113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

download 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

of 16

Transcript of 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    1/16

    Fluorescence Method Development

    HandbookHolger Franz and Verena Jendreizik; Germering, Germany

    Handbook

    ey Words

    UltiMate 3000

    System

    3D Synchro,

    Excitation, and

    Emission Scans

    Variable

    Emission Filter

    Sensitivity

    Xenon Flash

    Lamp

    Eluent Quality

    Test

    Introduction

    Fluorescence spectroscopy is the most sensitive optical

    detection technique used with high-perormance liquid

    chromatography (HPLC). Fluorescence o a molecule can

    occur naturally or can be achieved by labeling the molecule

    with a uorescent tag in a derivatization reaction. A classic

    application o such a derivatization is or amino acid analysis.

    The chemical modifcation can be perormed in an automated

    way prior to the chromatographic separation (precolumn)1 or

    ater the separation (postcolumn).2

    In the ow cell o a uorescence detector (FLD), the

    active molecule is exposed to light o a defned wavelength

    originating rom a high energy light source, typically a xenon

    lamp. Ater a lietime o up to several nanoseconds (ns), the

    excited analyte relaxes and emits the energy at a less energetic,

    longer wavelength.

    A wavelength-selective FLD usually utilizes a photo-

    multiplier which is positioned at an angle o 90 to the

    light source and detects the light that is emitted rom the

    uorescing compounds.

    In contrast to UV-vis detectors, a uorescence detector

    measures a very weak light signal rather than the dierence

    between light intensities (absorbance). A related challenge

    is to separate high-intensity excitation radiation rom low-intensity emission light; a related beneft o uorescence

    detection compared to UV detection is its ability to

    discriminate an analyte rom intererences or background

    peaks.3 Generally speaking, uorescence detection has

    dierent requirements compared to the almost ubiquitous

    UV detection.

    This handbook demonstrates how to optimize

    uorescence detection using the Thermo Scientifc Dionex

    UltiMate 3000 FLD-3000 Fluorescence Detector series.

    This document guides the analyst through the dierent

    method development and optimization steps, recommends

    the most suitable procedures, and explains optical eects

    that might be observed. The handbook does not discussspecial requirements when combining uorescence detection

    with UHPLC. For additional inormation on this topic, see

    Reerences 4 and 5. A number o application notes using the

    FLD-3000RS series uorescence detector are also available.

    See Reerences 6 to 10 or details.

    Equipment

    UltiMate 3000 Binary Rapid Separation System including

    SRD-3400 Solvent Rack with our degasser channels

    HPG-3400RS Binary Pump

    WPS-3000RS Wellplate Sampler

    TCC-3000RS Thermostatted Column Compartment

    DAD-3000RS Diode Array UV-vis Detector with

    semi-micro ow cell

    FLD-3400RS Fluorescence Detector with

    Dual-PMT and micro ow cellThermo Scientifc Dionex Chromeleon Chromatography

    Data System (CDS) sotware version 7.1

    All modules were connected using 0.005 in. (0.13 mm) i.d

    Thermo Scientifc Dionex Viper fttings.

    Chromatographic Conditions

    Column: Thermo Scientifc Acclaim RSLC 120, C18

    2.2 m, 3.0 100 mm (P/N 071604)

    Standard: PAH-Mix 221, Dr. Ehrenstorer GmbH,

    P/N XS20950221AL, Lot 91124AL

    12/2011 including naphthalene,

    anthracene, uoranthene, pyrene anddibenzo(a,h)anthracene; 100 ng/L in

    CH3CN, Dilution 1:10 (standard 1) or

    1:100 (standard 2) injection volumes 1 L

    or 10 L

    Eluent A: Water, Fisher Scientifc Optima LC/MS

    (P/N W6-212)

    Eluent B: Acetonitrile, Fisher Scientifc Optima

    LC/MS (P/N A955-212)

    Gradient: Time (min) % Acetonitrile

    0.0 75

    2.0 75

    4.0 955.0 95

    5.5 75

    7.5 75

    Column Temp.: 45 C

    Fluorescence Detector Flow

    Cell Temp.: 45 C

    Flow Rate: 1 mL/min

    Inj. Volume: 1 L or 10 L

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    2/162

    Experimental

    Initial Steps and Overview

    As or all scientifc work, method development or UHPLC

    with uorescence detection starts with a literature review.

    This typically provides valuable inormation on the column,

    eluents, and gradients used by other researchers. One must

    consider i the analyte uoresces, and i so, what are the

    published excitation and emission wavelengths, and what isthe UV absorption spectrum? An absorption spectrum is a

    good starting point or selecting the excitation wavelength

    o an analyte because the spectrum indicates which energy

    is absorbed to excite an electron to a higher quantum state.

    Ultraviolet-active substances transorm electromagnetic

    radiation into heat, uorescent compounds emit light o

    longer wavelength during the relaxation process.

    In preparation or the initial experiments, make sure

    that the mobile phase is compatible with highly sensitive

    uorescence measurements. Incompatible solvents may

    create background uorescence or stray light. This reduces

    the dynamic range o the detector, directly contributes to

    baseline noise, and thereore adversely aects the limit o

    detection (LOD). Solvent suppliers typically speciy i

    solvents are suitable or uorescence spectroscopy.

    Figure 1 provides an overview o the method

    development steps discussed in this document.

    Initial Experiments

    A suitable UHPLC system may use an FLD as the only

    detector, or an FLD in combination with a diode array

    detector (DAD). This combination is not required or routi

    use, but is useul or method development. The ollowing

    sections explain how to work efciently with both system

    confgurations.

    System with DAD and FLD

    The frst step is to optimize the chromatographic separatio

    The easiest way is to use a DAD, which can be considered

    universal detector or uorophores, because the absorption

    spectrum or the molecule is typically similar to the

    excitation spectrum. Fluorophores can be detected by UV-

    detectors and at the same time the obtained spectra indicat

    how to excite them. Note that the applied standard must b

    typically 50100 times more concentrated than or the late

    uorescence detection.

