Calibration Testing

Patrick Sutton, withMichael Landry, Gabriela Gonzalez, & Brian

O'Reilly

Page 2: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

2LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

Idea

We have two ~independent means of calibration, which can be compared for consistency.

`Official' calibration: » Measure H, C during dedicated calibration run (hours).

Eg. Done Feb 14-15.

» Fit measured H data to model, derive H, C, Cb for that time.

» Dynamic: Rescale H, C over time using alpha measured from calibration lines; update Cb.

Page 3: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

3LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

Idea (cont'd)

AutoCalibrator:

» Inject swept sine into ETMX/Y and measure transfer function to AS_Q (minutes).

» Empirical (no use of model) - independent of `official' calibrations up to treatment of actuation function A(f).

» Gives static snapshot of Cb.

Proposal: Use frequent AutoCalibrator sweeps as check on Official calibration.

This talk: show examples for S2

Page 4: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

4LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

Example: H1

Page 5: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

5LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

Ratio: H1 Official/AutoCal

Page 6: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

6LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

More H1 Sweeps

Page 7: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

7LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

Example: H2

Page 8: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

8LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

Ratio: H2 Official/AutoCal

Page 9: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

9LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

More H2 Sweeps

Page 10: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

10LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

Lonely L1 Example

Page 11: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

11LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

Lonely L1 Example

Page 12: 1 LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18 Calibration Testing Patrick Sutton, with Michael Landry, Gabriela Gonzalez,

12LIGO-G030093-00-Z Penn State / LIGO Scientific Collaboration 2003 March 18

Summary: Huh?

See very similar discrepancies between Official and AutoCal calibrations in all three IFOs.

Discrepancy increases with frequency. At 2kHz:» Magnitude: 40%

» Phase: 2 radians

Source unknown; will require further investigation.