1-s2.0-S0019057814002997-main
-
Upload
madalinaamg -
Category
Documents
-
view
214 -
download
0
Transcript of 1-s2.0-S0019057814002997-main
-
8/18/2019 1-s2.0-S0019057814002997-main
1/8
A low cost instrumentation system to analyze different types
of milk adulteration
Siuli Das a,n, Mulinti Sivaramakrishna b, Karabi Biswas c, Bhaswati Goswami d
a Department of Instrumentation Engineering, RAIT, Nerul, Navi Mumbai, Indiab Department of ECE, Polytechnic College, Pillaripattu, Andhra Pradesh, Indiac Department of Electrical Engineering, IIT Kharagpur, Indiad Department of Instrumentation and Electronics Engineering, Jadavpur University, Kolkata, India
a r t i c l e i n f o
Article history:
Received 25 January 2014
Received in revised form
17 September 2014
Accepted 30 November 2014Available online 20 December 2014
This paper was recommended for publica-
tion by Dr. Didier Theilliol
Keywords:
Milk adulteration
Instrumentation system
Synthetic milk
Liquid-whey
PMMA-lm
a b s t r a c t
In this paper, the design of a complete instrumentation system to detect different types of milk
adulteration has been reported. A simple to use indicator type readout device is reported which can be
used by milk community people. A low cost microcontroller based automatic sensing system is also
reported to detect ‘synthetic milk’, which has been reconstructed after adulterating the milk with ‘liquid-
whey’.
& 2014 ISA. Published by Elsevier Ltd. All rights reserved.
1. Introduction
Food quality monitoring is a serious issue and addressed by
different researchers [1–4]. Among these milk is one of the most
vulnerable food which is adulterated very easily. Adulteration
achieved with low value ingredients like water and whey into natural
milk is known as ‘economic adulteration’. It is a very common
practice by the milk community people to add water or ‘liquid-
whey’ [5,6] to increase the volume, this in turn reduces the milk
quality. Additionally whey retains many natural properties of milk, so
preparation of synthetic milk from liquid-whey is simple and can
camouage the natural milk easily. Hence, this becomes a serious
concern to the dairy rms who buy milk from thousands of different
milk suppliers and need a simple, robust and bio-compatible auto-
mated sensing system for quality control.There are several methods [7–9] available in the literature to
detect different types of adulterants in milk. Methods to determine
additional water present in the milk are explained in [10–12]. The
volume of water mixed in milk is detected by electrical conduc-
tivity (EC) method. In this method [11], quantity of ions decreases
so the resistance is higher and the EC is changed. To detect fat and
water content present in milk, electrical admittance spectroscopy
[7,10] method has been used. Another method used to determinethe additional water content in milk is freezing point osmometry
[12]. In [13–16], different methods are proposed for determination
of whey as an adulterant. Low priced whey [14,16] and rennet
whey solid [13,15] are often mixed with liquid milk and milk
powder. To detect the fraudulent addition of rennet (enzymes for
curdling of milk) whey solid in ultra high temperature (UHT) milk
[13], capillary electrophoresis is used. Sometimes urea [17–21] is
added to milk to increase the shelf life. In [17] a potentiometric
biosensor method is reported to detect urea adulteration in milk.
Measure of change of pH [19,21] is another method of urea
detection in milk. Enzyme based piezoelectric sensor [18] is also
used to detect urea in milk by measuring the pressure of the gas
evolved in the milk sample. Addition of urea in milk is also
detected by near infrared spectroscopy method [20]. But most of these methods are expensive and time consuming and need
skilled manpower. It is, therefore, necessary to develop a new
and easy to use instrumentation system for the rapid and reliable
detection of this kind of fraud.
So the aim of this paper is to develop a low cost, user friendly
instrumentation system for measurement of these kinds of adul-
teration and can be easily installed in dairy industry.
In this paper a constant phase element (CPE) sensor is used to
detect the adulteration in milk. The sensor is a stick type two
terminal device and when dipped inside a medium the phase
Contents lists available at ScienceDirect
journal homepage: www.elsevier.com/locate/isatrans
ISA Transactions
http://dx.doi.org/10.1016/j.isatra.2014.11.021
0019-0578/& 2014 ISA. Published by Elsevier Ltd. All rights reserved.
n Corresponding author. Tel.: þ 91 022 27709574; fax: þ 91 022 27709573.
E-mail address: [email protected] (S. Das).
