions

Page 14: 12706A Test Fixture OPERATION MANUAL

2-2

12706A Text Fixture Operation Manual

2. Part Descriptions

3. LID SIGNAL: contact output by opening and closing the cover

When the cover is closed, the terminals are short-circuited. When it is opened, the termin-als are open-circuited. Used to turn the applied voltage ON/OFF and to start measurement.

4. Test socket: socket to mount a sample

Voltage is applied to the V SOURCE and measurement is performed on the MEASURE. Sample conditions: Axial lead part

Width: 23 mm or less, diameter: 14 mm or less Lead diameter: 1.14mm or less

5. LID SIGNAL switch: switch of the LID SIGNAL [3]

6. Applied voltage switch: switch that shuts off the voltage applied to the V SOURCE ter-minal of the test socket when the cover is opened.

The applied voltage switch has mechanical durability of 250,000 times operation. How-ever, after it reaches the end of its service life, it will be always OFF regardless of open-ing and closing of the cover. If a higher resistance or lower current value than an ex-pected value is obtained in kwon measurement, the end of service life of the switch may be reached.

a. Flash symbol Indicates high voltage danger. Printed for a terminal to which voltage higher than 1 kV is input.

b. Warning symbol Printed at a location where the operation manual needs to be referred to for protecting the 12706A from damage.

CAUTION

Page 15: 12706A Test Fixture OPERATION MANUAL

3. M

3.1 AT

P

1

2

Methods

Attaching This section d

The test soURE term

Procedure:

1. Check tha) Axib) Widc) Lea

2. Attach a

of Use

a Sampledescribes the

ocket has thinal for mea

hat a sample tial lead part dth: 23 mm oad diameter:

sample as sh

e e procedure o

e V SOURCasurement. B

to be measur

or less, diam1.14 mm or

hown below:

Figur

of attaching a

CE terminal fBe careful wh

red meets the

meter: 14 mm less

re 3-1 Atta

CAUTION

12706A Te

a sample to b

for voltage hen attachin

e following c

or less

aching a Sam

est Fixture

3.1

be measured

application ng a polar sa

conditions:

mple

Operation M

Attaching a

to the test so

and the MEample.

3-1

Manual

a Sample

ocket.

EAS-

Page 16: 12706A Test Fixture OPERATION MANUAL

3-2

12706A Text Fixture Operation Manual

3.2 Connecting to the Measuring Instruments

3.2 Connecting to the Measuring Instruments This section describes the procedure of connecting the 12706A to measuring instruments. For the details of each instrument, refer to its operation manual.

3.2.1 Preparation before Connection

The following table shows connectable instruments and necessary cables.

Table 3-1 Connectable Measuring Instruments and Necessary Cables/Accessories

Model Terminal

Measuring instrument Voltage source

8340A 8240 8252 6161 6144

6243

For INPUT A01009

or A01239

A01009 A01009 - -

For V SOURCE A01240 - A01044*

+ A08531*

A01240 A01044*

+ A08531*

For LID SIGNAL A01036 - A01036 - -

A01009: TRIAX-TRIAX cable

A01239: High-voltage TRIAX-TRIAX cable

A01240: Banana-banana cable with cover

A01036: BNC-BNC cable

*: Supplied accessory

Page 17: 12706A Test Fixture OPERATION MANUAL

3-3

12706A Test Fixture Operation Manual

3.2 Connecting to the Measuring Instruments

3.2.2 Connecting with the 8340A

This section describes the procedure of connecting the 12706A with the 8340A. See Figure 3-2.

Procedure:

(1) Attach the short bar to the input section of the 8340A. (2) Connect the INPUT of the 8340A to the INPUT of the 12706A by using the TRI-

AX-TRIAX cable (A01009 or A01239). (3) Connect the V SOURCE HI of the 8340A to the V SOURCE HI of the 12706A by using

the banana-banana cable (A01240). The V SOURCE LO does not need connecting.

