sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 1
2
Digital Imaging and Communications in Medicine (DICOM) 4
Supplement 216 6
Second Generation Radiotherapy –
Brachytherapy Radiation Objects 8
10
12
14
16
DICOM Standards Committee, Working Group 7, Radiation Therapy 18
1300 N. 17th Street, Suite 1752
Rosslyn, Virginia 22209 USA 20
VERSION: Draft – Revision 01 22 February 12, 2019
Developed pursuant to DICOM Work Item TDB 24
This is a draft document. Do not circulate, quote, or reproduce it except with the approval of NEMA. 26
28
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 2
Table of Contents
Table of Contents ..................................................................................................................... 2 30
Table of Figures ........................................................................................................................ 3
Table of Tables ......................................................................................................................... 3 32
Open Issues and Discussion Points ......................................................................................... 4
Closed Issues ........................................................................................................................... 4 34
Foreword ................................................................................................................................... 5
Scope and Field of Application ................................................................................................. 5 36
Part 2 Addendum ...................................................................................................................... 6
Part 3 Addendum ...................................................................................................................... 7 38
7.14 ...... EXTENSION OF THE DICOM MODEL OF THE REAL-WORLD FOR RADIOTHERAPY SECOND GENERATION INFORMATION OBJECTS ....................................................... 7 40
A.86 ...... RT SECOND GENERATION ............................................................................ 10 A.86.1 ...... RT Second Generation Objects ............................................................... 10 42
A.86.1.1 ....... RT Second Generation Common Information ............................... 10 A.86.1.1.1 ............ RT Second Generation Entity-Relationship Model .... 10 44
A.86.1.a1 ..... Brachytherapy HDR Radiation Information Object Definition ........ 12 A.86.1.a1.1 .......... Brachytherapy HDR Radiation IOD Description ........ 12 46 A.86.1.a1.2 .......... Brachytherapy HDR Radiation IOD Entity-Relationship Model ................................................................................. 12 48 A.86.1.a1.3 .......... Brachytherapy HDR Radiation IOD Module Table .... 12 A.86.1.a1.4 .......... Brachytherapy HDR Radiation IOD Constraints........ 12 50
A.86.1.a2 ..... Brachytherapy LDR Radiation Information Object Definition ......... 13 A.86.1.a2.1 .......... Brachytherapy LDR Radiation IOD Description ........ 13 52 A.86.1.a2.2 .......... Brachytherapy LDR Radiation IOD Entity-Relationship Model ................................................................................. 13 54 A.86.1.a2.3 .......... Brachytherapy LDR Radiation IOD Module Table ..... 13 A.86.1.a2.4 .......... Brachytherapy LDR Radiation IOD Constraints ........ 14 56
A.86.1.a3 ..... Brachytherapy Seeds Information Object Definition ...................... 15 A.86.1.a3.1 .......... Brachytherapy Seeds IOD Description...................... 15 58 A.86.1.a3.2 .......... Brachytherapy Seeds IOD Entity-Relationship Model15 A.86.1.a3.3 .......... Brachytherapy Seeds IOD Module Table .................. 15 60 A.86.1.a3.4 .......... Brachytherapy Seeds IOD Constraints ..................... 15
C.36 ...... RT SECOND GENERATION MODULESERROR! BOOKMARK NOT DEFINED. 62
C.36.1 ...... RT Second Generation Concepts .............. Error! Bookmark not defined. C.36.2 ...... RT Second Generation Macros ............................................................... 18 64
C.36.2.m ...... RT Second Generation Device Macros ......................................... 18 C.36.2.m.x1 ......... X1 Macro ................................................................... 18 66 C.36.2.m.x2 ......... X2 Macro ................................................................... 18
C.36.m1 ... Brachytherapy HDR Delivery Device Module ........................................... 19 68 C.36.m1.1 .... Brachytherapy HDR Delivery Device Attribute Description ............ 20
C.36.m2 ... Brachytherapy HDR Beam Module .......................................................... 20 70 C.36.m2.1 .... Brachytherapy HDR Beam Attribute Description ........................... 21
C.36.m2.1.1 ......... Source Roll Continuous Angle .................................. 21 72 C.36.m3 ... Brachytherapy LDR Delivery Device Module ........................................... 21 C.36.m4 ... Brachytherapy LDR Beam Module ........................................................... 21 74 C.36.m5 ... Brachytherapy Seeds Delivery Device Module ........................................ 21 C.36.m6 ... Brachytherapy Seeds Module .................................................................. 22 76
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 3
