1Standards and Calibration Laboratory, SCL
Evaluation of Flicker Severity in Standards and Calibration
Laboratory (SCL)
Speaker: Chung Yin, PoonStandards and Calibration Laboratory (SCL)The Government of the Hong Kong Special
Administrative Region
2Standards and Calibration Laboratory, SCL
Learning Objectives
• To introduce the meaning of flickers in power line
• To describe a method for calibrating flicker source and flickermeter
3Standards and Calibration Laboratory, SCL
Introduction
• Flicker is the change in brightness of a lamp due to rapid fluctuations in the voltage of the power supply
• Cause disturbance to human visual perception system
• Voltage change characteristic at the terminals of the equipment under test, such as microwave ovens, washing machines
• All types of voltage fluctuations may be assessed by direct measurement using a flickermeter
4Standards and Calibration Laboratory, SCL
Introduction
Source: Figure B.1, IEC 61000-4-15-2010 Flickermeter - Functional and design specifications
5Standards and Calibration Laboratory, SCL
Evaluation of flicker parameters
• Flickermeter (Functional and design specification in IEC 61000-4-15)
• Flicker source (Limits of voltage changes produced by an equipment in IEC 61000-3-3)
• SCL developed a system to evaluate the accuracy of flicker source and flicker measurement functions of flickermeter
6Standards and Calibration Laboratory, SCL
Operating Principle• Flicker waveform is modeled as an amplitude
modulated (AM) signal in power line (120 V / 230 V)
7Standards and Calibration Laboratory, SCL
Operating Principle
• Relative voltage change / Voltage fluctuation
• Vupper: Amplitudes of upper sideband
• Vlower: Amplitude of lower sideband
• Vcarrier: Amplitude of the carrier
8Standards and Calibration Laboratory, SCL
Calibration of flicker source - Setup
9Standards and Calibration Laboratory, SCL
Resistive voltage divider
10Standards and Calibration Laboratory, SCL
Calibration of flicker source - Setup
11Standards and Calibration Laboratory, SCL
Calibration of flicker source
• Voltages of lower sideband (Vlower), carrier (Vcarrier) and upper sideband (Vupper) in dB are measured by signal analyzer
• Accuracy of the voltage fluctuation of standard flicker source can be determined
12Standards and Calibration Laboratory, SCL
Calibration of flickermeter
• Flickermeters under test are calibrated by direct comparison with the standard flicker source
• Calibration in terms of voltage fluctuation and modulating frequency
• Instantaneous flicker sensation (Pinst) will be compared against the theoretical values as illustrated in IEC 61000-4-15
13Standards and Calibration Laboratory, SCL
Calibration of flickermeter - Setup
14Standards and Calibration Laboratory, SCL
What to be measured?
15Standards and Calibration Laboratory, SCL
Measurement uncertainties
• Depends on– Amplitude accuracy of signal analyzer– Performance of standard flicker source– Calibration setup
16Standards and Calibration Laboratory, SCL
Maximum uncertainty of voltage fluctuation measurement
17Standards and Calibration Laboratory, SCL
Measurement uncertainties
• Flicker source: Measurement uncertainty < 0.03% at 95% confidence level
• Flickermeter: Measurement uncertainty < 0.3% for determination of flicker severities at 95% confidence level
• Type testing of Class F1 flickermeters according to IEC 61000-4-15– Tolerance of 8% in Pinst for sinusoidal voltage changes
18Standards and Calibration Laboratory, SCL
Ratio between the specifications of standard flicker source and maximum measurement uncertainty (TUR)
19Standards and Calibration Laboratory, SCL
Thank you