(4) Connect the LID SIGNAL of the 8340A to the LID SIGNAL of the 12706A by using the BNC-BNC cable (A01036-1500). (The V SOURCE of he 8340A is turned ON/OFF by closing/opening the cover of the 12706A.)

The 8340A outputs applied voltage of maximum 1000 V. Perform connection or measurement with great care. Fix the short bar of the 8340A between the GND, GUARD and LO terminals.

CAUTION

Page 18: 12706A Test Fixture OPERATION MANUAL

3-4

12706A

3.2 Co

A Text Fixtu

onnecting to t

ure Operatio

the Measurin

F

on Manual

ng Instrumen

Figure 3-2

nts

Connecting with the 83440A

Page 19: 12706A Test Fixture OPERATION MANUAL

3-5

12706A Test Fixture Operation Manual

3.2 Connecting to the Measuring Instruments

3.2.3 Connecting with the 8240

This section describes the procedure of connecting the 12706A with the 8240. See Figure 3-3.

Procedure:

(1) Connect the INPUT of the 8240 to the INPUT of the 12706A by using the TRIAX-TRIAX cable (A01009).

(2) Connect a voltage source such as the 6161 or 6144 to the V SOURCE (binding post) of the 12706A by using the banana-banana cable (A01240).

1. The output voltage of the voltage source must not exceed the maximum input voltage of the 8240. For the maximum input voltage of the 8240, refer to the 8240 Operation Ma-nual.

2. Be sure to turn ON the driving guard switch of the 8240 3. The maximum output voltage of the 6161 is 1100 V. Perform connection or

measurement with great care

CAUTION

Page 20: 12706A Test Fixture OPERATION MANUAL

3-6

12706A

3.2 Co

A Text Fixtu

onnecting to t

ure Operatio

the Measurin

Figure 3

on Manual

ng Instrumen

3-3 Connec

nts

cting with thhe 8240 and tthe 6161

Page 21: 12706A Test Fixture OPERATION MANUAL

3-7

12706A Test Fixture Operation Manual

3.2 Connecting to the Measuring Instruments

3.2.4 Connecting with the 8252

This section describes the procedure of connecting the 12706A with the 8252.

Procedure:

(1) Connect the INPUT of the 8252 to the INPUT of the 12706A by using the TRIAX-TRIAX cable (A01009).

(2) Connect the HI and LO of the 8252 (short circuit between LO and GND) to the V SOURCE of the 12706A by using the output cable (A01044) and the banana adapter (A08531).

(3) Connect the INTERLOCK of the 8252 to the LID SIGNAL of the 12706A by using the BNC-BNC cable (A01036-1500). (The voltage application of the 8252 can be controlled by closing/opening the cover of the 12706A.)

Page 22: 12706A Test Fixture OPERATION MANUAL

3-8

12706A Text Fixture Operation Manual

3.3 Measurement

3.3 Measurement Procedure:

(1) Attach a sample, referring to Section 3.1.

(2) Connect the 12706A with a measuring instrument, referring to Section 3.2.

(3) For the 8340A or 8252, set to voltage source current measurement (VSIM) or voltage source resistance measurement (VSRM). For the 8240, set to DC current measurement.

(4) Closing the cover of the 12706A will turn on the applied voltage switch, allowing mea-surement.

1. 12706A adopts a safety design. However, be careful when using high voltage as applied voltage.

2. Do not give vibrations to the 12706A or connected measuring instrument and cables during low current or high resistance measurement, otherwise measured values will vary.

CAUTION

Page 23: 12706A Test Fixture OPERATION MANUAL

3.4 STP8

1

ample PrThe followinPC9801 serie8340A, and l

1. Sample p

rogram ng shows a pes as personaeak current i

program list

program exaal computer is measured a

ample for min Figure 3

as i-t charact

12706A T

easurement -2. Voltage iteristic.