Part 4 Addendum .................................................................................................................... 23
Part 6 Addendum .................................................................................................................... 24 78
6 ........... REGISTRY OF DICOM DATA ELEMENTS ...................................................... 24
ANNEX A REGISTRY OF DICOM UNIQUE IDENTIFIERS (UID) (NORMATIVE) . 25 80
Part 16 Addendum .................................................................................................................. 26
CID SUP216001 <TITLE> ........................................................................................... 26 82
CID SUP216011 BRACHYTHERAPY HDR PROCEDURE TECHNIQUE .................. 26
CID SUP216012 BRACHYTHERAPY LDR PROCEDURE TECHNIQUE .................. 26 84
CID SUP216013 BRACHYTHERAPY SEEDS PROCEDURE TECHNIQUE ............. 27
ANNEX D DICOM CONTROLLED TERMINOLOGY DEFINITIONS (NORMATIVE)27 86
Table of Figures 88
Figure 7.14-1 DICOM MODEL OF THE REAL WORLD – RADIOTHERAPY ........................................9 90
Figure A.86.1.1.1-1 — RT Second Generation IOD information model ...............................................11
92
Table of Tables
94
Table C.36.2.m.x1-1 X1 Macro Atributes .............................................................................................18
Table C.36.2.m.x2-1 X2 Macro Atributes .............................................................................................18 96
Table C.36.m1-1 Brachytherapy HDR Delivery Device Module Attributes ...........................................19
Table C.36.m2-1 Brachytherapy HDR Beam Module Attributes ...........................................................20 98
Table C.36.m3-1 Brachytherapy HDR Delivery Device Module Attributes ...........................................21
Table C.36.m4-1 Brachytherapy LDR Beam Module Attributes ...........................................................21 100
Table C.36.m5-1 Brachytherapy Seeds Delivery Device Module Attributes .........................................21
Table C.36.m6-1 Brachytherapy Seeds Module Attributes ...................................................................22 102
104
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 4
Open Issues and Discussion Points 106
# Item
1
2
3
4
5
7
8
9
10
Closed Issues
# Item
108
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 5
110
Foreword
This Supplement specifies additional IODs necessary to support the new Second Generation 112 Radiotherapy IODs and operations.
Scope and Field of Application 114
Introduction
This Supplement introduces RT Radiation IODs and RT Radiation Set IODs. A Radiation Set IOD 116 defines a Radiotherapy Treatment Fraction as a collection of instances of RT Radiation IODs. RT Radiation IODs represent different treatment modalities. This Supplement introduces the 118 representation of the C-Arm techniques.
This Supplement is based on the real-world model and specifications defined in Supplement 147. 120 References, definitions etc. not present in this Supplement can be found in Supplement 147.
General Architectural Principles 122
• Different types of data are encoded in different IODs. This is in contrast to First Generation objects, where different types of data are encoded in a single IOD, such as RT Plan. 124
• The new IODs are designed to support all current treatment modalities and be extensible for future modalities and new equipment. 126
• Compatibility with First-Generation IODs: It will be possible for the content of First Generation IODs to be represented in Second Generation IODs. However, information beyond the content of 128 a First Generation SOP Instance will be needed to create a valid Second Generation SOP Instance. 130
• IODs specific to use cases: Explicit separate IODs have been developed for specific treatment modalities with the concept of RT Radiation IOD – for example, Tomotherapeutic treatments, C-132 Arm beams, Robotic beams are modeled separately. This allows more stringent conditions to be applied to the presence or absence of Attributes within those IODs, and thereby increases the 134 potential for interoperability.
• Treatment techniques already in use but not yet covered in First Generation (such as robotic 136 therapy and tomotherapy) have been taken into account.
138
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 6
Part 2 Addendum 140
Add new SOP Classes to PS3.2 Table A.1-2 UID Values:
142
UID Value UID Name Category
1.2.840.10008.5.1.4.1.1.481.S216.1 Brachytherapy HDR Radiation
Storage
Transfer
1.2.840.10008.5.1.4.1.1.481.S216.2 Brachytherapy LDR Radiation
Storage
Transfer
1.2.840.10008.5.1.4.1.1.481.S216.3 Brachytherapy Seeds Storage Transfer
144
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 7
Part 3 Addendum
7.14 EXTENSION OF THE DICOM MODEL OF THE REAL-WORLD FOR RADIOTHERAPY 146
SECOND GENERATION INFORMATION OBJECTS
For the purpose of RT Second Generation SOP Classes the DICOM Model of the Real-World is 148 described in this section. This subset of the real-world model covers the requirements for transferring information about planned and performed radiotherapeutic treatments and associated data. 150
Figure 7.14-1 describes the most important elements involved in the radiotherapy domain in DICOM.
