10014_ftp

download 10014_ftp

of 8

Transcript of 10014_ftp

  • 8/6/2019 10014_ftp

    1/8

    Original Research

    Evaluation of the Rotator Cuff and Glenoid Labrum

    Using a 0.2-Tesla Extremity Magnetic Resonance

    (MR) System: MR Results Compared to Surgical

    Findings

    Frank G. Shellock, PhD, FACSM,1* Jack M. Bert, MD,2 Hollis M. Fritts, MD,3

    Cooper R. Gundry, MD,3 Ruth Easton, RT(R),2 and John V. Crues III, MD4

    The purpose of this investigation was to evaluate the diag-nostic capabilities of magnetic resonance imaging (MRI)

    performed using a dedicated-extremity MR system in de-tecting lesions of the rotator cuff and glenoid labrum. Thisretrospective study compared the MR results obtained in47 patients that underwent MRI using a 0.2-Tesla extrem-ity MR system (E-scan) to the surgical findings. MR imagesof the shoulder were obtained as follows: shoulder coil, T1-weighted, coronal-oblique and axial images; short Tauinversion recovery (STIR), coronal-oblique images; and T2-weighted, coronal-oblique, sagittal-oblique, and axial im-ages. The MR examinations were interpreted by threehighly experienced, musculoskeletal radiologists. Opensurgical (N 26) or arthroscopic (N 21) procedures wereperformed within a mean time of 33 days after MRI. Thesurgical findings revealed rotator cuff tears in 28 patientsand labral lesions in 9 patients. For the rotator cuff tears,

    the sensitivity, specificity, positive predictive value, andnegative predictive value were 89%, 100%, 100%, and90%, respectively. For the labral lesions, the sensitivity,specificity, positive predictive value, and negative predic-tive value were 89%, 95%, 80%, and 97%, respectively. The findings indicated that there was good agreementcomparing the MR results obtained using the low-field ex-tremity MR system to the surgical findings for determina-tion of lesions of the rotator cuff and glenoid labrum. No-tably, the statistical values determined for the use of thisMR system were comparable to those reported in the peer-reviewed literature for the use of whole-body, mid- andhigh-field-strength MR systems. J. Magn. Reson. Imaging 2001;14:763770. 2001 Wiley-Liss, Inc.

    Index terms: shoulder, MR; magnetic resonance (MR), com-parative studies; shoulder, surgery; lowfield MRI; extremity;rotator cuff

    MAGNETIC RESONANCE IMAGING (MRI) is the nonin-

    vasive procedure of choice for evaluation of diseases of

    the shoulder, especially with regard to identifying le-

    sions of the rotator cuff and glenoid labrum (114).

    Shoulder MRI is typically performed with conventional,

    whole-body MR systems.

    The creation of new metal alloys has permitted the

    development of smaller permanent magnets that, in

    turn, have enabled the construction of MR systems that

    are substantially smaller than whole-body scanners

    (1524).

    In 1993, the first extremity MR system became avail-

    able on a commercial basis (Artoscan, Esaote, Genoa,

    Italy; General Electric (GE) Medical Systems/Lunar

    Corporation, Madison, WI). This scanner uses a low-

    field-strength (0.2-Tesla) permanent magnet to image

    feet, ankles, knees, hands, wrists, and elbows. Acquir-

    ing MR examinations with this MR system offers several

    distinct advantages over whole-body scanners, includ-

    ing reduced overall costs, more convenient installation

    and siting, and greater patient comfort and safety (15

    26). Importantly, the diagnostic performance of the

    present-day extremity MR system for evaluation of

    musculoskeletal pathology has been reported to rival

    that of mid- and high-field-strength MR scanners (15

    18,2224). Disadvantages of the extremity MR system

    are comparable to those of whole-body, low-field-

    strength MR systems and include the inherent lower

    signal-to-noise ratio (SNR) and longer overall examina-

    tion times.