ISA Transactions 56 (2015) 268–275
http://www.sciencedirect.com/science/journal/00190578http://www.elsevier.com/locate/isatranshttp://dx.doi.org/10.1016/j.isatra.2014.11.021mailto:[email protected]://dx.doi.org/10.1016/j.isatra.2014.11.021http://dx.doi.org/10.1016/j.isatra.2014.11.021http://dx.doi.org/10.1016/j.isatra.2014.11.021http://dx.doi.org/10.1016/j.isatra.2014.11.021mailto:[email protected]://crossmark.crossref.org/dialog/?doi=10.1016/j.isatra.2014.11.021&domain=pdfhttp://crossmark.crossref.org/dialog/?doi=10.1016/j.isatra.2014.11.021&domain=pdfhttp://crossmark.crossref.org/dialog/?doi=10.1016/j.isatra.2014.11.021&domain=pdfhttp://dx.doi.org/10.1016/j.isatra.2014.11.021http://dx.doi.org/10.1016/j.isatra.2014.11.021http://dx.doi.org/10.1016/j.isatra.2014.11.021http://www.elsevier.com/locate/isatranshttp://www.sciencedirect.com/science/journal/00190578
-
8/18/2019 1-s2.0-S0019057814002997-main
2/8
angle between the two terminals remains constant (hence, the
name constant phase element (CPE)). But if the property of the
measuring medium is changed the phase angle also changes [22].
This means that the phase angle of the sensor will be different for
plain milk and the milk with some impurity [9]. The change of
phase angle is then detected with the help of a phase detector
circuit [23]. Indicator LEDs are used to display the type of
adulteration. An automatic detection system for synthetic milk is
also developed and studied. The advantages of using such sensorare that the electrodes are coated with the poly-methylmetaha-
crylate(PMMA), which makes it bio-compatible and the milk
property does not get changed when the sensor is dipped in.
Moreover, it is a stick type rigid probe and can be easily dipped
inside the measuring medium, an essential requirement for auto-
matic detection.
The paper is organized as follows: Section 2 describes the
building blocks of the instrumentation system which includes
sensor design, signal processing block with the detector circuit and
display unit. Section 3 presents the automatic detection system
and conclusion is given in Section 4.
2. Design of the instrumentation system
Fig. 1 shows the block diagram of the developed instrumenta-
tion system to detect the adulteration of pure milk with different
types of adulterants. It includes sensor, signal conditioning circuit
and display circuit.
2.1. Development of the sensor
The sensor, shown in Fig. 1 [9], is a copper cladded printed
circuit board, cut into a suitable dimensions and coated with a thin
lm of Polymethyl-methacrylate (PMMA) by using spin coating
technique [24]. For determination of different adulteration of milk
the sensor has to be dipped inside the test medium as it has been
observed that its impedance changes with the interaction of the
sensor with the test medium [9]. In particular, the phase angle
changes when milk sample is adulterated with different materials
[9,24,25]. Hence in this measurement “change of phase angle” is
considered as the sensor output. Moreover, this sensor has the
property that for a particular measurement the phase angle remains
constant over almost one decade of frequency which is an essential
requirement for sensing system where change of signal frequency
for interference or other effect will not hamper the measurement.
The sensor with coating thickness 13 μm, named as FOE13, isselected for designing the instrumentation system for detection of
tap water and urea as an adulterant in pure milk in the frequency
range 5–12 kHz. It will be worth to mention that the phase angle
of the FOE13 sensor remains constant in this frequency range at a
particular test medium. Other sensors with different lm coatingcan be fabricated which gives constant phase angle in different
frequency zones [9].
FOE13 sensor is considered for development of the instrumen-
tation system because the circuit elements used are suitable for
the frequency range 5–12 kHz. So, here for development of the
instrumentation system, measurements are performed at 7 kHz
which is almost at the middle of the bandwidth of the sensor.
Hence the choice of frequency depends on the bandwidth of the
circuit element and the range of frequency where phase angle
remains constant. This value is chosen because due to environ-
mental or other effect, if the frequency of the detector circuit
shifts, the output of the sensor will remain the same.
2.1.1. CalibrationBefore calibrating the instrumentation system, each block has
to be characterized. For that purpose, the resolution, sensitivity,
repeatability and reproducibility of the sensor block are checked
with 7 kHz signal frequency.
The phase angle of FOE13 sensor is measured with LCR meter
(Agilent Precision Impedance Analyser 4294A), after dipping 2 cm
of the sensor separately in different test mediums. Four different
buffer pH solutions (pH 2.0, pH 4.0, pH 7.0 and pH 9.2), pure milk,
milk adulterated with different % of tap water and milk adulter-
ated with different concentration of urea are the testing mediums.
For measurement purpose, the frequency of the sinusoidal signal is
varied from 100 Hz to 4 MHz with a peak to peak voltage of 1 V.