152
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 8
Patient
RT Treatment Phase
1
0-n
1
0-1
0-n
1
0-1
0-n
has
0-n
RT Radiation Set
0-n
1-n
0-n
0-n
RT Segment Annotation
1
RT Radiation
0-n
0-n
0-1
0-n
0-n
RT Dose Image
1
results in
0-n
0-n
0-n
0-n
0-n
1
0-n
is subject of
RT Physician Intent
RT Course
belongs to
RT Radiation Record
subdivided by
prescribes
0-1 0-1 0-n
0-n
contains
0-1
prescribes to to
1
Conceptual Volume
0-1
0-n
RT Dose Histogram
RT Dose Samples
calculated for
constitutes fraction composed of
has
0-n
1
1
0.n annotates
defines
applies to
volumetrically represented
by
references references
RT Prescription
contains
1
-n
0-n
Volumes
Surfaces
Lines
OR
OR
No Baseline
OR
No Baseline CID
Points
154
Note 1: IODs which contain a representation of Volumes, Surfaces, Lines, Points can be annotated by an RT Segment Annotation. 156
Note 2: For better readability the diagram only contains the most important relationships, e.g. all objects have a relation to the Patient, but not all of these relationships are part of this diagram. 158
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 9
Figure 7.14-1 DICOM MODEL OF THE REAL WORLD – RADIOTHERAPY
160
Add the following columns in PS3.3 Section A.1.4, Table A.1-1 COMPOSITE INFORMATION
OBJECT MODULES OVERVIEW – RADIOTHERAPY 162
IODs
Modules
Brach
ythera
py
HDR
Radiat
ion
Storag
e
Brach
ythera
py
HDR
Radiat
ion
Storag
e
Brach
ythera
py
HDR
Radiat
ion
Storag
e
Patient M M M
Clinical Trial Subject U U U
General Study M M M
Patient Study U U U
Clinical Trial Study U U U
General Series M M M
Clinical Trial Series U U U
Enhanced RT Series
M M M
General Equipment M M M
Enhanced General Equipment
M M M
Frame Of Reference
M M M
General Reference Module
M M M
…
Radiotherapy
Common Instance
M M M
RT Delivery Device
Common
M M M
RT Radiation
Common
M M M
Brachytherapy
HDR Delivery
Device Module
M
Brachytherapy
HDR Beam Module
M
Brachytherapy
LDR Delivery
Device Module
M
Brachytherapy
LDR Beam Module
M
Brachytherapy
Seeds Delivery
Device Module
M
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 10
IODs
Modules
Brach
ythera
py
HDR
Radiat
ion
Storag
e
Brach
ythera
py
HDR
Radiat
ion
Storag
e
Brach
ythera
py
HDR
Radiat
ion
Storag
e
Brachytherapy
Seeds Module
M
…
Common Instance Reference Module
M M
SOP Common M M
164
Add the following to PS3.3 Chapter 10 Miscellaneous Macros:
166
Add the following to PS3.3 Annex A:
A.86 RT SECOND GENERATION 168
A.86.1 RT Second Generation Objects
This section provides a brief description of the IODs of RT Second Generation. Specifically, this 170 description includes:
• The Real-World Object which is represented by the IOD 172
• Information as to the scope of the represented object if appropriate
A.86.1.1 RT Second Generation Common Information 174
This section provides a description of the Module structure which is shared by the RT Second Generation IODs. 176
A.86.1.1.1 RT Second Generation Entity-Relationship Model
The E-R Model in Figure A.86.1.1.1-1 depicts those components of the DICOM Information Model 178 that are relevant to second-generation RT IODs.