    In 1999, another extremity MR system was developed

    to image the shoulder in addition to the aforementioned

    body part (E-scan, Esaote, Genoa, Italy; GE Medical

    Systems/Lunar Corporation, Madison, WI). To our

    knowledge, there has been no prior evaluation of the

    diagnostic performance of this MR system for shoulder

    MRI. Therefore, the goal of this investigation was to

    determine the diagnostic capabilities of MRI performed

    using this extremity MR system in detecting lesions of

    the rotator cuff and glenoid labrum.

    1University of Southern California, Los Angeles, California.2Summit Landmark Orthopedics, St. Paul, Minnesota.3Center for Diagnostic Imaging, Minneapolis, Minnesota.4Radnet, Los Angeles, California.

    Contract grant sponsor: GE Medical Systems/Lunar Corporation.

    *Address reprint requests to: F.G.S., 7511 McConnell Ave., LosAngeles,CA 90045. E-mail: [email protected]

    Received June 15, 2001; Accepted August 21, 2001.

    JOURNAL OF MAGNETIC RESONANCE IMAGING 14:763770 (2001)

    2001 Wiley-Liss, Inc. 763DOI 10.1002/jmri.10014

  • 8/6/2019 10014_ftp

    2/8

    MATERIALS AND METHODS

    The study population for this investigation was selectedfrom a consecutive group of patients with suspectedshoulder pathology who underwent MRI and surgery

    between August 1999 and September 2000. Patients were not included if they had prior surgery or if thesurgical procedure was more than 180 days after MRI.

    The study group consisted of 47 patients, 31 men and16 women, ranging in age from 1683 years old (meanage, 52 years old). Open surgical (N 26) or arthro-scopic (N 21) procedures were performed within amean time interval of 33 days (range, 8155 days) afterMRI by eight different surgeons specially trained in thediagnosis and treatment of shoulder disorders. The pa-tients rotator cuff and glenoid labrum were routinelyinspected by the surgeon in each case. MRI reports

    were available to the surgeons at the time of surgery.

    MRI

    MRI was performed with a 0.2-Tesla extremity MR sys-tem (E-scan, Esaote, Genoa, Italy; GE Medical Sys-tems/Lunar Corporation, Madison, WI) and a dedi-cated, linear shoulder coil that has an integratedpreamplifier (Fig. 1). The gradient magnetic fields forthis MR system operate at 20 mT/m (slew rate, 25mT/m/msec).

    The patients were imaged supine with the arm in aneutral position or mild external rotation. The followingprotocol was used: T1-weighted (TR/TE, 540/20 msec),coronal-oblique (i.e., oriented parallel to the longitudi-nal course of the supraspinatus tendon), and axialplanes; T2-weighted (TR/TE, 2,200/80 msec), coronal-oblique, sagittal-oblique (i.e., oriented orthogonal to

    those of the coronal-oblique images), and axial planes;and short Tau inversion recovery (STIR) (TR/TI/TE,1,970/75/30 msec) coronal-oblique plane (40 of 47 pa-tients). The section thickness was 5 mm with an inter-section gap of 0.5 mm for all pulse sequences, with theexception of the STIR sequence, which used a 7-mmsection thickness and 1-mm inter-section gap. The fieldof view (FOV) was 1416 cm, the matrix size was 192128, and the number of excitations was 2. The imagingtime for this procedure ranged from 5070 minutes.Notably, the MR facility that conducted all of the MRI ofthe shoulder examinations is accredited for musculo-skeletal MRI by the Intersocietal Commission for the

    Accreditation of Magnetic Resonance Laboratories(ICAMRL).

    Interpretation of the MR Images

    The MR images were interpreted by three board-certi-fied radiologists with extensive experience (two radiol-ogists were MR musculoskeletal fellowship trained with10 years of experience; one radiologist had 15 yearsof MR musculoskeletal reading experience) in muscu-loskeletal MRI. The radiologists interpreted the MR ex-aminations in an independent, prospective manner.

    They had knowledge of the patients age, sex, and pre-sumptive diagnosis. The radiologists were not provided

    instructions or guidance with regard to specific criteriato use to interpret the MR images, nor were concessions

    made for the low-field-strength MR images. However,standard, previously described, well-accepted interpre-tation criteria were used (1 6,1013,2732).