    This handbook does not ocus on optimization o the

    separation but uses constant separation conditions with

    respect to column selection and gradient profle. All workis perormed using a mixture o fve polycyclic aromatic

    hydrocarbons (PAHs). The PAHs are easy to separate and

    detect at higher concentrations with a DAD, as shown

    in Figure 2. Note that the our inital peaks are eluted

    isocratically; the gradient is applied to speed the elution o

    the fth peak.

    Absorption and fluorescence spectra

    Excitation and emission wavelengths

    Chromatographic separationof compounds

    Literature Review

    29094

    Initial Experiments

    3D Emission Scan

    3D Excitation Scan

    Variable EmissionFilter

    Sensitivity Settingof the PMT

    Data Collection Rateand Response Time

    Flash LampFrequency

    Eluent Quality Test

    Obtain UV spectra with DAD

    Without DAD: Use zero order mode or

    3D synchro scan of FLD

    Use suitable scan speed and PMT

    Specific test using optimized parameters

    Smart use improves noise andlamp lifetime

    Enter peak width in Chromeleoninstrument method wizard, optimizingnoise, precision, and data storage size

    Adjust the gain of the photomultiplier

    for best S/N

    Use the auto setting or optimize manually

    Scan for optimum excitation wavelengths

    Scan for optimum emission wavelengths

    Figure 2. Separation o ve PAHs with UV detection at 240 nm, standard 1,10 L injection.

    0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 7

    Minutes

    0

    600

    ACN:

    75.0 %

    95.0

    75.01

    2

    3

    4

    5

    mAU

    Peaks:

    1. Naphthalene

    2. Anthracene

    3. Fluoranthene

    4. Pyrene5. Dibenzo-

    (a,h)anthracene

    290

    Figure 1. Overview o fuorescence method development steps discussed in

    this document.

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    3/16 3

    The DAD now provides easy access to UV spectral

    inormation or each compound. Figure 3 shows all

    fve spectra (obtained at peak apex) and determines the

    wavelength o the absorption maximum. As the absorption

    maximum is almost identical with the excitation maximum o

    the analyte, the excitation wavelength is typically chosen close

    to the absorption maximum.

    System with FLD Only

    The FLD-3000 series uorescence detectors support two

    operation modes that are recommended or initial detection

    when little is known about the nature o the analytes and their

    retention times. Note that all the ollowing FLD experiments

    were perormed with the standard photomultiplier tube

    (PMT) that can be used in the emission wavelength

    range o 220650 nm because all compounds o the test

    application all within this range. The FLD-3000 detectors

    are also available in a unique Dual-PMT confguration.

    This confguration adds a second PMT which expands the

    emission wavelength range up to 900 nm. The beneft o theDual-PMT is that the wavelength range can be expanded

    without sacrifcing sensitivity in the UV range, as commonly

    occurs with other FLD detectors (Figure 4).

    The Dual-PMT can be ordered with the detector or as an

    upgrade. I analytes are present that require detection in the

    visible or near-inrared range, use the extended wavelength

    range o the Dual-PMT.

    200 250 300 350 400 450 500

    200

    1600 218.6

    nm

    mAU

    NaphthaleneMax Abs: 219 nm

    200 250 300 350 400 450 500

    200

    1400 251.7

    nm

    mAU

    AnthraceneMax Abs: 252 nm

    200 250 300 350 400 450 500

    50

    350 235.4

    nm

    mAU

    FluorantheneMax Abs: 235 nm

    200 250 300 350 400 450 500

    50

    450 239.9

    nm

    mAU

    PyreneMax Abs: 240 nm

    200 250 300 350 400 450 500

    100

    800 297.3

    nm

    mAU

    Dibenzo(a,h)-anthraceneMax Abs: 297 nm

    29096

    Figure 3. The UV absorption spectra o the ve analytes, which also indicate the

    excitation maximum o each.

    Figure 4. The Dual-PMT option o the FLD-3000 detector is a unique solution

    or providing optimum sensitivity across all wavelengths.

    Zero Order Mode

    In zero order mode, the grating o the emission mono-

    chromator reects the entire emission spectrum o the sample

    onto the PMT, rather than only a defned wavelength. Theexcitation monochromator is set to a single wavelength, as

    usual. To detect peaks in zero order mode, the excitation

    light must have a wavelength somewhere near the excitation

    maximum o the target analytes based on literature review

    or UV-absorption inormation. For instance, PAHs can be

    excited at 220250 nm, although the optimum excitation

    wavelength may be quite dierent.

    Figure 5 compares two chromatograms obtained in zero

    order mode with (A) excitation at 220 nm and (B) excitation

    at 250 nm. The responses or the fve peaks in chromatogram

    (A) are similar, whereas chromatogram (B) is dominated by

    the size o the second peak, anthracene. This compound has

    its excitation maximum at 250 nm; however, all other peaksstill can be detected. At both 220 and 250 nm, all analytes

    can be excited and their retention times can be determined

    and optimized. Zero order mode allows simultaneous

    detection o all emitted light, independent o the emission

    spectrum.

    0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0

    Minutes

    0

    20 105

    0

    55 105

    Response

    (counts)

    29097

    B

    A

    Response

    (counts)

    Figure 5. Dierent excitation wavelengths can be used or detection in zero

    order mode, as long as the analytes can be excited at all. Depending on the

    characteristics o the analytes, the response varies.

    200 300 400 500 600 700 800 900

    Standard Wavelength RangeExtended Wavelength RangeDual PMT

    LogarithmS

    ensitivity

    26787

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    4/164

    3D Synchro Scans

    The 3D synchro scan is a tool that provides uorescence

    response as long as any combination o excitation and oset

    results in a measurable light emission. This mode is even more

    generic than the zero order mode because knowledge about

    the excitation range o the analyte is not required. Use this

    mode in the early chromatographic development stage: or

    instance, to optimize the chromatographic resolution. It is not

    a tool or discovering the most suitable excitation or emission

    wavelengths.

    A user-defned excitation wavelength range is scanned,

    whereas the emission wavelength is synchronously scanned

    with a fxed distance to the excitation wavelength. This scan

    is repeated over the entire data acquisition time, resulting

    in a DAD-like 3D feld. The oset must take a reasonable

    wavelength shit, such as 30100 nm, into account. This is

    another approach to determine the retention times o the

    target analytes. In addition, rough excitation and emission

    wavelengths pairs can be obtained; however, optimum

    excitation and emission wavelengths can only be determined

    by perorming individual excitation and emission scans.

