Lab Equipment
![Page 2: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/2.jpg)
2
Topic Slides
DC Power Supply 3-4
Digital Multimeter 5-8
Function Generator 9-12
Scope – basic controls 13-20
Scope – cursors 21-24
Scope – triggering 25-32
Scope – exploring the waveform 33-34
Scope – trace rotation and probe compensation 35-40
Scope – rise/fall time measurements 41-47
Scope – computer communication problems 48
Contents
![Page 3: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/3.jpg)
3
DC Power Supply
All three supplies share a common terminal, which is NOT connected to the power supply “ground”. This terminal should normally be connected to your circuit “ground”.
The negative 0 to 20 volt supply is proportional (by an adjustable ratio of up to -1) to the positive supply
Vneg = (0 to –1) times Vpos
Don’t use; connects to chassis and power line gndDC = 0 to Vpos
DC = 0 to 6 DC = 0 to 20
+
++
Vpos
![Page 4: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/4.jpg)
4
DC Power Supply
DC = 0 to Vpos
DC = 0 to 6 DC = 0 to 20
+
++
CBA D
Since the COM terminal is isolated from the power supply ground, you are not limited to using point B for your circuit ground. If you use point A, C, or D as your circuit ground you can use series combinations of the three internal voltage sources to obtain voltages higher than +20V and lower than –20V.
Vpos
![Page 5: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/5.jpg)
5
Digital Multimeter (DMM)
Power switchDC voltage
AC voltage
DC current
AC currentResistance
AC = RMS value of AC part of waveform only
DC = value of DC part of waveform only
To measure TRUE RMS value of the complete waveform, press both DC and AC
Range is set automatically
![Page 6: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/6.jpg)
6
Diode test (passes small current through diode and measures the resulting voltage drop)
Connect the RED lead to the diode anode and the black lead to the cathode.
![Page 7: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/7.jpg)
7
Use the RATE switch to set the speed of measurement and the resolution of the displayed values:
F = fast speed = low resolution
M = medium speed = medium resolution
S = slow speed = high resolution
![Page 8: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/8.jpg)
8
10 AMP current input has low internal resistance and introduces the lowest error; use this input whenever possible
100 milliamp current input has considerable internal resistance and may introduce considerable error; use the 10 A input whenever possible.
For current measurements, move the RED lead to a current input
![Page 9: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/9.jpg)
9
Function Generator
outputSelect waveform
Set frequency, amplitude, and DC offset
![Page 10: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/10.jpg)
10
To enter values, press “Enter Number”, then the number keys, then “Enter”
Observe decimal point; ignore the commas
![Page 11: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/11.jpg)
11
When entering values, use these keys to define the
units. Notice that voltage values can be defined as
either RMS or peak-to-peak.
![Page 12: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/12.jpg)
12
If the amplitude of the waveform measured by the scope or digital multimeter does not agree with the amplitude shown on the function generator display, reset the output termination to “HIGH Z”
“SYS Menu”>
“OUT TERM”
EnterShift >
“50 OHM ” “HIGH Z ”
> >
>
>
Enter
![Page 13: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/13.jpg)
13
OscilloscopeAlways adjust the scope to obtain as large of a waveform as possible (to maximize accuracy)
Adjust the FOCUS and INTENSITY to obtain a clear, readable display.
When not actively observing waveforms, turn down the intensity to extend the life of the scope screen
Function buttons
![Page 14: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/14.jpg)
14The AUTO SET button helps the scope find the waveform, but it also sets the waveform signal coupling to AC (undesirable)
After using the AUTO SET button, reset the waveform signal coupling to DC using the AC/DC buttons
General Rule: ALWAYS use DC coupling on the vertical channels so that you can observe the complete (AC & DC) waveform
![Page 15: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/15.jpg)
15
The VAR controls must be in the CAL position for the display values to be valid
![Page 16: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/16.jpg)
16
Use the X POS control to start the scope trace at the left edge of the grid
![Page 17: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/17.jpg)
17
Set the zero volt positions for ch A and ch B to known positions on the grid, usually the center horizontal line, by first selecting GND and then adjusting the Y POS controls
![Page 18: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/18.jpg)
18
Zero voltage reference line set to known point (usually center line, but may anywhere you desire)
Once positioned, be careful not to change it.
Trace starts at first vertical grid line
Example of Setting Zero Volt Reference
![Page 19: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/19.jpg)
19
vertical and horizontal values per division (cm) are shown on the display
Use these to set the vertical (voltage) values
per division
Use this control to set the horizontal (time) value per division
![Page 20: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/20.jpg)
20
DIGITAL MEMORY mode must be selected to be able to use cursors and transfer waveforms to the computer
![Page 21: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/21.jpg)
21
Use function buttons to control cursors
Scope Cursors
![Page 22: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/22.jpg)
22
Cursor Positioning Buttons
Controls lower cursor
Controls upper cursor
Controls left
cursor
Controls right
cursor
RETURN key takes you back to the previous higher level
![Page 23: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/23.jpg)
23
Setting the Cursor Mode and Reference Channel
Use to turn cursors on/off and to select what they display (vertical measurement = voltage or ratio;
horizontal measurement = time, phase angle, or ratio)
Voltage cursor will use either channel A or B “voltage per division” setting, as selected.