    Data Collection

    The MRI and operative reports were reviewed to recordthe pertinent findings. Thus, data for the MR findings

    were extracted directly from the original preoperativereports, while data for the surgical findings were ex-tracted from the original operative reports obtainedfrom the orthopedic surgeons. Notably, this basic studydesign (i.e., retrospective review of MR and surgicalreports) has been used in several prior evaluations per-formed to compare MRI to surgical findings (13,8,12).

    Terms such as fibrillation, degeneration, tendonopa-thy, attenuation, fringe, and scuffing were categorizedas no tear in this study.

    Statistical Analysis

    The standard definitions for sensitivity, specificity, pos-itive predictive value, and negative predictive value

    were applied to the results of this study with regard tothe use of MRI for diagnosis of a rotator cuff tear and/orlabral lesion, with the surgical findings used as the goldstandard. Sensitivity was defined as the quotient oftrue-positive diagnoses on MRI (numerator) and true-positive plus false-negative diagnoses (denominator).Specificitywas defined as the quotient of true-negativediagnoses on MRI (numerator) and true-negative plusfalse-positive diagnoses (denominator).

    Positive predictive value was defined as the quotientof true-positive diagnoses (numerator) and all positivediagnoses (true-positive plus false-positive diagnoses

    (denominator)). Negative predictive value was definedas the quotient of true-negative diagnoses (numerator)and all negative diagnoses (true- plus false-negativediagnoses (denominator)).

    RESULTS

    Using the surgical findings as the gold standard, therewere 28 rotator cuff tears (i.e., 21 full-thickness tears; 2small tears, 8 10 mm; and 5 partial tears) and 9 labrallesions (anatomic distribution: 2 superior-labrum an-terior-posterior (SLAP); 3 superior; 1 anterior and su-perior; 2 anterior, superior, and posterior; and 1 re-

    ported massive tear) found in the patients. Ten patientshad neither a rotator cuff tear nor a labral lesion, whileone patient had both. The sensitivity, specificity, andpositive and negative predictive values for the MR read-ings for the rotator cuff and glenoid labrum are pre-sented in Table 1. Figures 24 show examples of shoul-der MR images selected to illustrate the results of thisstudy. Notably, MR images are representative examplesof the images that were acquired and that formed the

    basis for the accuracy rates presented in our paper.

    Rotator Cuff

    For the rotator cuff, MRI correctly identified 25 of the 28

    tears (89%). There were no false positives. For the threefalse negatives, partial-thickness tears of the rotator

    764 Shellock et al.

  • 8/6/2019 10014_ftp

    3/8

    cuff were missed by MRI that were seen on arthroscopicprocedures. Thus, none of the full-thickness tears weremissed by MRI. The positive predictive value, the per-centage of all (i.e., both full-thickness and partial-thick-ness) rotator cuff tears diagnosed on MRI that were

    subsequently diagnosed at surgery was 100%. The neg-ative predictive value was 90%.

    Glenoid Labrum

    For the glenoid labrum, MRI correctly identified eight of

    nine lesions (89%). For the two false positives, MRI

    reported small tears seen on the superior aspect of the

    labrum that were found to be normal on open surgicalprocedures. For the one false negative, a SLAP lesion

    Figure 1. a: The 0.2-Tesla extremity MR system (E-Scan) used to perform shoulder MRI in this study. Note the linear shouldercoil positioned in the center of the scanner. b: Example of patient positioned in the extremity MR system for MRI of the shoulder.

    Evaluation of Rotator Cuff and Glenoid Labrum 765

  • 8/6/2019 10014_ftp

    4/8

  • 8/6/2019 10014_ftp

    5/8

    may be indistinguishable from that which occurs withdegeneration, tendinitis, or tendonopathy (2,30 32).