180
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 11
182
Figure A.86.1.1.1-1 — RT Second Generation IOD information model
Add the following Section to Annex A: 184
Patient
is the subject of
Study
contains
Series
1
1,n
1
1,n
contains
creates
1,n
Equipment
1 1
spatially or temporally
defines
Frame of Reference
RT Segment
Annotation RT Radiation
RT Physician
Intent
0,n 0,n 0,n 0,n
0-1
1,n
RT Radiation
Set
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 12
A.86.1.a1 Brachytherapy HDR Radiation Information Object Definition 186
A.86.1.a1.1 Brachytherapy HDR Radiation IOD Description
The Ion Modulated Scanning Radiation IOD describes a radiotherapy treatment on a C-Arm delivery 188 device using ion particles with a modulated scanning technique.
A.86.1.a1.2 Brachytherapy HDR Radiation IOD Entity-Relationship Model 190
See Figure A.86.1.1.1-1.
A.86.1.a1.3 Brachytherapy HDR Radiation IOD Module Table 192
Table A.86.1.a1-1
Brachytherapy HDR Radiation IOD Modules 194
IE Module Reference Usage
Patient Patient C.7.1.1 M
Clinical Trial Subject C.7.1.3 U
Study General Study C.7.2.1 M
Patient Study C.7.2.2 U
Clinical Trial Study C.7.2.3 U
Series General Series C.7.3.1 M
Clinical Trial Series C.7.3.2 U
Enhanced RT Series C.36.A1 M
Equipment General Equipment C.7.5.1 M
Enhanced General Equipment C.7.5.2 M
Frame of Reference
Frame of Reference C.7.4.1 M
RT Radiation General Reference C.12.4 M
RT Delivery Device Common C.36.E1 M
RT Radiation Common C.36.E2 M
Brachytherapy HDR Delivery Device C.36.m1 M
Brachytherapy HDR Beam C.36.m2 M
SOP Common C.12.1 M
Common Instance Reference C.12.2 M
Radiotherapy Common Instance C.36.A2 M
Note: The Frame of Reference identifies the Patient Coordinate System used to define the geometric setup of the radiation beam with respect to the patient. The relationship of the patient-based 196 coordinates to the Equipment Frame of Reference is specified by a transformation (see 10.A10).
198
A.86.1.a1.4 Brachytherapy HDR Radiation IOD Constraints
A.86.1.a1.4.1 Modality Attribute 200
The value of Modality (0008,0060) shall be RTRAD.
A.86.1.a1.4.2 RT Delivery Device Common Module 202
The Equipment Frame of Reference UID (gggg,51A0) shall be 1.2.840.10008.1.4.RRR.1.
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 13
Code Sequence CID
Treatment Machine Special Mode Sequence (gggg,9C97)
Defined CID SUP175003 “Radiotherapy Treatment Machine Modes”
Radiation Dosimeter Unit Sequence (gggg,5113) Defined CID SUP175012 “C-Arm Photon-Electron Dosimeter Unit”
204
A.86.1.a1.4.3 RT Radiation Common Module
The value of RT Radiation Content Type (gggg,5013) shall not be NONE. 206
The value of RT Record Flag (gggg,5014) shall be NO.
The following code sequences shall have values from the identified CIDs: 208
Code Sequence CID
RT Treatment Technique Code Sequence (3010,0080)
Defined CID SUP216011 “Brachytherapy HDR Procedure Techniques”
A.86.1.a1.4.4 Radiotherapy Common Instance Module 210
Code Sequence CID
Author Identification Sequence (3010,0019) Defined CID for Organizational Role is CID SUP175015 “Radiotherapy Treatment Planning Person Roles”
A.86.1.a2 Brachytherapy LDR Radiation Information Object Definition 212
A.86.1.a2.1 Brachytherapy LDR Radiation IOD Description
The Ion Modulated Scanning Radiation IOD describes a radiotherapy treatment on a C-Arm delivery 214 device using ion particles with a modulated scanning technique.
A.86.1.a2.2 Brachytherapy LDR Radiation IOD Entity-Relationship Model 216
See Figure A.86.1.1.1-1.