    Figure 6. Contour plot o a synchro scan with 30 nm oset between excitation

    and emission wavelengths.

    Minutes

    ExcitationWavelength(nm)

    0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5

    250

    450

    29098

    Figure 6 shows how a uorescence 3D scan data is

    displayed in Chromeleon sotware. In a synchro scan, the

    y-axis represents the excitation wavelength. The emission

    wavelength is Em=Ex+oset. In the example shown in

    Figure 6, the oset is 30 nm. Using this plot, wavelength

    specifc chromatograms can be extracted.

    Figure 7. Synchro scan o pyrene at 30 nm oset (black), overlaid with the

    respective excitation (blue) and emission spectra (green). The synchro scan

    maximum represents the excitation wavelength at which the best response is

    obtained with the given oset.

    Figure 7 shows an example o a single synchro scan.

    The black synchro spectrum is a result o the combination

    o the excitation and the emission spectra also provided

    in Figure 7. The synchro spectrum maximum at 337 nm

    represents the most suitable excitation and emission wavelength

    combination at the given oset o 30 nm. The excitation

    wavelength o 337 nm is close to pyrenes excitation

    maximum at 333 nm. The emission wavelength o 367 nm

    (337 nm + 30 nm) provides a air light intensity.

    A shit to a shorter wavelengthor instance, to

    333 nmimproves the excitation but decreases the emissio

    response. A slight shit toward 345 nm does the opposite:excitation reduces as emission improves. Given the 30 nm

    oset, an excitation at the excitation maximum o 238 nm

    would result in almost zero response.

    For all 3D scans, the selection o a suitable scan speed is

    important. In Chromeleon sotware, the scan can be set to

    our dierent speeds: slow, medium, ast, and superast. Slow

    provides the lowest noise because it efciently combines sma

    wavelength steps with a signifcant number o data point

    repetitions at each wavelength selection. This speed is ideal

    monitoring the quality o a mobile phase, but it is typically to

    slow to be used or scanning peaks.

    386.7

    238.1

    270.5

    332.9

    336.9

    280 320 360 400 440 480 52

    Wavelength (nm)

    240200

    390.0

    375.5

    Emission Spectrum Pyrene, Ex 240 nm Excitation Spectrum Pyrene, Em 390 nm Synchro Spectrum Pyrene, Offset 30 nm

    290

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    5/16 5

    Figure 8 identifes the recommended scan speed based

    on the peak width and the scan range. I the peak width/scan

    range coordinates are within a certain scan speed range, a ew

    repeated scans per peak are guaranteed, which is enough to

    detect a peak and to assign related spectra.

    Figure 9. Overlay o dibenzo(a,h)anthracene excitation spectra obtained with

    medium (blue), ast (black), and superast scan speed (red). Medium and ast

    spectra are similar; the superast spectrum is not as precise but supports

    scanning o very narrow peaks.

    Thereore, the majority o UHPLC and HPLC peaks,

    even at sub-2 second peak widths, can be detected by the ast

    scan speed.

    Varying scan speeds will impact data quality.

    Figure 9 compares dibenzo(a,h)anthracene excitation spectra

    obtained at dierent speeds. The very similar medium

    and ast excitation spectra vary to some extent rom the

    superast spectrum. Medium and ast scan acquisitions

    dier by a wavelength step dierence o only 0.3 nm. Theirmain dierentiation is the number o ashes obtained per

    wavelength step, inuencing noise.

    For the sake o speed, the superast scan provides only

    1/10 o the wavelength steps o the ast scan. The lack o

    wavelength steps explains why the shape o the superast

    excitation spectrum in Figure 5 is similar but not very close

    to the spectra obtained at slower scan speeds. The excitation

    maxima, however, only vary by 5 nm. This dierence is

    close enough to the correct maximum, considering the

    20 nm spectral resolution o the FLD-3000 series. The

    superast scan is thereore suited or an initial screening o

    very narrow peaks obtained in ultraast separations. I very

    narrow peaks require superast scans, it is possible to repeat

    the experiment at ast scan speed and a reduced wavelength

    range o 4060 nm around the maximum.

    In general, the ast scan speed is a good deault setting. It

    is suitable to scan typical peak widths o both conventional

    HPLC and UHPLC separations. By selecting a smaller

    wavelength range, the scan rate and thereore the support o

    narrower analyte bands can be urther improved.

    Optimizing the Emission Wavelength

    Once the chromatographic method development is per-

    ormed with the help o a UV detector, the zero order mode,

    or the 3D synchro scans o the FLD-3000 detector series,

    the next step is to optimize the detection parameters o the

    uorescence detector to achieve the best signal-to-noise (S/N)

    ratio. The frst tool or this optimization process is the 3D

    emission scan eature.

    The FLD-3000 Series supports the acquisition o

    permanent 3D emission scans. During the scan, the excitation

    remains constant at a defned wavelength. Across the entire

    acquisition time, the emission monochromator repeatedly

    scans a user-defned emission wavelength range. Similar to

    the synchro 3D scan, this results in a DAD-like 3D feld.

    However, the synchro scan is a more generic approach

    because an excitation wavelength does not have to be defned.

    Instead, only a resonable Stokes shit is entered (also see: 3DSynchro Scans).

    Figure 8. The 3D scan speed can be selected based on expected peak width and

    scan range. Even UHPLC peaks are easily supported. Note that the y-axis uses a

    logarithmic scale.

    40 100 160 220 280 340 400 460 520 580

    Scan Range (nm)

    640

    29100

    7000

    Stop-Flow Scan Recommended

    Fluorescence 3D Scan Speeds

    Slow ScanRecommended for Solvent Monitoring

    Medium Scan

    Fast Scan

    Super-Fast Scan

    1000

    100

    10

    1

    2

    5

    0.5

    PeakWidth(s)

    50

    20

    500

    200

    200 300 400

    Wavelength (nm)

    0

    55Medium

    Fast

    Super-Fast

    Rel.Intensity(%)

    29101

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    6/166

    Figure 10 shows emission scans extracted rom a single

    separation. The data were obtained by setting the emission

    scan speed to ast and the constant excitation wavelength to

    219 nm, because the naphthalene UV spectrum in

    Figure 3 suggested that 219 nm is a good excitation

    wavelength or this compound. However, it is likely that

    exciting at 219 nm is not ideal or the other analytes. When

    compounds show quite dierent excitation spectra, it is

    questionable whether the excitation wavelength chosen has

    an impact on the resulting emission spectra.