The readings are repeated 5 times to check the repeatability and
the average values are considered for analysis. It is observed that
the deviation from its average values are negligibly small.
FOE13 sensor is characterized by standard buffer solutions of
pH 2.0, pH 4.0, pH 7.0 and pH 9.2. The graph of phase angle with
respect to pH value is shown in Fig. 2 and the average values are
tabulated in Table 1. A line t, y ¼mx þ c yields ‘m’, ‘c ’ and t factor
Fig. 1. Block diagram of the instrumentation system.
S. Das et al. / ISA Transactions 56 (2015) 268 – 275 269
-
8/18/2019 1-s2.0-S0019057814002997-main
3/8
R2 values which are shown as
y ¼ 3:760 x 2:594 with R2 ¼ 0:995 ð1Þ
From Eq. (1) it is observed that R2 value is close to 1 which
indicates that the relation of phase angle and pH values of FOE13
sensor (Fig. 2) is almost linear, slope is negative that is the phase
angle value decreases with an increase of the pH value.
2.1.2. Sensing adulterated milk
FOE13 sensor is tested by dipping into milk and milk adulter-
ated with different percentages of tap water and urea for three
consecutive days. The outputs are shown in Figs. 3 and 4 and the
average values of the sensor are tabulated in Tables 2 and 4
respectively.
2.1.3. Tap water
From Fig. 3, it is evident that sensor FOE13 has the capability of
identifying the presence of tap water when at least 5 ml of tap wateradded in 50 ml of milk. There is approximately 21 change of phase
angle with pure milk and milk adulterated with 10% tap water.
It has been found from Table 2 that FOE13 sensor is showing
reproducibility in consecutive three days. It is also observed from
Table 3 that the t factors in all the 3 days are approximately 0.8. The
slopes and intercept are almost the same (for all the 3 days) and the
values are negative ( 0.06 and 25.5), that is with an increase of %
of tap water, phase angle decreases. From Table 2, the threshold value
of phase with 10% tap water adulteration is almost 1.61.
2.1.4. Urea
To prepare synthetic milk and to increase the SNF (Solid Not Fat)
value, urea added to natural milk is 0.2–
0.7 mg/ml [26,27]. In our
2 4 6 8 10−40
−35
−30
−25
−20
−15
−10
pH value
P h a s e a n g l e ( d e g r e e )
y = −3.760x − 2.594
R 2 = 0.995
Fig. 2. Phase angle behavior of FOE13 sensor in different ionizing medium.
Table 1
Phase angle in different buffer solutions of FOE13 sensor at 7 kHz.
pH Avg θ (deg)
2.0 10.86
4.0 16.52
7.0 29.13
9.2 37.36
0 10 20 30 40−29
−28
−27
−26
−25
−24
Different % of tap water adulteration
P h a s e a n g l e ( d e g
r e e )
1st day
2nd day
3rd day
y(1st day) = −0.065x−25.50
y(2nd day) = −0.060x−25.76
y(3rd day) = −0.055x−26.51
Fig. 3. Phase angle versus % of tap water adulteration.
0 0.2 0.4 0.6 0.8 1 1.2−27
−26
−25
−24
−23
−22
−21
−20
−19
−18
−17
Different concentration of urea adulteration (mg)
P h a s e
a n g l e ( d e g r e e )
1st day
2nd day
3rd day
y(1st day) = 5.974x − 24.33
y(2nd day) = 5.858x − 24.65
y(3rd day) = 5.769x − 25.41
Fig. 4. Phase angle versus % of urea adulteration.
Table 2
Phase angle in milk and milk adulterated with tap water of FOE13 sensor at 7 kHz.
Day Test medium pH value Avg θ
(deg)
1 Pure milk 6.82 24.96
Added 10% tap water 6.90 26.73
Added 20% tap water 6.97 27.05
Added 40% tap water 7.07 27.85
Tap water 7.34 28.91
2 Pure milk 6.85 25.30
Added 10% tap water 6.94 26.88
Added 20% tap water 7.00 27.15
Added 40% tap water 7.11 27.98
Tap water 7.86 29.01
3 Pure milk 6.83 26.11
Added 10% tap water 6.92 27.50
Added 20% tap water 6.98 27.75
Added 40% tap water 7.09 28.53Tap water 7.46 28.95
Table 3
Slope, zero error and residuals with tap water adulteration.
