Tackling the Techniques Tandem Mass Spectrometry

Post on 26-May-2022

4 views 0 download

Transcript of Tackling the Techniques Tandem Mass Spectrometry

Tackling the Techniques Tandem Mass Spectrometry

Marie Appleton

Paediatric Metabolic Section

Royal Victoria Infirmary

Newcastle-upon-Tyne

Introduction

Background and history of MS/MS

How MS/MS works

Common errors and troubleshooting

Examples

Mass Spectrometry - History

“I feel sure that there are many problems in Chemistry which could be solved with far greater ease by this than

by any other method. The method is surprisingly sensitive – more so even than that of spectrum analysis – requires an infinitesimal amount of material and does

not require this to be specially purified”…

(Sir J.J Thomson 1921)

Tandem Mass Spectrometry - History

Most significant advance in metabolic disease/NBS screening

Exponential increase in use to biosciences during 1980s

• Development of ionisation methods (ESI, MALDI)

First used in clinical setting in 1982

Inherited metabolic diseases in early 1990s

Short analysis time (2-3 mins)

Broad spectrum – 1 test/ many disorders

Basic Principles

Sample ionisation

(Parent/Precursor ions)

Mass Filtration (m/z)

Fragmentation into smaller ions

(Daughter/Product ions)

Mass Filtration (m/z)

Detection (Mass spectrum)

Types of Mass Spectrometers

Mass Spectrometer

Sample

Solid

Liquid

Gas

Ionisation

ESI, APCI, CI, EI, FAB, MALDI

Mass Analyser

TOF

Quadrupole

Ion trap

FT-ICR

Detector

Electron multiplier

Photo multiplier

Faraday cup

ESI triple quadrupole tandem mass spectrometer LC system

Acetonitrile/methanol/water

Autosampler

Mass spectrometer

ESI triple quadrupole tandem mass spectrometry

Argon gas

(CID)

sample

Mass spectra MS1 MS2

Vacuum

Ion Source – Electrospray Ionisation

Quadrupoles – MS1

Collision-induced Dissociation

+ + +

Ar

Quadrupoles – MS2

Collison cell

Detection – Mass spectra

Molecular ion

Base peak ion

ESI +ve triple quadrupole mass spectrometer

Modes of MS/MS operation

Q1 Q2 Q3

A, B Scan

A-n B-n Scan

A→(A-n) +n B→(B-n) + n

A Select

A A1 A2 A3

A1 A2 A3 Scan

A, B Scan

A→(A-x) + x B→(B-x) + x

X Select

A Select

A1 select

A A1

A2

A3

Product (daughter) ion scan

Precursor (parent) ion scan

Neutral loss scan

Selected ion monitoring (SIM)

ESI +ve triple quadrupole mass spectrometer

Modes of MS/MS operation

Q1 Q2 Q3

A, B Scan

A-n B-n Scan

A→(A-n) +n B→(B-n) + n

A Select

A A1 A2 A3

A1 A2 A3 Scan

A, B Scan

A→(A-x) + x B→(B-x) + x

X Select

A Select

A1 select

A A1

A2

A3

Product (daughter) ion scan

Precursor (parent) ion scan

Neutral loss scan

Selected ion monitoring (SIM)

Acylcarnitine m/z 85 α-Amino acids NL m/z 46 (102 deriv) NBS - MRM

Precursor of m/z 85 scan C:\Xcalibur\...\2015\20150416std2 16/04/2015 17:14:13 CN Std - 020415

RT: 0.00 - 2.98

0.0 0.5 1.0 1.5 2.0 2.5

Time (min)

0

10

20

30

40

50

60

70

80

90

100

Re

lative

Ab

un

da

nce

0.260.36

1.22 1.971.62 1.82 2.23 2.33 2.58 2.88

NL:1.31E9

TIC MS 20150416std2

20150416std3 #6-15 RT: 0.26-0.59 AV: 5 SM: 7B NL: 3.52E6F: + p ESI Full pr 85.000 [200.000-550.000]

250 300 350 400 450 500 550

m/z

0

10

20

30

40

50

60

70

80

90

100

Re

lative

Ab

un

da

nce

260.13

263.14

221.14

456.34

277.21 482.52353.30374.30

288.20 318.30 428.41

Papers

Common Errors & Troubleshooting

Sample Preparation

• Appropriate extraction, internal standard and derivatisation performed

• Eluent compatible with mobile phase

• Good laboratory practice - accuracy

UP TO YOU!

