Post on 16-Jan-2016
THE MEASUREMENT OF pH
lseidman@matcmadison.edu
I GENERAL PRINCIPLES
Convenient way to express hydrogen ion concentration, or acidity
pH = - log[ H+]
Where concentration is expressed in moles/liter
lseidman@matcmadison.edu
pH SCALE
Is logarithmic Example, pH 5.0 solution has ten times more
hydrogen ions than pH 6.0 solution
As hydrogen ion concentration, or acidity, increases, the pH value decreases
lseidman@matcmadison.edu
CONSIDER pH OF PURE WATER
The [H+] of pure water is 1 X 10-7 mole/L What is pH?
lseidman@matcmadison.edu
pH = - log[ H+]
The log of 1 X 10-7 is -7
The negative log of 10-7 is -(-7) = 7
The pH of pure water is 7, which we call neutral
lseidman@matcmadison.edu
QUESTIONS
a. What is the pH of a solution with an H+ ion concentration of 10-4 mole/L?
b. What is the pH of solution with an H+ ion concentration of 5.0 X 10-6 mole/L?
lseidman@matcmadison.edu
ANSWER
a. pH = - log [H+] = - log 10-4 = -(-4) = 4
b. pH = - log [H+] = - log 5.0 X 10-6 =
-(-5.3) = 5.3
lseidman@matcmadison.edu
QUESTION
What is the concentration of H+ ions in a solution with a pH of 9.0?
lseidman@matcmadison.edu
ANSWER
pH = - log [H+]
9.0 = - log [H+]
- 9.0 = log [H+]
antilog (-9.0) = 1 X 10-9 mole/L
lseidman@matcmadison.edu
DESIGN OF pH METER/ELECTRODE MEASURING SYSTEMS
pH meter systems measure hydrogen ion concentration; electrochemical
pH S
cale
0
7
14
NEUTRAL
STRONGLY ACIDIC
STRONGLY BASIC
Drain cleaner
Milk of magnesia
Tap water
Mild shampoo
Orange juiceVinegar
Battery acid
lseidman@matcmadison.edu
ANIMATIONS
http://www.umd.umich.edu/casl/natsci/slc/slconline/PHM/theory.html
lseidman@matcmadison.edu
Design of pH meter system
lseidman@matcmadison.edu
II CHARACTERISTICS AND TYPES OF ELECTRODES AND pH METERS
Many types of electrodes and meters Meters:
Conventional Portable Solid state Microprocessor controlled
lseidman@matcmadison.edu
ELECTRODES
Heart of the system Can buy separate measuring and reference
electrodes or Combination; both electrodes are in one
housing Combination almost always found in biology
labs But not necessarily right
lseidman@matcmadison.edu
SEPARATE ELECTRODES
Difficult samples Require different type of glass electrode Different type of junction
Junctions have different flow rates Tendency to clog
lseidman@matcmadison.edu
REFERENCE ELECTRODES
Two major types: Ag/AgCl and calomel (mercury)
General purpose usually silver/silver chloride
lseidman@matcmadison.edu
Calomel electrodes are said to be more accurate for measuring pH of Tris buffers But calomel electrodes cannot be used in
solutions above 80o C and contain mercury
Both types are available in combination electrodes or separately
lseidman@matcmadison.edu
III PROPER OPERATION; AVOIDING ERROR Consider electrodes Consider sample Consider meter operation
lseidman@matcmadison.edu
ELECTRODES
Maintain electrodes properly Filling Solution. Filling solution, often saturated
KCl or AgCl/KCl Refillable electrodes are periodically filled to nearly the
top Filling hole must be open to pH and closed for storage Gel-filled electrodes contain gelled filling solution, never
refilled
lseidman@matcmadison.edu
ELECTRODES
Storage. Consult the manufacturer's instructions Don’t store electrodes in distilled water New combination electrodes often stored dry
Conditioned before use by soaking at least 8 hours in pH 7 buffer
lseidman@matcmadison.edu
SAMPLE
Important part of the system
Must be homogenous
May be chemical changes in sample
May be temperature effects
lseidman@matcmadison.edu
DIFFICULT SAMPLES
1. Non-aqueous solvents
2. High purity water. High purity water does not readily conduct current and it absorbs CO2 from the atmosphere
3. High salt samples. Sample ions compete with the reference filling solution ions
4. Sample-electrode compatibility. Tris buffer, sulfides, proteins, Br-, and I- , can complex with silver in Ag/AgCl electrodes leading to a clogged junction. A calomel electrode is often recommended
5. Slurries, sludges, viscous and colloidal samples. Use a fast flowing junction. Keep measuring electrode bulb clean
lseidman@matcmadison.edu
OPERATION; CALIBRATION
User calibrates frequently Two buffers of known pH; forms calibration
line. Try this in lab Older pH meter first calibration buffer should
be pH 7.00 For acidic samples, second is pH 4.00 For basic samples, second is pH 10.00
lseidman@matcmadison.edu
Newer, microprocessor-controlled meters: Any two standard buffers that bracket pH of
samples may be used Some microprocessor-controlled meters allow use
of more than two standard buffers
lseidman@matcmadison.edu
CALIBRATION BUFFERS
Accuracy of pH meter depends on calibration buffers
Some buffers react with CO2 from air Keep buffer containers closed Throw away buffer after use pH 10.0 buffer is particularly sensitive to CO2
lseidman@matcmadison.edu
BUFFERS
For best accuracy, do not use buffers after their expiration date
Avoid contamination
pH of a buffer will change as its temperature changes
lseidman@matcmadison.edu
TEMPERATURE AND pH
Temperature has two important effects: Measuring electrode's response to pH is affected
by the temperature
pH of solution that is being measured change as its temperature changes
lseidman@matcmadison.edu
COMPENSATION FOR TEMPERATURE Meter needs to "know" temperature of solution Can measure solution temperature with
thermometer and "tell" pH meter Alternatively, use ATC (automatic temperature
compensating) probes that is connected to the pH meter
Compact devices may have a temperature probe built into the electrode housing
lseidman@matcmadison.edu
TEMPERATURE
The pH of some solutions changes with temperature
Usually measure pH of samples at the temperature at which they will be used
lseidman@matcmadison.edu
A CONVENTIONAL METHOD FOR MEASURING pH
1. Warm-up meter
2. Open filling hole; check filling solution level
3. If meter has a "standby” mode, use it when the electrodes are not immersed in sample
lseidman@matcmadison.edu
4. Calibrate each day or before use:
a. Adjust temperature setting.