A.86.1.a2.3 Brachytherapy LDR Radiation IOD Module Table 218
Table A.86.1.a2-1
Brachytherapy LDR Radiation IOD Modules 220
IE Module Reference Usage
Patient Patient C.7.1.1 M
Clinical Trial Subject C.7.1.3 U
Study General Study C.7.2.1 M
Patient Study C.7.2.2 U
Clinical Trial Study C.7.2.3 U
Series General Series C.7.3.1 M
Clinical Trial Series C.7.3.2 U
Enhanced RT Series C.36.A1 M
Equipment General Equipment C.7.5.1 M
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 14
Enhanced General Equipment C.7.5.2 M
Frame of Reference
Frame of Reference C.7.4.1 M
RT Radiation General Reference C.12.4 M
RT Delivery Device Common C.36.E1 M
RT Radiation Common C.36.E2 M
Brachytherapy LDR Delivery Device C.36.m1 M
Brachytherapy LDR Beam C.36.m2 M
SOP Common C.12.1 M
Common Instance Reference C.12.2 M
Radiotherapy Common Instance C.36.A2 M
Note: The Frame of Reference identifies the Patient Coordinate System used to define the geometric setup of the radiation beam with respect to the patient. The relationship of the patient-based 222 coordinates to the Equipment Frame of Reference is specified by a transformation (see 10.A10).
224
A.86.1.a2.4 Brachytherapy LDR Radiation IOD Constraints
A.86.1.a2.4.1 Modality Attribute 226
The value of Modality (0008,0060) shall be RTRAD.
A.86.1.a2.4.2 RT Delivery Device Common Module 228
The Equipment Frame of Reference UID (gggg,51A0) shall be 1.2.840.10008.1.4.RRR.1.
Code Sequence CID
Treatment Machine Special Mode Sequence (gggg,9C97)
Defined CID SUP175003 “Radiotherapy Treatment Machine Modes”
Radiation Dosimeter Unit Sequence (gggg,5113) Defined CID SUP175012 “C-Arm Photon-Electron Dosimeter Unit”
230
A.86.1.a2.4.3 RT Radiation Common Module
The value of RT Radiation Content Type (gggg,5013) shall not be NONE. 232
The value of RT Record Flag (gggg,5014) shall be NO.
The following code sequences shall have values from the identified CIDs: 234
Code Sequence CID
RT Treatment Technique Code Sequence (3010,0080)
Defined CID SUP216011 “Brachytherapy LDR Procedure Techniques”
A.86.1.a2.4.4 Radiotherapy Common Instance Module 236
Code Sequence CID
Author Identification Sequence (3010,0019) Defined CID for Organizational Role is CID SUP175015 “Radiotherapy Treatment Planning Person Roles”
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 15
A.86.1.a3 Brachytherapy Seeds Information Object Definition 238
A.86.1.a3.1 Brachytherapy Seeds IOD Description
The Ion Modulated Scanning Radiation IOD describes a radiotherapy treatment on a C-Arm delivery 240 device using ion particles with a modulated scanning technique.
A.86.1.a3.2 Brachytherapy Seeds IOD Entity-Relationship Model 242
See Figure A.86.1.1.1-1.
A.86.1.a3.3 Brachytherapy Seeds IOD Module Table 244
Table A.86.1.a3-1
Brachytherapy Seeds IOD Modules 246
IE Module Reference Usage
Patient Patient C.7.1.1 M
Clinical Trial Subject C.7.1.3 U
Study General Study C.7.2.1 M
Patient Study C.7.2.2 U
Clinical Trial Study C.7.2.3 U
Series General Series C.7.3.1 M
Clinical Trial Series C.7.3.2 U
Enhanced RT Series C.36.A1 M
Equipment General Equipment C.7.5.1 M
Enhanced General Equipment C.7.5.2 M
Frame of Reference
Frame of Reference C.7.4.1 M
RT Radiation General Reference C.12.4 M
RT Delivery Device Common C.36.E1 M
RT Radiation Common C.36.E2 M
Brachytherapy HDR Delivery Device Module
C.36.m1 M
Brachytherapy HDR Beam C.36.m2 M
SOP Common C.12.1 M
Common Instance Reference C.12.2 M
Radiotherapy Common Instance C.36.A2 M
Note: The Frame of Reference identifies the Patient Coordinate System used to define the geometric setup of the radiation beam with respect to the patient. The relationship of the patient-based 248 coordinates to the Equipment Frame of Reference is specified by a transformation (see 10.A10).
250
A.86.1.a3.4 Brachytherapy Seeds IOD Constraints
A.86.1.a3.4.1 Modality Attribute 252
The value of Modality (0008,0060) shall be RTRAD.
A.86.1.a3.4.2 RT Delivery Device Common Module 254
The Equipment Frame of Reference UID (gggg,51A0) shall be 1.2.840.10008.1.4.RRR.1.