LC Pressure System

RT: 0.00 - 1.70 SM: 7G

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

20

40

60

80

100

0

20

40

60

80

100

0

20

40

60

80

100

Re

lative

Ab

un

da

nce

0

20

40

60

80

100

0

20

40

60

80

1000.43

1.11 1.31 1.48

1.20

0.15

0.86 0.970.49 0.71

0.40

1.08 1.311.19 1.53

1.00

1.620.710.18

0.40

0.07

NL: 3.76E6

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkA140513

NL: 7.49E5

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkB140513

NL: 4.35E6

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkC140513

NL: 3.83E6

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkD140513

NL: 4.53E6

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkE140513

RT: 0.00 - 1.70 SM: 7G

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

20

40

60

80

100

0

20

40

60

80

100

0

20

40

60

80

100

Re

lative

Ab

un

da

nce

0

20

40

60

80

100

0

20

40

60

80

1000.43

1.11 1.31 1.48

1.20

0.15

0.86 0.970.49 0.71

0.40

1.08 1.311.19 1.53

1.00

1.620.710.18

0.40

0.07

NL: 3.76E6

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkA140513

NL: 7.49E5

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkB140513

NL: 4.35E6

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkC140513

NL: 3.83E6

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkD140513

NL: 4.53E6

TIC F: + p sid=-10.00 SRM ms2 166.00@-15.00 [ 119.50-120.50] MS NSSsampchkE140513

Autosampler

• Common symptom is Low intensity

• Syringe

• Blockages - sample port/stata valve

• Leaks – rotor seal

• Wrong sample injection – user error!

Contamination

• Highly sensitive instrument

• Contamination may results from

• Sample preparation – other reagents and handcreams

• Bad laboratory practice

• Mobile phase – microbe growth in water

• Dirt build up in source

• Improper cleaning

Mass spectrometer • Physical

• Clean cone shield and ion transfer tube – reduced signal intensity

• Probe correct position from the cone – proper nebulization

Mass spectrometer • Physical

• Acquisition Parameters

• ESI Probe/source settings – temperature and gas flow optimised per compound (polar

mobile phases need high temperatures)

• Collision energy too high or too low –

no product ions formed

• Incorrect transitions or scan mode

• Incorrect flow rates

Mass spectrometer • Physical

• Acquisition Parameters

• Ion suppression

• Sample matrix and mobile phase additives (acids, salts) alter signal of target analyte.

Mass spectrometer • Physical

• Acquisition Parameters

• Ion suppression

• Data Processing

ARE YOU LOOKING AT THE RIGHT THING??

RT: 0.00 - 2.98

0.0 0.5 1.0 1.5 2.0 2.5

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

0.26

0.36

1.22 1.971.62 1.82 2.23 2.33 2.58 2.88

NL:1.31E9

TIC MS 20150416std2

RT: 0.00 - 2.98

0.0 0.5 1.0 1.5 2.0 2.5

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

0.26

0.36

1.22 1.971.62 1.82 2.23 2.33 2.58 2.88

NL:1.31E9

TIC MS 20150416std2

20150416std2 #31-41 RT: 1.52-2.02 AV: 11 SM: 7B NL: 1.74E5T: + p ESI Full pr 85.000 [200.000-550.000]

250 300 350 400 450 500 550

m/z

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lativ

e A

bu

nd

an

ce

260.1

456.4

291.1

218.1323.3

237.2

20150416std3 #6-17 RT: 0.26-0.68 AV: 6 SM: 7B NL: 3.23E6T: + p ESI Full pr 85.000 [200.000-550.000]