Day Slope (m) Zero error (c ) R2
1st 0.065 25.50 0.831
2nd 0.060 25.76 0.854
3rd 0.055 26.51 0.867
S. Das et al. / ISA Transactions 56 (2015) 268 – 275270
-
8/18/2019 1-s2.0-S0019057814002997-main
4/8
study, milk is adulterated with different concentration of urea and
change of phase angle is tabulated in Table 4. From Table 4, it is found
that the sensor is showing reproducibility and there is approximately
51 change of phase angle with pure milk and milk adulterated with
0.6 mg urea per ml milk. Hence the threshold value of phase with
0.6 mg urea per ml milk is almost 51. In this case the t factor is
approximately 0.9 which is close to 1 that means the relationship is
almost linear and shown in Table 5. The slopes are also the same for
3 consecutive days and are positive, so phase angle increases with an
increase of adulteration.
It is worth to mention that addition of urea has a reverse
response compared to the addition of water. The reason may be
that milk is a biological substance and it decomposes urea into
different ions [17] and makes the milk more acidic, as a result the
output phase angle decreases whereas, addition of water makes
the milk more alkaline and phase angle increases.
2.2. Phase detector circuit
The phase angle change with different adulterations of milk can be
tracked if the output signal from the sensor is converted into voltageby some signal conditioning circuit, so the second block of the
instrumentation system is a phase detector circuit [23]. The principle
of operation of the circuit is that output square pulse will be obtained
whose width depends upon the zero crossing of the voltage signal and
phase angle can be measured with respect to the zero crossing of a
reference voltage signal. Fig. 5 shows the phase detector circuit [23]
used for the sensing system. A low pass lter is added to this circuit to
get a dc voltage equivalent to the difference in the width of the pulse.
The milk samples are now tested by connecting the output of
the sensor with a phase detector circuit and it has been observed
that voltage also remains constant, in the frequency range of
5–12 kHz. The voltage obtained from signal conditioning block is
tabulated in Table 6 and plotted in Fig. 6 at frequency 7 kHz when
the sensor is dipped into milk and milk adulterated with tap water.
From Table 6, it is observed that as the % of adulteration increases;
voltage reading from phase detector circuit also increases.
Then milk is adulterated with different concentration of urea
and the voltage of the phase detector circuit is noted. It has been
observed that for a particular adulteration, circuit is giving con-
stant voltage in the frequency range 5–12 kHz. The voltage
obtained for consecutive three days from phase detector circuit
is tabulated in Table 7 and plotted in Fig. 7. From Table 7 it is found
that as the concentration of urea adulteration increases, voltagevalue decreases. Also phase angle and pH value decrease for the
same urea adulteration and are shown in Tables 7 and 4.
2.2.1. Sensitivity evaluation
The sensitivity of the sensor in phase detector circuit was calcu-
lated from the difference of pH between 10% tap water and pure milk
(Table 2) and it was found to be 0.08 only. Again from Table 6, it is
found that difference of voltage from phase detector circuit between
10% tap water and pure milk is 0.15 V. Hence, the sensitivity of the
phase detector circuit when FOE13 sensor dipped in tap water
adulterated milk is 1.875 V/pH. Similarly from Tables 4 and 7 it is
clear that the sensitivity of FOE13 sensor is 12.5 V/pH when dipped in
urea adulterated milk.
2.3. Display unit
In the nal prototype of the sensing system, a display circuit shown
in Fig. 8 has been used to display a different color LED array for
different types of milk adulteration. Three LEDs are used to indicate
three different voltage levels. The ORANGE LED turns on when the
input voltage is within the high voltage and low voltage limit. The RED
LED will glow when the input voltage is above the high voltage limit
and GREEN LED will glow when the input voltage is below the low
voltage limit. The combination of resistance values (R1 and R2) will set
the high and low limit of voltage values.
2.3.1. LED display for different percentages of tap water adulteration
The voltage obtained for pure milk is r4:28 V and milk adulter-ated with tap water lies between 4.28 V and 5.17 V as observed from
Table 8. Hence the high voltage limit and low voltage limit as shown in
Fig. 8 are 5.17 V and 4.28 V respectively. The proper resistance values of
the display circuit have been chosen so that the voltage of pure milk is
less than low voltage range, addition of 40% tap water is above the high
voltage range and milk adulterated with 10–40% tap water is in
between these two. The LED indication and the voltages obtained for
different types of adulteration are shown in Table 8. To keep the voltage
range within 5.17 V and 4.28 V of display circuit, the calculated R1 and
R2 values are 6.8 kΩ and 16.54 kΩ respectively. With these combina-tion of resistance values, the display unit (Fig. 8) is connected to the
output of the phase detector circuit. It is found that when the sensor
FOE13 is dipped in pure milk, GREEN LED turns on. When milk is
adulterated with 10–
40% tap water, ORANGE LED is glowing and whenmilk is adulterated with more than 40% tap water, RED LED turns on.