    Figure 11. Dibenzo(a,h)anthracene emission spectra extracted rom our

    dierent 3D emission scans. Each spectrum was obtained with a dierent

    excitation wavelength. The relative emission spectra are not infuenced by th

    excitation wavelength selection, except or the highlighted areas which are a

    consequence o second order diraction light.

    360 400 450 500 550 60

    Wavelength (nm)

    0

    60391.6

    412.5

    Dibenzo(a,h)anthraceneEx: 324 nm

    0

    55 391.5

    412.5

    Dibenzo(a,h)anthracene

    Ex: 235 nm

    0

    55 391.9

    412.7

    573.5

    Ex: 287 nm

    0

    55

    Rel.Intens.

    (%)

    391.8412.7

    553.4

    Dibenzo(a,h)anthraceneEx: 277 nm

    Rel.Intens.

    (%)

    Rel.Intens.

    (%)

    Rel.Intens.

    (%)

    Dibenzo(a,h)anthracene

    29

    Figure 11 shows our emission spectra o dibenzo(a,h)-

    anthracene obtained in our dierent 3D emission scans. The

    excitation wavelength was alternated between 235 nm,

    277 nm, 287 nm, and 324 nm. All spectra look qualitatively

    the same, except or the peaks indicated by the blue rectangle

    These peaks are a result o second order diraction light o

    the applied excitation wavelengths. Light diraction at a

    grating occurs as a frst, second, or higher order process. Thes

    diraction orders requently overlap. Second order diraction

    light, or example, is always visible at twice the excitation

    wavelength; so when exciting at 287 nm, second order light isvisible at 574 nm.

    In an optical detector, higher order diraction light

    is typically suppressed by optical long-pass flters. The

    FLD-3400RS uorescence detector eatures a unique variab

    emission flter that automatically selects a suitable emission

    cut-o flter element.

    Figure 10. Emission spectra obtained rom a single chromatographic run with

    3D emission acquisition. The optimum emission wavelength or each compoundis at the absolute maximum o the emission spectrum.

    250 300 400 500

    Wavelength (nm)

    0

    55 327.9

    441.3

    Naphthalene

    500250 300 400

    0

    55 399.8380.0

    423.2

    Anthracene

    250 300 400 500

    0

    55 443.2Fluoranthene

    250 300 400 500

    0

    55

    Rel.Intensity(%)

    390.0376.0

    Pyrene

    250 300 400 500

    0

    55 396.0

    417.1

    438.5

    Dibenzo(a,h)anthracene

    Ex Wvl: 219 nm

    Em Wvl Range: 250500 nm

    Scan Speed: Fast

    Scan Sensitivity:1

    Emission Filter: Open

    Wavelength (nm)

    Wavelength (nm)Wavelength (nm)

    Wavelength (nm)

    Rel.Intensity(%)

    Rel.Intensity(%)

    Rel.Intensity(%)

    Rel.Intensity(%)

    29102

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    7/16 7

    The ideal flter cut-o wavelength is between the

    excitation and the emission wavelengths. In the case o

    Figure 11, the flter was set to 280 nm or all runs to provoke

    the occurence o second order diraction light. For excitation

    at 277 nm, the second order eect is visible because the flter

    does not cut o at 280 nm sharp but has a certain transition

    range. This allows light o 277 nm to partially pass. The ideal

    emission flter depends on the excitation and the emission

    wavelength. The nominal flter wavelength should be betweenthe excitation and the emission wavelength. Thereore, or

    excitation wavelengths o 277 nm or 287 nm, the selection o

    the next higher emission flter (370 nm) would suppress higher

    order eects. The emission range must then be adapted not to

    overlap with the excitation. During 3D scans, light o higher

    diraction orders typically can be recognized by the band

    that is present during the complete separation (Figure 12).

    The FLD-3100 uorescence detector uses a fxed

    280 nm emission flter, similar to other FLD detectors. With

    this confguration, second order light rom excitation wave-

    lengths around 280 nm and higher cannot be suppressed.

    Returning to the 3D emission scan example o

    dibenzo(a,h)anthracene, Figure 13 shows an excitation

    spectrum o this molecule. The excitation maximum is

    at 293 nm. Below this wavelength, excitation decreases

    but remains present. It is thereore possible to excite

    dibenzo(a,h)anthracene across the range o 220300 nm.

    Typically, the obtained emission intensity will vary with the

    selected excitation wavelength, but the qualitative appearanceo the emission spectrum remains unchanged.

    In summary, optimum emission wavelengths can easily be

    extracted rom 3D emission scans. The FLD-3400RS detector

    is most efcient in suppressing higher order diraction light

    because it automatically selects the best suitable emission

    flter. The emission spectrum is typically not aected by

    the excitation wavelength (also known as Kashas rule).11

    Customers using the FLD-3000 detector series can thereore

    typically extract optimum emission wavelengths or all analytes

    rom a single 3D emission scan.

    Optimizing the Excitation Wavelength

    The second part o the wavelength optimization is the

    identifcation o suitable excitation wavelengths. The easiest

    way to do this is to acquire permanent 3D excitation

    scans. During the scan, the emission wavelength remainsconstant. Across the entire acquisition time, the excitation

    monochromator repeatedly scans a user-defned excitation

    wavelength range. Similar to synchro and emission 3D

    scans, this results in a DAD-like 3D feld.

    The detector logic requires constant settings o the

    scan range and the emission wavelength across the entire

    chromatogram. For a separation o quite dierent analytes,

    as in the example, there are options to run a 3D scan

    sample with the optimized emission wavelength or each

    compound, or to use a generic emission wavelength across

    all analytes (i.e., a single run or all peaks). This leads to a

    question similar to that raised or the emission scan: does

    the selection o the constant emission wavelength inuencethe shape o the observed excitation spectrum?