    The similarity and overlap of signal patterns for theselesions generally creates interpretation problems re-

    gardless of the field strength of the MR system, theshoulder coils, or the pulse sequences used for imaging

    (17,30).In a recent study that involved an analysis of inter-

    pretive errors related to the use of high-field MRI todiagnose rotator cuff tears, Balich et al (30) reported

    that detection of partial tears was much less sensitiveand this low sensitivity could not be improved withcurrent imaging techniques or increased reader experi-

    ence. While the use of intra-articular gadolinium orsaline (i.e., MR arthrography) may improve the sensi-tivity and specificity of MRI for partial tears (30), it is

    unlikely to influence the ability of MRI to detect intra-subtance or bursal-sided partial-thickness tears, and

    this invasive procedure increases the time and expenseof the examination. Additionally, the overall impact on

    patient care is debatable.It is crucial to differentiate between full-thickness

    and partial-thickness tears because the treatments forthese rotator cuff lesions differ. A full-thickness tear

    typically requires surgery. Typical clinical managementof a partial-thickness tear initially involves conservative

    treatment, with surgical intervention utilized if symp-toms persist for more than 6 months. Thus, from apatient management consideration, there may be no

    major implications for missing partial tears on MRI,especially since therapeutic decisions for these rotator

    cuff disorders are predominantly guided by clinicalfindings. However, some surgeons may elect to treat a

    partial-thickness tear in the dominant arm of an activeindividual with debridement and, when appropriate,

    acromioplasty or excision of acromioclavicular (AC)joint spurs.

    Shoulder MRI is generally considered useful for de-tection of labral pathology (813,33,34). Studies usingconventional MRI (i.e., without the use of intra-articu-lar contrast) performed with mid- and high-field-strength MR systems have reported sensitivities rang-

    ing from 44%95% and specificities ranging from 63%91% for the glenoid labrum (including identification ofSLAP lesions) (811,15,33,34). Differences in imagingprotocols and techniques probably contribute to these

    variances in sensitivity and specificity. Additionally, anumber of interpretation pitfalls exist that may be re-sponsible for the wide differences in diagnostic capabil-ities reported for the use of MRI in diagnosing labrallesions (12).

    Using the extremity MR system, the sensitivity andspecificity for assessment of the glenoid labrum were89% and 95%, respectively. Again, this diagnostic per-formance was comparable to that reported for mid- andhigh-field MR systems. For the two false positives, the

    MR reports indicated the presence of ill-defined (i.e.,the specific term used by the interpreting radiologist inthe MRI report) tears that were found to be normal onopen surgical procedures. For the one false negative,the MR report indicated that the glenoid labrum ap-pears intact without visualization of a discrete labraltear, while a type I SLAP lesion was seen on arthros-copy.

    The difficulty of using conventional MRI to diagnoseand characterize SLAP lesions is well known

    Figure 4. Axial, T1-weighted MR image of the shoulder show-

    ing a tear of the anterior glenoid labrum, confirmed by surgicalfindings.

    Figure 3. Coronal-oblique, T1-weighted MR image of theshoulder showing a SLAP tear, confirmed by surgical findings.

    Evaluation of Rotator Cuff and Glenoid Labrum 767

  • 8/6/2019 10014_ftp

    6/8

    (8,9,11,12). In general, the use of intra-articular con-trast improves the diagnostic capabilities of MRI for thisapplication (14). Therefore, in our opinion, it may beadvisable to use this technique for patients who present

    with physical findings suggestive of superior labral le-sions.

    In general, the findings of this investigation showed

    the ability of a low-field extremity MR system to accu-rately identify lesions of the rotator cuff and glenoidlabrum. Studies comparing the performance of conven-tional MRI using high-field vs. low-field MR systems forassessment of labral lesions have been conducted by anumber of research groups (8,3335). Allmann et al (33)reported that shoulder MRI at 0.2- and 1.0-Tesla wascomparable. In their study group of 35 patients, thesensitivity and specificity were 91% and 91%, respec-tively, for labral pathology for both the 0.2- and 1.0-

    Tesla MR systems (33).Shih et al (34) studied patients that underwent MRI

    performed with 0.3- and 1.5-Tesla MR systems. Withregard to labral lesions, the performance of the high-

    field and low-field MR systems was almost equal (34).However, for the rotator cuff, Shih et al (34) reportedthat the high-field MR system was better in the differ-entiation of tendinitis from tears, in the confirmation of

    bursitis, and in the detection of subscapularis lesions.Notably, Shih et al (34) did not use fat suppression orSTIR techniques for the shoulder imaging protocols,

    which may explain these findings (to be discussedlater).