From Table 8, it can be said that the display circuit shown in
Fig. 8 is able to indicate different percentages of tap water
adulteration by using different colored LED.
2.3.2. LED display for different concentration of urea adulteration
Similar tests were performed with milk adulterated with urea.
From Table 9, it has been observed that the voltage obtained for pure
milk and milk adulterated with different concentration of urea lies
between 4.31 V and 3.09 V respectively. To keep the voltage range
within 4.31 V and 3.09 V of display circuit, the calculated R1 and R2
values are 16.32 kΩ and 28.68 kΩ respectively. With these resistancevalues connecting in display circuit and FOE13 sensor dipping in pure
milk, it is found that RED LED turns on. FOE13 sensor is next dipped in
Table 4
Phase angle in milk and milk adulterated with urea of FOE13 sensor at 7 kHz.
Day Test medium pH value Avg θ (deg)
1 Pure milk 6.82 24.96
Added 0.6 mg/ml urea 6.78 19.77
Added 0.8 mg/ml urea 6.75 19.08
Added 1.0 mg/ml urea 6.73 18.46
Added 1.2 mg/ml urea 6.69 17.88
2 Pure milk 6.85 25.29
Added 0.6 mg/ml urea 6.77 20.15
Added 0.8 mg/ml urea 6.75 19.50
Added 1.0 mg/ml urea 6.72 18.93
Added 1.2 mg/ml urea 6.68 18.33
3 Pure milk 6.83 26.10
Added 0.6 mg/ml urea 6.76 20.85
Added 0.8 mg/ml urea 6.73 20.34
Added 1.0 mg/ml urea 6.71 19.80
Added 1.2 mg/ml urea 6.67 19.23
Table 5
Slope, zero error and residual with urea adulteration.
Day Slope (m) Zero error (c ) R2
1st 5.974 24.33 0.935
2nd 5.858 24.65 0.932
3rd 5.769 25.41 0.919
S. Das et al. / ISA Transactions 56 (2015) 268 – 275 271
-
8/18/2019 1-s2.0-S0019057814002997-main
5/8
milk adulterated with 0.6 mg urea to 1.2 mg urea and ORANGE LED
glows. Again FOE13 sensor dipped in milk which is adulterated with
more than 1.2 mg urea and found GREEN LED turns on (Fig. 8). It is to
be noted that in case of pure milk RED LED turns on. But a switching
arrangement can be made so that GREEN LED turns on when FOE13
sensor is dipped in pure milk and RED LED turns on when dipped in
milk which is adulterated with more than 1.2 mg urea.
From Tables 8 and 9 it may be said that if we design two LED
circuits which can identify either tap water adulteration or urea
adulteration the two blocks together may be able to give an
indication. So using this LED circuit it may be possible to
differentiate among pure milk and milk adulterated with tap
water or urea.
Fig. 5. Phase detector circuit.
Table 6
Output voltage in milk and milk adulterated with tap water at 7 kHz.
Test medium Output voltage at
day 1 (V)
Output voltage at
day 2 (V)
Output voltage at
day 3 (V)
Pure milk 4 .28 4.31 4.43
Added 10% tap
water
4.43 5.00 5.15
Added 20% tap
water
4.66 5.23 5.22
Added 40% tap
water
5.17 5.69 5.73
0 10 20 30 404
4.5
5
5.5
6
Different % of tap water adulteration
V o l t a g e ( V )
1st day
2nd day
3rd day
Fig. 6. Output voltage versus % of tap water adulteration.
Table 7
Output voltage in milk and milk adulterated with urea at 7 kHz.
Test medium Output voltage at
day 1 (V)
Output voltage at
day 2 (V)
Output voltage at
day 3 (V)
Pure milk 4.31 4.43 4.29
Added
0.6 mg
urea
3.81 3.72 3.71
Added
0.8 mg
urea
3.55 3.45 3.49
Added
1.0 mg
urea
3.27 3.29 3.26
Added1.2 mg
urea
3.09 3.10 3.01
0 0.2 0.4 0.6 0.8 1 1.23
3.5
4
4.5
5
Different concentration of urea adulteration (mg)
V o l t a g e ( V )
1st day
2nd day
3rd day
Fig. 7. Output voltage versus concentration of urea adulteration.
S. Das et al. / ISA Transactions 56 (2015) 268 – 275272
-
8/18/2019 1-s2.0-S0019057814002997-main
6/8
3. Automatic detection system
A low cost automated sensing system is designed in which the
sensor detects the adulterated milk and a micro-controller based
circuit closes the valve installed at the outlet of the milk supply
line to prevent mixing. The block diagram is shown in Fig. 9. As acase study ‘the results with the synthetic milk (reconstructed after
adulterating with liquid-whey)' is presented in the paper.