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 16
Code Sequence CID
Treatment Machine Special Mode Sequence (gggg,9C97)
Defined CID SUP175003 “Radiotherapy Treatment Machine Modes”
Radiation Dosimeter Unit Sequence (gggg,5113) Defined CID SUP175012 “C-Arm Photon-Electron Dosimeter Unit”
256
A.86.1.a3.4.3 RT Radiation Common Module
The value of RT Radiation Content Type (gggg,5013) shall not be NONE. 258
The value of RT Record Flag (gggg,5014) shall be NO.
The following code sequences shall have values from the identified CIDs: 260
Code Sequence CID
RT Treatment Technique Code Sequence (3010,0080)
Defined CID SUP216011 “Brachytherapy HDR Procedure Techniques”
A.86.1.a3.4.4 Radiotherapy Common Instance Module 262
Code Sequence CID
Author Identification Sequence (3010,0019) Defined CID for Organizational Role is CID SUP175015 “Radiotherapy Treatment Planning Person Roles”
264
266
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 17
268
Add the following to PS3.3 Annex C:
C.36 RT SECOND GENERATION MODULES 270
The following Attribute Macros and Modules are used by the RT Second Generation IODs.
C.36.1 RT Second Generation Concepts 272
This section dicusses general concepts used in RT Second Generation Modules.
… 274
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 18
C.36.2 RT Second Generation Macros 276
C.36.2.m RT Second Generation Device Macros
C.36.2.m.x1 X1 Macro 278
Table C.36.2.m.x1-1 280
X1 Macro Atributes
Attribute Name Tag Type Description
Number of Sources (gggg,TODO) 1 Number of Sources in the Sources Definition Sequence (gggg,TODO).
Sources Definition Sequence (gggg,TODO) 1C Sources…
The number of Items included in this Sequence shall equal the value of Number of Sources (gggg,TODO).
Required if the Number of Sources (gggg,TODO) has a non-zero value.
>Include Table C.36.2.m.3-1 “RT Accessory Device Identification Macro Attributes”
Defined CID SUPnn4002 “TODO”.
>
282
C.36.2.m.x2 X2 Macro
284
Table C.36.2.m.x2-1
X2 Macro Atributes 286
Attribute Name Tag Type Description
Number of X2 (gggg,TODO) 1 Number of X2 in the X2 Definition Sequence (gggg,TODO).
X2 Definition Sequence (gggg,TODO) 1C X2...
The number of Items included in this Sequence shall equal the value of Number of X2 (gggg,TODO).
Required if the Number of X2 (gggg,TODO) has a non-zero value.
>
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 19
Attribute Name Tag Type Description
288
C.36.m1 Brachytherapy HDR Delivery Device Module
The following is an excerpt from C-Arm for illustration. The content has to be replaced by applicable 290 Brachytherapy content.
The C-Arm Photon-Electron Module defines constant C-Arm-specific parameters pertaining to the 292 physical device used to deliver external ion beams using modulated scanning techniques.
Table C.36.m1-1 294
Brachytherapy HDR Delivery Device Module Attributes
Attribute Name Tag Type Description
Virtual Source-Axis Distances (300A,030A) 1 Distance (in mm) from virtual source position to gantry rotation axis or nominal isocenter position (fixed beam-lines) of the equipment to be used for beam delivery.
Specified by a numeric pair - the VSAD in the IEC Gantry X direction followed by the VSAD in the IEC Gantry Y direction.
The VSAD is commonly used for designing apertures in contrast to the effective source-axis-distance (ESAD) that is commonly used with the inverse square law for calculating the dose decrease with distance.
See Section C.8.8.25.4.
Scanning Type Code (TODO)
Include Table C.36.2.m.7-1 “Radiation Generation Mode Macro Attributes”
Defined CID for Radiation Type Code Sequence (gggg,51C4) is CID 9526 “Ion Therapy Particle”.
Defined CID for Energy Unit Code Sequence (gggg,51C9) is CID SUPnn4001 “Energy Unit for Particle Therapy”.
No baseline CID defined for Radiation Fluence Modifier Code Sequence (gggg,51C8).
Include Table C.36.2.m.8-1 “RT Beam Limiting Device Definition Macro Attributes”
Defined CID for included ‘RT Accessory Device Identification Macro’ is CID SUP175001 “Beam Limiting Device Types”.
