1LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Calibration Testing
Patrick Sutton, withMichael Landry, Gabriela Gonzalez, & Brian
O'Reilly
2LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Idea
We have two ~independent means of calibration, which can be compared for consistency.
`Official' calibration: » Measure H, C during dedicated calibration run (hours).
Eg. Done Feb 14-15.
» Fit measured H data to model, derive H, C, Cb for that time.
» Dynamic: Rescale H, C over time using alpha measured from calibration lines; update Cb.
3LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Idea (cont'd)
AutoCalibrator:
» Inject swept sine into ETMX/Y and measure transfer function to AS_Q (minutes).
» Empirical (no use of model) - independent of `official' calibrations up to treatment of actuation function A(f).
» Gives static snapshot of Cb.
Proposal: Use frequent AutoCalibrator sweeps as check on Official calibration.
This talk: show examples for S2
4LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Example: H1
5LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Ratio: H1 Official/AutoCal
6LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
More H1 Sweeps
7LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Example: H2
8LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Ratio: H2 Official/AutoCal
9LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
More H2 Sweeps
10LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Lonely L1 Example
11LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Lonely L1 Example
12LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18
Summary: Huh?
See very similar discrepancies between Official and AutoCal calibrations in all three IFOs.
Discrepancy increases with frequency. At 2kHz:» Magnitude: 40%
» Phase: 2 radians
Source unknown; will require further investigation.