3.1. Design of the automated sensing system
The rst block of Fig. 9 consists of the FOE sensor dipped inside
the test sample. The second block is the phase detector circuit. The
output of the sensor is used as an input to the phase detector
circuit as shown in Fig. 5, which converts the phase angle into its
corresponding voltage output. The next is the microcontroller
block. The output voltage from phase detector circuit is fed into
microcontroller. The voltage corresponding to the pure milk is
stored in the micro-controller and whenever a deviation occurs
with respect to this stored voltage an error signal is generated and
the driver circuit drives the control valve (solenoid valve). So,
whenever adulteration is detected, the control valve at the outlet
of the milk line shuts off the ow of milk to prevent mixing. The
display unit consists of LCD display and a LED array system. The
output voltage can be read from the LCD and a comparator based
LED driver system is used to glow the ‘RED’ or ‘GREEN’ LED to
indicate the status of the milk.
3.1.1. Controller block
The heart of the control block is an 8 bit microcontroller (ATMEGA 8)
as shown in Fig. 10. It has an inbuilt ADC which is connected to Port C.There are 3 ports in this microcontroller. Port C is used to collect the
analog signal from phase detector circuit. Port D gives the control signal
to the solenoid valve. Port B and Port D are used to interface LCD
display.
ATMEGA 8 microcontroller does the following functions:
Converts the output voltage from phase detector circuit to itscorresponding digital form.
It compares the output from the sensor with the referencesignal which corresponds to pure milk. A tolerance level of 5%
is allowed for the comparison. This is 0 :95 V ref oV ino1:05 V ref . Generates an error signal to activate the solenoid valve. When
the voltage exceeds the desired limits, sends that to the
control valve. Sends the corresponding voltage value to be displayed in the
LCD display. Sends the signal to glow the corresponding LED. If exceeds the desired limit after sending the error signal waits
for starting fresh.
The microcontroller is programmed accordingly. Output signal
from PD.4, either 0 or 1, is fed to the solenoid valve through a relay.
3.1.2. Working of solenoid valve
The output of ATMEGA 8 microcontroller is used to energize a
24 V relay via ampliers. These relays have been used for switch-
ing the solenoid valve. Separate signal conditioning has been
designed by using a bridge rectier circuit for conversion from
Fig. 8. Display circuit used to indicate different adulterants.
Table 8
LED indication for different types of tap water adulteration and their corresponding
voltages.
Test medium LED ON Voltage
Pure milk GREEN Less than or equal to 4.28 V
Added 10–40% tap water ORANGE Within 5.17 V and 4.28 V
Added Z40% tap water RED Greater than or equal to 5.17 V
Table 9
LED indication for different concentration of urea adulteration and their corre-
sponding voltages.
Test medium LED ON Voltage
Pure milk RED Greater than or equal to 4.31 V
Added 0.6–1.2 mg/ml urea ORANGE Within 4.31 V and 3.09 V
Added Z1:2 mg=ml urea GREEN Less than or equal to 3.09 V
S. Das et al. / ISA Transactions 56 (2015) 268 – 275 273
-
8/18/2019 1-s2.0-S0019057814002997-main
7/8
220 V supply. Normally solenoid valve remains open. The voltage
from the phase detector circuit is kept within the specied
allowable range 0:95 V ref oV ino1:05 V ref , to keep the valve in
open condition. The microcontroller output gives a low output i.e.,
0 to the solenoid circuitry and remaining same state of ow in
pipe, i.e., valve remains open and allows the milk ow. When the
voltage from phase detector circuit is out of the allowable range
0:95 V ref oV ino1:05 V ref , the output of the pin PD4 of microcon-
troller will change from 0 to 1. The valve will be actuated and will
be closed and will block the ow of adulterated milk.
3.2. Results
Table 10 shows the pH value of the test samples and the
corresponding phase angle and voltage obtained by the FOE13 sensor
at 7 kHz.
It is to be noted from Table 10 that after adding NaOH to the
whey adulterated milk though the pH value is brought to the same
value of the pure-milk, the sensor could identify between these
two test samples and show different phase angles. This is also
reected in the voltage output.
Fig. 9. Block diagram of the automatic sensing system.
Fig. 10. ATMEGA8 microcontroller conguration.
S. Das et al. / ISA Transactions 56 (2015) 268 – 275274
-
8/18/2019 1-s2.0-S0019057814002997-main
8/8
From Table 10, it is also observed that the voltage obtainedfrom pure milk is 4.38 V so the stored value in the microcontroller
i.e., V ref is 4.38 V. So the allowable range of voltage is 4.16–4.59 V.
