Revolution Centrifugal Pump
Performance Overview Rev 2.1
Bearing Failures Leaks Decouple Noise & Fractured
Magnets Thrombus Generation &
Hemolysis Air Entrapment Vane Impeller = Blender
Flow Sensor Issues Ease of Set-up
Centrifugal Pump Challenges
Designed by the experts in centrifugal pump design.
Unique molded magnet driver:
• Fills internal space
• Assures balanced magnet
• Consistent performance
Pump Assembly Manufacturing Process
100% Quality Control Process Designed for manufacturability Simple assembly Minimal parts Ultrasonic weld
Simple Assembly = High Quality
Assembly
1. Patented Curved Inlet
Port
3. Patented “Spin-Inducer”
2. Patented Seal-less, Low Friction Upper
Bearing
5. Ultrasonic Weld
Upper & Lower
Housing
Internal Features
4. Impeller floats while
pumping
Flow enters the inlet rinsing the top bearing
Centrifugal force moves fluid to the outside
Fluid not leaving the pump moves
back to the low pressure center,
rinsing the bottom bearing Recirculation is
completed by fluid moving up thru the central channel
As fluid moves up thru the central channel, it lifts the impeller off the bottom bearing, opposing the magnetic forces, reducing
bearing load and bearing friction. The impeller literally floats.
Flow Path
Pump-Head Internal Surface Area
190.0 215.0
270.0 281.0
400.0
590.0
0.0
100.0
200.0
300.0
400.0
500.0
600.0
CM
2
Rotaflow Delphin Revolution Isoflow Capiox BP80
Revolution’s small surface
area reduces platelet
activation.
Testing documented in Sorin Group Engineering Report SPN0263
Model Prime (ml) Rate Flow (lpm)
Rated Press (mm Hg)
Max Speed (rpm)
Revolution 57 8 800 3500
Medtronic Biopump 89 8 900 4400
Sarns Delphin 52 9.9 700 3600
Lifestream Isoflow 65 9.99 800 3500
Jostra Rotaflow 32 10 750 5000
Terumo Capiox 45 8 800 3000
General Specifications data source: Manufacturers IFU
2 cc of air injected
Volume collected in chamber
Test Circuit
Computational Flow Dynamics
Large changes in velocity would indicate vortexing.
Note steady increases from center to end of vane tips.
Computational Flow Dynamics
Analysis featured in Dynamics magazine
Click here to read the article or go to:http://www.cd-adapco.com/press_room/dynamics/18/
Dr. Robin ShandasProfessor of Pediatric Medicine & Mechanical Engineering – University of Colorado
Validates CFD: No Vane Tip Vortices
Gentle acceleration from center to outer diameter
Flow Visualization
Click on picture to start/stop video
Unorganized flowupstream of inlet
Upstream of inlet
Flow Visualization
Click on picture to start/stop video
Organized flowapproaching inlet
Near inlet
Flow Visualization
Click on picture to start/stop video
Smooth rotationalflow at inlet due tospin-inducer
Inlet
Flow Visualization
Click on picture to start/stop video
Rotational flowcreated by spin-inducer
Spin-inducer
Flow Visualization
Click on picture to start/stop video
0
20
40
60
80
100
m
g%
Capiox SP Rotaflow Isoflow Revolution BPX-80 Sarns BP-80
Plasma Free Hemoglobin at 360 min
Low Hemolysis
P=.047
P=.073
P=.007
P=.472
P=.762P=.316
Testing documented in Sorin Group Engineering Report SPN0263
Plasma Free Hemoglobin
Low Hemolysis
0
10
20
30
40
50
60
70
mg
%
Capiox SP Rotaflow Isoflow Revolution BPX-80 Sarns BP-80
30 min
180 min
360 min
Testing documented in Sorin Group Engineering Report SPN0263
Index of Hemolysis
Low Hemolysis
0
10
20
30
40
50
60
70
mg
/10
0L
Capiox SP Rotaflow Isoflow Revolution BPX-80 Sarns BP-80
P=.047
P=.799
P=.464
P=.007
P=.076
P=.308
Testing documented in Sorin Group Engineering Report SPN0263
Temperature at 360 min
Low Heat Generation
26.0
27.0
28.0
29.0
30.0
31.0
32.0
°C
Rotaflow Revolution Sarns BP-80
Testing documented in Sorin Group Engineering Report SPN0263
Pressure-Flow (@ Maximum RPM)
600
650
700
750
800
850
900
950
1000
0 2 4 6 8
LPM
mm
Hg
Revolution 3500 BP80 Sarns
Terumo Jostra Isoflo
Testing documented in Sorin Group Engineering Report SPN0263
0
100
200
300
400
500
600
700
800
900
0 1000 2000 3000 4000 5000
RPM
mm
Hg
Revolution
BP-80
Isoflow
Rotaflow
Capiox
Sarns
Pressure generated at 5 LPM
High Efficiency: Highly efficient pump reduces hemolytic heat generation.Testing documented in Sorin Group Engineering Report SPN0263