RETURN key takes you back to the previous higher level
cursor reference channel
Reference channel selection available only when both channels A and B are displayed
![Page 24: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/24.jpg)
24
Value of differences between horizontal
(voltage) cursors
Value of difference between vertical (time) cursors
Frequency computed from inverse of time interval (valid only if interval is one cycle)
![Page 25: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/25.jpg)
25
Scope Triggering
The waveform on the scope display is drawn from left to right (repetitively), beginning at a point when one of the signals being displayed meets a specified criteria. This criteria is defined by the settings of the horizontal (time base) trigger controls:
Mode - the standard mode is AUTO
Source - which signal is being used to trigger the scope
Coupling - normally P-P (AC), but sometimes DC gives more control
Slope - the scope trace begins on the upward or downward slope of the wave
Level - the voltage level of the wave at which the trace begins
![Page 26: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/26.jpg)
26
Sets the trigger mode (normally set to AUTO)Displays the trigger mode
Displays trigger slope Sets the trigger slope
Displays trigger source and coupling
Sets trigger coupling
Sets trigger source
Adjusts the trigger level
Horizontal (time base) trigger controls
![Page 27: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/27.jpg)
27
To capture a non-repetitive waveform, set the desired trigger conditions, select the SINGLE TRACE trigger mode, then press RESET to “arm” the scope.
After the scope has been “armed”, a single trace will be drawn when the signal level matches the trigger conditions (slope and level) you have selected. The trace will remain displayed until you press RESET or transfer the waveform to the computer.
Single Trace Operation
![Page 28: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/28.jpg)
28
Trigger level set to zero volts and a positive (upward) slope
Example Waveform
![Page 29: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/29.jpg)
29
Trigger level set to +1 volt, with a positive slope
Zero volt reference line
Example Waveform
![Page 30: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/30.jpg)
30
Trigger level set to –1 volts, with a
positive slope
Zero volt reference line
Example Waveform
![Page 31: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/31.jpg)
31
Trigger level set to zero volts, with a
negative (downward) slope
Example Waveform
![Page 32: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/32.jpg)
32
The trigger delay can be used to shift where the trigger point appears on the screen; this allows you to observe the waveform up to 10 divisions prior to the trigger point on repetitive waveforms.
Delayed Trigger
![Page 33: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/33.jpg)
33
Use the X-magnification button to zoom in on part of the waveform; the magnification factor is changed each time you press the button.
Waveform Zoom
![Page 34: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/34.jpg)
34
Use the Display Part rocker switch to move the display window across the waveform image stored in the scope. (The scope stores two screen widths of data at X1 magnification)
Move the display window
![Page 35: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/35.jpg)
35
Adjust to make trace parallel
with horizontal grid
Trace Rotation
![Page 36: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/36.jpg)
36
Scope Probe Compensation Adjustment
Connect probe to calibration
signal
Make sure channel A and B signal coupling are set to DC
![Page 37: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/37.jpg)
37
Scope Probe Compensation Adjustment
Each probe has an adjustment screw
that changes the compensating
capacitor
![Page 38: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/38.jpg)
38
Scope Probe Compensation Adjustment
Adjustment capacitor value needs to be increased until waveform is rectangular
![Page 39: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/39.jpg)
39
Scope Probe Compensation Adjustment
Adjustment capacitor value needs to be decreased until waveform is rectangular
![Page 40: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/40.jpg)
40
Scope Probe Compensation Adjustment
Rectangular waveform with correct probe adjustment
![Page 41: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/41.jpg)
41
0%10%
90%100%
For rise or fall time measurements, use the 0%, 10%, 90%, and 100% lines on the scope grid
Making Rise and Fall Time Measurements on Pulse Waveforms
![Page 42: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/42.jpg)
42
Using the vertical scale control to set your waveform amplitude so that it is greater than the range of 0% to 100%
Then use the var and the Ypos controls to line up your waveform with the 0% and 100% grid lines.
![Page 43: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/43.jpg)
43
100%90%
10%0%
The adjusted waveform display
![Page 44: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/44.jpg)
44
Expand the waveform horizontally using the time base (TB) control
Set the trigger slope to positive (upward) for rise
time measurements
Set trigger coupling to
DC
Adjust trigger level to the 10% point or below
For rise time measurements
![Page 45: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/45.jpg)
45
Adjust the vertical cursors so that they pass the 10% and 90% points on the waveform, and read off the rise time.
10%
90%
![Page 46: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/46.jpg)
46
Expand the waveform horizontally using the time base (TB) control
Set the trigger slope to negative (downward) for rise time measurements
Set trigger coupling to
DC
Adjust trigger level to the 90% point or above
For fall time measurements
![Page 47: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/47.jpg)
47
Adjust the vertical cursors so that they pass the 10% and 90% points on the waveform, and read off the fall time.
10%
90%
![Page 48: 1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.](https://reader031.fdocuments.us/reader031/viewer/2022032704/56649d625503460f94a43cb9/html5/thumbnails/48.jpg)
48
Use OPTION function button to check or correct communications settings
If scope fails to communicate with the computer:
OPTION
IEEE
IEEE-ADDR: 8 MODE: talker/listener
If you have to correct the communications settings, cycle the scope power to reset the communications interface.