250 300 350 400 450 500 550

m/z

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lativ

e A

bu

nd

an

ce

260.1

221.1

456.3

277.2 482.5

353.3

374.3288.2 318.3

428.4

Examples

Example 1

Courtesy of Liverpool

Misaligned microtitre plate

Example 2

• Plate positioned correctly

• Sample in well

• LC-solvents correct

• No Leaks/blocks

• Correct acquisition

parameters

RT: 0.00 - 1.70

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Ab

un

da

nce

0.911.410.21 0.88 1.400.23

1.431.070.80 1.560.18 0.980.02 0.36 0.750.600.56 1.331.291.651.150.33

NL:4.35E1

TIC F: MS NSSsampchkA141028

Example 2

Power supply 3 assembly failure Call an engineer

RT: 0.00 - 1.70

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Ab

un

da

nce

0.911.410.21 0.88 1.400.23

1.431.070.80 1.560.18 0.980.02 0.36 0.750.600.56 1.331.291.651.150.33

NL:4.35E1

TIC F: MS NSSsampchkA141028

Example 3 RT: 0.00 - 1.69 SM: 5B

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

0.40

0.40

1.14

NL:3.45E7

TIC F: MS SPA140108RC1

NL:1.14E7

TIC F: MS SPAR601759

RT: 0.00 - 1.69 SM: 5B

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

0.40

0.40

1.14

NL:3.45E7

TIC F: MS SPA140108RC1

NL:1.14E7

TIC F: MS SPAR601759

RT: 0.00 - 1.69 SM: 5B

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Ab

un

da

nce

0.40

1.14

NL:3.45E7

TIC F: MS SPA140108RC1

NL:1.14E7

TIC F: MS SPAR601759

Example 3 RT: 0.00 - 1.69 SM: 5B

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Re

lative

Ab

un

da

nce

NL:1.14E7

TIC F: MS SPAR601759

NL:1.14E7

TIC F: MS ICIS SPAR601759

Area 159777277

Area 188572354

Incorrect seating of ferrule causing dead volume

Example 4 RT: 0.00 - 1.69 SM: 7B

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Ab

un

da

nce

0.64

0.11

NL:1.09E8

TIC F: MS SPAR596134A

Sample overload

Example 5

Courtesy of Birmingham

Improper nebulization due to ESI capillary misaligned following maintenance.

Did not extend beyond the probe tip

Example 6 RT: 0.00 - 1.70

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

20

40

60

80

1000

20

40

60

80

1000

20

40

60

80

100

Re

lative

Ab

un

da

nce

0

20

40

60

80

1000

20

40

60

80

100 NL:1.25E7

TIC F: MS NSSsampchkD140513

NL:1.38E7

TIC F: MS NSSsampchkA140513

NL:2.61E6

TIC F: MS NSSsampchkB140513

NL:1.56E7

TIC F: MS NSSsampchkC140513

NL:1.77E7

TIC F: MS NSSsampchkE140513

Example 6 RT: 0.00 - 1.70

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

20

40

60

80

1000

20

40

60

80

1000

20

40

60

80

100

Re

lative

Ab

un

da

nce

0

20

40

60

80

1000

20

40

60

80

100 NL:1.25E7

TIC F: MS NSSsampchkD140513

NL:1.38E7

TIC F: MS NSSsampchkA140513

NL:2.61E6

TIC F: MS NSSsampchkB140513

NL:1.56E7

TIC F: MS NSSsampchkC140513

NL:1.77E7

TIC F: MS NSSsampchkE140513

Example 6 RT: 0.00 - 1.70

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

20

40

60

80

1000

20

40

60

80

1000

20

40

60

80

100

Re

lative

Ab

un

da

nce

0

20

40

60

80

1000

20

40

60

80

100 NL:1.25E7

TIC F: MS NSSsampchkD140513

NL:1.38E7

TIC F: MS NSSsampchkA140513

NL:2.61E6

TIC F: MS NSSsampchkB140513

NL:1.56E7

TIC F: MS NSSsampchkC140513

NL:1.77E7

TIC F: MS NSSsampchkE140513

Pump running at low flow rate 50ul/min. Fix = Back pressure/retention loop (2m length peek) added

Example 7 RT: 0.00 - 8.00 SM: 9B

0 1 2 3 4 5 6 7

Time (min)

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Re

lative

Ab

un

da

nce

MMA

SA3.50

4.36

MMA

SA4.28

5.58

NL:1.09E5

TIC F: - c ESI SRM ms2 117.100 [72.850-73.350] MS MMA040714test22

NL:8.66E4

TIC F: - c ESI SRM ms2 117.100 [72.850-73.350] MS MMA040714test01

Example 7 RT: 0.00 - 8.00 SM: 9B

0 1 2 3 4 5 6 7

Time (min)

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Re

lative

Ab

un

da

nce

MMA

SA3.50

4.36

MMA

SA4.28

5.58

NL:1.09E5

TIC F: - c ESI SRM ms2 117.100 [72.850-73.350] MS MMA040714test22

NL:8.66E4

TIC F: - c ESI SRM ms2 117.100 [72.850-73.350] MS MMA040714test01

Mobile phase composition

Example 8

Courtesy of Liverpool

Poor response for Vit D3 but D2 and IS# are OK

Fresh mobile phase....Formic acid in methanol should be SHAKEN not STIRRED!