From Table 10, the output voltage of the synthetic milk, V in, is
3.70 V, which is out of the limit, so the valve closes and shuts off
the ow of milk.
4. Conclusion
In this paper, a simple and inexpensive instrumentation system
has been reported to detect milk adulteration. The system is
capable to differentiate between pure milk and milk adulterated
with water and urea. For water adulteration, it has been noted that
the change is signicant between pure milk and milk adulterated
with 10% water, though for higher concentration of adulterationthe change is not that signicant. Further research work is
required in this direction to increase the sensitivity of the sensor
by different polymer coating. This is also true for urea adulteration
where 0.6 mg/ml shows maximum slope but decreases for higher
adulteration.
It is to be noted that phase detector circuit faithfully reproduces
the change in phase angle and gives output voltage. The sensor
output is noted at 7 kHz as this is in the middle of the frequency
range (5–12 kHz) where the opamp works properly and the sensor
behaves as a CPE.
A low cost automatic sensing system is also designed to detect
synthetic milk. The synthetic milk is reconstructed after adulter-
ating pure milk with ‘liquid-whey’. The main difference between
the pure milk and the synthetic milk is the presence of differentions which the sensor is capable to detect. The phase angle
remains constant over a band of frequency, so measurement can
be frequency independent, which is an added advantage for such
automatic sensing system.
Acknowledgments
The authors are thankful to UGC-DAE Consortium for Scientic
Research, Kolkata, India and also to PURSE project, JU. The authors
are also thankful to the Department of Science and Technology,
West Bengal, India, for their nancial support. The authors also
like to thank Prof. S.K. Gangopadhaya of Mohanpur Dairy Institute,
WB, for his help and suggestions.
References
[1] Syaie S, Tadeoa F, Villan M, Alonso AA. Learning control for batch thermalsterilization of canned foods. ISA Trans 2011;50:2–90.
[2] Holy R, Pozivil J. Batch control system for a pharmaceutical plant. ISA Trans2002;41:245–54.
[3] Careri M, Bianchi F, Corradini C. Recent advances in the application of massspectrometry in food-related analysis. J Chromatogr A 2002;970:3–64.
[4] Berna Amalia. Review: metal oxide sensors for electronic noses and their
application to food analysis. Sensors 2010;10:3882–910.[5] Galloway JA. Great fare of London. Lancet 2000;355:323–4.[6] Gem H. The adulteration of milk. Lancet 1857;69:156.[7] Mabrook MF, Petty MC. Application of electrical admittance measurements to
the quality control of milk. Sens Actuators B: Chem 2002;84:136–41.[8] Kasemsumran S, Thanapase W, Kiatsoonthon A. Feasibility of near-infrared
spectroscopy to detect and to quantify adulterants in cow milk. Anal Sci
2007;23(7):907–10.[9] Das S, Sivaramakrishna M, Biswas K, Goswami B. Performance study of a
‘constant phase angle based’ impedance sensor to detect milk adulteration.
Sens actuators A : Phys 2011;167:273–8.[10] Mabrook MF, Petty MC. A novel technique for the detection of added water to
full fat milk using single frequency admittance measurements. Sens Actuators
B: Chem 2003;96:215–8.[11] Alonso VLM, Minor MA. Design and construction of a system for measuring
the concentration of water in milk. In: Conference on electronics, robotics andautomotive mechanics, 2006. p. 47–51.
[12] Buttel B, Fuchs M, Holz B. Freezing point osmometry of milk to determine the
additional water content–an issue in general quality control and German food
regulation. Chem Central J 2008;2:6. http://dx.doi.org/10.1186/1752-153X-2-6.[13] Recio I, Garcia-Risco MR, Lopez-Fandino R, Olano A, Ramos M. Detection of
rennet whey solids in UHT milk by capillary electrophoresis. Int Dairy
J 2000;10:333–8.[14] Das S, Sivaramakrishna M, Dey M, Goswami B, Biswas K. Performance of a
constant phase element (CPE) sensor to detect adulteration in cow-milk with
whey. In: Proceedings of the IEEE sensors 2009, New Zealand, 2009. p. 745–50.[15] Bremer MGEG, Kemmers-Voncken AEM, Boers EAM, Haasnoot RFW. Enzyme-
linked immunosorbent assay for the detection of bovine rennet whey powder
in milk powder and buttermilk powder. Int Dairy J 2008;18:294–302.[16] Wolfschoon-Pombo AF, Furtado MAM. Detection of adulteration of pas-
teurised milk with whey by determination of the casein-bound phosphorus
and protein nitrogen content. Z Lebensm Unters Forsch 1989;188:16–21.[17] Trivedi UB, Lakshminarayana D, Kothari IL, Patel NG, Kapse HN, Makhija KK,
et al. Potentiometric biosensor for urea determination in milk. Sens ActuatorsB: Chem 2009;140:260–6.