    Figure 12. A 3D emission scan with an excitation wavelength o 287 nm.

    Second order stray light is visible as a permanent band at 554 nm.

    Figure 13. Excitation spectrum o dibenzo(a,h)anthracene. The molecule

    can be excited best at 293 nm but also in the range down to approximately

    220 nm.

    Minutes

    nm

    1.5 2.0 2.5 3.0 3.5 4.0 4.5

    360

    600

    29104-01

    200 220 240 260 280 300 320 340 360 380 400 423

    Wavelength (nm)

    0

    55

    R

    el.Intensity(%)

    293.3

    228.7

    246.0

    Excitation low

    but present

    29105

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    8/16

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    9/16 9

    applying the auto setting o the VEF, the built-in logic o

    the detector selects the best flter or the vast majority o

    analytes and wavelength combinations. Figure 17 provides

    an example. The graph shows the relationship between flter

    setting and the S/N ratio or two analytes. The flter which

    achieves the highest S/N is labeled with the related ratio. In

    both cases, the 370 nm flter provides the best result. This

    flter also would have been selected by the auto setting.

    Minutes

    Log.MaxPMTSaturation

    Respo

    nse(

    counts)

    ClearPMTSaturation

    Log.MaxPMTSaturation

    ClearPMT-

    Saturation

    29110

    Figure 19. MaxPMTSaturationis a useul parameter to optimize the sensitivity

    o the PMT.

    Note that the 435 nm cut-o flter still allows a certain

    part o the light to pass, although the emission wavelengths

    o uoranthene and pyrene are below the cut-o flter

    wavelength (443 nm and 390 nm). The reason or this is the

    20 nm bandwidth o the emission monochromator and the

    transmission characteristics o the flter. At the specifed cut-

    o wavelength, 50% o the light can pass. As indicated in

    Figure 18, the usable emission wavelength range starts at

    15 nm below the nominal flter wavelength (i.e., or a 280

    nm flter, at 265 nm).

    80007000600050004000300020001000

    0280

    9000

    no filter

    Signal-to-

    NoiseRatio

    29108

    435370

    Fluranthene (Ex 232, Em 433) Pyrene (Ex 238, Em 390)

    2423

    8504

    Figure 17. Signal-to-noise ratio or two analytes with dierent cut-o ltersettings. The best result is obtained with the 370 nm lter, which would also be

    selected by the auto setting.

    50%

    15 nm

    Useable Em wavelength range

    220 240 260 280 300 340 360

    100%

    Wavelength (nm) 29109

    320

    Filtertransmission

    Figure 18. At the specied cut-o wavelength o the lter, 50% o the light

    is transmitted. The usable emission wavelength range starts at 15 nm below

    the nominal lter wavelength.

    In summary, the FLD-3400RS uorescence detector

    automatically picks the best suitable flter or a selected

    wavelength pair, reduces stray light, and thereore

    improves S/N due to its unique variable emission flter.

    Optimizing the Sensitivity Setting

    All commercial FLDs or use with HPLC use a PMT or

    measuring the intensity o the emitted light. It is thereore

    important to understand the capabilities and limitations o

    this device. A PMT can be regarded as a current source that

    generates a current proportional to the light intensity to

    which it is exposed.

    A PMT consists o a photocathode, several dynodes, and

    an anode. When photons hit the photocathode, electrons

    are ejected rom its surace. Between the photocathode and

    the frst dynode is an electrostatic potential dierence that

    accelerates these electrons towards the dynode. Based on

    the phenomenon o secondary emission, this dynode emits a

    larger number o electrons. A ollowing cascade o dynodes

    unctions as an electron multiplier. Depending on the voltage

    dierence between the dynodes, more or less electrons

    are emitted at each incident (i.e., the gain o the chain o

    dynodes can be adjusted by the applied voltage). Finally, a

    current pulse arrives at the anode.12

    Depending on the wavelength o the detected light and its

    intensity, it is useul to tune the voltage applied to the PMT.Too large a voltage can cause an electron overow or even

    damage the light-sensitive photocathode. Too low a voltage

    can result in increased noise and thereore in a reduced S/N

    ratio. The FLD-3000 series property or adjusting the gain o

    the PMT is called Sensitivity. At any time o data acquisition,

    this setting can be altered between 1 and 8; 1 represents the

    lowest and 8 the highest settable voltage (i.e., the sensitivity

    o the PMT to light). Typically, the best limit o detection can

    be obtained with a setting o 5 or 6.

    The intensity o the emission light (measured by the

    PMT) is normalized with the intensity o excitation light

    passing through the sample measured by the reerence diode

    (Figure 16). Thereore, the values (counts) that the emissionchannel displays cannot be used or optimizing the sensitiv-

    ity. Instead, a dierent procedure is required to tune the

    sensitivity setting or the best S/N ratio. The schematic o

    this procedure is depicted in Figure 19.

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    10/1610

    At this point o the method development, elution

    windows o peaks o interest are known. Between the elution

    o the peaks, a pair o commands can be placed to determine

    the PMT saturation, i.e. the ratio between light actually

    received and the maximum allowed light level. Use a sample

    or standard with highest expected concentration to optimize

    this setting.

    The parameterMaxPMTSaturation indicates the highest

    saturation o the PMT signal measured since the last reset.Logging this value ater the elution o a peak with the

    command log.MaxPMTSaturationwill create an entry in

    the sample audit trail. Beore the next eluting peak, reset

    the PMT saturation with the ClearMaxPMTSaturation

    command. Repeat the process ater each peak.

    The saturation o the PMT is ideal when it is between

    3080% (Table 1). I the saturation is below 30%, increase

    the sensitivity so that it reaches the optimum range. A

    one-step increase o the sensitivity increases the peak height

    by a actor o approximately 2. When the PMT is 100%

    saturated, the detector automatically reduces the sensitivity

    setting. This ensures that the PMT is not damaged; however,

    the peak which causes the PMT overow cannot be used or

    quantitative analysis.

    Note that a PMT overow at sensitivity 1 does not lead

    to a reduced sensitivity, but reports an error in the audit

    trail, and potentially leads to a non-optimal peak shape and

    probably incorrect quantifcation o the peak. Thus, or a

    successul quantifcation, reinject the sample and reduce the

    sensitivity setting as suggested by the Chromeleon sotware

    sample audit trail, or dilute the sample i sensitivity 1 is still

    too sensitive.