Example 9 C:\Xcalibur\...\Sept\08\ExtScreenUnk190 09/09/2014 06:35:52 Unk190

RT: 0.01 - 1.75 SM: 7G

0.5 1.0 1.5

Time (min)

0

50

100

0

50

100

0

50

100

0

50

100

0

50

100

Re

lative

Ab

un

da

nce

0

50

100

0

50

100

0

50

1000.33

0.33

0.34

0.41

0.35

0.35

0.29

0.29

NL: 1.03E7

TIC F: + p sid=-10.00

SRM ms2 132.00@-15.00

[ 85.50-86.50] MS

ExtScreenUnk190

NL: 1.48E7

TIC F: + p sid=-10.00

SRM ms2 135.00@-15.00

[ 88.50-89.50] MS

ExtScreenUnk190

NL: 8.90E4

TIC F: + p sid=-10.00

SRM ms2 150.00@-15.00

[ 103.50-104.50] MS

ExtScreenUnk190

NL: 2.42E6

TIC F: + p sid=-10.00

SRM ms2 153.00@-15.00

[ 106.50-107.50] MS

ExtScreenUnk190

NL: 3.68E6

TIC F: + p sid=-10.00

SRM ms2 166.00@-15.00

[ 119.50-120.50] MS

ExtScreenUnk190

NL: 1.55E7

TIC F: + p sid=-10.00

SRM ms2 171.00@-15.00

[ 124.50-125.50] MS

ExtScreenUnk190

NL: 1.31E6

TIC F: + p sid=-10.00

SRM ms2 182.00@-15.00

[ 135.50-136.50] MS

ExtScreenUnk190

NL: 2.80E6

TIC F: + p sid=-10.00

SRM ms2 186.00@-15.00

[ 139.50-140.50] MS

ExtScreenUnk190

RT: 0.00 - 1.75 SM: 7G

0.0 0.5 1.0 1.5

Time (min)

0

20

40

60

80

100

0

20

40

60

80

100

0

20

40

60

80

100

0

20

40

60

80

100

Re

lative

Ab

un

da

nce

0

20

40

60

80

100

0

20

40

60

80

100

0

20

40

60

80

100 1.46

0.30

1.53

1.401.14

1.211.54

0.25

1.601.35

1.02

0.32

1.611.22

0.32

1.16

NL: 4.13E5

TIC F: + p sid=-10.00

SRM ms2

246.00@-35.00 [

84.50-85.50] MS

ExtScreenUnk190

NL: 3.33E5

TIC F: + p sid=-10.00

SRM ms2

255.00@-35.00 [

84.50-85.50] MS

ExtScreenUnk190

NL: 2.27E4

TIC F: + p sid=-10.00

SRM ms2

276.00@-35.00 [

84.50-85.50] MS

ExtScreenUnk190

NL: 1.22E5

TIC F: + p sid=-10.00

SRM ms2

279.00@-35.00 [

84.50-85.50] MS

ExtScreenUnk190

NL: 2.56E4

TIC F: + p sid=-10.00

SRM ms2

288.20@-35.00 [

84.50-85.50] MS

ExtScreenUnk190

NL: 6.10E5

TIC F: + p sid=-10.00

SRM ms2

291.30@-35.00 [

84.50-85.50] MS

ExtScreenUnk190

NL: 4.78E4

TIC F: + p sid=-10.00

SRM ms2

316.20@-35.00 [

84.50-85.50] MS

ExtScreenUnk190

Amino acids Acylcarnitines

Example 9

RT: 0.00 - 1.75 SM: 7G

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

50

100

0

50

100

0

50

100

0

50

100

Re

lative

Ab

un

da

nce 0

50

100

0

50

100

0

50

100 1.46

0.30

1.53

1.401.14

1.21 1.54

0.25

1.601.35

1.02

0.32

1.611.22

0.32

1.16

NL: 4.13E5

TIC F: + p sid=-10.00 SRM ms2 246.00@-35.00 [ 84.50-85.50] MS ExtScreenUnk190

NL: 3.33E5

TIC F: + p sid=-10.00 SRM ms2 255.00@-35.00 [ 84.50-85.50] MS ExtScreenUnk190

NL: 2.27E4

TIC F: + p sid=-10.00 SRM ms2 276.00@-35.00 [ 84.50-85.50] MS ExtScreenUnk190

NL: 1.22E5

TIC F: + p sid=-10.00 SRM ms2 279.00@-35.00 [ 84.50-85.50] MS ExtScreenUnk190

NL: 2.56E4

TIC F: + p sid=-10.00 SRM ms2 288.20@-35.00 [ 84.50-85.50] MS ExtScreenUnk190

NL: 6.10E5

TIC F: + p sid=-10.00 SRM ms2 291.30@-35.00 [ 84.50-85.50] MS ExtScreenUnk190

NL: 4.78E4

TIC F: + p sid=-10.00 SRM ms2 316.20@-35.00 [ 84.50-85.50] MS ExtScreenUnk190

RT: 0.00 - 1.72 SM: 7G

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Time (min)