b. Rinse electrodes and blot dry. Don’t wipe.
c. Immerse electrodes in pH 7.00 calibration buffer. Junction must be immersed and level of sample must be below filling solution level. Allow reading to stabilize.
lseidman@matcmadison.edu
d. Adjust meter to read 7.00.
e. Remove electrodes, rinse, blot dry.
f. Place electrodes in second standardization buffer. Adjust meter to pH. Remove, rinse, and blot.
g. Older meters: Recheck pH 7.00 buffer and readjust as necessary. Recheck second buffer and readjust as necessary.
h. Readjust as needed up to three times. If readings are not within 0.05 pH units of what they should be, electrode probably needs cleaning.
lseidman@matcmadison.edu
QUALITY CONTROL
5. Optional: Quality control checks.Try this in lab.a. Linearity Check. Take reading of a third calibration buffer. For example, if you calibrated with pH 7.00 and 10.00 buffers, check pH 4.00 buffer.
b. Allow the reading to stabilize and record value. Do not readjust meter. If reading is outside proper range, service electrodes.
lseidman@matcmadison.edu
QUALITY CONTROL
c. Check pH of a control buffer whose pH is known and that has a pH close to the pH of the sample.
Common to set maximum allowable error of control buffer to + 0.10 pH units. Do not adjust meter to pH of control buffer.
If pH of control buffer is not within required tolerance, service electrodes.
lseidman@matcmadison.edu
SAMPLE
6. Set meter to temperature of sample or use ATC probe.
lseidman@matcmadison.edu
7. Place electrodes in sample; allow reading to stabilize.
a. Wait too long, pH of some samples will change due to air, chemical reactions, etc.
b. Difficult solutions may require longer to stabilize.
c. Many new pH meters have an “autoread” feature.
lseidman@matcmadison.edu
9. Record all relevant information.
10. Remove electrodes from sample, rinse and blot, store properly with filling hole closed (refillable electrodes only).
lseidman@matcmadison.edu
IV TROUBLE-SHOOTING
First step is to know you have trouble. Symptoms of pH system problems include:
Reading drifts and won't stabilize. Reading fluctuates. Meter cannot be adjusted to both calibration
buffers. pH value for a buffer or sample seems wrong.
lseidman@matcmadison.edu
TROUBLE-SHOOTING TIPS
Look for and correct simple (embarrassing) mistakes: Electrode measuring bulb and junction are not immersed in
sample. Meter is not turned on or plugged in, or the electrode cables
are not connected to meter. Reference electrode is not filled with electrolyte. Reference electrode filling hole is closed. Sample is not well-stirred. Calibration buffers are not good. Electrode is cracked or broken.
lseidman@matcmadison.edu
NEXT
What is most likely cause of problems? Problems can arise in:
The reference electrode The measuring electrode The calibration buffers The sample The meter
lseidman@matcmadison.edu
REFERENCE ELECTRODE
Reference electrode junction most common source of problems.
Occluded junction causes long stabilization time; reading drifts slowly towards correct pH.
If junction is completely plugged, reading may never stabilize. This is also caused by broken electrodes and by some
problems within meter.
lseidman@matcmadison.edu
Slow equilibration also caused by changes in sample temperature, reactions in sample, or sample-electrode incompatibility.
lseidman@matcmadison.edu
Poor calibration buffers will cause inaccuracy. This may be detected if solution of known pH is checked.
If sample is not homogenous, or if its temperature is
unstable, then pH readings will fluctuate or drift.
Difficult samples may be slow to equilibrate or give incorrect readings.
lseidman@matcmadison.edu
Complete lack of response likely caused by problems with meter, but meter is least likely component to cause problems.
To check which component is at fault substitute in a new reference or combination electrode.