Include Table C.36.2.m.12-1 “Compensators Definition Macro Attributes”
Include Table C.36.2.m.13-1 “Blocks Definition Macro Attributes”
Include Table C.36.2.m.14-1 “RT Accessory Holders Definition Macro Attributes”
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 20
Attribute Name Tag Type Description
Include Table C.36.2.m.15-1 “General Accessories Definition Macro Attributes”
Include Table C.36.2.m.16-1 “Boluses Definition Macro Attributes”
Include Table C.36.2.m-x1-1 “X1 Macro Attributes”
296
C.36.m1.1 Brachytherapy HDR Delivery Device Attribute Description
298
C.36.m2 Brachytherapy HDR Beam Module
The following is an excerpt from C-Arm for illustration. The content has to be replaced by applicable 300 Brachytherapy content.
The C-Arm Photon-Electron Beam Module specifies how a C-Arm photon or electron treatment beam 302 is to be delivered.
Table C.36.m2-1 304
Brachytherapy HDR Beam Module Attributes
Attribute Name Tag Type Description
C-Arm Photon-Electron Control Point Sequence
(gggg,9C00) 1 Control Points used to model the beam delivery.
Two or more Items shall be included in this Sequence.
>Include Table C.36.2.m.x2-1 “X2 Macro Attributes” Defined CID SUP175010 “C-Arm Photon-Electron Delivery Dose Rate Unit”
>Referenced Radiation Generation Mode Index
(gggg,9124) 1C Radiation Generation Mode Index (gggg,9113) in the Radiation Generation Mode Sequence (gggg,51C0) in this IOD.
Required if the conditions in Section C.36.2.m.5.1.1 are satisfied.
>Include Table C.36.2.m.9-1 “RT Beam Limiting Device Opening Macro Attributes”
>Source Roll Continuous Angle (gggg,51B5) 1C Continuous gantry roll angle in degrees of the radiation source at the Control Point with respect to the Equipment Frame of Reference.
See C.36.G2.1.1, C.36.1.8 and C.36.E1.1.1.
Required if the conditions in Section C.36.2.m.5.1.1 are satisfied.
>RT Beam Limiting Device Continuous Angle
(gggg,51B4) 1C Angle in degrees of the Beam Modifier Coordinate System about the Z-axis relative to the parent coordinate system.
See C.36.1.8 and C.36.G2.1.2.
Required if the conditions in Section C.36.2.m.5.1.1 are satisfied.
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 21
Attribute Name Tag Type Description
>Source to Patient Surface Distance
(gggg,9C63) 2C Source to Patient Surface (skin) distance in mm.
Required if the conditions in Section C.36.2.m.5.1.1 are satisfied.
>Source to External Contour Distance
(gggg,9C62) 2C Source to External Contour distance in mm including devices associated with the patient anatomy model. For dosimetric purposes this value may differ from the Source to Surface Distance (300A,0130).
See C.36.C2.1.4.
Required if the conditions in Section C.36.2.m.5.1.1 are satisfied.
306
C.36.m2.1 Brachytherapy HDR Beam Attribute Description
C.36.m2.1.1 Source Roll Continuous Angle 308
For an Equipment Frame of Reference UID (gggg,51A0) 1.2.840.10008.1.4.RRR.1 the source roll angle is the rotation of the IEC 61217 GANTRY coordinate system about the Y-axis of the IEC 61217 310 FIXED coordinate system.
C.36.m2.1.2 RT Beam Limiting Device Continuous Angle 312
For an Equipment Frame of Reference UID (gggg,51A0) 1.2.840.10008.1.4.RRR.1 the RT Beam Limiting Device Continuous Angle (gggg,51B4) is the rotation of the IEC 61217 BEAM LIMITING 314 DEVICE system about the Z-axis of the IEC 61217 GANTRY system.