    Tung et al (8) determined the performance character-istics of high-field (1.5-Tesla) and low-field (0.2-Tesla)MRI for diagnosis of SLAP tears. The sensitivity/speci-ficity of high-field vs. low-field MRI was 95%/63% and

    64%/79%, respectively (8). Thus, the performancecharacteristics of high-field MRI were reported to besuperior to those of low-field MRI. However, Tung et al(8) obtained MR images in oblique sagittal and obliquecoronal planes using a frequency-selected fat-saturatedpulse sequence, without acquiring MR images with acomparable technique (e.g., STIR) for the low-field MRexaminations. Additionally, they used smaller sectionthicknesses (as Tung et al (8) indicated, thinner slicesshould also increase the sensitivity of MRI for detectingtears of the glenoid labrum) and smaller FOVs for thehigh-field MR studies. These major protocol differencesdetract from the relative significance of their finding forhigh- vs. low-field MR systems.

    The relationship between image quality and magneticfield strength has created a strong bias in favor of theuse of high-field MR systems. However, advances inlow-field MR technology, including new magnet de-signs, improved RF coils, faster gradient systems, andmore sophisticated pulse sequences, have helped tocounter the previous limitations of low-field-strengthMR systems (16). Even though the SNR is directly pro-portional to field strength, the inherent lower SNR may

    be addressed on low-field MR systems by using addi-tional signal averages or excitations, narrower band-

    widths, better RF coil designs, and optimized pulse se-quences (16). The protocol that was used for shoulder

    MRI was devised to accomplish this important aspect oflow-field MRI. Furthermore, for musculoskeletal MRI,

    the contrast-to-noise ratio (CNR) is a more clinically

    relevant parameter because it determines the extent to

    which adjacent structures can be distinguished from

    one another and the general conspicuity of pathologic

    findings (16). Unlike SNR, the CNR for MRI does not

    increase substantially with the strength of the static

    magnetic field (16). The CNR is primarily dependent on

    imaging parameters.Of further note regarding the issue of SNR for low-

    field MR systems, we would like to emphasize that,

    despite obtaining MR images of the shoulder that may

    be less aesthetically pleasing compared to what is rou-

    tinely acquired using a high-field MR system, the diag-

    nostic accuracy of the dedicated-extremity MR system

    was comparable to that reported for high-field MR sys-

    tems.

    Focal and/or diffuse increases in signal intensity on

    T2-weighted images are associated with rotator cuff

    and labral pathology (112,2734). For shoulder MR

    examinations, it is important to perform fat-sup-

    pressed, T2-weighted imaging for optimal diagnostic

    performance because this technique appears to be

    more sensitive to increases in signal intensity as a re-

    sult of the extended gray scale (5,7,28,29,36). For ex-

    ample, Reinus et al (5) reported that use of the fat

    suppression technique improved the detection of both

    full-thickness and partial-thickness tears of the rotator

    cuff, compared with standard spin-echo pulse se-

    quences.

    Using a low-field MR system, it is difficult to acquire

    fat-suppressed MR images using frequency-selective

    techniques because the difference between the fat and

    water spectral peaks is field-strength dependent (16).

    Therefore, in this study, the MR protocol involved ac-

    quisition of fat-suppressed images using a STIR pulsesequence (i.e., for 40 of the 47 patients imaged using

    this pulse sequence). Notably, the use of the STIR se-

    quence has been reported to be more sensitive than

    fat-suppressed, T2-weighted techniques for detection of

    musculoskeletal lesions (3739). Thus, it is possible

    that the use of the STIR sequence contributed to the

    overall positive findings of this investigation.

    Our investigation had a number of possible limita-

    tions. First, it was a retrospective study that included a

    large number of symptomatic patients (based on a thor-

    ough examination by the referring orthopedic surgeon).