    Note that it is not mandatory to optimize the sensitivity;

    it is possible to operate the PMT at a lower-than-optimal

    setting.

    Figure 20 provides an example o how much the signal

    height, noise, resulting S/N, and PMT saturation change,

    depending on the sensitivity setting. When changing rom

    1 to 5, the noise increases rom 2501 to 10,759 counts,

    equating to a actor o 4 (calculated with equation 1). At th

    same time, the peak height increases by a actor o almost

    As a consequence, the S/N ratio improves by a actor o 4.

    Sensitivity 5 provides the best S/N ratio; however, it is only

    slightly better than sensitivity 4. So sensitivity 4 may be the

    better choice because it provides a larger dynamic range; it

    can handle unknowns with concentrations 2 higher than

    those o sensitivity 5.

    Equation 1:

    Table 1. Recommended actions to optimize the sensitivity setting.

    S/N = 2.Noise

    Peak Height

    10 106

    70 106

    0

    Height Noise PMTSensitivity dibenzo(a,h)anthracene (counts) S/N Saturatio

    (counts) (%)

    1 3.2 106 2501 2537 4

    2 7.0 106

    1754 7929 9

    3 15.1 106 3518 8576 19

    4 28.0 106 5749 9755 36

    5 53.8 106 10759 9998 69

    6 PMT overflow

    Minutes

    Response(

    counts)

    291

    Figure 20. Eect o the sensitivity setting on the peak height, noise, and

    S/N ratio o dibenzo(a,h)anthracene. The S/N improves by a actor o 4 rom

    sensitivity 1 to sensitivity 5. At sensitivity 6, the PMT overfows

    (standard 2, 1 L injection).

    Value Action to Optimize

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    11/16 11

    In summary, the sensitivity setting allows tuning o the

    PMT or the best S/N ratio. I dierent sensitivities provide the

    same S/N ratio, it is benefcial to use the lower setting because

    it supports higher concentrated samples. The optimization can

    easily be done with the command pair oMaxPMTSaturation

    and ClearMaxPMTSaturation. I a PMT overows in routine

    operation, the detector automatically reduces the sensitivity

    setting. Rerun the same sample with the adjusted sensitivity

    in order to quantiy the peak that previously caused thePMT overload.

    Data Collection Rate and Response Time

    Data collection rate (DCR) and response time are settings

    which inuence data storage size and S/N ratio.

    The DCR is the number o data points per second gener-

    ated by the detector electronics. For good precision during

    quantitative analysis, a peak should be defned by 2030

    data points. I the data collection rate is too low, the start

    and end points o peaks will not be determined accurately.

    I the data collection rate is too high, data fles may occupy

    excessive disk space and post-run analyses may require more

    processing time.

    The DCR is typically paired with a mathematical data

    flter called response time (or simply response or rise time).

    The response time is a measure o how quickly the detector

    responds to a change in signal. It is defned as the time it

    takes the detectors output signal to rise rom 10% to 90%

    o its fnal value. The input value is postulated to directly

    jump rom 0% to 100% (Figure 21).

    Initial experiments should be perormed at relatively

    high data collection rates and short response times. Once

    separations are optimized, the widths o relevant peaks can

    be extracted rom the chromatogram. As in Chromeleon

    sotware version 6, Chromeleon sotware version 7 provides

    guidance on the selection o data collection rates and response

    times. Calculated based on the peak width, Chromeleon

    sotware makes sure that a sufcient number o data points

    across a peak are obtained and optimizes the related response

    time. To access this guidance, open the instrument method

    (Chromeleon sotware version 7), click on the FLD symbol

    on the let, and select the rider Channel Settings. In the Data

    Collection section, the peak width can be entered (Figure 22).

    Select DCR and response times as suggested by the sotware.

    Selecting appropriate data collection settings is thereore aneasy task.

    Signal

    10%

    Time

    90%

    Response Time (or Rise Time)

    Filtered Signal

    Input Signal

    29112

    Figure 21. The response time measures the time until the detector responds to

    a change in signal. It is dened as the time it takes the detector's output signalto rise rom 10% to 90% o its nal value ollowing an instantaneous change o

    the input signal.

    29113

    Figure 22. The Chromeleon instrument method wizard calculates suitable DCR

    and response times based on the width o the narrowest peak o interest.

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    12/1612

    Effect of Changing the Xenon Flash Lamp Frequency

    The FLD-3000 detector series uses a xenon ash lamp as a

    light source. The expected lietime o this lamp is related to

    the lamp operation time and the used ash requency. Xenon

    ash lamp operation time is similar to data acquisition

    time because the lamp only requires a ew seconds to warm

    up, unlike continuous xenon and mercury lamps that need

    time to reach a thermal equilibrium and typically must be

    switched on 1 h prior to the frst injection; these lamps

    usually remain on between samples. Leaving the lamp on

    obviously reduces the number o samples analyzed during the

    lietime o a continuous lamp. A ash lamp with a lietime

    o 4000 h can thereore process ar more than 4 times the

    number o samples o a 1000 h continuous lamp.

    The FLD-3000 detector series xenon ash lamp oers

    three dierent ash requencies: HighPower (300 Hz),

    Standard (100 Hz), and LongLie (20 Hz). A higher ash

    requency improves the S/N ratio; a lower ash requency

    improves the lietime o the lamp. Operating the lamp in

    LongLie mode results in >15,000 h o chromatography.

    Note that the lietime o the lamp is defned as such that thespecifcations o the Raman S/N (an internal measurement to

    judge lamp health) will still be achieved. Thus, at the end o

    the lamp lietime, the sensitivity o the detector is expected

    to reach a limit, but the lamp may be used longer without

    any problem.

    Next, consider the inuence o the lamp mode on

    both the analyte peak and the baseline. Figure 23 shows

    an overlay o three dibenzo(a,h)anthracene peaks detected

    using HighPower, Standard, and LongLie modes. All other

    settings remained constant. The lamp mode change does

    not aect the peak height or area, but the baseline noise

    increases with a smaller ash requency.