C.36.m3 Brachytherapy LDR Delivery Device Module 316
Table C.36.m3-1
Brachytherapy LDR Delivery Device Module Attributes 318
Attribute Name Tag Type Description
C.36.m4 Brachytherapy LDR Beam Module 320
Table C.36.m4-1
Brachytherapy LDR Beam Module Attributes 322
Attribute Name Tag Type Description
C.36.m5 Brachytherapy Seeds Delivery Device Module 324
Table C.36.m5-1
Brachytherapy Seeds Delivery Device Module Attributes 326
Attribute Name Tag Type Description
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 22
C.36.m6 Brachytherapy Seeds Module 328
Table C.36.m6-1
Brachytherapy Seeds Module Attributes 330
Attribute Name Tag Type Description
332
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 23
Part 4 Addendum 334
Add the following to PS3.4, Appendix B.5, Table B.5-1
336
SOP Class Name
SOP Class UID IOD Spec
(defined in PS 3.3)
Brachytherapy HDR Radiation
Storage
1.2.840.10008.5.1.4.1.1.481.S216.1 Brachytherapy HDR
Radiation IOD
Brachytherapy LDR Radiation
Storage
1.2.840.10008.5.1.4.1.1.481.S216.2 Brachytherapy LDR
Radiation IOD
Brachytherapy Seeds Storage 1.2.840.10008.5.1.4.1.1.481.S216.3 Brachytherapy Seeds IOD
338
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 24
Part 6 Addendum 340
Add the following data elements to PS3.6:
342
Editorial Note:
Use Range (gggg,A800) – (gggg,AFFF). 344
6 REGISTRY OF DICOM DATA ELEMENTS 346
(gggg,A800)
(gggg,A801)
348
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 25
Add the following to PS3.6 Annex A: 350
ANNEX A REGISTRY OF DICOM UNIQUE IDENTIFIERS (UID) (NORMATIVE) 352
Table A-1 UID Values
354
UID Value UID NAME UID TYPE Part
1.2.840.10008.5.1.4.1.1.481.S216.1 Brachytherapy HDR
Radiation Storage
SOP Class PS3.4
1.2.840.10008.5.1.4.1.1.481.S216.2 Brachytherapy LDR
Radiation Storage
SOP Class PS3.4
1.2.840.10008.5.1.4.1.1.481.S216.3 Brachytherapy Seeds
Storage
SOP Class PS3.4
356
Table A-3 Context Group UID Values
358
Context UID Context
Identifier
Context Group Name
1.2.840.10008.6.1.S216.001 SUP216001 <Title>
1.2.840.10008.6.1.S216.011 SUP216011 Brachytherapy HDR Procedure Technique
1.2.840.10008.6.1.S216.012 SUP216012 Brachytherapy LDR Procedure Technique
1.2.840.10008.6.1.S216.013 SUP216013 Brachytherapy Seeds Procedure
Technique
360
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 26
Part 16 Addendum 362
Add the following new CIDs to PS3.16, Annex B:
CID SUP216001 <TITLE> 364
Context ID SUP216001
<Title> 366
Resources: HTML | FHIR JSON | FHIR XML | IHE SVS XML
Type: Extensible 368
Version: yyyymmdd
UID: 1.2.840.10008.6.1.S216.001 370
Coding Scheme
Designator
(0008,0102)
Code Value
(0008,0100)
Code Meaning
(0008,0104)
372
CID SUP216011 BRACHYTHERAPY HDR PROCEDURE TECHNIQUE
Context ID SUP216001 374
Brachytherapy HDR Procedure Technique
Resources: HTML | FHIR JSON | FHIR XML | IHE SVS XML 376
Type: Extensible
Version: yyyymmdd 378
UID: 1.2.840.10008.6.1.S216.011
380
Coding Scheme
Designator
(0008,0102)
Code Value
(0008,0100)
Code Meaning
(0008,0104)
CID SUP216012 BRACHYTHERAPY LDR PROCEDURE TECHNIQUE 382
Context ID SUP216013
Brachytherapy LDR Procedure Technique 384
Resources: HTML | FHIR JSON | FHIR XML | IHE SVS XML
Type: Extensible 386
sup216_01_BrachytherapyRadiationObjects.doc: 2nd Generation RT - Brachytherapy Radiation Objects Page 27
Version: yyyymmdd
UID: 1.2.840.10008.6.1.S216.012 388
Coding Scheme
Designator
(0008,0102)
Code Value
(0008,0100)
Code Meaning
(0008,0104)
390
CID SUP216013 BRACHYTHERAPY SEEDS PROCEDURE TECHNIQUE
Context ID SUP216013 392
Brachytherapy Seeds Procedure Technique
Resources: HTML | FHIR JSON | FHIR XML | IHE SVS XML 394
Type: Extensible
Version: yyyymmdd 396
UID: 1.2.840.10008.6.1.S216.013
398
Coding Scheme
Designator
(0008,0102)
Code Value
(0008,0100)
Code Meaning
(0008,0104)
400
Add the following to the table in PS3.16, Annex D:
402
ANNEX D DICOM CONTROLLED TERMINOLOGY DEFINITIONS (NORMATIVE)
404
Code Value Code Meaning Definition Notes
S216001
S216002
S216003