    Therefore, bias was introduced in favor of diagnosing

    rotator cuff and labral lesions on MRI, even though thereviewing radiologists were unaware of the final surgi-

    cal findings.

    Second, this investigation was subject to verification

    or work-up bias (8). That is, it is well-known that when

    a study is restricted to patients who require an invasive

    test to verify disease, then the study population may be

    biased toward patients with more severe disease. In this

    study, the reference standard was the surgical findings

    for the patients. Obviously, patients with negative find-

    ings on MRI are unlikely to undergo surgical proce-

    dures, and this tendency may result in a report of rel-

    atively higher sensitivity for MRI than is true because

    the number of negative tests (both false negative andtrue negative) is decreased. The lack of an adequate

    768 Shellock et al.

  • 8/6/2019 10014_ftp

    7/8

    control group is understandable and unavoidable con-sidering the invasive aspects of the gold standard test.

    Third, the radiologists who interpreted the MR im-ages were experienced, musculoskeletal MR specialists.In general, more knowledgeable readers have a greaterability to identify shoulder pathology on MR images(28,29,31). A considerable degree of experience is re-

    quired before one becomes adept at interpreting MRimages, particularly with regard to shoulder MR, due tothe various interpretive pitfalls present for this joint(1,12,29).

    Fourth, the use of arthroscopic procedures (arthro-scopic procedures were used for 21 of 47 patients, 45%of the patients in this study, as opposed to open surgi-cal procedures) as a gold standard of reference may bea weakness because there may be surgeon-dependent

    variations, and more importantly, subtle findings mayescape arthroscopic visualization (29,40). Further-more, some arthroscopists routinely explore the sub-acromial space, while others do not. Such problems of

    visualization may be compounded by the presence of

    synovitis and marked degeneration of the glenohumeraljoint, as well as bursitis and marked degeneration af-fecting the subacromial space (29).

    Finally, there was no attempt to assess inter- or in-traobserver variability for the results of this study. This

    was not a primary goal of this investigation (i.e., similarto other previously published, peer-reviewed articlesthat evaluated the diagnostic performance of shoulderMRI that, likewise, did not evaluate these factors) (13).

    Additionally, it was deemed unnecessary to determineinter- and intraobserver variability because of the ex-tensive experience of the reading radiologists. Previousstudies performed to assess intra- and interobserver

    variability for MRI of the shoulder indicated that high values are directly related to the level of training andexperience of the radiologist (29,30).

    In conclusion, the findings of this study indicatedthat shoulder MRI performed with a low-field extremityMR system exhibited acceptable diagnostic accuracy indepicting lesions of the rotator cuff and labrum basedon a comparison between the MR results and surgicalfindings. The overall diagnostic capabilities were com-parable to those reported in the peer-reviewed literaturefor the use of whole-body and mid- and high-field MRsystems.

    ACKNOWLEDGMENT

    An unrestricted research grant was provided by GEMedical Systems/Lunar Corporation, Madison, WI, toF.G. Shellock to support this investigation.

    REFERENCES

    1. Zlatkin MB, Ionnotti JP, Roberts MC, et al. Rotator cuff tears:

    diagnostic performance of MR imaging. Radiology 1989;172:223

    229.

    2. Rafii M, Firooznia H, Sherman O, et al. Rotator cuff lesions: signal

    patterns at MR imaging. Radiology 1990;177:817823.

    3. Farley TE, Neumann CH, Steinbach LS, Jahnke AJ, Petersen SS.

    Full-thickness tears of the rotator cuff of the shoulder: diagnosis

    with MR imaging. AJR Am J Roentgenol 1992;158:347351.

    4. Stoller DW, Fritz RC. Magnetic resonance imaging of impingement

    and rotator cuff tears. Magn Reson Imaging Clin N Am 1993;1:4763.

    5. Reinus WR, Shady KL, Mirowitz SA, Totty WG. MR diagnosis of

    rotator cuff tears of the shoulder: value of using T2-weighted fat-

    saturated images. AJR Am J Roentgenol 1995;164:14511455.