[18] Renny EF, Daniel DK, Krastanov AI, Zachariah CA, Elizabeth R. Enzyme based
sensor for detection of urea in milk. Biotechnol Biotechnol Equip
2005;19:198–201.[19] Singh M, Verma N, Garg AK, Redhu N. Urea biosensors. Sens Actuators B: Chem
2008;134:345–51.[20] Jha SN, Matsuoka T. Detection of adulterants in milk using near infrared
spectroscopy. J Food Sci Technol - Mysore 2004;41(3):313–6.[21] Sharma R, Rajput YS, Kaur S, Tomar SK. A method for estimation of urea using
ammonia electrode and its applicability to milk samples. J Dairy Res
2008;75:466–70.[22] Biswas K, Sen S, Dutta PK. A constant phase element sensor for monitoring
microbial growth. Sens Actuators B: Chem 2006;119:186–91.[23] Wobsschall D. Circuit design for electronic instrumentation. USA: McGraw-
Hill; 1979.[24] Sivaramakrishna M, Das S, Biswas K, Goswami B. Fabrication of a fractional
order capacitor with desired specication: a study on process identication
and characterization. IEEE Trans Electron Devices 2011;58:4067–
73.[25] Das S, Sivaramakrishna M, Goswami B, Biswas K. Design of a low-cost
‘constant phase angle’ based sensing system to detect natural milk and
‘synthetic-milk’ reconstructed from ‘liquid-whey’. In: The fth international
conference on sensor technologies and applications, Nice, France, 2011.[26] Paradkar MM, Singhal SR, Kulkarni RP. An approach to the detection of
synthetic milk in dairy milk: detection of urea. Int J Dairy Technol
2007;53:87–91.[27] Bansal P, Bansal N. Synth Milk Genes Curr Sci 1997;73.
Table 10
Performance of the sensing system in the test medium at 7 kHz frequency.
Test samples pH value Phase angle Voltage
Pure milk 6.87 25.08 4.38
Synt heti c mi lk (r econ structed a fter a ddi ng 3 0% whey a nd N aO H) 6 .8 7 20.17 3.70
S. Das et al. / ISA Transactions 56 (2015) 268 – 275 275
http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref2http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref2http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref2http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref2http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref2http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref3http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref3http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref3http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref3http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref3http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref4http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref4http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref4http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref4http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref4http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref5http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref5http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref5http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref5http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref6http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref6http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref7http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref7http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref7http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref7http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref7http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref8http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref8http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref8http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref8http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref8http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref8http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref10http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref10http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref10http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref10http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref10http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref10http://localhost/var/www/apps/conversion/tmp/scratch_4/dx.doi.org/10.1186/1752-153X-2-6http://localhost/var/www/apps/conversion/tmp/scratch_4/dx.doi.org/10.1186/1752-153X-2-6http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref15http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref15http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref15http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref15http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref15http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref15http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref16http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref16http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref16http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref16http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref16http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref16http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref17http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref17http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref17http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref17http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref17http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref17http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref18http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref18http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref18http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref18http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref18http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref18http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref19http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref19http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref19http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref19http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref19http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref20http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref20http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref20http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref20http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref20http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref21http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref21http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref21http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref21http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref21http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref21http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref22http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref22http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref22http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref22http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref22http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref23http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref23http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref23http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref26http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref26http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref26http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref26http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref26http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref26http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref27http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref27http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref27http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref26http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref26http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref26http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref24http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref23http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref23http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref22http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref22http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref21http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref21http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref21http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref20http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref20http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref19http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref19http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref18http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref18http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref18http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref17http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref17http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref17http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref16http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref16http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref16http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref15http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref15http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref15http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref13http://localhost/var/www/apps/conversion/tmp/scratch_4/dx.doi.org/10.1186/1752-153X-2-6http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref10http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref10http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref10http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref9http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref8http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref8http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref8http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref7http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref7http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref6http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref5http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref4http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref4http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref3http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref3http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref2http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref2http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1http://refhub.elsevier.com/S0019-0578(14)00299-7/sbref1