Test Fixture

3

by GPIB cois applied to

Operation M

3.4 Sample

ontrol using o the capacito

3-9

Manual

Program

the NEC or by the

Page 24: 12706A Test Fixture OPERATION MANUAL

3-10

12706A Text Fixture Operation Manual

3.4 Sample Program

2. Sample program descriptions

Line No. Description 120

130 140 150 160

170 180 190 200 210 220 230 240 250 260 270 280 290 300 310

Address 1 of the 8340A. (A variable does not start with a number. So, "R8340A" is used here.) Interface clear, remote enabled. Initialize the 8340A. Initialize the 8340A parameters. Set the 8340A parameters. "MO1"……………………….Sampling HOLD "PVS25"……………………..Source voltage 25 V ' Set the 8340A to discharge status. Make discharge time. Set the 8340A to voltage source and charge status. Make charge time. Set the 8340A to measurement status. ' Measure 30 times.

Give a measurement trigger to the 8340A. Capture data and print. Make measurement interval time.

To line 230. ' Set the 8340A to discharge status. Program end

Page 25: 12706A Test Fixture OPERATION MANUAL

3

3. Sample pprogram resu

ult

12706A TTest Fixture

3

Operation M

3.4 Sample

3-11

Manual

Program

Page 26: 12706A Test Fixture OPERATION MANUAL

3-12

12706A Text Fixture Operation Manual

3.5 Replacing the Test Socket

3.5 Replacing the Test Socket The test socket contact has durability of connecting/disconnecting 50,000 or more times. If contact failure due to abrasion occurs, the socket needs replacing. Please contact ADC Corpo-ration or your sales representative.

Replace the test socket in accordance with the following procedure:

Table 3-2 Test Socket Standard

Part name Stock No. Standard name Test socket JCE-AB004EX01-1 201-0080-000605

Textool (3M) Note: Quote the stock number of the part when ordering.

Procedure:

(1) Remove the six screws, and remove the cover. See Figure 3-4 (a).

(2) Remove the cable soldered on the test socket terminal. See Figure 3-4 (b).

(3) Remove the screws and the nuts, and replace the socket. (The screws and the nuts will be used again.) See Figure 3-4 (c).

(4) Attach the socket by the reversing above procedure.

Page 27: 12706A Test Fixture OPERATION MANUAL

F

Pan

Figure 3-4

Cabl

Panel

nel

Flat he

Replacing th

S

le

Test s

ead screw M2

12706A T

he Test Socke

oldering iron

socket

Nut

Spring wa

Test socke

2 x 12

Test Fixture

3.5 Repl

et

n

asher

et

Operation M

lacing the Te

3-13

Manual

est Socket

Page 28: 12706A Test Fixture OPERATION MANUAL
Page 29: 12706A Test Fixture OPERATION MANUAL

4-1

12706A Test Fixture Operation Manual

4. Specifications

4. Specifications

Dielectric strength INPUT signal line – V SOURCE HI 1.5 kV DC for a minute INPUT signal line – Chassis 1.5 kV DC for a minute V SOURCE HI – Chassis 1.5 kV DC for a minute

Insulation resistance Temperature: 0 to 40°C, relative humidity: 70% or less INPUT signal line – V SOURCE HI 1 × 1015 Ω or higher INPUT signal line – Chassis 1 × 1015 Ω or higher V SOURCE HI – Chassis 1 × 109 Ω or higher

Connector INPUT: triaxial connector V SOURCE HI: biding post (red) V SOURCE LO: biding post (black) LID SIGNAL: BNC connector

Socket Socket for an axial lead type part Attachable sample: 23 mm wide, 14 mm dia. 1.14 mm lead dia.

Operating environment Temperature range: -20°C to +80°C Relative humidity: 80% or less

Storage temperature -25°C to +80°C Dimensions Approx. 150 (W) x 80 (H) x 150 (D) mm Mass 1.5 kg

Page 30: 12706A Test Fixture OPERATION MANUAL
Page 31: 12706A Test Fixture OPERATION MANUAL

A1.1

Internall Connect

tion

12706A TTest Fixture Operation M

A

A1-1

Manual

Appendix

Page 32: 12706A Test Fixture OPERATION MANUAL

A1-2*

12706A

Appendix

A1.2

A Text Fixtu

x

12706A E

ure Operatio

External

on Manual

View