    6. Zlatkin MB, Falchook FS. Magnetic resonance pathology of the

    rotator cuff. Top Magn Reson Imaging 1994;6:94120.

    7. Needell SD, Zlatkin MB. Comparison of fat-saturated fast spin echo

    versus conventional spin-echo MRI in the detection of rotator pa-

    thology. J Magn Reson Imaging 1997;7:674 677.

    8. Tung GA, Entzian D, Green A, Brody JM. High-field and low-field

    MR imaging of superior glenoid labral tears and associated tendoninjuries. AJR Am J Roentgenol 2000;174:11071114.

    9. Cartland JP, Crues JV, Stauffer A, Nottage W, Ryu RK. MR imaging

    in the evaluation of SLAP injuries of the shoulder: findings in 10

    patients. AJR Am J Roentgenol 1992;159:787792.

    10. Garneau RA, Renfrew DL, Moore TE, el-Khoury GY, Nepola JV,

    Lemke JH. Glenoid labrum: evaluation with MR imaging. Radiology

    1991;179:519522.

    11. Monu J, Pope TL Jr, Chabon SJ, Vanarthos WJ. MR diagnosis of

    superior labrum-anterior to posterior (SLAP) injuries of the glenoid

    labrum: value of routine imaging without intraarticular injection of

    contrast material. AJR Am J Roentgenol 1994;163:14251429.

    12. Tuite MJ, Cirillo RL, De Smet AA, Orwin JF. Superior labrum

    anterior-posterior (SLAP) tears: evaluation of three MR signs on

    T2-weighted images. Radiology 2000;215:841845.

    13. Fritts HM, Craig EV. MRI of the shoulder. Semin Ultrasound CT MR

    1994;15:341365.14. Grainger AJ, Elliott JM, Campbell RS, et al. Direct MR arthrogra-

    phy: a review of current use. Clin Radiol 2000;55:163176.

    15. Kersting-Sommerhoff B, Hof N, Lenz M, Gerhardt P. MRI of periph-

    eral joints with a low-field dedicated system: a reliable and cost-

    effective alternative to high-field units? Eur Radiol 1996;6:561

    565.

    16. Peterfy CG, Roberts T, Genant HK. Dedicated extremity MR imag-

    ing: an emerging technology. Radiol Clin N Am 1997;35:120.

    17. Franklin PD, Lemon RA, Barden HS. Accuracy of imaging the me-

    nisci on an in-office, dedicated, magnetic resonance imaging ex-

    tremity system, Am J Sports Med 1998;25:382388.

    18. Ahn JM, Sartoris DJ, Kank HS, et al., Gamekeeper thumb: com-

    parison of MR arthrography with conventional arthrography and

    MR imaging in cadavers. Radiology 1998;206:737741.

    19. Shellock FG, Walgenbach A, Stone K, Crues JV. Determination of

    chondral lesions using a 0.2 Tesla extremity MR system: diagnosticfindings compared to arthroscopy [Abstract]. Radiology 2000;214:

    P07H.

    20. Shellock FG, Stone K, Crues JV. Development and clinical appli-

    cations of kinematic MRI of the patellofemoral joint using an ex-

    tremity MR system. Med Sci Sport Exer 1999;31:788791.

    21. Shellock FG, Mullen M, Stone K, Coleman M, Crues JV. Kinematic

    MRI evaluation of the effect of bracing on patellar positions: qual-

    itative assessment using an extremity MR system. J Athletic Train-

    ing 2000;35:4449.

    22. Steinborn M, Heuck A, Jessel C, Bonel H, Reiser M. Magnetic

    resonance imaging of lateral epicondylitis of the elbow with a 0.2-T

    dedicated system. Eur Radiol 1999;9:13761380.

    23. Merhemic Z, Breitenseher M, Trattnig S, Happel B, Kukla C, Rand

    T, Imhof H. MRI of the ankle joint. Comparison of the 1.0-T and the

    0.2-T units. Radiologe 1999;39:4146.