    3.65 4.15

    0

    HighPowerStandard

    LongLife

    10 106

    Minutes

    Response

    (counts)

    29114

    17 104

    15 104

    Figure 23. Changing the fash lamp requency has an impact on the baseline

    noise, but the peak height remains constant. The lamp mode thereore only

    infuences the noise and respective S/N (standard 2, 1 L).

    Figure 24. Signal-to-noise ratio or dibenzo(a,h)anthracene obtained with

    HighPower, Standard, and LongLie lamp modes.

    Noise (F2) = Noise (F1)F2

    F1

    The related S/N ratios are shown in Figure 24. A rough

    estimate o the noise change can be calculated using

    Equation 2.

    Equation 2:

    With

    F1= Previous ash lamp requencyF2= Planned ash lamp requency

    Changing the lamp mode rom HighPower (300 Hz) to

    Standard (100 Hz) thereore increases the baseline noise by

    approximately 3 which is a actor o ~1.7. In the example

    the noise changes by a actor o 1.53. From Standard

    (100 Hz) to LongLie (20 Hz), the predicted actor is 2.24.

    The actor calculated rom real results is 1.7. In summary,

    equation is not precise, but allows to estimate the eect o

    the lamp requency change.

    9482

    29

    36326178S/N

    8000

    7000

    6000

    5000

    4000

    3000

    2000

    1000

    0

    9000

    10000

    HighPower LongLifeStandard

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    13/16 13

    Time (min) Lamp Mode Wavelengths Optimized For Excitation (nm) Emission (nm) Sensitivity Filter

    0.00 LongLie Naphthtalene 219 330 6 280

    0.75 HighPower

    1.50 Anthracene 247 400 3 370

    1.85 Fluoranthene 232 443 6 370

    2.13 Pyrene 238 39 4 370

    2.60 Dibenzo(a,h)anthracene 294 396 4 370

    4.25 LongLie

    Table 2. Timed settings across the chromatogram.

    Smart Use of the Xenon Lamp Mode

    The ash lamp o the FLD-3000 detector series is so exible

    that the lamp mode can even be changed during a chromato-

    graphic run. This eature gives the analyst a smart way to

    optimize their method.

    A chromatogram typically shows parts in which no peaks

    o interest elute. In gradient separations, this includes the time

    until the elution o the frst peak and the wash/re-equilibration

    time at the end o the run. A ash lamp could even be

    switched o during this time, but the variable lamp modes o

    the FLD-3000 detector series allow an even smarter way to

    achieve both the best S/N ratio and an increased lietime o

    the light source. During chromatographic idle times, the lamp

    mode can be switched to LongLie; during peak elution, the

    mode can be changed to Standard or HighPower.

    Figure 25 applies this approach to the separation

    used throughout this document. Total run time, including

    re-equilibration, is 7.5 min. The authors applied LongLie

    mode rom 00.75 min and rom 4.257.5 min. In total,

    the LongLie mode was active or 4 min. During the

    remaining 3.5 min o the chromatogram, the lamp modewas set to HighPower. The number o ashes decreased by

    50%, compared to i HighPower had been used across the

    entire chromatogram. Thus, this approach leads to 100%

    longer lietime o the ash lamp without any sacrifce to the

    obtained data quality. Data acquisition is always on, and

    during idle times, the baseline is monitored at reasonable

    sensitivity.

    Summary of Optimizations

    Several optimized settings were obtained or the separation

    o the PAH standard. The data collection rate and the

    response time are settings which are applied across the entire

    chromatogram:

    Data collection rate: 20 Hz

    Response time: 0.4 s

    Other settings are optimized based on the characteristics

    o a uorophore such as excitation and emission wave-length,

    sensitivity setting, and the emission flter. The lamp mode is

    also varied to apply HighPower mode during peak elution

    and LongLie mode to sections o the chromatogram with no

    peaks o interest. Table 2 summarizes these timed settings.Sensitivities were evaluated or 1 L injections o standard 2.

    Figure 25. Smart use o the lamp modes urther extends the lietime o the

    xenon fash lamp (standard 2, 1 L) injection.

    0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 7.5

    0

    HighPower,

    300 Hz

    LongL

    ife,

    20Hz

    4.5 106

    Minutes

    Respo

    nse(

    counts)

    29116

    LongLife, 20 Hz

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    14/1614

    Test of Eluent Quality

    This chapter demonstrates a way to test the eluent quality,

    particularly when switching to a new lot or even a new

    supplier.

    As mentioned at the beginning o the Experimental

    section, it is important to use solvents o compatible quality

    or uorescence detection. Noncompatible solvents may

    create background uorescence and thereore stray light.

    This stray light reduces the dynamic range o the detector,

    directly contributes to baseline noise, and thereore adversely

    aects the limit o detection (LOD).

    It is not only important to buy a solvent product that is

    nominally uorescence compatible, but it is also useul to

    make a relative comparison between dierent suppliers. In

    addition, a brie test when switching to a new batch or

    a new supplier gives confdence that the solvent quality

    is suitable.

    Results may dier with stock time o the water, so reshly

    purifed lab water may be the best choice.

    In a typical run, the signal is set to zero by an autozero

    command. The disadvantage o this procedure is that thelevel o background uorescence created by the solvent

    cannot be evaluated; or any solvent quality, the baseline

    starts at zero counts. With the FLD-3000 detector series,

    use the ClearAutoZero command to release the baseline. A

    certain level o stray light is always present, so the baseline

    will rise to a higher level. The worse the eluent quality,

    the higher the baseline level observed. Figure 26 shows an

    overlay o three experiments with water rom three dierent

    sources, obtained with a backpressure capillary instead o a

    column. A backpressure o a ew bar ensures that the pump

    logic operates seamlessly. As the stray light level is inuenced

    by the sensitivity setting, it must be kept constant during

    comparison experiments. Here, the sensitivity was set to 5throughout the run.

    At 0 min, the run starts with the excitation and emission

    wavelengths or the frst analyte, naphthalene. At 0.5 min,

    the baseline is released with the ClearAutoZero command.

    Deionized water reaches the highest background uorescence

    with ~400,000 counts under these conditions. Lab water

    and the LC/MS grade water are o similar quality at ~100,000

    counts.