0

50

100

0

50

100

0

50

100

0

50

100

Re

lative

Ab

un

da

nce 0

50

100

0

50

100

0

50

1000.51

0.35

0.77

0.35

1.00

0.51

0.32

1.321.10

0.32 0.45

0.97

0.32

0.52

0.840.71

0.36

0.39

0.68

NL: 7.55E4

TIC F: + p sid=-10.00 SRM ms2 246.00@-35.00 [ 84.50-85.50] MS NSA15372828

NL: 4.43E5

TIC F: + p sid=-10.00 SRM ms2 255.00@-35.00 [ 84.50-85.50] MS NSA15372828

NL: 5.37E3

TIC F: + p sid=-10.00 SRM ms2 276.00@-35.00 [ 84.50-85.50] MS NSA15372828

NL: 7.39E4

TIC F: + p sid=-10.00 SRM ms2 279.00@-35.00 [ 84.50-85.50] MS NSA15372828

NL: 1.33E4

TIC F: + p sid=-10.00 SRM ms2 288.20@-35.00 [ 84.50-85.50] MS NSA15372828

NL: 8.06E5

TIC F: + p sid=-10.00 SRM ms2 291.30@-35.00 [ 84.50-85.50] MS NSA15372828

NL: 3.16E4

TIC F: + p sid=-10.00 SRM ms2 316.20@-35.00 [ 84.50-85.50] MS NSA15372828

Formic acid in sample NOT mobile phase

Example 10

Courtesy of Birmingham

Contamination of reconstituting reagent (ACn) with Internal standard

m/z 403

Partial derivatisation • Not drying samples properly • Insufficient reagent/heating • Derivatising reagent ‘gone off’

Example 11 20140721002 #6-17 RT: 0.26-0.68 AV: 6 SM: 7B NL: 2.89E6T: + p ESI Full pr 85.000 [200.000-550.000]

250 300 350 400 450 500 550

m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative A

bu

nd

an

ce

263.2

221.1

465.5

353.3277.2

288.3316.3

SM: 7B

275 280 285 290 295

m/z

0

50

100

0

50

100

0

50

100

Re

lative A

bu

nd

an

ce

0

50

100

0

50

100277.1

282.3

288.3284.4280.1

282.3

277.2

284.3280.2 288.1

277.2

284.4

277.2

275.0 284.4

277.3

284.2279.3

NL: 5.46E5

20140722023#6-17 RT: 0.26-0.68 AV: 6 T: + p ESI Full pr 85.000 [200.000-550.000]

NL: 7.20E5

20140724002#6-17 RT: 0.26-0.68 AV: 6 T: + p ESI Full pr 85.000 [200.000-550.000]

NL: 7.25E5

20140722027#6-17 RT: 0.26-0.68 AV: 6 T: + p ESI Full pr 85.000 [200.000-550.000]

NL: 5.75E5

20140722029#6-17 RT: 0.26-0.68 AV: 6 T: + p ESI Full pr 85.000 [200.000-550.000]

NL: 8.44E5

20140722025#6-17 RT: 0.26-0.68 AV: 6 T: + p ESI Full pr 85.000 [200.000-550.000]

Contamination from

Conclusion

MSMS are robust analytical instruments

Sensitive to small changes

Contamination will be detected – Good Lab Practice performed

Common problems with leaks and blocks in peek tubing

Take care with connection during maintenance

Conclusion

Contamination – sample preparation, dirty source

Peak shifts and shape – blocks or leaks, mobile phase

Low intensity- sample injection, clogged transfer capillary

No peaks – acquisition parameters, sample prep/position

Troubleshooting Strategy

• Start from easiest to fix first

• Eliminate the obvious

• Examine your

chromatographic profile

RT: 0.00 - 3.00

0.0 0.5 1.0 1.5 2.0 2.5 3.0

Time (min)

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Rela

tive A

bundance

0.26

0.36

0.31

0.480.22

0.56

0.64

0.73

0.81

0.90

NL:6.98E8

TIC MS 20150408std1

NL:4.20E8

TIC MS 20150408std3

• Familiarise yourself

with YOUR TICs – tells

you alot!

• If it doesn’t look

right…..Do something

about it!

Troubleshooting Strategy