    24. Hollister MC. Dedicated extremity MR imaging of the knee: how lowcan we go? Magn Reson Imaging Clin N Am 2000;8:225241.

    25. Shellock FG, Crues JV. Aneurysm clips: assessment of magnetic

    field interaction associated with a 0.2-T extremity MR system. Ra-

    diology 1998;208:407409.

    26. Shellock FG, ONeil M, Ivans V, et al. Cardiac pacemakers and

    implantable cardiac defibrillators are unaffected by operation of an

    extremity MR system. AJR Am J Roentgenol 1999;172:165172.

    27. Iannotti JP, Zlatkin MB, Esterhai JL, Kressel HY, Dalinka MK,

    Spindler KP. Magnetic resonance imaging of the shoulder: sensi-

    tivity, specificity, and predictive value. J Bone Joint Surg Am 1991;

    73:1729.

    28. Quinn SF, Sheley RC, Demlow TA, Szulmowski J. Rotator cuff

    tendon tears: evaluation with fat-suppressed MR imaging with ar-

    throscopic correlation in 100 patients. Radiology 1995;497501.

    29. Robertson PL, Schweitzer ME, Mitchell DG, et al. Rotator cuff dis-

    orders: interobserver and intraobserver variation in diagnosis withMR imaging. Radiology 1995;194:831835.

    Evaluation of Rotator Cuff and Glenoid Labrum 769

  • 8/6/2019 10014_ftp

    8/8

    30. Balich SM, Sheley RC, Brown TR, Sauser DD, Quinn SF. MR im-

    aging of the rotator cuff tendon: interobserver agreement and anal-

    ysis of interpretive errors. Radiology 1997;204:191194.

    31. Wnorowski DC, Levinsohn EM, Chamberlain BC, McAndrew DL.

    Magnetic resonance imaging assessment of the rotator cuff: is it

    really accurate? Arthroscopy 1997;13:710719.

    32. Shellock FG, Stoller D, Crues JV. MRI of the shoulder. A rational

    approach to the reporting of findings. J Magn Reson Imaging 1996;

    6:268270.

    33. Allmann KH, Walter O, Laubenberger J, et al. Magnetic resonancediagnosis of the anterior labrum and capsule. Effect of field

    strength on efficacy. Invest Radiol 1998;33:415420.

    34. Shih TT, Chen WG, Su CT, Huang KM, Ericson F, Chiu LC. MR

    patterns of rotator cuff and labral lesions: comparison between

    low-field and high-field images. J Formos Med Assoc 1994;92:146

    151.

    35. Merl T, Scholz M, Gerhardt P, et al. Results of a prospective mul-

    ticenter study for evaluation of the diagnostic quality of an open

    whole-body low-field MRI unit. A comparison with high-field MRI

    measured by the applicable gold standard. Eur J Radiol 1999;30:

    4353.

    36. Mirowitz SA. Normal rotator cuff: MR imaging with conventional

    and fat-suppression techniques. Radiology 1991;180:735740.

    37. Hilfiker P, Zanetti M, Debatin JF, McKinnon G, Hodler J. Fast

    spin-echo inversion-recovery imaging versus fast T2-weighted

    spin-echo imaging in bone marrow abnormalities. Invest Radiol

    1995;30:110114.

    38. Nakatsu M, Hatabu H, Itoh H, et al. Comparison of short inversion

    time inversion recovery (STIR) and fat-saturated (chemsat) tech-

    niques for background fat intensity suppression in cervical and

    thoracic MR imaging. J Magn Reson Imaging 2000;11:5660.

    39. Steinborn M, Heuck A, Maier M, Schnarkowski P, Scheidler J,

    Reiser M. MRI of plantar fasciitis. Rofo Fortschr Geb Rontgenstr

    Neuen Bildgeb Verfahr 1999;170:4146.

    40. Thomazeau H, Gleyze P, Lafosse L, Walch G, Kelberine F, Coudane

    H. Arthroscopic assessment of full-thickness rotator cuff tears.

    Arthroscopy 2000;16:367372.

    770 Shellock et al.