    At 1 min, wavelengths are switched to the ones used or

    detecting the second analyte, anthracene, ollowed by an

    AutoZero. As a consequence, the baseline returns to zero

    counts response. At 1.5 min, the baseline is released again,

    resulting in increased baseline levels. This process is repeatedor all the wavelength/flter combinations.

    Figure 26. Three dierent water qualities are tested or their background

    fuorescence at analyte-relevant detection parameters. This quick test makes

    eluent quality monitoring an eective yet easy task.

    0.17 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.10100000

    0

    500000

    Naphthalene

    Anthracene

    Fluoranthene

    Pyrene

    Dibenzo(a,hanthracen

    Deionized tap water

    Purified lab waterLC/MS grade water

    Minutes

    Response(

    counts)

    29

    This test ensures that the quality o the solvent is not just

    checked at a single wavelength combination that might not

    even be relevant or a given analysis. All relevant parameters

    are tested quickly and can easily be compared when overlayinthem with earlier results or other eluent products.

    Note that the peak at 2 min is a consequence o the

    wavelength switch to uoranthene detection conditions. Fo

    a short time, the background uorescence already reaches

    the levels that are permanently shown between 2.5 and

    3 min. The AutoZero just ater the wavelength switch mak

    the baseline return to zero counts. The consequence is a

    peak-like signal.

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    15/16 15

    Conclusion

    UltiMate3000FLD-3100andFLD-3400RSuorescencedetectors are designed or highest perormance and methoddevelopment exibility.

    Zeroordermodeand3Dsynchroscansareeffectivetools or initial experiments without the need o a diodearray detector.

    3Demissionand3Dexcitationscansreducetheeffortto determine the most suitable wavelength pair toa minimum.

    Thevariableemissionlterandthesensitivitysettingo the photomultiplier can signifcantly improve theS/N ratio (FLD-3400RS only).

    Chromeleonsoftwarecalculatesoptimaldatacollectionrates and response times based on the narrowest peakwidth or the best S/N ratio and data storage size.

    Thisstudydemonstratestheeffectofdifferentxenonlampash requencies on noise and shows how tocombine the modes within a chromatogram or best limito detection and urther-increased lamp lietime.

    Asimpleyetapplication-tailoredtestforevaluatingtheeluent quality is explained.

    References1. Technical Note 107: Automated In-Needle Derivatizat ion Applying a User-

    Defned Program or the Thermo Scientifc Dionex WPS-3000 Split-LoopAutosampler, LPN2849, 2011. Thermo Scientifc: Documents, TechnicalNotes. www.dionex.com/ (accessed Dec 5, 2011).

    2. Pickering, M.V. Ion-Exchange Chromatography o Free Amino Acids,LC-GC19897(6), 484490.

    3. Snyder, L.R.; Kirkland, J.J.; Glajch, J.L. Practical HPLC MethodDevelopment, 2nd ed.; John Wiley & Sons, Inc.: New York, 1997; p 82.

    4. Technical Note 92: Combining Fluorescence Detection with UHPLC:An Overview o the Technical Requirements, LPN2716-01, 2011.Thermo Scientifc: Documents, Technical Notes. www.dionex.com/(accessed Dec 5, 2011).

    6. Application Brie 115: Determination oN-Methylcarbamates, LPN2587,2010. Thermo Scientifc: Documents, Applications. www.dionex.com/(accessed Dec 5, 2011).

    7. Application Update 177: Faster and More Sensitive Determination oN-Methylcarbamates in Drinking Water by HPLC, LPN2675, 2010.Thermo Scientifc: Documents, Applications.www.dionex.com/. (accessed Dec 5, 2011).

    8. Application Note 266: Determination o Sialic Acids Using UHPLC withFluorescence Detection, LPN2662, 2011. Thermo Scientifc: Documents,Applications. www.dionex.com/ (accessed Dec 5, 2011).

    9. Application Note 272: Faster Yet Sensitive Determination oN-Methylcarbamates in Rice, Potato, and Corn by HPLC, LPN2728, 2011.

    Thermo Scientifc: Documents, Applications. www.dionex.com/(accessed Dec 5, 2011).

    10. Application Note 278: UHPLC Determinat ion o Sialic Acids withFluorescence Detection, LPN2817, 2011. Thermo Scientifc: Documents,Applications. www.dionex.com/(accessed Dec 6, 2011).

    11. Kasha, M. Characterization o Electronic Transitions in Complex MoleculesDisc Faraday Soc.1950,9, 1419.

    12. Lakowicz, J.R. Principles of Fluorescence Spectroscopy, 3rd ed.; Springer

    Science+Business Media, LLC: New York, 2006; p. 44.

  • 7/27/2019 113591 Handbk FLD 3000 Method Develop 22May2012 LPN2988

    16/16

    Thermo Scientifc Dionex produ

    designed, developed, and manu

    under an ISO 9001 Quality Syste

    www.thermoscientific.com

    All trademarks are the property o Thermo Fisher Scientifc Inc. and its subsidiaries. This inormation is presented as an example o the capabiliti es o ThermoFisher Scientifc Inc. products. It is not intended to encourage use o these products in any manners that might inringe the intellectual property rights o others.Specifcations, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative or details.

    FLDHandbook_E 02/12 S

    In addition to these

    ofces, Thermo Fish

    Scientifc maintains

    a network o represe

    tative organizations

    throughout the worl

    Dionex Products(408) 737-0700

    North America:U.S./Canada(847) 295-7500

    South America:Brazil(55) 11 3731 5140

    Europe:Austria(43) 616 51 25

    Benelux

    (31) 20 683 9768(32) 3 353 4294

    Denmark(45) 36 36 90 90

    France(33) 1 39 30 01 10

    Germany(49) 61125 991 0

    Ireland(353) 644 0064

    Italy(39) 02 51 62 1267

    Sweden(46) 8 473 3380

    Switzerland(41) 62 205 9966

    United Kingdom(44) 1276 691722

    Asia Pacific:Australia(61) 2 9420 5233

    China(852) 2428 3282

    India(91) 22 2764 2735

    Japan(81) 6885 1213

    Korea(82) 2 2653 2580

    Singapore(65) 6289 1190

    Taiwan(886) 2 875 6655